R NINGBO DAVID SERVICE MANUAL YP-2000 INFANT INCUBATOR NINGBO DAVID MEDICAL DEVICE CO., LTD. ADD: NO.158 DAQING ROA
Views 210 Downloads 0 File size 7MB
R
NINGBO DAVID
SERVICE MANUAL
YP-2000 INFANT INCUBATOR
NINGBO DAVID MEDICAL DEVICE CO., LTD. ADD: NO.158 DAQING ROAD, SHIPU, NINGBO, ZHEJIANG PROVINCE, CHINA POST CODE: 315731
FAX: 0086-574-65962111
MARKETING CENTER: NO. 100, JINGHUA ROAD, HI-TECH INDUSTRIAL DEVELOPMENT ZONE, NINGBO, CHINA POST CODE: 315000 TEL: 0086-574-87800008, 87800009 E-MAIL: [email protected]
FAX: 0086-574-87801111 [email protected]
WEBSITE: http://www.nbdavid.com EUROPEAN REPRESENTATIVE: SHANGHAI INTERNATIONAL HOLDING CORP. GMBH (EUROPE) ADD: EIFFESTRASSE 80, 20537 HAMBURG, GERMANY TEL: 0049-40-2513175
FAX: 0049-40-255726 EDITION/REVISION A/0
Guidance and manufacture’s declaration – electromagnetic emissionsfor all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission The YP-2000 Infant Incubator is intended for use in the electromagnetic environment specified below. The customer of the user of the YP-2000 Infant Incubator should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
The YP-2000 Infant Incubator uses RF energy
CISPR 11
only for its internal function. Therefore, its RF Group 1
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11 Harmonic emissions IEC 61000-3-2
Class A Class A
all
establishments,
other
than
domestic
establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/ flicker emissions
The YP-2000 Infant Incubator is suitable for use in
Complies
IEC 61000-3-3
Page 1of 4
Guidance and manufacture’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity The YP-2000 Infant Incubator is intended for use in the electromagnetic environment specified below. The customer or the user of YP-2000 Infant Incubator should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood,
discharge (ESD)
±8 kV air
±8 kV air
concrete or ceramic tile. If
IEC 61000-4-2
floor are covered with synthetic
material,
the
relative humidity should be at least 30%. Electrical fast
±2 kV for power supply
±2kV for power
Mains
transient/burst
lines
supply lines
should be that of a typical
IEC 61000-4-4
power
commercial
quality
or
hospital
power
quality
environment. Surge
±1 kV differential mode
±1 kV differential
Mains
IEC 61000-4-5
±2 kV common mode
mode
should be that of a typical
±2 kV common mode
commercial
or
hospital
power
quality
environment. Voltage dips, short
95% dip in UT)
should be that of a typical
for 0.5 cycle
for 0.5 cycle
commercial
power supply input
or
hospital
environment. If the user of
lines
40% UT
40% UT
the
IEC 61000-4-11
(60% dip in UT)
(60% dip in UT)
Incubator
requires
for 5 cycles
for 5 cycles
continued
operation
YP-2000
during
power
mains
70% UT
70% UT
interruptions,
(30% dip in UT)
(30% dip in UT)
recommended
for 25 cycles
for 25 cycles
YP-2000 Infant Incubator
95% dip in UT)
supply or a battery.
for 5 sec
for 5 sec
be
NOTE
Infant
UT is the a.c. mains voltage prior to application of the test level.
Page 2 of 4
powered
it
is
that from
the an
power
Guidance and manufacture’s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS that are LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity The YP-2000 Infant Incubator is intended for use in the electromagnetic environment specified below. The customer or the user of YP-2000 Infant Incubator should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the YP-2000 Infant Incubator, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance
Conducted RF
3Vrms
IEC 61000-4-6
150 kHz to 80 MHz
3 Vrms
outside ISM bands a
10 Vrms
⎡ 3 .5 ⎤ d =⎢ ⎥ P ⎣ V1 ⎦
10 Vrms
150 kHz to 80 MHz in ISM bands
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
10 V/m 80 MHz to 800 MHz
800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). b
Field strengths from fixed RF transmitters, as c determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.d Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a The ISM(industrial, scientific and medical) bands between 150kHz and 80MHz are 6.765MHz to 6.795MHz; 13.553 MHz to 14.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. b The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the frequency range 80 MHz to 2.5GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges. c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the YP-2000 Infant Incubator is used exceeds the applicable RF compliance level above, the YP-2000 Infant Incubator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the YP-2000 Infant Incubator . d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Page 3 of 4
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the YP-2000 Infant Incubator The YP-2000 Infant Incubator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the YP-2000 Infant Incubator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the YP-2000 Infant Incubator as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter (m)
Rated maximum
150 kHz to 80 MHz
150 kHz to 80 MHz
80 MHz to 800
output power
outside ISM bands
in ISM bands
MHz
of transmitter (W)
⎡ 3 .5 ⎤ d =⎢ ⎥ P ⎣ V1 ⎦
0.01
0.1167
0.12
0.12
0.23
0.1
0.3689
0.3795
0.3795
0.7273
1
1.1667
1.2
1.2
2.3
10
3.6893
3.7947
3.7947
7.2732
100
11.6667
12
12
23
800 MHz to 2.5 GHz
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 The ISM(industrial, scientific and medical) bands between 150kHz and 80MHz are 6.765MHz to 6.795MHz; 13.553 MHz to 14.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. NoTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could couse interference if it is inadvertently brought into patient areas. NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Page 4 of 4
SERVICE MANUAL FOR INFANT INCUBATOR
WARRANTY The product being described in this manual is warranted against defects in materials or workmanship for one year from the date of shipment, with the following exceptions. 1. All consumable and disposable products are guaranteed to be free from defects upon shipment only. 2. Calibrations are considered normal maintenance and are not included in the 1-year warranty. During the warranty period any defective parts other than those listed above will be replaced at no charge to the customer. This warranty is rendered void and our company cannot be held liable for conditions resultant therefrom if: 1. Damage to the unit is incurred as a result of mishandling. 2. The customer fails to maintain the unit in a proper manner. 3. The customer uses any parts, accessories, or fittings not specified or sold by our company. 4. Sale or service is performed by the non-certified service/dealer agency. This warranty is in lieu of all other warranties, expressed or implied, and our company shall in no event be liable for incidental or consequential damages including loss of use, property damage, or personal injury resulting from breach of warranty. The Accreditation Manual for Hospitals requires each piece of equipment to be tested prior to initial use and at least every 12 months. To comply with this standard, we recommend that you participate in our accreditation Testing compliance Program during the warranty period. This service can be performed through our company and authorized dealers.
SERVICE For optimal performance, product service should be performed only by qualified service personnel. Technical Services representatives can be reached for fixations and are dispatched for required maintenance by calling 0086-574-87800008, 87800009. Customers outside China should contact their local factory-authorized distributor for service.
COMPLEMENTARY NOTICE Since our factory, conducts a continuous product improvement program, circuit and component improvements are sometimes incorporated into equipment before they can be incorporated into the printed manuals. This manual contains proprietary information, repairs and authorized modifications should be performed only by qualified service personnel to maintain your warranty and to avoid creating safety hazards. We cannot assume responsibility for any conditions affecting the proper operation of this equipment which may result from unauthorized repair modification. Therefore, some parts used in your equipment may be different than those which appear in the parts list of this manual. This sometimes occurs due to difficulty in parts procurement, but does not alter the function of the equipment The device, accessories and the packaging have to be disposed of waste correctly at the end of the usage. Please follow Local Ordinances or Regulations for disposal. i I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE OF DEFINITIONS AND SYMBOLS TECHNICAL DEFINITIONS SKIN TEMPERATURE SENSOR: A sensing device including the link with the equipment intended to measure the infant’s skin temperature. INCUBATOR TEMPERATURE: Air temperature at a point 10cm above and centered over the mattress surface. CONTROL TEMPERATURE: The temperature set at the temperature control. AVERAGE INCUBATOR TEMPERATURE. The average of the maximum and minimum Incubator temperatures achieved during Temperature Condition. STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature does not vary by more than 1℃ over a period of 1 hour. TEMPERATURE ALARM CHECKOUT STATE: The difference between real temperature and control temperature is within ±0.5℃ and such state lasts for over 10minutes. The equipment must stay in such state when check up the alarm about temperature TEMPERATURE UNIFORMITY: The amount by which the average temperature at each of four points 10 cm above the mattress surface differs from the Average incubator Temperature at steady Temperature Condition. The four points are the centers of four quadrants formed by lines that divide the width and length of the mattress surface. TEMPERATURE VARIABILITY: The variability of the Incubator Temperature that will be observed over a one hour period after Incubator Temperature Equilibrium has been reached. TEMPERATURE RISING TIME: The time required for the Incubator Temperature to rise 11℃, when the Air Control Temperature is at least 12℃ above ambient. STEADY HUMIDITY CONDITION: A condition that the disparity between the indicated humidity value and control value is less than ±5%RH, and maintains over 2 min.
NOTE, IMPORTANT, CAUTION AND WARNING NOTE: A note is inserted in text to point out procedures or conditions, which may otherwise be misinterpreted or overlooked. A note may also be used to clarify apparently contradictory or confusing situations. IMPORTANT: Similar to a Note but be used where greater emphasis is required. CAUTION: A caution is inserted in text to call attention of a procedure which, It not followed exactly, can lead to damage or destruction of the equipment. WARNING: A warning is inserted in text to call attention to dangerous or hazardous conditions inherent to the operation, cleaning, and maintenance of the equipment which may result in personal Injury or death of the operator or patient. ii I5
SERVICE MANUAL FOR INFANT INCUBATOR
SYMBOLS Attention; Consult
CLASS I
accompanying documents
Class I Equipment
Type BF Applied Part
Main power on
Main power off
On (only for a part of equipment)
Off (only for a part of equipment)
User-Defined Key
Set Key
Set Temperature Up Key
Set Temperature Down Key
Silence / Reset Key
RS232
Socket for skin temperature 1
Socket for skin temperature 2
High Water Level
Low Water Level Mattress Tilt Direction
Caution, scald Low
The VHA Pedestal Rising Key 220-230V~ 50Hz
AC Power 220-230V, 50Hz
T 1AL//250V Fuse Type T 1AL/250V 220-230V~ 50Hz MAX:3A
SN
High
The VHA Pedestal Falling Key F 6.3AH/250V Fuse Type F 6.3AH/250V T 160mAL//250V Fuse Type T 160mAL/250V
Assistant net power outlet, MAX: 3A T 800mAL/250V Fuse Type T 800mAL/250V Serial Number
Date Of Manufacture
Weight Limit
CE Marking
iii I5
RS232 Connector
SERVICE MANUAL FOR INFANT INCUBATOR
SEASONAL SAFETY CHECK 1. Please clean the plug of power cord at least once a year. Too much dust on plug may cause the fire. 2. The following safety checks should be performed at least every 12 months by a qualified person who has adequate training, knowledge, and practical experience to perform these tests. The data should be recorded in an equipment log. If the device is not functioning properly or fails any of the above tests, the device has to be repaired. 1 . Inspect the equipment and accessories for mechanical and functional damage. ○ 2 . Inspect the safety relevant labels for legibility. ○ 3 . Inspect the fuse to verify compliance with rated current and breaking characteristics. ○ 4 . Verify that the device functions properly as described in the instructions for use. ○ 5 . Test the protection earth resistance according IEC 60601-1:1988 + A1:1991 + A2:1995: ○
Limit 0.1Ω. 6 . Test the earth leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: ○
NC 500μA, SFC: 1000μA. 7 . Test the enclosure leakage current according to IEC 60601-1:1988 + A1:1991 + A2:1995: ○
Limit: NC 100μA, SFC: 500μA. 8 . Test the patient leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: ○
for a.c.: 100μA (BF), for d.c.: 10μA (BF). 9 . Test the patient leakage current under single fault condition with mains voltage on the ○
applied part according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: for a.c.:500μA (BF), for d.c.: 50μA (BF). 10 . According to the test methods of IEC 60601-1:1988 + A1:1991 + A2:1995, the patient ○
leakage current (net voltage should be added on the applied part) of the testing device must less than 5000μA. 11 . Test the patient auxiliary leakage current according IEC 60601-1:1988 + A1:1991+ ○
A2:1995: Limit: NC for a.c.: 100μA (BF), for d.c.: 10μA (BF).SFC 500μA (BF), for d.c.: 50μA (BF).
iv I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE OF CONTENTS SECTION
PAGE
1. GENERAL INFORMATION…………………………………………………………………………1-1 1.1 INTRODUCTION……………………………………………………………………………………1-1 1.2 DESCRIPTION………………………………………………………………………………………1-1 1.3 ACCESSORIES……………………………………………………………………………………1-1 2. INSTALLATION………………………………………………………………………………………2-1 2.1 UNPACKING…………………………………………………………………………………………2-1 2.2 ASSEMBLING ………………………………………………………………………………………2-1 2.3 OPERATION CHECKOUT PROCEDURE………………………………………………………2-6 3.TECHNICAL INFORMATION………………………………………………………………………3-1 3.1 SPECIFICATION……………………………………………………………………………………3-1 3.2 WORKING PRINCIPLE……………………………………………………………………………3-4 3.3 ALARM ………………………………………………………………………………………………3-7 4. CLEANING AND MAINTENANCE………………………………………………………………4-1 4.1 GENERAL……………………………………………………………………………………………4-1 4.2 CLEANING…………………………………………………………………………………………4-1 4.3 STERILIZATION……………………………………………………………………………………4-8 4.4 MAINTENANCE……………………………………………………………………………………4-8 5.SERVICE……………………………………………………………………………………5-1 5.1 GENERAL…………………………………………………………………………………5-1 5.2 HUMIDITY SYSTEM………………………………………………………………………5-1 5.3 OXYGEN SYSTEM…………………………………………………………………………………5-2 5.4 REPLACEMENT OF ELEMENT……………………………………………………………5-3 5.5 SYSTEM SETTING PROCEDURES…………………………………………………………5-6 5.6 PROCEDURE OF TROUBLE-SHOOTING……………………………………………………5-12 6. SPARE PARTS LIST AND ITS NO..………………………………………………………………6-1 6.1 GENERAL……………………………………………………………………………………………6-1 7. DIAGRAMS……………………………………………………………………………………………7-1 7.1 GENERAL……………………………………………………………………………………………7-1
v I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 1 GENERAL INFORMATION 1.1 INTRODUCTION This manual provides instructions for installation, maintenance and repair for the Infant Incubators. This manual is intended to use as reference only by trained, qualified service personnel. Instructions for the operator of the equipment are provided in a separate operator’s manual.
1.2 DESCRIPTION YP-2000 Infant Incubator includes oxygen concentration system, humidity control system, Vertical Height Adjustment cabinet (VHA cabinet), baby scale (for optional function) and monitoring shelf (for optional function). If you don’t purchase the optional part, please omit it.
1.3 ACCESSORIES The following diagram shows the main parts of the Infant Incubator.
I.V. Pole
Monitoring shelf Baby compartment
Shelf
Base
Controller
Mobile Cabinet
I5
1-1
SERVICE MANUAL FOR INFANT INCUBATOR
DESCRIPTION OF PART
EXPLANATION
I.V. pole
A kind of bearing part, which is used for hanging the infusion bottle. Max. Load: 2Kg
Controller
The core part with two kinds of temperature control modes: air mode, baby mode, moreover, it also have the function of weighing, humidity control and oxygen concentration controlling system which are used for autocontrol the heat output and the humidity and oxygen concentration in the incubator. The detail operation please refer to the user's manual.
Monitoring shelf
A kind of bearing part, which is used for putting Infant monitor system. Max. Load: 8Kg
Shelf
A kind of bearing part, which is used for putting some small objects. Max. Load: 3.5Kg
Baby compartment
It is used for placing the infant inside, including the Double-wall Acrylic hood, bassinet, and so on. The bassinet can be tilted at the request of clinical needs. The baby scale is optional part, and the Max. Load of bassinet is 10 Kg. Size of mattress: 740mm×376mm
Base
An important part of Infant Incubator, which mainly composed of the oxygen device, humidity chamber, air filter, and so on.
Mobile Cabinet
A part which can support the main body of Infant Incubator. Cabinets are supplied in VHA stand which including large and small storage compartment. Big storage compartment Max. load: 4 Kg. Small storage compartment Max. load: 2 Kg. Storage cover Max. load: 1 Kg.
1-2 I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 2 INSTALLATION 2.1 UNPACKING Generally, the Infant Incubator is usually packed to two cartons: main body and Vertical Height Adjustment cabinet (VHA cabinet). When taking out the equipment from the cartons, take care not to damage the spare parts of the Infant Incubator.
2.2 ASSEMBLING At least two professionals are required to do the installation of the Infant Incubator with spanners. A. Install the main body onto the Vertical Height Adjustment cabinet Mount the main body on the VHA Pedestal using the four hexangular bolts, spring washer, flat washer by the equipment provided figured as 2.1.
ENLARGED ILLUSTRATION I
Flat Washer M6 Spring Washer M6 Hexangular Bolt M6X45 I
TEMPERATURE CONTROLLER POWER PLUG
2-1 I5
SERVICE MANUAL FOR INFANT INCUBATOR
2 1
Insert the temperature controller power WHEEL UNLOCKED
WHEEL LOCKED
Cord to the socket.
NOTE: The Pedestal front locking casters must be facing down to be locked indicated by arrow 1. The Pedestal front locking casters must be facing down to be unlocked indicated by arrow 2. FIGURE 2.1 IMPORTANT: The main body can be installed on the VHA Pedestal only if the power cord of temperature controller from the VHA Pedestal is located on the same side as the temperature controller. WARNING: 1. The incubator must be attached to the VHA Pedestal using the bolts provided. Failure to do so could result in the incubator separating from the VHA Pedestal if the VHA Pedestal is tilted, particularly when the hood is opened. 2. To prevent the incubator from sliding when parked on an incline, the front locking caster for the VHA Pedestal must be facing down the incline and locked. B. Install the I.V. Pole
Seat for the monitor shelf
Step 1: Fix the seat for monitoring shelf on the right and left sides of the main body with the inner hexagon bolts.
2-2 I5
SERVICE MANUAL FOR INFANT INCUBATOR
I.V. Pole Support pole for I.V. Pole
Clamp seat
Support pole for I.V. Pole
Step 2: Fix the I.V. Pole on the Seat for the monitor
Step 3: Adjust the proper height of I.V.
shelf with latch seat and the inner hexagon
Pole, and tighten it as the arrow
bolts as the arrow indicates.
indicates. FIGURE 2.2
C. Install Shelf As figure 2.3 indicates, insert the fixed seat to the infusion rod. Ensure the top of fixed seat is uniform with the position line on the infusion rod as enlarged figure A show. Then tighten the hexagon bolts. ENLARGED ILLUSTRATION A Line for Tray Position Shelf
A Fix Seat 1
Inner Hexagon Bolt
NOTE: The side with note label of shelf must be installed in front of the incubator. FIGURE 2.3 D. Install Monitoring Shelf See figure 2.4, the fixing method of support pole for I.V. pole, fix the support pole for shelf on the seat for monitoring. Insert the one terminal of tray with hole into the hole of support pole of monitor, and fasten it with inner hexagon bolt tightly. 2-3 I5
SERVICE MANUAL FOR INFANT INCUBATOR
Monitoring Shelf The support pole for monitoring shelf The inner hexagon bolts M5×30
FIGURE 2.4 E. Install the big and small storage Put Φ6 spring washer, Φ6 flat washer and Φ6 outter sawtooth tighten washer onto the four inner hexagon bolt by turn as figure 2.5 shows. Then fix the small storage and big storage to both sides of VHA cabinet seprately by inner hexagon spanner.
