Service Manual DIG-610 SM01-02-01 SERIVEC MANUAL MODEL: DIG-610 (PORTABLE X-RAY UNIT) * ALL THE INFORMATION PRESCR
Views 152 Downloads 9 File size 537KB
Service Manual
DIG-610
SM01-02-01
SERIVEC MANUAL MODEL:
DIG-610
(PORTABLE X-RAY UNIT)
* ALL THE INFORMATION PRESCRIBED INSIDE THE MANUAL IS PRESENTED ONLY AS A GUIDE FOR THE OPERATION OF OUR PRODUCTS. AND IT MAY BE CHANGED WITHOUT PRIOR NOTICE. * DONGMUN CO., LTD.: 717-HO, UNITECHVIL B/D, 1141-2, BAEGSUK-DONG, ILSAN-GU, KOYANG CITY, SOUTH KOREA TEL: 82-31-907-7222, FAX: 82-31-904-2231,E-MAIL: [email protected], URL: dong-mun.co.kr
Service Manual
DIG-610
CONTENTS
1
Notice
-----------------------------------------------------------------------
27
2
Operation Sheet
-----------------------------------------------------------------------
27
3
Block Diagram
-----------------------------------------------------------------------
28
4
Calibration Process
-----------------------------------------------------------------------
29
5
Calibration Code and Data
-----------------------------------------------------------------------
33
6
Error code & Countermeasure
-----------------------------------------------------------------------
33
7
Schematics
-----------------------------------------------------------------------
34
8
Maintenance
-----------------------------------------------------------------------
40
TABLES 1
mA Calibration Code
-----------------------------------------------------------------------
33
2
Filament Preheat Calibration
-----------------------------------------------------------------------
33
3
Technical guide
-----------------------------------------------------------------------
41
2
Service Manual
DIG-610
1. NOTICE 1.1. If the output calibration is needed after periodical inspections or repairs, please observe following rules and methods. CAUTION The ionized radiation is dangerous for the operator if the following safety measures are not strictly observed. 1.2. This manual is designed to ensure correct use and operation of DIG-610. Please read all the lines thoroughly before you use this equipment. 1.3. Incorrect use and operation exceeding described conditions in this manual may occur damage of the machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. 1.4. This equipment should be used only by the legally qualified persons and practitioners. 1.5. Regarding the AC cable and software, only the one developed and supplied by Dongmun should be used. 1.6. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which many results from the use of this manual. 1.7. Always keep the manual at hand for your reference.
2. OPERATION SHEET
3
Service Manual
DIG-610
3. BLOCK DIAGRAM
READY
PXP60HF-CTL-001
1 2
X-RAY
CON1
3 35312-03
HAND SWITCH
3V/35mA LASER POINTER
CONTROL BOARD OP PANEL
1 2
CON6
1 2
CON2
* EXPORT : NORMAL * DOMESTIC : NOT USED
LASER
35312-02
& HT CONT.
