Citation preview

SIREMOBIL Compact / SIREMOBIL Iso-C

VB01C.1

Copyright (C) Siemens AG 2005

SIREMOBIL Compact / SIREMOBIL Iso-C

Tabla de contenido Contents view ...................................................................................................... 11 SIREMOBIL overview ............................................................................................ 12 Serial Port ............................................................................................................ 13 Error messages, general informations ................................................................... 13 Error Class ........................................................................................................... 14 D1 Software ......................................................................................................... 14 Error Organization ................................................................................................ 15 Host 7-segment display ........................................................................................ 16 Host 7-segment display, standby messages ........................................................... 16 Host 7-segment display, status messages ............................................................. 17 Host 7-segment display, error codes ..................................................................... 18 Host, D1 switches ................................................................................................ 20 Error messages .................................................................................................... 20 Multispot error messages ...................................................................................... 21 User console messages ........................................................................................ 21 Host - Job control messages ................................................................................. 21 Host - User console messages .............................................................................. 22 Host - Bios messages ........................................................................................... 22 Host - Memoskop messages ................................................................................. 23 Memoskop fatal error messages ............................................................................ 24 Memoskop error messages ................................................................................... 24 Memoskop warning messages ............................................................................... 24 Memoskop information messages ......................................................................... 25 Host error classification error message .................................................................. 25 Generator hardware messages ............................................................................. 25 EEProm data checks overview ............................................................................... 26 EEProm data checks - Error log ............................................................................. 26 EEProm data checks - Info datas ........................................................................... 26 EEProm data checks - Generator ........................................................................... 26 EEProm data checks - Parameters ......................................................................... 27 EEProm access error messages ............................................................................. 27 Host - Slot diaphragm motor driver error messages ............................................... 28 Host - Slot diaphragm rotation motor driver error messages ................................... 28 Host - Camera rotation motor driver error messages .............................................. 28 Host - X-Iris motor driver error messages .............................................................. 29 Host - TV-Iris motor driver error messages ............................................................ 29 Host - Videomed DC messages ............................................................................. 29 Videomed DC messages ....................................................................................... 30 Host - Radiation release control messages ............................................................ 30 Host - Collimator, slot diaphragm messages .......................................................... 30 Host - Diamentor messages .................................................................................. 31 Host - Power up message ..................................................................................... 31 E0001 .................................................................................................................. 31 F0002 .................................................................................................................. 31 W0003 ................................................................................................................. 32 W0100 ................................................................................................................. 32 W0101 ................................................................................................................. 32

2 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W0102 ................................................................................................................. 32 E0103 .................................................................................................................. 33 W0110 ................................................................................................................. 33 W0111 ................................................................................................................. 33 W0113 ................................................................................................................. 34 W1001 ................................................................................................................. 34 W1002 ................................................................................................................. 34 W1003 ................................................................................................................. 34 I1004 .................................................................................................................. 35 I1005 .................................................................................................................. 35 I1006 .................................................................................................................. 35 I1007 .................................................................................................................. 35 I1008 .................................................................................................................. 36 I1009 .................................................................................................................. 36 W1010 ................................................................................................................. 36 W1101 ................................................................................................................. 37 W1102 ................................................................................................................. 37 W1103 ................................................................................................................. 37 W1104 ................................................................................................................. 37 W1105 ................................................................................................................. 38 W1106 ................................................................................................................. 38 E1107 .................................................................................................................. 38 E1108 .................................................................................................................. 39 E1109 .................................................................................................................. 39 E1110 .................................................................................................................. 39 E1111 .................................................................................................................. 39 E1112 .................................................................................................................. 40 E1113 .................................................................................................................. 40 E1114 .................................................................................................................. 40 W1115 ................................................................................................................. 40 E1116 .................................................................................................................. 41 E1117 .................................................................................................................. 41 E1118 .................................................................................................................. 41 E1119 .................................................................................................................. 42 E1120 .................................................................................................................. 42 E1121 .................................................................................................................. 42 E1122 .................................................................................................................. 42 W2000 ................................................................................................................. 43 E2001 .................................................................................................................. 43 I2002 .................................................................................................................. 43 W2003 ................................................................................................................. 44 W2004 ................................................................................................................. 44 E0112 .................................................................................................................. 44 E2005 .................................................................................................................. 44 E2006 .................................................................................................................. 45 E2007 .................................................................................................................. 45 W2008 ................................................................................................................. 45 E2010 .................................................................................................................. 45 E2011 .................................................................................................................. 46 E2012 .................................................................................................................. 46

3 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E2013 .................................................................................................................. 46 E2020 .................................................................................................................. 46 W2030 ................................................................................................................. 47 W2031 ................................................................................................................. 47 W2032 ................................................................................................................. 47 W2033 ................................................................................................................. 48 W2034 ................................................................................................................. 48 W2035 ................................................................................................................. 48 W2040 ................................................................................................................. 48 W2041 ................................................................................................................. 49 F2050 .................................................................................................................. 49 F2051 .................................................................................................................. 49 E2060 .................................................................................................................. 49 E2061 .................................................................................................................. 50 F2070 .................................................................................................................. 50 F2071 .................................................................................................................. 50 E2080 .................................................................................................................. 51 F2105 .................................................................................................................. 51 F2106 .................................................................................................................. 51 F2201 .................................................................................................................. 51 F2202 .................................................................................................................. 52 F2203 .................................................................................................................. 52 F2206 .................................................................................................................. 52 F2207 .................................................................................................................. 52 F2401 .................................................................................................................. 52 F2402 .................................................................................................................. 53 F2403 .................................................................................................................. 53 F2404 .................................................................................................................. 53 F2405 .................................................................................................................. 53 F2406 .................................................................................................................. 53 F2407 .................................................................................................................. 54 F2408 .................................................................................................................. 54 W2600 ................................................................................................................. 54 W2601 ................................................................................................................. 54 E2601 .................................................................................................................. 55 W2602 ................................................................................................................. 55 E2603 .................................................................................................................. 55 W2604 ................................................................................................................. 55 W2700 ................................................................................................................. 56 F28zz ................................................................................................................... 56 F29zz ................................................................................................................... 56 W4009 ................................................................................................................. 56 W4013 ................................................................................................................. 57 W4016 ................................................................................................................. 57 E4017 .................................................................................................................. 57 E4018 .................................................................................................................. 57 W4020 ................................................................................................................. 58 E4021 .................................................................................................................. 58 E4024 .................................................................................................................. 58 W4027 ................................................................................................................. 59

4 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E4028 .................................................................................................................. 59 E4029 .................................................................................................................. 59 E4030 .................................................................................................................. 59 E4031 .................................................................................................................. 60 W4032 ................................................................................................................. 60 E4040 .................................................................................................................. 60 I4501 .................................................................................................................. 61 I4507 .................................................................................................................. 61 W4601 ................................................................................................................. 61 W4602 ................................................................................................................. 61 W4606 ................................................................................................................. 62 E4701 .................................................................................................................. 62 E4702 .................................................................................................................. 62 E4703 .................................................................................................................. 62 E4704 .................................................................................................................. 63 E4705 .................................................................................................................. 63 E4709 .................................................................................................................. 63 E4710 .................................................................................................................. 64 E4711 .................................................................................................................. 64 E4712 .................................................................................................................. 64 E4713 .................................................................................................................. 64 E4714 .................................................................................................................. 65 E4715 .................................................................................................................. 65 E4716 .................................................................................................................. 65 E4717 .................................................................................................................. 65 E4718 .................................................................................................................. 66 E4719 .................................................................................................................. 66 E4720 .................................................................................................................. 66 E4721 .................................................................................................................. 66 E4722 .................................................................................................................. 67 E4723 .................................................................................................................. 67 E4724 .................................................................................................................. 67 F4808 .................................................................................................................. 67 F4809 .................................................................................................................. 68 E4999 .................................................................................................................. 68 W5000 ................................................................................................................. 68 W5001 ................................................................................................................. 68 W5002 ................................................................................................................. 69 F5003 .................................................................................................................. 69 F5004 .................................................................................................................. 69 E5005 .................................................................................................................. 69 E5006 .................................................................................................................. 70 E5007 .................................................................................................................. 70 F5010 .................................................................................................................. 70 F5011 .................................................................................................................. 71 F5012 .................................................................................................................. 71 F5013 .................................................................................................................. 71 F5014 .................................................................................................................. 71 E5015 .................................................................................................................. 72 E5016 .................................................................................................................. 72

5 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W5017 ................................................................................................................. 72 E5018 .................................................................................................................. 73 E5019 .................................................................................................................. 73 F5030 .................................................................................................................. 73 F5031 .................................................................................................................. 73 F5032 .................................................................................................................. 74 F5033 .................................................................................................................. 74 E5034 .................................................................................................................. 74 E5035 .................................................................................................................. 75 E5080 .................................................................................................................. 75 E5081 .................................................................................................................. 75 E5090 .................................................................................................................. 76 F5901 .................................................................................................................. 76 E6000 .................................................................................................................. 76 W6001 ................................................................................................................. 76 W6002 ................................................................................................................. 77 W6003 ................................................................................................................. 77 W6004 ................................................................................................................. 77 W6005 ................................................................................................................. 78 W6006 ................................................................................................................. 78 W6007 ................................................................................................................. 78 E6008 .................................................................................................................. 78 E6009 .................................................................................................................. 79 E6100 .................................................................................................................. 79 E6101 .................................................................................................................. 79 E6102 .................................................................................................................. 79 E6103 .................................................................................................................. 80 E6110 .................................................................................................................. 80 E6200 .................................................................................................................. 80 E6201 .................................................................................................................. 80 E6202 .................................................................................................................. 81 E6203 .................................................................................................................. 81 E6204 .................................................................................................................. 81 E6205 .................................................................................................................. 82 E6206 .................................................................................................................. 82 E6207 .................................................................................................................. 82 E6208 .................................................................................................................. 83 E6209 .................................................................................................................. 83 E6210 .................................................................................................................. 83 E6211 .................................................................................................................. 84 E6212 .................................................................................................................. 84 E6213 .................................................................................................................. 84 E6214 .................................................................................................................. 85 E6215 .................................................................................................................. 85 E6216 .................................................................................................................. 85 E6217 .................................................................................................................. 85 E6218 .................................................................................................................. 86 E6219 .................................................................................................................. 86 E6220 .................................................................................................................. 86 E6221 .................................................................................................................. 87

6 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E6222 .................................................................................................................. 87 E6223 .................................................................................................................. 87 W6224 ................................................................................................................. 87 E6225 .................................................................................................................. 88 E6231 .................................................................................................................. 88 E6232 .................................................................................................................. 88 E6233 .................................................................................................................. 88 E6234 .................................................................................................................. 89 E6235 .................................................................................................................. 89 E6236 .................................................................................................................. 89 E6301 .................................................................................................................. 89 E6302 .................................................................................................................. 90 E6303 .................................................................................................................. 90 E6304 .................................................................................................................. 90 E6305 .................................................................................................................. 90 E7000 .................................................................................................................. 91 E7001 .................................................................................................................. 91 E7002 .................................................................................................................. 91 E7003 .................................................................................................................. 92 E7004 .................................................................................................................. 92 E7006 .................................................................................................................. 92 E7008 .................................................................................................................. 93 E7010 .................................................................................................................. 93 E7100 .................................................................................................................. 93 E7101 .................................................................................................................. 94 E7102 .................................................................................................................. 94 E7103 .................................................................................................................. 94 E7104 .................................................................................................................. 95 E7106 .................................................................................................................. 95 E7108 .................................................................................................................. 95 E7110 .................................................................................................................. 95 E7150 .................................................................................................................. 96 E7151 .................................................................................................................. 96 E7152 .................................................................................................................. 96 E7200 .................................................................................................................. 97 E7201 .................................................................................................................. 97 E7202 .................................................................................................................. 97 E7203 .................................................................................................................. 98 E7204 .................................................................................................................. 98 E7206 .................................................................................................................. 98 E7208 .................................................................................................................. 99 E7210 .................................................................................................................. 99 E7300 .................................................................................................................. 99 E7301 ................................................................................................................ 100 E7302 ................................................................................................................ 100 E7303 ................................................................................................................ 100 E7304 ................................................................................................................ 101 E7306 ................................................................................................................ 101 E7308 ................................................................................................................ 101 E7309 ................................................................................................................ 101