Hexangular bolt
Storage bracket
FIGURE 2.5 F. Connect the sensor Connect the sensor connector with sensor socket which is on the side of the incubator’s main body as figure 2.6 directed. Then tighten the bolt in the connector. NOTE:Connection between the sensor connector and socket should in the right orientation for these two parts have their own direction.
Fixed bolt
Sensor connector FIGURE 2.6 2-4 I5
SERVICE MANUAL FOR INFANT INCUBATOR
G. Install oxygen sensor Install oxygen sensor on the incubator as figure 2.7 directed. Stator Sensor Cover
Stopper Pull out the stator on the left side of sensor box as the arrow indicates, and pull out the sensor box, unscrew 2 Cross groove countersunk head bolts, and take out the sensor cover, and then unscrew the stopper on it. Oxygen Sensor Sensor Cover
Screw 2 oxygen sensors into the sensor cover clockwise.
Pull the connecting wire of oxygen sensor out and connect the sensor with the double wire of green and black (if the device is equipped with oxygen control system, please connect another oxygen sensor with the other double wire of yellow and black). After correctly connecting them, screw the sensor cover board down with two m3×8 cross groove head bolt, then push the sensor box into the original place in the hood. CAUTION: 1. Do not block any of the holes on the sensor box. 2. Please pull out and push in the sensor box in correct way. 3. Sensor box is the important part of controlling the temperature of the incubator, please treat it carefully. FIGURE 2.7 2-5 I5
SERVICE MANUAL FOR INFANT INCUBATOR
H. Install weighing system Install the weighing system in the incubator as figure 2.8 directs. 2 1
Adjust the bed to the lowest position, and open the front door, place the bed with baby scale on the wavering mechanism, and it can be pulled out and push in freely as the arrow 1 and 2 indicates. Arrow
Gap
Make the weight sensor can be inserted into
Weight Sensor
The gap of socket correctly 1. When plug in or out the power supply cord, the plug must be taken hold and the wring should not be pulled. 2. Do not bend the connection of sensor. Connect the plug of sensor on the baby scale with the socket on sensor box correctly. FIGURE 2.8 I. Insert the power cord Insert the power cord into the socket of general power supply. Check the INCUBATOR according to the instruction in section 2.3.
2.3 OPERATION CHECKOUT PROCEDURE Please proceed the following checkout procedure each time before operation. 1. Check caster Check whether the caster can drop when lifting the incubator 2cm above the ground. The drop of the caster will cause danger during transporting. IMPORTANT: At least two persons are required, one lifts the incubator, and the other has to check the casters. It needs Min. Please support the stand with block to avoid the incubator inclining if the casters drop. 2-6 I5
SERVICE MANUAL FOR INFANT INCUBATOR
2. START-UP THE CONTROLLER Connect the power cord of the Base to the electrical source and turn the power of the controller on. Then the device sounds like "di..." and company's label and the system edition will be displayed. Then self-check being carried for RAM, SRAM, REAL TIME CLOCK, EEPROM and communication system by the main body of the device as well as the LED indication lights on the board totally light. The whole self-check process lasts about 7 seconds. After passing self-check procedure, the device will enter operation and control interface, or else, the screen indicates the reason for not passing the self-check procedure and display the words of "Please repair it by professional maintenance man". 3. CHECK POWER FAILURE ALARM Pull off the power cord of whole unit, the Power failure alarm should active, and the “Power” light will flash with continuous audible sound. This operation is used for checking if the power failure is normal or not. Insert the power cord again after finishing checking. IMPORTANT: Make sure that the rechargeable battery is full before usage. If not full, it may cause the power failure without the alarming indication. If full, and there is no any indication after disconnecting the main power supply, please refer to the qualified service personnel. 4. CHECK HEATER Control the environment temperature at 21℃ ~ 26℃, and choose the air mode, and set the temperature at 33.0 ℃, and all heat power indicator are on, and heater will output heat completely. NOTE: When the incubator works under the setting state, and continue the following operation procedure. 5. CHECK THE ACCURACY OF TEMPERATURE CONTROL Select the Air Mode, and set the air temperature at 36℃, after the air temperature enters into the STEADY TEMPERATURE CONDITION, put the calibrated temperature measuring device on the position above 10cm from the center of mattress to measure the air temperature, compared with the indicated air temperature to check whether the deviation between them is within 0.8℃. 6. CHECK BABY MODE Insert two skin temperature sensors into the socket of skin sensor1 and skin sensor2 seprately, the arrow sign on the plug should aim to the opening on the sensor socket so that the sensor is inserted correctly.
2-7 I5
SERVICE MANUAL FOR INFANT INCUBATOR
Skin temperature Sensor 1 Arrow
Opening
Sensor plug A
Oxygen concentration sensor
Humidity sensor and air temperature sensor are located in the position A 1.Insert or pull out the skin sensor, you must hold the plug of skin sensor, pulling the leads is forbidden. 2.Please do not bend the connection of sensor. When the Air Mode works stablly, change it into the Baby Mode, set temperature at 37℃. Keep the temperature detected by skin temperature sensor is 2℃ lower than the set temperature while making the displayed air temperature 4℃ lower than the set skin temperature. The sensor failure alarm should be actived then. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Skin Sensor 1 is placed wrong”. Pull off the skin temperature sensor 1, the sensor failure alarm should active. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm skin sensor 1 fault”. 7. CHECK PRECISION OF SKIN TEMPERATURE SENSOR Put the skin temperature sensor with the mercury thermograph for the accuracy with ±0.1ºC into the water cup with the temperature as 30ºC±5ºC. Make the probe of the skin temperature sensor and the mercury ball as closely as possible and stir enough, then read the value of mercury thermograph. Compare the value of the skin temperature sensor and the mercury thermograph, and the deviation must be within 0.5ºC. NOTE: Please check again if the accuracy of the skin temperature sensor exceeds the permissibility deviation. Please let the professional maintenance man service machine if the accuracy the skin temperature sensor exceeds the permissibility of deviation again. 2-8 I5
SERVICE MANUAL FOR INFANT INCUBATOR
8. CHECK THE SENSOR BOX
Check the connection between the modular and its socket. Only when the interface of the sensor box is correctly facing the interface of the sensor socket, they can be successfully connected. STATOR
Undraw the left stator of sensor box as the arrow indicates; under the condition of stator undrawing, undraw the sensor box outwards, to check if the Sensor box slides in or out of the hood smoothly. When the modular falls off from the hood, The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Sensor Box is placed wrong”. After the modular being put into the correct location and being fixed by stator, the alarm cancels automaticly as well as the sensor connect indicating light off. 1. All gaps in the sensor module must not be blocked. 2. Please pull out and plug in the sensor module correctly. 3. Sensor Module is an important part which supply Temperature Control to Incubator, and must operate carefully. 9. CHECK FAN MOTOR ALARM Refer to paragraph 4.2.1, open the Hood, and take out Infant Mattress, Mattress Tilt Mechanism, Mattress Tray, and the heater cover. As the follow figure, hold the Motor Fan on hand carefully, turn on the switch of the controller power, the Alarm Indicator shall be activated, the message “Alarm Motor Fault” should appear in place of the Trend Display and the audible alarm should sound. Press the Silence/Reset key, alarm is over and the Controller return to operation status. 2-9 I5
SERVICE MANUAL FOR INFANT INCUBATOR
NOTE: It must take at least 45 minutes after the incubator stopped working to perform this operation. 10. CHECK OVER-TEMPERATURE ALARM In the mode of Air Mode or Baby Mode, when the unit enters into the working condition, Press the air temperature self-definite function key and set value increasing key in figure 4.2A, at this time, all heating indicators are on, If the set temperature is no indication, it means that the controller has entered into the over-temp test state. After a long time, the device can alarm, the alarming lights are on; “Alarm Over Temperature” information can replace the curve indication on the screen. Press the silence/reset key to cancel the over-temp alarming, and then the device can return to the working condition. 11. CHECK DEVIATION ALARM In Air Mode, close all door and panels, set the temperature at 32 ºC. Enter TEMPERATURE ALARM CHECKOUT STATE(refer to descriptions of terms and symbols concerned), fan into the hot air inside the hood. When the air temperature indicates 35.1ºC, the high deviation alarm will active. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Temperature Deviation”; set the temperature at 35℃, after the device enters into the temperature alarming checkout state, open the front access panel, when the air temperature indicates 31.9 ºC, the low deviation alarm should active. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Temperature Deviation”. NOTE: If the system can not enter into the TEMPERATURE ALARM CHECKOUT STATE or the air Temperature does vary within ±3℃ than the setting temperature, the deviation alarm can not occur. In Baby Mode, set the temperature at 35 ºC. Enter TEMPERATURE ALARM CHECKOUT STATE, put the skin sensor into the water cup at 37 ºC. When the skin temperature indicates 36.1 ºC, the high deviation alarm will active. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Temperature Deviation”; Set the temperature at 35℃, after the device enters into the temperature alarming checkout state, put the skin sensor into the water cup at temperature 33℃, when the skin temperature indicates to 33.9ºC, the low deviation alarm should active. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Temperature Deviation”. 2-10 I5
SERVICE MANUAL FOR INFANT INCUBATOR
NOTE: If the system can not enter into TEMPERATURE ALARM CHECKOUT STATE or the skin temperature does vary within ±1℃ than the setting temperature, the deviation alarm can not occur. 12. CHECK THE WATER LACKING ALARM When humidifying system is on working, pour out the water in the water tank. After a while, the incubator will alarm. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Low Water”. 13. CHECK THE WATER RESERVOIR POSITION WRONG ALARM When the incubator is in the working state, pulling out the water tank, the incubator will alarm. The “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Water Reservoir is placed wrong”. 14. CHECK THE TIMER When the controller is normally working, the message shall displays the current real time. If the displayed time is not correct, please reset the time. The setting method is as following: The time can be indicated year (four figures, the former two are the century, the rest are the year), month, data, week, hour, minute, second. Press the first self-definite function key, and switch on the controller, and the controller can enter into the interface in follow fugure to set the recording time of controller.
Press set self-definite function key, the underlined century can indicate, and then press the select self-definite function key, the underline can move with the time indication, press the increasing key or the decreasing key can change the underlined information, after correction, press the store key to change the time and record the time, press Exit key, the controller can enter into the first interface. 15. CHECK OXYGEN FEEDING SYSTEM 1 The system support with two kinds of oxygen concentration calibration: 21% oxygen ○
concentration and 100% oxygen concentration. When the oxygen concentration of the environment is below 21%, we suggest using 100% oxygen concentration calibration. 2-11 I5
SERVICE MANUAL FOR INFANT INCUBATOR
21% OXYGEN CALIBRATION Keep the incubator in the environment which oxygen concentration is 21%. a. Take out the sensor modular and press the switch of position monitor on the sensor modular to avoid alarming caused by improper place the sensor modular. b. After press the second self-definite function key, turning the controller to enter the interface of oxygen concentration calibration. Press “21% Cal.” Key, see below figure. The controller can calibrate oxygen sensor automatically, and the screen indicates “O2 sensor calibration”. After calibration, controller can enter into the first interface. If calibration work fails, the indicator displays the failure hint, and then users can solve problems according to the hint. If these failures can not be canceled, please ask the authorized and qualified personnel for repair.
100% OXYGEN CALIBRATION a. Pull the sensor box out as below figure shows, connect the oxygen calibration device which equipped with oxygen connecting pipe to the sensor box, supplying the device with 100% oxygen for more than 2min. b. After press the second self-definite function key, turning on the controller, the indicator display the interface of oxygen concentration calibration. Press “100% Cal.” Key, see the before figure. The controller can calibrate oxygen sensor automatically, and the screen indicates “O2 sensor calibration”. After calibration, controller can enter into the first interface. If calibration work fails, the indicator displays the failure hint, and then users can solve problems according to the hint. If these failures can not be canceled, please ask the authorized and qualified personnel for repair.
Sensor modular
Oxygen concentration calibration device 2-12 I5
SERVICE MANUAL FOR INFANT INCUBATOR 2 CHECK THE OXYGEN CONCENTRATION MONITOR FUNCTION ○
a. Place a calibrated oxygen concentration analyzer on the middle of the bassinet. b. Input oxygen to the device, wait until the oxygen concentration displays steady, the value deviation displayed between the value displays in the Oxygen Concentration indicator and the oxygen concentration analyzer should in the range of ±3%. 3 CHECK FOR OXYGEN CONCENTRATION DEVIATION ALARMING FUNCTION ○
Connect the oxygen-offering system and manual oxygen input port, set the oxygen concentration datum mark 30%, and then you can change the oxygen flow according to the oxygen concentration table posted on the incubator, you can increase or decrease the flow rapidly when the indication of oxygen concentration reaches the datum mark; while the indication of oxygen concentration deviates from the datum mark ±5%, the alarming can occur. “Alarm O2 Deviation” can replace the curve trend on screen. Press Silence/Reset key to cancel the alarming, but the indicator of alarming can flash continuously until the indication value does not exceed the datum mark ±5%, the indicator can be turn off. Connect the oxygen input interface of oxygen –offering system and the one of serve control. The standard of setting oxygen concentration of 30%, when the oxygen concentration reaches the standard one, rise the oxygen concentration inside of the hood (transfer the extra oxygen concentration into the hood ), to check the upper deviation alarm (+5% deviation alarm ); When the oxygen concentration inside of hood resume the standard one, open the front panel to check the down deviation alarm (-5% deviation alarm ) , when there is deviation alarm, “Alarm O2 Deviation” can replace the curve trend. Press Silence/Reset key to cancel the alarming, but the alarming light can flash without stopping until the indicate value dose not exceed the standard ±5%, the light is off automatically. 16. CHECK WEIGHT SYSTEM 1 Calibration of weight sensor ○
a. Take out all the things on the bed and make sure the bed cleans. b. Press the zero self-definite function key, weight indicator can read 0000g±5g, if exceeds this range, press ZERO key again. c. Put the weight of 5000g on the baby scale, and then press Calibration self-definite function key, the indication window can indicate the following figure, press No key or do not pressing any key within 15s, the system can exit the calibration procedure automatically. Press Yes key, the system can enter into the calibration state, the indication window can indicate the weight within 5000g±5g, when exceeds this range, repress Calibration self-definite function key until the weight is in the right range. If failed, the baby scale should be repaired by qualified personnel. 2-13 I5
SERVICE MANUAL FOR INFANT INCUBATOR
2 CHECKING THE CONNECTING OF WEIGHT SENSOR ○
Disconnect the connection between the weight sensor and the sensor box, enter into the interface for weight, weight indicator can hint the failure of weight sensor, the controller can cancel the failure indication information automatically and indicate the weight of sensor. 17. CHECK HUMIDITY CONTROL SYSTEM 1 Fulfill the humidity chamber to the highest level with water, then place a calibrated ○
humidity analyzer in the middle of the bassinet. 2 Heating the incubator to 36℃, set the humidity value in 50%. ○ 3 Waiting until the value in the indicator rose to 50%RH, the Humidity indicator should ○
display 50%RH±10%RH. 4 Check the humidity deviation alarm ○
In the humidity set state, set the humidity value to 40%RH. Waiting until the humidity display value reach the set value, then increase the humidity display value to make it 15%RH higher than the set value, the incubator alarming, the “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Humidity Deviation”. Set the humidity value to 50%RH. Waiting until the humidity display value reach the set value, then decrease the humidity display value to make it 15%RH lower than the set value, the incubator alarming, the “Alarm” light will flash with continuous audible sound, and the screen indicator will show “Alarm Humidity Deviation”. Press Silence /Reset key to stop alarming, but the alarm indicator light continuously shines until the deviation is less than ±15%RH. If 4m later, the alarming condition still exists, it will alarm again. NOTE:If the displayed value won’t rise anymore or can’t decline to the set value in the deviation rate more than ±15%, the alarm won’t occur. When checking the low-deviation alarm, please make sure that the environmental humidity is lower than the humidity in the hood so that the alarm can be active.
2-14 I5
SERVICE MANUAL FOR INFANT INCUBATOR
18. CHECK THE HOOD OPERATION NOTE: Make sure all cables are disconnected from the all accessories before raising the hood in order to avoid interfering with raising the Hood.
Disconnect the cord of sensor module.
THE KNOB LOCATED Disconnect the cord of the Sensor Module. Slowly tilt the Hood back until the Hood locks in place. Release the Hood by pulling on and holding the knob located on the right rear hinge while closing the Hood. 19. CHECK THE ACCESS PANEL
The Pawl Latch
UNLOCK
The Latch Seat LATCHED Rotate the panel Latches and make it stays opening. Press the latch seat to open the Access Panel to the full open position (rotate downward). Close the Access Panel, the seat latch fastened automatically and then rotate both latches until they are fully locked. Both latches and the seat latch must be fully locked to avoid accidental opening of the panel. 2-15 I5
SERVICE MANUAL FOR INFANT INCUBATOR
20. CHECK IRIS ENTRY PORTS
Rotate the outer ring of each enrty Port, the port should be open and close as rotation is continued through 360 degree. 21. CHECK INNER WALLS ARE PROPERLY LATCHED
UNLATCHED
LATCHED
Open the Access Panels and check that the front and rear inner walls are properly latched. 22. CHECK THE LATCH AND GASKET OF ACCESS DOOR
Press on-off of each access door to open it automatically and check the seal of gaskets.
2-16 I5
SERVICE MANUAL FOR INFANT INCUBATOR
23. CHECK MATTRESS TILT MECHANISM
Rotate the mattress tilt mechanism to adjust it. 24. CHECK BASSINET
Open the Access Panel and slide it out to the fully extended position indicated by the arrow. Lean on Mattress Tray to make sure it is properly supported to provide a firm infant platform. Return the mattress and close Access Panel. 25. CHECK THE AIR INTAKE FILTER Loosen the two thumb screws of the Air Intake Filter Cover and remove the cover and inspect the filter, if dirty, it should be cleaned. Put the filter cover back.
WARNING: A dirty Air Intake filter may affect oxygen concentration and /or cause Carbon Dioxide build-up. The filter must be checked on a routine basis and changed at least every two months. 2-17 I5
SERVICE MANUAL FOR INFANT INCUBATOR
26. CHECK THE X-RAY TRAY
Open the Access Panel and withdraw
The X-Ray Tray can be pull out easily Indicated
the X-Ray Tray.
as the arrow.
27. CHECK THE HIGH ADJUSTMENT OF THE CABINET Step the Up and Down button with your feet to adjust the height of whole unit.
Down foot button
Up foot button
CAUTION: Please adjust the height of incubator to the lowest position during moving the incubator.
2-18 I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 3 TECHNICAL INFORMATION 3.1 SPECIFICATION Specifications for the Incubators are provided in Table 3.1. All specifications maybe subject to change without advance notice. Open Access Doors and Front Panel or change Mattress obliquity, which can alter the air flow pattern, may affect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and infant skin temperature.
