EVA-HF325/525 Troubleshooting Technical Publication Ver.3.2 Troubleshooting EVA-HF325/525 generator COMED Medical Sys
Views 84 Downloads 4 File size 6MB
EVA-HF325/525 Troubleshooting
Technical Publication Ver.3.2
Troubleshooting EVA-HF325/525 generator
COMED Medical Systems Co., Ltd.
0
EVA-HF325/525 Troubleshooting
ADVISORY SYMBOLS The following advisory symbols will be used throughout this manual. Their application and meaning are described below.
DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEED OR AVOIDED WILL CAUSE SERIOUS PERSONAL INJURY OR DEATH.
ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED
OR
AVOIDED
COULD
CAUSE
SERIOUS
PERSONAL INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR DATA.
ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED COULD CAUSE PERSONAL INJURY OR DAMAGE TO EQUIPMENT OR DATA.
COMED Medical Systems Co., Ltd.
1
EVA-HF325/525 Troubleshooting
CONTENTS
1. INTRODUCTION 2. ERROR
CODES
3. FLOW
CHART
4. ERROR 5.
9. MAIN 10. 11.
--------------------------------------
4 6
-------------------------------------
TABLE WAVEFORM
7. JOB 8.
------------------------------------
CHECK
DATA
6.
------------------------------------3
-------------------------------------
11
DATA---------------------------------
13
CODE---------------------------------------X-RAY CARD
CHECK-----------------------------------BOARD(MOTHER
TECHNICAL JOB
CODE
BOARD)DRAWING----------
INFORMATION-------------------------INFORMATION--------------------------
COMED Medical Systems Co., Ltd.
9
16 40 45 46 56
2
EVA-HF325/525 Troubleshooting
1. INTRODUCTION The
generator
contains
many
self-diagnostic
routines
which
greatly
facilitates
troubleshooting. The self-diagnostic functions require that all microprocessors function properly. Each microprocessor contains LEDs, LCD that indicates their proper operation.
THE MAIN STORAGE CAPACITORS RETAIN A LARGE PORTION OF THEIR FOR APPROXIMATELY 5MIN. AFTER THE UNIT IS TURNED OFF.
ALWAYS HAVE “IPM DRIVER PCB” CONNECTED WHEN MAIN POWER IS ACTIVATED IN THE GENERATOR. WHITOUT “IPM DRIVER PCB” CONNECTED, PERMANENT DAMAGE WILL OCCUR TO IPM’S.
DO NOT USE THE GENERATOR POWER SUPPLIES TO POWER EQUIPMENT EXTERNAL TO THE GENERATOR. ALL GENERATOR INPUTS AND OUTPUTS MUST BE TTL LOGIC LEVELS (TYPICALLY ACTIVE LOW) AND ISOLATED FROM OTHER SUBSYSTEMS OPTICALLY OR WITH DRY CONTACT RELAYS.
IF CHANGE EEPROM ON CONSOLE DISPLAY BOARD, WILL LOSE CALIBRATION DATA, APR DATA, AND SOME DATA. SO MUST CHECK EEPROM DATA. AND IF CHANGE UV-PROM ON HT CONTROL CPU BOARD AND CONSOLE DISPLAY BOARD, WILL CHECK THE VIRSION OF
COMED Medical Systems Co., Ltd.
PROGRAM.
3
EVA-HF325/525 Troubleshooting
2. ERROR CODES System error codes are displayed on the kVp display LCD(condition display LCD) Each code is preceded by the letter “ E”(i.e. E16). Error codes are described in below tables.
ERROR CODE
DESCRIPTION A system failure has been detected. The system must be turned
Display E xx
OFF and then ON to clear the error. Besides display “E xx”, another error code could be shown on the Console display LCD.
E16
E17
E18
E19
KV DATA is more than “FD”(see JOB CODE 020) Check KV DATA located APR data in EEPROM KV DATA is less than “20”(see JOB CODE 020) Check KV DATA located APR data in EEPROM mA DATA is more than “FD”(see JOB CODE 020) Check mA DATA located APR data in EEPROM mA DATA is less than “20”(see JOB CODE 020) Check mA DATA located APR data in EEPROM TIME DATA is abnormal(see JOB CODE 021)
E20
E21/E22
Check TIME DATA located APR data in EEPROM COMP DATA is abnormal(see JOB CODE 020) Check COMP DATA located Calibration data in UV-PROM ROTOR is abnormal(see JOB CODE 022)
E26/E27
Check ROTOR sensing circuit section or ROTOR DATA located in UV-PROM address CHARGE is abnormal(see JOB CODE 008 ~010)
E28
Charge capacitor(4700uF) is not charge Check charge sensing circuit section and control cpu board Filament sensing “LOW” (see JOB CODE 023)
E29
Low current input to filament Check filament sensing circuit section and connection of HV cable
COMED Medical Systems Co., Ltd.
4
EVA-HF325/525 Troubleshooting
Filament sensing “HIGH”(see JOB CODE 023) E30
High current input to filament Check filament drive board section TEMP is abnormal(see JOB CODE 024)
E32
Tube temperature is overheated Check Tube temperature section and signal connection. IPM fault(see JOB CODE 025)
E00
Check IPM fault section
COMED Medical Systems Co., Ltd.
5
EVA-HF325/525 Troubleshooting
3. FLOW CHART Power ON
Text 1 LED 9 ON LED10,11,12 ON
Check 1
LED 7,8 ON
Check 2
Text 2
Text 3 LED 1 OR 2 ON LED 3 OR 4 OR 5 ON
Check 3
Contactor 1 Contactor 2
Check 4
Text 4
Text 5 LED 4 ON
Initialize OK
COMED Medical Systems Co., Ltd.
6
EVA-HF325/525 Troubleshooting
TEXT 1 Led 9 ON is meaning of system power input OK. Led 10,11,12 ON is meaning of power supply +5V,+12V,-12V OK. That Leds located on Relay power board.
TEXT 2 Led 7 ON is meaning of charge of capacitor ON. Led 8 ON is meaning of Filament power input(AC60) ON. Bucky power input(AC110) ON. Lock power input(DC30) ON. Collimeter power input(AC20) ON. And relay power(DC24V) in relay power board That Leds located on Relay power board.