7 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E7310 ................................................................................................................ 102 I7350 ................................................................................................................. 102 I7351 ................................................................................................................. 102 I7352 ................................................................................................................. 103 E7353 ................................................................................................................ 103 E7400 ................................................................................................................ 103 E7401 ................................................................................................................ 104 E7402 ................................................................................................................ 104 E7403 ................................................................................................................ 104 E7404 ................................................................................................................ 105 E7406 ................................................................................................................ 105 E7408 ................................................................................................................ 105 E7409 ................................................................................................................ 105 E7410 ................................................................................................................ 106 E8000 ................................................................................................................ 106 W8002 ............................................................................................................... 106 W8003 ............................................................................................................... 107 E8005 ................................................................................................................ 107 W8006 ............................................................................................................... 107 E8007 ................................................................................................................ 108 E8008 ................................................................................................................ 108 E8020 ................................................................................................................ 108 W8030 ............................................................................................................... 108 E8031 ................................................................................................................ 109 E8032 ................................................................................................................ 109 E8033 ................................................................................................................ 109 E8050 ................................................................................................................ 110 W8500 ............................................................................................................... 110 W8501 ............................................................................................................... 110 W8502 ............................................................................................................... 110 W8503 ............................................................................................................... 111 E8510 ................................................................................................................ 111 E8511 ................................................................................................................ 111 E8512 ................................................................................................................ 112 E8513 ................................................................................................................ 112 E8514 ................................................................................................................ 112 E8515 ................................................................................................................ 112 E8516 ................................................................................................................ 113 E8517 ................................................................................................................ 113 E8518 ................................................................................................................ 113 W9010 ............................................................................................................... 113 W9011 ............................................................................................................... 114 W9012 ............................................................................................................... 114 W9013 ............................................................................................................... 115 W9014 ............................................................................................................... 115 W9015 ............................................................................................................... 116 W9022 ............................................................................................................... 116 W9023 ............................................................................................................... 117 W9024 ............................................................................................................... 117 E9102 ................................................................................................................ 118

8 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E9101 ................................................................................................................ 118 E9701 ................................................................................................................ 118 E9702 ................................................................................................................ 119 IFFFF ................................................................................................................. 119 Emergency operation .......................................................................................... 119 Parameter overview ........................................................................................... 120 Brightness nominal value .................................................................................... 120 Buzzer and blocktime-settings ............................................................................. 121 camera rotation ................................................................................................. 122 Settings for service PC ........................................................................................ 122 Control curve selection (FL/PFL) .......................................................................... 123 Foot Switch ........................................................................................................ 123 Laser ................................................................................................................. 123 Diamentor ......................................................................................................... 124 Error log ............................................................................................................ 124 Limitation of skindoserate (FL) ............................................................................ 124 Memoskop parameters overview ......................................................................... 125 Memoskop parameters User Setup ...................................................................... 125 Memoskop parameters fluoro .............................................................................. 127 Memoskop parameters IFL .................................................................................. 128 Memoskop parameters DR .................................................................................. 129 Memoskop parameters Subtraction ..................................................................... 129 Memoskop parameters ROADMAP ....................................................................... 131 Memoskop parameters display ............................................................................ 131 Memoskop parameters Hardcopy and VCR ........................................................... 132 Memoskop parameters Diamentor ....................................................................... 132 Memoskop parameters Date and Time ................................................................ 133 Memoskop parameters Special functions ............................................................. 133 Memoskop parameters Technical Setup ............................................................... 134 Screen display Slot diaphragm ............................................................................ 136 Parameters Doserate adjustments ...................................................................... 136 Doserate adjustment overview ........................................................................... 138 Doserate adjustment for SIREMOBIL Compact from serial no. 01001 to 01416, with board D1, part no. 30 99 632 .............................................................................. 138 Doserate adjustment for SIREMOBIL compact from serial no. 01417 to 01999 ....... 140 Screen display X-Iris ........................................................................................... 141 Settings for "STOP-MODE" .................................................................................. 142 Slot diaphragm ................................................................................................... 143 X-Iris ................................................................................................................. 143 Monitoring overview ........................................................................................... 144 Generator adjustment ........................................................................................ 144 KV offset adjustment .......................................................................................... 145 Doserate adjustment for SIREMOBIL Compact from serial no. 02001 and SIREMOBIL Iso-C ................................................................................................................. 146 Actual values ..................................................................................................... 148 Calibration overview ........................................................................................... 149 Camera rotation calibration ................................................................................ 149 Camera rotation, Adjustment actual value potentiometer ..................................... 151 Diamentor calibration ......................................................................................... 151 X-Iris calibration ................................................................................................. 152

9 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Calibration Screen display X-Iris .......................................................................... 153 Slot diaphragm calibration .................................................................................. 155 Correction of the 0 degree position of the slot diaphragm display ......................... 156 Calibration Screen display slot diaphragm ............................................................ 156 Tests overview ................................................................................................... 158 Cassette simulation ............................................................................................ 158 Internal tests ..................................................................................................... 159 Memoskop tests ................................................................................................. 159 Videomed DC tests ............................................................................................. 159 Download .......................................................................................................... 160 Download Memoskop C / C-E / SUB ..................................................................... 160 Download board D1 ............................................................................................ 160 Memoskop C-E, Keyboard emulation ................................................................... 161 Memoskop video frequency configuration ............................................................ 165 Memoskop, configuration to the number of existing monitors ............................... 165 Deleting Error log ............................................................................................... 166 Backup .............................................................................................................. 166

10 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Contents view SIREMOBIL Iso-C / SIREMOBIL Compact Helptext Version VB01C Contents view · SIREMOBIL overview · Error messages, general informations Error classes Error organization Error messages · Host 7-segment display Host 7-segment display, error codes Host 7-segment display, status messages Host 7-segment display, standby messages · Host, D1 service switches · Parameter overview Brightness nominal value Buzzer and blocktime settings Camera rotation Control curve selection (FL/PFL) Diamentor Parameters Doserate adjustment Error log Foot switch Laser Limitation of skin dose rate (FL) Screen display Slot diaphragm Screen display X-Iris Settings for "STOP-MODE" Settings for Service PC Slot diaphragm X-Iris Memoskop · Calibration overview Camera rotation Diamentor Screen display X-Iris Screen display Slot diaphragm Slot diaphragm X-iris Memoskop · Tests overview Cassette simulation Internal tests (not available for service) Memoskop test Videomed DC test · Monitoring overview Doserate adjustment overview Generator adjustment KV offset adjustment Actual values ·

Memoskop C / C-E

11 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Video frequency configuration Configuration to the number of existing monitors Memoskop C-E, Keyboard emulation Memoskop test patterns · Download Download board D1 Download Memoskop C / C-E · Backup · Serial port Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

SIREMOBIL overview

SIREMOBIL Iso-C / SIREMOBIL Compact Overview

SIREMOBIL Compact

SIREMOBIL Iso-C

Go to Contents view

12 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Easily create EBooks

Serial Port

Connecting cable: 99 00 440 RE999 (Null modem cable, 9pin cannon plugs ) Default settings: Connection: Baudrate: FIFO:

COM1 19200 ON *)

*) For to turn ON select "Standard" in Start/ Settings/ Control Panel/ System/ Hardware/ Device Manager/ Ports/ COM1/ Properties/ Port Settings/ Advanced: Use FIFO If needed, select another serial port during callup of the service program. Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

Error messages, general informations

Error class The SIREMOBIL Error classes are: · Informations · Warnings · Errors · Fatal Errors Error organization The SIREMOBIL Error messages are organized in: · User console · Host · Memoskop · Generator · EEProm data checks · Videomed DC · Radiation release control · Collimator, slot diaphragm · Camera rotation Videomed DC · Collimator X iris · Diamentor (if present) See also: Deleting error log Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

13 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Error Class

The SIREMOBIL · I · W · E · F

Error classes are: = Informations = Warnings = Errors = Fatal Errors

· Informations: Description: Displayed on basic unit? Stored in Error Log? Remarks See Error Log .

Operating error no no (default service software setting) For Unit observation, activate "Log warnings and infos"

Description: Displayed on basic unit? Stored in Error Log? Remarks: See Error Log .

sporadic disturbance, no function limitations no no ( default service software setting) For Unit observation, activate "Log warnings and infos"

Description: Displayed on basic unit? Stored in Error Log? Remarks:

disturbance, (strong) function limitations yes yes Error reset possible, in the case of Videomed DC actual brightness errors, maybe "emergency operation" possible

Description: Displayed on basic unit? Stored in Error Log? Remarks:

disturbance, Unit use not possible yes yes If a fatal error will occur, radiation will stop immediately. Error reset is possible.

· Warnings:

· · Errors:

· Fatal Errors:

see also Error organization Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

D1 Software

· Boot software: The host starts with the boot software. The boot software is stored in a unerasable area of the Flash Prom. After the successful initialization routine and test of the D1 hardware, the user software is started. Exceptions: User software not present, or switch D1. S3.3 set to position ON. See also Host 7-segment display, standby messages . · User software: After the successful D1 hardware initialization, the user software ist started. The user software is stored in Flash PROMs.

14 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Using Download, the user software can be loaded into the Flash PROMs via the service interface. See also Host 7-segment display, standby messages . · Download: Updating the user software using a service PC. (Connect the service PC to the serial interface D1. Start the service program and select the Download menu). See also Download board D1. Go to Contents view Created with the Personal Edition of HelpNDoc: Free Kindle producer

Error Organization

The SIREMOBIL errors are organized as follows and also used in the service pc: Error code XYYZZ reference list with X: with YY: YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ = YYZZ =

Error class 00ZZ 01ZZ 10ZZ 20ZZ 40ZZ 45ZZ 46ZZ 47ZZ 48ZZ 49ZZ 50ZZ 60ZZ 61ZZ 611Z 62ZZ 63ZZ 70ZZ 71ZZ 72ZZ 73ZZ 74ZZ 80ZZ 85ZZ 90ZZ 91ZZ 97ZZ FFFF

and ZZ:

integration to a component User console messages Host - User console messages Host - Job control messages Host Bios messages Host - Memoskop messages Memoskop information messages Memoskop warning messages Memoskop error messages Memoskop fatal error messages Host - Error classification error messages Generator hardware messages EEProm data checks - Error log EEProm data checks - Generator EEProm data checks - Info datas EEProm data checks - Parameters EEProm access error messages Host - Slot diaphragm motor driver error messages Host - Slot diaphragm rotation motor driver error messages Host - Camera rotation motor driver error messages Host - X-Iris motor driver error messages Host - TV-Iris motor driver error messages Host - Videomed DC messages Videomed DC messages Host - Radiation release control messages Host - Collimator, slot diaphragm messages Host - Diamentor messages (if diamentor is present) Host - Power up message error numbers within a component.

The component mentioned first registers the error and reports the error code to the host. The host enters the error in the error memory. According to Error class , the error is stored and/ or displayed on the user console. Go to Contents view

15 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Produce electronic books easily

Host 7-segment display

The 7-segment display shows informations about the host: These informations are divided into: Error codes Status messages Standby messages Error codes: 0 = Processor error D1. 1 = Checksum error boot code D1. 2 = RAM error D1. 4 = Watchdog error 8 = Telegram NAK error, serial port Service PC. 9 = Telegram overrun error, serial port Service PC. C = Telegram checksum error, serial port Service PC. P = Telegram parity error, serial port Service PC. F = Telegram framing error, serial port Service PC. b = Telegram break error, serial port Service PC. E = AD converter error D1. S = serial port to service pc defective. L = CPU timer error Status messages 1 = Checksum error boot code D1. 2 = RAM test D1 active. 3 = Checksum test user software D1 active. 4 = Watchdog test active. 6 = Internal messages 7 = Internal messages d = Init EEProm Standby messages Circulating segment clockwise Circulating segment counter clockwise Horizontal segment, vertical jumping Flashing middle horizontal segment Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Web Help sites

Host 7-segment display, standby messages

The standby messages indicate the operating status of the host after the end of the internal tests.

Circulating segment clockwise Host user software started from flash prom. User software is active, normal case.

16 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Circulating segment counter clockwise Possible trouble: host boot software started from boot prom. - Boot software is active, normal during Download. - Normal in replacement boards with empty flash proms . Possible help: Try Download Host. D1 defective.

Horizontal segment, vertical jumping Possible trouble: D1 service switch S3.3 in position "on". - Boot software is active - Download is possible. Possible help: Start Download of the Host (if needed) or switch to "off".

Flashing middle horizontal segment Possible trouble: D1 service switch S3.2 in position "on". Download Memoskop is possible. Possible help: Start Download of the Memoskop (if needed) or switch to "off". See also Host 7-segment display, error codes and Host 7-segment display, status messages. Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

Host 7-segment display, status messages

The messages 1 to 4, 6, 7 and "d" represent status messages of the host. These are displayed only briefly in the normal case. If the test was completed successfully, the processor jumps to the next test. Once all tests are completed successfully, the momentarily valid standby messages are output. 1=

Checksum test boot software D1 active. Host is testing the boot prom checksum.

2=

RAM test D1 active. Host is testing the RAM.

3=

Checksum test user software D1 active. Host is testing the flash prom checksum.

4=

Watchdog test active. Host is testing the watchdog timer.

6=

Internal status message Initializing message

7=

Internal status message Host is calculating the kV/mA curve

d=

Init EEProm Host is deleting the EEProm. The procedure takes about 5 minutes. Only active, after the first download, if the board D1

17 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

was changed and the EEProm on board D1 was not empty. Attention ! Do not switch off the unit! Do nit start other service activities! Wait, till the 7 segment display is changing to the normal standby messages (see standby messages). If one of the status messages is displayed permanently, this test could not be ended successfully. See Host 7-segment display, error codes. Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

Host 7-segment display, error codes

The messages 1 to 4 and 7 are displayed only briefly in the normal case and represent only status messages . If these messages are displayed permanently, the relevant test could not be ended successfully. The message must therefore be assessed as an error message. The messages 8 to C, F and P indicate errors of the serial interface to the service PC. If these messages are displayed only sporadically and briefly, no further measures have to be taken. Further information: see Host 7-segment display, standby messages. The following error codes are displayed: 0=

Processor error D1. Possible trouble: Possible help:

Processor detects internal error. Test +5 V power pack D1 defective.