TABLE 3.1
SPECIFICATIONS
This equipment belongs to Class I, Type BF applied part, normal device (IP20) continuously operated. POWER REQUIREMENTS………………………………………………AC220V-230V, 50Hz,1300VA Heater Power Output……………………………………………………………………………350W/240V Working power of VHA cabinet……………………………………………………………120W/220V Working power of humidifier…………………………………………………………………190W/240V POWER output socket……………………………………AC220V-230V, 50Hz, MAX. CURRENT 3A Heater power display………………………………………0 to 100%, adjustable in 10% increments Temperature control modes………………………………………………………………………Air mode Baby mode Air Temperature Control range………………………………………………………………25℃~37℃ override mode: 37℃~39℃ (if applicable) Baby Temperature Control range……………………………………………………………34℃~37℃ override mode: 37℃ ~38℃ (if applicable) Temperature sensor display range……………………………………………………………5℃~65℃ Temperature rise Time*(environment temperature is +22℃) ……………………………… ≤40min Temperature Variation*………………………………………………………………………………≤0.5℃ Temperature Uniformity*(level mattress) ………………………………………………………≤0.8℃ Temperature Uniformity *(Tilt mattress) ……………………………………………………≤1.0℃ Deviation between the indicated air temperature and the real temperature…………≤0.8℃ Accuracy of skin temperature sensor……………………………………………………………±0.3℃ PROTECTIVELY EARTHED…………………………………………………………………………< 0.1Ω EARTH LEAKAGE CURRENT………………………………………………Normal condition: < 0.5 m A Single Fault Condition: < 1mA PATIENT LEAKAGE CURRENT……………………………………………Normal Condition: < 0.1mA Single Fault Condition: < 0.5mA PATIENT LEAKAGE CURRENT …………………………………………Single Fault Condition: < 5mA (Mains Voltage On The Applied Part )
ALARM (See section 3.2) Power failure alarm……………………………………………………………Actives if power failure Motor failure alarm………………………………………Motor failure, or motor run speed too low I5
3-1
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE3.1 SPECIFICATIONS (Continued) Fan failure alarm………………………………………Cool fan failure inside of controller or sensor box Sensor alarm……………………………………………………………..…Air temperature sensor failure Skin temperature sensor 1 is out of position Skin temperature sensor 1 failure Isolated temperature sensor failure Air temperature sensor is different from Isolated temperature sensor Air flow temperature sensor failure Sensor box placed wrong alarm………..………………….………………Sensor box is placed wrong ROM failure alarm…………………………………ROM error in the main MCU inside of the controller ROM error in the client MCU inside of the sensor box Internal system failure alarm…………Internal system error in the main MCU inside of the controller Internal system error in the client MCU inside of the sensor box SRAM failure alarm……………………………………………………SRAM error inside of the controller EEPROM failure alarm……………………………………………EEPROM error inside of the controller Real time clock failure alarm………………………………Real time clock error inside of the controller Communication failure alarm……………Communication error inside of the controller or sensor box ADC alarm………………………………..Analog-to-Digital Converters error inside of sensor box Keyboard failure alarm……………………………..………………………Short circuit on the Keyboard Battery failure alarm……………………………..……………………………Battery failure or no battery LCD display failure alarm……………………………..………………………LCD display error
Humidity control Running time when humidity chamber is full of distilled water………………humidity ≤60% RH, At least 24hrs Capacity of humidity chamber……………………………………………………………………1300ml Humidity display range………………………………………………………………0%RH ~ 99%RH Humidity control range……………………………………0% ~ 90%( adjustable in 1% increments) Humidity display accuracy (ambient temperature 20℃~40℃, ambient humidity 10% ~ 90%)………………………±10% RH NOTE: Under the very high ambient humidity, it may not reach the relative lower humidity control.
Oxygen control Oxygen display range…………………………………………………………………………0% ~ 99% Oxygen resolution………………………………………………………………………………………1% Oxygen display accuracy ………………………………………………………………………±3 vol.% Oxygen control accuracy ………………………………………………………………………±4 vol.% Oxygen set range…………………………………………20% ~ 60%( adjustable in 1% increments) Lifetime of oxygen sensor………………Max.: 10000hrs under the oxygen concentration at 100%
3-2 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE3.1
SPECIFICATIONS (Continued)
WEIGHT INDICATION Weight range……………………………………………………………………………………100 ~ 8000g Weight display accuracy …………………………………………………………………………………1g Weight indication accuracy……………………………………………………………………………±1%
ENVIRONMENT TEMPERATURE Operating Range…………………………………………………………………………………+20~+30℃ Storage Range……………………………………………………………………………………-20~+55℃
ENVIRONMENT HUMIDITY Operating Range…………………………………………………………………………………30~75%RH Storage Range……………………………………………………………………………………≤ 93%RH
ATMOSPHERIC PRESSUR Shipment and Store atmospheric pressure range…………………………………500hPa~1060hPa Operating atmospheric pressure range……………………………………………700hPa~1060hPa
AIR VELOCITY Ambient Air Velocity………………………………………………………………………………< 0.3m/s
OTHER SPECIFICATION Noise level within Hood Environment………………………………Environment noise ≤45dB(A) noise inside hood ≤55dB(A) Carbon dioxide(CO2) level within the hood…………………Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress. Air Velocity Over Mattress………………………………………………………………less than 0.35m/s Lifetime of product…………….…………………………….……………………….………….…8 years
SILENCE/RESET KEY Press to silence the alarming, press twice to reset alarming. Silence……………………………………………………………………………silence alarming for 4min, alarming will silence if there is any other alarming actives Reset…………………………………………………silence alarming to come back to the setting state * Refer to Table of Definitions and Symbols
I5
3-3
SERVICE MANUAL FOR INFANT INCUBATOR
3.2 WORKING PRINCIPLE 3.2.1 General This section contains a functional description and detailed theory of operation of the equipment. A system block diagram of the Controller is shown in Figure 3.2 and 3.3. The control of temperature is achieved by means of air circulation system as shown in Figure 3.1. A controlled amount of room air is drawn through the air intake filter by means of the motor-driven impeller on the Controller and then drawn back to motor though air return filter. Circulated air flows through temperature sensing sensor. Temperature Controller automatically modulates the temperature inside Hood according to the temperature sensed by the air temperature sensing sensor. Heat output control under air mode: control the output of heater according to the air temperature automatically; Heat output control under baby mode: control the output of heater according to the skin temperature automatically. Oxygen feeding control: control the switch of oxygen valve automatically according to the oxygen concentration from the oxygen sensor. Humidity output control of humidity chamber: control the hater inside humidity chamber according to the humidity from humidity sensor.
FIGURE 3.1 AIR/OXYGEN CONCENTRAION/HUMIDITY CYCLE SYSTEM
3-4 I5
SERVICE MANUAL FOR INFANT INCUBATOR
Motor test
Keyboard
Humidity control
Displayer Main MCU
O2 servo control
I/O Control
Temperature heater control
A
Air temperature sensor
Skin temperature sensor 1 Skin temperature sensor 2 Humidity sensor
ADC Client MCU A
Oxygen sensor 1
Oxygen sensor 2 Weight sensor module
Isolated temperature sensor ADC
Air flow temperature sensor 2
Client MCU B Temperature heater control
FIGURE 3.2 CONTROL BLOCK DIAGRAM
I5
3-5
B
SERVICE MANUAL FOR INFANT INCUBATOR
Start INT1 Initialization Real timer
Data display
Return
INT0
YES
Key input
ADC Sampling Key handling NO Return Data handling Timer interruption
YES
Timer
Trouble checking Return NO
Alarm identification Series port interruption
Heating control
Data transfer Data display Return
FIGURE 3.3 PROCEDURE BLOCK DIAGRAM
3-6 I5
SERVICE MANUAL FOR INFANT INCUBATOR
3.2.2 Temperature Control In Air Mode, the controller will control the incubator temperature automatically to keep the air temperature close to the set value. Comparing the air temperature and the air set temperature, the controller will adjust the heat output proportion to keep the heat balance inside hood. The other probe inside of air sensor is regarded as the isolated temperature sensor, when set temperature is < 37ºC incubator temperature must be ≤38.0ºC (while when set temperature is > 37ºC incubator temperature must be ≤40.0ºC ), the heater is cut off automatically. In baby Mode, the controller will control the incubator temperature automatically to keep the skin temperature close to the set value. Skin sensor is attached to the skin of patient. Comparing the skin temperature and the skin set temperature, the controller will adjust the heat output proportion to keep the heat balance inside hood. In the whole baby mode, the isolated temperature sensor of air sensor is always monitor the inside temperature, when the incubator temperature is ≤40ºC, the heater is cut off automatically. When the baby mode is working, if the skin temperature sensor 1 drops from the socket, the skin temperature indicator displays “--.-”, the sensor alarm actives and the heater is cut off automatically.
3.3 ALARM Alarm is used for monitoring the temperature control system state. The system will check the state by the alarm checkout procedure. When the checked data is abnormal, it will alarm. Table 3.2 listed alarm condition.
TABLE 3.2 ALARM ALARM MESSAGE Power Failure Alarm Alarm Sensor box is placed wrong Alarm Air Sensor Fault
“Alarm” light is on
Alarm isolated Sensor Fault Alarm Air Sensor is different from isolated Sensor Alarm Air flow Temp. Sensor Fault
I5
DESCRIPTION If the mains power supply fails or no power supply, the alarm with audible and visual will active. If the sensor box wrongly placed, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. In the air mode, if the temperature sensor short-circuit, open-circuit or bad connected, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. If the isolated temperature sensor short-circuit, open-circuit or bad connected, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. If the temperature difference detected between air temperature sensor and isolated temperature sensor over 0.8℃, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. If the air flow sensor short-circuit, open-circuit or bad connected, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. 3-7
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 3.1 ALARM (CONTINUE) ALARM MESSAGE Alarm Skin Sensor 1 Fault
Alarm Skin sensor 1 is placed wrong Alarm Humidity Sensor Fault
Alarm O2 Sensor 1 Fault
Alarm O2 Sensor 2 Fault
“Alarm” light is on
Alarm O2 Sensors are Different
Alarm Motor Fault
Alarm Fan 1 Fault
Alarm Fan 2 Fault
DESCRIPTION In Baby Mode, if the skin sensor 1 short-circuit, open-circuit or bad connected, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. In Baby Mode, if the temperature detected by skin sensor 1 always 2℃ below the set temperature, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. If the humidity sensor short-circuit, open-circuit or bad connected, at the same time, heater keeps working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. In oxygen module, if short-circuit or open-circuit or failure occurs on the oxygen concentration sensor 1, at the same time, heater keeps working, and the oxygen valve cuts off the oxygen feeding, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. In oxygen module, If short-circuit or open-circuit or failure occurs on the oxygen concentration sensor 2, at the same time, heater keeps working, and the oxygen valve cuts off the oxygen feeding, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. In oxygen module, the oxygen concentration deviation from oxygen sensor 1 and sensor 2 is 3%O2, at the same time, heater keeps working, and the oxygen valve cuts off the oxygen feeding, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. When the wind speed is lower than 1000rpm, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. Fan is failed inside control, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. Fan is failed inside sensor module, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. In air mode, If the air temperature is higher 3℃ than the set temperature, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist.
Alarm Temperature Deviation
In baby mode, if the skin temperature is higher 1℃ than the set temperature, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. In air mode, if the air temperature is lower 3℃ than the set temperature. At this time, the heater keeps working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. In baby mode, if the skin temperature is lower 1℃ than the set temperature. At this time, the heater keeps working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. 3-8 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 3.1 ALARM (CONTINUE) ALARM MESSAGE
Alarm Humidity Deviation
Alarm O2 Deviation
Alarm Over Temperature “Alarm” light is on
Humidity from humidity sensor is higher 15%RH than the set humidity. At this time, the heater keeps working, while evaporator stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. Humidity from humidity sensor is lower 15%RH than the set humidity. At this time, the heater keeps working, and evaporator keeps working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist. The oxygen concentration from oxygen sensor 1 is higher 5%O2 than set value, at the same time, heater keeps working, and the oxygen valve cuts off the oxygen feeding, press Silence/Reset Key to clear alarming for 110s, it can reset when the failure does not exist. The oxygen concentration from oxygen sensor 1 is lower 5%O2 than set value, at the same time, heater keeps working, and the oxygen valve keeps the oxygen feeding, press Silence/Reset Key to clear alarming for 110s, it can reset when the failure does not exist. In the air mode, when the incubator temperature is less than 38℃(set temperature <37℃) or less than 40℃(set temperature > 37℃ ) , at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can not reset after clearing the failure, press Silence/Reset Key to clear alarming. In the Baby mode, when the incubator temperature is less than 40℃, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can not reset after clearing the failure, press Silence/Reset Key to clear alarming.
Alarm Air flow Temp. Over Fault
The temperature detected by air-flow temperature sensor is overhigh, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can not reset after clearing the failure, press Silence/Reset Key to clear alarming.
Alarm Water Reservoir is placed wrong
When the water tank is not in the correct location, at the same time, the heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist.
Alarm ROM of main MCU Fault Alarm ROM of client MCU A Fault Alarm ROM of client MCU B Fault Alarm Internal system of main MCU Fault Alarm Internal system of client MCU A Fault I5
DESCRIPTION
ROM inside of main MCU fails. At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. ROM inside of auxiliary MCU A fails. At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. ROM inside of auxiliary MCU B fails. At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. The ALU or inner RAM in the main MCU failure, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. The ALU or inner RAM in the auxiliary MCU A failure, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. 3-9
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 3.1 ALARM (CONTINUE) ALARM MESSAGE Alarm Internal system of client MCU B Fault Alarm EEPROM Fault Alarm SRAM Fault Alarm Communication A Fault Alarm Communication B Fault
“Alarm” light is on
DESCRIPTION The ALU or inner RAM in the auxiliary MCU B failure, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. EEROM inside main MCU fails, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. SRAM failure (including system cord failure). At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. Communication of main MCU and Auxiliary MCU A fails, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. Communication of Auxiliary MCU A and Auxiliary MCU B fails, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Alarm ADC1 Fault
The Analog-to-Digital Converter TLV2548 inside of the sensor box is error, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Alarm ADC2 Fault
The Analog-to-Digital Converter TLV2544 inside of the sensor box is error, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Alarm ADC3 Fault
The internal Analog-to-Digital Converter of the client MCU MS1210 is error inside of the sensor box, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Alarm Keyboard Fault Alarm Battery Fault Alarm Temperature heater system F lt Alarm Real Time Clock Fault Alarm water shortage
The keyboard circuit is broken, at this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. The battery inside of the controller is error, at this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. The temperature heater system inside of controller is error, at this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. The timer RTC inside the system fails or is not accurate, at this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part. When the water in the water tank is lower than the detect point, at the same time, the heater will working, the heater in the water reservoir stop working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist.
3-10 I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 4 CLEANING AND MAINTENANCE 4.1 GENERAL This section provides cleaning and maintenance instructions. WARNING: Disconnect all the connections with oxygen feeding device before cleaning and maintenance. Cleaning or maintenance in the environment full with oxygen will cause fire or explosion.
4.2 CLEANING This device must be cleaned and sterilized for the first time for initial use, or after used it for one week. 4.2.1 DISASSEMBLY BEFORE CLEANING 1. Turn off the Power Switch on the VHA Pedestal and remove the Power Cord. Disconnect the cables from the Sensor Module and slowly raise the Hood. CAUTION: Before lifting Incubator Hood for cleaning, ensure that all mounted accessories have been removed to prevent possible interference with the raised Hood. 2. Refer to Paragraph 2.3, Step Y, and remove the X-ray Tray from the Mattress Tray. 3. Refer to Figure 4.1 and remove the Mattress Tray. (For the bed with the baby scale, please see the figure 2.8, disconnect the weight sensor and take the baby scale out. Then lift the raise the Hood.)
Push the Mattress Tray to end following as the arrow 1, raising the Mattress Tray like the arrow 2, then pull out it as the arrow 3. FIGURE 4.1
I5
4-1
SERVICE MANUAL FOR INFANT INCUBATOR
4. Refer to Figure 4.2 and remove the Mattress Tilt Bars.
FIGURE 4.2 5. Refer to Figure 4.3 and pull out the Main Deck by loosening the screw.
FIGURE 4.3 6. Refer to Figure 4.4 and remove the Heater Cover.
FIGURE 4.4 The Heater can be sufficiently hot to cause burns. Avoid removing or touching the Heater until the unit has been switched off for at least 45 minutes.
4-2
I5
SERVICE MANUAL FOR INFANT INCUBATOR
7. Refer to figure 4.5 and take out the inner walls. The Inner Walls are hinged on the Access Panel or rear wall of the Incubator. Refer to Paragraph 2.3 Step T, and release the Inner Wall by pressing on the catches located at the top of the Inner Wall. Press the inner wall, and at same time, move it as the arrow 1 indicates to take out the inner wall.
1
FIGURE 4.5 8. Refer to figure 4.6 and take out the water reservoir.
Handle Reservoir
Rotate the spanner of the water tank to the horizontal position and pull the water tank out then. (Please keep the water tank flat and straight and don't move it up towards when it been pulled out, or else it will be hard to pulled out). FIGURE 4.6 CAUTION:1. Because the temperature of the water tank is high after the power of the device is on as well as the heating groups are tighten then, so please take out the water tank after it is totally cooling down to avoid damage to the heating group. 2. The humidifying system is in the bottom of the reservoir. Don’t put hands into the reservoir to avoid being burned.
I5
4-3
SERVICE MANUAL FOR INFANT INCUBATOR
9. Refer to figure 4.7 and take out the fan. NOTE: The direction for unscrewing the heat dissipation slice in way of anti-clock wise.
Cover For Bolt
Hold the fan, and loose the cover for bolt, draw the fan upward FIGURE 4.7 10. As the figure 4.8, take out the oxygen transfusion tube.
Unscrew the fixing nut of the oxygen transfusion tube, and take it out FIGURE 4.8 11. Remove all Access Door Sleeves, Cuff and Gaskets from each side of the Hood by pulling them free. 12. Remove Tubing Access Ports from the Hood by pulling them free. 13. Remove the Air Intake Micro filter Cover by loosening the two thumbscrews. Refer to Paragraph 2.3 Step X. 4.2.2 CLEANING PROCEDURE CAUTION: Some chemical cleaning agents may be conductive and/or leave a residue which may permit a build-up of dust or dirt which may be conductive. Do not permit cleaning agents to contact electrical components. Do not spray cleaning solutions onto any of those surfaces.
4-4
I5
SERVICE MANUAL FOR INFANT INCUBATOR
A. Clean the skin temperature sensor and sensor modular Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or paper towel. CAUTION: Do not steep Sensor into disinfectant-detergent and avoid the disinfectantdetergent leak into the sensor modular. B. Clean the Soft tubing, Gaskets, Iris Port Sleeves Allow parts to soak as recommended by the cleaning solution’s manufacturer, then remove them and dry completely with a clean cloth or paper tower. C. Clean the Hood and inner wall Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or paper towel. NOTE: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol, acetone, or any organic solvents for cleaning. Do not expose the panel assembly to direct ultraviolet radiation. D. Clean the controller Use detergent to clean the all surfaces thoroughly. Then dry it with clean cloth. E. Clean the heater radiator and fan impeller Remove any lint build-up on the Heater Radiator and Fan Impeller. WARNING:Not clean the heater and fan will result in accumulate of fiber dust, baffle the air flow. That will influence the temperature control and result in high CO2 concentration. F. Clean the mattress, mattress tilt bars, main deck, shelf, storage compartment and heater cover. Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or paper towel. G. Clean mattress cover Use a disinfectant-detergent to thoroughly clean it and sun it after washing it with clean water. H. Clean the infant bed Separate the upper bed and the lower bed, clean the dirt’s inside the bed; clean the surface of bed and baby scale (accessory) with the cloth dipping into the detergent, and dry it with clean cloth. NOTE: There is weight sensor inside of bed, clean and place carefully, do not put it into the detergent and sterilized liquid. I. Clean Air filter Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or paper towel. If the air filter is dirt or is used for more than 2 months, please replace it. I5
4-5
SERVICE MANUAL FOR INFANT INCUBATOR
J. Clean Water tank and Humidity chamber fill spout Take the humidifying chamber out from the drawer group as figure 4.7 shows. Open the two cover boards on the humidifying chamber and pull up the water input baffle. Use the national registered neutral disinfectant-detergent to clean the all surfaces and cupped place thoroughly. Clean it again with clean water and then dry it with clean cloth or paper. Please pay special attention that clean the inner and outter surface of the probe with the around the detect point can control the microorganism breed in the water tank and ensure the accuracy of the water level detecting. When the scale appears visibly on the radiating cover inside the water tank, it is recommended to soak it with table vinegar or the scale cleaning detergents which is used exclusively for casting water dispenser. The soaking time is about 15min to 20 min. (The longer the radiating cover be soaked, the better) .In the process of soaking, make sure that the cover is immersed entirely. After soaking, scrub with bottle brush and rinse the cover till the scale is cleaned completely. Reek outlet
Cover board
Detecting probe
Radiating cover Water proof Gap of the water proof
Humidity chamber
Drawer groups
Water inlet hole on the bottom
CAUTION:1. Please clean the water tank after the radiating cover inside the water tank totally cooleddown. 2. Make sure the water inlet hole on the bottom and the reek outlet hole of tank throughable after cleaning. Two gaps on the water proof must placed downwards. 3. If the water tank is uncleaned, especially there're dusts near the probe, there will be a possibility that water-lacking alarm failure. 4. Please use soft cloth to wipe the surface of the probe when it is covered by nobel metal. Do not use the hard material to scrape the probe so as not to damage it. 5. When the scale appears on the radiating cover visibly, you must clean it in time, or it will easily breed ground for bacterium and affect the humidification. FIGURE 4.9 K. Clean the oxygen infusion tube Clean the surface of oxygen infusion tube with clean cloth. Replace it when it is damaged. L. Clean surface of device Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations, Access Doors and Iris Ports then dry them with a clean cloth or paper towel.