TEXT 3 Led 1 ON is meaning of selected large focus filament. Led 2 ON is meaning of selected small focus filament. Leds ON will select Led 1 or Led 2. Led 3 ON is meaning of selected bucky 1. Led 4 ON is meaning of selected bucky 2. Led 5 ON is meaning of selected bucky 3. Led 3,4,5 all OFF is meaning of selected NO bucky and That leds ON will selected Led 3 or Led 4 or Led 5. That Leds located on Relay Power Board.
TEXT 4 Contactor 1, 2 is meaning of Main contactor and charge contactor ON.
TEXT 5 That meaning is System initializing OK. Led 4 located on HT contol cpu board.
COMED Medical Systems Co., Ltd.
7
EVA-HF325/525 Troubleshooting
4. ERROR CHECK
CHECK 1 -
Line input power check(see JOB CODE 001)
-
Communication cable check(see JOB CODE 002)
-
Membrane key check(see JOB CODE 003)
-
Bridge diode in Relay power board(see JOB CODE 004)
-
AC12V input check(see JOB CODE 005)
CHECK 2 -
Power(+5V,+12V,-12V) in Relay power board(see JOB CODE 006)
-
Power ON signal check(see JOB CODE 007)
CHECK 3 Check ERROR CODE TABLE
-
CHECK 4 -
Check charge/discharge monitor board(see JOB CODE 008)
-
Check charge contactor(see JOB CODE 009)
-
Check charge signal in HT control cpu board(see JOB CODE 010)
Additional CHECK point - If stopped LCD display “ initialize … “ always. And No ERROR CODE display in LCD dispay(see JOB CODE 007)
COMED Medical Systems Co., Ltd.
8
EVA-HF325/525 Troubleshooting
5. DATA TABLE The Calibration data and some technique data stored in the U7-UV_PROM And APR data stored in the U23-EEPROM of the Console display Board. 5.1 CALIBRATION DATA TABLE IN UV_PROM
COMED Medical Systems Co., Ltd.
9
EVA-HF325/525 Troubleshooting
5.2 APR DATA TABLE IN EEPROM
COMED Medical Systems Co., Ltd.
10
EVA-HF325/525 Troubleshooting
6. WAVEFORM DATA
1. Control waveform
3. Preheating
2. Output in gate of Q3,Q4
4. Ready preheating (at high mA data)
COMED Medical Systems Co., Ltd.
11
EVA-HF325/525 Troubleshooting
5. Ready preheating (at low mA data)
7. KV data output (at high kV data)
COMED Medical Systems Co., Ltd.
6. KV data output (at low kV data)
8. Exposure enable
12
EVA-HF325/525 Troubleshooting
9. control waveform
11. –kV feedback
10.+kV feedback
12. dead time control
COMED Medical Systems Co., Ltd.
13
EVA-HF325/525 Troubleshooting
13. TL594 output (at low kV data)
15. KDR1~KDR4 output
14. TL594 output (at high kV data)
16. +mA feedback
COMED Medical Systems Co., Ltd.
14
EVA-HF325/525 Troubleshooting
17. –mA feedback
18. Rotor signal
COMED Medical Systems Co., Ltd.
15
EVA-HF325/525 Troubleshooting
7. JOB CODE 7.1 JOB CODE 001 Tools & Equipment required : DVM THIS GENERATOR IS PERMANENTLY CONNECTED TO THE POWER LINE. INTERNAL PARTS OF THE GENERATOR(LINE FUSE, ETC.) ARE PERMANENTLY
POWERED
ON
THROUGH
POWER
LINE
ALTHOUGH THE CONSOLE IS OFF. BE SURE THAT THE SAFETY SWITCH IS OFF BEFORE HANDLING ANY INTERNAL PART OF THE EQUIPMENT.
THIS 325/525 GENERATOR LINE POWER IS SINGLE PHASE 220VAC. FOR SINGLE PHASE GENERATOR, CONNECT THE POWER CABLE L1 AND L2 (N: NEUTRAL) AND GND TO THE TERMINAL BLOCK.
x-ray power terminal line power cable
system power terminal AC connection wire
Check 1. input power of line power cable L1 and L2 voltage is 215 ~230VAC Check 2. connection of AC connection wire are tight to each terminals
COMED Medical Systems Co., Ltd.
16
EVA-HF325/525 Troubleshooting
7.2 JOB CODE 002 Tools & Equipment required : DVM CHECK POWER ON/OFF/COM OF J2 IN COMMUNICATION CABLE SIGNAL
J1 J2 Pic1.view in mother board J1 +5V(H→C) GND(H→C) TXD RXD HX-RAY HREADY HINHIBIT HTRANSFER CREADY CX-RAY CTRANSFER CINHIBIT +12V(C→H) +12V(H→C)
J1
AND
J2
IS
J2
J1 Pic2. wiew in console J2
+KV -KV DC24V-2(H→C) TUBE2 READY TUBE2X-RAY FLUORO MODE POWER OFF POWER ON POWER COM C+5V C+12V C GND
CONNECTED
CONTROLLER
AND
CONSOLE
WITH
RS-232
COMMUNICATION CABLE SO, FIX TIGHTLY WITH SCREW WHEN CONNECT J1,J2. AND THEN CHECK IT.
COMED Medical Systems Co., Ltd.
17
EVA-HF325/525 Troubleshooting
ONESIDE COMMUNICATION CABLE LOCATED MOTHER BOARD INSIDE
CONTROLLER
INTERCONNECTION
ARE
FACTORY
MADE. ONLY OPPOSITE CONNECTION SHOULD BE CONNECT TO CONSOLE. AND COMMUNICATION CABLE LENGTH IS 10M. FOR CABLE LENGTH REFER TO STANDARD RS-232 COMMUNICATION WIRE LENGTH. IF ROOM REQUIRE MORE THAN LENGTH MUST NOT OVER THAN 15M. IF WIRE LENGTH MORE THAN 15M WILL BE OCCUR COMMUNICATION FAILURE.
COMED Medical Systems Co., Ltd.
18
EVA-HF325/525 Troubleshooting
7.3 JOB CODE 003 Tools & Equipment required : DVM Check SW49(power off), SW50(power on) connection. PB0 PB1 PB2 PB3 PB4 PB5 PB6 PB7 PC0 PC1 PC2 PC3 PC4 PC5 PC6 PC7 NC SWOFF SWON SWCOM
SW1
SW2
SW3
SW4
SW5
SW6
SW9
SW10
SW11
SW12
SW13
SW14
S17
SW18
SW19
SW20
SW21
SW25
SW26
SW27
SW28
SW29
SW30
SW33
SW34
SW35
SW36
SW37
SW38
SW41
SW42
SW43
SW44
SW7
SW8
SW39
SW49
SW50
power OFF
power ON
misprint silkscreen membrane key
COMED Medical Systems Co., Ltd.