1=

Checksum error boot code D1. Possible trouble: Processor detects checksum error of the boot prom. Possible help: Test +5 V power pack D1 defective.

2=

RAM error D1. Possible trouble: Possible help:

4=

Watchdog error Possible trouble: Possible help:

Processor detects checksum error of the flash prom. Test +5 V power pack D1 defective. Processor detects watchdog error. Jumper X98. 1-2 not plugged in. D1 defective.

8=

Telegram NAK error, serial port Service PC. Possible trouble: If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Possible help: Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

9=

Telegram overrun error, serial port Service PC.

18 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble: Possible help:

If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

b=

Telegram break error, serial port Service PC. Possible trouble: If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Possible help: Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

C=

Telegram checksum error, serial port Service PC. Possible trouble: If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Possible help: Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

E=

AD converter error D1. Possible trouble: AD converter test could not be completed successfully. Possible help: Check + 15 V / - 15 V power supplies. D1 defective.

F=

Telegram framing error, serial port Service PC. Possible trouble: If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Possible help: Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

L=

CPU timer error Possible trouble: Possible help:

Timer on D1 defective. D1 defective.

S=

serial port to service pc defective. Possible trouble: Serial interface to the service PC defective. Possible help: D1 defective.

P=

Telegram parity error, serial port Service PC. Possible trouble: If the boot prom is active, the serial interface to the service PC is controlled by the boot software. The host detects errors during the communication with the service PC. Possible help: Sporadic: Interface cable defective. Faults in the system. Check service PC. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

19 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Host, D1 switches

The following switches are provided on the board D1: Switch D1. S2

SS switch ON: OFF:

DIP switch D1. S3

SS switch is on if LED V85 lights up. Radiation release is possible. The radiation generation of the SIREMOBIL is disabled (for service). However, if radiation is released, the error E5006 will occur.

Default settings: All switches in position OFF S3.1

Always in OFF position.

S3.2 OFF S3.2 ON

Normal position. Only Download Memoskop possible. See also Download Memoskop.

S3.3 OFF S3.3 ON

Normal position. Prevents the user software starting, boot software is active. Only Download board D1 possible. See also Host 7-segment display, standby messages. See also Download board D1.

S3.4 OFF S3.4 ON

Normal position. DEMO mode, radiation is blocked. SIREMOBIL otherwise ready for demonstration.

Go back to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

Error messages

The SIREMOBIL error messages are organized as follows: See also Error organization and Error classes. · · · · · · · · · · · · · · · · ·

Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages Error messages

00zz 01zz 10zz 20zz 40zz 45zz 46zz 47zz 48zz 49zz 50zz 60zz 610z 611z 62zz 63zz 70zz

= = = = = = = = = = = = = = = = =

User console messages Host - User console messages Host Job control messages Host Bios messages Host - Memoskop messages Memoskop information messages Memoskop warning messages Memoskop error messages Memoskop fatal error messages Host - Error classification error message Generator hardware messages EEProm data checks - Error log EEProm data checks - Generator EEProm data checks - Info datas EEProm data checks - Parameters EEProm access error messages Host - Slot diaphragm motor driver slot diaphragm

20 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

· Error messages diaphragm rotation · Error messages · Error messages · Error messages · Error messages · Error messages · Error messages · Error messages · Error messages · Error messages

71zz

=

Host - Slot diaphragm rotation motor driver slot

72zz 73zz 74zz 80zz 85zz 90zz 91zz 97zz FFFF

= = = = = = = = =

Host - Camera rotation motor driver error messages Host - X-Iris motor driver error messages Host - TV-Iris motor driver error messages Host - Videomed DC messages Videomed DC messages Host - Radiation release control messages Host - Collimator, slot diaphragm messages Host - Diamentor messages (if present) Host - Power up message

=

Multispot error messages

· Error messages Go to Contents view

Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

Multispot error messages

· The Multispot hardcopy camera error messages are only displayed in the Multispot camera front display. · The error messages will not be stored in the SIREMOBIL error log. You will find additional informations in the Multispot camera user manual. Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

User console messages

E0001 F0002 W0003

User console no contact to host User console internal error User console internal watchdog reset

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

Host - Job control messages

W1001 W1002 W1003 I1004 I1005 I1006 I1007 I1008 I1009 W1010

User console internal system error Cassette switch pressed dur. radiation Cassette switch released dur. radiation User console reserve keycode 1 User console reserve keycode 2 User console reserve keycode 3 User console unknown keycode User console ROADMAP pressed User console SUB pressed Foot switch parameter wrong configurated

21 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W1101 W1102 W1103 W1104 W1105 W1106 E1107 E1108 E1109 E1110 E1111 E1112 E1113 E1114 W1115 E1116 E1117 E1118 E1119 E1120 E1121 E1122

Operating system driver int. syst. error Radiation release driver int. error Gen. driver int. system error Fluoro state machine int. system error DR state machine internal system error IDL state machine internal system error Videomed DC radiation release timeout Gen. learning internal system error Dose adjustment internal system error Cassette mode internal system error Acquisition mode internal system error Cassette switch simulation int. error Operating mode not valid Gen. learning routine int. error Gen. learning user break Diamentor adjustment routine int. error KV Offset adjustment routine int. error Memoskop timeout Host - Memoskop dose reduction not received RM state machine wrong state pointer SUB statemachine wrong state pointer 3D statemachine wrong state pointer

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Kindle producer

Host - User console messages

W0100 W0101 W0102 E0103 W0110 W0111 E0112 W0113

User console ACK timeout User console Byte timeout User console receive timeout User console initialization timeout User console internal system error User console internal system error User console does not answer User console internal system error

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

Host - Bios messages

W2000 E2001 I2002 W2003 W2004 E2005 E2006 E2007 W2008

Undefined processor interrupt System reboot after watchdog timeout Invalid acquisition end interrupt Serial port 0, invalid interrupt Serial port 1, invalid interrupt Videomed DC, ser. receive line error Memoskop, ser. receive line error User console, ser. receive line error Service, ser. receive line error

E2010 E2011

Videomed DC, ser. port init not possible Memoskop, ser. port init not possible

22 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E2012 E2013 E2020 W2030 W2031 W2032 W2033 W2034 W2035 W2040 W2041 F2050 F2051 E2060 E2061 F2070 F2071 E2080 F2105 F2106 F2201 F2202 F2203 F2206 F2207 F2401 F2402 F2403 F2404 F2405 F2406 F2407 F2408 W2600 E2601 W2601 W2602 E2603 W2604 W2700 F28zz F29zz

User console, ser. port init not possible Service, ser. port init not possible Internal system error User console serial port parity error User console serial port overflow error User console serial port framing error Memoskop serial port parity error Memoskop serial port overflow error Memoskop serial port framing error Host serial port buffer overflow Host serial port buffer overflow Host - filament configuration failed Host - filament code corrupted Host - ARC configuration failed Host - ARC code corrupted Host - ADC test failed Host - ADC Busy too long Host - powerfail signal active during bootup Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Internal system error Elapsed time meter initialization error Elapsed time meter uncorrectable error Elapsed time meter uncorrectable error Elapsed time meter correctable error EEProm area for elapsed time meter defective Elapsed time meter writing not more possible Host - powerfail signal active (sporadic) Delay task internal debug messages Delay task internal debug messages

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

Host - Memoskop messages

W4009 W4013 W4016 E4017 E4018 W4020 E4021 E4024 W4027 E4028 E4029

Internal system error Memoskop acq timeout Memoskop wait telegram timeout Memoskop busy signal timeout Memoskop acq start signal timeout Memoskop send timeout Memoskop does not answer Memoskop initialization timeout Memoskop unexpected reset Memoskop unexpected acq end signal Memoskop unexpected acq. reset

23 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E4030 E4031 W4032 E4040

Memoskop acq. active w.o. start signal Memoskop acq. too short time active Memoskop acq. too short time active Memoskop acq active dur. initialization

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

Memoskop fatal error messages

F4808 F4809

Memoskop harddisk drive image datas lost Memoskop battery ram failed

Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

Memoskop error messages

E4701 E4702 E4703 E4704 E4705 E4709 E4710 E4711 E4712 E4713 E4714 E4715 E4716 E4717 E4718 E4719 E4720 E4721 E4722 E4723 E4724

Memoskop corrupted patient directory Memoskop hardcopy timeout error Memoskop harddisk drive error Memoskop hardware error Memoskop harddisk drive not ready Memoskop battery ram failed Memoskop D/A 1 LUT error Memoskop D/A 1 0V error Memoskop D/A 2 LUT error Memoskop D/A 2 0V error Memoskop D/A 3 LUT error Memoskop D/A 3 0V error Memoskop memory 1 access failure Memoskop memory LSN access failure Memoskop memory 2 access failure Memoskop memory 3 access failure Memoskop no response from 3D PC Memoskop communication error with 3D PC Memoskop disk read or write error in 3D mode Memoskop memory allocation error in 3D mode Memoskop, error message from 3D PC received

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

Memoskop warning messages

W4601 W4602 W4606

Memoskop no HS_ACK within timeout period Memoskop realtime clock malfunction Memoskop disk full

Go to Contents view

24 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

Memoskop information messages

I4501 I4507

Memoskop disk transfer time too long Memoskop no stored images

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

Host error classification error message

E4999

Host received telegram with wrong error class

Go to Contents view Created with the Personal Edition of HelpNDoc: Free HTML Help documentation generator

Generator hardware messages

W5000 W5001 W5002 F5003 F5004 E5005 E5006 E5007 F5010 F5011 F5012 F5013 F5014 E5015 E5016 W5017 E5018 E5019 F5030 F5031 F5032 F5033 E5034 E5035 E5080 E5081 E5090 F5901

Gen. Cassette exposure user break Gen. mAs actual value too low Gen. Sirephos temp. value out of range Gen. ref. voltage +10V out of range Gen. Uz ist out of range Gen. tried X-Ray on without LIH signal Gen. tried X-Ray on without SS relais Gen. IH max exceeded Gen. filament inverter short circuit Gen. main inverter short circuit Gen. kV max exceeded Gen. +15 V or -15 V out of tolerance Gen. Sirephos max oil pressure exceeded Gen. Sirephos temp. lower limit obtained Gen. Sirephos max temp. exceeded Gen. Sirephos temp. fast level change Gen. kV offset below lower limit Gen. kV/mA courve selection, int. error Gen. tube current actual value too high Gen. tube current actual value too low Gen. high voltage actual value too high Gen. high voltage actual value too low Gen. IH actual value too high (standby) Gen. IH actual value too low (standby) Gen. learning pushfactor out of tolerance Gen. learning internal error Gen. Videomed DC not available Gen. tried X-Ray on w.o. XRAY_query signal

Go to Contents view

25 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

EEProm data checks overview

Error messages 60zz Error messages 61zz Error messages 611z Error messages 62zz Error messages 63zz

EEProm data checks - Error log EEProm data checks - Generator EEProm data checks - Info datas EEProm data checks - Parameters EEProm access error messages

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

EEProm data checks - Error log

E6000 W6001 W6002 W6003 W6004 W6005 W6006 W6007 E6008 E6009

Error log is not initialized Error log, reading error Error log, writing error Error log, end marking not found Error log, checksum error Error log, correctable error Error log, error could not be registrated Error log, EEProm blocked Error log, EEProm write failed Error log, EEProm write failed

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

EEProm data checks - Info datas

E6110

Infodatas checksum error

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

EEProm data checks - Generator

E6100 E6101 E6102 E6103

Generator needs learning Gen, kV offset out of range Gen, push factor out of range Gen, filament heat. factor out of range

Go to Contents view

26 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

EEProm data checks - Parameters

E6200

Parameter checksum error

E6201 E6202 E6203 E6204

Param. Cam. rot. -180 deg. out of range Param. Cam. rot. 0 deg. out of range Param. Cam. rot. -180 deg. out of range Param. Cam. rot. max position out of range

E6205 E6206 E6207 E6208 E6209 E6210

Param. X-iris min. pos. out of range Param. X-iris max pos. out of range Param. X-iris close pos. out of range Param. X-iris full format pos. out of range Param. X-iris zoom format pos. out of range Param. X-iris cassette pos. out of range

E6211

Param. brightness nom. value out of range

E6212 E6213

Param. kV/mA curve normal wrong Param. kV/ma curve push wrong

E6214

Param. settings for stop mode wrong

E6215

Param. Limit. of skindoserate wrong

E6216 E6217 E6218

Param. Buzzer/blocktime settings wrong Param. Buzzer/blocktime settings wrong Param. Buzzer/blocktime settings wrong

E6219 E6220

Param. Diamentor existing wrong Param. Diamentor constant wrong

E6221 E6222

Param. Error log is wrong Param. Laser existing is wrong

E6223 W6224 E6225

Param. Monitoring timeout is wrong Param. Test enable is wrong Param. Max error messages is wrong

E6231 E6232 E6233 E6234 E6235 E6236

Param. TV-Iris not plausible Param. Zoom area factor out of range Param. MID-HIGH Factor out of range Param. MID-LOW Factor out of range Param. Adjustment mode out of range Param. Settings for Stop parameter out of range