4-6
I5
SERVICE MANUAL FOR INFANT INCUBATOR
4.2.3 REASSEMBLY AFTER CLEANING NOTE: Before reassembling parts into Incubator, check carefully if there is any crack or damage. Some disinfectant-detergent may cause crazing of plastic parts in Incubator. A. Install Humidity chamber fill spout in the reverse steps as they are moved out. B. Install the Heater Cover. C. Install the Main Deck. CAUTION: Make sure the Hood is raised before attempting to install the Main Deck. Installing the Main Deck when the Hood is down may result in damage to the Main Deck and/or jamming of the Hood in place. D. Install the Mattress Tilt Bars on the Main Deck. E. Install the Mattress Tray and X-ray Tray. F. Install the Mattress. Visually and physically examine the mattress for any holes or cuts that will permit the entry of fluids onto the inner foam. If the mattress is damaged, it should be replaced. G. See figure 4.5 install inner walls in the reverse steps as they are moved out. H. Rotating the iris entry sleeves, see figure 4.10.
Install the smaller diameter elastic band of a new
Fold back and slip elastic band over
Sleeve over the inner ring of the port housing.
the outer ring of the port housing.
Rotate outer ring to close. If properly installed, the sleeve will open again if rotation is reversed. FIGURE 4.10 I5
4-7
SERVICE MANUAL FOR INFANT INCUBATOR
I. See figure 4.11, install the Access door gasket, and then install the Access door cuff.
Installation of Access Door Gasket
Installation of Access Door Cuff
FIGURE 4.11 J. Install Tubing Access Ports, as shown in Figure 4.12.
FIGURE 4.12 K. Install an Air Intake Micro filter. Refer to the Paragraph 2.3, Step X, place the Air Intake Micro filter into the Cover and tighten the two thumbscrews.
4.3 STERILIZATION NOTE: Please do not steam autoclave. Please sterilize under low temperature.
4.4 MAINTENANCE Please check the condition of the build-in rechargeable battery before the first use of device or in the alternation of device using. A. Operate the unit for a period of 12 to 24 hours. B. Trigger a power failure alarm by disconnecting the AC power cord. C. The power failure alarm should activate and continue to alarm for at least 10 minutes. D. Reconnect the unit to the AC line and recharge the battery. If the power failure alarm cannot last more than 10 minutes, please replace the rechargeable battery. For this battery, it should be replaced by authorized and qualified service personnel.
4-8
I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 5 SERVICE 5.1 GENERAL This section provides faliure detecting and overhauling for the humidity system failure, oxygen supplying system failure, inner rechargeable battery failure and the replacement for fuses.
5.2 HUMIDITY SYSTEM 5.2.1 Working Principle See following figure, and the humidity id mainly composed of the water saving part, humidifier, heater and so on. Water from water saving part filters from the baffle naturally to enter into the humidifier through the water inlet, and the heater will heat the heat conduction bush to make the water boil, which produce the vapor and then enter inside hood through the vaport outlet. With the output of vapor, there is always water to be added to the humidifier to keep the water in humidifier to consist with the one in water saving part. When the water level is lower, please add water from the water inlet. Water adding port
Water saving part
vapor outlet
humidifier
micro switch
temperature control switch
baffle water inlet shell of humidity chamber
5-1 I5
heat conduction bush
heater
SERVICE MANUAL FOR INFANT INCUBATOR
5.2.2 Service The following table lists the possible trouble failure, reason and remedy within the lifetime of humidity system: TABLE 5.1 TROUBLE FAILURE TROUBLE FAILURE
Abnormal humidity message
Distilled Water Can Not Enter Into The Humidifier
Water Leakage
REASON
REMEDY
1.Humidity Sensor Failure
See Flow Chart 5.11
2.Heater Is Damaged For Short Circuit
Replace Heater
3. Temperature Controller Switch For Short Circuit
Replace The Temperature Controller Switch
1.Baffle Hole Is Blocked
Clean Baffle
2.Water Inlet Is Blocked
Clean Water Inlet
Heat Conduction Bush Is Loose
Tighten The Screw Or Replace The Silicon Pad
Too Much Overflow
Distilled
Water
Will
Lower It To The Highest Water Level
5.3 OXYGEN SYSTEM 5.3.1 Working Principle Oxygen feeding system is mainly composed of seperation unit of air and oxygen, electric valve, MCU control system, and there is two connectors on the seperation unit of air and oxygen: one connects the system servo control oxygen port, and the other connects the inlet. The system will compare the oxygen concentration from oxygen sensor and the set oxygen concentration during supplying oxygen, and it will output the differenct control signals accoridng the result to control the on and off for the electric valve so that the oxygen concentration inside hood will reach the stable state. seperation unit of air and oxygen
electric valve
5-2 I5
SERVICE MANUAL FOR INFANT INCUBATOR
5.3.2 Service The following table lists the possible trouble failure, reason and remedy within the lifetime of oxygen system: TABLE 5.2 TROUBLE FAILURE TROUBLE FAILURE
Abnormal oxygen message
REASON
REMEDY
1.humidity sensor failure
See flow chart 5.12
2.heater is damaged for short circuit
See flow chart 5.12
3. temperature controller switch for short circuit
See flow chart 5.13
High devication of oxygen concentration
Check whether the oxygen flow rate is over high. Calibrate oxygen concentration sensor Check the seal condition of access doors and panels
Alarm O2 Deviation Low devication of oxygen concentration
Much noise of oxygen supply
Check whether the oxygen flow rate is over low. Check whether the output voltage of oxygen concentration sensor is normal, (output voltage in normal condition is >9mV), or else, replace oxygen concentration sensor Check oxygen concentration sensor
Connector of oxygen system may leak oxygen
Check the seal condition of all connectors
Spring inside the seperation unit of air and oxygen may change
Replace the seperation unit of air and oxygen
5.4 REPLACEMENT OF ELEMENT 5.4.1 DISASSEMBLY THE CONTROLLER 1. See figure 5.1, uncrew the fixing bolt beside of power switch of controller.
FIGURE 5.1 5-3 I5
SERVICE MANUAL FOR INFANT INCUBATOR
2. See figure 5.2, take out the controller.
FIGURE 5.2 3. Disconnect all connection on the back of controller, and unscrew the bolts on two sides of contrtoller, and you can open the cover for the controller. 5.4.2 REPLACEMENT OF FUSE AND INTERNAL BATTERY A. Replacement Of Internal Battery Open Controller and replace internal battery. See figure 5.3, disonnect the battery fastener and the internal charing battery, and take out the plastic warsher to replace the battery.
Battery fastene
Internal charing battery
Baffle
FIGURE 5.3 NOTE: 1. During replacing the internal battery, the plastic washer should be placed the original position, and keep correct connection between the anode/ cathode of battery and its fastener; the faulty connection will cause the damage to the battery and no hont for power failure alarm. 2. Please replace theb internal battery after using it for three years. 3. Battery will be automatically charged when the temperature controller works, charging the battery outside of the incubator is forbidden. 4. The charging battery inside the wartmer is GP17R8H Ni-MH8.4V/170mAh, to ensure safety, do not disassemble battery. Connect the battery correctly, do not use other type battery. 5. It will not influence battery charging and the battery capacity when the battery is full. (Switch on controller) 6. Please do not throw the old battery into fire or water to avoid explosion or leakage. 7. Please dispose the Old battery accoring to local law, and please do not throw them away randomly. 5-4 I5
SERVICE MANUAL FOR INFANT INCUBATOR
B. Replacement Of Fuse a. See figure 5.4, unscrew the fuse's cover to replace 2pcs fuse F6.3AH/250V
Fuse’S Seat
FIGURE 5.4 b. See figure 5.5, open the outer shell of controller.Open the fuse's cover T type fuse T1AL/250V and T800mAL/250V on power board with tools to replace the fuse. Unscrew he fuse's cover T type fuse T 160mAL/250V of power transformer to replace the fuse. Fuse’S Seat
Fuse’S Seat
Fuse’S Seat FIGURE 5.5 NOTE: You must cut off the power supply when replacing the fuse, and then pull off the power cable, replace the fuse according to the marked specification. 5.4.3 REPLACE HEATER See figure 5.6, rotate anticlockwise to replace heater.
FIGURE 5.6 5-5 I5
SERVICE MANUAL FOR INFANT INCUBATOR
5.4.4 REPLACEMENT OF SEPARATING UNIT FOR AIR AND OXYGEN See figure 5.7, take out the air filter cover, and unscrew the fixed bolt on separating unit for air and oxygen, pull it out from the port from the air filter, and unscrew the connectors from separating unit for air and oxygen to change it. connection of oxygen port inlet connection
FIGURE 5.7 NOTE: The inlet on the separating unit for air and oxygen and inlet of oxygen feeding are connected, therefore, please be careful to pull out the separating unit for air and oxygen, do not pull it out for lond distance. CAUTION: After replacing the separating unit for air and oxygen, make sure all tubes are correctly connected and sealed. 5.4.5 Replacement Of Baby Scale Because the baby scale is connected with bassinet together, therefore, please replace the whole part when the weight system is damaged.
5.5 SYSTEM SETTING PROCEDURES 5.5.1 General This paragraph provides controller system setting procedure. This procedure is only available to the qualified service personnel. NOTE: Please follow by the procedures of item 5.5.3. If it is unnecessary, please do not change data of system setting to avoid hazards. 5.5.2 Brief Introduction Controller system setting procedures listed below: 1. Air temp Compensation 2. Air flow Compensation 3. Skintemp 1 Compensation 4. Skintemp 2 Compensation 5. Humidity Compensation 6. Incubator Address 5-6 I5
SERVICE MANUAL FOR INFANT INCUBATOR
7. Alarm-set Fan Motor 8. Alarm-set Air temp Low 9. Alarm-set Air temp High 10. Alarm-set Skin temp Low 11. Alarm-set Skin temp High 12. Alarm-set Humidity Low 13. Alarm-set Humidity High 14. Alarm-set O2 Low 15. Alarm-set O2 High 16. Set Minimum Humidity 17. Set Maximum Humidity 18. Set Minimum O2 19. Set Maximum O2 20. Set Minimum Air temp 21. Set Maximum Air temp 22. Set Minimum Skin temp 23. Set Maximum Skin temp 24. Set Over-temp A for 37℃ 26. Alarm-set Air flow A for 37℃ 5.5.3 Procedure At first, connected power supply, press and hold and loose
key , LCD screen will indicate like the figure 5.8:
FIGURE 5.8 5-7 I5
key and turn on the power supply for 3s
SERVICE MANUAL FOR INFANT INCUBATOR
See figure 5.8, press
or
key to switch the number, and them press the muner key
to input the password 78722, press
key to enter into the system setting screen--first screen
infigure 5.9. At this time, press the select key to choose the different item to set, the left arrow can show the current item, at this time, press
or
key to set the data of current item. If not pressing
any key for 20s, the system will exit the setting screen.
The first screen
The second screen FIGURE 5.9
A. The first screen
Select Select Press this key to move the arrow downward“
” that indicated on the left side of screen. The arrow
can move circularly, if you miss the wanted item, you have to move the arrow to the last item and restarting the arrow-moving. After choose the wanted item, press
or
key to change the
setting value of the selected item.
Store Press this key to store the current changed setting value.
Menu Press this key to enter into the second screen:
Exit Press this key to exit the system setting interface, the controller enters into the working mood and starts self-testing then.
5-8 I5
SERVICE MANUAL FOR INFANT INCUBATOR
B. The second screen Default Press this key to reset the system data to the original set value. Resume Press this key to load the copied data. (namely the data which have been copied before) Backup Press this key to copy the current data, for the using of reset. Home
Press this key to exit the second display interface to the first display interface. z
1. Air temp Compensation Stipulate the temperature compensation of air sensor, and the setting value is +050~-050,
“+”and“-”mean the increasing and descreasing value. For example, “+002” means that the temperature meansured by air sensor increases by 0.2℃ as compensation, “-013” means that the temperature meansured by air sensor decreases by 1.3℃ as compensation. NOTE: This temperature compensation acts on isolated temperature sensor. z
2. Air flow Compensation Stipulate the temperature compensation of air flow temperature sensor, and the setting value is
+050~-050,“+”and“-”mean the increasing and descreasing value. For example, “+002” means that the temperature meansured by air flow temperature sensor increases by 0.2℃ as compensation, “-013” means that the temperature meansured by air flow temperature sensor decreases by 1.3℃ as compensation. z
3. Skin temp 1 Compensation Stipulate the temperature compensation of skin sensor 1, and the setting value is +050~-050,
“+”and“-”mean the increasing and descreasing value. For example, “+002” means that the temperature meansured by skin sensor 1 increases by 0.2℃ as compensation, “-013” means that the temperature meansured by skin sensor 1decreases by 1.3℃ as compensation. z
4. Skintemp 2 Compensation Stipulate the temperature compensation of skin sensor 2, and the setting value is +050~-050,
“+”and“-”mean the increasing and descreasing value. For example, “+002” means that the temperature meansured by skin sensor 2 increases by 0.2℃ as compensation, “-013” means that the temperature meansured by skin sensor 1decreases by 1.3℃ as compensation.
5-9 I5
SERVICE MANUAL FOR INFANT INCUBATOR
z
5. Humidity Compensation Stipulate the relative humidity compensation of humidity sensor, and the setting value is
+050~-050,“+”and“-”mean the increasing and descreasing value. For example, “+008” means that the relative humidity (%)meansured by humidity sensor increases by 8% as compensation, ”-118” means that the relative humidity (%)meansured by humidity sensor decreases by 18% as compensation. z
6. Incubator Address The address of the infant incubator, and the range of number is 0~40, For example, 001 means the
machine No. 1, 015 means the machine No. 15. This number is just used for the central monitoring operation. z
7. Alarm-set Fan Motor Stipulate the setting methods of alarm value for motor speed. Seeting range is 000~255,every 001
means 20rpm. For example, “050” means 1000rpm, if the speed of motor is lower than this speed, it will alarm. z
8. Alarm-set Air temp Low Stipulate the setting methods of alarm value for low deviation in air mode. Seeting range is
000~050,for example, “008” means 0.8℃, if the temperature measured by the air sensor is lower 0.8℃ than the setting temperature, it will alarm. z
9. Alarm-set Air temp High Stipulate the setting methods of alarm value for high deviation in air mode. Seeting range is
000~050,for example, “008” means 0.8℃, if the temperature measured by the air sensor is higher 0.8℃ than the setting temperature, it will alarm. z
10. Alarm-set Skin temp Low Stipulate the setting methods of alarm value for low deviation in skin mode. Seeting range is
000~050,for example, “008” means 0.8℃, if the temperature measured by the skin sensor 1 is lower 0.8℃ than the setting temperature, it will alarm. z
11. Alarm-set Skintemp High Stipulate the setting methods of alarm value for high deviation in skin mode. Seeting range is
000~050,for example, “008” means 0.8℃, if the temperature measured by the skin sensor 1 is higher 0.8℃ than the setting temperature, it will alarm. z
12. Alarm-set Humidity Low Stipulate the setting methods of alarm value for humidity low deviation. The setting value is
000~020. For example, “008” means that the relative humidity 8% for low deviation, if thehumidity measured by the humidity sensor is lower 8% than the setting value, it will alarm. 5-10 I5
SERVICE MANUAL FOR INFANT INCUBATOR
z
13. Alarm-set Humidity Higih Stipulate the setting methods of alarm value for humidity high deviation. The setting value is
000~020. For example, “008“ means that the relative humidity 8% for high deviation, if thehumidity measured by the humidity sensor is higher 8% than the setting value, it will alarm. z
14. Alarm-set O2 Low Stipulate the setting methods of alarm value for oxygen concentration low deviation. The setting
value is 000~010. For example, “005” means that the oxygen concentration 5% for low deviation, if thehumidity measured by the oxygen concentration sensor is lower 5% than the monitoring value, it will alarm. z
15. Alarm-set O2 High Stipulate the setting methods of alarm value for oxygen concentration high deviation. The setting
value is 000~010. For example, “005” means that the oxygen concentration 5% for high deviation, if thehumidity measured by the oxygen concentration sensor is higher 5% than the monitoring value, it will alarm. z
16. Set Minimum Humidity Stipulate the setting methods of lower limit for humidity setting. The lower limit value is 000~050.
For example, “020”means that Min. setting value is 20% rlative humidity. z
17. Set Maximum Humidity Stipulate the setting methods of lower limit for humidity setting. The lower limit value is 030~100.
For example, “090” means that Min. setting value is 90% relative humidity. z
18. Set Minimum O2 Stipulate the setting methods of lower limit for oxygen concentration. The lower limit value is
000~030. For example, “020” means that Min. setting value is 20% oxygen concentration. z
19. Set Maximum O2 Stipulate the setting methods of upper limit for oxygen concentration. The upper limit value is
030~065. For example, “060” means that Min. setting value is 60% oxygen concentration. z
20. Set Minimum Airtemp Stipulate the setting methods of lower limit for air temperature. The lower limit value is 000~340.
For example, “250” means that Min. setting value is 25.0℃. z
21. Set Maximum Aritemp Stipulate the setting methods of upper limit in the mode of air setting temperature>37℃. The
upper limit value is 370~390. For example, “390” means that Min. setting value is 39.0℃. 5-11 I5
SERVICE MANUAL FOR INFANT INCUBATOR
z
22. Set Miminum Skintemp Stipulate the setting methods of lower limit for skin temperature. The lower limit value is 300~340.
For example, “340” means that Min. setting value is 34.0℃. z
23. Set Maximum Skintemp Stipulate the setting methods of upper limit in the mode of skin setting temperature>37℃. The
upper limit value is 370~390. For example,” 380” means that Min. setting value is 38.0℃. z
24. Set Ovet-temp A for 37℃ Stipulate the setting methods of over-temp alarm in the mode of air temperature ( setting
temperature >37.0℃)or in the skin mode. The alarm value is 385~400. For example, “395” means that over-temp alarm value is 39.5℃. z
26. Alarm-set Air flow A for 37℃ Stipulate the setting methods of air flow temperature alarm in the mode of air temperature(setting
temperature >37.0℃)or in the skin mode. The alarm value is 405~520. For example, “395” means that over-temp alarm value is 39.5℃.