19
EVA-HF325/525 Troubleshooting
7.4 JOB CODE 004 Tools & Equipment required : DVM Check D9(bridge diode) of Relay power board
bridge diode
COM/ON/OFF
LED9
Check diode input AC12V and output DC20V
R40 1.2K
D16
RY9 SW
1N4001 LD9 LED
J1-21,43(AC220V) J1-19,41(AC220V) J1-20,42(AC220V)
-
R34
+
R35 1K
D9 DF06
+
C12
220uF/35V
R37
Q6 2N2222A
1K R38 4.7K
Q5 2N2222A
1K
R36 4.7K
to console control J1-1(S/W ON)) J1-2,24(S/W COM) J1-23(S/W OFF) J1-22(AC12V) J1-44(AC12V)
LED9 ON SWITCH ON/COM/OFF LINE IS CONNECTED WITH COMMUNICATION AND CONSOLE.
7.5 JOB CODE 005 COMED Medical Systems Co., Ltd.
20
EVA-HF325/525 Troubleshooting
Tools & Equipment required : DVM Check AC12V in Pictures Pic 1. SUB TRANSFORMER AC12V(to main power board) AC20V SYSTEM POWER TERMINAL Pic 2. MAIN POWER BOARD J5(11P)
J5
J9(6P) J9 AC220V AC220V AC12V (From AC12V AC20V AC20V
(To mother board) sub-transformer)
DC24V-2 GND DC24V-1 AC220V(N) AC220V(N) IPM AC220V IPM AC220V AC12V AC12V AC220V AC220V
mother board drawing refer
7.6 JOB CODE 006 Tools & Equipment required : DVM Check TEST POINT +5V,+12V,-12V in Relay Power Board
COMED Medical Systems Co., Ltd.
21
EVA-HF325/525 Troubleshooting
Check +5V(4.95V ~5.10V) Check +12V(11.95V ~12.10V) Check –12V(11.95 ~12.10V)
LED10 ON : +5V LED11 ON : +12V LED12 ON : -12V
R10 : +5V ADJUST R25 : +12V ADJUST R33 : -12V ADJUST
R10 R33
R25
7.7 JOB CODE 007 Tools & Equipment required : DVM Check POWER ON signal. 1. Check output signal 17pin of U19 in HT Control Cpu Board
COMED Medical Systems Co., Ltd.
22
EVA-HF325/525 Troubleshooting
17pin of U19(power on signal) is High→Low
2. Check input signal 2pin of J3 in Main Power Board J3
ROTOR ON POWER ON 220/40 CHANGE TUBE SELECT ROTOR SENSE ROTOR SENSE BUCKY AC110V BUCKY AC110V
Relay 3 and D5
SP4
TO MAIN TRANSFORMER J2-4
FROM SUB TRANSFORMER
F7
J2-6
RY3
20A
D5
+24V
POWER ON
1N4007
Pic. Main power board drawing One wire of sub transformer (Input power 220VAC) and main transformer (220VAC) is interconnected by Relay 3. So, operating by Relay 3 will give Input power to system. 3. Check input signal to Relay8 in Relay Power Board
COMED Medical Systems Co., Ltd.
23
EVA-HF325/525 Troubleshooting
J2-13,35(DC24V-1) J1-4(DC24V-2)
R39 2.2K
LD8 LED
D17
RY8
J1-12,34(DC24V-2 OUTPUT) to 1pin of J11and 2pin of J2 in mother board and
1N4001
J2-26(POWER ON) To Relay Power
4. Check input power DC24V-2, DC24V-1(All power is from Main power board)
If stopped LCD display “ initialize … “ always and No ERROR CODE display in LCD dispay , Check this section
7.8 JOB CODE 008 Tools & Equipment required : DVM Check Led1 light ON and Charge Capacitor storge voltage(about 310VDC)
COMED Medical Systems Co., Ltd.
24
EVA-HF325/525 Troubleshooting
P2(charge sense) P3(discharge signal) R3- R9(33K) this resistor endure at 230VAC Must check input power voltage(210 – 230VAC)
Discharge signal refer to Relay Power Board drawing LED1
J1 + -
1 2
R1 68K/3W
NOTE1 R2 27k/3w
R4 15K/3W
R3 100K/3W
R5 15K/3W
R6 15K/3W
D1 IN4764
R7 15K/3W
R8 15K/3W
R9 15K/3W
Q1 IRFP450
LED1 D3 1N4733
D2 IN4764
U1 1
6
P3
R4
1 2
3.3M 2
5
3
4
TLP531
NOTE1 :
2005.10.7
DIS CHARGE
To 3,4pin of J10 in Mother Board
P2 1 2 3
To 5,6pin of J10 in Mother Board
CHARGE SEN
AC220/1PHASE - > 27K/3W AC380/3PHASE - > 18K/3W
Note2 : AC 220/1Phase -> 33K AC 380/3Phase -> 9K
Pic. Charge/discharge Monitor Board THIS GENERATOR IS PERMANENTLY CONNECTED TO THE POWER LINE. INTERNAL PARTS OF THE GENERATOR(LINE FUSE, ETC.) ARE PERMANENTLY POWERED ON THROUGH POWER LINE ALTHOUGH THE CONSOLE IS OFF. BE SURE THAT THE SAFETY SWITCH IS OFF BEFORE HANDLING ANY INTERNAL PART OF THE EQUIPMENT. THE MAIN STORGE CAPACITORS RETAIN A LARGE PORTION OF THEIR CHARGE FOR APPROXIMATELY 3 MIN. AFTER THE UNIT IS TURNED OFF.
7.9 JOB CODE 009 Tools & Equipment required : DVM Check main and charge contactor action.
COMED Medical Systems Co., Ltd.