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Qt Help files

EEProm access error messages

E6301 E6302

EEProm reading error EEProm writing error

27 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E6303 E6304 E6305

EEProm reading blocked EEProm writing blocked EEProm writing failed

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate EPub eBooks with ease

Host - Slot diaphragm motor driver error messages

E7000 E7001 E7002 E7003 E7004 E7006 E7008 E7010

Slot diaphragm movement blocked Slot diaphragm movement blocked Slot diaphragm movement to wrong direction Slot diaphragm movement to wrong direction Slot diaphragm incremental encoder out of range Slot diaphragm incremental encoder out of range Slot diaphragm unexpected movement Slot diaphragm, programmed limit exceeded

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

Host - Slot diaphragm rotation motor driver error messages

E7100 E7101 E7102 E7103 E7104 E7106 E7108 E7110 E7150 E7151 E7152

Slot diaphragm rotation blocked Slot diaphragm rotation blocked Slot diaphragm rotation to wrong direction Slot diaphragm rotation to wrong direction Slot diaphragm rotation actual value out of range Slot diaphragm rotation actual value out of range Slot diaphragm rotation unexpected movement Slot diaphragm rotation limits exceeded Slot diaphragm rotation index not found Slot diaphragm rotation 0 deg.. pos. not found Slot diaphragm rotation unexpected position values

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

Host - Camera rotation motor driver error messages

E7200 E7201 E7202 E7203 E7204 E7206 E7208 E7210

Camera rotation blocked Camera rotation blocked Camera rotation to wrong direction Camera rotation to wrong direction Camera rotation actual value out of range Camera rotation actual value out of range Camera rotation unexpected movement Camera rotation limits exceeded

Go to Contents view

28 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

Host - X-Iris motor driver error messages

E7300 E7301 E7302 E7303 E7304 E7306 E7308 E7309 E7310 I7350 I7351 I7352 E7353

X-Iris movement blocked X-Iris movement blocked X-Iris movement to wrong direction X-Iris movement to wrong direction X-Iris actual value out of range X-Iris actual value out of range X-iris unexpected movement X-Iris wrong position X-Iris limits exceeded X-Iris stored positions not plausible X-Iris monitor display values not accessible X-Iris stored position values not accessible X-Iris, I.I. format not plausible

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Qt Help files

Host - TV-Iris motor driver error messages

E7400 E7401 E7402 E7403 E7404 E7406 E7408 E7409 E7410

TV-Iris movement blocked TV-Iris movement blocked TV-Iris movement to wrong direction TV-Iris movement to wrong direction TV-Iris actual value out of range TV-Iris actual value out of range TV-Iris unexpected movement TV-Iris wrong position TV-Iris limits exceeded

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

Host - Videomed DC messages

E8000 W8002 W8003 E8005 W8006 E8007 E8008 E8020 W8030 E8031 E8032 E8033

VID DC initialization timeout VID DC receive error VID DC Internal system error VID DC camera mode not confirmed VID DC internal system error VID DC no data transfer possible VID DC no data transfer possible VID DC unknown error VID DC internal system error VID DC brightness nominal value wrong VID DC brightness nominal value rejected VID DC brightness nominal value wrong

29 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E8050

VID DC, Service switch not in default position

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

Videomed DC messages

W8500 W8501 W8502 W8503 E8510 E8511 E8512 E8513 E8514 E8515 E8516 E8517 E8518

VID DC parity error VID DC Internal system error VID DC Internal system error VID DC Internal system error VID DC act. brightness device error VID DC manual gain out of range VID DC black level out of range VID DC video signal out of range VID DC +5V digital out of range VID DC -9V out of range VID DC +5V analog out of range VID DC +15V out of range VID DC CCD sensor defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

Host - Radiation release control messages

W9010 W9011 W9012 W9013 W9014 W9015 W9022 W9023 W9024

Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed Radiation release conditions not allowed

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

Host - Collimator, slot diaphragm messages

E9101 E9102

Slot diaphragm lift magnet too hot Slot diaphragm lift magnet test failed

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

30 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Host - Diamentor messages

E9701 E9702

Diamentor High voltage error Diamentor Wrong test value

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

Host - Power up message

IFFFF

Host - Power up message

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

E0001 User console no communication to host Possible trouble:

User console recognized that the serial port to the host does not work.

Possible help:

Check power supply. Check user console connector and wires. Check host 7-segment display. D1 defective. User console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

F0002 User console internal system error Possible trouble:

User console recognized an internal RAM or EEProm error.

Possible help:

Check power supply. Check user console connector and wires. User console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

31 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W0003 User console Internal watchdog reset Possible trouble:

User console recognized an internal watchdog reset

Possible help:

On sporadic entry in the error memory possibly faults in the system. Check power supply. Check user console connector and wires. User console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

W0100 User console ACK timeout Possible trouble:

Host registered communication error of the user console.

Possible help:

On sporadic entry in the error memory possibly faults in the system. Check connection cable to the user console. Check power supply.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

W0101 User console byte timeout Possible trouble:

Host registered communication error of the user console

Possible help:

On sporadic entry in the error memory possibly faults in the system. Check shieldings to the user console. Check connection cable to the user console. Check power supply.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

W0102 User console receive timeout Possible trouble:

Host registered communication error of the user console

32 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

On sporadic entry in the error memory possibly faults in the system. Check connection cable to the user console. Check power supply.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

E0103 User console initialization timeout Possible trouble:

Host registered a timeout of the user console during initialization of the system.

Possible help:

Check power supply. Check connection cables to the user console. User console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

W0110 User console internal system error Possible trouble:

Host registered an internal error. Host software problem

Possible help:

Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

W0111 User console internal system error Possible trouble:

Host registered an internal system error. Host software problem

Possible help:

Faults in the system, Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

33 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W0113 User console internal system error Possible trouble:

Host registered a telegram with unknown contents. Host software problem.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

W1001 User console internal system error Possible trouble:

Host registered an internal system error. Host software problem.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

W1002 Cassette switch pressed during radiation Possible trouble:

Cassette switch for direct exposure selection was pressed during radiation.

Possible help:

Operating error, Check cables to the cassette switch for insulation damage. Microswitch for direct exposure selection defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

W1003 Cassette switch released dur. radiation Possible trouble:

Cassette switch for direct exposure selection was released during radiation.

Possible help:

Operating error, Cassette holder damaged mechanically.

34 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Check leads to the cassette switch for wire breakage. Microswitch for direct exposure selection defective. Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

I1004 User console reserve keycode 1 Possible trouble:

Host registered reserved key code on the user console.

Possible help:

Sporadic: Communication error Host - User console. Maybe user console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

I1005 User console reserve keycode 2 Possible trouble:

Host registered reserve key code on the user console.

Possible help:

Sporadic: Communication error Host - User console. Maybe user console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

I1006 User console reserve keycode 3 Possible trouble:

Host registered reserve key code on the user console.

Possible help:

Sporadic: Communication error Host - User console. Maybe user console defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

I1007 User console unknown keycode

35 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

Host registered unknown keyboard code on the user console. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free HTML Help documentation generator

I1008 User console ROADMAP pressed Possible trouble:

Host registered pressed key ROADMAP on the user console, but no MEMOSKOP SUB present.

Possible help:

User fault, no error. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

I1009 User console SUB pressed Possible trouble:

Host registered pressed key SUB on the user console, but no MEMOSKOP SUB present.

Possible help:

User fault, no error. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

W1010 Foot switch parameter wrong configurated Possible trouble:

Host registered a wrong configurated Foot switch. Programmed is a single pedal, but radiation was released from the second pedal.

Possible help:

Check the parameter Foot switch and programme it to the correct type (single pedal / double pedal). See also Parameter overview and Parameter Foot switch.

Go to Contents view

36 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

W1101 Operating system driver int. syst. error Possible trouble:

Host registered internal system error in the operating system driver. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

W1102 Radiation release driver int. error Possible trouble:

Host registered internal system error in the driver for radiation release. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

W1103 Gen. driver int. system error Possible trouble:

Host registered internal system error in the host driver for the generator. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

W1104 Fluoro state machine int. system error Possible trouble:

Host registered internal system error in the software control for the generator.

37 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Host software error. Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

W1105 DR state machine internal system error Possible trouble:

Host registered internal system error in the software control for the generator. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

W1106 IDL state machine internal system error Possible trouble:

Host registered internal system error in the software control for the generator. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: iPhone web sites made easy

E1107 Videomed DC radiation release timeout Possible trouble:

Host sends request for radiation release, but the Videomed DC does not answer.

Possible help:

Service switch of the Videomed DC in wrong position. Check power supply for Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

38 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E1108 Gen. learning internal system error Possible trouble:

Host started generator learning and an internal system error occurred. Host software error.

Possible help:

Faults in the system. Execute generator learning once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E1109 Dose adjustment internal system error Possible trouble:

Host started routine for dose rate setting, and an internal system error occurred. Host software error.

Possible help:

Faults in the system. Perform dose rate setting once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

E1110 Cassette mode internal system error Possible trouble:

Host registered internal system error in the software control for executing direct exposure. Software error D1

Possible help:

Faults in the system. Perform direct exposure selection once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

E1111 Acquisition mode internal system error Possible trouble:

Host registered internal system error in the software control on querying the acquisition cable. Host software error.

39 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Faults in the system, check shieldings.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

E1112 Cassette switch simulation int. error Possible trouble:

Host started routine for simulation of the cassette switch, (direct exposure selection without attachment of the cassette holder), and an internal system error occurred. Host software error

Possible help:

Faults in the system. Perform simulation of the cassette switch once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E1113 Operating mode not valid Possible trouble:

Host registered invalid operating mode. Host software error.

Possible help:

Faults in the system.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

E1114 Gen. learning routine int. error Possible trouble:

Host started routine for generator learning, and an internal system error occurred. Host software error

Possible help:

Faults in the system. Perform generator learning once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

W1115 Gen. learning user break

40 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

Host registered break of the generator learning service routine.

Possible help:

Operating error (radiation release switch let go during generator learning). Perform generator learning once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E1116 Diamentor adjustment routine int. error Possible trouble:

Host started service routine for Diamentor adjustment, (area dose product - measuring device) and an internal system error occurred. Host software error.

Possible help:

Faults in the system. Perform Diamentor adjustment once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E1117 KV Offset adjustment routine int. error Possible trouble:

Host started routine for setting the kV offset, and an internal system error occurred. Host software error.

Possible help:

Faults in the system. Perform kV offset adjustment once again.

Go to Contents view Created with the Personal Edition of HelpNDoc: iPhone web sites made easy

E1118 Memoskop timeout Possible trouble:

Host registered Memoskop timeout.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view

41 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E1119 Host - Memoskop dose reduction not received Possible trouble:

Host did not receive the dose reduction from the Memoskop, but was waiting for.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

E1120 RM state machine wrong state pointer Possible trouble:

Host registered internal system error in the software control for the RM state machine. Internal system error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

E1121 SUB statemachine wrong state pointer Possible trouble:

Host registered internal system error in the software control for the SUB statemachine. Internal system error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E1122 3D statemachine wrong state pointer Possible trouble:

Host registered internal system error in the software control for the 3D statemachine. Internal system error.

42 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

W2000 Undefined processor interrupt Possible trouble:

Host registered an undefined processor interrupt, hardware error D1.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate EPub eBooks with ease

E2001 System reboot after watchdog timeout Possible trouble:

Host registered a system reboot, The watchdog on D1 triggered the reboot (system overload).

Possible help:

Faults in the system. Check power supply (voltage and ripple).

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

I2002 Invalid acquisition end interrupt Possible trouble:

Even though the Memoskop acq end line is not active, the host registers level changes on this line.

Possible help:

Faults in the system, check shieldings of the acq end line. Check shieldings to the single tank. Check shieldings of the board D2. Check shieldings of the monitor carriage.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

43 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W2003 Serial port 0, invalid interrupt Possible trouble:

The module for the serial interface to the Videomed DC and service registered an invalid interrupt.

Possible help:

Faults in the system, check shieldings of the serial interface. Possibly D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

W2004 Serial port 1, invalid interrupt Possible trouble:

The module for the serial interface to the console and to the Memoskop registered an invalid interrupt.

Possible help:

Faults in the system, check shieldings of the serial interfaces. Possibly D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E0112 User console does not answer Possible trouble:

Host sent 3 telegrams to the user console, but this did not answer.

Possible help:

Check power supply. Check connection cable to the user console. Possibly user console or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

E2005 Videomed DC, ser. receive line error Possible trouble:

The receive line has a constant level.

Possible help:

Check receive line between D1 and Videomed DC. Check service switches on D4. In normal use, all switches s1 to S8 should be set to OFF. Possibly D4 or D1 defective.

44 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

E2006 Memoskop, ser. receive line error Possible trouble:

The receive line has a constant level.

Possible help:

Check connector "X-Ray Interface" at the Memoskop. Check receive line between D1 and Memoskop. Check contact pins of connector X10 (Monitor cart cable) With installed option DICOM Bridge or 3D Reconstruction, check the Y-cable. Possibly wires broken. Possibly Memoskop or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E2007 User console, ser. receive line error Possible trouble:

The receive line has a constant level.

