GEMSS Medical NHFG Service Manual 1.SPECIFICATIONS MODEL RMX-550R RMX-750R RMX-850R R R R HighFrequency Inverter
Views 71 Downloads 0 File size 2MB
GEMSS Medical NHFG Service Manual
1.SPECIFICATIONS MODEL
RMX-550R
RMX-750R
RMX-850R
R
R
R
HighFrequency Inverter(400KHz)
HighFrequency Inverter(400KHz)
HighFrequency Inverter(400KHz)
CONSTANT OUTPUT(Kw)
50Kw
50Kw
80Kw
Kvp
40~150Kv
40~150Kv
40~150Kv
mA Range
10~500mA
10~700mA
10~800mA
Exposure Time
0.001~10 sec
0.001~10 sec
0.001~10 sec
AEC
Enable
Enable
Enable
APR
250
250
250
Power Line
3-Phase 380VAC
3-Phase 380VAC
3-Phase 380VAC
ROTOR
OPTIONAL LOW / DUAL
OPTIONAL LOW / DUAL
DUAL SPEED STARTER (HIGH / LOW)
Input Frequency(Hz)
50/60
50/60
50/60
TYPE
1. POWER LINE REQUIREMENTS
50Kw Three-Phase Line Voltage: 380VAC±5% Line Frequency: 50/60Hz Momentary Current: 100Amps Stanby Current: 5Amps
80Kw Three-Phase Line Voltage: 380VAC±5% Line Frequency: 50/60Hz Momentary Current: 150Amps Stanby Current: 5Amps
PAGE 1 / 27
GEMSS Medical NHFG Service Manual
Mode l
RMX550R (50K w) RMX750R (50K w) RMX850R (80K w)
Main Voltag e
Minimum Recommend ed Main Disconnect to HFG
HFG Momenta ry Line Current
Minimum Recommend ed Generator Service Rating
Minimum Recommend ed Distribution Transformer Rating
Groun d Wire Size
Apparent Mains Resistan ce
380VA C
#6 (15mm2)
100A
100A
65kVA
#4 (25mm 2)
0.17Ω
380VA C
#6 (15mm2)
100A
100A
85kVA
#4 (25mm 2
0.13Ω
380VA C
#6 (15mm2)
150A
100A
105kVA
#4 (25mm 2
0.11Ω
2..INSTALL
INSTALL POWER LINE
Securely fix the Power Line to the Terminal. Connect Ground Line to the Bolt where GND Lines are gathered using 5mm NUT st 1 type Ground is recommended.
PAGE 2 / 27
GEMSS Medical NHFG Service Manual
TRANSFORMER CONNECTION
In the case of RMX-850R, Transformer is connected as the picture. Depending on Main Voltage, Transformer’s Connection shall be changed.
< When input voltage is not 380VAC> Separate the socket which is connected to 380VAC, and change to match the Main Voltage. Changeable Voltages are 360VAC 380VAC 400VAC.
RMX-850R Transformer Connection
PAGE 3 / 27
GEMSS Medical NHFG Service Manual
RMX-550/750R(Low Rotor) Transformer Connection
ROTOR CONNECTION Tube Side
Generator Side
1
5.W(Com)
2
4.B(Main)
3
3.G(Phase)
5
1.Tube Temp
6
2.Tube Temp
PAGE 4 / 27
GEMSS Medical NHFG Service Manual
3. SETTING
DIP SWITCH Check(Product set up check)
CPU BOARD
DIP SWITCH (SW1)
PIN
SW1
SW2
SW3
SW4
ON
BUCKY1 SPRING
BUCKY2 SPRING
BUCKY3 SPRING
AEC OFF
OFF
BUCKY1 MOTOR
BUCKY2 MOTOR
BUCKY3 MOTOR
AEC ON
SW5 ROTOR ERROR NO CHECK
SW6 FILAMENT ERROR NO CHECK
SW7 CHARGE ERROR NO CHECK
SW8 TEMP ERROR NO CHECK
ROTOR ERROR CHECK
FILAMENT ERROR CHECK
CHARGE ERROR CHECK
TEMP ERROR CHECK
PAGE 5 / 27
GEMSS Medical NHFG Service Manual
DIP SWITCH (SW2)
MODEL SETTING RMX-550R RMX-750R RMX-850R
SW1 ON ON OFF
SW2 OFF ON ON
SW3 ON OFF OFF