25
EVA-HF325/525 Troubleshooting
main
charge
precharge resistor
Sub Relay Board J2
J3
DC24V-2 GND MAIN RELAY CHARGE RELAY PRECHARGE RELAY
MAIN MAG MAIN MAG CHARGE CHARGE MAGELAY
7.10 JOB CODE 010
Tools & Equipment required : DVM Check each signal actions in HT control cpu board. U19
U19 1 2 3 4 5 6 7 8 9
I1 I2 I3 I4 I5 I6 I7 I8 GND
R39-46 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 COM
18 17 16 15 14 13 12 11 10
10 J2-27(MAIN MAG) J2-5(POWER ON) J2-28(PRECHARGE) J2-6(CHARGE MAG) J2-7(TUBE SEL) J2-29(220/60) J2-30(HIGH SPEED RTON) J2-8(ROTOR ON) 1K
ULN2803
POWER OFF
POWER ON
MAIN MAG
CHARGE MAG
18PIN
low
high
low
low
17PIN
high
low
low
low
COMED Medical Systems Co., Ltd.
26
EVA-HF325/525 Troubleshooting
16PIN
low
high
low
high
15PIN
low
high
high
low
If power not charge in storge capacitor will displayed “E28” in console display.
7.20 JOB CODE 020 Tools & Equipment required : ROM-WRITER, CALCULATOR Check kV, mA, comp data in Calibration Mode or read and re_write Serial-ROM with ROMWRITER. See Serial-ROM TABLE(chapter.6) address
description
address
description
0007600 ~ 0007610
KVp data at 25mA
0007820 ~ 0007830
mA data at 50mA
0007620 ~ 0007630
kVp data at 50mA
0007840 ~ 0007850
mAp data at 75mA
0007640 ~ 0007650
KVp data at 75mA
0007860 ~ 0007870
mA data at 100mA
0007660 ~ 0007670
KVp data at 100mA
0007880 ~ 0007890
mA data at 150mA
0007680 ~ 0007690
KVp data at 150mA
00078A0 ~ 00078B0
mA data at 200mA
00076A0 ~ 00076B0
KVp data at 200mA
00078C0~ 00078D0
mA data at 250mA
00076C0~ 00076D0
KVp data at 250mA
00078E0 ~ 00078F0
mA data at 300mA
00076E0 ~ 00076F0
KVp data at 300mA
0007900 ~ 0007910
mA data at 400mA
0007700 ~ 0007710
KVp data at 400mA
0007920 ~ 0007930
mA data at 500mA
0007720 ~ 0007730
kVp data at 500mA
0007A00 ~ 0007A00
Comp data
0007800 ~ 0007810
mA data at 25mA
0007A20 ~ 0007A20
Technical data
See EEPROM TABLE(chapter.6) address
description
0003000 ~ 00030D0
KVp, mA, time,bucky selecting data at very big human
0003100 ~ 00031D0
KVp, mA, time,bucky selecting data at big human
0003200 ~ 00032D0
KVp, mA, time,bucky selecting data at small human
0003300 ~ 00033D0
KVp, mA, time,bucky selecting data at very small human
Pic. Calibration Mode LCD display
KVp data, mA data, comp data
COMED Medical Systems Co., Ltd.
27
EVA-HF325/525 Troubleshooting
If need change Calibration data with new data saved in Serial-ROM. You will read Serial-ROM with Rom_writer and re_save data in same address. if you save abnormal data, it will be displayed ERROR message in Console Display LCD. If APR data brocken by noise. You can re_save with pushing set switch(save switch) in APR MODE. Pic. Serial-ROM and EEPROM location(Console display board(operating_panel_console board)
U7(M27C256B) ) U25(25256)
COMED Medical Systems Co., Ltd.
28
EVA-HF325/525 Troubleshooting
7.21 JOB CODE 021 Tools & Equipment required : ROM-WRITER, CALCULATOR Check reference of calibration data in Serial-ROM with ROM-WRITER.
Reference of Calibration data
All calibration data of KV(MA)SET is sum of reference data and calibration data. Must not change value of reference data at Serial-ROM. Only changeable calibration data. KV-SET = reference data + calibration data MA-SET = reference data + calibration data MA-COMP = calibration data Pic. Reference address
and
data in SerialROM
KVp reference data at 200mA
mA
reference
data at 25mA
COMED Medical Systems Co., Ltd.
29
EVA-HF325/525 Troubleshooting
mA reference data at 50mA
mA reference data at 75mA
Pic. Reference data address table in Serial-ROM address 0006E00 ~ 0006E55 0006E80 ~ 0006ED5 0006F00 ~ 0006F55 0006F80 ~ 0006FD5 0007000 ~ 0007055 0007080 ~ 00070D5 0007100~ 0007155 0007180 ~ 00071D5 0007200 ~ 0007255 0007280 ~ 00072D5 0007300 ~ 0007355
description KVp reference data at 200mA mA reference data at 25mA mA reference data at 50mA mA reference data at 75mA mA reference data at 100mA mA reference data at 150mA mA reference data at 200mA mA reference data at 250mA mA reference data at 300mA mA reference data at 400mA mA reference data at 500mA
If you want detail calibration information see CALIBRATION MANUAL
7.22 JOB CODE 022 Tools & Equipment required : DVM,OSC 1. Check ROTOR connecton Pic. Main power board and drawing
COMED Medical Systems Co., Ltd.
30
EVA-HF325/525 Troubleshooting
J10
SSR CT J7(To Interface Terminal)
J10
1 2
SP(spark killer)
CT1 0VAC
ROTOR COM/MAIN/PHASE
DC24V-1 SP1
SSR1
220VAC SSR
D4 1N4007 J7-1 J7-2 J7-3 J7-4 J7-5 C4
24uF
Pic. Interface Terminal
ROTOR ON
To ROTOR of TUBE
Pic. Rotor(anode of tube) TEMP G(PHASE) B(MAIN)
COMED Medical Systems Co., Ltd.
31
EVA-HF325/525 Troubleshooting
W(COM)
Check connection of ROTOR wire.(Checking point is interface terminal) Must power OFF when check the ROTOR connection. Will accuse electrical shock.
Checking Point
Value
Accuracy limit
COM - PHASE
Ω
58 ~ 62Ω
COM - MAIN
Ω
29 ~ 32Ω
MAIN - PHASE
Ω
85 ~ 95Ω
Check ROTOR input signal when pushed ready switch. Checking Point
Value
Accuracy limit
COM - PHASE
VAC
280 ~ 310VAC
COM - MAIN
VAC
210 ~ 230VAC
MAIN - PHASE
VAC
280 ~ 310VAC
Check ROTOR feedback signal and waveform(use OSC equipment) Checking Point
Value
TP9
Accuracy limit VDC
0.8 ~ 0.95VDC
Rotor feed back signal must check with Osciloscope. Rotor rotating signal is too short time to exam. ON 0.2sec
OFF 3sec
ON
OFF
0.2sec
Pic.Rotor feedback signal in HT control control board
COMED Medical Systems Co., Ltd.
