Toshiba xario service manual_maintanance.pdf

No. 2D730-503E SERVICE MANUAL
 FOR
 DIAGNOSTIC ULTRASOUND SYSTEM
 ! 
 MODEL SSA-660A
 [MAINTENANCE VOLUME]
 (2D730-503E

Views 604 Downloads 16 File size 11MB

Report DMCA / Copyright

DOWNLOAD FILE

Recommend stories

Citation preview

No. 2D730-503E

SERVICE MANUAL
 FOR
 DIAGNOSTIC ULTRASOUND SYSTEM
 ! 
 MODEL SSA-660A
 [MAINTENANCE VOLUME]
 (2D730-503E)

!

© TOSHIBA MEDICAL SYSTEMS CORPORATION 2004 ALL RIGHTS RESERVED

No. 2D730-503E

IMPORTANT! 1. No part of this manual may be copied or reprinted, in whole or in part, without written permission. 2. The contents of this manual are subject to change without prior notice and without our legal obligation.

C-!1

No. 2D730-503E



C-!2

*

No. 2D730-503E

REVISION RECORD REV.

DATE

REASON

(MM/YY)

INI.

09/'04

/AUTHOR Mr.

Muto

PAGE

SER.

DOC.

CHANGE D

No.

PRODUC T.

-------

TMWP7



R-!1

*

CONTENTS

No. 2D730-503E Page

- !a -

1.

No. 2D730-503E

INTRODUCTION 1 1.1

Maintenance Menu Configuration (For Users) 1

1.2

Service Menu Configuration (For Service Engineers) 2

2.

OUTLINE OF MENUS 1 2.1

Outline of Points Common to the Maintenance and Service Menus 1

2.2

Outline of the Maintenance Menu 1

2.3

Outline of the Service Menu 2

2.4

Reboot Function 4

3.

MAINTENANCE MENU FUNCTIONS 1 3.1

Activation and Termination 1

3.2

Archiving 2

3.3

System Maintenance 3

3.4

Remote Access Management 8

3.5

Image Quality Report 9

4.

SERVICE MENU FUNCTIONS 1 4.1

Activation 1

4.2

Termination - !b -

No. 2D730-503E

4 4.3

Information Function 5

4.4

Test Tools Function 83

4.5

Maintenance 93

4.6

Utilities 111

4.7

Shortcuts 103

- !c -

1.

INTRODUCTION The test system described in this manual is used for maintenance and servicing of diagnostic ultrasound systems. It consists of a maintenance menu which assists the user in maintenance of diagnostic ultrasound systems and a service menu that assists the service engineers in servicing diagnostic ultrasound systems. Note: The service menu functions are not available to users.

1.1

Maintenance Menu Configuration (For Users) The maintenance menu is composed of the following 6 items. ⎯

Menu name

Description (use and purpose)

1

Archiving

• Used to regularly back up user-specific data on separate media as a measure against problems such as HDD failure. • Allows the use of removable media.

2

System Test

• Used to check if the hardware is operating normally. • When something appears to be wrong with the operation of a device, this function can be used to check the status of the device. • This function can identify the malfunctioning board. • The test results can be saved in the HDD according to the specified settings

3

System Maintenance

• A template is provided for the user's daily check record.

4

Communication

• This function allows the user to transmit requests and notifications to the service company. • This function is also used to check and read the response from the service company.

5

Remote Access Management

• With this function, the user can permit the service company to perform remote maintenance on the system via a network or telephone line.

6

Image Quality Report

• When a problem (such as noise) occurs for an image, the settings, such as the imaging conditions, can be saved on the HDD together with the problem image

Note: 1. Data recorded on the HDD must be backed up regularly. 2. Menus 2 and 4 in the table above will be supported in the future.

1-!1

1.2

Service Menu Configuration (For Service Engineers) The service menu is composed of the following items: Item Information

Description (use and purpose)

1) Site

Looking up and updating site data

2) System

Looking up and updating system data

3) HW/SW a)

Common I/F

Looking up and updating hardware data

b)

Peripherals

Looking up and updating peripheral device data

c)

Software

Looking up and updating software data

4) Network a)

Network I/F (TCP/IP)

Looking up and updating network setting data for the diagnostic ultrasound system

b)

Modem

Looking up and updating modem setting data

c)

IV Setup

Looking up and updating InnerVision setting data

5) Configuration

Test Tools

Maintenance

a)

DICOM

Basic settings for DICOM communication

b)

Advanced DICOM

Advanced setting for DICOM communication

c)

File Transfer

Basic settings for data transfer

6) Service Config.

Looking up and updating user menu show/hide settings

1) Diagnostics

Diagnostic ultrasound system testing

2) HDD

Scandisk

3) DICOM

DICOM test

4) Probe*1

Transducer deterioration test

5) Front End*1

Performance check of the front-end analog sections

1) File Maintenance a)

User Data Backup/ Restore

b)

Backup and restoration of user-specific data

File Transfer

File transfer using an FTP

a)

Event Viewer

Log display using Event Viewer

b)

Original Log Viewer*1

Log display of service results

2) Log

3) Report

Viewing the information entered using the Communication user menu and sending responses.

4) Service History

Looking up and updating service history data

1-!2

Utilities



Short Cut

5) Performance*1

Display of phantom image acquisition procedure

1) Command Line

Execution of commands

2) Color Adjust*1

Color adjustment

3) Remote Application

Display of pop-up windows on the diagnostic ultrasound system

1) Smart Backup

Backup of user-specific data and preset data

2) Software Upgrade

Software version upgrade

*1: These functions will be supported in the future.

1-!3

*

2. 2.1

OUTLINE OF MENUS Outline of Points Common to the Maintenance and Service Menus (1)

Operation When the menu screen is opened, a full-screen dialog box for specifying the functions to be started up is displayed. When service mode is started up, the touch panel screen and the panel are placed in service mode until service mode is terminated. Usual panel operation and touch panel screen operation are disabled in service mode.

(2)

Restarting interrupted software If service mode is started when other software components are in the interrupted status, it may not be possible to restart the software after exiting service mode. The system may have to be turned OFF and restarted after servicing is completed.

2.2

Outline of the Maintenance Menu Operation of the maintenance menu is possible only at the diagnostic ultrasound systems. The menu is provided as a pop-up menu. The Image Quality Report function, however, is provided in the OTHER menu on the touch panel. (1)

Basic maintenance menu window The maintenance menu window is configured with the following basic format.

Cancel

!

Go

Exit

Each window has [Cancel], [Go], and [Exit] buttons. The buttons have the following functions: (a)

[Go]

: Executes the specified process.

(b)

[Cancel] : Stops the process in progress immediately. The process in progress is returned to the status before execution.

(c)

[Exit]

: Closes the window.

2-!1

(2)

Setting of whether menus are to be displayed It is possible to set whether or not the System Test, System Maintenance, Image Quality Report, and Communication menus are displayed. The settings are specified in Service Config in the service menu (refer to subsection 4.3.8). In the default status, the System Test and Communication menus are not displayed. These menus should not be displayed until the corresponding function is supported.

2.3

Outline of the Service Menu The service menu can be operated in two ways: by displaying it on the system monitor or by accessing it using a remote PC.
 


The service menu can be operated in two ways: operation at the system by displaying it on the system monitor or remote-access operation by displaying it on the InnerVision system.

2-!2

(1)

Service menu window Service menu main window Main Menu

Service Software Information Site

Test Tools System

Configration

HW/SW

Network

Service Config

Maintenance File Maintenance

HDD

DICOM

Stationing Application

Remote Application

Utilities

Log

Report

Service History

Short Cut Smart Backup

Diagnostic

Command Line

Service Exit Software Upgrade

Exit

! Service menu item window: When an item is selected, a window with the following configuration is displayed.

!

2-!3

2.4

Reboot Function The system is rebooted automatically in the following cases: (1)

When system operation becomes unstable

(2)

When the basic system settings are changed



2-!4

*

3. 3.1

MAINTENANCE MENU FUNCTIONS Activation and Termination (1)

Activation Press the MENU switch on the panel, select Maintenance in the pop-up menu, and then select the desired function in the submenu.

! (a)

Archiving, System Maintenance, Remote Access Management These functions are activated by selecting the corresponding function name in the Maintenance submenu.

(b)

Image Quality Report This function is activated by pressing [IQ Report] in the OTHER menu on the touch panel.

(2)

Termination To terminate a function, select Exit in the dialog box for the function. Refer to the explanation of termination processing for each function.

3-!1

3.2

Archiving (1)

Purpose (a)

Used to regularly back up the user-specific data on separate media as a measure against problems such as HDD failure.

(b)

The data to be backed up is listed below:

• User-defined measurement data (values defined using Application Preset Editor in Exam Type Edit in the System Setting pop-up menu) • User-defined system presets (preset values defined in General Preset, Peripheral System, OUTPUT, and Exam Information in the System Setting pop-up menu) and Imaging presets (imaging parameter information that can be saved for each Exam Type displayed by pressing the Preset switch on the touch panel) • Service-related presets (preset values defined in the service menu) Note: Removable media must be used to back up the user-specific data.
 The system identification data must be stored on the media. Otherwise, the media is ejected automatically and the data cannot be saved. Each system requires dedicated media. (2)

Display window

!

3-!2

No.

(3)

3.3

Item

Description

1.

Drive

• Used to specify the drive (external media) for making the backup copy.

2.

Refresh Drive List

• Used to obtain the drive connection data

3.

Initialize Media

• Used to write system identification data onto backup media. This process cannot be aborted. • This button is disabled during backup operation

4.

Progress Status Bar

• Indicates the progress of archiving.

5.

[Cancel]

• Aborts backup operation. The data backed up before the process was aborted cannot be restored (the directory name is changed).

6.

[Go]

• Starts backup operation.

7.

[Exit]

• Closes the window.

Brief description of the procedure (a)

Prepare (insert) the media for backup.

(b)

Select a drive.

(c)

Press [Initialize Media].

(d)

Press [Go] in the bottom right part of the window.

(e)

The backup process starts.

(f)

The backup process ends. When the backup data is saved, the media is ejected automatically from the drive.

(g)

Press [Exit] in the bottom right corner to close the window.

System Maintenance (1)

Purpose • Provides a template for the daily check conducted by the user. • Provides a hard copy of the check record and saves the check record on the HDD.

3-!3

(2)

Display window

! (a)

Check window No.

Item

Description

1.

Check items

• Items to be checked. • In the pull-down menu, select No Execution, Pass, or Fail. The default is No Execution for all items.

2.

Site and system data

• Displays the site and system data (data is displayed automatically).

3.

Date

• Test date (Displayed automatically. The date format normally used in the system is used.)

4.

Tester Name

• Enter the name of the person doing the test. (If the person responsible has already been specified, his/ her name appears automatically.)

5.

[Display]

• Displays a test pattern on the monitor.

6.

[Execute]

• Starts the Operation Panel test screen.

7.

[Cancel]

• Initializes the entered data.

8.

[Print]

• Prints out the test results and stores the data on the hard disk.

9.

[Exit]

• Closes the window.

3-!4

(b)

Monitor CRT test screen When the [Display] button is pressed, a test pattern is displayed on the monitor for approximately 5 seconds.

