Technical Modifications DD+DIS015.01E Section 11 List of Contents 1 Replacing the new rear cover 3 and the top cover
Views 163 Downloads 5 File size 57KB
Technical Modifications
DD+DIS015.01E
Section 11
List of Contents 1
Replacing the new rear cover 3 and the top cover.................................... 1
1.1
Cause CURIX 60......................................................................................................1
1.2
Introduction as production standard ....................................................................1
1.3
Cause CP 1000 ........................................................................................................1
1.4
Introduction as production standard ....................................................................1
2
Modification kit for the strain relief ............................................................ 2
2.1
Cause Curix 60 / CP 1000 .......................................................................................2
2.2
Introduction as production standard ....................................................................2
3
Replacement of the cage cover .................................................................. 2
3.1
Cause CP 1000 ........................................................................................................2
3.2
Introduction as production standard ....................................................................2
4
Replacement of the valves of the replenisher bottles .............................. 3
4.1
Cause.......................................................................................................................3
4.2
Introduction as production standard ....................................................................3
5
Pump mounting with latches, introduced as production standard ......... 3
5.1
Cause.......................................................................................................................3
Edition 2, Revision 0
CURIX 60 / CP 1000 (Type 9462)
Section 11 / I
Technical Modifications
DD+DIS015.01E
1
Replacing the new rear cover 3 and the top cover
1.1
Cause CURIX 60 While the processor is developing films, undeveloped films which are placed on the dryer hood are fogged via the IR heaters. Error symptoms see section 6.3.
1.2
Introduction as production standard Cover 3 Top cover
1.3
CM+9.9463.1214.0 CM+9.9462.1185.2 Machine type
As of SN
9462/105
2219
9462/155
1031
9462/165
1056
Cause CP 1000 While the processor is developing films, undeveloped films which are placed on the dryer hood are fogged via the IR heaters. Error symptoms see section 6.3.
1.4
Introduction as production standard Top cover
Edition 2, Revision 0
CM+9.9462.1185.2 Machine type
As of SN
9462/205
2108
9462/285
1072
9462/305
1311
9462/845
2364
CURIX 60 / CP 1000 (Type 9462)
Section 11 / 1
Technical Modifications
DD+DIS015.01E
2
Modification kit for the strain relief
2.1
Cause Curix 60 / CP 1000 Due to spilling from the machine tanks, and also in case of careless handling in the area of the tank heater/machine tank, the chemicals may run along the cables and reach the Control Board where they can probably cause a short circuit.
2.2
Introduction as production standard Machine type
As of SN
9462/105
2179
9462/155
1026
9462/165
1056
9462/205
2059
9462/285
1072
9462/305
1311
9462/845
2176
3
Replacement of the cage cover
3.1
Cause CP 1000 When handling full machine tanks above the dryer, the chemicals may spill on the Control Board.
3.2
Introduction as production standard Closed cage cover CM+9.9463.1471.0
Section 11 / 2
Machine type
As of SN
9462/205
2108
9462/285
1072
9462/305
1311
9462/845
2364
CURIX 60 / CP 1000 (Type 9462)
Edition 2, Revision 0
Technical Modifications
DD+DIS015.01E
4
Replacement of the valves of the replenisher bottles
4.1
Cause The material of the valves has been affected by photochemicals.
4.2
Introduction as production standard Valve, complete (developer) CM+9.9432.1070.5 Valve, complete (fixer) CM+9.9432.1110.5 Valve, complete (water) CM+9.9432.1090.5 Machine type
As of SN
9462/105
2701
9462/155
-
9462/165
-
9462/205
2108
9462/285
1072
9462/305
1311
9462//845
2394
5
Pump mounting with latches, introduced as production standard
5.1
Cause The customers do not really accept that they have to open screws in order to clean the pumps. Developer pump Fixer pump Water pump
Edition 2, Revision 0
CM+9.9462.9110.0 CM+9.9462.9120.0 CM+9.9462.9130.0
Machine type
As of SN
9462/105
1881
9462/155
1006
9462/165
1036
9462/205
1803
9462/285
1054
9462/305
1131
9462/845
1716
CURIX 60 / CP 1000 (Type 9462)
Section 11 / 3