AquilionCXL - Trouble shooting, 15 June 2012

2.5 Trouble shooting © Toshiba Medical Systems Corporation

1

Contents Diagnostics and Escalation Gantry Couch X-ray system Data Acquisition system Console Software

© Toshiba Medical Systems Corporation

2

Diagnostics and Escalation For performance and long life - Proper operation and maintenance • • • • •

Operator training - applications training Trouble shooting guide - mainly for operator Service manuals - trouble shooting procedures and flowcharts Toshiba service Toshiba escalation - local - regional - global support It is essential for prompt system recovery that abnormal system behaviour is reported appropriately

© Toshiba Medical Systems Corporation

3

Gantry

© Toshiba Medical Systems Corporation

4

GTSA - Status

[GTS] sts [GTS] *** dualport RAM cyclic status *** done_sts=4e41(NA) off_sts=52(R) rote_sts=52(R) x_ray_sts=52(R) bed_sts=52(R) gns_sts=52(R) err sts=3030383a30303030(008:0000) slid_inter=30(0) hite_inter=30(0) tilt_inter=30(0) r_spd=303030(000) x_ray_pos=30303030(0000) sld_dsp=2b3030303030(+00000) hite_dsp=2b303030(+000) tilt_dsp=2b303030(+000) warning=40(@) free_lgt_sts=30(0) preheat_cut=30(0) req_sts=40(@) pre_err=40(@) sen_chk=3030303c32303030(000