!

3-!5

(c)

Operation Panel test screen When the [Execute] button is pressed, the following window is displayed. The ROM and RAM tests are executed automatically, and Button Tests and other tests are performed interactively. The highlight moves to the item being executed. For each test, one of the four statuses (Checking, Pass, Failed, and Indistinct) is displayed. "Checking" appears while the test is being executed. "Pass" appears upon normal completion of a test. "Failed" appears upon abnormal termination of a test. "Indistinct" appears when a test is aborted. ROM and RAM checks cannot be aborted.

!

!

The following window is displayed at the top of the test screen for the switch test.

Clicking this button results in moving to the previous test. The status of the previous test is Indistinct. Note that this button is invalid in the case of "Button Test 1".

Clicking this button results in moving to the next test. The status of the next test is Indistinct.

Push the illuminated Button. >

The following window is displayed at the top of the test screen for the keyboard test.

!

Clicking this button results in returning to "Button Test 5".

Start keyboard test Type something. >

Invalid key

• ROM
 The ROMs in the circuit in the panel are checked automatically. • RAM
 The RAMs in the circuit in the panel are checked automatically. • Panel switch test
 The LEDs on the switches of the panel are lit in sequence and the peripheral circuits for the switches are checked when the switches corresponding to the lit LEDs are pressed. The peripheral circuits for the following switches are checked. a) Button Test1 ⎯⎯⎯⎯⎯⎯⎯ New Patient switch b) Button Test2 ⎯⎯⎯⎯⎯⎯⎯ TRANSDUCER switch c) Button Test3 ⎯⎯⎯⎯⎯⎯⎯ MENU switch d) Button Test4 ⎯⎯⎯⎯⎯⎯⎯ FREEZE switch e) Button Test5 ⎯⎯⎯⎯⎯⎯⎯ CDI switch • Keyboard
 The items entered from the keyboard are displayed in a dialog box to enable verification of normal keyboard entry. (3)

Check items No.

Check item name

Check item

Check method

1.

Power Supply Cable

• Area around the power outlet to which the power cable is connected

• Visual inspection

2.

Smell/Noise

• Abnormal noise or smell

• Checking for noise and smell

3.

Air Filter

• Air filter

• Visual inspection

4.

Fixture

• Securing of peripheral devices

• Visual inspection

5.

Caster

• Caster operation

• Visual inspection

6.

Probe

• Transducer

• Visual inspection (checking for abnormality in appearance)

7.

Peripheral Equipment

• Peripheral devices

• Visual inspection

8.

Image Quality

• Image quality

• Visual inspection (checking for image abnormalities before the service menu is started)

9.

Monitor CRT

• Monitor

• Displaying a test pattern

• Operating panel

• Checking entries made from the panel • Displaying the operation test instruction dialog by pressing the operation test button

10. Operation Panel

3-!7

(4)

3.4

Brief description of the procedure (a)

User check.

(b)

Enter the name of the person who conducted the test.

(c)

Save the test results on an HDD and also make a hard copy.

(d)

Press [Exit] in the bottom right corner to close the window.

Remote Access Management (1)

Purpose The user can permit the service company to perform remote maintenance on the system via a network or telephone line.

(2)

Display window

! No. 1. (3)

Item [Exit]

Description • Closes the window. Reboots the system.

Brief description of the procedures Press [Exit] in the bottom right corner of the window to close the window. The system is rebooted.

3-!8

3.5

Image Quality Report (1)

Purpose When an abnormality occurs for an image, this function saves the image and the system data.

(2)

Display window No window is provided. The operation status is indicated by the "in-service" display on the touch panel.

(3)

Brief description of the procedure No special operation is required. When a user identifies an abnormal image, the image can be saved in the following manner. Freeze the image ! Select OTHER menu (touch panel) ! Press [IQ Report] ! The contents displayed on the TV monitor are saved.

3-!9

*

4. 4.1

SERVICE MENU FUNCTIONS Activation (1)

License key for activation A license key based on a fixed Vendor Host ID is required for activation. Note: For on-site license key activation, the service engineer prepares the media (CD, MO, etc.) containing the license key. If such a media is unavailable, the TAC issues a license key that is valid for one day only.

(2)

Activation processes

Activation at the diagnostic ultrasound system (a)

Select [Service] in the pop-up menu to display the corresponding submenu.

(b)

Press "Ctrl + Shift" on the keyboard and press the right trackball button at the same time.

(c)

The license key entry window appears.

(d)

Enter the license key.

(e)

The service menu is displayed.

Activation from a network PC (a)

Give the user the following instructions over the telephone. Select [Service] in the pop-up menu and select Maintenance. Start Remote Access Management and permit remote access.

(b)

After remote access is permitted, open the license entry window by entering the address in the browser on the InnerVision system screen.

(c)

The license entry window is displayed on the InnerVision system screen.

(d)

Enter the license key.

(e)

The service menu is displayed on the InnerVision system screen.

Activation at the diagnostic ultrasound system •

Operations on the diagnostic ultrasound system are identical with those for sites with a service contract and InnerVision.

Activation from a network PC •

Remote access is not supported.

4-!1

Select the desired function from the menu displayed by selecting [Service] in the pop-up menu.

Enter the special key combination

Activate Remote Access Management

At the system

Check the service license.

License key entry window

Enter the license key.

Service menu main window

When it is acceptable Maintenance operations

Service menu main window

Request the customer by telephone to permit remote access.

When it is acceptable

Check the service license. Remote access Enter the license key.

Specify the address in the browser.

License key entry window

!

Activation act the diagnostic ultrasound system (a)

Set a media (CD, MO etc.) that is ready for on-site use into the corresponding drive of the diagnostic ultrasound system.

(b)

Select [Service] in the pop-up menu to display the corresponding submenu.

(c)

Press "Ctrl + Shift" on the keyboard and press the right trackball button at the same time.

(d)

The license key entry window appears. Press the "License Read" button.

Activation from a network PC •

Remote access is not possible.

4-!2

(3)

License key entry window: Enter the license key in this field. Asterisks are displayed for the entered characters.

License Read

7CD9CF68768956 Ok

Cancel

Set As Default

! No.

Items

Description

LM

RM

For sites with a service license (with a service contract and InnerVision or with a service contract only) 1.

Key entry

• Used to enter the license key (asterisks (*) are displayed for the lower-order 6 characters entered).





2.

[License read]

• Used to read the license key from an external media. • The media is ejected when the system enters service mode.



X

3.

[License read]

• The button is not active.

X



4.

[Ok]

• Used to proceed to the next step after the key is entered/read correctly. • If media is set in the drive, it is ejected.





5.

[Cancel]

• Used to close the window. The system does not enter service mode. • If media is set in the drive, it is ejected.



X

6.

[Cancel]

• Used to close the window.

X



7.

[Set As Default

• Check this box to fix the higher-order 14 characters.



X



-

For sites without a service license (without a service contract) 8.

[Entry window]

LM: Local mode RM: Remote mode

• The window is not displayed.

O: Effective
 X: Not effective

A license key is a string of 20 alphanumeric characters. The higher-order 14 characters of the key can be stored in the system by selecting "Set as Default" in the license key entry window. Once the higher-order 14 characters are in the system, the service menu can be started by entering only the lower-order 6 characters.

4-!3

(4)

Service menu window Main Menu

Service Software Information Site

Test Tools System

Configration

HW/SW

Network

Service Config

Maintenance File Maintenance

Log

Report

Short Cut

Command Line

Stationing Application

Remote Application

Exit

Termination (1)

DICOM

Service Exit Software Upgrade

! 4.2

HDD

Utilities

Service History

Smart Backup

Diagnostic

Normal termination Select [Exit] in the service menu window.

4-!4

4.3 4.3.1

Information Function Site (1)

Purpose • Looking up and updating site information (information about the location where the system is installed). Note: When the HDD is replaced, enter the serial number of the new HDD. The first character of the serial number must always be replaced with "3". Example

(2)

: Serial number shown on the label

Serial number to be entered

A4572001

34572001

Display window

!

4-!5

No.

Item

Description

LM

RM

1.

Site Name

• Used to look up the hospital name.





2.

Location

• Used to look up the hospital department in which the system is installed (including installation location).





3.

Serial Number

• Used to set and look up the system serial number.





4.

Administrator • Name


• Used to look up the person responsible for the system at the hospital. • Used to look up and update the customer management ID number.





• ID 5.

Address

• Used to look up the address data (Hospital, Street, Street No., Zip Code, Phone No., City, State/Region, and Country).





6.

Memo

• Used to enter the hospital data (name of the chief ultrasonographer etc.).





7.

[Save]

• Saves the entered data in the system.





8.

[Cancel]

• Abandons all the data entered in this window and overwrites it with the data stored in the system. Note: Canceling after saving data does not restore the data before saving.





9.

[Home]

• Returns the system to the service menu main window.





LM: Local mode RM: Remote mode (3)

O: Effective
 X: Not effective

Brief description of the procedure (a)

Enter the customer name and customer ID.

(b)

Enter the hospital data etc. in Memo field.

(c)

Press [Save] to save the entered data.

4-!6

4.3.2

System (1)

Purpose Looking up and updating system data.

(2)

Display window

!

4-!7

No.

Item

Description

LM RM

1.

Product Name

System series name (factory preset)





2.

Product Serial No.

System serial number (factory preset)





3.

System Use Time

• Displays the time of system use relative to a specified time period.













• Use Time

• The use time is displayed in hours, with the minutes rounded up.

• Set/ • Initializes or sets the use time (unit = Rese hour). t 4.

Probe Use Time

5.

• The system inspection schedule can be Inspection Schedule specified. (Note: This function will be supported in the future.) • Information message Display

• Date, Target date
 • Hours, Use Time

• Cumulative ON time for each transducer. • Select the transducer from the list box. The cumulative ON time for the selected transducer is displayed in hours (the minutes are rounded up). • The list box displays only transducers that have been connected to the system.

• When an inspection is due, it is possible to display a message indicating this when the power is turned ON. When this check box is checked, an entry window is displayed automatically. • The inspection schedule is specified in terms of date. Enter the date of the next inspection. • The inspection schedule is specified in terms of use time. Enter the use time after which the next inspection is to be performed.

• Every Power OnTime

• The message is displayed every time the power is turned ON when an inspection is overdue.

• Just One Time

• The message is displayed only the first time the power is turned ON after an inspection becomes overdue.

6.

Installed from

Date of system installation





7.

Installed until

Expiration date of lease etc. (This must be entered even for non-lease systems.)





8.

Maintenance contract

Presence/absence of maintenance contract





9.

InnerVision support

Presence/absence of InnerVision contract





Saves the entered data in a file.





10. [Save]

4-!8

11. [Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in the system. Note: Canceling after saving data does not restore the data before saving.





12. [Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!9

(3)

(4)

Brief description of the procedure (a)

Specify the inspection date.

(b)

Enter the lease expiration date.

(c)

Enter whether or not there is a maintenance contract.

(d)

Enter whether there is an InnerVision contract.

(e)

Press [Save] to save the entered data.

Message display function If an inspection is not performed by the date specified in the Inspection Schedule, a message is displayed each time the system is started up.

