Sulfur Analysis with Compliance Flexibility Sindie® 2622 Gen 3 complies with ASTM D2622, D7039 and ISO 20884 methods, enabling complete flexibility in sulfur analysis. With no compromises in detection, performance and reliability, Sindie 2622 is the ideal sulfur analytical solution from ultra low sulfur diesel and gasoline to heavy fuel oil and crudes.
Applications • Total sulfur analysis from ultra low sulfur fuels to crudes • F or use in refinery labs, pipeline terminals, additive plants and inspection laboratories Features and Benefits • LOD: 0.15 ppm at 300 s • Dynamic Range: 0.15 ppm - 10 wt% • E asy to use - Intuitive touch screen - Just plug-in and measure - Measurement time: 30-900 s • L ow and high range calibrations available: - Low Range: 0.15 ppm - 3000 ppm - High Range: 0.3 wt% - 10 wt% • E xtremely low maintenance: no conversion gasses, heating elements, columns, or quartz tubing • 75 W air-cooled excitation tube • F its on any lab bench • Use Accucells for hassle-free sample prep Options • 8-cell Autosampler • Helium optical path available • LIMS data output compatible software
ASTM D2622 & D7039 ISO 20884
TRUSTED PRECISION
Monochromatic Wavelength Dispersive X-ray Fluorescence (MWDXRF®) utilizes state-of-the-art focusing and monochromating optics to increase excitation intensity and dramatically improve signal-to-background over high power traditional WDXRF instruments. This enables significantly improved detection limits and precision, and a reduced sensitivity to matrix effects. A monochromatic and focused primary beam excites the sample and secondary characteristic fluorescence X-rays are emitted from the sample. A second monochromating optic selects the sulfur characteristic X-rays and directs these X-rays to the detector. MWDXRF is a direct measurement technique and does not require consumable gasses or sample conversion.
X-RAYS FROM SAMPLE
SAMPLE
DETECTOR
POLYCHROMATIC X-RAYS FROM SOURCE OPTIC 1
OPTIC 2
SELECTED MONOCHROMATIC SAMPLE X-RAYS
MONOCHROMATIC EXCITATION
Precision
Typical repeatability (r) and reproducibility (R) values in diesel fuel, at 95% confidence. 300 s measurement time.
ACCUCELL SAMPLE CUPS • No assembly of separate film & cup components • Pre-vented sample cups • Eliminates sample & cup contamination • One discharge of 1 ml pipette will fill the cup
Sulfur Concentration (ppm)
r
R
2
0.3
0.7
5
0.5
0.8
8
0.6
1.0
15
0.8
1.4
100
2
4
500
5
10
High Range Calibration
Low Range Calibration
Product Specifications Model
Sindie 2622 Gen 3
Test Method
ASTM D7039, D2622 and ISO 20884
Dimensions
37 cm (w) x 50 cm (d) x 34 cm (h)
Power 100-120 VAC, 47-63 HZ at 6.0 Amps/ 200-240 VAC, 47-63 HZ at 6.0 Amps Sample Cup Volume
1 ml
Ambient Temperature Requirements
5-40° C (40-104° F)
Dynamic Range
Standard: 0.15 ppm - 10 wt%
Measurement
User selectable: 30-900 s
Calibration 8 calibration curves. Automatic and manual calibration functionality Optical Path
Vacuum (helium available)
Sindie uses a weighted least squares regression in low range which is extremely linear and easy to set up. Typical correlation (R value) is expected to be on the order of 0.999 or better.