TANK T 2
35312-02 CON4
35312-05 CON3
THERMAL EP_FB(-)
35312-05
35312-11 CON2
/100V_SEL
LAMP_CONT FIL_FB
EP_CONT2 FIL_CONT1 FIL_CONT2
IP_FB(+)
GND IP(+)
1 2 3 4 5 6 7 8 9 10 11
1 2 3 4 5
+5V DGND +12V AGND -12V
EP_CONT1
AGND
EP(-)
1
1 2 3 4 5
1 2 3 4 5 6 7 8 9 10 11
35312-11
1 2 3 4 5
CON5 35312-05
CON3
L101 HT XFMR
PXP-60HF-IGBT TB2
TB1
TB4
IGBT BOARD
T1
3TURNS
T1 20
35312 CON7
PXP60HF-PWR
35312 CON10 35312 CON8 35312
PWR & DRIVE
CON9
CN2 35312
CN3 35312
CN4 35312
1 2
1 2
35312-03
TB3 CN1 35312
1 2
TB1
TB2 CON5
1 2
1 2 3
COM D-205
1 2 1 2 FG 1 2 1 2
CON4
FILAMENT DRIVE CB1
575
T2
1 2 3 4
T2 C1
3 2
+12.2 SHIELD GND
SHIELD LINE
3 C2
53258-04
FIL XFMR
250V/16A
5 4 3 2 1
250V/20A
53258-05 CON1
COLLIMATOR DRIVECON6 1 2
COLLIMATOR SHIELD LINE
CN1 PLUG AC MALE
LAMP 53258-02
12V/50W
APPROVED BY
DESIGNED BY
DIG-610 Main Block
PXP-60HF-Main Block
CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev A
Sheet 1
4
of
1
Service Manual
DIG-610
-PREPARATION Following tools and instruments are needed. a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver b. Instruments: DVM, oscilloscope (Tektronics TDS 3032 and more than Equivalent) 4. CALIBRATION PROCESS 4.1. Line Power Supply -Line switch should be at off position. -Connect power line cord to the power outlet. -Turn on the line switch. -Then the LED on the operation panel illuminates. 4.2. Calibration procedures are composed of three steps. -Tube voltage confirmation and calibration. -Tube current confirmation and calibration. -Filament preheat calibration A. Calibration and Confirmation of Tube Voltage -Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP board appears. -All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually located in this location.) -Connect the channel one probe of the oscilloscope to TP5 (IP F/B). -Connect the channel another one probe of the oscilloscope to TP6 (EP F/B). Channel 1 is mA, and 60mA = 9 V Channel 2 is kV, and 40kV = 2V -Connect the ground to TP11 (GND). -Turn on the line switch and the display indicates as follows.
kV
mAs
-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the oscilloscope channel 1. -If it is not about 9V, adjust mA by VR3 on the OP board. Then observe the kV waveform of the oscilloscope channel 2. -If it is not 2V, adjust kV by VR2 on the OP board. After adjustment, set the tube voltage kV to 80KV by the kV up/down switch. -Set the kV and the mAs at 80kV and 0.4mAs(40mA).Then observe the mAs, kV waveform of the oscilloscope channel 1, 2.
5
Service Manual
DIG-610
-If it is not about 6V, adjust mA by VR3 on the OP board. -If it is not about 4V, adjust kV by VR2 on the OP board. B. Calibration and Confirmation of Tube Current -Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP board appears. -All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually in this position.) -Connect the channel one probe of the oscilloscope to TP5 (IP F/B). -Connect the channel another one probe of the oscilloscope to TP6 (EP F/B). Channel 1 is mA, and 60 mA = 9V Channel 2 is kV, and 40 kV = 2V -Connect the ground to TP11 (GND). -Turn on the line switch and the display indicates as follows.
mAs
kV
-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the oscilloscope channel 1. -If it is not 7V, adjust mA by the calibration data on the OP board. -Turn off the line switch, and wait until the display switch the light off. -DIP switch 1 on the PCB OP board is turned on. -Turn on the line switch and the display indicates as follows. -Then push the mAs up switch (see operation sheet). The display indicates as follows.
kV Calibration code
mAs Calibration data
C00 is calibration code of tube current (60mA at 40kV-50kV). 179 is calibration data of tube current (60mA at 40kV-50kV). -Adjust mA by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If it is not 9V, set the calibration data again. -DIP switch 1 on the PCB OP board is turned on. -Then push the kV up switch (see operation sheet).
6
Service Manual
DIG-610
The display indicates as follows.
mAs
kV Calibration code
Calibration data
C01 is calibration code of tube current (40mA at 51kV-70kV). 153 is calibration data of tube current (40mA at 51kV-70kV). -Adjust mA by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If it is not 6V, set the calibration data again. -Table .1 shows the calibration code and the calibration data against an each tube current. C. Filament preheat calibration -This is preheating filament for making tube current. -The overshoot or the undershoot may occur from the tube current if the preheat calibration data or the VR3(preheat gain volume) is set by mistake. -There are two methods to make preheat calibration setting. The calibration data is possible to be changed in accordance with the tube condition. -The VR3(preheat gain volume) adjusts the whole range of tube current. Filament preheat calibration data adjusts each tube current. -Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP Board appears. -DIP switch 1 on the PCB OP board is turned on. -Turn on the line switch. -Push APR switch ① and then the display indicates as follows.
mAs
kV Preheat calibration code
Preheat calibration data
P00 is Filament preheat calibration code (Preheat calibration code of 60mA). 17 is Filament preheat calibration data (Preheat calibration data of 60mA). -Adjust data by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. If overshoot or undershoot occur in the waveform, set the calibration data again. Then push the kV up switch (see operation sheet). The display indicates as follows.