4.3.3 4.3.3.1

HW/SW Common I/F (1)

Purpose Looking up board information. It is possible to save board information that cannot be detected automatically.

(2)

Display window Information HW/SW Information Site System HW/SW Common IF Peripherals Software Option Install Network Configuration Service Config Test Tools Maintenace Utilities Short Cut

Home

No

Name

P-No/Revision No (Auto) P-No/BSM No (Manual)

Report

01

TI

unknown

Automatic

02

TX0

unknown

Automatic

03

TX1

unknown

Automatic

04

RX

unknown

Automatic

05

RC

unknown

Automatic

06

BE

unknown

Automatic

07

VP

unknown

Automatic

08

MA

unknown

Automatic

09

IO

unknown

Automati

10

BP

BP

Manual

11

CB

unknown

Automatic

Cancel

! Ready

4-!10

Save

No .

Item

LM

R M

1.

No.

I/F numbers





2.

Name

Names of hardware items in the system





3.

P-No./Revision No. (Auto)

Display of the following board information. • The P-list number and revision are displayed for the boards for which automatic detection is possible (TI, TX0, TX1, RX,RM, IO, RC, BE, VP, CB, and IP boards).





P-No./BSM No. (Manual)

• The revision and BSM number can be entered for the boards for which automatic detection is not possible (BP, Physio, and Panel boards).

4.

Report

Indicates whether automatic detection is possible or not. For the PWBs for which "Manual" is displayed, enter the data manually.





5.

[Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in the system. Note: Canceling after saving data does not restore the data before saving.





6.

[Save]

Saves the entered data on the HDD. The automatically detected data is not saved.



X

7.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Description

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Enter the data manually for PWBs for which "Manual" is displayed.

(b)

Press [Save] to save the entered data.

4-!11

4.3.3.2

Detail (1)

Purpose Looking up detailed board information. (This function is applicable only to boards for which automatic information detection is possible.)

(2)

Display window Information HW/SW Information Site System HW/SW Common IF Detail Peripherals Software Option Install Network Configuration Service Config Test Tools Maintenace Utilities Short Cut

!

Common I/F Detail

1 TI SELECT 1 TI 2 TX0 3 TX1 4 RX 5 RC 6 BE 7 VP 8 MA 9 IO 10 BP 11 CB 12 IP 11 CB 12 IP 13 Physio 14 Panel

Cancel

Ready

Save

No.

Item

Description

LM

R M

1.

Selection combo box

Displays the I/F numbers and names for selection.
 Selection is possible only for the boards for which "Auto" is displayed in the "Report" column in the Common I/F window.





2.

Common I/F Detail

Displays data, such as P-list numbers, that can be detected automatically.





3.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Home

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Select the desired PWB name from the pull-down menu.

(b)

The detailed information for the selected PWB is displayed.

4-!12

4.3.4

Peripherals (1)

Purpose Looking up and updating the peripheral device information.

(2)

Display window

! No.

Item

LM RM

1.

No.

Number





2.

Equipment Name/ Memo

Enter the data for the peripheral devices connected to the system.





3.

[Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in the system.





4.

[Save]

Saves the entered data on the HDD.





5.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Description

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Enter the data for the peripheral devices.

(b)

Press [Save] to save the entered data.

4-!13

4.3.5

Software (1)

Purpose Looking up software information

(2)

Display window Main Menu Information Site System HW/SW Common IF Peripherals Software Option Install Network Configuration Service Config Test Tools Maintenace Utilities Short Cut

Home Software Windows 2000

Service Pack 4

Middleware Version : VX02A Product Software Version : 1.00*R000 Service Software Version : 1.00*R000 Data Version : 1.00*R000 Option Software USHI-770A : Enable Option Software USDF-770A : Enable Option Software USFR-770A : Enable Option Software USDI-770A : Enable Option Software USDI-772B : Enable Option Software USSE-770A : Enable Option Software USAT-770A : Enable Option Software USIA-770A : Enable Option Software USPV-770A : Enable

!

Ready

No.

Item

Description

LM RM

1.

Software

Software name and version data. The following software data is displayed: • Middleware version • System software version • Service software version • Data version • Optional software





2.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!14

4.3.6 4.3.6.1

Network Network I/F (TCP/IP) (1)

Purpose Looking up the network configuration for the diagnostic ultrasound system.

(2)

Display window

! No.

Item

Description

LM

R M

1.

Adapter

Network hardware data stored in the system. The hardware can be selected from the combo box. When the hardware is selected, all the hardware data is updated accordingly.





2.

Obtain an IP address from DHCP server
 or
 Specify an IP address

Select one of the two options. If "Obtain..." is selected, "IP address", "Subnet Mask", and "Default Gateway" cannot be specified.





3.

[Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in the system.





4.

[Save]

Saves the settings.



X

5.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

Note: When it is necessary to configure the diagnostic ultrasound system for connection to the hospital network or to the InnerVision system, set the IP address and other necessary data according to the installation manual.

4-!15

4.3.6.2

Modem (TCP/IP) (1)

Purpose Looking up and updating the modem setting information.

(2)

Display window

! No.

Item

Description

LM

R M

1.

Modem

Modem name





2.

Com Port

COM port number





3.

[Cancel]

All the data entered in this window is abandoned and replaced with the data stored in the system.





4.

Allow remote TCP/ IP clients to access

The access range of the Windows2000 RAS server can be looked up and updated.





• this computer only • Entire network

• Limits access to this PC through RAS. • Expands the range of access to PCs on a LAN through RAS. ○



5.

Allow remote TCP/ • When this is selected, the PC-RAS server IP clients to request assigns an address to the client when the a predetermined system is connected to the RAS. address • When this is not selected, the server allows connection to the RAS with the address requested by the client. • TCP/IP address pool Begin/End

• Enter the IP addresses to be assigned when "Allow --" is selected.

6.

[Save]

Saves the entered data ("TCP/IP Address pool" only) in the system (system reboot required)..





7.

[Home]

Returns the system to the service menu main window.





4-!16

LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!17

4.3.6.3

IV-Setup (1)

Purpose Looking up and updating the InnerVision setup data. (This data is required for the use of InnerVision.)

(2)

Display window

!

4-!18

No.

Item

Description

LM

RM

1.

Connecting Device

Method of connecting the system to the TAC center. Exclusive selection.





2.

Dialup Router • IP Address • Configuration button

Valid only when "Dialup Router" is selected for Connecting Device. • IP address of the Config window that is supplied by the dialup router • Connects to the dialup router using IE. Thereafter, the setup program of the dialup router is used.





3.

Modem (TA)


Valid only when "Modem (TA)" is selected for Connecting Device. • Telephone book entry name to be used for making the call • Telephone number of the TAC center





Configuration for 
 TAC-Center • TAC IP Address • Data Receive Dir







• Data Send Dir


• Transmission directory (for the site) to be used at the TAC center • FTP user name to be used for sending data to the TAC center

• Phonebook Entry
 • TAC Phone Number 4.

• Ftp User Name

• IP address to be used by the TAC center • Reception directory for the concerned site in the TAC center

• Ftp User Password

• FTP password to be used for sending data to the TAC center

5.

[Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in a file.





6.

[Save]

Saves the entered data in the system.





7.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Enter the IV-Setup data.

(b)

Press [Save] to save the entered data.

4-!19

4.3.7

Configuration

4.3.7.1

Basic configuration for DICOM communication Whether DICOM configuration is possible depends on whether the required optional software, such as USDI-770A or USDI-772B, is installed. Refer to the table below. Title

USDI-770A not installed

USDI-770A installed

USDI-770A and 
 USDI-772B installed

local AEs

o

o

o

character set

o

o

o

Archive devices

o

o

o

Query/Retrieve

x

o

o

Network nodes

x

Hardcopy devices

x

x

o

hc overview

x

x

o

LUT files

o

o

o

HIS/RIS nodes

x

x

o

MPPS

x

x

o

Advanced DICOM

x

x

o

o (Setting is possible for Storage only.)

o (Setting is possible for Storage and Query/Retrieve.)

o: DICOM configuration is possible. x: DICOM configuration is not possible. • The figure below shows the relationship between the USDI-770A and USDI-772B. SSA-660A USDI-770A

Remote Activities

USDI-772B

PRINT SCP

PRINT SCU Storage SCP

Storage SCU

Storage SCU

Storage SCP

Storage commitment SCU

Storage commitment SCP

Q/R SCU

Q/R SCP

Storage SCP

Storage SCU

MWM SCU

MWM SCP

!

MPPS SCU

Set through the DICOM Menu

Set through the Advanced DICOM menu

4-!20

MPPS SCP

4-!21

4.3.7.1.1

Local AEs (Application Entity titles) (1)

Purpose HIS/RIS and Study Transfer for the diagnostic ultrasound system and the application entity title for Print are set. The port number is fixed to 104.

(2)

Display window

!

No.

Item

Description

LM

R M

1.

HIS/RIS

Setting of the application entity title for the Modality Worklist Management (MWM) SCU of the diagnostic ultrasound system. 
 This item is enabled only when the USDI-772B is installed.





2.

Study Transfer

Setting of the application entity title of the Study Transfer (Storage,Q/R) SCU of the diagnostic ultrasound system.





3.

Print

AE title of the Print SCU of the diagnostic ultrasound system.





4.

Port

The port number for communication is displayed.





5.

Host Name

The computer name set in the main system is displayed.





6.

[Save]

Saves the changes in the file (system reboot required).



X

7.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!22

(3)

Help screen

! (4)

Brief description of the procedure (a)

Enter each AE title. Note that the titles must be unique within the network.

(b)

Press [Save] to save the data.

4-!23

4.3.7.1.2

Character set (1)

Purpose Setting the default character set to be used for DICOM communication and for saving patient information.

(2)

Display window

! No. 1.

Item

Description

DICOM default character Default character set set • 1-byte characters are used. (Fixed) • Allow only one character set

LM

R M





• All multiple character sets

• Multi-byte characters are used.
 (Note: This will be supported in the future.)

2.

Character set

This item is used to specify an additional character set by selecting the corresponding character code. (Only 1byte character codes are supported.) However, this is fixed to ISO_IR 100.





3.

[Save]

Saves the entered character set (system reboot required).



X

4.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!24

(3)

Help screen

! (4)

Brief description of the procedure (a)

Select 1-byte characters or multi-byte characters. (Fixed to "Allows only one character set".)

(b)

Select the character code. (Fixed to ISO_IR 100)

(c)

Save the settings.

4-!25

4.3.7.1.3

Archive devices (1)

Purpose Setting the archive device to be used by the diagnostic ultrasound system. The standard is magneto-optical disk drive (MOD).

(2)

Display window

! No.

Check item

Description

LM

R M

1.

Select logical name

Select the logical name to be created in this window.





2.

Logical Name

Enter the logical name to be given to the archive device in this system.





3.

Drive Type

Select the type of drive connected to this system.





4.

Drive Letter

Select the drive letter corresponding to the selected drive type.





5.

[Save]

Saves the settings (system reboot required).





6.

[Delete]

Deletes the settings (system reboot required).