32
EVA-HF325/525 Troubleshooting
Waveform of ROTOR feedback signal refer chapter 6 18.ROTOR signal
U19 1 2 3 4 5 6 7 8 9
I1 I2 I3 I4 I5 I6 I7 I8 GND
R39-46 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 COM
18 17 16 15 14 13 12 11 10
10 J2-27(MAIN MAG) J2-5(POWER ON) J2-28(PRECHARGE) J2-6(CHARGE MAG) J2-7(TUBE SEL) J2-29(220/60) J2-30(HIGH SPEED RTON) J2-8(ROTOR ON) 1K
ULN2803
If did not output sensing signal of ROTOR sense, it will display “E26/27” in Console display LCD.
COMED Medical Systems Co., Ltd.
33
EVA-HF325/525 Troubleshooting
7.23 JOB CODE 023 Tools & Equipment required : DVM 1. Check the feedback signal.(HT control control board) filament feedback signal(from filament drive board) +12V C49
+12V
104
0.6V 8
TP11 VR8
3
10k
2
U12A
+
D24 1
-
1N4148
4
R65
R63 4.7K
8
TL082A J1_13,35(FILSEN)
6
R64 4.7K
U6 ULN2004
11 J1_12(FILLOW)
10k C48 104
+12V
5 VR9
6
U12B
+ -
TL082A
5
U6 ULN2004
1N4148
12 J1_34(FILHIGH)
4
10k
D25 7
TP10
C50 3.3V -12V 104
Check time is not ready switch pushed. Check point
Value
Accuracy limit
Filament feedback sensing
VDC
0.6 ~ 0.9VDC
Sensing LOW limit(TP11)
VDC
0.1 ~ 0.5VDC
Sensing HIGH limit(TP10)
VDC
3.0 ~ 3.5VDC
If you want LOW limit change, use with VR8. and HIGH limit is VR9. If filament sensing value is lower than TP11, will display “E29” in Console Display LCD. And higher than TP10, will display “E30” in Console Display LCD.
COMED Medical Systems Co., Ltd.
34
EVA-HF325/525 Troubleshooting
7.24 JOB CODE 024 Tools & Equipment required : DVM Pic. TUBE(anode) TUBE temperature signal is mormal close but temperature about over than 70° will open.
Pic. Interface terminal Temp(from tube)
If TUBE temperature signal is open, will display “E32” in Console display LCD. J7
J7(10P)
BUCKY1 BUCKY1 BUCKY2 BUCKY2 BUCKY3 BUCKY3 TUBE1 TEMP TUBE1 TEMP TUBE2 TEMP
COMED Medical Systems Co., Ltd.
35
EVA-HF325/525 Troubleshooting
TUBE2 TEMP
Tube 2 temp not used 325/525 generator use two tube(option), so,tube2 temperature signal must come to ground
7.25 JOB CODE 025 Tools & Equipment required : DVM 1. Check that the wires maked “+kV, -kV, +mA, -mA” in the HV Tank are well placed and tighten. Pic. HV tank(High tension transformer) 1.-kV(black) 2.+kV(red) 3.+mA(green) 4.-mA(white) 5.GND(yellow/shield) 6.small 7.com 8.large
Pic. Mother board(Main card board)
From HV tank
Connection check point Check Point
Value
Accuracy limit
-kV – GND
kΩ
113 ~ 114kΩ
+kV – GND
kΩ
112 ~ 113kΩ
COMED Medical Systems Co., Ltd.
36
EVA-HF325/525 Troubleshooting
+mA – GND
kΩ
2 ~ 2.5kΩ
-mA - GND
kΩ
2 ~ 2.5kΩ
Must power OFF When check the connection 2. Check the IPM driver Pic. IPM Drive Board drawing. D3 FR107
PC2 2
8 6
3
5
SPR(+5V INPUT1) R1 6.8K
C7
J1-2
MAYER CAPACITOR
0.1
CP(-KV DR1)
HCPL4503
J1-3
VPC(GND1)
J1-4
PC1 8
2
6
3
5
VP1(+15VOUTPUT1) C5 0.1
J1-1
FPO(-FAULT1)
J1-5
HCPL4503
Table. Check Point(All power was setted originally in factory. Don’t need contrast) Checking point
Value
Accuracy limit
J1 – 1,4
VDC
14.9 ~ 15.1VDC
J1 – 2,4
VDC
4.95 ~ 5.05VDC
J2 – 1,4
VDC
14.9 ~ 15.1VDC
J2 – 2,4
VDC
4.95 ~ 5.05VDC
Pic.IPM and IPM drive board IPM IPM drive board J2 J1 J3 J1(J2) FAULT GND -KV DR +5V +15V
COMED Medical Systems Co., Ltd.
+5V GND KDR1 KDR2 FAULT AC220V AC220V
37
EVA-HF325/525 Troubleshooting
ALWAYS HAVE “IPM DRIVER PCB” CONNECTED WHEN MAIN POWER IS ACTIVATED
IN
THE
GENERATOR.
WHITOUT
“IPM
DRIVER
PCB”
CONNECTED, PERMANENT DAMAGE WILL OCCUR TO IPM’S.
7.26 JOB CODE 026 If “E00” display at console display LCD when x-ray shooting. Check point. 1. feedback waveform Pic. Testing point kV, mA
+mA
+kV GND When measure “+kV” wih OSC, real kV is 10kV per 1V. Also, real mA is 20mA per 1V. Case 1
IPM FAULT “E00” 1. HV cable bad(over than 80kVp) change HV cable. 2. Calibration data abnormal check calibration data 3. IPM damage(sound “ting”) change IPM 4. check IC TL594,ULN2004 in HT control control board and AD7224 in HT control cpu board KOUT, TL594 output signal waveform check.
COMED Medical Systems Co., Ltd.
38
EVA-HF325/525 Troubleshooting
U27
I1 I2 I3 I4 I5 I6 I7 GND
O1 O2 O3 O4 O5 O6 O7 COM
5
16 15 14 13 12 11 10 9
16 15
5.