7
Service Manual
DIG-610
kV Preheat calibration code
mAs Preheat calibration data
P01 is Filament preheat calibration code (Preheat calibration code of 40mA). 37 is Filament preheat calibration data (Preheat calibration data of 40mA). -Adjust data by the mAs up/down switch ⑥ (on the control panel). -After it is adjusted, push the Store switch ⑨ (on the control panel). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If overshoot or undershoot occur in the waveform, set the calibration data again. -Table. 2 shows the calibration code and the calibration data against an each tube current. The calibration data is possible to be changed in accordance with the tube condition.
8
Service Manual
DIG-610
5. CALIBRATION CODE AND DATA [Table 1. mA Calibration Code and Data] AC voltage
Code
Data
Tube voltage
Tube current
mAs
220V 220V 220V 220V 220V 220V 220V 220V 220V
C00 C01 C02 C03 C04 C05 C06 C07 C08
235 156 155 155 136 120 115 115 97
40kV - 50kV 51kV - 70kV 51kV - 70kV 71kV - 80kV 71kV - 80kV 81kV - 90kV 81kV - 90kV 91kV - 100kV 91kV - 100kV
60mA 40mA 40mA 40mA 35mA 30mA 30mA 30mA 25mA
0.4 - 100 0.4 - 64 80 - 100 0.4 - 32 40 -80 0.4 - 40 50 - 80 0.4 - 50 64 - 80
[Table 2. Filament Preheat Calibration Code and Data] AC voltage
Code
Data
Tube current
220V 220V 220V 220V 220V
P00 P01 P02 P03 P04
7 60 75 85 95
60mA 40mA 35mA 30mA 25mA
6. ERROR CODE & COUNTERMEASURE [Table 3. Code & Measures] Error Code Err. 07
Err. 08 Err. 10 Err. 11 Err. 12 Err. 14 Err. 15 Err. 16 Err. 17 Err. 19 Err. 21
Content
Action
When the line switch is turned on, this is inspected. The ready contactor of the hand switch is closed even it is not pushed. When the line switch is turned on, this is inspected. The x-ray contactor of the hand switch is closed even it is not pushed. It is displayed when the FIL FB signal is not detected in the state of standby. It is displayed when the EP FB signal is not detected in the state of standby. It is displayed when the IP FB signal is not detected in the state of standby. It is displayed when the FIL FB signal is not detected in the state of ready. It is displayed when the EP FB signal is low in the state of exposure. It is displayed when the IP FB signal is low in the state of exposure. It is displayed when the IP FB and EP FB signal is high in the state of exposure. It is displayed when the mAs is over in the state of exposure. It is displayed when the connector CON6 on the main board is not insert.