X

7.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!26

(3)

Help screen

! (4)

4.3.7.1.4 4.3.7.1.4.1

Brief description of the procedure (a)

Enter the logical name to be given to the archive device in this system.

(b)

Select the type of the archive device.

(c)

Select the drive letter for the archive device.

(d)

Save the settings.

Network nodes Host (1)

Purpose Setting the network information related to the DICOM transfer target. For connection via a network (LAN), set the host name and TCP/IP address. For connection via a modem (RAS), set the phone number, TCP/IP address, domain name, password, and login script of the remote host.
 


Note: Currently, only LAN is supported.

4-!27

(2)

Display window

!

4-!28

No.

Check item

Description

LM

RM

1.

Select Host

Selection of a host that has already been set. When a name is selected here, the corresponding host name and IP address that have already been set are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".





2.

Host Properties • Host Name

Setting of the host name • Searches for the specified computer name (if applicable) on the network or for the logical name assigned for convenience (valid only within the system).





Select the connection method (LAN or RAS). To use a LAN (fixed to LAN) To use a telephone circuit 
 (Note: This will be supported in the future.)





(Note: This will be supported in the future.)





• TCP/IP address 3.

Set Local Host

4.

connected by

When this box is checked, the NetBios computer name is displayed in the Host Name field. • LAN • RAS

5.

Data items required to use RAS • • • • • •

• Enter the IP address of the transfer target.

Phone No. PPP Login Script PPP Domain PPP Account PPP Password Verify PPP Password

• • • • • •

Phone number of the transfer target Login script, if applicable Domain name of the transfer target Login name (account) of the transfer target Password to connect to the transfer target Password to verify connection to the transfer target

6.

[Save]

Saves the entered data(system reboot required).



X

7.

[Delete]

Deletes the settings(system reboot required).



X

8.

[Find]

Searches for the specified host name on the network.





9.

[Test]

Searches for the specified TCP/IP address on the network.
 Executes the Ping test.





10.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!29

(3)

Help screen

! (4)

Brief description of the procedure (a)

Enter the logical name (computer name) of the DICOM transfer target.

(b)

Select the connection method.

(c)

For LAN connection, enter the IP address only. For RAS connection, enter the data items required to use RAS. (Only LAN connection is supported.)

(d)

Save the entered data.

(e)

Press [Test] to confirm the DICOM transfer target exists on the network.

4-!30

4.3.7.1.4.2

Remote (1)

Purpose Setting the AE data (title and port number) and DICOM service information of the DICOM transfer target. The image transfer syntax can be selected. If the USDI-770A only is installed, the DICOM transfer target is remote. If the 
 USDI-772B is also installed, the transfer target for Storage is Advanced DICOM and the transfer target for Query/Retrieve is remote.

(2)

Display window Main Menu

!

Information Site System HW/SW Network Configuration DICOM local AEs character set archive devices network nodes hosts remote Query and Retrieve hardcopy devices hc overview LUT files HIS/RIS nodes MPPS Advanced DICOM File Transfer Service Config Test Tools Maintenace

Home Select Logical Name General logi

Logical Name

logi

Host Selection

qqq

Application Entity Properties Select AE Title

define new aeae

AE Title

aeae

Port Number

1234

Verification

Supported Entity Properties Image Tranfer (Store) Byte Order Implicit Little Endian Explicit Little Endian

Structure Report Properties Standard

Ready

Private Save

4-!31

Delete

No.

Item

Description

LM

RM

1.

Select Logical Name

Selection of a logical name that has already been set. 
 When a name is selected here, the information for the corresponding remote system that has already been set is displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".





2.

Logical Name

Setting for the logical name of the remote system (The name set here is displayed on the system.)





3.

Host Selection

Selection of the host of the remote system (The host that is set in the DICOM-network nodes-host page should be selected.)
 For Storage, be sure to select the host name of this system.
 For Q/R, select the host name of the communication partner.





4.

Default Node

Selection of the default node (Do not check this box.)





5.

Select AE Title

Selection of an AE title that has already been set. When an AE title is selected here, the AE title and port No. that have already been set for and correspond to the logical name are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. If no settings have been entered yet, only "define new" is displayed.





6.

AE Title

Setting of the AE title.





7.

Port Number

Setting of the port number.
 For Storage, the port number is fixed to 2000.
 For Q/R, set the port number of the communication partner.





8.

Add/Del

Button for adding/deleting the AE title, port No., and Image Transfer information that correspond to the logical name (Do not use this button.)



X

9.

Verification

Button for executing verification for the remote system.



X

10.

Image Transfer (Store)

Setting of the service class support for Image Transfer (Store)





11.

Byte Order/Compression

Setting for the Transfer Syntax
 (Only Implicit VR Little Endian and Explicit VR Little Endian of Byte Order are supported. Explicit Big Endian is not supported and must not be set.) 
 If data compression is necessary, set it in Advanced DICOM. (Do not select Compression on this page.)





12.

Structure Properties

Selection for transfer of the Structure Report data Standard: Check this box to transfer the data using the SR standard protocol (default). Private: Check this box to transfer the data using the Toshiba protocol.





Report


LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective 4-!32

No.

Item

Description

LM

RM

13.

Save

Button for saving the settings (system reboot required)



X

14.

Delete

Button for deleting the settings (system reboot required)



X

15.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Help screen

!

4-!33

4.3.7.1.5 4.2.7.1.5.1

Query/Retrieve Host (1)

Purpose Setting the DICOM Query/Retrieve host.

(2)

Display window

! No.

Item

Description

LM

R M

The list of the host names stored using this window is displayed in the pull-down menu.









1.

Select Host

2.

Host Properties • Host Name • TCP/ IPaddress

3.

Set Local Host

When this box is checked, the NetBios computer name is displayed in the Host Name field.





4.

connected Host

Select the connection method (LAN or RAS).





• Enter the logical name of the communication partner on the network (computer name) or the logical name assigned for convenience (effective only in this system). • Enter the IP address of the communication partner on the network.

• To use a LAN (fixed to LAN at present) • To use a telephone line (Note: This will be supported in the future.).

4-!34

No. 5.

Item Data items required to use RAS

Description (Note: This will be supported in the future.)

• Phone NO.

• Phone number of the communication partner

• PPP Login Script

• Login script, if applicable

• PPP Domain

• Domain name of the communication partner

• PPP Account

• Login name (account) of the communication partner

• PPP Password

• Password to connect to the communication partner

• Verify PPP Password

• Password to verify connection to the communication partner

LM

RM





6.

[Save]

Saves the settings (system reboot required).



X

7.

[Delete]

Deletes the settings (system reboot required).



X

8.

[Find]

Searches for the specified computer name 
 (if applicable) on the network.





9.

[Test]

Searches for the specified TCP/IP address on the network.





10.

[Home]

Returns the system to the service menu main window.





4-!35

4.3.7.1.5.2

Remote (1)

Purpose Setting the DICOM Query/Retrieve partner and setting the detailed data required to communicate with the host specified in subsection 4.3.7.1.5.1.

(2)

Display window

! No.

Item

1.

Select Logical Name

2.

General

3.

4.

LM

R M

The list of the logical names of the communication partners stored using this window is displayed in the pull-down menu.

















• Logical

• Enter the logical name given for convenience (effective only in this system).

• Host Selection

• Select the host name specified in subsection 4.3.7.1.5.1

Application

Supported

Description

Entity

• Properties • Select AE Title

• The list of the AE titles of the communication partners stored using this window is displayed in the pulldown menu.

• AE Title

• Enter the AE title of the communication partner.

• Port Number

• Enter the port number used by the AE title.

DICOM • services

• Check this box to perform image transfer.

• Image Transfer (store) • Byte Order

• Select the DICOM data format to be used in image transfer.

4-!36

5.

[Save]

Saves the settings (system reboot required).



X

6.

[Delete]

Deletes the settings (system reboot required).



X

7.

[Home]

Returns the system to the service menu main window.





4-!37

4.3.7.1.6 4.3.7.1.6.1

Hardcopy devices Host (1)

Purpose Setting the network information related to hardcopy devices (imager and/or printer). For connection via a network (LAN), set the host name and TCP/IP address. For connection via a modem (RAS), set the phone number, TCP/IP address, domain name, password, and login script of the remote host.
 


Note: Currently, only LAN is supported. (2)

Display window

!

4-!38

No.

Item

1.

Select Host

2.

Host Properties • Host Name

Description

L M

R M

Selection of a host that has already been set. When a name is selected here, the corresponding host name and IP address that have already been set are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".









• Enter the logical name (computer name) of the hardcopy device selected in "Select Host" field or the logical name assigned for convenience (valid only within the system). Up to 14 1-byte characters can be entered.

• TCP/IP Address

• Enter the IP address of the hardcopy device selected in "Select Host" field.

3.

Set Local Host

When this box is checked, the NetBios computer name is displayed in the Host Name field.





4.

connected by

Select the connection method (LAN or RAS).









• LA N • RA S 5.

Data items required to use RAS

• To use a LAN (fixed to LAN) • To use a telephone circuit
 (Note: This will be supported in the future.)

(Note: This will be supported in the future.)

• Phone No.

• Phone number of the transfer target

• PPP Login Script • PPP Domain

• Login script, if applicable • Domain name of the transfer target

• PPP Account • PPP Password • Verify PPP Password

• Login name (account) of the transfer target • Password to connect to the transfer target • Password to verify connection to the transfer target

6.

[Save]

Saves the entered data(system reboot required).



X

7.

[Delete]

Deletes the settings(system reboot required).



X

8.

[Find]

Searches for the specified computer name (if applicable) on the network.





4-!39

9.

[Test]

10. [Home]

LM: RM:

Local mode Remote mode

Checks whether the system can connect to the address set for "TCP/IP address" on the network. Executes the Ping test.
 When "RAS" is selected for "connected by", nothing is performed.





Returns the system to the service menu main window.





O: Effective
 X: Not effective

4-!40

(3)

Help screen

!

4-!41

4.3.7.1.6.2

Remote (1)

Purpose Setting the AE data (title and port number) of the hardcopy device (imager and/or printer).

(2)

Display window

! No.

Check item

Description

1.

Select Logical Name

2.

General

LM: RM:

Selection of a logical name that has already been set. When a name is selected here, the information for the corresponding remote system that has already been set is displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".

• Logical Name

• Setting of the logical name of the remote system. The same name must be used in Logical Name, AE Title, and Hardcopy Device. Up to 20 1-byte characters can be entered.

• Host Selection

• Select the printer host name set in "Host" (in "Hardcopy devices").

Local mode Remote mode

O: Effective
 X: Not effective

4-!42

LM

R M









No. 3.

Item

Description

Application Entity Properties Select AE Title

Selection of an AE title that has already been set. When an AE title is selected here, the AE title and port No. that have already been set for and correspond to the logical name are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. If no settings have been entered yet, only "define new" is displayed.

• AE Title

• Enter the AE title of the transfer target Up to 16 1-byte characters can be entered

• Port Number

• Enter the port number used by the AE title.

LM

RM





4.

[Verification]

Executes verification of the specified host.





5.

[Add]/[Del]

Button for adding/deleting the AE title and port No. that correspond to the logical name



X

6.

[Save]

Stores the data (system reboot required).