5.6 PROCEDURE OF TROUBLE-SHOOTING 5.6.1 Introduction Section 5.6.3 and 5.6.4 will offer the repairing methods of control. In section 5.6.3, it offers the alarm code when the alarm of the controller happens; in section 5.6.4, it gives the details of the repairing methods by the way of flow chart. NOTE: Before deal with the troubles, you should confirm the correct set of the system set procedure. 5.6.2 Test device During the trouble-shooting procedure, it needs the following test devices, and you can replace it with the same effect device. 5-12 I5
SERVICE MANUAL FOR INFANT INCUBATOR
z
Figure clock, FLUKE 17B Type
z
31.0℃ Air Temperature Simulation
Resistor Resistor R=23166Ω/0.25W/0.5% z
31.0℃ Air flow Temperature Simulation
Resistor Resistor R=23166Ω/0.25W/0.5% z
31.0℃ Skin Temperature Simulation
Resistor Resistor R=23166Ω/0.25W/0.5% 5.6.3 Troubleshooting All alarms have the relevant alarm code except for the power ailure alarm, please refer to Table 5.3. TABLE 5.3 ERROR CODES TYPE
Air temperature sensor failure alarm
EXPLANATION AND CORRECTIVE ACTION 1. check whether the sensor module is placed right or not 2. check the connection between the micro switch on the bottom of sensor module and hood 3. replace sensor module Refer to Flow chart 5.1
Isolated temperature sensor alarm
Refer to Flow chart 5.2
temperature deviation failure alarm
Refer to Flow chart 5.3
Air flow sensor alarm
Refer to Flow chart 5.4
Skin temperature sensor 1 failure alarm
Refer to Flow chart 5.5
Over temperature failure alarm
Refer to Flow chart 5.6
Air flow over-temp failure alarm
Refer to Flow chart 5.7
Wrong position of skin sensor 1 alarm
Check the position of skin sensor 1
Fan motor failure alarm
Refer to Flow chart 5.8
Fan failure alarm inside controller
Refer to Flow chart 5.9
Fan failure alarm inside sensor module
Refer to Flow chart 5.10
High deviation temperature failure alarm
Check whether there is heat source nearby, far away from the heat source
Low deviation temperature failure alarm
Check the closing condition of all doors and panels
Humidity sensor failure alarm
Refer to Flow chart 5.11
Sensor module failure alarm
5-13 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 5.3 ERROR CODES (Continued) TYPE
EXPLANATION AND CORRECTIVE ACTION
High deviation alarm for humidity
Check whether there is water inside of humidity chamber Check the humidity condition around
Oxygen concentration sensor 1 failure alarm
Refer to Flow chart 5.12
Oxygen concentration sensor 2 failure alarm
Refer to Flow chart 5.12
Difference alarm of oxygen sensor 1 and sensor 2
Refer to Flow chart 5.13
High deviation alarm concentration sensor 1
for
Oxygen
Check whether outside oxygen flow rate is overhigh
Low deviation alarm concentration sensor 1
for
Oxygen
Low deviation alarm for humidity
Check whether oxygen flow rate is small
ROM failure alarm inside main MCU
Replace Main MCU
Internal system failure alarm inside main MCU
Replace Main MCU
ROM failure alarm inside auxiliary MCU A
Replace sensor board
Internal system failure alarm inside auxiliary MCU A
Replace sensor board
ROM failure alarm inside auxiliary MCU B
Check Auxiliary MCU B inside sensor module(AT89C4051, replace the sensor board)
Internal system failure alarm inside auxiliary MCU B
Check Auxiliary MCU B inside sensor module(AT89C4051, replace the sensor board)
ADC failure alarm outside Auxiliary MCU A
Refer to Flow chart 5.14
ADC failure alarm outside Auxiliary MCU B
Refer to Flow chart 5.15
ADC failure alarm inside Auxiliary MCU A
Check Auxiliary MCU A inside sensor module (MS1210), replace the sensor board
Temperature heater system failure alarm
Refer to Flow chart 5.16
Humidity heater system failure alarm
Refer to Flow chart 5.17
EEPROM inside main MCU failure alarm
Replace main MCU
SRAM failure alarm
1. check the connection between SRAM and MCU 2. check SRAM (U7 on main board)
Real timer RTC failure alarm Communication failure alarm between main MCU and auxiliary MCU A Communication failure alarm between auxiliary MCU A and auxiliary MCU B Keyboard fault alarm
Battery failure alarm Power failure alarm
1. check the connection between RTC and MCU 2. check RTC (U23 on main board) 1. Check the connection of circuit between the main MUC and auxiliary MCU 2. Replace main MUC and auxiliary MCU Replace sensor board When the key board of the control panel on controller short-circuit alarms, please replace the control panel. Inner storage battery failure alarm: 1. Check the wire connection between the inner storage battery and the power board. 2. Replace the inner storage battery. Refer to Flow chart 5.18 5-14 I5
SERVICE MANUAL FOR INFANT INCUBATOR
5.6.4 Troubleshooting Flow Chart Turn off the power supply before reparing, remove the controller and sensor box from incubator and check the connecting wire from sensor box to the controller. Befor turn on the controller and sensor box, keep the power supply +5V, +12V of controller, elements and leads normal. Connect the wire of sensor box to the socket at the back of controller. In working state, at the same time, press key 4 and upper key, it will display the real temperature of isolated sensor at the display area of air sensor, display the real temperature of air flow sensor at the display area of set value, display the battery voltage at the area of skin sensor 1, and display motor value at the area of skin sensor 2( “7.2” means 1440 rpm), and the oxygen concentration indication window can indicate the second oxygen concentration, so it is easy for the repairman to find the problem.
AIR TEMPERATURE SENSOR FAILURE ALARM
No
Good connection?
Check the relative connection of air temperature sensor
Yes
Connect resistance 23166Ωinstead of the Air temperature sensor to socket J3 on sensor prefix board, then open the power supply
the temperature display is 31.0℃ and add the compensation temperature
System alarm?
No
Replace the air temperature sensor
of system set procedure
Yes
On Sensor board
U4-6 is 2.63V?
Check R8, C2
Check U4
FLOW CHART 5.1 AIR TEMPERATURE SENSOR FAILURE
I5
5-15
SERVICE MANUAL FOR INFANT INCUBATOR
``
ISOLATED TEMPERATURE SENSOR FAILURE ALARM
No
Good connection?
Check the relative connection of isolated temperature sensor
Yes
Connect resistance 23166Ωinstead of the Air temperature sensor to socket J2 on sensor prefix board, then open the power supply
the temperature display is 31.0℃ and add the compensation temperature
System alarm?
No
Replace isolated temperature sensor
of system set procedure
Yes
On Sensor board
U7-6 is 2.63V?
Check R11, C4
Check U7
FLOW CHART 5.2 ISOLATED TEMPERATURE SENSOR FAILURE
5-16 I5
SERVICE MANUAL FOR INFANT INCUBATOR
DIFFERENCE FAILURE BETWEEN AIR AND ISOLATED TEMPERATURE ALARM ``
On Sensor prefix board
Voltage of J2-1 and
No
Check the socket J2 and J3 on Air and Isolated temperature sensor
Goodare conection? J3-1 same?
Yes
On Sensor board
Voltage of J1-1
No
Check the connection between sensor prefix board and sensor board
and J1-3 are same?
Yes
On Sensor board
Voltage of U4-6 and
No
U7-6 are same?
Check the connection of R8, R11, C2, C4
Yes Replace U4 and U7
Still alarming?
No
Exit
Yes
Replace sensor board
FLOW CHART 5.3 DIFFERENCE FAILURE ALARM BETWEEN AIR AND ISOLATED TEMPERATURE SENSOR
I5
5-17
SERVICE MANUAL FOR INFANT INCUBATOR
FAN MOTOR SENSOR FAILURE ALARM
``
Good connection?
No
Check the relative connection of fan motor sensor
Yes
Connect 31.0℃ air flow temperature simulation ( equal to 23166Ω) instead of air flow sensor, turn on power supply
The temperature display is 31.0℃ and add the
System alarm?
No
compensation temperature
Replace air flow sensor
of system set procedure Yes
On sensor board
Pin11 of U4 is about
No
Check R26, C12
2.63V?
Check U7
FLOW CHART 5.4 FAN MOTOR SENSOR FAILURE ALARM
5-18 I5
SERVICE MANUAL FOR INFANT INCUBATOR
SKIN TEMPERATURE SENSOR 1 FAILURE ALARM
``
Good connection?
No
Check the relative connection of skin temperature sensor 1
Yes
Connect 31.0℃ skin temperature simulation ( equal to 23166Ω) instead of skin sensor 1, turn on power supply
The temperature display is 31.0℃ and add the
System alarm?
No
compensation temperature
Replace skin temperature seneor 1
of system set procedure Yes
On sensor board
Pin7 of U4 is about
No
Check R9, C3
2.32V?
Check U4
FLOW CHART 5.5 SKIN TEMPERATURE SENSOR 1 FAILURE ALARM
I5
5-19
SERVICE MANUAL FOR INFANT INCUBATOR
OVER-TEMP FAILURE
Yes Is there heat source nearby?
Far away from the heat source
No
Yes Air and isolated temperature
check air and isolated temperature sensor
Sensor is OK? No
Replace Air and Isolated temperature sensor
FLOW CHART 5.6 OVER TEMPERATURE FAILURE
AIR CIRCULATION FAILURE ALARM
Check whether the inlet
Check the fan motor sensor
Or outlet is blocked
Check the air flow condition of inlet and outlet
FLOW CHART 5.7 AIR CIRCULAATION FAILURE
5-20 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FAN MOTOR FAILURE ALARM
Motor rotating
No To sheet 2
Yes
On the power
Pulse at J12-4?
No check the motor fan checking
supply board
Yes
Pulse at U5-2?
On the power
No
check U5
supply board
Yes
On the main
Pulse at U20-16?
No
check the connect condition between
board
power board and main board Yes
On the main board
The wind speed is lower than 1000rpm
Yes Replace the motor
No Replace U20
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM(SHEET 1 OF 2) I5
5-21
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
Whether the power board for motor fan works
No Check the power cord
Yes
On the power supply board
Whether the motor fan can be rotated by manual operation
No Replace the motor
Yes
Replace the capacitor CM
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM(SHEET 2 OF 2)
5-22 I5
SERVICE MANUAL FOR INFANT INCUBATOR
Alarm Fan in the controllor Fault
No Fan in the controller is
Check the fan and the connection
rotating?
Yes
On powerboard
Votage on pin 3 of
No
J2 is about 1.5V?
Adjust the registor regulator VR1 to make the voltage on pin 3 of J2 as 1.5 V
Yes
On powerboard
Voltage on pin 5 of
No
U3 is about 1.8V?
Check resistor R16, R17
Yes
On powerboard
Voltage on pin 2 of
No Check resistor R4, R14
U3 is about 0.5V?
Yes
On powerboard
Output of pin U3-1 is high level?
No Replace U3
Yes To sheet 2
FLOW CHART 5.9 ALARM FAN IN THE CONTROLLOR FAULT (SHEET 1OF 2) I5
5-23
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
On powerboard
Output of pin U3-7
No
is high level?
Repalce U3
Yes
Output of pin U5-8 On powerboard
No
is low level?
Check U5 and the connenction between U5 and U3
Yes
Intput of pin U11-4 On mainboard
is high level?
No
Check the connenction between mainboard and powerboard
Yes
Replace the mainboard
FLOW CHART 5.9 ALARM FAN IN THE CONTROLLOR FAULT (SHEET 2 OF 2)
5-24 I5
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM FAN IN THE SENSOR MODULE FAULT
Fan in the sensor
No
Check the fan and the connection
No
Adjust the registor regulator VR1
module is rotating?
Yes
On sensor board
Votage on pin 2 of J1 is about 1.5V?
to make the voltage on pin 2 of J1 as 1.5 V
Yes
On sensor board
Voltage on pin 5 of
No Check resistor R13, R14
U1 is about 1.8V?
Yes
On sensor board
Voltage on pin 2 of
No Check resistor R6, R12
U1 is about 0.5V?
Yes
On sensor board
Output of pin U1-1
No Replace U1
is high level?
Yes To sheet 2
FLOW CHART 5.10 ALARM FAN IN THE SENSOR MODULE FAULT (SHEET 1 OF 2) I5
5-25
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
On sensor board
Output of pin U1-7
No
Repalce U1
is high level?
Yes
On sensor board
Input of pin U5-46
No
is high level?
Check the connenction between U5 and U1
Yes
On sensor board
Intput of pin U5-47 is high level?
No
Check the connenction between U5 and U1
Yes Replace the senso rboard
FLOW CHART 5.10 ALARM FAN IN THE SENSOR MODULE FAULT (SHEET 2 OF 2)
5-26 I5
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM HUMIDITY SENSOR FAULT
Connection is ok?
No
Check connection humidity sensor
Yes
On prepositive sensor board
Pin J4-10 is within the range of 0.6V~4.5V?
No
Check
the
humidity
sensor and the power Vcc
Yes
On sensor board
The voltage on pin U4-10 is same as pin
No
Check C1, R27, R1
J1-5 ?
Check U4
FLOW CHART 5.11 ALARM HUMIDITY SENSOR FAULT
I5
5-27
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM O2 SENSOR 1 FAULT
Connection is ok?
No
Check connection oxygen sensor 1
Yes Mesure the voltage between pin J6-4 and pin J6-3, and named “O2 Vol.” below
O2 Vol. > 9mV?
No
Replace the O2 sensor 1
Yes Calibration the O2 sensor. (See the second 6 of operator’s manual)
Alarm again?
No
Exit
Yes
The voltage on pin On Sensor board
U11-1 is 0V?
No
Check R28, R29, R32, R34, C6, C40, U11
Yes
To sheet 2
FLOW CHART 5.12 ALARM O2 SENSOR 1 FAULT (Sheet 1of 2) 5-28 I5
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet
On Sensor board
The voltage on pin U4-12 and U11-1 are same?
No
Check U11
Check U4
FLOW CHART 5.12 ALARM O2 SENSOR 1 FAULT (Sheet 2 of 2)
I5
5-29
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM O2 SENSORS ARE DIFFERENT
The difference value of
No
two O2 sensor is>3%?
Check the main MCU (U16 on the mainboard)
Yes Yes Mesure the voltage between pin J6-4 and pin J6-3, and named “O2 Vol.” below
O2 Vol. > 9mV?
No
Replace the O2 sensor 1
Yes Calibration the O2 sensor. (See the section 6 of operator manual)
Alarm again?
No
Exit
Yes
Replace the mainboad and sensor board
FLOW CHART 5.13 ALARM O2 SENSORS ARE DIFFERENT 5-30 I5
SERVICE MANUAL FOR INFANT INCUBATOR
ADC FAILURE ALARM OUTSIDE AUXILIARY MCU A
Voltage on the joint On sensor board
of R18 and R19 is
No
Check R18, R19
2.5V?
Yes
On sensor board
Votage on pin 13
No
of U4 is 2.5V?
Check the connection between U4 and R18, R19
Yes
The connection On sensor board
between U4 and U5 is ok?
No
Connecting coreectly or replace the sensor board
Yes Replace the sensor board
FLOW CHART 5.14 ADC FAILURE ALARM OUTSIDE AUXILIARY MCU A
I5
5-31
SERVICE MANUAL FOR INFANT INCUBATOR
ADC FAILURE ALARM OUTSIDE AUXILIARY MCU B
Voltage on the On sensor board
No
joint of R22 and
Check R22, R23
R23 is 2.5V?
Yes
On sensor board
Votage on pin 9 of
No
U7 is 2.5V?
Check the connection between U7 and R22, R23
Yes
The connection On sensor board
between U7 and
No
Connecting coreectly or replace the sensor board
U9 is ok?
Yes Replace the sensor board
FLOW CHART 5.15 ADC FAILURE ALARM OUTSIDE AUXILIARY MCU B
5-32 I5
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM TEMPERATURE HEATER SYSTEM FAULT
Is heater On powerboard
No Connect well
connected with pin 1 and pin 4 of J9 ?
Yes Switch off the power of incubator, and disconnect the heater from the controller, then measure the resistance value of the heater.
The resistance
No
value of heater is
Replace the heater
abour 165 Ohms?
Yes Disconnect the sensor box and the mainboard, connect pin 11 of J11 and pin 7 of J13 on powerboard to VCC power, and then switch on the controller.
Voltage on Q1, Q3 On powerboard
are about 0.7V?
No
Check R1, Q1, R3, R8, Q3
Yes To sheet 2
FLOW CHART 5.16 ALARM TEMPERATURE HEATER SYSTEM FAULT (SHEET 1OF 3) I5
5-33
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
Voltage on control On powerboard
No
Replace Q1, Q3
port of relay S1, K1 is about 12V?
Yes Cnnect pin 11 of J11 and pin 7 of J13 on powerboard to Gnd power instead, and then switch on the controller.
Voltage on B port of On powerboard
audion Q1, Q3 is about
Yes
Check R1, Q1, R3, R8, Q3
0.7 V?
No
Voltage on control On powerboard
Yes
port of relay S1, K1
Replace Q1, Q3
is about 12V?
No Disconnect these connections for testing, and connect the mainbaord and the sensor box, then switch on the controller.
To sheet 3
FLOW CHART 5.16 ALARM TEMPERATURE HEATER SYSTEM FAULT (SHEET 2 OF 3)
5-34 I5
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 2
Alarm again?
No
Exit
Yes Replce the relay S1 and K1
Alarm again?
No Exit
Yes Check the connection between powerboard and mainboard, or replace the mainboard.
Alarm again?
No
Exit
Yes Check the connection between powerboard and sensor board, or replace the sensor board.
FLOW CHART 5.16 ALARM TEMPERATURE HEATER SYSTEM FAULT (SHEET 3 OF 3)
I5
5-35
SERVICE MANUAL FOR INFANT INCUBATOR
ALARM HUMIDITY HEATER SYSTEM FAULT
Is heater On powerboard
No Connect well
connected with pin 2 and pin 5 of J9 ?
Yes Switch off the power of incubator, and disconnect the heater from the controller, then measure the resistance value of the heater.
The resistance
No
value of heater is
replace the humidity heater.
abour 360 Ohms?
Yes Disconnect the sensor board and the main board, connect pin 12 of J11 on powerboard to VCC power, and then switch on the controller.
Voltage on B port On powerboard
of audion Q2 is
No Check R2, Q2
about 0.7V?
Yes To sheet 2
FLOW CHART 5.17 ALARM HUMIDITY HEATER SYSTEM FAULT (SHEET 1OF 3) 5-36 I5
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
On powerboard
Voltage on control
No
port of relay S2 is
Replace Q2
about 12V?
Yes Cnnect pin 12 of J11 on powerboard to Gnd power instead, and then switch on the controller.
Voltage on B port of On powerboard
Yes Check R2, Q2
audion Q2 is about 0.7 V?
No
Voltage on control On powerboard
Yes
port of relay S2 is
Replace Q2
about 12V?
No Disconnect these connections for testing, and connect the mainbaord and the sensor box, then switch on the controller.
To sheet 3
FLOW CHART 5.17 ALARM HUMIDITY HEATER SYSTEM FAULT (SHEET 2 OF 3)
I5
5-37
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 2
Alarm again?
No Exit
Yes Replce the relay S2
Alarm again?
No
Exit
Yes Check the connection between powerboard and mainboard, or replace the mainboard.
FLOW CHART 5.17 ALARM HUMIDITY HEATER SYSTEM FAULT (SHEET 3 OF 3)
5-38 I5
SERVICE MANUAL FOR INFANT INCUBATOR
POWER FAILURE ALARM
The power of the
No Check the power
controller is ok?
Yes
On DC powerboard
The fuse F3 is ok?
No
Replace the fuse F3
Yes
On DC powerboard
The output voltage on
No
second part of the
Check the fuse F1, F2, and the transfromer
transform is correct?
Yes
On DC powerboard
The output voltage
No
Check U2, U3, C2, C4, C5, C7, C8, C10,
on J1 is correct?
C11, C12 Yes
Cross point of R10, On main board
R25 is +2.5V?
No
Check R10, R25
Yes To sheet 2
FLOW CHART 5.18 POWER FAILURE ALARM (SHEET 1OF 2) I5
5-39
SERVICE MANUAL FOR INFANT INCUBATOR
From sheet 1
On power board
Crosspoint of R11, R26 is 2.53V?
No Replace R11, R26
Yes
Crosspoint of On power board
No Replace R24, R27
R24, R27 is 2.1V?
Yes
On power board
Pins7, 1 of U4 is high level?