TUBE SETTING E7239X E7252X RAD13 E7869X USER USER USER USER
SW4 ON OFF ON OFF ON OFF ON OFF
SW5 ON ON OFF OFF ON ON OFF OFF
SW6 ON ON ON ON OFF OFF OFF OFF
FREQUENCY SETTING 50Hz 60Hz
SW7 OFF ON
ROTOR SPEEP SETTING HIGH SPEED ROTOR (DUAL ROTOR) LOW SPEED ROTOR
SW8 OFF ON
After DIP SW CHECK, Check if it is matching to the Setting for Hardware Information of CONSOLE
PAGE 6 / 27
GEMSS Medical NHFG Service Manual
4.CHECK POINT After INSTALL
3MAIN POWER BOARD
LD2
LD1
VDD
GND
TEST POINT LD2 LD1 VDD-GND
VALUE ON ON DC520V~DC550V
REMARK Generator Power On Main Voltage On Fault < DC520V
PAGE 7 / 27
GEMSS Medical NHFG Service Manual
LOW POWER BOARD
TEST POINT LD1 LD2 TP1-GND TP2-GND
VALUE ON ON +12VDC -12VDC
REMARK +12VDC ON -12VDC ON Adjustment By VR1 FIXED
PAGE 8 / 27
GEMSS Medical NHFG Service Manual
CPU POWER BOARD
TEST POINT LD1~LD6 +5V-GND +24V-GND +15V-GND +12V-GND -15V-GND
VALUE ALL ON +5VDC +24VDC +15VDC +12VDC -15VDC
REMARK ALL DC VOLTAGE ON
PAGE 9 / 27
GEMSS Medical NHFG Service Manual
CPU BOARD
TEST POINT LD5~LD9
VALUE ALL ON
LD1~LD4
OFF
REMARK ALL DC VOLTAGE ON ON only during Program Writing on CPU
PAGE 10 / 27
GEMSS Medical NHFG Service Manual
FILAMENT BOARD
TEST POINT LD1
VALUE ON
LD2~LD3
ALL ON
TP1, TP2
TP4
TP7-TP8
REMARK POWER FROM FILAMENT SMPS +12VDC, -12VDC ON FROM LOW POWER BOARD-
Normal: Large Focus(200mA)=0.9VDC Small Focus(100mA)=0.6VDC
Ready Button Push: Large Focus(200mA)=2.5VDC Small Focus(100mA)=1.7VDC
Normal: Large Focus(200mA)=1.8VDC Small Focus(100mA)=1.5VDC
Ready Button Push: Large Focus(200mA)=3VDC Small Focus(100mA)=2.3VDC
52VDC
When it is less than 52VDC, more than 500mA cannot normally exposed.
PAGE 11 / 27
GEMSS Medical NHFG Service Manual
TEST POINT
VALUE
REMARK
Adjust to VR1 (Factory Setting)
TP3-GND
1V
10usec
15usec per one cycle
NEW CONTROL BOARD
TEST POINT LD1~LD8
VALUE OFF
LD9~LD10
ALL ON
LD13
ON Large Focus: ON Small Focus: OFF
LD12
REMARK It is on when ERROR occurs. +12VDC, -12VDC ON FROM LOW POWER BOARD-
PAGE 12 / 27
GEMSS Medical NHFG Service Manual
POWER RELAY BOARD
TEST POINT J4
VALUE
REMARK
360~380VAC
From Transformer
200~220VAC
From Transformer
100~110VAC
From Transformer
J4
J4
J5 High Rotor : 360VAC Low Rotor : 200VAC
Rotor
Power
J6 200~220VAC
Fan
Power
PAGE 13 / 27
GEMSS Medical NHFG Service Manual
J7 200~220VAC
Use only when it is High(Dual) Rotor
ROTOR POWER SWITCHING BOARD(NOT USE LOW ROTOR)
TEST POINT
VALUE
LD1
OFF
LD2
ON
REMARK High Rotor : ON Low Rotor : OFF POWER for Collimator (J2,J6 USE)
J4 200~380VAC
High Rotor Low Rotor
: 360~380VAC : 200~220VAC
PAGE 14 / 27
GEMSS Medical NHFG Service Manual
DUAL ROTOR BOARD
LED DESCRIPTION START RUN STOP 180Hz 60Hz ON/OFF TEST POINT
VALUE
REMARK
START LED ON
It is ready before press “Ready” button.
START LED ON STOP LED ON 60Hz LED ON
It is Standby during LOW ROTOR driving condition.