X

7.

[Delete]

Deletes the stored logical name (system reboot required).



X

8.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Help screen

!

4-!43

4.3.7.1.6.3

Details (1)

Purpose Setting individual parameters supported by the hardcopy devices (imager and/or printer), such as the film type and film destination.

(2)

Display window

!

4-!44

No.

Item

Description

LM

R M

1.

Select HC device

Selection of the HC drive name that has already been set. When a name is selected here, the information for the corresponding remote system that has already been set is displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".





2.

HC Device


Setting of the HC device name of the remote system (The name set here is displayed on the system.) The same name must be used in Logical Name, AE Title, and Hard copy Device. Check the box to set the entered hardcopy device as the default. (This function is not currently supported.)














 


Default Camera

3.

General Settings • Type

• Select the type of printer.
 (Normally, select DICOM Printer.)

• Class

• Specify the class of the cameraspecific filming properties and the hardcopy device. (Refer to table 1 for the supported hardcopy devices.)

• DICO M Node

• Enter the logical name set in "Remote" (in "Hardcopy devices").

• Locatio • Enter a comment describing the n location. • Comm ent 4.

5.

• Enter a comment if necessary.

Filming Properties • Hold printed film jobs

• Select the number of jobs that can be held.

• Pixel Size 
 [1/1000 mm]

• Enter the pixel size.

• Film Sheet Formats

• Select the film sheet format.

• Film Sheet Format

• Define a new film sheet format.

• Number of Pixel

• Define the number of pixels (resolution).

• Medium type

• Select the film type to be used.

• Film destination

• Select the film supply method.

• Transformation

• Select the interpolation method to be used for output to the film

[Add]/[Del]

Adds or deletes a film sheet format.

4-!45

6.

[Save]

Saves the settings (system reboot required).



X

7.

[Delete]

Deletes the settings (system reboot required).



X

8.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!46

Table 1 List of supported hardcopy devices • • • • • • • • ) • • • • • • • • •

(3)

agfadrystar2000Color agfadrystar2000 agfadrystar3000 agfalr2100 agfalr3300 agfalr5200 codonicsnp1660(1600) codonicsnp1660Mono(1600 FujiDRYPIX1000 Fujifl_dp_3543 FujiFNPS551 imation8100 imation8300 imation8500 imation8700 imation959 imation969HQHQS

Help screen

!

4-!47

• • • • • • • • • • • • • • • •

Kodak1120 Kodak190 Kodak2180 KonicaDRYPRO722 KonicaLi10A KonicaLi1417DH KonicaLi21 KonicaLi62P KonicaLi7Li7ALi7DDLi8 Seiko1720DColor Seiko1720DMono SonyUPAD3Color SonyUPAD3Mono Sterling1417 Steringinkjet400 Steringlp300400

4.3.7.1.7

HC overview (1)

Purpose Setting the default hardcopy device (imager and/or printer).

(2)

Display window

! No.

Item

Description

LM

R M

1.

Select logical name

Select the HC device name from among the HC device names.





2.

Name

The HC device name set in "Details" (in "Hardcopy devices") is displayed.





3.

Type

The printer type set in "Details" (in "Hardcopy devices") is displayed.





4.

Comment

The comment entered in "Details" (in "Hardcopy devices") is displayed.





5.

Location

The location entered in "Details" (in "Hardcopy devices") is displayed.





6.

Hold printed film jobs

The number of film jobs set in "Details" (in "Hardcopy devices") is displayed.





7.

Default Camera

If this box is checked, it indicates that the printer is set as default in "Details" (in "Hardcopy devices"). (This will be supported in the future.)





8.

Network access [Test]

Checks whether the corresponding hardcopy device is connected on network. Executes the Ping test.





9.

DICOM access [Test]

Checks the AE access. 
 Executes the DICOM Validation test.





10.

[Save]

Saves the settings (system reboot required).



X

11.

[Home]

Returns the system to the service menu main window.





4-!48

LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!49

(3)

Help screen

!

4-!50

4.3.7.1.8 4.3.7.1.8.1

LUT files Import (1)

Purpose Adding or removing a look-up table (LUT).

(2)

Display window

! No. 1.

Check item

Description

Import LUT 
 Source

Used to import an external LUT.

• drive

• Drive in which the desired LUT data is stored

• path

• Directory in which the desired LUT data is stored

Destination • LUT directory

LM

R M







X



X





• Import destination of the external LUT data

[copy] Executes import. 2.

Remove LUT • LUT directory

• System LUT data directory to be removed

• LUT file

• LUT data file name

[remove] Executes removal. 3.

LM: RM:

[Home]

Local mode Remote mode

Returns the system to the service menu main window. O: Effective
 X: Not effective

4-!51

(3)

Help screen

!

4-!52

4.3.7.1.8.2

Set (1)

Purpose Selecting a LUT and setting the maximum density.

(2)

Display window

! No.

Check item

Description

LM

R M

1.

Image type

Specify the LUT data type and density for each image.





2.

[Save]

Saves the settings (system reboot required).



X

3.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!53

(3)

Help screen

!

4-!54

4.3.7.2

HIS/RIS nodes This function is enabled only when the USDI-772B is installed.

4.3.7.2.1

Host (1)

Purpose Setting the network information of the host (the station from which the examination information is obtained). For network connection (LAN), set the host name and TCP/IP address.
 Modem connection is not supported.

(2)

Display window

!

4-!55

No.

Item

1.

Select Host

2.

Host Properties • Host Name

• TCP/IP address 3.

Set Local Host

4.

Connected by

LM

R M

Selection of a host that has already been set. When a name is selected here, the corresponding host name and IP address that have already been set are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".





















Enter the logical name (computer name) of the transfer target, or the logical name given for convenience (valid only within the system). Enter the IP address of the transfer target. When this box is checked, the NetBios computer name is displayed in the Host Name field.





5.

Description

LA N

To use a LAN (fixed to LAN) To use a telephone circuit
 (Note: This will be supported in the future.)

RA S

Data items required to use RAS

(Note: This will be supported in the future.)

• Phone No.

Phone number of the transfer target

• PPP Login Script

Login script, if applicable

• PPP Domain

Domain name of the transfer target

• PPP Account

Login name for connection to the transfer target

• PPP Password

Password for connection to the transfer target

• Verify PPP Password

Password to verify connection to the transfer target

6.

[Save]

Saves the settings (system reboot required).



X

7.

[Delete]

Deletes the settings.



X

8.

[Find]

Searches for the specified computer name (if applicable) on the network.





9.

[Test]

Searches for the specified TCP/IP address on the network.
 Executes the Ping test.





10.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective 4-!56

4-!57

(3)

Help screen

! (4)

Brief description of the procedure (a)

Enter the logical name (computer name) of the station with which the system communicates through the network.

(b)

Select the connection method.

(c)

For LAN connections, set the IP address only. For RAS connections, set the items required to use RAS. (Currently, only LAN connection is supported.)

(d)

Save the settings.

(e)

Press [Test] to confirm that the communication partner exists on the network.

4-!58

4.3.7.2.2

Remote (1)

Purpose Setting the DICOM transfer target and setting the detailed data required to communicate with the HIS/RIS station specified in "Host" in the previous subsection.

(2)

Display window

!

4-!59

No.

Item

1.

Select Logical Name

2.

General

3.

4.

Description

LM

R M

Selection of a logical name that has already been set. When a name is selected here, the information for the corresponding remote system that has already been set is displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. To add new settings, select "define new".













Setting of the MWM method (active MWM or passive MWM)
 Check the box for Basic Worklist only.





• Logical Name

Enter the logical name.

• Host Selection

Select the HIS/RIS station specified in "Host" in the previous subsection.

Application Entity Properties • Select AE Title

Selection of an AE title that has already been set. When an AE title is selected here, the AE title and port No. that have already been set for and correspond to the logical name are displayed. When these settings are to be changed, change them and then click the Save button to save the new settings. If no settings have been entered yet, only "define new" is displayed.

• AE Title

Enter the AE title of the transfer target.

• Port Number

Enter the port number used by the AE title.

Supported DICOM services • Basic Worklist

5.

[Add]/[Del]

Adds or deletes the AE title, port number, and Supported DICOM services corresponding to the selected logical name.



X

6.

[Verification]

Executes DICOM Verification (C-ECHO).





7.

[Save]

Saves the settings. This button becomes effective when any of above items is set (system reboot required).



X

8.

[Delete]

Deletes the stored data. This button becomes effective when Logical Name is selected (system reboot required).



X

9.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!60

(3)

Help screen

!

4-!61

4.3.7.2.3

HIS/RIS (1)

Purpose Selecting the AE title of the host (the station from which the examination information is obtained), and setting the timing when the information is obtained from the Basic Worklist SCP.

(2)

Display window

!

4-!62

No. 1.

Item

Description

Application Entity Title Selection • Basic Worklist SCP

2.

Select a predefined worklist.

LM

R M









Select the AE title (that supports Basic Worklist) to be actually connected from among the AE titles set for Basic Worklist in the HIS/RIS nodes/remote page. If no AE title is set in the HIS/RIS nodes/remote page, the message "no Application Entity selected" is displayed.

Basic Worklist SCP Query Properties • Query Waiting Time

Enter the wait time between the query for the scheduled examinations and the query response.

• Max Query Match Number

Enter the maximum number of scheduled examinations to be obtained (maximum 200).

• Query Interval

Enter the polling interval for automatic periodic execution of MWM. (Note: This function will be supported in the future.)

3.

[Save]

Saves the settings (system reboot required).



X

4.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!63

4.3.7.2.4

MPPS

4.2.7.2.4.1 Host (1)

Purpose Setting the information of the MPPS station.

(2)

Display window

! No.

Item

Description

1.

Select Host

The list of the host names stored using this window is displayed in the pull-down menu.

2.

Host Properties • Host Name

• TCP/IP address 3.

Set Local Host

4.

Connected by

Enter the logical name of the communication partner on the network (computer name) or the logical name assigned for convenience (effective only in this system)..

LM

R M

















Enter the IP address of the communication partner on the network. When this box is checked, the NetBios computer name is displayed in the Host Name field. •



LA N

To use a LAN. To use a telephone line (Note: This will be supported in the future.)

RA S

4-!64

No. 5.

Item Data items required to use RAS

Description (Note: This will be supported in the future.)

• Phone No.

Phone number of the communication partner

• PPP Login Script

Login script, if applicable

• PPP Domain

Domain name of the communication partner

• PPP Account

Login name (account) of the communication partner

• PPP Password

Password to connect to the communication partner

• Verify PPP Password

LM

RM





Password to verify connection to the communication partner

6.

[Save]

The list of the host names stored using this window is displayed in the pull-down menu.





7.

[Delete]

Saves the settings (system reboot required).



X

8.

[Find]

Deletes the settings (system reboot required).



X

9.

[Test]

Searches for the specified TCP/IP address on the network.



X

10.

[Home]

Returns the system to the service menu main window.



X

4-!65

4.3.7.2.4.2 Remote (1)

Purpose Setting the DICOM communication partner and setting the detailed data required to communicate with the MPPS specified in subsection 4.3.7.2.4.1.

(2)

Display window

! No.