No Replace U4
Yes
On power board
Pin 6 of U5 is
No Replace U5
low level?
Yes
On main board
Pin6 of U11 is high level?
No
Check connection between power
board
and
main
board Yes Replace main board
FLOW CHART 5.18 POWER FAILURE ALARM (SHEET 2 OF 2)
5-40 I5
SERVICE MANUAL FOR INFANT INCUBATOR
SECTION 6 MAIN PARTS LIST AND PARTS NUMBER 6.1 GENERAL This section provides parts lists for infant incubator. Part numbers of accessories are provided below:
05
03 04 02
01
FIGURE 6.1 PARTS LOCATION DIAGRAM, THE WHOLE DEVICE ASSEMBLY TABLE 6.1 REPLACEMENT PARTS LIST FOR WHOLE SPARE PARTS ITEM No.
DESCRIPTION
1
VHA Basinet
13.10.000.00
2
Controller
13.07.000.00
3
Hood
13.02.001.01S
4
Base
13.04.008.02S
5
I. V. Pole Groupware
13.11.001.00S 6-1
I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.1 PARTS LOCATION DIAGRAM, HOOD ASSEMBLY (SHEET 1-2)
6-2 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.1 PARTS LOCATION DIAGRAM, HOOD ASSEMBLY (SHEET 2-2)
6-3 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.2 HOOD ASSEMBLY REPLACEMENT PARTS LIST ITEM No.
DESCRIPTION
PART NUMBER
1
Hood
13.02.002.00
2
Access panel
13.02.006.00
3
Window touch head
05.02.013.00
4
Latch seat
05.02.015.00
5
Cross groove oval head bolt M3×10
86.44.148.00
6
Bolt cover M3
05.02.046.00
7
Clamp
13.02.008.00
8
Iron scrap
13.02.009.00
9
Spring
13.02.010.00
10
Cross groove oval head bolt M3×14
86.44.150.00
11
Nut M3
86.44.334.00
12
Soft port for I.V. Pole
13.02.011.00
13
Stopper
12.02.046.00
14
B type round pin3×25
86.44.471.00
15
Left gemel for front door
13.02.012.00
16
Cross groove oval head bolt M5×16
86.44.184.00
17
Flat washer 5
86.44.444.00
18
Spring washer 5
86.44.418.00
19
Cross groove oval head bolt M3×12
86.44.149.00
20
Right pad (I)
13.02.013.00
21
Left pad (II)
13.02.014.00
22
Back Left gemel (I)
13.02.015.00
23
Back Left gemel (II)
13.02.018.00
24
Front right gemel (I)
13.02.017.00
25
Front right gemel (II)
13.02.019.00
26
Front left gemel (II)
13.02.020.00
27
Position limit pole
13.02.021.00
28
Cross groove countersunkhead bolt M3×8
86.44.216.00
29
Inner front panel
13.02.022.00
30
Heat insulation board for Inner front panel
13.02.035.00
31
Cross groove head bolt ST2.9×9.5
86.44.268.00 6-4 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.2 HOOD ASSEMBLY REPLACEMENT PARTS LIST (CONTINUE) ITEM No.
DESCRIPTION
32
Fix fittings for inner wall
13.02.026.00
33
Special bolt
13.02.034.00
34
Right pad (II)
13.02.043.00
35
Back right gemel (I)
13.02.028.00
36
Back right gemel (II)
13.02.029.00
37
Fastener
13.02.030.00
38
Shield cap
13.02.031.00
39
Door panel sheath
12.02.038.00
40
Shield sheath
13.02.042.00
41
Cross groove head bolt ST2.9×9.5
86.44.256.00
42
Door latch
12.02.040.00
43
Spring (Ⅳ)
12.02.039.00
44
Latch handle
05.02.035.01
45
Spring (I)
05.02.034.00
46
Roller latch sheath
05.02.033.00
47
Latch stopper
05.02.032.00
48
O type seal circleφ12×1.9
86.44.547.00
49
Access door’s cuff
05.02.029.00
50
Rubber seal for window(I)
13.02.032.01
51
Rubber seal for window(II)
13.02.033.02
52
Torsional Spring
05.02.027.00
53
Window’s shaft
05.02.026.00
54
Bolt cover M5
05.02.047.00
55
Moving assembly
05.02.022.00
56
Fixing circle
05.02.023.00
57
Set assembly
05.02.024.00
58
Iris door’s cuff
05.02.025.00
59
Cross groove oblate head bolt 5×16
86.44.290.00
60
Locating bush
13.02.038.00
61
O type seal φ6.3×1.8
86.44.509.00 6-5
I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.2 HOOD ASSEMBLY REPLACEMENT PARTS LIST (CONTINUE) ITEM No.
DESCRIPTION
PART NUMBER
62
Hexagon nutM5
86.44.338.00
63
Special bolt (II)
13.02.039.00
64
Rotating gasket
13.02.040.00
65
Gemel blank cap
13.02.041.00
66
Lock fastener
05.02.040.01
67
Latch stopper
05.02.043.00
68
Cross groove countersunk bolt 3×25
86.44.212.00
69
Lock supporting block
05.02.042.00
70
Locked base
05.02.044.00
71
Flat washer 3
86.44.440.00
72
Door lock shaft
05.02.041.00
73
Transition pad
13.02.044.00
74
Axle sleeve
05.02.045.00
75
Thread locking agent
86.61.119.00
6-6 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.2 PARTS LOCATION DIAGRAM, BASE ASSEMBLY (SHEET 1-2)
6-7 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.2 PARTS LOCATION DIAGRAM, BASE ASSEMBLY (SHEET 2-2) 6-8 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.3 BASE ASSEMBLY REPLACEMENT PARTS LIST ITEM No.
DESCRIPTION
1
Handle
13.04.009.00
2
Cross groove head bolt M5×16
86.44.184.00
3
Spring washer 5
86.44.418.00
4
Flat washer 5
86.44.444.00
5
Inner hexagon column bolt M6×20
86.44.099.00
6
Spring washer 6
86.44.420.00
7
Flat washer 6
86.44.446.00
8
Catheter
13.03.001.00
9
Water reservoir
10
Upper shell A
13.04.011.00
11
Bottom shell
13.04.015.01
12
Air-oxygen seperation controlling device
13
Air filter
13.04.018.00
14
Spring washer 4
86.44.414.00
15
Flat washer 4
86.44.442.00
16
Cross groove head bolt M4×20
86.44.167.00
17
Cover for box
13.04.021.00
18
Flat washer 3
86.44.440.00
19
Spring washer 3
86.44.412.00
20
Cross groove oval head bolt M3×10
86.44.271.00
21
Protection mantle
12.07.024.00
22
Checking board
23
Cover for Protection mantle
12.07.025.00
24
Raster plate
12.07.026.00
25
Notch clamp bolt M3×5
86.44.130.00
26
Cross groove head bolt ST2.9×9.5
86.44.256.00
13.05.001.00S
13.16.007.00S
12.07.084.00S
6-9 I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.3 BASE ASSEMBLY REPLACEMENT PARTS LIST (CONTINUE) ITEM No.
DESCRIPTION
PART NUMBER
27
Cross groove head bolt M3×10
86.44.148.00
28
Two-Segment Adjustable Wiring stator TS-0607
86.45.018.00
29
Label for warning
13.04.081.00
30
Start-up CBB61-500V-0.27uF±5%
86.51.032.00
31
Fixing pin for capacitor
12.07.048.00
32
Left shaft
13.04.025.00
33
Right tilt machanism
34
Cross groove countersunk head bolt M5×8
86.44.235.00
35
Cross groove head bolt M4×10
86.44.161.00
36
Charles clip CC-3
86.45.005.00
37
Impeller groupware
38
Shell plug
39
Sensor connecting wire
40
Cross groove oval head bolt M3×16
86.44.274.00
41
Hex nut M3
86.44.334.00
42
Motor 90YY2
86.60.002.00
43
Fixed circle for motor
44
Cross groove head bolt M4×12
86.44.162.00
45
Cross groove countersunk head bolt M3×20
86.44.211.00
46
Heater group
47
Pad
13.04.053.00
48
Heat insulation board
13.04.071.00
49
Heat dissipation slice
13.04.054.00
50
Cover
13.04.055.00
51
Cover board
13.04.056.00
52
Lock groupware
53
Non-shedding bolt
13.04.072.00
54
O type seal circleφ3.75×1.8
86.44.507.00
13.04.026.00S
12.07.077.00S 12.05.004.00 13.04.044.00S
13.04.046.01S
13.04.047.00S
13.04.090.00S
6-10 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.3 BASE ASSEMBLY REPLACEMENT PARTS LIST (CONTINUE) ITEM No.
DESCRIPTION
55
Self-control oxygen inlet groupware
13.16.032.01S
56
Cover groupware
13.04.057.00S
57
Right tilt machanism
13.04.059.00S
58
Cross groove head bolt M5×8
59
Left shaft groupware
60
Handwheel
13.04.067.00
61
Cover cap
13.04.069.00
62
Drawer group
63
Seal of humidifier rim
13.04.023.00
64
Press board for seal
13.04.024.00
65
Cross groove oval head bolt M3×6
86.44.158.00
66
Heat dissipation slice label
13.04.079.00
67
Label for high-temp warning
13.04.070.00
68
Air flow sensor
13.07.082.00
69
Earthed lable
12.07.047.00
70
Cross groove head bolt M3×8
86.44.159.00
71
Warning label
12.07.029.00
72
Motor absorbing pad
13.04.073.00
73
Upper shell B
13.04.074.00
74
Handle
13.04.075.00
75
Seal for shell
13.04.076.00
76
Plastic washer
13.07.008.00
77
Cross groove countersunk head bolt M3×8
86.44.216.00
78
Earthed clamp
13.04.082.00
79
Cross groove countersunk head bolt M3×6
86.44.215.00
80
Cross groove countersunk head bolt M4×16
86.44.220.00
81
Connecting column
13.04.083.00
86.44.195.00 13.04.064.00S
13.05.009.00S
6-11 I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.4 PARTS LOCATION DIAGRAM, IV POLE ASSEMBLY TABLE 6.4 I.V. POLE ASSEMBLY REPLACEMENT PARTS LIST ITEM No.
DESCRIPTION
PART NUMBER
1
Flat washer 8
86.44.448.00
2
Spring washer 8
86.44.423.00
3
Inner hexagon colume bolt M8×40
86.44.118.00
4
Seat for the monitor shelf
13.12.012.00
5
Fixing seat
13.11.013.00
6
Support pole for I.V. Pole
13.04.008.00
7
Tray
06.11.005.00
8
Spring washer 6
86.44.420.00
9
Flat washer 6
86.44.446.00
10
Inner hexagon colume bolt M6×20
86.44.099.00
11
I.V.Pole
13.11.001.00S
12
Monitor shelf
13.12.001.00S
13
Inner hexagon colume bolt M5×30
86.44.089.00
14
The support pole for monitoring shelf
13.12.005.00
15
Clamp seat
13.12.009.00
16
Cross groove head bolt M6×16
86.44.201.00
17
Inner hexagon colume bolt M8×35
86.44.116.00
6-12 I5
SERVICE MANUAL FOR INFANT INCUBATOR
40
FIGURE 6.5 PARTS LOCATION DIAGRAM, LIFT STAND 6-13 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.5 LIFT STAND REPLACEMENT PARTS LIST ITEM No.
DESCRIPTION
PART NUMBER
1
Left drawer
13.10.042.00S
2
Weight warning lableA
13.10.015.00
3
Outer serrate clamp washer 6
86.44.398.00
4
Inner hexagon bolt M6×25
86.44.100.00
5
Flat washer 6
86.44.445.00
6
Spring washer 5
86.44.418.00
7
Flat washer 5
86.44.444.00
8
Outer serrate clamp washer 5
86.44.397.00
9
Cross groove head bolt M5×12
86.44.180.00
10
Connection board
13.10.016.00
11
Protectin mantle goupware
13.10.017.00
12
Weight warning lable B
13.10.019.00
13
Right drawer
14
Winding frame
13.10.022.00
15
Cross groove head bolt M4×10
86.44.161.00
16
Spring washer 4
86.44.414.00
17
Flat washer 4
86.44.442.00
18
Outer serrate clamp washer 4
86.44.395.00
19
power cable
86.46.013.00
20
Protective bottom shell
13.10.023.00
21
Lift support
22
Spring washer 12
13.10.052.00S
13.10.060.00S 86.44.404.00
6-14 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.5 LIFT STAND REPLACEMENT PARTS LIST (CONTINUE) ITEM No.
DESCRIPTION
23
Flat washer 12
86.44.431.00
24
Flat Axletree wheel
86.52.002.00
25
Inner hexagon bolt M8×25
86.44.112.00
26
Spring washer 8
86.44.422.00
27
Flat washer 8
86.44.447.00
28
Switch rack groupware
29
Hex nut M3
86.44.334.00
30
Spring washer 3
86.44.412.00
31
Blade terminal OT1.5-3
86.55.129.00
32
Earthed label
12.07.047.00
33
Stand
13.10.038.00
34
Adjustable wiring clamp
86.45.032.00
35
Power cable for the controller
86.46.002.00
36
Weight warning lable C
13.10.040.00
37
Flat Axletree wheel with brakes
86.52.006.00
38
Outer serrate clamp washer 8
86.44.399.00
39
Spring washer 6
86.44.419.00
40
Power cable
86.46.012.00
13.10.058.00S
6-15 I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.6 PARTS LOCATION DIAGRAM1, CONTROLLER ASSEMBLY TABLE 6.6 CONTROLLER ASSEMBLY REPLACEMENT PARTS LIST 1 ITEM No.
DESCRIPTION
PART NUMBER
1
Middle frame groupware
13.07.037.00S
2
Groupware for indicating frame
13.07.053.00S
3
Electric nameplate
13.07.032.00
4
Cross groove countersunk head bolt M3×8
86.44.216.00
5
Sensor box
13.07.059.01S
6
RS232 label
13.07.036.00
7
Earthed label
12.07.047.00
8
Skin sensor
13.07.088.00
6-16 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.7 PARTS LOCATION DIAGRAM 2, CONTROLLER ASSEMBLY
6-17 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.7 CONTROLLER ASSEMBLY REPLACEMENT PARTS LIST 2 ITEM No.
DESCRIPTION
PART NUMBER
1
Shell
13.07.001.00
2
Cross groove countersunk head bolt M3×8
86.44.216.00
3
Groupware for power socket
4
Earthed label
12.07.047.00
5
Flat washer 3
86.44.440.00
6
Spring washer 3
86.44.412.00
7
Cross groove head bolt M3×6
86.44.158.00
8
Wiring terminal
86.55.145.00
9
Hex nut M3
86.44.334.00
10
Label for switch
13.07.002.00
11
Power module
13.07.044.00S
12
Cross groove countersunk head bolt M3×5
86.44.214.00
13
Cross groove countersunk head bolt M3×4
86.44.519.00
14
Ferrite
86.60.082.00
15
Groupware for power socket
12.07.089.00S
16
Groupware for power switch
13.07.051.00S
17
Cross groove countersunk head bolt M3×16
86.44.210.00
18
Fan cover
13.07.037.00
19
Fan cushion
13.07.038.00
20
Axial-flow fan groupware
13.07.052.00S
13.07.043.00S
6-18 I5
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.5 PARTS LOCATION DIAGRAM 3, CONTROLLER ASSEMBLY TABLE 6.5 CONTROLLER ASSEMBLY REPLACEMENT PARTS LIST 3 ITEM No.
DESCRIPTION
1
Groupware for front panel
13.07.010.00
2
LCD protectibve board
13.07.011.00
3
Main board
13.07.058.00
4
Front panel
13.07.014.00
5
Cross groove countersunk head bolt M3×8
86.44.216.00
6
Hexagon Isolated Roller S6×16
13.07.016.00
7
Nut cover
13.07.006.00
8
Cross groove pan head bolt M3×10
86.44.148.00
9
Hex bolt M3
86.44.334.00
10
Spring washer 3
86.44.412.00
11
Flat washer 3
86.44.440.00
12
LCD indication board
86.58.051.00
13
Cross groove countersunk head bolt M3×10
86.44.207.00
6-19 I5
PART NUMBER
SERVICE MANUAL FOR INFANT INCUBATOR
FIGURE 6.6 PARTS LOCATION DIAGRAM 4, CONTROLLER ASSEMBLY
6-20 I5
SERVICE MANUAL FOR INFANT INCUBATOR
TABLE 6.6 CONTROLLER ASSEMBLY REPLACEMENT PARTS LIST 4 ITEM No.
DESCRIPTION
1
Bottom shell for sensor
13.07.024.00
2
Top shell for sensor
13.07.025.00
3
Socket for skin sensor 2
4
Label for sensor box
13.07.021.00
5
Screw stopper
13.07.027.00
6
Cross groove countersunk head bolt M3×8
86.44.216.00
7
Sensor cover board
13.07.026.00
8
Cross groove head bolt M3×18
86.44.152.00
9
Label for skin temperature
13.07.019.00
10
Socket for skin sensor 1
13.07.066.00S
11
Position limit switch
13.07.068.00S
12
Groupware for prepositive board of sensor
13.07.062.00S
13
LED reflector shell
13.07.028.00
14
Groupware for sensor board
13.07.060.00
15
Wire clip 6P- 4
86.45.002.00
16
Wiring for sensor
13.07.061.00S
17
Socket for baby scale
13.07.067.00S
18
Label for baby scale
13.07.065.00S
13.07.030.00
6-21 I5
PART NUMBER
SERVICE MANUAL FOR YP-2000 INCUBATOR
SECTION 7 DIAGRAMS 7.1 GENERAL The section provides schematic and wiring diagrams for Infant Incubator and circuit diagrams, connection diagram for temperature controller boards.
S:Power switch for controller
G:Internal battery GP17R8H Ni-MH8.4V/170mAh
X1:Powr input socket for controller
F:DC12V axial-flow fan
X2,X4:Wiring terminal
T:Transformer R20
A1:Power board
A2: Main board
A3: DC power board WIRING DIAGRAM OF TEMPERATURE CONTROLLER I5
7-1
D1
XZ33
D2 XZ29
X3
XZ16
XZ17
XZ14
2(N)
1(L)
XZ27
XZ28 1(L)
XZ24 XZ26
XZ29
SB2
1
2
3
X2
XZ15
XZ13 XZ17
XZ12 XZ16
XZ25 XZ27
2(N)
XZ26
2
SB1
3
1
2 XZ2
XZ12 XZ13 XZ14 XZ15
XZ8-1
2 XZ1 4
1
F3
XZ2 2 XZ24
1
XZ9
XZ8-2
1
3
XZ25
2
XZ3
1
X1
L
XZ1
F1
CT
G
XZ32
1
T
N
XZ4
XZ8-1 XZ6 2
1 3 4
1
XZ11
XZ10
S1
XZ30
F2
1
XZ5
1
XZ4 1
D
M
1
4
XZ5
XZ9 XZ7
XZ18 XZ19 XZ20 XZ21
XZ18 XZ22
XZ19 XZ23
XZ20
2
SB3
3
1
SB4
1
2
3
XZ21
XZ23
XZ22
SERVICE MANUAL FOR YP-2000 INCUBATOR
3
3
SB1~SB4: Jiggle switch
S1: Power switch
T: Transformer
F3: 800mAL/250V type T fuse
M: Lift motor
F1, F2: 6.3AH/250V type F fuse
CT: Power supplying plug
X1,X2,X3: Socket
WIRING DIAGRAM OF LIFT STAND
7-2
I5
XZ31
XZ33
SERVICE MANUAL FOR YP-2000 INCUBATOR
K1: Temperature control switch KSD-II-85 P1~P4:φ6.3retagular plug
K2: Temperature control switch KSD-II-85 M:Motor for controller
J1:Plug MC100-50806
J2:Plug MC420-35010
J5:Plug YT-005-4( To Motor) J8:Plug YT-005-5(To humidifier )
J6:Plug YT-005-2( To Steamer) J10,J10’:Plug socket JD28050-3P, JD28050-3R
K3: Minor switch Z15G1305 (Applied for reservior location detecting) Start-up capacitance:0.27μ/400V J7:Plug YT-005-3( To O2 Sensor ) J3,J3’:Plug socket JD28050-6P, JD28050-6R
INCUBATOR WIRE CONNECTING DIAGRAM
7-3
SERVICE MANUAL FOR YP-2000 INCUBATOR
DC POWER BOARD ASSEMBLY DIAGRAM 1
7-4 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
DC POWER BOARD ASSEMBLY DIAGRAM 2
7-5
SERVICE MANUAL FOR YP-2000 INCUBATOR
DC POWER BOARD PARTS LIST No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
1
C1~C5,C12
Capacitor 0805-0.1uF±20%
6
2
C6~C8
Capacitor CD11-25V-470uF±20%
3
3
C9~C11
Capacitor CD11-16V-2200uF±20%
3
4
B1,B2
Rectification bridge GBU10005
2
5
U1,U2
Integrated circuit MC7805CT
2
6
U3
Integrated circuit MC7812CT
1
Radiator GHSI-45
3
7 8
F1,F2,F3
Seat for fuse 520101
3
9
F1
Fuse T1AL/250V
1
10
F2
Fuse T800mAL/250V
1
11
F3
Fuse T160mAL/250V
1
12
T1
Transformer R20
1
13
J1
Socket XY4-6S
1
14
J2,J4
Socket XY4-3S
1
15
J3
Socket XY4-4S
1
7-6 I5
7-7
3PIN
J4
PRI.