START LED ON STOP LED ON 60Hz LED ON RUN LED ON ON/OFF LED ON START LED ON STOP LED ON 180Hz LED ON
“READY” BUTTON is pressed during LOW ROTOR driving condition.
During HIGH ROTOR driving condition.
PAGE 15 / 27
GEMSS Medical NHFG Service Manual
START LED ON STOP LED ON 180Hz LED ON RUN LED ON ON/OFF LED ON
“READY” BUTTON is pressed during HIGH ROTOR driving condition.
J10 200~380VAC
High Rotor Low Rotor
: 360~380VAC : 200~220VAC
60Hz ROTOR BOARD(Only Low Rotor Speed)
TEST POINT LD1
VALUE OFF
LD2, LD3
ON
REMARK On when it’s Ready +12VDC, -12VDC ON FROM LOW POWER BOARD
J1 200~220VAC
Low Rotor Power
PAGE 16 / 27
GEMSS Medical NHFG Service Manual
NEW CONTROL VCO BOARD
TEST POINT
VALUE
REMARK
Level does not change. Only Frequency will be changed according to Kv, and mA.
TP1-GND TP2-GND
5V
10usec
PAGE 17 / 27
GEMSS Medical NHFG Service Manual
Level does not change. Only pulse interval will be changed according to Time.
TP3-GND
2V
10msec
5.ERROR CODES AND TROUBLE SHOOTING GUIDE Error Code Error 01
Description Fault_Rotor_Signal
Error 03
☞Rotor Power Check, Rotor Cable Check, Tube Check Check the voltage of ROTOR BD. Check if the LED of ROTOR BD is normally ON. Check if the ROTOR Cable is properly connected to GENERATOR. Fault_FET_Power
Error 04
☞Check Mosfet Board Input Power The error cannot be identified in the field. Returning to factory is required. Fault_FET_A
Error 05
☞Mosfet Board Overcurrent >>>Change Board The error cannot be identified in the field. Returning to factory is required. Fault_FET_B
Error 06
☞Mosfet Board Overcurrent >>>Change Board The error cannot be identified in the field. Returning to factory is required. Fault_FET_C
Error 07
☞Mosfet Board Overcurrent >>>Change Board The error cannot be identified in the field. Returning to factory is required. Fault_KV ☞Check New Control KV BD or Control Main BD Replacement of NEW CONTROL MAIN BD MODULE is recommended. PAGE 18 / 27
GEMSS Medical NHFG Service Manual
Error 08
Fault_ mA
Error 10
☞Check New Control mA BD or Control Main BD Replacement of NEW CONTROL MAIN BD MODULE is recommended. Fault_Fil_Power ①Check if the voltage of J4 on N-Filament BD is measured as the order of +12VDC,-12VDC, and GND ②Check if DC48~52V is measured on TP7 and TP8.
Error 11 Error 12
Fault_Fil_Over Replacement of N-FILAMENT BD is required. Fault_Fil_Low ①Check if FILAMENT of X-RAY TUBE is Pre-heating ②HIGHTENSION VOLTAGE CHECK
Check the Voltage supplying to J2 of Tank [LG COM SM] In 40KV 100mA condition, measure the AC Voltage at COM and SM with DVM First AC7~9V >>> When it’s READY, it will be changed to AC13~15V In 40KV 200mA condition, measure the AC Voltage at COM and LG with DVM. First AC7~9V >>> When it’s READY, it will be changed to AC13~16V. ③FILAMENT BD VOLTAGE CHECK ERRRO will be occurred when the voltage of TP4 and GND is more than 5V. At LARGE FOCUS, 1.8V→3V(READY) At SMALL FOCUS, 1.5V→2.3V(READY) Above voltage allows little difference, and it is used only for checking the change of the level. ☞ If Pre-heating in ① is checked, but the result of checking of ②③ is not measured, replacement of FILAMENT BD is required.
Error 15
If the result of ②③ are measured but Pre-heating of ① is not checked, replacement of Tank is required. Fault_Charge ☞Check Main Power (3Phase power)
PAGE 19 / 27
GEMSS Medical NHFG Service Manual
The voltage shall be measured as above picture and more than DC520V shall be measured. If the voltage is less than 520VDC, an ERROR can be occurred when X-ray is exposure on high condition.