Item

1.

General • Logical Name

Description

Enter the logical name given for convenience (effective only in this system).

LM

R M









• Host Selection Enter the HIS/RIS name specified in subsection 4.3.7.2.4.1. 2.

Application Entity The list of the AE titles of the Properties • Select AE Title communication partners stored using this window is displayed in the pull-down menu. • AE Title

Enter the AE title of the communication partner.

• Port Number

Enter the port number used by the AE title.

3.

[Save]

Saves the settings (system reboot required). This is enabled when at least one data is set.



X

4.

[Delete]

Deletes the settings (system reboot required). This is enabled when the data for Logical Name is set.



X

5.

[Home]

Returns the system to the service menu main window.





4-!66

4.3.7.3

Advanced DICOM

4.3.7.3.1

Main Configuration (1)

Purpose Setting the AE titles, port numbers, etc. of Advanced DICOM and Basic DICOM in order to connect them to each other.

(2)

Display window

!

4-!67

No.

Item

Description

LM

R M

1.

AE Title of Advanced DICOM

Enter the AE title of Advanced DICOM.
 Set the AE title specified in the Study Transfer field in the window displayed by selecting DICOM ! local AEs (subsection 4.3.7.1.1).





2.

Advanced DICOM SCP Port Number

Port number of Advanced DICOM. This is fixed to 2000.





3.

Basic DICOM Host Name

Enter the Xario host name.





4.

Basic DICOM Storage SCP Port Number

TCP/IP port number listened to by Advanced DICOM in the Local Protocol function used between Advanced DICOM and Basic DICOM. This is fixed to 3000..





5.

Basic DICOM Storage SCO Port Number

TCP/IP port number listened to by Basic DICOM in the Local Protocol function used between Advanced DICOM and Basic DICOM. This is fixed to 5000.





6.

Remaining Free Space on Hard Disk (M)

Remaining available space. This is fixed to 1000 MB.





7.

CD-R Driver

Drive name for CD-R. This is fixed to G.





8.

MO Driver

Drive name for MO. This is fixed to F.





9.

Cache 1

Drive name for cache for CD writing. This is fixed to U.





10.

Cache 2

Drive name for the USB HD.





11.

Enable Removable Media

Be sure to check this box.





12.

[Save]

Saves the settings.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!68

4.3.7.3.2

DICOM Destination (1)

Purpose For setting Basic DICOM Storage SCP and Remote (Storage, Storage Commitment, MPPS).

(2)

Display window

!

4-!69

No.

Item

Description

LM

RM

1.

Number of Configured Destination

The total number of destinations that are currently set is displayed.





2.

Select AE Title

Select the AE title of the destination.





3.

AE Title

Enter the AE title of the destination. Up to 16 characters can be entered using alphanumeric characters and underscores (_).





4.

Host

Enter the host name of the destination. Set the host name specified in the window displayed by selecting DICOM ! network nodes ! host (subsection 4.3.7.1.4.1).





5.

Port

Enter the port number of the destination.





6.

Alias

Enter an alias (an alternative name for the destination or a comment that identifies the destination).





7.

Packet Size

Enter the packet size.





8.

[First]

Jumps to the top of the list.





9.

[Prev]

Goes to the previous list.





10.

[Next]

Goes to the next list.





11.

[Last]

Jumps to the bottom of the list.





12.

[Add]

Adds a new destination.





13.

[Delete]

Deletes a destination.





14.

[Save]

Saves the settings.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!70

4.3.7.3.3

Archives (1)

Purpose For setting Basic DICOM Storage SCP and Remote Storage SCP.

(2)

Display window

!

4-!71

No.

Item

Description

LM

RM

1.

Number of Archives

The total number of currently set Storage destinations is displayed.





2.

Number of Retries

Set the number of retries for Storage and MPPS processing.





3.

Commitment association 
 timeout (sec)

Set the Storage Commitment timeout value.





4.

Archive Number

Set the list number of the Storage partner.





5.

Called AE Title

Select the AE title of the destination from the AE title list set on the DICOM Destinations page.





6.

Description

Enter a comment.





7.

AE Title

Enter the AE title of the destination. Set the AE title of the Storage destination specified in the window displayed by selecting DICOM ! network nodes ! remote (subsection 4.3.7.1.4.2).





8.

Compression

Select the compression method from among:





• • • • • •

As it YBR_FULL YBR_FULL + RLE YBR_FULL_422 + JPEG RLE YBR_PARTIAL_422 + JPEG

9.

JPEG Compression Ratio

Set the compression ratio for JPEG compression. This field is enabled only when JPEG compression is selected for "Compression Type".





10.

Timeout (sec)

Enter the timeout value for the association response from the destination.





11.

Storage Commitment

Check this box when Storage Commitment is executed automatically after Storage processing.





12.

USS = >Retired US

Check this box when US Image is converted to Retired US Image.





13.

Additional Script

Select the data conversion script. Normally, select "Blank.scr".





14.

[Prev]

Goes to the previous list.





15.

[Next]

Goes to the next list.





16.

[Add]

Adds a new destination.





17.

[Delete]

Deletes a destination.





18.

[Save]

Saves the settings.





19.

[Test Association]

Executes verification (C-ECHO).





This should be performed for the set destination. LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!72

4.3.7.3.4

MPPS (1)

Purpose Setting the AE title of the destination to which the examination start and end notifications are sent.

(2)

Display window

! No.

Item

Description

LM

R M

1.

Enable Modality Performed Procedure Step (MPPS)

Selection of whether MPPS is set to effective or not effective. Check this box to make MPPS effective.





2.

AE Title of MPPS

Selection of the MPPS connection destination Select one from the AE title list set in the DICOM Destination page.





3.

[Save]

Saves the settings.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!73

4.3.7.3.5

Logging (1)

Purpose Setting the method for acquiring the DICOM-related logs.

(2)

Display window

! No.

Item

Description

LM

R M

1.

DICOM Logging

Log setting for Merge.log Checked : All operations are logged. Not checked : Only the operations in which an error occurs are logged.





2.

General Logging

Log setting for other logs Checked : All operations are logged. Not checked : Only the operations in which an error occurs are logged.





3.

Maximum Size [KB]

Specify the maximum size of the log file.





4.

Save

Saves the settings.





5.

Download Log Files

The downloaded log files are listed in this area. When each hyperlink is clicked, the corresponding log window is opened and the contents of the file are displayed.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!74

4.3.7.3.6

Import/Export (1)

Purpose This procedure should not be performed.

(2)

Display window

! No.

Item

Description

LM

RM

1.

IMPORT TO APLIO

Enter the file name to be imported (include the extension).





2.

Browse...

Used to display a list of file names.





3.

File name at Aplio

Enter the file name to be saved in Xario (include the extension).





4.

Copy to Aplio

Used to import the file specified in item 3 above.





5.

EXPORT FROM APLIO

The files registered in Xario are listed in this area.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!75

4.3.7.4

File Transfer Basic setup for offline data transfer

4.3.7.4.1

Targets (1)

Purpose Setting the network data of a host that supports data transfer to the system. (This setting is required to transmit/receive files via a LAN off line.)

(2)

Display window

!

4-!76

No.

Check item

Description

1.

Select Host

The list of the hosts stored using this window is displayed in the pull-down menu.

2.

Host Properties • Host Name

• Enter the logical name (computer name) of the transfer target, or the logical name given for convenience (effective only in this system).

• TCP/IP address 3.

Set Local Host

4.

connected by

















• Enter the IP address of the transfer target.

Select the connection method (LAN or RAS). LAN • To use a LAN RAS • To use a telephone circuit (Note: This will be supported in the future.)

Data items required to use RAS • • • • • •

RM

When this box is checked, the NetBios computer name is displayed in the Host Name field. • •

5.

LM

Phone No. PPP Login Script PPP Domain PPP Account PPP Password Verify PPP Password

(Note: This will be supported in the future.) • • • • • •

Phone number of the transfer target Login script, if applicable Domain name of the transfer target Login name (account) of the transfer target Password to connect to the transfer target Password to verify connection to the transfer target

6.

[Save]

Saves the settings (system reboot required).



X

7.

[Delete]

Deletes the saved logical name (system reboot required).



X

8.

[Find]

Searches for the specified computer name (if applicable) on the network.





9.

[Test]

Searches for the specified TCP/IP address on the network.





10.

[Home]

Returns the system to the service menu main window. LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!77

(3)

Help screen

! (4)

Brief description of the procedure (a)

Enter the host name to be registered.

(b)

Enter the IP address.

(c)

Select "RAS" to connect the host via RAS (telephone circuit).

(d)

When RAS is to be used, enter the required data.

(e)

Press [SAVE] to save the entered data. The newly defined host is registered to the Select Host pull-down menu.

4-!78

4.3.7.4.2

Directories (1)

Purpose Setting details of the DICOM transfer host. (This setting is required to transmit/receive files via a LAN off line.)

(2)

Display window

! No.

Check item

Description

LM

R M

1.

Select Logical Name

The list of the logical names of the transfer targets stored using this window is displayed in the pull-down menu.





2.

Logical Name

Enter the logical name given for convenience (effective only in this system).





3.

FTP Host Selection

Select the host name specified on the "Targets" page of "File Transfer".





4.

[Save]

Stores the entered data (system reboot required).



X

5.

[Delete]

Deletes the stored data (system reboot required).



X

6.

[Home]

Returns the system to the service menu main window.

LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!79

(3)

Help screen

! (4)

Brief description of the procedure (a)

Enter the logical name to be registered in the Logical Name field.

(b)

Select the host name to be associated from the FTP Host Selection pull-down menu.

(c)

Press the [Save] button to save the settings in the system.

(d)

The newly defined logical name is registered in the Select Logical Name pull-down menu.

4-!80

4.3.8

Service Config. (1)

Purpose Looking up and updating the user menu item display settings and automatic transmission settings.

(2)

Display window

! No. 1.

Check item

Description

Service • Represent ative

LM RM ○







Looking up and updating the name of the nearest service representative.

• Phone No. Looking up and updating the phone number of the nearest service representative. 2.

System Test


LM: RM:

Note: Do not remove the check mark for "Hide" until the System Test function is supported. Hide • Check the box for "Hide" to hide System Test in the maintenance menu. Setting of automatic transmission for System Test in maintenance menu

• No send

• To not perform automatic transmission

• Send event only

• To transmit the event only

• Send with full attachment

• To transmit the results

Local mode Remote mode

O: Effective
 X: Not effective

4-!81

No. 3.

4.

Item

Description

System Maintenance

Hide • Check the box for "Hide" to hide System Maintenance in the maintenance menu. Setting of automatic transmission for System Maintenance in the maintenance menu

• No send

• To not perform automatic transmission

• Send event only

• To transmit the event only

• Send with full attachment

• To transmit the results

Image Quality Report

LM

RM













Hide • Check the box for 'Hide" to hide Image Quality Report in the maintenance menu. Setting of automatic transmission for Image Quality Report in the maintenance menu

5.