Possible help:

Check receive line between D1 and user console. Possibly user console or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

W2008 Service, ser. receive line error Possible trouble:

The receive line has a constant level.

Possible help:

Check receive line between D1 and service interface. Possibly service-PC, serial interface or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

E2010 Videomed DC, ser. port init not possible

45 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

The initialization of the serial interface D1 - Videomed DC is not possible.

Possible help:

D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

E2011 Memoskop, ser. port init not possible Possible trouble:

The initialization of the serial interface D1-Memoskop is not possible.

Possible help:

D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E2012 User console, ser. port init not possible Possible trouble:

The initialization of the serial interface D1 - user console is not possible.

Possible help:

D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

E2013 Service, serial port init not possible Possible trouble:

The initialization of the serial interface D1 - service interface îs not possible.

Possible help:

D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E2020 Internal system error Possible trouble:

Error queue overflow.

46 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

.

Host software error. Overloaded system.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

W2030 User console serial port parity error Possible trouble:

Host registered a parity error of the host serial port. (Host - User console)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

W2031 User console serial port overflow error Possible trouble:

Host registered an overflow error of the host serial port. (Host - User console)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

W2032 User console serial port framing error Possible trouble:

Host registered a framing error of the host serial port. (Host - User console)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

47 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W2033 Memoskop serial port parity error Possible trouble:

Host registered a parity error of the Host serial port. (Host - Memoskop)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

W2034 Memoskop serial port overflow error Possible trouble:

Host registered an overflow error of the Host serial port. (Host - Memoskop)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

W2035 Memoskop serial port framing error Possible trouble:

Host registered a framing error of the Host serial port. (Host - Memoskop)

Possible help:

Check shieldings. Check cables. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

W2040 Host serial port buffer overflow Possible trouble:

Host registered a buffer overflow of the host serial port. (Host - User console)

48 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Internal system error Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: iPhone web sites made easy

W2041 Host serial port buffer overflow Possible trouble:

Host registered a buffer overflow of the host serial port. (Host - Memoskop)

Possible help:

Internal system error Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

F2050 Host - filament configuration failed Possible trouble:

During power up, the filament configuration failed.

Possible help:

Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

F2051 Host - filament code corrupted Possible trouble:

During power up, the host registered a corrupted filament code.

Possible help:

Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

E2060 Host - ARC configuration failed Possible trouble:

During power up, the ARC encoder

49 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

(motor positioning system) configuration (Software code) failed. Possible help:

Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

E2061 Host - ARC code corrupted Possible trouble:

During power up, the host registered a corrupted ARC encoder code. (Motor positioning system)

Possible help:

Try download. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

F2070 Host - ADC tests failed Possible trouble:

The AD Converter configuration (Software code) failed.

Possible help:

Check the voltage supply M14 (+/-15V) Check cables from voltage supply M14 to board D1 M14 defective Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

F2071 Host - ADC Busy too long Possible trouble:

The AD Converter on board D1 did not deactivate the ADC busy signal. The analog signals cannot longer be digitalized.

Possible help:

Try download. Board D1 defective.

Go to Contents view

50 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Easily create Qt Help files

E2080 Host - powerfail signal active during bootup Possible trouble:

The powerfail signal was active during bootup.

Possible help:

Check powerfail signal wire (from power supply M14) Power supply M14 defective Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

F2105 Internal system error Possible trouble: .

System queue overflow. Host software error. Overloaded system.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

F2106 Internal system error Possible trouble:

Generator queue overflow Host software error Overloaded system

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

F2201 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

51 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

F2202 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

F2203 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

F2206 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

F2207 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Documentation generator

F2401 Internal system error Possible trouble:

Host software error.

Go to Contents view

52 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

F2402 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

F2403 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate Kindle eBooks with ease

F2404 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate EPub eBooks with ease

F2405 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

F2406 Internal system error Possible trouble:

Host software error.

53 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

F2407 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

F2408 Internal system error Possible trouble:

Host software error.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EBook and documentation generator

W2600 Elapsed time meter initialization error Possible trouble:

New board D1: no fault.

Possible help:

no action necessary.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

W2601 Elapsed time meter uncorrectable error Possible trouble:

The elapsed time meter is uncorrectable deleted. It will be set to 0.

Possible help:

Sporadic: no action necessary. Permanent: board D1 defective.

Go to Contents view

Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

54 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E2601 Elapsed time meter uncorrectable error Possible trouble:

The elapsed time meter is uncorrectable deleted. It will be set to 0.

Possible help:

Sporadic: no action necessary. Permanent: board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Qt Help files

W2602 Elapsed time meter correctable error Possible trouble:

If the SIREMOBIL is switched off exactly at the time when the elapsed time meter was updated, this error comes up.

Possible help:

no action necessary.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

E2603 EEProm area for Elapsed time meter defective Possible trouble:

The EEProm area for the Elapsed time meter is defective.

Possible help:

Sporadic: no action necessary Permanent: Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

W2604 Elapsed time meter writing not more possible Possible trouble:

After an elapsed time of more than 83200 hours, the elapsed time meter stops. That is necessary because the maximum EEProm write cycles are reached.

Possible help:

no action necessary.

Go to Contents view

55 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

W2700 Host - powerfail signal active (sporadic) Possible trouble:

The powerfail signal was active for less than 300ms.

Possible help:

Check mains powerline. Check connector X10. Check power supply M14. Check powerfail wire and it´s connector pins.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

F28zz Delay task internal debug messages Possible trouble:

This group of error messages (from w2800 to W2899) are internal debug messages.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

F29zz Delay task internal debug messages Possible trouble:

This group of error messages (from W2900 to W2999) are internal debug messages.

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

W4009 Internal system error Possible trouble:

Host registered internal system error in the software control for the communication with the Memoskop. Host software error.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view

56 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Benefits of a Help Authoring Tool

W4013 Memoskop acq timeout Possible trouble:

Host registered timeout of the Acq end task.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

W4016 Memoskop wait telegram timeout Possible trouble:

Host registered timeout of the wait telegram task.

Possible help:

Faults in the system. Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E4017 Memoskop busy signal timeout Possible trouble:

Host registered timeout of the busy telegram task,

Possible help:

Sporadic:

Faults in the system. Memoskop timing too slow. Continuous: D1 - Memoskop software version not compatible Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E4018 Memoskop acq start signal timeout Possible trouble:

Host registered timeout of the acquisition start signal task.

Possible help:

Sporadic: Continuous:

Faults in the system. Memoskop overloaded. D1 - Memoskop software version not compatible Check timing between Memoskop start signal (D1.X10.27) and Memoskop acquisition signal (D1.X10.26).

57 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Time offset must be < 40 ms (in fluoroscopy). Possibly Memoskop or D1 defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

W4020 Memoskop send timeout Possible trouble:

Host registered timeout of the send task.

Possible help:

Faults in the system. Internal debug messages.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E4021 Memoskop does not answer Possible trouble:

Host repeats a task 3 times, error E4021 is then entered.

Possible help: "Erase Harddisk".

If the Memoskop C / CE 100 / C-MOD / C-SUB / C-SUB &MOD was replaced, try (Memoskop, Technical setup, menu "Diagnostics" - "Erase Harddisk") Note: Ask your customer before you start erasing ! All images will be lost ! Check for additional error messages E2006 and E4024. Possibly Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E4024 Memoskop initialization timeout Possible trouble:

Host registered a timeout of the Memoskop during the initialization of the system Waiting time for the Memoskop reporting exceeded.

Possible help:

Check the reset switch at the Memoskop. With installed option "DICOM Bridge" or "3D Reconstruction": - Check the programmed Bridge delay (Memoskop Technical setup). If present, disconnect the external "SCSI" cable from the Memoskop. Disconnect the "X-Ray Interface" cable from the Memoskop. Reboot the Memoskop. - If the Memoskop is still not working, try Download of the Memoskop. - Possibly Memoskop defective. - If the Memoskop is working, possibly D1 defective.

58 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Additional info: See Memoskop Service software help file, "Check communication". Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

W4027 Memoskop unexpected reset Possible trouble:

Host registered unexpected reset of the Memoskop.

Possible help:

Memoskop has reset itself independently. Faults in the system. Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

E4028 Memoskop unexpected acq end signal Possible trouble:

Host registered unexpected acq end signal of the Memoskop.

Possible help:

For failure investigation check the error log for additional error messages. Check faults on the Memoskop acquisition cable (D1.X10.27). Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E4029 Memoskop unexpected acq. reset Possible trouble:

Host registered unexpected acq. reset

Possible help:

Check faults on the Memoskop acquisition cable (D1.X10.27). Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EBook and documentation generator

E4030 Memoskop acq. active without start signal

59 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

Host registered acq. signal, before start signal is active.

Possible help:

Check faults on the Memoskop acquisition cable (D1.X10.27). Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Documentation generator

E4031 Memoskop acq. too short time active Possible trouble:

Host registered the acq. signal, but only for a too short time. (10% below the calculated time)

Possible help:

Check faults on the Memoskop acquisition cable (D1.X10.27). Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

W4032 Memoskop acq. too short time active Possible trouble:

Host registered the acq. signal, but only for a too short time. (10% below the calculated time)

Possible help:

Following error. Check for additional VIDEOMED DC errors (e.g. E8000)

Go to Contents view

Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

E4040 Memoskop acq. active dur. initialization Possible trouble:

Host registered active level on the acquisition line of the Memoskop.

Possible help:

Check faults on the Memoskop acquisition line (D1.X10.27). Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view

60 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Easily create EPub books

I4501 Memoskop disk transfer time too long Possible trouble:

Memoskop registered a timeout during the image transfer of the harddisk drive.

Possible help:

Memoskop harddisk drive or harddisk controller busy or defective Check Memoskop. Check connection cables to the Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

I4507 Memoskop no stored images Possible trouble:

There are no stored images, but the operator tied to select one.

Possible help:

no fault, no action necessary.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

W4601 Memoskop no HS_ACK within timeout period Possible trouble:

The Memoskop registered a timeout of the signal HS_ACK.

Possible help:

Sporadic: timing problems Memoskop - Host. Continuos: check cables to Memoskop.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free HTML Help documentation generator

W4602 Memoskop realtime clock malfunction Possible trouble:

The realtime clock sent a wrong date or time.

Possible help:

Set the realtime clock, date and time to the correct values.

61 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

If not possible, Memoskop defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

W4606 Memoskop disk full Possible trouble:

The harddisk drive is full, all stored images are protected.

Possible help:

Unprotect images.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Kindle producer

E4701 Memoskop corrupted patient directory Possible trouble:

The Memoskop registered a corrupted patient directory,

Possible help:

Normal after download of the Memoskop. Format harddisk drive. (see Memoskop technical setup, CTRL + T)

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E4702 Memoskop hardcopy timeout error Possible trouble:

With Viking interface: The Memoskop registered a viking interface hardcopy timeout failure

Possible help:

Select the Memoskop Technical Setup (CTRL + T) and programme the Viking interface to "NO".

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Kindle producer

E4703 Memoskop harddisk drive error Possible trouble:

The Memoskop registered a harddisk drive problem.

Possible help:

Try erase of the MEMOSKOP harddisk.

62 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Note: Ask your customer before you start erasing ! All images will be lost ! Memoskop defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E4704 Memoskop hardware error Possible trouble:

The Memoskop registered a internal hardware error.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

E4705 Memoskop harddisk drive not ready Possible trouble:

The Memoskop registered a harddisk drive problem.

Possible help:

Operator fault. If the image transfer buttons or the "All Pat" button were pressed during startup of the Memoskop, this error comes up. Wait till the SIREMOBIL initialization is finished completely. ( > 30 seconds)

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

E4709 Memoskop battery ram failed Possible trouble:

During the selftest, the Memoskop registered a battery ram error. The Memoskop parameters will be set to the default values.

Possible help:

Sporadic: check the Memoskop parameters. Continuos: Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

63 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E4710 Memoskop D/A 1 LUT error Possible trouble:

During the selftest, the Memoskop registered a digital to analog converter (Monitor A video output) failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

E4711 Memoskop D/A 1 0V error Possible trouble:

During the selftest, the Memoskop registered a digital to analog converter (Monitor A video output) failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E4712 Memoskop D/A 2 LUT error Possible trouble:

During the selftest, the Memoskop registered a digital to analog converter (Monitor B video output) failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E4713 Memoskop D/A 2 0V error Possible trouble:

During the selftest, the Memoskop registered a digital to analog converter (Monitor B video output) failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Documentation generator

64 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E4714 Memoskop D/A 3 LUT error Possible trouble:

During the selftest, Memoskop registered a digital to analog converter failure. (Hardcopy / Video tape recorder video output)

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

E4715 Memoskop D/A 3 0V error Possible trouble:

During the selftest, the Memoskop registered a digital to analog converter failure. (Hardcopy / Video tape recorder video output)

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

E4716 Memoskop memory 1 access failure Possible trouble:

During the selftest, the Memoskop registered a memory access failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

E4717 Memoskop memory 1 LSN access failure Possible trouble:

During the selftest, the Memoskop registered a memory access failure.

Possible help:

Memoskop defective.