9
Change the hand switch
Change the hand switch
Replace operation board or Replace HV tank Replace operation board or Replace HV tank Replace operation board or Replace HV tank Replace main board or Replace power board Replace operation board Replace main board or Replace power board Replace main board or replace power board Check the IP FB level and adjust it Replace operation board Check CON3
Service Manual
DIG-610
7. SCHEMATICS
TD62304
+5V
SEG08_OUT1
(1M-6)
LED4 WAIT 2 1 RED
LED5 M1 2
LED10 M6 2
LED11 M7 2
LED12 M8 2
LED13 ERROR 2 1 RED
1 GREEN
1 GREEN
LED6 M2 2
uD_SR105K
1 GREEN
LED7 M3 2
SEG5
5
9
1
2
4
6
10
DP
G
F
E
D
C
B
7 3 8
SEG4
CC CC
A
10
9
1
2
4
5 DP
G
F
E
D
7
6
C
B
CC CC
A
10
9
1
2
4
6
7
5 DP
G
F
E
D
C
B
CC CC uD_SR105K
LED3 X-RAY 2 1 YELLOW
1 GREEN
A
5
10
9
1
2
4
6
SEG3
LED2 READY 2 1 GREEN
1 GREEN
DP
G
F
E
D
C
B
7
uD_SR105K
3 8
SEG2
3 8
uD_SR105K
3 8
SEG1
CC CC
A
10
9
1
2
4
6
7
5 DP
G
F
E
D
C
B
A
A
2
5
R28 15 1/4W 1
10
2
9
R27 15 1/4W 1
1
2
F
R26 15 1/4W 1
2
2
E
2
R25 15 1/4W 1
4
2
R24 15 1/4W 1
CC CC
16 15 14 13 12 11 10 9
3 8
Q1 Q2 Q3 Q4 Q5 Q6 Q7 VCC
DP
I1 I2 I3 I4 I5 I6 I7 GND
G
1 2 3 4 5 6 7 8
D
(1B-13) (1B-13) (1B-12) (1B-12) (1B-12) (1B-12) (1B-12)
3 8
SEG01_OUT SEG02_OUT SEG03_OUT SEG04_OUT SEG05_OUT SEG06_OUT SEG07_OUT
6
2
R23 15 1/4W 1
CC CC
U8
C
2
R22 15 1/4W 1
7
R21 15 1/4W 1
(2F-5) (2G-5) (2H-5) (2J-5) (2K-5) (2L-5) (2N-5) (2O-5)
B
SEG_A SEG_B SEG_C SEG_D SEG_E SEG_F SEG_G SEG_DP
uD_SR105K
LED8 M4 2
1 GREEN
1 GREEN
SEG6
uD_SR105K
LED9 M5 2
1 GREEN
LED14 REVERSE 2 1 GREEN
APPROVED BY
DESIGNED BY
HTP-24 PORTABLE X-RAY CONTROL BOARD OP PANEL & HT CONT. CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev 0
HTP24-CTL-001 Sheet 3
10
of
7
Service Manual
DIG-610
1
+5V RA7
2 3 4 5
4.7K
CON6 ISO_PWR
(1B-10),(7G-13) (1B-10),(5N-13)
2 3
/FAULT IP_FB_IN
R29 270
ISO_GND 1
2
35312-0310
2
1
D33 1N4148
PC1 8
2
7
3
6
(1B-10)
HS_READY
1
1 CON1 1 2 3 4
(1B-9)
35312-0410
1
2
2
R30 270 D34 1N4148
4
HS_X-RAY
5
1
TLP620-2
+5V D35
D36
CON2 1
2
1 2 1N4148
2
1 1N4148 (1M-7)
/LASER
35312-0210
APPROVED BY
DESIGNED BY
HTP-24 PORTABLE X-RAY CONTROL BOARD OP PANEL & HT CONT. CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev
HTP24-CTL-001
0
Sheet 4
11
of
7
Service Manual
DIG-610
(7B-13)
IP_FB R31
10K 1% (1J-5)
IP_FB_ADC
R32 +12V 2
10K 1%
1
39K
14 1
3 VR1 10K
2
12 4
1
-12V
1SC CPI COM
+VS -VS
FOUT VOUT
5 10 11
BOX332 2
(1B-10),(4J-11)
7 13
IP_FB_IN
VFC32
MN104
1
C22