• No send

• To not perform automatic transmission

• Send event only

• To transmit the event only

• Send with full attachment

• To transmit the results

Communication


Note: Do not remove the check mark for "Hide" until the Communication function is supported. Hide • Check the box for "Hide" to hide Communication in the maintenance menu. Setting of automatic transmission for Communication in the maintenance menu

• No send

• To not perform automatic transmission

• Send/Receive available

• Transmission and reception

• Receive only

• Reception only

6.

[Cancel]

Abandons all the data entered in this window and overwrites it with the data stored in the system or in a file.





7.

[Save]

Saves the entered data in the system.





8.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Specify the maintenance menu show/hide setting.

(b)

Specify the automatic transmission condition for the maintenance menu.

(c)

Press [Save] to save the settings.

4-!82

4.4 4.4.1

Test Tools Function Diagnostics (1)

Purpose Performing tests on diagnostic ultrasound systems. Note: To perform automatic tests, press the Freeze switch to stop transmission and disconnect all of the transducers from the system before entering the service menu.

CAUTION:

Disconnect the transducers only after stopping transmission by pressing the Freeze button. Otherwise, the transducers may be damaged.

When the automatic tests are completed, terminate the service menu and shut down the system. It is recommended that the manual tests be executed first and then the automatic tests. The manual tests cannot be started after the automatic tests. If each board has to be tested in more detail, perform testing according to the procedure described in section 8 "STARTING THE BOARD TESTS". (2)

Display window (a)

Test screen Test Tool Diagnostic Information Test Tools Diagnostic HDD DICOM Probe Front End Maintenace Utilities Short Cut

Home Date

2004/04/01 15:01

Engineer Automatic TI

Manual Pass

Fail

Monitor/VCR

TX

Execute

Panel/TCS

RX

IO (RM)

RC

Speaker

BE

MIC

VP

USB RS232C

CB

Ether

IP Physio

!

All Select

Times New

Ready

4-!83

Save

No.

Item

Description

LM

R M

1.

Date

The current date is displayed automatically. Previous test execution dates are displayed in a pull-down menu.





2.

Engineer

Enter the name of the engineer executing the tests.





3.

Automatic tests • Check boxes







X

4.

• Select the tests to be executed.

• Result display area

• The test result for each unit is indicated as Pass or Fail.

• [All Select]

• Selects all the test items.

• Test count

• Any value from 1 to 100 can be entered for the number of executions. The default is 1.

• [Execute]

• Executes the automatic tests for each PWB and unit the specified number of times (system reboot required).

• [Execute and disconnect line]

• Executes the board test (even if the system is disconnected from the network) and sends the results to the appropriate folder.

• [Cancel]

• Cancels execution of tests after completion of the test in progress.

• Information

• Instructions, such as those for disconnecting the transducer, are displayed.

Manual tests • Pass and Fail check boxes

• Perform visual checks and enter the results. Some test results are checked automatically.

• [Execute]

• Displays supporting data for test execution. Voice is output.

5.

[New]

• Acquires the current time.





6.

[Save]

• Saves the test results in the HDD. Up to 64 files can be saved. • When the number of files exceeds 64, the oldest file is deleted automatically.





7.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!84

(b)

Panel/TCS test screen When the [Execute] button is pressed, the screen below is displayed and the PANEL/ TCS test is started.

!

4-!85

No.

Item

Description

1.

ROM

The ROMs in the circuit in the panel are checked automatically.

2.

RAM

The RAMs in the circuit in the panel are checked automatically.

3.

Button Test 1-5

The LEDs on the switches of the panel are lit in sequence and the peripheral circuits for the switches are checked when the switches corresponding to the lit LEDs are pressed. The peripheral circuits for the following switches are checked. • Button Test1-----------New Patient switch • Button Test2-----------TRANSDUCER switch • Button Test3-----------MENU switch • Button Test4-----------FREEZE switch • Button Test5-----------CDI switch

4.

Track Ball Test

Move the cursor on the screen using the trackball and check the coordinates.
 The result should be entered by the engineer.

5.

STC Test
 STC 1-8

Move each STC sliding VR from the minimum to maximum and check whether the value varies from 0 to 255.
 The result should be entered by the engineer.

6.

Vol Test
 Vol 1-6

Rotate the VR knobs from the minimum to maximum and check whether the value varies from 0 to 255. 
 The result should be entered by the engineer. • Vol1 ---------------------PCG POSI VR knob • Vol2 ---------------------ECG POSI VR knob • Vol3 ---------------------AUDIO BALANCE VR knob • Vol4 ---------------------PCG GAIN VR knob • Vol5 ---------------------ECG GAIN VR knob • Vol6 ---------------------AUDIO volume knob

7.

TCS Test

Press the switches displayed on the touch panel screen and check the corresponding responses.
 The test result is checked automatically.

8.

Full Key Board Test

Press the keys of the keyboard and check whether the corresponding characters are displayed on the screen.
 The result should be entered by the engineer.

4-!86

(3)

Other tests Monitor/VCR : Check whether the images are recorded and played back correctly, and enter the result.

(4)

Speaker

: Check whether sound is output from the speaker and enter the result.

MIC

: Check whether the voice input from the microphone is recorded and played back correctly by the VCR, and enter the result.

USB

: Connect a USB mouse and check whether it operates correctly, and enter the result.

Brief description of the procedure [Automatic tests] (a)

Confirm that all of the transducers are disconnected.

(b)

Select the board to be tested.

(c)

Enter the test counts.

(d)

Press [Execute] to start the tests. Note: The test process may take several minutes for some boards. Pay close attention to the message at the bottom-left of the screen.

(e)

Press [Save] to save the test results.

(f)

When the tests are completed, press [Home] to return to the service menu screen, and then press [Exit] to close the service menu.
 On the screen displayed after the service menu is closed, select Shutdown and then turn the power OFF.

[Manual tests] (a)

Press [Execute] to start the tests or perform the designated checks.

(b)

Enter the test results in the check boxes.

(c)

Press [Save] to save the test results.

4-!87

4.4.2

HDD (1)

Purpose Executing Scandisk and displaying past scandisk results.

(2)

Display window

! No.

Item

LM RM

1.

Date

The current date is displayed automatically.
 Previous test execution dates are displayed in a pull-down menu.
 Select a past execution date to display the HDD test results for that day.





2.

Engineer

Enter the name of the engineer executing the tests.





3.

Output

Display the results of scandisk.





4.

[Cancel]

Registers a new test execution date without saving the test results.





5.

[Save]

Saves the test results.





6.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Description

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Select the drive to be tested.

(b)

Press [Execute].

(c)

The test is started and the result is displayed in the text box.

(d)

Press [Save] to save the test results.

4-!88

4.4.3

DICOM The following tests are executed according to the type of DICOM server. (a)

Verify Network: The network settings are tested.

(b)

Storage: The file and image transmission system is tested. (This function will be supported in the future.)

(c)

Basic print: The print settings are tested. (This function will be supported in the future.)

(1)

Purpose

Testing the connection with the remote host (executing ping).

(2)

Testing the connection with the remote DICOM AE-Title (C-ECHO). Display window (a)

Display window for Verify Network

! No.

Item

Description

LM RM

1.

Host Name

Select the server to be tested.





2.

Logical Name

Select the logical name associated to the host name.





3.

AET

Select the AE-Title to be tested.





4.

[GO]

Starts the test.



X

5.

[Abort]

Aborts the test.



X

6.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!89

(b)

Display window for Storage

! No.

Item

Description

LM RM

1.

Extend Trace Level

The trace test is performed for the DICOM Storage application. The Transfer Syntax pull-down menu displays All, Implicit Little Endian (ILE), Explicit Little Endian (ELE), Explicit Big, Endian (EBE), JPEG Lossless (JLL), and JPEG Lossy (JLO).





2.

Send Images

Image transfer to DICOM Storage SCP is tested. The Transfer Syntax pull-down menu displays All, Implicit Little Endian (ILE), Explicit Little Endian (ELE), Explicit Big, Endian (EBE), JPEG Lossless (JLL), and JPEG Lossy (JLO). The selectable items are , , , , and .





3.

Sim. Rem. Receiver

DICOM image reception is tested. and can be selected. can be specified.





4.

Sim. Loc. Receiver

Images from the specified sender are received. and can be selected. can be specified.





5.

Add. Loc. Receiver

Additional Storage SCP is started. can be specified.





6.

[GO]

Starts the test.



X

7.

[Abort]

Aborts the test



X

8.

[Home]

Returns the system to the service menu main window.





4-!90

LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!91

(c)

Display window for Basic Print

! No.

Item

Description

LM RM

1.

Extend Trace Level

The DICOM Basic Printer is started in trace mode.





2.

Simulate Camera

The DICOM Reception Basic Printer simulator is started.





3.

[GO]

Starts the test.



X

4.

[Abort]

Aborts the test.



X

5.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!92

4.5

Maintenance

4.5.1 4.5.1.1

File Maintenance User Data Backup/Restore (1)

Purpose Backing up user-specific data. This function enables backing up (data saving and restoring) of the various files created and updated by user operation. This function allows the user-specific data to be recovered when the data file is lost due to an incorrect operation or a problem with the magnetic disk. The following three types of data can be backed up. MCR

: User-defined measurement data (values set in Application Preset Editor in Exam Type Edit in the System Setting pop-up menu)

Preset

: User-defined system presets (the preset values set in General Preset, Peripheral System, OUTPUT, and Exam Information) and imaging presets (imaging parameter information that can be saved for each Exam Type by pressing the Preset switch on the touch panel)

Service Preset : Presets related to servicing (the values set using the service menu) Note:

Removable media must be used to back up the user-specific data.
 


The system identification data must be stored on the media. Otherwise, the media is ejected automatically and the data cannot be saved. Each system requires dedicated media. (2)

Display window

!

4-!93

No.

Item

1.

Description

Backup/Restore • Ba cku p

LM

R M





• Data is transferred from the system to media (including a remote HDD).

• Re • Data is transferred from media stor (including a remote HDD) to the e system. 2.

Drive & Dir

Specify the media (including a remote HDD)and directory to be used for backup or restoration.





3.

[New Folder]

Creates a new directory





4.

[Refresh Drive List]

Initializes (clears) the area (specified in the Drive&Dir field) to which the backup data is to be copied.





5.

[Initialize Media (Folder)]

The medium for backup is logically formatted using system identification data. A medium (directory) not formatted with this function cannot be used for backup.





6.

[Eject MO]

Ejects the medium from the MO drive.





7.

[Cancel]

Before [Go] is selected: The selections are returned to the initial status. During execution (after [Go] is selected): Backup/restoration is aborted. Copying of data is aborted after the current copy operation is completed.





8.

[Go]

Executes Backup/Restore operation.





9.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Select Backup or Restore. (Only the procedure for backup is given below.)

(b)

In the Drive&Dir field, select the medium or directory to which the data is to be backed up. To create a directory, use [Make Dir].

(c)

Initialize the directory using [Initialize (Partition)].

(d)

Select the type of data to be backed up from among MCR, Preset, and ServicePreset.

(e)

Press [Go] to execute the backup operation. When backup is completed successfully, a directory is created in the directory specified above, with the date and time of backup of the data.

4-!94

4.5.1.2

File Transfer (1)

Purpose Transferring files concerning various data items for the system using FTP. 
 The following "sources" are provided.