Go to Contents view

65 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

E4718 Memoskop memory 2 access failure Possible trouble:

During the selftest, the Memoskop registered a memory access failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E4719 Memoskop memory 3 access failure Possible trouble:

During the selftest, the Memoskop registered a memory access failure.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

E4720 Memoskop no response from 3D PC Possible trouble:

The Memoskop registered an communication error with the 3D PC

Possible help:

Is the 3D PC running? Check the 3D PC and restart the system. Check connectors Memoskop - 3D PC

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate Kindle eBooks with ease

E4721 Memoskop communication error with 3D PC Possible trouble:

The Memoskop received an unknown or corrupted telegram code

Possible help:

Check the cables and shieldings.

Go to Contents view

66 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

E4722 Memoskop disk read or write error in 3D mode Possible trouble:

The Memoskop registered a harddisk drive problem during 3D mode (3 times)

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E4723 Memoskop memory allocation error in 3D mode Possible trouble:

Internal software error

Possible help: Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E4724 Memoskop, error message from 3D PC received Possible trouble:

The 3D PC reported an error message.

Possible help:

Check the error log files of the 3D PC

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

F4808 Memoskop harddisk drive image datas lost Possible trouble:

The Memoskop harddisk drive lost image datas.

Possible help:

Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Benefits of a Help Authoring Tool

67 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

F4809 Memoskop battery ram failed Possible trouble:

The Memoskop battery ram test failed. All user parameters will be set to the default values.

Possible help:

Sporadic: The user setup must be programmed new. Continuos: Memoskop defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

E4999 Host received telegram with wrong error class Possible trouble:

The Host received a telegram with wrong error class.

Possible help:

Fault in the system Internal debug message

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

W5000 Gen. Cassette exposure user break Possible trouble:

During direct exposure, the radiation switch was released.

Possible help:

Operating error or radiation switch (hand switch) defective. Check cable of the handswitch.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

W5001 Gen. mAs actual value too low Possible trouble:

In direct exposure, despite maximum exposure time, the selected mAs value was not reached, or the exposure was ended prematurely.

Possible help:

Check line resistance. Operating error.

Go to Contents view

68 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

W5002 Gen. Sirephos temp. out of range Possible trouble:

Single tank temperature outside the valid range.

Possible help:

In combination with E5015: Check temperature sensor cable for breakage. In combination with E5016: Check temperature sensor cable for short circuit. Possibly hardware error D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

F5003 Gen. ref. voltage +10V out of range Possible trouble:

The reference voltage +10V for the analog circuits of the generator deviate by more than 1 % from the nominal value.

Possible help:

Check +15V power supply on board D1. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

F5004 Gen. Uz ist out of range Possible trouble:

The unregulated intermediate circuit voltage is outside +200V to +320V.

Possible help:

Check voltage Uz on board D1. Measuring divider: 50 V corresponds to 1 V line voltage wrongly programmed on transformer T1. Check fuse D2/F1 and D2/F2. D1 defective. D2 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

E5005 Gen. tried X-Ray on without LIH signal

69 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

Generator attempts radiation without active LIH signal.

Possible help:

Hardware error on board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E5006 Gen. tried X-Ray on without SS relais Possible trouble:

Generator attempts radiation with SS relay switched off.

Possible help:

D1, switch S2 in wrong position. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

E5007 Gen. IH max exceeded Possible trouble:

Generator exceeded maximum filament current.

Possible help:

Actual value acquisition on D1 defective. D1 defective. D2 defective. Sirephos defective (filament transformer).

Go back to Contents view

Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

F5010 Gen. filament inverter short circuit Possible trouble:

Filament circuit inverter has short circuit.

Possible help:

Filament circuit defective (board D2).

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

70 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

F5011 Gen. main inverter short circuit Possible trouble:

Main inverter has short circuit.

Possible help:

Main inverter defective (board D2).

Go to Contents view Created with the Personal Edition of HelpNDoc: Free HTML Help documentation generator

F5012 Gen. kV max exceeded Possible trouble:

The maximum KV values were exceeded.

Possible help:

KV actual value acquisition defective (board D1). Possibly board D2 defective. Oscilloscope + KV and - KV. Values must be symmetrical. (Measuring divider: 10 KV corresponds to 1 V). With asymmetric KV values check actual value cables, possibly Sirephos defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

F5013 Gen. +15 V or -15 V out of range Possible trouble:

The +15V or -15V power supply of the generator is outside the tolerance band.

Possible help:

Check +15V and -15V power supply. Switching power supply defective. Check supply lines of the power supply to board D1. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

F5014 Gen. Sirephos max oil pressure exceeded Possible trouble:

The oil pressure switch on the Sirephos has responded.

Possible help:

Temperature of the Sirephos too high. Check oil pressure switch.

71 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Check cables to the oil pressure switch. In the case of defective oil pressure switch, replace the Sirephos. Monitoring of the oil pressure switch defective (board D1). Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

E5015 Gen. Sirephos temp. lower limit obtained Possible trouble:

Single tank temperature lower temperature range not reached (< 5 deg. Celsius).

Possible help:

In combination with E5002: Check temperature sensor cable for interruption. Temperature of the Sirephos < 5 deg. Celsius - unallowable operating temperature.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

E5016 Gen. Sirephos max temp. exceeded Possible trouble:

Single tank temperature maximum temperature range exceeded (> 75 deg. Celsius).

Possible help:

In combination with E5002: Check temperature sensor cable for short circuit. Temperature acquisition defective (board D1).

Go to Contents view Created with the Personal Edition of HelpNDoc: Benefits of a Help Authoring Tool

W5017 Gen. Sirephos temp. fast level change Possible trouble:

Temperature acquisition on D1 registered abrupt temperature changes.

Possible help:

Wrongly determined AD value. Possibly D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

72 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E5018 Gen. kV offset below lower limit Possible trouble:

The KV offset is below the valid value range.

Possible help:

Perform automatic KV offset adjustment or generator learning. If unsuccessful: D1 or D2 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E5019 Gen. kV/mA courve selection, int. error Possible trouble:

Generator registered the selection of an inadmissible KV/mA curve.

Possible help:

Check parameters for KV/ma curves. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

F5030 Gen. tube current actual value too high Possible trouble:

Actual tube current (Iroe_ist) too high.

Possible help:

Sirephos tube arcing, Sirephos no longer high-voltage-proof. Measure Iroe_ist at D1."mA". Check actual value cables. Measured value acquisition on D1 defective. Sirephos defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

F5031 Gen. tube current actual value too low Possible trouble:

Actual tube current (Iroe_ist) too low.

Possible help:

Measure Iroe_ist at D1. "mA". Check actual value cables.

73 / 166

.

SIREMOBIL Compact / SIREMOBIL Iso-C

Measured value acquisition on D1 defective. Check filament circuit cables (short circuit or interruption). Sirephos defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

F5032 Gen. high voltage actual value too high Possible trouble:

Actual high voltage (sum voltage KV I) too high.

Possible help:

Try KV offset adjustment / Generator adjustment. Measure KV I at D1 (measuring divider 20 kV corresponds to 1 V). Measured value acquisition on D1 defective. Oscilloscope + KV and - KV. Values must be symmetrical (measuring divider: 10 KV corresponds to 1 V). With asymmetrical KV values, check actual value cables, possibly Sirephos defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

F5033 Gen. high voltage actual value too low Possible trouble:

Actual high voltage (KV I) too low.

Possible help:

Try KV offset adjustment / Generator adjustment. Measure KV I at D1. Check actual value cables. Measured value acquisition on D1 defective. Check line resistance. Check connection terminals. Sirephos defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

E5034 Gen. IH actual value too high (standby) Possible trouble:

Actual filament current (IH_ist) in standby mode too high.

Possible help:

Measure IH_ist at D1. Faults in the system. Check actual value cables. Measured value acquisition on D1 defective.

74 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

E5035 Gen. IH actual value too low (standby) Possible trouble:

Actual filament current (IH_ist) in standby mode too low.

Possible help:

Measure IH_ist at D1. Check actual value cables. Measured value acquisition on D1 defective, possibly D2 defective. Filament broken.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring tool

E5080 Gen. learning pushfactor out of tolerance Possible trouble:

During generator learning a pushfactor value out of the tolerance limits was set.

Possible help:

Check the kV offset adjustment. Switch off the unit and repeat generator learning. possibly D2 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E5081 Gen. learning internal error Possible trouble:

During generator learning the internal exposure counter was uncorrect set.

Possible help:

Internal software error. Switch off the unit and repeat generator learning. possibly D2 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

75 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E5090 Gen. Videomed DC not available Possible trouble:

Generator recognized Videomed DC error. If the Videomed DC is defective and no contact to it is possible, this error will be stored after each radiation.

Possible help:

For failure investigations, check the error log for additional error messages. Check Power supply Videomed DC. Check Videomed DC serial port. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

F5901 Gen. tried X-Ray on without XRAY_query signal Possible trouble:

Generator attempts radiation but the signal XRAY_query was not present. Software error D1.

Possible help:

Faults in the system. Sirephos defective D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

E6000 Error log is not initialized Possible trouble:

D1 is new and EEProm is empty. Normal after change of the Board D1.

Possible help:

After exchange of the board D1: no actions necessary. Continuos: Change board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

W6001 Error log, reading error Possible trouble:

EEProm error log could not be read.

76 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Sporadic : timing problem, no actions necessary. Continuous: change board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate Kindle eBooks with ease

W6002 Error log, writing error Possible trouble:

Error datas could not be written and stored in the EEProm.

Possible help:

Sporadic : timing problem, no actions necessary. Continuous: change board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

W6003 Error log, end marking not found Possible trouble:

EEProm error log end marking was not found. The error log will be reorganized. Maybe the error log is lost.

Possible help:

Sporadic : timing problem, no actions necessary. Continuous: change board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

W6004 Error log, checksum error Possible trouble:

EEProm error log checksum is not correct. The error log will be reorganized. Maybe the error log datas are lost.

Possible help:

Sporadic : timing problem, no actions necessary. Continuous: change board D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

77 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

W6005 Error log, correctable error Possible trouble:

EEProm error log recognized a correctable error. The error log will be reorganized. No error datas will be lost.

Possible help:

Sporadic : Power off during writing to the error log. No actions necessary.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

W6006 Error log, error could not be registrated Possible trouble:

An error could not be registrated and could not be stored in the error log. This error datas will be lost.

Possible help:

Sporadic : timing problems, no actions necessary. Continuous: change D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

W6007 Error log, EEProm blocked Possible trouble:

An error could not be registrated and could not be stored in the error log. This error datas will be lost.

Possible help:

Software corrupted or software error Try download.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E6008 Error log, EEProm write failed Possible trouble:

An error could not be stored in the error log. This error datas will be lost.

78 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Board D1 defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E6009 Error log, EEProm write failed Possible trouble:

An error could not be stored in the error log. This error comes up, if the error E6008 could not be stored in the error log.

Possible help:

Board D1 defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate Kindle eBooks with ease

E6100 Generator needs learning Possible trouble:

The following generator learning datas are invalid: . kV offset. Push factor, Filament heating factor.

Possible help:

Start Generator learning.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

E6101 Generator, kV offset out of range Possible trouble:

The kV offset is out of range .

Possible help:

Start Generator learning.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

E6102 Generator, push factor out of range Possible trouble:

The push factor is out of range .

79 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Start Generator learning.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E6103 Generator, filament heating factor out of range Possible trouble:

The kV offset is out of range .

Possible help:

Start Generator learning.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E6110 Infodata checksum error Possible trouble:

The infodata checksum , stored in the EEProm is wrong.

Possible help:

Sporadic : timing problems, no actions necessary. Continuous: change D1.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EBook and documentation generator

E6200 Parameter checksum error Possible trouble:

The EEProm checksum for the parameter area is wrong

Possible help:

Load / reload parameter file to EEProm. If not possible: change D1. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E6201 Parameter Camera rotation -180 deg. out of range Possible trouble:

The EEProm data for camera rotation -180 degree position is out of range.

80 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Adjust / readjust camera rotation. Camera rotation mechanic faults. Camera actual value uncorrect (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E6202 Parameter Camera rotation 0 deg. out of range Possible trouble:

The EEProm data for camera rotation 0 degree position is out of range.

Possible help:

Adjust / readjust camera rotation. Camera rotation mechanic faults. Camera actual value uncorrect (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E6203 Parameter Camera rotation +180 deg. out of range Possible trouble:

The EEProm data for camera rotation +180 degree position is out of range.

Possible help:

Adjust / readjust camera rotation. Camera rotation mechanic faults. Camera actual value uncorrect (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

E6204 Parameter Camera rotation max position out of range Possible trouble:

The EEProm data for camera rotation maximum position (+220 degree position / -220 degree position) is out of range.

Possible help:

Adjust / readjust camera rotation.

81 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Camera rotation mechanic faults. Camera actual value uncorrect (potentiometer...) Maybe D1 defective. See also Parameter overview Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

E6205 Parameter X-iris minimal position out of range Possible trouble:

The EEProm data for x-iris minimal position (mechanical end position) is out of range.

Possible help:

Adjust / readjust x-iris X-iris mechanic faults. X-iris actual value not correct (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

E6206 Parameter X-iris maximum position out of range Possible trouble:

The EEProm data for x-iris maximum position (mechanical end position) is out of range.