+IN -IN
+12V
C62
1N4148 1
R33
1
8
D38
+ R35 1/2W 200
C21 1
U10
2
1
2
2
3
+
2
R34 4.7k
LM358 -
1N4148 40mA(0 ~ 8V)
U9A
4
D37
C24 1
6.8uF/16V tantal
BOX103 2
CON3 1
IP_FB(+)
2 3
EP_FB(+)
C60 2
39P 1
4 R75
5
100K 1%
35312-0510 (1A-10)
(6B-14),(7B-11)
/HT_CHECK
8
D39 1N4148
100KV(0 ~ -5V)
(1J-5)
C58 2
R76 1K
EP_FB_ADC
R84 R
U9B MN104 1
1
1
1
1N4148
1K
2
BOX223 BOX223 1
1
R36 7
4
D40 BOX223
LM358
2
C27
+
6
2
2
2
2
C26
5 100K 1%
1
R37 C25
EP_FB
+12V C28 MN104 2 1
2
+12V
-12V
C63
VR2
MN104
2
1
-12V
2K 1
3
R38 10K 1% 1
C29 2
C30 2
TP3 MN104 2
MN101 1
TP4
EP_REF 1
+12V C31 1
2
MN101 1
1 IP_CONT
R42 10M
8
EP_REF
U11A LM358 MN104 2
C33 MN101
U11B
5
+
7
6
-
(6C-5)
IP_REF
LM358 R44 10K 1%
2
2
1
C59 1
2 1
-
R40 10K 1% 1
(1F-7)
4
R43 10K 1%
BOX224
(6B-14)
1
1
1
2
2
+
1
C32 R41 10M
IP_REF
8 3
1
2
4
R39 10K 1% 1
(1F-7)
2
-12V
2
2
EP_CONT
APPROVED BY
DESIGNED BY
HTP-24 PORTABLE X-RAY CONTROL BOARD OP PANEL & HT CONT. CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev 0
HTP24-CTL-001 Sheet 5
12
of
7
Service Manual
DIG-610
+12V +12V R45 5.6K (5J-9),(7B-11) (5F-4)
U12
1
2
1 1 1
EP_FB EP_REF
2 R46 5.6K
1
2
2
1
2
1
2
C36
3 C34 R48 BOX682 470K 2
C1 E1
12 8 9
R47 1
910
R50 1
910
CON4 2
1 2
EP_CONT1 2
3 4
EP_CONT2 1 C37 BOX102 2 R51 6.2K 1%
1
VCC
-IN1 F.B.
R49 680
2
BOX102
+IN1
5
CT
6
C2 E2
11 10 FIL_CONT1 FIL_CONT2
RT
4
5 6
DT
7 15 16
GND -IN2 +IN2
VR OC
7 8
LAMP_CONT
14 13 FIL_FB
9 10
C38
+
11
/100V_SEL
R52 47K
1uF/25V TANTAL 2 2
1
1
35312-1110 2
U13A 14066
1
1
TL494
(6B-7)
2
13
R53 2K
(1M-6)
/HV_ON
VR3 2K
R54 750
R55 15K
2 1
R56 10K 1%
3 1
2
1
(1J-5)
2
FIL_FB_ADC
2
2
1
+5V
R57 10K 1% 1
-12V
1
2
2 2
1 R82 10K
2
1
2 3 2
3 U17 1PN2222
1
1 C42 BOX102 2 R64 5K 1%
R81 1
3.3K
2
1
1
R79 5.6K 1
5 6 4 7 15 16
2 R70 470K 2 2
-IN1
C1 E1
10
8 9
R58 1K
F.B. 2
R63 10K
12
12
2
2
C39 R62 BOX473 470K
VCC
CT
C2 E2
11 10
RT DT GND -IN2 +IN2
1
R61 5.6K
1K
+IN1
1
2
2 1
R80 C41 MN104 2 1
+12V PN2222 DRV1
1 1
1
2
R65 1K VR OC
14 13 2
1
1
C40 BOX104
6
1
3
11
U15
7 1
2 2
R66 1K /LAMP_ON
(1M-6)
1 TL494
(5L-4)
IP_REF
R78 30K 1
IP_FB
C61 BOX104
2
1
1
R77 5.6K
U13C 14066
R67 47K 2
3
1
4
C43 BOX473
2
(6K-11) 1
U13B 14066
+12V R59 5.6K
OP07
2
4 8
U14
+
R68 2K 2
5
FIL_FB
2
-
R60 4.7K
/PREP
APPROVED BY
DESIGNED BY
(1M-6)
HTP-24 PORTABLE X-RAY CONTROL BOARD OP PANEL & HT CONT. CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev 0
HTP24-CTL-001 Sheet 6
13
of
7
Service Manual
+12V
DIG-610
+12V TP5
2
+12V C44 R69
R85 100k
1