Error 16
Check if above Magnetic SW is pressed. If two of Magnetic SW are not pressed, ERROR can be occurred because of abnormal power supply. Fault_Temp
Error 22
☞Tube overheating It can be happened when Tube is overheated or mal-connection of cables between TUBE and ROTOR BD. Fault_AD_FIL
Error 23
Error 24
☞Check Filament Board It’s not possible to check in the field as it’s the problem with AD Converter of FILAMENT BD. Replacement of FILAMENT BD is recommended. Fault_PW_CPU1 ☞Check CPU Power Board Check if the LED of CPU POWER BD as above picture is ON. Check the output of voltage for each TEST POINT. If the output of voltage is not correct, CPU POWR BD shall be replaced. Fault_PW_CPU2 ☞Check CPU Power Board Same action is required as ERROR 23. PAGE 20 / 27
GEMSS Medical NHFG Service Manual
Error 25
Fault_PW_CPU3
Error 26
☞Check CPU Power Board Same action is required as ERROR 23.. Fault_PW_CPU4
Error 27
☞Check CPU Power Board Same action is required as ERROR 23. Fault_PW_CTL1
Error 28
☞Check New Control Main BD Check if the voltage of +12V -12V GND is supplied from J13 on NEW CONTROL BD. Fault_PW_CTL2
Error 31
☞Check New Control Main BD Check if the voltage of +12V -12V GND is supplied from J13 on NEW CONTROL BD. Fault_AD_WORK ☞Check CPU Board Replacement of CPU BD is required. Fault_KV_EEPROM ☞Check CPU Board EEPROM It occurs when the KV data at EEPROM of CPU BD is over than its limited value. Replacement of EEPROM is recommended. Fault_MA_EEPROM
Error 32
☞Check CPU Board EEPROM It occurs when the mA data at EEPROM of CPU BD is over than its limited value. Replacement of EEPROM is recommended. Fault_S_PREHEAT_EEPROM
Error 33
☞Check CPU Board EEPROM It occurs when the SMALL FOCUS PREHEATING data at EEPROM of CPU BD is over than its limited value. Replacement of EEPROM is recommended. Fault_L_PREHEAT_EEPROM
Error 29
Error 30
Error 35
☞Check CPU Board EEPROM It occurs when the LARGE FOCUS PREHEATING data at EEPROM of CPU BD is over than its limited value. Replacement of EEPROM is recommended. Fault_Feedback_High ☞Check CPU Board & New Control BD It occurs when the value of FEEDBACK is too high. Replacement of CPU BOARD and NEW CONTROL MAIN BD is recommended.
PAGE 21 / 27
GEMSS Medical NHFG Service Manual
6.CALIBRATION
mA CALIBRATION
NEW CONTROL MAIN BOARD
TEST POINT TP1 TP2 TP3 TP4
VALUE Small Focus mA Testing Point Small Focus GND Point Large Focus mA Testing Point Large Focus GND Point
REMARK 20mA / 1V Oscilloscope GND 100mA / 1V Oscilloscope GND PAGE 22 / 27
GEMSS Medical NHFG Service Manual
VR2 VR3 VR1
mA offset adjust Small mA feedback value adjust Large mA feedback value adjust
Factory setting Factory setting Factory setting
Large Focus mA adjust
Small Focus mA adjust
N-FILAMENT BOARD
FIL OUTPUT ADJ BOARD
TEST POINT
VALUE
VR1
Small Focus mA feedback value adjust
VR2
Large Focus mA feedback value adjust
REMARK Adjust Volume. Adjust the value of Small Focus mA to 20mA/1V. Adjust Volume. Adjust the value of Large Focus mA to 100mA/1V.
PAGE 23 / 27
GEMSS Medical NHFG Service Manual
Kv CALIBRATION
TEST POINT
VALUE
REMARK
TP1
VCO Control Votage
factory testing point
TP2
kVp offset voltage
factory testing point
TP3
kVp GND Point
Oscilloscope GND
TP4
kVp Testing Point
20kVp / 1V
VR1
kVp output adjust
Factory setting
VR1
Large mA feedback value adjust
Factory setting PAGE 24 / 27
GEMSS Medical NHFG Service Manual
FEEDBACK WAVEFORM
SMALL FOCUS TEST
20Kv/1V 20Ma/1V TIME: 0.05S
VALUE
WAVEFORM
80Kv 100mA
120Kv 100mA
140Kv 100mA
PAGE 25 / 27
GEMSS Medical NHFG Service Manual
LARGE FOCUS TEST
20Kv/1V 100mA/1V TIME: 0.05S
VALUE
WAVEFORM
80Kv 200mA
110Kv 320mA
80Kv 400mA
PAGE 26 / 27
GEMSS Medical NHFG Service Manual
7.BLOCK DIAGRAM