Possible help:

Adjust / readjust x-iris. X-iris mechanic faults. X-iris actual value not correct (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

E6207 Parameter X-iris close position out of range Possible trouble:

The EEProm data for x-iris close position (smallest size if the x-iris will be closed with the x-iris close button) is out of range.

Possible help:

Adjust / readjust x-iris.

82 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

X-iris mechanic faults. X-iris actual value not correct (potentiometer...) Maybe D1 defective. See also Parameter overview Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

E6208 Parameter X-iris full format position out of range Possible trouble:

The EEProm data for x-iris full format position (II full format size) is out of range.

Possible help:

Adjust / readjust x-iris. X-iris mechanic faults. X-iris actual value not correct (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

E6209 Parameter X-iris zoom format position out of range Possible trouble:

The EEProm data for x-iris zoom format position (II zoom format size) is out of range.

Possible help:

Adjust / readjust x-iris. X-iris mechanic faults. X-iris actual value not correct (potentiometer...) Maybe D1 defective. See also Parameter overview

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E6210 Parameter X-iris cassette exposure position out of range Possible trouble:

The EEProm data for x-iris cassette exposure position is out of range.

Possible help:

Adjust / readjust x-iris. X-iris mechanic faults. X-iris actual value not correct (potentiometer...)

83 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Maybe D1 defective. See also Parameter overview Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E6211 Parameter brightness nominal value out of range Possible trouble:

The EEProm data for brightness nominal value is out of range.

Possible help:

Set the parameter brightness nominal value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Help generator

E6212 Parameter kV / mA curve normal wrong Possible trouble:

The EEProm data for kv / mA curve normal is wrong.

Possible help:

Set the parameter kv / mA curve normal to a correct value. The programmed curves are described in the user manual. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E6213 Parameter kV / mA curve push wrong Possible trouble:

The EEProm data for kv / mA curve push is wrong.

Possible help:

Set the parameter kv / mA curve push to a correct value. The programmed curves are described in the user manual. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EPub and documentation editor

84 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E6214 Parameter settings for stop mode wrong Possible trouble:

The EEProm data for settings for stop mode is wrong.

Possible help:

Set the parameter settings for stop mode to a correct value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

E6215 Parameter Limitation of skindoserate wrong Possible trouble:

The EEProm data for Limitation of skindoserate is wrong.

Possible help:

DHHS countries: Adjust / readjust the limitation of skindoserate. Other countries: Set the value to 50. (no limitation) See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E6216 Parameter Buzzer and blocktime settings wrong Possible trouble:

The EEProm data for Buzzer and blocktime settings, Buzzer/Blocking Timing, is wrong.

Possible help:

Set the parameter Buzzer and blocktime settings, Buzzer/Blocking timing to a correct value. ( depending on the national regulations.) See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E6217 Parameter Buzzer and blocktime settings wrong Possible trouble:

The EEProm data for Buzzer and blocktime settings, Buzzer Mode (FL/IFL/DR), is wrong.

Possible help:

Set the parameter Buzzer and blocktime settings,

85 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Buzzer Mode (FL/IFL/DR) to a correct value. ( depending on the national regulations.) See also Parameter overview. Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

E6218 Parameter Buzzer and blocktime settings wrong Possible trouble:

The EEProm data for Buzzer and blocktime settings, Buzzer mode (CEX) is wrong.

Possible help:

Set the parameter Buzzer and blocktime settings, Buzzer mode (CEX) to a correct value. ( depending on the national regulations.) CEX = Cassette Exposure See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create iPhone documentation

E6219 Parameter Diamentor existing wrong Possible trouble:

The EEProm data for Diamentor, Diamentor existing, is wrong.

Possible help:

Set the parameter Diamentor, Diamentor existing to a correct value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

E6220 Parameter Diamentor constant wrong Possible trouble:

The EEProm data for Diamentor, Constant is wrong.

Possible help:

Set the parameter Diamentor to the correct value. (Qt x Kg) Readjust the Diamentor. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

86 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E6221 Parameter Error log is wrong Possible trouble:

The EEProm data for Error log is wrong.

Possible help:

Set the parameter Error log to a correct value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Web Help generator

E6222 Parameter Laser existing is wrong Possible trouble:

The EEProm data for Laser existing is wrong.

Possible help:

Set the parameter Laser, Laser existing to the correct value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E6223 Parameter Monitoring timeout is wrong Possible trouble:

The EEProm data for Monitoring timeout is wrong.

Possible help:

Set the parameter Monitoring timeout to the correct value. See also Parameter overview.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

W6224 Parameter Test enable is wrong Possible trouble:

The EEProm data for Test enable is wrong. This parameter is only for internal debug routines and cannot be set from service engineers.

Possible help:

no action necessary, if this parameter is wrong, it will be automatically set to a valid value.

Go to Contents view

87 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E6225 Parameter Max error messages is wrong Possible trouble:

The EEProm data for maximum errors is wrong.

Possible help:

Set the parameter Maximum errors to the correct value. See also Parameter overview , Settings for service PC

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce electronic books easily

E6231 Parameter TV-Iris position not plausible Possible trouble:

The EEProm datas for TV-Iris opening are not plausible.

Possible help:

Readjust the doserate. See also Doserate adjustment.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

E6232 Parameter Zoom area factor out of range Possible trouble:

The EEProm data for zoom area factor is out of range.

Possible help:

Set the parameter Zoom factor to the correct value. See also Parameter Dose adjustment

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E6233 Parameter MID-HIGH Factor out of range Possible trouble:

The EEProm data for MID-HIGH factor is out of range.

Possible help:

Set the parameter MID-HIGH Factor to the correct value. See also Parameter Dose adjustment

Go to Contents view

88 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Produce electronic books easily

E6234 Parameter MID-LOW Factor out of range Possible trouble:

The EEProm data for MID-LOW factor is out of range.

Possible help:

Set the parameter MID-LOW Factor to the correct value. See also Parameter Dose adjustment

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EBook and documentation generator

E6235 Parameter Adjustment mode out of range Possible trouble:

The EEProm data for Adjustment mode is out of range.

Possible help:

Set the parameter Adjustment mode to the correct value. See also Parameter Dose adjustment

Go to Contents view Created with the Personal Edition of HelpNDoc: Free PDF documentation generator

E6236 Parameters Settings for "STOP MODE" out of range Possible trouble:

The EEProm datas Settings for "STOP Parameter" are out of range.

Possible help:

Set the parameters Settings for Stop parameter to the correct values. See also Parameters Dose adjustment, , Parameters X-Iris or Parameters

Camera rotation. Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E6301 EEProm reading error Possible trouble:

The processor system tried reading dates from the EEProm outside the addressable EEProm size.

Possible help:

internal system error, internal debug message.

Go to Contents view

89 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

E6302 EEProm writing error Possible trouble:

The processor system tried writing dates to the EEProm outside the addressable EEProm size .

Possible help:

internal system error, internal debug message

Go to Contents view Created with the Personal Edition of HelpNDoc: Create cross-platform Qt Help files

E6303 EEProm reading blocked Possible trouble:

The processor system tried reading dates from the EEProm but reading was not possible.

Possible help:

internal system error, internal debug message. try rebooting maybe board D1 defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

E6304 EEProm writing blocked Possible trouble:

The processor system tried to write datas to the EEProm but could not.

Possible help:

Sporadic: internal system error, internal debug message try rebooting Continuous: Board D1 defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E6305 EEProm writing failed Possible trouble:

The processor system tried to write datas to the EEProm but could not.

90 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

internal system error, internal debug message try rebooting maybe board D1 defective

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EPub books

E7000 Slot diaphragm movement blocked Possible trouble:

The host activated the slot diaphragm motor, but the actual value did not change.

Possible help: Sporadic:

Permanent:

Slot diaphragm mechanic needs too much force. Actual value increment disk defective. Actual value increment disk mechanically blocked. Check wires. D1 defective. See Error E7001.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub producer

E7001 Slot diaphragm movement blocked Possible trouble:

The host activated the slot diaphragm motor, but the actual value did not change.

Possible help:

Slot diaphragm mechanically blocked. Actual value increment disk defective. Actual value increment disk mechanically blocked. Check wires. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

E7002 Slot diaphragm movement to wrong direction Possible trouble:

The host detected a movement of the slot diaphragm to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Incremental encoder wrong connected.

91 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Board D1 defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

E7003 Slot diaphragm movement to wrong direction Possible trouble:

The host detected a movement of the slot diaphragm to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Incremental encoder wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create EBooks

E7004 Slot diaphragm incremental encoder out of range Possible trouble:

The increment / decrement counter of the slot diaphragm positioning system is below the minimum value.

Possible help:

Slot diaphragm moved by hand ? Wrong slot diaphragm parameters. Incremental encoder defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create CHM Help documents

E7006 Slot diaphragm incremental encoder out of range Possible trouble:

The increment / decrement counter of the slot diaphragm positioning system is above the maximum value.

Possible help:

Slot diaphragm moved by hand ? Wrong slot diaphragm parameters. Incremental encoder defective or wrong connected. Board D1 defective.

Go to Contents view

92 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free EPub producer

E7008 Slot diaphragm unexpected movement Possible trouble:

The slot diaphragm motor was off, but the positioning system detected a movement.

Possible help:

Slot diaphragm moved by hand ? Motor driver short circuit, board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

E7010 Slot diaphragm, programmed limit exceeded Possible trouble:

The slot diaphragm exceeded the programmed limit

Possible help:

Slot diaphragm moved by hand? Check the programmed slot diaphragm parameters Check the slot diaphragm adjustment. Check the incremental encoder wires. Incremental encoder defective. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E7100 Slot diapragm rotation blocked Possible trouble:

The host activated the slot diapragm rotation, but the actual value did not change.

Possible help: Sporadic:

Permanent:

Slot diaphragm mechanic needs too much force. Lifting magnet blocked. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. D1 defective. See Error E7101.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

93 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E7101 Slot diaphragm rotation blocked Possible trouble:

The host activated the slot diaphragm rotation, but the actual value did not change.

Possible help:

Slot diaphragm mechanic needs too much force. Lifting magnet blocked. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. Collimator defective. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate Kindle eBooks with ease

E7102 Slot diaphragm rotation to wrong direction Possible trouble:

The host detected a movement of the slot diaphragm rotation to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

E7103 Slot diaphragm rotation to wrong direction Possible trouble:

The host detected a movement of the slot diaphragm rotation to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

94 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E7104 Slot diaphragm rotation actual value out of range Possible trouble:

The actual value of the slot diaphragm rotation is below the minimum value. (AD converter value)

Possible help:

Slot diaphragm rotation moved by hand ? Wrong slot diaphragm rotation parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EPub and documentation generator

E7106 Slot diaphragm rotation actual value out of range Possible trouble:

The actual value of the slot diaphragm rotation is above the maximum value. (AD converter value)

Possible help:

Slot diaphragm rotation moved by hand ? Wrong slot diaphragm rotation parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create PDF Help documents

E7108 Slot diaphragm rotation unexpected movement Possible trouble:

The slot diaphragm rotation was not activated, but the system detected a movement.

Possible help:

Slot diaphragm rotation moved by hand ? Lifting magnet driver short circuit, board D1 defective. Motor driver short circuit, board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

E7110 Slot diaphragm rotation limits exceeded Possible trouble:

The slot diaphragm rotation exceeded the programmed limit

95 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Slot diaphragm moved by hand? Check the programmed slot diaphragm rotation parameters Check the slot diaphragm rotation adjustment. Check the incremental encoder wires. Incremental encoder defective. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create help files for the Qt Help Framework

E7150 Slot diaphragm rotation index not found Possible trouble:

The slot diaphragm rotation index position was not found during initialization.

Possible help:

Check the index switch Check the index switch wires. Check the motor. Check the motor wires. Check the mechanical movement. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

E7151 Slot diaphragm rotation 0 degr. pos. not found Possible trouble:

The slot diaphragm rotation 0 degree position was not found during initialization.

Possible help:

Check the mechanical movement. Check the motor. Check the motor wires. Check the actual value potentiometer wires. Check the actual value potentiometer. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Help documents

E7152 Slot diaphragm rotation unexpected position values Possible trouble:

The host started the slot diaphragm rotation movement, but the position values did not change.

Possible help:

Check incremental encoder.

96 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Check incremental encoder wires. Check the motor. Check the motor wires. D1 defective. Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

E7200 Camera rotation blocked Possible trouble:

The host activated the camera rotation, but the actual value did not change.

Possible help: Sporadic:

Permanent:

Camera rotation, mechanic needs too much force. Cable drum is blocking the camera rotation. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. D1 defective. See Error E7201.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

E7201 Camera rotation blocked Possible trouble:

The host activated the camera rotation, but the actual value did not change.

Possible help:

Camera rotation, mechanic needs too much force. Cable drum is blocking the camera rotation. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. Camera rotation (I.I. optical system) defective. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

E7202 Camera rotation to wrong direction Possible trouble:

The host detected a movement of the camera rotation rotation to the wrong direction.

