Technical Modifications DD+DIS015.01E Section 11 List of Contents 1 Replacing the new rear cover 3 and the top cover

Views 163 Downloads 5 File size 57KB

Report DMCA / Copyright

DOWNLOAD FILE

Recommend stories

  • Author / Uploaded
  • IM
Citation preview

Technical Modifications

DD+DIS015.01E

Section 11

List of Contents 1

Replacing the new rear cover 3 and the top cover.................................... 1

1.1

Cause CURIX 60......................................................................................................1

1.2

Introduction as production standard ....................................................................1

1.3

Cause CP 1000 ........................................................................................................1

1.4

Introduction as production standard ....................................................................1

2

Modification kit for the strain relief ............................................................ 2

2.1

Cause Curix 60 / CP 1000 .......................................................................................2

2.2

Introduction as production standard ....................................................................2

3

Replacement of the cage cover .................................................................. 2

3.1

Cause CP 1000 ........................................................................................................2

3.2

Introduction as production standard ....................................................................2

4

Replacement of the valves of the replenisher bottles .............................. 3

4.1

Cause.......................................................................................................................3

4.2

Introduction as production standard ....................................................................3

5

Pump mounting with latches, introduced as production standard ......... 3

5.1

Cause.......................................................................................................................3

Edition 2, Revision 0

CURIX 60 / CP 1000 (Type 9462)

Section 11 / I

Technical Modifications

DD+DIS015.01E

1

Replacing the new rear cover 3 and the top cover

1.1

Cause CURIX 60 While the processor is developing films, undeveloped films which are placed on the dryer hood are fogged via the IR heaters. Error symptoms see section 6.3.

1.2

Introduction as production standard Cover 3 Top cover

1.3

CM+9.9463.1214.0 CM+9.9462.1185.2 Machine type

As of SN

9462/105

2219

9462/155

1031

9462/165

1056

Cause CP 1000 While the processor is developing films, undeveloped films which are placed on the dryer hood are fogged via the IR heaters. Error symptoms see section 6.3.

1.4

Introduction as production standard Top cover

Edition 2, Revision 0

CM+9.9462.1185.2 Machine type

As of SN

9462/205

2108

9462/285

1072

9462/305

1311

9462/845

2364

CURIX 60 / CP 1000 (Type 9462)

Section 11 / 1

Technical Modifications

DD+DIS015.01E

2

Modification kit for the strain relief

2.1

Cause Curix 60 / CP 1000 Due to spilling from the machine tanks, and also in case of careless handling in the area of the tank heater/machine tank, the chemicals may run along the cables and reach the Control Board where they can probably cause a short circuit.

2.2

Introduction as production standard Machine type

As of SN

9462/105

2179

9462/155

1026

9462/165

1056

9462/205

2059

9462/285

1072

9462/305

1311

9462/845

2176

3

Replacement of the cage cover

3.1

Cause CP 1000 When handling full machine tanks above the dryer, the chemicals may spill on the Control Board.

3.2

Introduction as production standard Closed cage cover CM+9.9463.1471.0

Section 11 / 2

Machine type

As of SN

9462/205

2108

9462/285

1072

9462/305

1311

9462/845

2364

CURIX 60 / CP 1000 (Type 9462)

Edition 2, Revision 0

Technical Modifications

DD+DIS015.01E

4

Replacement of the valves of the replenisher bottles

4.1

Cause The material of the valves has been affected by photochemicals.

4.2

Introduction as production standard Valve, complete (developer) CM+9.9432.1070.5 Valve, complete (fixer) CM+9.9432.1110.5 Valve, complete (water) CM+9.9432.1090.5 Machine type

As of SN

9462/105

2701

9462/155

-

9462/165

-

9462/205

2108

9462/285

1072

9462/305

1311

9462//845

2394

5

Pump mounting with latches, introduced as production standard

5.1

Cause The customers do not really accept that they have to open screws in order to clean the pumps. Developer pump Fixer pump Water pump

Edition 2, Revision 0

CM+9.9462.9110.0 CM+9.9462.9120.0 CM+9.9462.9130.0

Machine type

As of SN

9462/105

1881

9462/155

1006

9462/165

1036

9462/205

1803

9462/285

1054

9462/305

1131

9462/845

1716

CURIX 60 / CP 1000 (Type 9462)

Section 11 / 3