Medidor Coriolis

INSTRUMENTACION INDUSTRIAL: MEDIDORES MASICOS Ing. CARLOS E. COTRINO B. M. Sc. Fuerza de Coriolis Cor · i · o · lis (k

Views 172 Downloads 6 File size 1MB

Report DMCA / Copyright

DOWNLOAD FILE

Recommend stories

Citation preview

INSTRUMENTACION INDUSTRIAL: MEDIDORES MASICOS Ing. CARLOS E. COTRINO B. M. Sc.

Fuerza de Coriolis Cor · i · o · lis (k ô r´ë-ö´lis). Una fuerza ficticia usada para describir matemáticamente un movimiento, como el de un avión o una formación de nubes, relativo a un marco de referencia no inercial con movimiento rotatorio uniforme, como la tierra. [ Gaspard G. de Coriolis (1792-1843), Matemático francés.]

C.COTRINO-NOV-2011

2

Page 1

Fuerza de Coriolis3 CORIOLIS FORCE = 2 x BALL MASS x BALL VELOCITY x PLATFORM ANGULAR VELOCITY

C.COTRINO-NOV-2011

3

Fuerza de Coriolis3 Fuerza Coriolis = 2 . m . V x ω

C.COTRINO-NOV-2011

4

Page 2

Principio de operación1 Teoría de operación: Midpoint Measuring Tube with flow

Measuring Tube without flow

Fc

Fc

Flow w

Direction of Oscillation

Sensors A & B

5

C.COTRINO-NOV-2011

Principio de operación1

w

v

Fc Coriolis Force

= –2 • m • [ Mass

ω

×

Angular Speed of Rotation

C.COTRINO-NOV-2011

v

]

Velocity

6

Page 3

Principio de operación1 Sensor A

Mass Flow = 0 Sensor B

Sensor A

Sensor B

Sensor A

Mass Flow > 0 Sensor B

7

C.COTRINO-NOV-2011

Principio de operación4 



C.COTRINO-NOV-2011

Un tubo lleva un magneto y el otro lleva una bobina colectora o pickup. Las ondas seno generadas representan el movimiento de un tubo respecto al otro.

8

Page 4

Principio de operación4  



Los tubos oscilan con una diferencia de fase de 180°. Los “pick ups” interno y externo generan ondas seno continuamente. Cuando el tubo esta vacío no se genera diferencia de fase

9

C.COTRINO-NOV-2011

Principio de operación4 



C.COTRINO-NOV-2011

Cuando pasa fluido a través de los tubos se generan fuerzas de Coriolis. Cuando el tubo tiende a subir durante el semiciclo correspondiente el fluido tiende a impulsarlo hacia abajo

10

Page 5

Principio de operación4 





Esta diferencia de momentum angular produce una torsión del tubo La torsión del tubo hace que las señales generadas por los “pickups” estén fuera de fase. La diferencia de fase se mide por un ΔT

C.COTRINO-NOV-2011

11

Principio de operación4 

 

K es el factor de Calibración de cada tubo. Depende del material y la temperatura. Q = K*ΔT Coeficiente de temperatura : % de cambio de la elasticidad por 100 °C

C.COTRINO-NOV-2011

12

Page 6

Principio de operación4 



Cuando el volumen de fluido en el tubo permanece constante, los cambios en la masa son debidos a los cambios de densidad. La frecuencia de oscilación del tubo depende de la masa y de la densidad

13

C.COTRINO-NOV-2011

Principio de operación4 



C.COTRINO-NOV-2011

Cuando la masa del fluido aumenta, la frecuencia de oscilación decrece. Cuando la masa del fluido disminuye, la frecuencia de oscilación aumenta

14

Page 7

Principio de operación4 La señal de frecuencia se toma del pickup izquierdo (LPO).  Se mide el periodo de la oscilación en microsegundos  La densidad es directamente proporcional al periodo 

C.COTRINO-NOV-2011

15

Características ◦ Medición directa de gran exactitud:    

Flujo másico Densidad Temperatura Medición indirecta de Flujo volumétrico

◦ En la medición no afectan cambios de: Presión, Temperatura, Densidad, Viscosidad, Régimen de flujo, Conductividad, Velocidad. ◦ Tipo de montaje (no es necesario definir tramos de medición).

C.COTRINO-NOV-2011

16

Page 8

Aplicaciones   

    

Slurries Viscosidad variable Líquidos sanitarios Líquidos abrasivos Líquidos de densidad variable Líquidos sensibles al cizallamiento Líquidos adheribles Gases

17

C.COTRINO-NOV-2011

Masa vs. Volumen 

Ejemplo 200 L

195 L

GASOLINA A 15 DEG C

GASOLINA A -6.5 DEG C

146 Kilogramos

146 Kilogramos

C.COTRINO-NOV-2011

18

Page 9

Flujo bifásico3 Flujo bifásico anular: El liquido fluye por las paredes y el gas circula por el centro de la tubería a gran velocidad.