8
1
(5I-15)
1
2
+
2
-
C45 BOX473 (1B-10),(4J-11)
2
R86 100k
1 LM393
1
R71 100K
3
4
IP_FB
2 MN104 U16A
1
64K
IP_FB
TP6
/FAULT
1
14066
1
TD62304AP
C50 MN104
C51 MN104 2
C49 MN104
2
2
TD62304AP C48 MN104
2
27512 C47 MN104
1
4
1
74HC573
100K
1
1
LM393 C46 BOX103
R74 R73 + C35 + C23 10u/16V 56K 10u/16V
-
2
6
+5V
1
2
7
2
1 (5J-9),(6B-14) R72 100K
2
2
EP_FB
U16B
+
1
5
8
1
EP_FB
TP7
+5V
1
+5V
H1
H2
H3
H4
H5
H6
H7
H8
H9
H10
H11
H12
1
HOLE
HOLE
2
1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE 1
HOLE
1
DGND
1
C53 100uF/50V
2
+
1
1
TP8 C52 MN104
CON5
TP9
1 2 3 4 5
1
TP11
1
1
TP10
1
2 1
TP12 C57 MN104 -12V 1
2
C56 + 100uF/50V
AGND
AGND 1
2
C55 MN104
1
C54 + 100uF/50V
2
35312-0510
+12V
+12V
-12V
APPROVED BY
DESIGNED BY
HTP-24 PORTABLE X-RAY CONTROL BOARD OP PANEL & HT CONT. CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev
HTP24-CTL-001
0
Sheet 7
14
of
7
Service Manual
DIG-610
CAPACITOR BANK BOARD 53258-05 CON1
224/AC250V
5 4 3 2 1
224/AC250V SC-10-D050
TB1 68K 1W
330/400V
TB3
1
DC+ OUT
1
R7 1/2W470K R8
L2
C8 102/AC250V
C13
C53 C54 C55
C9 C10 C15 C16 C11 C14 -
C12
+
+
102/AC250V C17
+
+
+
+
+
+
FET GATE DRIVE CIRCUIT
R9 R12 NTC10D22
BD1 D25XB60
TB2 DC- OUT
TB4
1
68K 1W
S1
1
+24V GND
330u/400V/30mm*40mm SIG1(+)
BTA26-600 Q11
R13 10K
F1
50T2A L3 SQE20
C18
+24V GND
100/400V
D14 1N4937
SIG1(-) T1
PQ32/30
1 C19 M104/100V
+ R14 1W100
D6 1N4937 C20 47/25V 2 4
1 2 3 4 5 6 7 8
CON7
DRV2A
G
S1
S
1 2 35312
1 2 3 4 5 6 7 8
CON10 G
DRV2A
1 2
S C47
R15 2W24K
R16 2W 100K
1
IN
OUT
C22 2200u/25V
NP C48 22/25V
C45
+
R18 220
8 5
4
3
BOX104
C28 TT22/16V
PG_S 3
C27
TC7662A1 1
6
8
Q12 IXIF 16N90Q 1
C29
D15 S2L60 9
C30
BOX102
R19 3.9K
R22 C31
-12V 20mA
+24V
TP4 +24V 0.2A 1
470/25V
+24V +24V GND SIG2(-)
FIL_CONT LAMP_CONT
1
BOX102
D9 D10LC20U
+15V
35312
CON9 G
1 2
S
35312
100V_SEL
+5V
+ C33 2200/16V(13pie)
+ C34 2200/16V
1
R25 5.1K
1
3
Q18 PN2222
Q16 C1815
Q17 PN2907
R30 + C40 22/25V
1W1K
20K 1N4746
R32 10
CON4 53258-4
Q19 IRF640
Z7 Q20 PN2907
R33 C41 M472/100V
D13 1W1K
R34 20K
R35 5W/0.1 1N4937
2
4
1 2 3 4
Z6 1N4937
R31 1%9.3K
2
+15V
D12 R28
+15V
100
Q15 IRF640
1N4746
R24 1/2W470
U5 PC817 C39 10uF/25V
R27 10
R41 47K
12 R26
M472/100V C37
Q14 PN2222
R23 1W680 2200/16V
1
GND
1 2
1 2 3 4 5 6 7 8 DRV2A
C35 +
6
2
S
+12.2V
U4 78L05
+
G
S1
+15V
1
1N4937 47/25V
+
CON8
DRV2A
FIL_FB
TP7 +15V 6A
10
5
+ C36
OUT
1 2 3 4 5 6 7 8
35312-11
D10
C38 100uF/25V
TP5
C32
BOX103
IN
CON3 1 2 3 4 5 6 7 8 9 10 11
C51
PG