(2)

(a)

Image

: To transfer images from the system

(b)

File

: To transfer files from the system

(c)

Download List : To view the list of files transferred from the remote device to the system

(d)

Extract List

: To view the list of files in the transfer directory in the system

Display window (a)

Display window for Image

!

4-!95

No. 1.

Item

Description

LM

RM





Action • Transfer Online

• Specify the patient name, study, and series to display thumbnail images. The displayed images are transferred to a remote terminal.

• Extract Only

• Specify the patient name, study, and series to display thumbnail images. The specified file is copied to the transfer directory.

• Display Header

• Specify the patient name, study, and series to display thumbnail images.

2.

Patient

The list of patient numbers is displayed in the pulldown menu.





3.

Study

The list of study numbers is displayed in the pulldown menu.





4.

Series

The list of series numbers is displayed in the pulldown menu.





5.

Selected Images

The list of selected images is displayed.





6.

Target

Select the transfer destination remote terminal.





7.

Show Thumb Nails

Select whether or not thumbnail display is to be enabled.





8.

[Delete]

Deletes the selected images.





9.

[Reset]

Initializes image selection.





10.

[Go]

Transfers the selected images.





LM: RM: (b)

Local mode Remote mode

O: Effective
 X: Not effective

Display window for File The menu screen differs depending on the selected action. No.

LM: RM:

Item

Description

L M

R M

1.

Transfer Online

Specify the patient name, study, and series for thumbnail image display. The displayed images can be transferred directly to a remote terminal (operating terminal).

-

-

2.

Extract Only

Specify the patient name, study, and series for thumbnail image display. The files are copied to the transfer directory.

-

-

3.

Download

The specified files are downloaded from a remote device to the system.

-

-

Local mode
 Remote mode

4-!96

(b)-1 Menu screen for Transfer Online

! No. 1.

Item

Description

File Type • Any file

LM

R M





• Any of the files on the system disk are transferred to a remote terminal.

• Most common logfiles (Note: This will be supported in the future.) • Service related logfiles (Note: This will be supported in the future.) • All available logfiles (Note: This will be supported in the future.) 2.

Filter

File types are filtered.





3.

[Refresh]

Refreshes the display of the directory information.





4.

[Go]

Starts the file transfer operation.





5.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!97

(b)-2 Menu screen for Download

! No. 1.

Item File Type •

LM: RM:

Description

L M

R M





Any file • Any of the files on a remote terminal are downloaded to the system.

2.

Source

Select the remote terminal.





3.

[Refresh]

Refreshes the display of the download directory information.





4.

[Go]

Starts the download operation.





5.

[Home]

Returns the system to the service menu main window.





Local mode Remote mode

O: Effective
 X: Not effective

4-!98

(b)-3 Menu screen for Extract Only

! No. 1.

Item

Description

File Type • Any file

L M

R M





• Any of the files on the system disk are copied to the transfer directory.

• Most common logfiles (Note: This will be supported in the future.) • Service related logfiles (Note: This will be supported in the future.) • All available logfiles (Note: This will be supported in the future.)

LM: RM:

2.

[Refresh]

Refreshes the display of the directory information.





3.

[Go]

Copies the file to the transfer directory.





4.

[Home]

Returns the system to the service menu main window.





Local mode Remote mode

O: Effective
 X: Not effective

4-!99

(c)

Display window for Download List

! No.

LM: RM:

Item

Description

L M

R M





1.

Action

2.

Site Directory

The location of the directory in the system is displayed.





3.

Filter

File types are filtered.





4.

[Refresh]

Refreshes the display of the directory where the downloaded files are to be stored.





5.

[Delete]

Deletes the downloaded files.





6.

[Home]

Returns the system to the service menu main window.





Local mode Remote mode

• Show/ The list of files is displayed. Or the files Delete are deleted.

O: Effective
 X: Not effective

4-!100

(d)

Display window for Extract List

! No. 1.

LM: RM:

Item

Description

Action • Transfer Online

• The files in the transfer directory are transferred.

• Show/ Delete

• The list of files in the transfer directory is displayed. Or the files in the transfer directory are deleted.

L M

R M





2.

Filter

File types are filtered.





3.

[Refresh]

Refreshes the display of the transfer directory information.





4.

[Go]

Executes the file transfer operation.





5.

[Home]

Returns the system to the service menu main window.





Local mode Remote mode

O: Effective
 X: Not effective

4-!101

4.5.2 4.5.2.1

Log Original Log Viewer (1)

Purpose Enables the results of maintenance operations, such as System Test, Diagnostic, System Maintenance, and Image Quality Report in the maintenance menu, to be viewed.

(2)

Display window (a)

Initial window

!

4-!102

(b)

System Test Log window

! (c)

Diagnostic Log window

!

4-!103

(d)

System Maintenance Log window

! (e)

Image Quality Report Log window

!

4-!104

(f)

Operation Log window

!

4-!105

4.5.2.2

Event Viewer (1)

Purpose Viewing the event log through simplified operation.

(2)

Display window

!

4-!106

No. 1.

Item

Description

Domain • Application • Security • System

2.

Severity • • • •

Error Warning Information All

LM

RM

Select the type of log. • Application log • Security log • System log





Select the severity (error level). • Errors are displayed. • Warnings are displayed. • Information is displayed. • All levels of messages are displayed.





3.

Searchorder • Newest first • Oldest first

Select the display order. • The newest data is displayed first. • The oldest data is displayed first.





4.

Message Limit

Specify the maximum number of messages to be displayed.





5.

Time Range

Specify the date range for display of the messages. • Start date • End date





• From • To 6.

Time

Check this box to narrow down the range to particular hours on the specified date.





7.

[Exit]

Closes the window.





8.

[Abort]

Aborts the search operation.





9.

[Go]

Starts the search operation.





LM: RM: (3)

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Select the log type.

(b)

Select the desired message in the search result list.

(c)

Details are displayed for the selected message.

4-!107

4.5.3

Report (1)

Purpose Enables electronic data from users and service engineers at the user site to be viewed.

(2)

Display window

! No.

Item

Description

LM RM

1.

Date/Time

Date and time of data creation





2.

Subject/Directory

Data title





3.

From

Name of writer





4.

Category

Data category





5.

Subject

Title given by responding service engineer





6.

Name

Name of responding service engineer





7.

[Cancel]

All the data entered in this window is abandoned.





8.

[Save]

The entered data is transferred to the system at the user site and saved there. The corresponding data that was used previously is deleted.





9.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!108

4.5.4

Service History (1)

Purpose Creating the service history.

(2)

Display window

! No.

Item

Description

LM RM

1.

Engineer

Enter name of the engineer in charge of service.





2.

Date

If "New" is specified, the current date is automatically read and displayed. If a past date is specified, the data corresponding to the specified date is displayed.





3.

Type of Maintenance

Select either proactive maintenance or reactive maintenance.





4.

Contents

Check the boxes for the target service items.





5.

Frequency of Phenomenon

Select the frequency of recurrence of the problem.





6.

Current Condition

Select the servicing status.





7.

Working time

Enter the actual time required for the work.





8.

Memo

Enter the work contents and comments.





9.

[Cancel]

If "New" is specified, all the data entered in the window is deleted.





10.

[New]

Used to enter the data for the work performed on the current date and to initialize other data.





11.

[Save]

Transfers and saves the entered data to the user system.





12.

[Home]

Returns the system to the service menu main window.





4-!109

LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective 


4-!110

(3)

4.6 4.6.1

Brief description of the procedures

(a)

Enter the name of the service engineer.

(b)

Check the boxes in the Maintenance Work section according to the current servicing work.

(c)

Press [Save] to save the data in the Aplio system

Utilities Command Line (1)

Purpose Executing a command line (DOS command) and displaying results of command execution.

(2)

Display window

! No.

Item

Description

LM RM

1.

Command Line

Enter the command to be executed.





2.

Output

The command execution results are displayed.





3.

[Explorer]

Displays Explorer.



X

4.

[Control Panel]

Displays the Control Panel.



X

5.

[MO Eject]

Ejects the MO from the drive.





6.

[Go]

Executes the entered command.





7.

[Abort]

Aborts the command being executed.





8.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!111

4.6.2

Stationing application (1)

Purpose Enables the α protocol and the card reader to be turned ON/OFF.

(2)

Display window

! No.



1.

Item

Description

LM RM

Alpha Protocol

Specify whether α protocol is to be used or not.





Card Reader

Specify whether the card reader is to be used or not.





3.

[Cancel]

Initializes the settings.





4.

[SAVE]

Saves the settings made in this window.





5.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-101!A

*

4.6.3

Remote Application (1)

Purpose Opening a window on the diagnostic ultrasound system to display messages.

(2)

Display window

! No.

Item

LM RM

1.

Output

Enter the message to be displayed on the diagnostic ultrasound system.





2.

[Cancel]

Clears the entered message.





3.

[Display]

Displays the message on the diagnostic ultrasound system.





4.

[Home]

Returns the system to the service menu main window.





LM: RM: (3)

Description

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedures (a)

Enter the message to be displayed on the diagnostic ultrasound system.

(b)

Press [Display].

(c)

The entered message is displayed on the diagnostic ultrasound system.

4-!102

4.7 4.7.1

Shortcuts Smart Backup (1)

Purpose Backing up the service information through simplified operation.

(2)

Display window

!

4-!103

No.

Item

LM RM

1.

Backup/Restore

Backup : System to media (including remote HDD) Restore: Media (including remote HDD) to system





2.

Drive & Dir

Specify the media (including remote HDD) and directory to be used for backup or restore.





3.

[New Folder]

Creates a new directory.





4.

[Refresh Drive List]

Initializes (clears) the area (specified in the Drive&Dir field) to which the backup data is to be copied.





5.

[Initialize Media (Folder)]

The medium for backup is logically formatted using system identification data. A medium (directory) that is not formatted with this function cannot be used for backup.





6.

[Eject MO]

Ejects the medium from the MO drive.





7.

[Cancel]

Before [Go] is selected: The selections are returned to the initial status. During execution (after [Go] is selected): Backup /Restore is aborted. Copying of data is aborted after the current copy operation is completed.





8.

[Go]

Executes the Backup /Restore operation.





LM: RM: (3)

Description

Local mode Remote mode

O: Effective
 X: Not effective

Brief description of the procedure (a)

Select Backup.

(b)

In the Drive&Dir field, select the medium or directory to which the data is to be backed up. To create a directory, use [Make Dir].

(c)

Initialize the directory using [Initialize (Partition)].

(d)

Select the type of data to be backed up from among MCR, Preset, and ServicePreset.

(e)

Press [Go] to execute the backup operation. When backup is completed successfully, a directory is created in the directory specified above, with the date and time of backup of the data.

4-!104

4.7.2

Software Upgrade (1)

Purpose The software upgrade procedure is displayed and the automatic login process is terminated.

(2)

Display window

! No.

Item

Description

LM RM

1.

Software Upgrade

The procedure for upgrading the software is displayed.





2.

[Go]

The automatic login process is terminated (system reboot required).





3.

[Home]

Returns the system to the service menu main window.





LM: RM:

Local mode Remote mode

O: Effective
 X: Not effective

4-!105