3 2 1 T160mA
F3
SEC.
TRANSFORMER
3 2 1
4 3 2 1
3PIN TRANS-IN
J2
4PIN TRANS-OUT
J3
4
3
T800mA/250V
F2
T1A/250V
F1
-
+
4
3
-
+
GBU10005
AC
AC
B2
GBU10005
AC
AC
B1
2
1
2
1
Va
C12 104
C2 104
C1 104
IN
U3 MC7812CT
Vin
U2 MC7805CT
Vin
+12 VCC VLCD+
1 2 3 4 5 6
C8 Comment: 470uF
1
C7 470uF
1
C6 470uF
1
U1 MC7805CT
GND 2 GND 2 GND
ELEMENTS DIAGRAM FOR DC POWER BOARD 6PIN
J1
2
Va
OUT
+5V
+5V
3
3
3
C11 2200uF
+12
C10 2200uF
VCC
C9 2200uF
C5 104
C4 104
C3 104
VLCD+
SERVICE MANUAL FOR YP-2000 INCUBATOR
SERVICE MANUAL FOR YP-2000 INCUBATOR
POWER BOARD ASSEMBLY DIAGRAM 1
7-8 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
POWER BOARD ASSEMBLY DIAGRAM 2
7-9
SERVICE MANUAL FOR YP-2000 INCUBATOR
POWER BOARD PARTS LIST No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
1
R1~R3,R6,R7,R9,R35~R38, R43
Resistor 0805-1KΩ±5%
11
2
R4
Resistor 0805-1KΩ±1%
1
3
R5
Resistor 0805-2KΩ±5%
1
4
R8,R12,R13,R18
Resistor 0805-10KΩ±5%
4
5
R10,R25,R26
Resistor 0805-2KΩ±1%
3
6
R11
Resistor 0805-7.5KΩ±1%
1
7
R14
Resistor 0805-9.1KΩ±1%
1
8
R15,R41
Resistor 0805-100KΩ±1%
2
9
R16
Resistor 0805-33KΩ±1%
1
10
R17
Resistor 0805-18KΩ±1%
1
11
R19
Resistor 0805-100Ω±5%
1
12
R20~R23, R44
Resistor 0805-10KΩ±5%
5
13
R24
Resistor 0805-3KΩ±1%
1
14
R27
Resistor 0805-2.2KΩ±1%
1
15
R28
Resistor 0805-10KΩ±1%
1
16
R29~R34
Resistor 0805-200Ω±5%
6
17
R39
Resistor 0805-510Ω±5%
1
18
R40
Resistor 0805-30KΩ±1%
1
19
R42
Resistor 0805-499KΩ±5%
1
20
VR1
Potentiometer 3296-100Ω
1
21
C1~C3,C10
Capacitor SMD-25V-150uF±20%
4
22
C4,C5,C14,C15,C18,C24,C25,C33
Capacitor 0805-0.1uF±20%
8
23
C6~C9,C11,C32
Capacitor 0805-0.01uF±20%
6
24
C19,C20
Capacitor 0805-2200pF±20%
2
25
C21,C22,C23
Capacitor CA45B-10uF±20%
3
26
C28~C31
Capacitor CA45A-25V-1uF±20%
4
27
C34
Capacitor 1210-25V-10uF±20%
1
28
D1~D7
Diode SM4007-D0214
7
29
D8
Diode HZ5.6BP
1
30
D10
Diode 1N5339B
1
31
D11
Diode HZ11B1
1
32
Q1~Q5,Q10,Q11
Diode MMBT3904LT1
7
7-10 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
POWER BOARD PARTS LIST (Continue) No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
33
Q6
Diode MMBT3906LT1
1
34
Q7
Diode 2SD667
1
35
U1,U2
Integrated circuit MC7805BDTG
2
36
U3,U4,U9,U15
Integrated circuit LM358D
4
37
U5
Integrated circuit SN74AHCT00D
1
38
U8
Integrated circuit 4N35
1
39
U10,U11
Integrated circuit K1010
2
40
U12
Integrated circuit MAX232CSE
1
41
U13
Integrated circuit KP1020
1
42
U14
Integrated circuit MAX489CSD
1
43
U16
Integrated circuit SN74LVC1G3157DBVR
1
44
J1
Socket ME040-50803
1
45
J2
Socket XY2.5-3S
1
46
J3
Socket XY4-3S
1
47
J4
Socket XY2.5-2S
1
48
J5
Socket XY4-2S
1
49
J6
Socket XY2.5-4S
1
50
J7
Socket ME040-50802
1
51
J8
Socket XY4-6S
1
52
J9
Socket ME040-50806
1
53
J10
Buzzer EFM230
1
54
J11
Socket DC-2-26S
1
55
J12
Socket ME040-35010
1
56
J13,J14
Socket DD09SZG
2
57
T1,T2
Inductor TVA1421-01
2
58
DC2
DC-DC F0505S-1W
1
59
K1
Solid relay JW1aFSN-DC12V
1
60
S1,S2
Solid relay CX240D5
2
61
F1,F2
Fuse F 6.3AH/250V
2
62
F3,F4
Fuse 1812L050-C
2
7-11
1 2 3
3 2 1
VCC
VCC
VCC
VCC
VCC
6 5 4 3 2 1
RENB
200
R33
WENB
200
R32
TOTXD2
200
R30
1K
R35
Vout-
Vout+
KP1020
U13B
KP1020
U13A
K1010
U11
K1010
U10
H0505S-1W
GND
VIN
DC2
TORXD2
2
1
Bell
+5
200
R29
5
7
8050
Q4
R39 510
JW1aFSN
K1B
1K
R38
1K
R37
1K
R36
200
+5D
+5D
+5D
+5D
R34 200
+5D
ctr_bell
D8 HZ5.6BP
R31
1K
R6
C31 1uF
INDUCTOR2
T2
INDUCTOR2
T1
6
5
2 1
2 1
C28 1uF
LM358
+ -
+ -
7
1 3 4 5 T1OUT T2OUT
VEE
2
5
14 13 12 11 10 9 8
6
13 8
14 7
1uF
C30
1uF
A B Z Y
+5D
+5D
8050
Q1
8050
Q3
0.1uF
C25
Q11 8050
C24 0.1uF
ctr_hum
ctr_ssr
ctr_relay
cool fan
+12
DB9
J14
3 2 1
+12
R4 1K
R17 18K
R16 33K
VR1 100
R14 9.1K
VCC
C23 10uF3
RS 232 interface
1 6 2 7 3 8 4 9 5
3PIN
J2
humheater_state
VCC
1K
R2
R21 10K
8050
1K
R1
1K
R3
Q10 8050
Q2
C29 +5D
C22 10uF
1K
R9
2 16
+5D
VCC NC A B Z Y NC
SN74LVC1G3157
B1VCC B2 A S GND
U6
MAX489CSD 3 1 4 6
VDD VCC
1N4007
D2 1N4007
+12
D1 1N4007
+12
1N4007
D7
MAX232CSE
GND
C21 10uF
JW1aFSN D3
K1A
1N4007
R1OUT R1IN R2OUT R2IN
T1IN T2IN
U14 1 NC 2 RO 3 /RE 4 DE 5 DI 6 GND 7 GND
15
12 9
C1+ C1C2+ C2-
U12
3 4
3 4
+12
LM358
11 10
2200pF
10K C20
R23
U9B
RP1A40D5
~ ~
S2
RP1A40D5
~ ~
S1
2200pF
10K C19
R22
1K
R7
2
6PIN/empty
J10
CON6
1 2 3 4 5 6
AC-L
power switch
J7
AC-N
D4
1
6
5
2
3
R26 2K
R25 2K
LM358
LM358 U3B
U3A
R11 7.5K tvb
+12
C14 0.1uF
VCC
7
1
R27 2.2K
R24 3K tvc
+5
+12
1N4002
1N4002 D6
D5
R10 2K tva
VCC
bat+
PTC 500mA
F3
U5C 9
6
+5
LM358
C15 0.1uF
ctr_bat
SN74AHC00D
8
7
1
10K
R12
fan_state
0.1uF
LM358 U4B
U4A
ctr_o2 R5 2K
5
tvc
2
tva
tvc
3
+5
C5
U1 MC7805BDTG 1 3 IN OUT C2 GND 150uF
tvb
10
J8 6PIN
U2 MC7805BDTG 1 3 IN OUT C1 GND 150uF
+12 VCC VLCD+
4
3-6 Motor
2-5 Steamer
1-4 Heater
J9
T6.3A
T6.3A F2
F1
1
heater_state
VCC R20 10K
4
AC POWER OUTPUT For DC Powerbaord
3PIN
J3
3PIN
J1
2
3
0.1uF U9A
C18
6 5 4 3 2 1
AC POWER INPUT
+12
4N35
R13 10K
Q5 8050
R18 10K
VCC
100
R19
Q6 8550
+12
VCC
R28 10K
R40 30K
bat+
C34 10uF
bat_voltage C32 0.01uF
2
1 3
1K
R43
10K
R8
VCC
R44 10K
C7 0.01uF
SN74AHC00D
U5A
tank_state
BATTERY
J4
+5
Q7 2SD667
100K
R15
+12
R45 R42 1M 499K
PTC 500mA
F4
6
5
J6 battery switch
6 5 4
power_state
D10 1N5339B
D11 HZ11B1
LM358
U15B
0.1uF
C33
7
anode base cathode collector n.c. emitter
U8
6
C4 0.1uF
SN74AHCT00
water_state
1 2 3
4
+5
U5B
C3 150uF
5
+5A
C10 150uF
To DC Powerboard
sen_airflow
+12
+12
DB9
J13
J12
To Sensorboard
1 6 2 7 3 8 4 9 5
10PIN C9 C11 0.01uF 0.01uF
R1IN bat_voltage T1OUT ctr_relay
C8 0.01uF
R41 sen_airflow
2PIN
J5
For LCD power
2 1
C6 0.01uF 1 2 VCC 1M 3 motor signal 4 5 6 water 7 oxygen valve 8 9 10
VCC
VLCD+
26PIN
J11
To Mainboard
1 2 3 4 5 6 7 8 ctr_bell 9 10 ctr_bat ctr_ssr 11 12 ctr_hum ctr_o2 13 14 15 motor_pulse 16 water_state 17 18 fan_state heater_state 19 humheater_state 20 power_state 21 tank_state 22 23 VCC 24 25 +5 26 +5A
WENB RENB TOTXD2 TORXD2 T1OUT R1IN
SERVICE MANUAL FOR YP-2000 INCUBATOR
ELEMENTS DIAGRAM FOR POWER BOARD
7-12
I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
MAIN BOARD ASSEMBLY DIAGRAM 1
7-13
SERVICE MANUAL FOR YP-2000 INCUBATOR
MAIN BOARD ASSEMBLY DIAGRAM 2
7-14 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
MAIN BOARD PARTS LIST No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
1
R2~R5,R7,R8,R21,R22, R24~R29
Resistor 0805-1KΩ±5%
14
2
R1,R6,R9~R13,R18
Resistor 0805-10KΩ±5%
8
3
R19,R30~R32
Resistor 0805-10KΩ±5%
4
4
R14
Resistor 0805-5.6KΩ±5%
1
5
R20,R23
Resistor 0805-200Ω±5%
2
6
R15,R16
Resistor 0805-180Ω±5%
2
7
R17
Potentiometer 3296-500Ω
1
8
C21,C36~C38,C40~C42
Capacitor 0805-0.01uF±20%
7
9
C1~C6,C10,C12~C15
Capacitor 0805-0.1uF±20%
11
10
C17~C20,C22~C27
Capacitor 0805-0.1uF±20%
10
11
C32,C33
Capacitor SMD-25V-150uF±20%
2
12
C16
Capacitor 0805-1500PF±20%
1
13
C7,C8
Capacitor 0805-15PF±20%
2
14
C34,C35
Capacitor 0805-22PF±20%
2
15
C28~C31
Capacitor CA45A-25V-1uF±20%
4
16
C9
Capacitor CA45B-16V-10uF±20%
1
17
C11
Capacitor CD11-25V-10uF±20%
1
18
C39
Capacitor CD11-16V-470uF±20%
1
19
U22
Integrated circuit TPS3824-50
1
20
U18,U19
Integrated circuit SN74LVC1G3157DBVR
2
21
U12
Integrated circuit SN74AHCT00D
1
22
U4
Integrated circuit SN74AHCT04D
1
23
U20
Integrated circuit AT89C51ED2-U
1
24
U7
Integrated circuit DS1230Y-100
1
25
U8,U9
Integrated circuit SN74AHCT574DW
2
26
U3
Integrated circuit SN74AHCT273DW
1
27
U17
Integrated circuit MAX232CSE
1
7-15
SERVICE MANUAL FOR YP-2000 INCUBATOR
MAIN BOARD PARTS LIST (Continue) No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
28
U10
Integrated circuit MC34063AP1
1
29
U23
Integrated circuit DS12C887
1
30
U13
Integrated circuit RA8835P3N
1
31
U14
Integrated circuit CY62256LL-70SNC
1
32
U21
Integrated circuit SN74AHCT373DW
1
33
U11,U16
Integrated circuit SN74LS244
2
34
U5,U6
Integrated circuit SN74AHCT139D
2
35
U1,U2
Integrated circuit SN74AHCT32D
2
36
U26,U27
Integrated circuit K1010
2
37
JZ1
Crystal oscillator HC-49/S-SMD-11.0592MHz
1
38
JZ2
Crystal oscillator HC-49/S-SMD-10.000MHz
1
39
L1
Winding inductance 100uH
1
40
L2~L7
Indsuctance 1210-22uH±20%
6
41
D1
Diode 1N5817
1
42
H1
LED-φ5-G/Y
1
43
H2
LED-φ5-G
1
44
H3
LED-φ5-Y
1
45
H4~H5
LED-φ5-R
2
46
TDK1
Inverter CXA-L10A
1
47
J3
Single row DPCZ-60T
6
48
J2
Socket XY4-2S
1
49
J1
Socket DC-2-26S(WS)
1
50
UZ1(U20)
integrated circuit socket
PLC-44
1
51
UZ2(U23)
integrated circuit socket
IC-24J
1
52
UZ3(U7)
integrated circuit socket
IC-28J
1
53
JP1
Socket 5597-12APB
1
54
J4
Socket SM03-BHS-1-TB
1
7-16 I5
C12 0.1uF
VCC
D7 D6 D5 D4 D3 D2 D1 D0 A8 CS8 /RD /WR
+5
C8
C13 0.1uF
D7 D6 D5 D4 D3 D2 D1 D0 A8 CS8 /RD /WR
ELEMENTS DIAGRAM 1 FOR MAIN BOARD
D2 D3 D4 D5 D6
XD CS A0 VDD D0 D1
+20
R14 5.6K
R17 500
1 2 3 4 5
D2 D3 D4 D5 D6
C14 0.1uF
55 56 57 58 59 60
JZ2 XTAL
CS8 A8 +5 D0 D1
0.1uF 15pF
C15
15pF
1
10K 10uF
/VCE 2
1
180
R16
VCC
5 7 2
180
R15
1500pF
C16
MC34063AP1(8)
8 1 3
LCD Driver
DRVC SWC CT
100uH
L1
CMPR IPK SWE
U10
JP1 5597-12APB(MOLEX )
VA8 VA9 VA10 VA11 VA12 VA13 VA14 VA15 VD0 VD1 VD2
35 34 33 32 31 30 29 28 27 26 25 24
1N5817
D1
SED1335F
NC VA14 VA15 VD0 VD1 VD2
VA8 VA9 VA10 VA11 VA12 VA13
U13
SN74AHCT00
U12A
VA5 VA14 VA7 VA13 VA12 VA11 VA10 VA9 VA1 VA2 VA4 VA3 VA8 VA0 VA6
VA5 VA14 VA7 VA13 VA12 VA11 VA10 VA9 VA1 VA2 VA4 VA3 VA8 VA0 VA6
VA15
C11 10uF
3
+20V +20
10 9 8 7 6 5 4 3 25 24 21 23 2 26 1
10 9 8 7 6 5 4 3 25 24 21 23 2 26 1
5
4
HM62256
A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14
U15
HM62256
A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14
U14
WE OE CE
GND
D0 D1 D2 D3 D4 D5 D6 D7 VCC
WE OE CE
GND
D0 D1 D2 D3 D4 D5 D6 D7 VCC
SN74AHCT00
U12B 6
27 22 20
14
11 12 13 15 16 17 18 19 28
27 22 20
14
11 12 13 15 16 17 18 19 28
/VWR RAM3CS
VCC
VD0 VD1 VD2 VD3 VD4 VD5 VD6 VD7
RAM2CS
/VWR
VCC
VD0 VD1 VD2 VD3 VD4 VD5 VD6 VD7
RAM3CS
10 11
R6
/WR /RD
D7 XD3 XD2 XD1 XD0
C7
+5
/RES CP GND LP
2
/VCE /VWR VA0 VA1 VA2 VA3 VA4 VA5 VA6 VA7
GND VCC XD3 XD2 XD1 XD0 DOFF
8 12
54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 38 37 36
XG SEL1 SEL2 WR RD NC NC RES VRD VCE VWR VA0 VA1 VA2 VA3 VA4 VA5 VA6 VA7 DOFF FLM
U12D SN74AHCT00
U12C SN74AHCT00
3
470uF
C39
J2 2PIN
From powerboard
TDK1 CXA-L10A
J4 SMO3-BHS-1-TB(JST)
OUT1
1 2 VCC 1 2 1
D7 XD3 XD2 XD1 XD0 XECL XSCL VSS LP WF YDIS YD YSCL VD7 VD6 VD5 VD4 VD3 VD7 VD6 VD5 VD4 VD3
6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 CP LP FLM
7-17 12 11 10 9 8 7 6 5 4 3 2 1
OUT2
4 OGND
5 GND 2
C9
SERVICE MANUAL FOR YP-2000 INCUBATOR
9
13
+5
2
10K
R1
RST
WDI
U22 TPS3824-50
VDD
RST
wdi
A14 A15
GND
4
C21 0.01uF 5
+5
+5
+5
3 1 4 6
rst
/rst
2
5
15 14
43 42 41 40 39 38 37 36
/RD /WR
rst
19 18
10
21 20
17 motor_pulse 16 +5 35
RTC_INT
D0 D1 D2 D3 D4 D5 D6 D7
A10 A11
A12 A13
1
2 3
15
14 13 Y0 Y1 Y2 Y3
Y0 Y1 Y2 Y3
12 11 10 9
2
5
4 5 6 7
P1.0 P1.1 P1.2 P1.3 P1.4 P1.5 P1.6 P1.7
P3.0/RXD P3.1/TXD ALE/P PSEN
P2.0/A8 P2.1/A9 P2.2/A10 P2.3/A11 P2.4/A12 P2.5/A13 P2.6/A14 P2.7/A15
+5
11 13 33 32
24 25 26 27 28 29 30 31