3
8 7 6 5
C25 C26 100/16V TT22/16V +
U3 +
120
BOX472 R20 20K
R21 5W 0.2 1N4747 TP6
+
Z5
1 2 3 4
1 2 3 4 5 CON2 35312-5
1
2
7
R17 4.7
TP4
+
UC3844
1
D11 1N4937
100/16V +
C23 100/16V
HS1 U7 2
TG
C46 104
HS3 104
+5V GND +12.2V GND -12V
C24 +
2
R29 27K
3
1
7
1
C2235
+24V GND
GND 1
D7 1N4937
U2 LM7805CT
3 Q21
TP3
+5V D8 S2L60 TP2
35312
S1 TP1 +5V 0.5A
GND
1/400V
C21 222/1KV
1
TH
3
C42 101/3KV
R37 4.7K 1
C43 104 R38 10K
R36 5D11
50N60
APPROVED BY
DESIGNED BY
PXP-60HF PWR & DRV DIG-610 PWR &DRV
2
ISO_5V
HS4 Q13
t 2
U6 TL431 ISO_GND
R39 1%1.8K
2 1
222/AC250V
3 2 1
C44 HAND SW PWR
COLIMETER CON5
CON6
35312-3
53258-2
CHECKED BY
DATE Size A3
FSCM NO.
DWG NO
Rev A
PXP60HF-PWR
Sheet 1
15
of
1
Service Manual
DIG-610
8. Maintenance 8-1 Maintenance and periodic operator Maintenance has checked periodically with manual by the user. User checks items as following. -
Cleaning of product is clean on the outside surface when external form is dirty and you clean by cleaning material etc… with soft cloth. But do not clean surface of product with water.
-.
Check If the bolt of Hand switch housing under hand switch is loosed when you used the product.
-
When main power of product is on, and display will show values (kV and mAs)
as following .
And this product is normal, please use safely. .
kV
mAs
-
If you have a problem, show error code on display and see page 28 and check it. After the error code on display check it, if this error code is not appear Please just let me know. Tel +82-31-907-7222 Fax +82-31-904-2231 Homepage: www.dong-mun.co.kr E-mail: [email protected]
16
Service Manual
DIG-610
8. Technical Technical date
TECHNICAL GUIDE LOCATION
DISTANCE
GRID
KVP
mA
TIME
SIZE
CHEST
100-110
+
65
30
0.2-0.3
10 × 12
CHEST(LAT)
100-110
+
65
30
0.3-0.4
10 × 12
CHEST
100-110
+
75
25
0.5-0.8
14 × 14
SKULL(AP)
100
75
25
0.8-1
8 × 10
SKULL(LAT)
100
75
25
0.3
8 × 10
ABDOMEN
100
75(80)
25(20)
1.5-2
14 × 17
FOREARM
100
65
30
0.2 - 0.3
10 × 12
WATER'S VIEW
80
75
25
0.8 - 1
8 × 10
ANKIE
80-100
65
30
0.2-0.3
8 × 10
KNEE
80-100
65
30
0.2 - 0.3
10 × 12
WRIST
80-100
65
30
0.14-0.18
8 × 10
RIB LOWER
100
80
20
1.7 - 2
14 × 17
C-SPINE(AP)
100
75
25
0.8-1
8 × 10
C-SPINE(LAT)
100
75
25
0.3
8 × 10
T-SPINE(AP)
100
+
75
25
2
14 × 14
T-SPINE(LAT)
100
+
80
20
2 - 2.5
7 × 17
L-SPINE(AP)
100
+
90
15
2.5 - 3
14 × 14
L-SPINE(LAT)
100
+
90
15
3-5
7 × 17
PEVRIS
100
+
80
20
2.5 - 3
14 × 17
HIP JOINT
100
90
15
2.5 - 3
10 × 12
+
+
17