97 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

E7203 Camera rotation to wrong direction Possible trouble:

The host detected a movement of the camera rotation rotation to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free HTML Help documentation generator

E7204 Camera rotation actual value out of range Possible trouble:

The actual value of the camera rotation is below the minimum value. (AD converter value)

Possible help:

Camera rotation moved by hand ? Wrong camera rotation parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write eBooks for the Kindle

E7206 Camera rotation actual value out of range Possible trouble:

The actual value of the camera rotation is above the maximum value. (AD converter value)

Possible help:

Camera rotation moved by hand ? Wrong camera rotation parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

98 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E7208 Camera rotation unexpected movement Possible trouble:

The camera rotation was not activated, but the system detected a movement.

Possible help:

Camera rotation moved by hand ? Motor driver short circuit, board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

E7210 Camera rotation limits exceeded Possible trouble:

The camera rotation exceeded the programmed limit

Possible help:

Camera rotation moved by hand? Check the programmed camera rotation parameters Check the camera rotation adjustment. Check the actual value potentiometer wires. Actual value potentiometer defective. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce Kindle eBooks easily

E7300 X-Iris blocked Possible trouble:

The host activated the X-Iris movement, but the actual value did not change.

Possible help: Sporadic:

Permanent:

X-Iris, mechanic needs too much force. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. D1 defective. See Error E7301.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

99 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E7301 X-Iris blocked Possible trouble:

The host activated the X-Iris movement, but the actual value did not change.

Possible help:

X-Iris, mechanic needs too much force. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. Collimator defective. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

E7302 X-Iris movement to wrong direction Possible trouble:

The host detected a movement of the X-Iris to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

E7303 X-Iris movement to wrong direction Possible trouble:

The host detected a movement of the X-Iris to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

100 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E7304 X-Iris actual value out of range Possible trouble:

The actual value of the X-Iris is below the minimum value. (AD converter value)

Possible help:

X-Iris moved by hand ? Wrong X-Iris parameters? Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

E7306 X-Iris actual value out of range Possible trouble:

The actual value of the X-Iris is above the maximum value. (AD converter value)

Possible help:

X-Iris moved by hand ? Wrong X-Iris parameters programmed? Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

E7308 X-Iris unexpected movement Possible trouble:

The X-Iris movement was not activated, but the system detected a movement.

Possible help:

X-Iris moved by hand ? Motor driver short circuit, board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E7309 X-Iris wrong position Possible trouble:

If the error 7300 to 7308 were detected, this error is displayed after each radiation release.

101 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

User warning: The user must delete this error before radiation can be released again. Possible help:

For more informations, read out the previous error messages. Check the X-Iris adjustment. Check the actual value potentiometer. Check the actual value potentiometer wires.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

E7310 X-Iris limits exceeded Possible trouble:

The X-Iris exceeded the programmed limit

Possible help:

X-Iris moved by hand? Check the programmed X-Iris parameters Check the X-Iris adjustment. Check the actual value potentiometer wires. Actual value potentiometer defective. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

I7350 X-Iris stored positions not plausible Possible trouble:

The X-Iris position values are not plausible

Possible help:

Check the programmed X-Iris parameters Check the X-Iris adjustment. Check the actual value potentiometer wires. Actual value potentiometer defective. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Produce online help for Qt applications

I7351 X-Iris screen display values not accessible Possible trouble:

The X-Iris screen display values cannot be read.

Possible help:

Check the programmed X-Iris screen display parameters Check the X-Iris screen display adjustment.

102 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Type of Memoskop does not support the X-Iris screen display functions. Parameter value for Type of diaphragm wrong set. Board D1 defective. Go to Contents view

Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

I7352 X-Iris stored position values not accessible Possible trouble:

The X-Iris position values cannot be read.

Possible help:

Check the programmed X-Iris parameters Check the X-Iris adjustment. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Benefits of a Help Authoring Tool

E7353 X-Iris, I.I. format not plausible Possible trouble:

The X-Iris values for I.I format are not plausible.

Possible help:

Check the programmed X-Iris parameters Check the X-Iris adjustment. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create HTML Help documents

E7400 TV-Iris movement blocked Possible trouble:

The host activated the TV-Iris movement, but the actual value did not change.

Possible help: Sporadic:

Permanent:

TV-Iris, mechanic needs too much force. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. D1 defective. See Error E7401.

Go to Contents view

103 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: iPhone web sites made easy

E7401 TV-Iris movement blocked Possible trouble:

The host activated the TV-Iris movement, but the actual value did not change.

Possible help:

TV-Iris, mechanic needs too much force. Actual value potentiometer defective. Actual value potentiometer mechanically blocked. Check wires. TV-Iris (I.I. optical system) defective. D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E7402 TV-Iris movement to wrong direction Possible trouble:

The host detected a movement of the TV-Iris to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E7403 TV-Iris movement to wrong direction Possible trouble:

The host detected a movement of the TV-Iris to the wrong direction.

Possible help:

Motor wires wrong connected. Check motor wires. Cable harness, pins wrong connected. Actual value potentiometer wrong connected. Board D1 defective.

Go to Contents view

104 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

E7404 TV-Iris actual value out of range Possible trouble:

The actual value of the TV-Iris is below the minimum value. (AD converter value)

Possible help:

TV-Iris moved by hand ? Wrong doserate parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

E7406 TV-Iris actual value out of range Possible trouble:

The actual value of the TV-Iris is above the maximum value. (AD converter value)

Possible help:

TV-Iris moved by hand ? Wrong doserate parameters. Actual value potentiometer defective or wrong connected. Board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

E7408 TV-Iris unexpected movement Possible trouble:

The TV-Iris movement was not activated, but the system detected a movement.

Possible help:

TV-Iris moved by hand ? Motor driver short circuit, board D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free CHM Help documentation generator

E7409 TV-Iris wrong position

105 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

If the error 7400 to 7408 were detected, this error is displayed after each radiation release. User warning: The user must delete this error before radiation can be released.

Possible help:

For more informations, read out the previous error messages. Check the doserate adjustment. Check the actual value potentiometer. Check the actual value potentiometer wires.

Go to Contents view Created with the Personal Edition of HelpNDoc: Qt Help documentation made easy

E7410 TV-Iris limits exceeded Possible trouble:

The limits for the TV-Iris movement were exceeded.

Possible help:

Check programmed parameters Check doserate adjustment. Check actual value potentiometer. Check actual value potentiometer wires.

Go to Contents view Created with the Personal Edition of HelpNDoc: News and information about help authoring tools and software

E8000 Videomed DC initialization timeout Possible trouble:

Host registered timeout during initialization of the Videomed DC.

Possible help:

Check power supply of the Videomed DC. Check cables to the Videomed DC. Possibly Videomed DC or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

W8002 Videomed DC receive error Possible trouble:

Host registered a receive timeout of the Videomed DC.

Possible help:

For sporadic entry in the error memory possibly faults in the system. Check connection cable to the Videomed DC.

106 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Check power supply of the Videomed DC. Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

W8003 Videomed DC Internal system error Possible trouble:

Host registered unknown ID of the Videomed DC.

Possible help:

For sporadic entry in the error memory possibly faults in the system. Check connection cable to the Videomed DC. Check power supply of the Videomed DC.

Go to Contents view Created with the Personal Edition of HelpNDoc: Write EPub books for the iPad

E8005 Videomed camera mode not confirmed Possible trouble:

Host registered no acknowledgement of the Videomed DC on operating mode selection.

Possible help:

DIP switch of the Videomed DC in wrong position. Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easily create Qt Help files

W8006 Videomed DC Internal system error Possible trouble:

Host registered communication error of the Videomed DC

Possible help:

For sporadic entry in the error memory possibly faults in the system. Check connection cable to the Videomed DC. Check power supply of the Videomed DC.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create HTML Help, DOC, PDF and print manuals from 1 single source

107 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E8007 Videomed DC no data transfer possible Possible trouble:

Host registered the failure of the Bright Actual telegrams of the Videomed DC.

Possible help:

Host selects KV Stop automatically Emergency operation of the system may possible. Check power supply of the Videomed DC. Check cables to the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Benefits of a Help Authoring Tool

E8008 Videomed DC no data transfer possible Possible trouble:

Host registered the failure of the Bright Actual telegrams of the Videomed DC.

Possible help:

Host selects KV Stop automatically Emergency operation of the system may possible. Check power supply of the Videomed DC. Check cables to the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free iPhone documentation generator

E8020 Videomed DC unknown error Possible trouble:

Host registered unknown error number of the Videomed DC.

Possible help:

Internal debug message.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

W8030 Videomed DC Internal system error Possible trouble:

Host registered communication error of the Videomed DC.

108 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

For sporadic entry in the error memory possibly faults in the system. Check connection cable to the Videomed DC. Check power supply of the Videomed DC.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free PDF documentation generator

E8031 Videomed DC brightness nominal value wrong Possible trouble:

Host sent the brightness nominal value but the VIDEOMED DC answered with another brightness nominal value.

Possible help:

Host selects KV Stop automatically Emergency operation of the system may possible. Check parameter brightness nominal value. Check power supply of the Videomed DC. Check cables to the Videomed DC.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E8032 Videomed DC brightness nominal value rejected Possible trouble:

Host sent the brightness nominal value but the VIDEOMED DC rejected it.

Possible help:

Host selects KV Stop automatically Emergency operation of the system may possible. Communication problems. Check power supply of the Videomed DC. Check cables to the Videomed DC.

Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E8033 Videomed DC brightness nominal value wrong Possible trouble:

VIDEOMED DC sent an incorrect brightness nominal value to the host.

Possible help:

Host selects KV Stop automatically Emergency operation of the system may possible. Communication problems. Check power supply of the Videomed DC.

109 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Check cables to the Videomed DC. Go to Contents view Created with the Personal Edition of HelpNDoc: What is a Help Authoring tool?

E8050 VID DC, Service switch not in default position Possible trouble:

If the VIDEOMED DC service switch D4.S301.1 to D4.S301.8 are not in position off, this error occurs.

Possible help:

Set all to position off.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free help authoring environment

W8500 Videomed DC parity error Possible trouble:

Videomed DC reports parity error.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Create iPhone web-based documentation

W8501 Videomed DC Internal system error Possible trouble:

The Videomed DC records the wrong order of messages.

Possible help:

Sporadically: Internal system errors - check the connection cables/ Videomed DC-D1. Possibly Videomed DC or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy CHM and documentation editor

W8502 Videomed DC Internal system error Possible trouble:

The Videomed DC records a message not defined.

110 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Sporadically: Internal system errors - check the connection cables /Videomed DC-D1. Possibly Videomed DC or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free EBook and documentation generator

W8503 Videomed DC Internal system error Possible trouble:

The Videomed DC records a message with a wrong content.

Possible help:

Sporadically: Internal system errors - check the connection cables/ Videomed DC-D1. Possibly Videomed DC or D1 defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured Kindle eBooks generator

E8510 Videomed DC act. brightness device error Possible trouble:

Videomed DC Bright Actual value acquisition defective.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Free Qt Help documentation generator

E8511 Videomed DC manual gain out of range Possible trouble:

Videomed DC automatic adjustment of the manual gain not possible or outside the valid tolerance band.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy EBook and documentation generator

111 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

E8512 Videomed DC black level out of range Possible trouble:

Videomed DC black level of the video signal outside the valid tolerance band.

Possible help:

Check black level in the video signal of the camera. Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E8513 Videomed DC video signal out of range Possible trouble:

Videomed DC B signal in the BAS outside the valid tolerance range.

Possible help:

Check video cable between Videomed DC and Memoskop Check 75 ohm termination of the Videomed DC. Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EBook editor

E8514 Videomed DC +5V digital out of range Possible trouble:

Videomed DC +5V power supply for the digital part outside the valid tolerance range.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured EPub generator

E8515 Videomed DC -9V out of range Possible trouble:

Videomed DC -9V power supply for analog circuits outside the valid tolerance range.

112 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Single source CHM, PDF, DOC and HTML Help creation

E8516 Videomed DC +5V analog out of range Possible trouble:

Videomed DC +5V power supply for the analog circuits outside the valid tolerance range.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Full-featured multi-format Help generator

E8517 Videomed DC +15V out of range Possible trouble:

Videomed DC +15V power supply for the analog circuits outside the valid tolerance range.

Possible help:

Check power supply of the Videomed DC. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Easy to use tool to create HTML Help files and Help web sites

E8518 Videomed DC CCD sensor defective Possible trouble:

Videomed DC CCD Sensor defective.

Possible help:

Test LED for test of the Videomed DC defective or CCD sensor defective. Videomed DC defective.

Go to Contents view Created with the Personal Edition of HelpNDoc: Generate EPub eBooks with ease

W9010 Radiation release conditions not allowed

113 / 166

SIREMOBIL Compact / SIREMOBIL Iso-C

Possible trouble:

Release conditions for radiation release are not fulfilled. Error detected with radiation. SW RELEASE signal active, but neither HAND/MANUAL or FOOT signal detected

Possible help:

Error in the radiation release circuit. Check the supply voltage +15V at board D3 (D30). Check the signals SW RELEASE, HAND/MANUAL and FOOT (see circuit diagram, page RADIATION RELEASE). If the signal SW RELEASE is active (approx. +5V), board D3 (D30) is defective. If the signal SW RELEASE is not active ( or