2 3 4 5 6 7 8 9 A8 A9 A10 A11 A12 A13 A14 A15
R11 10K
+5
RXD TXD ALE
wdi
S0
C24 0.1uF
VCC
CS5 CS6 CS7 CS8
CS1 CS2 CS3 CS4
SN74AHCT32
5
SN74LVC1G3157
B1VCC B2 A S GND
U19
0.1uF
C23
3 1 4 6
SN74LS139A
G
A B
U6A
SN74LS139A
G
A B
U5B
AT89C51ED2-P44
P3.7/RD P3.6/WR
RST
XTAL1 XTAL2
EA/VP
P3.5/T1 P3.4/T0
P3.3/INT1 P3.2/INT0
P0.0/AD0 P0.1/AD1 P0.2/AD2 P0.3/AD3 P0.4/AD4 P0.5/AD5 P0.6/AD6 P0.7/AD7
U20
RXD1 RXD2 C22 RXD 0.1uFS0
VCC
SN74LVC1G3157
B1VCC B2 A S GND
U18
0.1uF
C5
JZ1 11.0592M
+5
R10 10K
1
22pF
C35
Y0 Y1 Y2 Y3
4 5 6 7
SN74AHCT32
CS7 /RD
2
/RD
13
U1D 12
CS6
/WR
CS5
SN74AHCT32
5
U2B 4
G1
CLK3
/WR
CS3
SN74AHCT32
3
8
CLK1
U2A 1
0.1uF
C4
U1C 9
10
3
U1B 4
SN74AHCT32
/WR
CS2
/WR
SN74AHCT32
2
CS1
SN74LS139A
G
A B
U5A
R9 10K
2
22pF
C34
3
1
+5
TXD1 TXD2 TXD S0
+5
1
2 3
+5
0.1uF
C3
0.1uF
C1
U1A 1
6
11
6
D0 D1 D2 D3 D4 D5 D6 D7
ALE
D0 D1 D2 D3 D4 D5 D6 D7
A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14
G2
CLK4
CLK2
4 5 6 7 8 9 10 11 12
1
1 11
3 4 7 8 13 14 17 18
10 9 8 7 6 5 4 3 25 24 21 23 2 26 1 WE OE CE
Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7
0.1uF
C25
0.1uF
R/W AS CS
IRQ RST DS
DS12C887
AD0 AD1 AD2 AD3 AD4 AD5 AD6 AD7 GND
MOT VCC SQW
U23
SN74AHC373 C26
OE LE
D0 D1 D2 D3 D4 D5 D6 D7
U21
15 14 13
19 18 17
24 23
2 5 6 9 12 15 16 19
VCC
A0 A1 A2 A3 A4 A5 A6 A7
/WR /RD A15
D0 D1 D2 D3 D4 D5 D6 D7
/WR ALE CS4
RTC_INT /rst /RD
VCC
VCC
27 22 20
11 12 13 15 16 17 18 19 28
VCC D0 D1 D2 D3 D4 D5 D6 D7 VCC
0.1uF
DS1230Y-100
A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14
U7
C17
VCC
R12 10K
D4 D5 D6 D7
D0 D1 D2 D3
D0 D1 D2 D3
D0 D1 D2 D3 D4 D5 D6 D7
CLK3
D0 D1 D2 D3 D4 D5 D6 D7
CLK2
CLK1 /rst
D0 D1 D2 D3 D4 D5 D6 D7
9 7 5 3
18 16 14 12
18 16 14 12
2 3 4 5 6 7 8 9
1 11
2 3 4 5 6 7 8 9
1 11
11 1
3 4 7 8 13 14 17 18 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8
2 5 6 9 12 15 16 19
0.1uF
1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q
VCC
0.1uF
G A1 A2 A3 A4
Y1 Y2 Y3 Y4
SN74LS244
A1 A2 A3 A4
G
SN74LS244 U16B
Y1 Y2 Y3 Y4
U16A
G A1 A2 A3 A4
SN74LS244
Y1 Y2 Y3 Y4
U11A
11 13 15 17
19
2 4 6 8
1
2 4 6 8
1
1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q
VCC
SN74HCT574DW
1D 2D 3D 4D 5D 6D 7D 8D
OC CLK
U9
C19
SN74HCT574DW
1D 2D 3D 4D 5D 6D 7D 8D
OC CLK
U8
C18
SN74AHC273DW
CLK CLR
D1 D2 D3 D4 D5 D6 D7 D8
U3
key_out1 key_out2 key_out3
led_batchg led_batfull led_set led_37
+5
key_in1 key_in2 key_in3
G2 0.1uF
C27 VCC
humheater_state tank_state
G2
heater_state fan_state power_state water_state
G1
19 18 17 16 15 14 13 12
19 18 17 16 15 14 13 12
0.1uF
C10
led_pwr
ctr_hum ctr_o2
ctr_ssr ctr_bat
2
R22 K1010 1K
VCC
C36
KC1
R13 10K
KR1
19
C38
KC3
R19 10K
KR3
KC6
KC5
KC4
KC3
KC2
KC1
C40
KC4
R30 10K
KR4
1K R26 1K R27 1K R28 1K R29 1K
1K R25
R24
C6
C41
KC5
R31 10K
KR5
KR6
KR5
KR4
KR3
KR2
KR1
C42
KC6
R32 10K
KR6
SN74LS244
A1 A2 A3 A4
G Y1 Y2 Y3 Y4
C28 1uF
0.1uF U11B
+5A
R23 200
R21 1K
+5A
+5
TXD2 11 TORXD2 13 15 17
0.1uF led_alm
C2
ctr_bell
C37
KC2
R18 10K
KR2
22uH
22uH L7
22uH L6
22uH L5
22uH key_out3 L4
key_out1 L2 22uH key_out2 L3
VCC
RXD1
U27
K1010
R20 200 U26
VCC
SN74AHC04
4
SN74AHC04+5 U4B
U4A
TXD1
3
1
1K
R4
1 2 3 4 5 6
1K
R5
1K
R8
1K
R7
1K
R3
MAX232
C1+ C1T1IN T2IN R1OUT R2OUT
Keyboard
TOTXD2 RXD2 WENB RENB
1 3 11 10 12 9
C29 1uF U17 2 VS+
6PIN
J3
9 7 5 3
+5A
led_37
led_pwr
led_batfull
led_batchg R2 1K
led_set
0.1uF
C20
LED
LED H3
LED H5
VS-
6
C30 1uF WENB C31 RENB 1uF TOTXD2 TORXD2 T1OUT R1IN
D7 D6 D5 D4 D3 D2 D1 D0 A8 CS8 /RD /WR
D7 D6 D5 D4 D3 D2 D1 D0 A8 CS8 /RD /WR
26PIN
J1
From Powerboard
1 2 3 4 C2+ 4 5 C25 14 T1OUT 6 7 T2OUT 7 13 R1IN 8 8 R2IN ctr_bell 9 10 ctr_bat ctr_ssr 11 12 ctr_hum 13 ctr_o2 14 15 motor_pulse 16 water_state 17 18 fan_state heater_state 19 humheater_state 20 power_state 21 tank_state 22 +5 VCC 23 VCC 24 25 C32 C33 +5 26 150uF 150uF +5A
LED2 H4
LED H1
H2
SERVICE MANUAL FOR YP-2000 INCUBATOR
ELEMENTS DIAGRAM 2 FOR MAIN BOARD
7-18
I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD ASSEMBLY DIAGRAM 1
7-19
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD ASSEMBLY DIAGRAM 2
7-20 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD PARTS LIST No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
1
R3,R4,R6
Resistor 0805-1KΩ±1%
3
2
R1,R5,R41,R48,R43,R44
Resistor 0805-1KΩ±5%
6
3
R12
Resistor 0805-9.1KΩ±1%
1
4
R2,R15~R17,R20, R21,R24
Resistor 0805-10KΩ±5%
7
5
R25,R36~R40
Resistor 0805-10KΩ±5%
6
6
R14
Resistor 0805-18KΩ±1%
1
7
R13
Resistor 0805-33kΩ±1%
1
8
R7,R42,R45
Resistor 0805-200Ω±5%
3
9
R27,R32~R35
Resistor 0805-499KΩ±5‰
5
10
R50,R51
Resistor 0805-100Ω±1%
2
11
R8~R11,R18,R19
Resistor 0805-20KΩ±1‰
6
12
R22,R23,R26,R46
Resistor 0805-20KΩ±1‰
4
13
R28~R31
Resistor 0805-13KΩ±5‰
4
14
R47
Resistor 0805-30KΩ±1‰
1
15
R49
Resistor 0805-510Ω±5%
1
16
VR1
Potentiometer 3296-100Ω
1
17
C16,C17
Capacitor 0805-10PF±20%
2
18
C1~C4,C5~C12,C40,C43
Capacitor 0805-0.01uF±20%
14
19
C18~C27,C32~C34
Capacitor 0805-0.1uF±20%
13
20
C44,C45,C47~C49
Capacitor 0805-0.1uF±20%
5
21
C35
Capacitor 1206-1uF±20%
1
22
C28~C31
Capacitor 0805-0.1uF±20%
5
7-21
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD PARTS LIST (Continue) No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
23
C13~C15,C36
Capacitor CA45B-16V-10uF±20%
4
24
C41,C42
Capacitor 0805-22PF±20%
2
25
C37~C39,C46
Capacitor SMD-25V-150uF±20%
4
26
U10
Integrated circuit SN74AHCT74D
1
27
U11,U15
Integrated circuit TLC277CD
2
28
U4
Integrated circuit TLV2548CDW
1
29
U5
Integrated circuit MSC1210Y5
1
30
U6
Integrated circuit MAX232CSE
1
31
U1,U16
Integrated circuit LM358D
2
32
U8,U13
Integrated circuit MC7805BDTG
2
33
U2,U3
Integrated circuit TPS3824-50
2
34
U17
Integrated circuit TLV2544CD
1
35
U9
Integrated circuit AT89C4051
1
36
U12,U14
Integrated circuit K1010
2
37
U17
Integrated circuit SN74AHC08D
1
38
Q1~Q3
Triode MMBT3904LT1
3
39
JZ1,JZ2
Crystal oscillator HC-49/S-SMD-11.0592MHz
2
40
LS1
Buzzer TMB12A05
1
41
S1
Plucking switch A6S-4101
1
42
H1
LED RL-803ARC30-0
1
43
F1,F2
Fuse 1812L050-C
2
44
J1
Strip needle with round hole RCD-40P
1
45
J8
Socket ATY1.25-6S
1
7-22 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD PARTS LIST (Continue) No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
46
J3
Socket ATY1.25-2S
1
47
J6
Socket XY2.5-4S
1
48
J4,J5,J7
Socket ATY1.25-4S
3
49
J2
Socket XY2.5-10S
1
50
JP1
Single row header DPCZ-60T
1
7-23
SERVICE MANUAL FOR YP-2000 INCUBATOR
THE PREPOSITIVE SENSOR BOARD ASSEMBLY DIAGRAM THE PREPOSITIVE SENSOR BOARD PARTS LIST No.
SYMBOL
NAME AND SPECIFICATION
QUANTITY
1
J2,J3
Thermistor 14A3002-C3
2
2
J4
Humidity sensor HIH-4000-003
1
3
J5
Socket FD-10P
1
4
R1
Resistor RJ-0.25-68Ω±1%
1
J2
+5B
+5C
sen_duli smallfan
+5C sen_hum
+5B
J3
1 2
1 2
3 2 1
1 2 +12
J4 smallfan
J1
+5B
+12 1 2 3 4 5 6 7 8 9 10 J5
ELEMENTS DIAGRAM FOR THE PREPOSITIVE SENSOR BOARD 7-24 I5
SERVICE MANUAL FOR YP-2000 INCUBATOR
1. Oxygen sensor
2. Connecting wire for oxygen sensor
3. Connecting wire for Baby scale sendor
SENSOR WIRE-CONNECTING DIAGRAM
7-25
10PIN
4PIN
4PIN
SDO2 SDI2 SCLK2 ADC2INT
0.01uF
16 15 14 13 12 11 10 9 R23 20K
+5C
C15 10uF
C11 0.01uF
C9 0.01uF
2
+5C
C33 0.1uF 5
K1010 U14
22pF
C42
22pF
3
JZ2 XTAL
RST
U3 TPS3824-50 C41
VDD
RST
GND
1
R44 K1010 1K
+5B
RXD2
1
+5C
R22 20K
CSTART2
ADC2CS
ch_O2_2
0.1uF
2
TLV2544
SDO /CS SDI REFP SCLK REFM EOC/(-INT) FS VCC /PWDN A0 GND A1 /CSTAR A2 A3
0.1uF
C32
TLC277
499K C7
R35
1
C45
TXD2
R42 200 U12
+5B
+5B
+5B
WDI
ch_O2_2
1 2 3 4 5 6 7 8
0.01uF
R34 499K C6
ch_O2_1
0.1uF
C12 0.01uF
4
U7
2
1
TLC277
+5C C43 0.01uF U15A 3
2
+5B C40 0.01uF U11A 3
C5 0.01uF
C10 R18 R19 0.01uF 20K 20K
+5B
C14 10uF
+5C
1 T2IN 2 R2OUT 3 4 5 ADC2INT 6 7 8 9 10
10K
R16
10K
R20
clear
RST
C22 0.1uF
RXD1 TXD1 ADC1INT
JZ1 XTAL
20 19 18 17 16 15 14 13 12 11
R21 10K
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
CSTART2 ADC2CS SDO2 SDI2 SCLK2
+5C
R15 +5B 10K
+5B
VCC P1.7 P1.6 P1.5 P1.4 P1.3 P1.2 P1.1 P1.0 P3.7
0.1uF
C34
AT89C4051
RST RXD TXD XTAL2 XTAL1 /INT0 /INT1 T0 T1 GND
U9
+5C
R45 200
R2OUT
T2IN
3
+5B
10pF
1
R43 1K
+5C
U2 TPS3824-50
VDD
RST
GND
C17
10pF
10K
R25
10K
R24
R47 30K
R46 20K
+5C
XOUT XIN P3.0/RxD0 P3.1/TxD0 P3.2/INT0 P3.3/INT1/TONE/PWM P3.4/T0 P3.5/T1 P3.6/WR P3.7/RD DVdd DGND RST DVdd DVdd NC U5 MSC1210
clear
+5C
+5C +5C
4 3 2 1
ch_isolated
J8 scaler
6
5
2
3
Q Q
LM358
U16B
+5C
LM358
U16A
+5B
0.1uF
SN74LS74A C27
PRE CLK D CLR
U10A
1uF +5B
R3 R4 1K 1K C35
5 6
7
1
0.1uF
C48
over_alarm
C36 C25 0.1uF 10uF
48 47 46 45 44 43 42 41 40 39 38 37 36 35 34 33
2
315mA
+5B
3
R49 510
+5C
0.1uF
C49
SN74AHC08D
U17A
1K
R48
315mA
+5B
R17 10K
+5B
2PIN
J3
4 3 2 1
+5B
2 1
1 C46 150uF
1 C37 150uF
RXD1
IN
U8
relay_ctr
MAX232
C1+ C1T1IN T2IN R1OUT R2OUT
0.1uF
C24
U13 MC7805BDTG 3 IN OUT GND
MC7805BDTG 3 OUT GND
1 3 11 10 12 9
C29 U6 1uF 2 VS+
2PIN PSEL
JP1
LS1 TMB12A05 Q3 8080
C47 0.1uF
C26 0.1uF
Q1 8050
over_alarm R41 1K
TXD1
C28 1uF
C23 0.1uF
PSEN
2 1
A6S-4101
S1
coolfan_st1 coolfan_st2
+5B
F2
5 6 7 8
R36 10K
+5B
F1
1K
R5
1
+5C
+12
+12
EA P0.6/AD6 P0.7/AD7 ALE PSEN/OSCCLK/MDOCLK P2.7/A15 DVdd DGND P2.6/A14 P2.5/A13 P2.4/A12 P2.3/A11 P2.2/A10 P2.1/A09 P2.0/A08 NC
SW_BOX
LAMP_ALM
SW_HUM SW_O2 SW_O2CTR SW_WEIGHT
4
+5C
13K
R31
R33 499K
13K
R29
R32 499K
R26 20K
CSTAR1 ADC_CHECK ch_O2_1 ch_airflow
ADC1CS
C20 0.1uF 5
2
R2 10K
+5B
C16
6 5 4 3 2 1
13K
R30
13K
R28
C44
TLV2548
20 19 18 17 16 15 14 13 12 11
coolfan_st2
0.1uF
C19
coolfan_st1
0.1uF
C18
4
+5C ch_isolated
O2 Sensors
4 3 2 1
100 C8 0.01uF
+5B ch_air ch_skin1 ch_skin2 bat_voltage ch_hum
SDO CS SDI REFP SCLK REFM EOC/(INT) FS VCC PWDN A0 GND A1 CSTAR A2 A7 A3 A6 A4 A5
7
1
+5B
NC 32 REF OUT 31 REF IN+ 30 REF IN29 AVdd 28 AGND 27 AINCOM 26 AIN7/EXTA 25 AIN6/EXTD 24 AIN5 23 AIN4 22 AIN3 21 AIN2 20 AIN1 19 AIN0 18 AGND 17
J6
Skin 2 sensor
R10 20K
R51
1 2 3 4 5 6 7 8 9 10
0.1uF
+5B
LM2903
U1B
LM2903
U1A
WDI
4PIN
+5B
SDO1 SDI1 SCLK1 ADC1INT
U4
+5B
C21
6
5
2
3
4
4 3 2 1
100 C3 0.01uF
ch_skin1
+5B
R13 R14 33K 18K
+5B
R12 R6 9.1K 1K
VR1 100
P0.5/AD5 P0.4/AD4 P0.3/AD3 P0.2/AD2 P0.1/AD1 P0.0/AD0 P1.0/T2 P1.1/T2EX DGND DVdd P1.2/RxD1 P1.3/TxD1 P1.4/INT2/SS P1.5/INT3/MOSI P1.6/INT4/MISO P1.7/INT5/SCLK
J5
R9 20K
R50
C4 0.01uF
C13 10uF
2
2
Skin 1 sensor
+5B
C2 R11 0.01uF 20K
1K
ch_hum
1
4 3 2 1
R8 20K
ch_isolated smallfan ch_air
R27 499K C1 0.01uF R1
+5B
49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64
J4
+5C +5B
+12
3
To Prepositive sensor board
10 9 8 7 6 5 4 3 2 1 1 ADC1CS SCLK1 SDI1 SDO1 TXD2 RXD2
J1
CSTAR1
+5B R37 R38 R39 R40 10K 10K 10K 10K
+5C
+5B
C38 150uF
C39 150uF
C2+ C2T1OUT T2OUT R1IN R2IN
VS-
Q2 8050
LED
H1
4 5 14 7 13 8
6
+5B
4PIN
J7
C31 1uF
To powerboard of controller
10PIN
J2
serial port
1 2 3 4
1 2 3 +12 4 ch_airflow 5 bat_voltage 6 relay_ctr 7 8 9 10 R1IN T1OUT
T1OUT
R1IN
C30 1uF
TXD1 RXD1
+5B
200
R7
SERVICE MANUAL FOR YP-2000 INCUBATOR
SENSOR BOARD PRINCIPLE DIAGRAM
7-26
I5