ULN2004
3 1 2 7
CT +IN1 -IN1 FB +IN2 -IN2 GND U3 TL594
RT
1 2 3 4 5 6 7 8
13 12 11 10
6
U6
VCC C1 C2 E1 E2 REF CTL DTC
12 11 8 9 10 14 13 4
15
HV
9 8 7 6
D0 D1 D2 D3 D4 D5 D6 D7
WR
VDD RS
VREF CS LDAC AGND DGND VSS VOUT
18 17
3 14 16 4 5 1 2
AD7224
transformer
bad
change HV transformer. Case 2
Sometimes IPM FAULT “E00” 1. Input power(kW) “LOW” Change exposure condition 100kVp 300mA 0.3sec 100kVp 100(150)mA 0.9(0.6)sec This is input current low, so change mA condition. 2. IC SG3524 in filament drive board is damage. change new IC. 2. Aanother factor. 1. If IPM driver board broken, always display “E00”(power on state ,not exposure state) change IPM driver board check IPM drive input power(220VAC) and signal connection. 2. If IPM was damaged, sound “ ting—“and display “E00” at high kVp exposing change IPM
COMED Medical Systems Co., Ltd.
39
EVA-HF325/525 Troubleshooting
8. X-RAY CHECK 1. FILAMENT DRIVE BOARD CHECK Tools & Equipment required : DVM,OSC Pic. Filament drive board ① filament enable signal (high low) ② control waveform (see waveform 1) ③ output in gate of Q3,Q4(see waveform 2) ④ preheating wave(see waveform 3) ready wave with high mA data(see waveform 4) ready wave with low mA data(see waveform 5)
②
③
DC0V - 100V
+12V
C22
+
ZD3
100uF/35V
1N4745
D9
T3
D10
FR102
D7
FR102
R27 1K
FR102
Q11 IRFP250
D11
Q7 2N2907
1N4745
C24 473
Q8 CLK R29
Q3 IRF630
10 4 5 8 U2 LM 3524
SHTDWN +SENSE -SENSE GND
1K R9
15
9
VIN COMP
OSCOUT CT
CA CB EA EB
12 13
U3 1 2 3 4
11 14
NC INA GND INB
NC OTA VDD OTB
R19 1K
FR102
R32 1K
R31 1K
FR102
D15
FR102
FR102
Q10
5K
D17
FSENSE
R49 10K
C28 1uF
+
R47 10K
D19 1N4148
R45 20K
R42 R38 47K/3W
1K
473
C16
C19
104
104
R35
104
R36 100
560
R39 0.5/5W
1K 20K
C35
T5
R43
R48
D21 1N4148
470uF/200V
C32
10uF
1N4148
224
Q11 2P4M
20K C34
+ C30
+12V
R46
D22
R44 1K
D20
+12V
D18 1N4148
C33 102 -12V
0.1
1N4148
J 1-4,5,26,27(FIL OUTA)
U5B
-
7
100uF/35V
+
U5A
6
+
2
J 1-6,7,28,29(FIL OUTB)
3
4
TL082 4
TL082 C36
-
1
5
8
C37 8
+
ZD10
C13
C27
1N4745
4.7 1/2W
20K
R50 10K
C26 102
R34
FR102
R51
1N4745
2N2907
1 1/2W ZD8 1N4733
1K
ZD9 Q9 2N2907
R33
1K
J2-19,20,41,42(-12V)
D14
Q4 IRF630
R22
0.5/3W
222/1KV
D13
T4
D8
R18 1K
100/3W R41
Q12 IRFP250
4.7 1/2W
FR102
1N4745
104
R30
ZD4
8 7 6 5
TC4427
VR3
R23
J 2-30(M SY)
V-
1
472
RT
VREF
3 7
V+
6 4.7K
C18
2
R52
C15
104
C25 1uF
16
5K
D12
1N4745
C31
ZD5 1N4733
C21 0.1
R37 47K/3W
R40 C14
ZD7
2N2907
1 1/2W
+12V
VR2
C23 102
470uF/200V
224
ZD6
FR102
R28 1K
+ C29
C12
-12V 0.1
COMED Medical Systems Co., Ltd.
40
EVA-HF325/525 Troubleshooting
①
④
Filament control frequency check Checking point
Value
Accuracy limit
CLK in U2(SG3524)
65 ~ 75㎲ec
㎲ec
Adjust point : VR2
NF1
Q1
R1
Q2
R2
Q3
R3
Q4
R4
0.5/3W
BUX48A
BD1 + KBU10M
0.5/3W
BUX48A
J2-13,14,35,36(AC60V) NOISE FILTER
0.5/3W
BUX48A J2-15,37(EARTH)
R11 47K/3W
+ C1
+ C2
470uF/200V
470uF/200V
R10 0.5/3W
BUX48A Q5
1K/3W
R24 5.2K
2.2K/1/2W
1N4148
C17
+
R17
R13 390
R15 820
47K
C4 C5 0.1
J2-21,22,43,44(GND) R7 2.4K
R14 12K
47K
5
+ 4
+12V
R12 12K R16
6
-
1N4148
D3
Q6 2N3440
104
U6B LM358 7
D1
R6
J2-17,18,39,40(+12V)
C7
R5
TIP42
DC0 - 100V
0.2/5W
+12V
8
-
R8 2.7K
U6A LM358 1
D2
100uF/35V
0.1 8
J2-10,11,32,33(AC60V)
R20
2
-
47K
3
+
1N4148
R21 4
47K C3 0.1
R54 10K
J2-31(FOUT)
C8
VR1 2K
8
+12V R57
104
2 10K
3
-
1
+
U4A TL082
4
C9 R58 10K
R55 R53 1K
120K 8
-12V R56
104
6 100K
R26 4.7K
C6 0.1
-
7
+
U4B TL082
4
C10
5
R25 10K
8
+12V +
6 5
1
-
U1A TL082
+
2 3
FSENSE
-12V
4
4
U1B TL082 C11
8
104 7
J1-13,24,35(FSEN)
104
⑤
⑥
VR1
⑤ Filament feedback signal (to HT control control board) ⑥ FOUT signal (from HT control cpu board)
COMED Medical Systems Co., Ltd.
41
EVA-HF325/525 Troubleshooting
All mA step condition will
change uniformly with VR1 contrasting FOUT
signal. And, If you want each mA step condition change, use Calibration mode
BECAREFUL WHEN YOU CHECKING “FOUT SIGNAL”.