19

C.COTRINO-NOV-2011

Flujo bifásico3 Flujo bifásico con burbujas:

Alta viscosidad

Baja viscosidad C.COTRINO-NOV-2011

20

Page 10

Flujo bifásico3 Flujo bifásico con “slug flow”:

C.COTRINO-NOV-2011

21

Tipos de tubo3

C.COTRINO-NOV-2011

22

Page 11

Flujo dividido4

C.COTRINO-NOV-2011

23

Tubos rectos5

C.COTRINO-NOV-2011

24

Page 12

Tubo recto4

C.COTRINO-NOV-2011

25

Especificaciones1

C.COTRINO-NOV-2011

26

Page 13

Especificaciones1

C.COTRINO-NOV-2011

27

Especificaciones1

C.COTRINO-NOV-2011

28

Page 14

Especificaciones1

C.COTRINO-NOV-2011

29

Especificaciones1

C.COTRINO-NOV-2011

30

Page 15

Especificaciones1

C.COTRINO-NOV-2011

31

Pérdida de presión 9

C.COTRINO-NOV-2011

32

Page 16

Pérdida de presión 9

C.COTRINO-NOV-2011

33

Pérdida de presión 9

1: Standard; 2: High pressure

C.COTRINO-NOV-2011

34

Page 17

Vibración7,10 “If the external vibration is close to the working frequency of the Coriolis mass flowmeter, measurement errors will occur”.  “Pulsation is not only critical at the working drive frequency fE of the mass flowmeter but also at frequencies f = fc − fe , where fc is the Coriolis frequency” 

C.COTRINO-NOV-2011

35

Vibración7,10 



“Coriolis mass flowmeters with high working frequencies are much less sensitive to pulsation and external vibrations than others”. “For severely vibrating applications, where the low working frequency of the Coriolis mass flowmeters might become critical, the influence of the external vibration can be considerably reduced by using flexible piping and vibration isolating pipe supports”. C.COTRINO-NOV-2011

36

Page 18

Instalación9 



Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors. Avoid the following locations: ◦ Highest point in a run. ◦ Directly upstream from a free pipe outlet in a down pipe.

C.COTRINO-NOV-2011

37

Instalación3,9 

 



Recommended orientation with upward direction of flow. Entrained solids sink down. Gases rise away from the measuring-tube when fluid is not flowing. The measuring tubes can be completely drained and protected against solids build-up.

C.COTRINO-NOV-2011

38

Page 19

Instalación9 



Transmitter housing above or below the pipe. This arrangement means that no gas and air accumulations and solid deposits can form in the curved measuring tube (single-tube system).

C.COTRINO-NOV-2011

39

Instalación3,9 •Do not install the sensor in such a way that it is suspended in the pipe, in other words without support or attachment. •This is to avoid excessive strain at the process connection. •The base plate of the sensor housing is designed for mounting on a tabletop, wall or post.

C.COTRINO-NOV-2011

40

Page 20

Instalación9 



Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress Installation in a down pipe (e.g. for batching applications): ◦ ◦ ◦ ◦ ◦

1 = Supply tank, 2 = Sensor, 3 = Orifice, pipe restrictions 4 = Valve, 5 = Batching tank

C.COTRINO-NOV-2011

41

Instalación5

Reduzierung

Reduzierung

Instalación con reducción

Instalación con codos

C.COTRINO-NOV-2011

42

Page 21

Calentamiento por vapor5

Verificar máxima temperatura de calentamiento. (Típica: 150°C). Solución para productos viscosos, con tendencia a solidificarse 43

C.COTRINO-NOV-2011

Aplicaciones6 



C.COTRINO-NOV-2011

Engine consumption Measurement Fuel conditioning and consumption measurement

44

Page 22

Aplicaciones8 Dairy plants process raw milk to separate the cream (or milk fat) from the milk to create skim milk.  Standardization of the milk, where the cream and skim milk are reblended to create a “standardized” milk with a desired fat percentage for the product being made. 

C.COTRINO-NOV-2011

45

Aplicaciones8

C.COTRINO-NOV-2011

46

Page 23

Aplicaciones

Fuente: Ing. Jairo Velasco

C.COTRINO-NOV-2011

47

Ventajas / Limitaciones 

Ventajas:  Gran Exactitud  Gran Rangeabilidad  Flujo Bi direccional.  No requiere tramos rectos de tubería  Factor de calibración universal.  Múltiples medidas  Fluidos viscosos, no newtonianos  Gases C.COTRINO-NOV-2011

48

Page 24

Ventajas / Limitaciones 

Limitaciones: ◦ ◦ ◦ ◦ ◦ ◦ ◦

Costo alto Costo de instalación alto Rango de temperatura limitado Restricción en fluidos bifásicos Peso Caída de presión Vibración

C.COTRINO-NOV-2011

49

Especificaciones ..\Soporte\Especificaciones\E+H coriolis massflowmeter.pdf  ..\Soporte\Especificaciones\ISA_TR20_00 _01_2001_ Coriolis.pdf  ..\Soporte\Especificaciones\Micro_Motion _Tube.pdf  ..\Soporte\Especificaciones\Micro_Motion _Transmitter.pdf 

C.COTRINO-NOV-2011

50

Page 25

Referencias 1.

2.

3. 4. 5. 6.

Micro Motion. F-Series Coriolis Flow and Density Meters Product Data Sheet PS-00603, Rev. E. June 2007. Micro Motion. Model D and DT Mass Flow and Density Sensors Product Data Sheet. PS-00388, Rev. B. March 2005. Foxboro. I/A Series mass Flowmeters. PS 1-2B4 A. Micro Motion. Coriolis mass meters tutorial. Khrone. OPTIMASS 1300 - Product data Sheet. SIEMENS. Sitrans FC Transmitters. WS 080810.0. Siemens AG 2008

C.COTRINO-NOV-2011

51

Referencias 7.

Cheesewright R. et. al. The identification of external factors which influence the calibration of Coriolis massflow meters. Flow measurement and instrumentation.Vol 11. (2000) pages 1-10

8.

Micro Motion. AN-001203 Coriolis Flow Meters Improve Accuracy of Standardization Process in Milk Preparation. 2009.

9.

Endress and Hauser. Coriolis Mass Flow Measuring System Promass 80/83 A .TI 054D/06/en 2010

10.

Martin Anklina, Wolfgang Drahmb,, Alfred Riederb. Coriolis mass flowmeters: Overview of the current state of the art and latest research. Flow Measurement and Instrumentation 17 (2006) 317– 323.

C.COTRINO-NOV-2011

52

Page 26