IF
ELECTRICAL
SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE BROCKEN.
Pic.Filament drive board
COMED Medical Systems Co., Ltd.
42
EVA-HF325/525 Troubleshooting
SomeTimes, U1(TL082) is broken by filament noise. This IC operate feedback sensing regulator. If U1 broken, E30 display in Console LCD. Because, filament feedback signal is pull-off(12VDC) by broken IC. So filament feedback signal feedback “HIGH” by filament sensing circuit in HT control control board.(see JOB CODE 023)
2. HT CONTROL CONTROL BOARD CHECK Tools & Equipment required : DVM,OSC Pic.
HT
control control board
COMED Medical Systems Co., Ltd.
43
EVA-HF325/525 Troubleshooting
Test exposure condition : 40kV/70kV, 200mA ,0.025sec) ④
⑧ +12V R38 4.7K
R37 4.7K
8
KV
2
C8
R19
100P
J1_27(-MASEN) J1_5(+MASEN)
ZD3
102 J1_26(-KVSEN) J1_4(+KVSEN)
6 1M
C5
D5
1N4744
1N4148
T2 100uH
+12V
1N4148
U2B
C9
C10
273
100P
R17
7
D7
D8
1N4148
1N4148
R1 100K
VR2
3
1M
2 D 1 IN
S
20K ADG201A U4B
VR1
154K
14
ADG201A
8
J2_32(KOUT)
3
+12V
D10 1N4148
TL082A C11
-12V
R23 52K
C12 273
R24 100K
100P
D11
D12
1N4148
1N4148
16 15 3 1 2 7
24.9K
-
C13
C14 104
R25 1K
1N4752
1N4148
1N4148
+12V
U1B
-
7
+ TL082A
C2
ZD2
D3
D4
104
1N4752
1N4148
1N4148
R4
C3 103
100K
J1_23(M SEN) 10K
R26 1K
CT +IN1 -IN1 FB +IN2 -IN2 GND U3 TL594
VCC C1 C2 E1 E2 REF CTL DTC
104
12 11 8 9 10 14 13 4
TP3
TP1 C16
R27
0.22
10K
VR3 10K
C17
U6
TP4
R29 1K
R30 1K
1 2 3 4 5 6 7 8
I1 I2 I3 I4 I5 I6 I7 GND
O1 O2 O3 O4 O5 O6 O7 COM
16 15 14 13 12 11 10 9
R33-R36
J1_7(KDR1) J1_29(KDR2) J1_8(KDR3) J1_30(KDR4)
10
ULN2004
104
R21 10K
4.99K D14
J2_9(EXP) J2_31(KSY)
U5A 2 1
1N4148 R31 1K
D13
R32 1K
D IN
S
3
13 U4 ADG202A
ADG201A U5C
1N4148 10 9
+5V
D IN
7 8
D IN
C24 104
4
C25 104
S
11 13
ADG201A U5D S
6
U5 ADG201A
C26 104
4
C27 104
ADG201A
①
1
②
③
⑤
⑥
⑦
kV output data voltage from 2 of U27 in Ht control cpu board 1(when exposured) kV low (40kV see waveform 6) kV high(70kV see waveform 7)
2
exposure/kV enable signal high low (see waveform 8)
3
TP 2 is control waveform(see waveform 9)
COMED Medical Systems Co., Ltd.
R7 10K
R5
+MA
+12V
MA
-12V
103
R20
R6 10K
6
5 103
R22
1
4
D9 1N4148
2
+
104
5
C23
C15
ADG201A
683
U2A
D2
1K
TP2
RT
C18
15 D 16 IN
S
D1
1K
R28
U5B
ZD1
R2 15 D 16 IN
S
10K
+KV
6 100K
C1
T1 100uH
U4A
R14
14
R3 1K
104
100P
R13
3
S
103
C22
C7
ZD4 1N4744
D IN ADG201A
-M A
R18 34K
+
-12V
-12V
102 10K
10K
TL082A
C6
R11
-
5
D6
R16
104
4
10K
2 1
8
R12
1M
J1_1(KSEN)
14
S
4
C21 8
R10
R8 ULN2004
1
D IN ADG201A U4A
C4
154K +12V
J2_26(R/F)
+
1M
R15
-KV
-
3
U4B 15 16
10
4
ZD3,ZD4À» 1N4744¿¡¼ 1N4746·Î ±³Ã¼ÇÔ 2001³â 8¿ù 29ÀÏ
U1A TL082A
R9 10K
U6
7
44
EVA-HF325/525 Troubleshooting
Checking point
Value
TP2
Accuracy limit kHz
39 ~ 41kHz
Adjust point : VR1 ④ test point : -kV, +kV feedback signal waveform +kV feedback (see waveform 10) -kV feedback(see waveform 11) 1V ≒ 10kV ⑤TP1(dead time control) Adjust point : VR3 (see waveform 12, when exposure) ⑥TP3,TP4 : TL594 output point kV data low(see waveform 13) kV data high(see waveform 14) E1 is 90 degree phase ahead of E2 ⑦KDR1,KDR4 (see waveform 15) KDR1,4 is 90 degree phase ahead of KDR2,3 KDR1 ~ KDR4 signal input to IPM DRIVE BOARD ⑧ test point : -mA, +mA feedback signal waveform +mA feedback (see waveform 16) -mA feedback(see waveform 17) 1V ≒ 20mA BECAREFUL WHEN YOU CHECKING “KOUT SIGNAL”.
IF
ELECTRICAL
SHORT HAPPEN, U27(AD7224) IN HT CONTROL CPU BOARD WILL BE BROCKEN.
9. MAIN CARD BOARD DRAWING J3(3P) J4(4P) J5(11P)
J6(8P)
J7(10P) J8(4P) J9(7P)
COMED Medical Systems Co., Ltd.
45
EVA-HF325/525 Troubleshooting
J11(5P) J10(6P)
J14(10P) J1(15P) J3
J7
TUBE2 AC220V TUBE2 AC220V
J4
J5
J6
TUBE2 READY TUBE2 X-RAY FLUORO MODE GND
J8
AC220V AC220V AC12V AC12V IPM AC220V IPM AC220V J9 AC220V(N) AC220V(N) DC24V-1 GND DC 24V-2
J2(15P) J15~J18(NOT USED) J19(6P) J13(7P) J12(7P) J11
J14
BUCKY1 BUCKY1 BUCKY2 BUCKY2 BUCKY3 BUCKY3 TUBE1 TEMP TUBE1 TEMP J12 TUBE2 TEMP TUBE2 TEMP
J19
AC220V(N) AC220V(N) AC60V AC60V
J13 AC220V AC220V AC15V AC15V AC20V AC20V RTN AC20V
ROTOR ON POWER ON J10 220/40 CHANGE TUBE SELECT ROTOR SENSOR GND AC110V AC110V
DC24V-2 GND PRECHARGE CHARGE MAIN
GND +5V KDR4 KDR3 IPM FAULT AC220V AC220V
-KV +KV +MA -MA GND SMALL COM LARGE TUBE2 AC220V TUBE2 AC220V AC220V AC220V +12V GND +12V +5V GND +5V
GND +5V KDR2 REFER JOBKDR1 CODE002 IPM FAULT AC220V AC220V
J1,J2 :
DISCHARGE4 DISCHARGE3 DISCHARGE2 DISCHARGE1 CHARGE SENSE GND
10. TECHNICAL IMFORMATION 10.1 MAIN SIGNAL DRAWING
COMED Medical Systems Co., Ltd.
46
EVA-HF325/525 Troubleshooting
CONSOLE
DISPLAY AND MEMBRANE KEY
CONTROLLER CONSOLE CPU BOARD COMMUNICATION CABLE
HT CONTROL CPU BOARD
KOUT
Feedback HT CONTROL CONTROL BOARD
+
TUBE
―
IPM driver e
IPM
FILAMENT DRIVEFOUT BOARD
HV cable RELAY POWER BOARD
HV transformer + terminal Filament power
MAIN POWER BOARD
T1 and T2 ROTOR power
10.2 MAIN SIGNAL EXPLANATION 10.2.1 communication Console display CPU and HT Control CPU communicate with cable.
COMED Medical Systems Co., Ltd.
47
EVA-HF325/525 Troubleshooting
This communication included below like as.
kV
mA
1ms
8bit
Time
1ms
1ms
8bit
Bucky
8bit
8bit
Console display CPU give kV data,mA data,time data,bucky select data to HT control CPU. And HT control CPU re-give same data to Console display CPU. So, Console display CPU check data error or good. If error detected, display EROOR code. Ex) KV data error(E16,E17), mA data error(E18,E19), time data(E20) … Pic. Console display CPU board
TP1(Txd), TP2(Rxd)
Pic. HT control CPU board TP1(Txd) TP2(Rxd)
COMED Medical Systems Co., Ltd.
48
EVA-HF325/525 Troubleshooting
TP3(Txd) TP4(Rxd)
TP3
TP4
TP1
TP2
If push KV up/down key in Console membrane. mA kV
mA , time, bucky select signal not transfer. Only kV data data transfer cf . kV data 8bit = 2 8 = P
P
256 information
COMED Medical Systems Co., Ltd.
49
EVA-HF325/525 Troubleshooting
kV data(138+1 = 139)
KV-SET data “139” in Console display cpu board give to HT control cpu board And, HT control cpu give data to AD7224 When communicate HT control cpu board with console display cpu board, sometimes error detected by electrical noise. 1. Check ground state. 2. Check shield state of communication cable. 3. Check noise of Power line.
]
10.2.2 Exposure data When exposed, kV data, mA data input to AD7224 with digital value( 0 – 256 information). And then, analog value(Kout and Fout) output from AD7224.
COMED Medical Systems Co., Ltd.
50
EVA-HF325/525 Troubleshooting
Pic. HT control CPU board
AD7224(8-Bit DAC with Output Amplifiers) ESD (electrostatic discharge) sensitive device. Electrostatic charges as high as 4000 V readily accumulate on the human body and test equipment and can discharge without detection. Although the AD7224 features proprietary ESD protection circuitry, permanent damage may occur on devices subjected to high energy electrostatic discharges. Therefore, proper ESD precautions are recommended to avoid performance degradation or loss of functionality.
COMED Medical Systems Co., Ltd.
51
EVA-HF325/525 Troubleshooting
Analog Kout data give to HT control control board
Analog Fout data give to Filament drive board
Pic. HT control cpu board
COMED Medical Systems Co., Ltd.
52
EVA-HF325/525 Troubleshooting
TP5 : Analog Kout data TP6 : not used TP7 : Analog Fout data AD7224 If you want to see waveform TP5,TP7(when push x-ray s/w) see Chap.6 6 and7 waveform. BECAREFUL WHEN YOU CHECKING “FOUT SIGNAL”.
IF
ELECTRICAL
SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE BROCKEN. BECAREFUL WHEN YOU CHECKING “KOUT SIGNAL”.
IF
ELECTRICAL
SHORT HAPPEN, U27(AD7224) IN HT CONTROL CPU BOARD WILL BE BROCKEN
10.2.3 Feedback signal Pic. Ht control control board
1
2 1. KOUT signal
3
4 2. Feedback signal
COMED Medical Systems Co., Ltd.
53
EVA-HF325/525 Troubleshooting
3. TL594 output signal
4. IPM drive signal
TL594(494) receive two signal (KOUT(-ini) and feedback signal(+ini))and give output signal(E1,E2) to IPM drive. If KOUT(analog data) is much, width of TL594 output signal would be widen. TL494(594) Technical data
COMED Medical Systems Co., Ltd.
54
EVA-HF325/525 Troubleshooting
COMED Medical Systems Co., Ltd.
55
EVA-HF325/525 Troubleshooting
10.
JOB CODE IMFORMATION
ERROR CODE OR
DESCRIPTION
JOB CODE
ERROR NOT POWER ON/OFF E16/17/18/19/20/21/22 E26/27 E28 E29/E30
Power on/off circuit abnormal
JOB CODE 001/002/003/004
Calibration and APR data broken
JOB CODE 020/021
Rotor Rotating abnormal
JOB CODE 022
Charge and discharge abnormal
JOB CODE 008/009/010
Filament drive board abnormal
JOB CODE 007/023
E32
Tube temperature abnormal
JON CODE 024
E00
IPM fault
JOB CODE 025/026
Only display “Initializing.” NOT KEY UP/DOWN
Not communication control and console Membrane key abnormal
COMED Medical Systems Co., Ltd.
JOB CODE 002/004/005/006/007 JOB CODE 003
56