FloWEB Gas Flow Metering System

User Manual Firmware version 10905 ModBus version R11

Edition 2

Revision 4

Date 05/11/2010

FloWEB User Manual

Page 2 di 2

FloWEB_UM_UNICIG_R24 ENG.doc

Pietro Fiorentini pursues a policy of continuous development and product improvement. The information contained in this document is, therefore, subject to change without notice. To the best of our knowledge, the information contained in this document is accurate. However, Pietro Fiorentini cannot be held responsible for any errors, omissions or inaccuracies, or any losses incurred as result of them.

IMPORTANT NOTICE Because Pietro Fiorentini is continuously improving its products, some of the menus which appear on your instrument’s display may not be exactly as illustrated and described in this manual. However, because the menus are simple and intuitive, this should not cause any major problems.

This manual is concurrent with embedded software version 10905

FloWEB User Manual – Edition 2 Revision 4 Page 2/ 2 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 3 di 3

FloWEB_UM_UNICIG_R24 ENG.doc

Static precautions Some parts of the instrument (such as circuit boards) may be damaged by static electricity. Therefore, when carrying out any work which involves the risk of static damage to the instrument, the instructions show the following notice:

While carrying out this procedure you must wear an earthed wrist strap at all times to protect the instrument against static shock.

At such times you must wear an earthed wrist-strap to protect the instrument.

Important Warnings!! 1. FloWEB is not ATEX approved and, therefore, can only be used in officially designated safe (non-hazardous) areas. 2. All examples of connections in this manual are designed to be attempted in a non-hazardous area. 3. Hazardous area considerations are outside the scope of this manual. Always refer to installation documentation supplied by the manufacturer for their instrumentation.

FloWEB User Manual – Edition 2 Revision 4 Page 3/ 3 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 4 di 4

Table of Contents 1.

INTRODUCTION .....................................................................................................................................8

1.1. BACKGROUND ............................................................................................................................................8 1.1.1. Product Range..................................................................................................................................8 1.1.2. Product identification ........................................................................................................................8 1.2. REFERENCES .............................................................................................................................................9 1.2.1. Hardware Configurations ..................................................................................................................9 1.2.2. Firmware Release.............................................................................................................................9 2.

OPERATOR’S PANEL ..........................................................................................................................10

2.1. FRONT PANEL ..........................................................................................................................................10 2.1.1. LCD Display....................................................................................................................................10 2.1.2. Diagnostic LEDs .............................................................................................................................11 2.1.3. Operator Keypad ............................................................................................................................11 2.1.4. Ethernet Port...................................................................................................................................11 2.1.5. Programming button .......................................................................................................................11 2.1.6. Functional Keys ..............................................................................................................................11 2.1.7. Memory Card ..................................................................................................................................11 3.

BACK PANEL........................................................................................................................................12

3.1. BACK PANEL TERMINATIONS.....................................................................................................................12 3.1.1. Terminal Identification.....................................................................................................................13 3.1.2. FloWEB configurations ...................................................................................................................13 3.1.2.1. 3.1.2.2. 3.1.2.3. 3.1.2.4. 3.1.2.5.

4.

FloWEB single stream .......................................................................................................................... 14 FloWEB two streams............................................................................................................................. 15 FloWEB single stream with Analogue Output option board .................................................................. 16 FloWEB two streams with Analogue Output option board .................................................................... 17 FloWEB two streams with Analogue Output and FieldBus option boards ............................................ 18

FLOWEB INPUT CONNECTIONS........................................................................................................19

4.1. OVERVIEW OF FLOWEB INPUTS ...............................................................................................................19 4.2. CHANNELS CONFIGURATION .....................................................................................................................19 4.2.1. Analogue Input Channels Assignments..........................................................................................21 4.3. CONNECTIONS TO FLOW METER ...............................................................................................................22 4.3.1. Turbine (pulses) Flow Meter Inputs ................................................................................................22 4.3.2. Ultrasonic (AGA 9) Flow Meter Inputs ............................................................................................23 4.3.3. Orifice Flow Meter Inputs................................................................................................................23 4.4. CONNECTIONS TO TEMPERATURE TRANSDUCERS .....................................................................................24 4.4.1. mA type Temperature Transmitter..................................................................................................24 4.4.2. RTD type Temperature Transducer................................................................................................24 4.5. CONNECTIONS TO PRESSURE TRANSDUCERS ...........................................................................................25 4.5.1. mA type Pressure Transmitter ........................................................................................................25 4.6. ANALOGUE INPUTS BOARD .......................................................................................................................26 4.7. DIGITAL INPUT CONNECTIONS ...................................................................................................................26 5.

FLOWEB OUTPUT CONNECTIONS....................................................................................................27

5.1. OVERVIEW OF FLOWEB OUTPUTS ...........................................................................................................27 5.1.1. Digital/Pulse Outputs Connections .................................................................................................28 5.1.2. Analogue Outputs Connections ......................................................................................................29 5.2. POWER SUPPLY BOARD ...........................................................................................................................30 FloWEB User Manual – Edition 2 Revision 4 Page 4/ 4 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 5 di 5

5.3.

ANALOGUE OUTPUTS BOARD....................................................................................................................31

6.

FLOWEB INSTALLATION ....................................................................................................................32

6.1. OVERVIEW ...............................................................................................................................................32 6.2. SAFE AND HAZARDOUS ZONES .................................................................................................................32 6.3. INSTALLATION PROCEDURE .......................................................................................................................32 6.3.1. Step 1: Drawing-up a wiring schedule ............................................................................................32 6.3.2. Step 2: Unpacking the FloWEB ......................................................................................................33 6.3.3. Step 3: Setting the DIP switches ....................................................................................................33 6.3.4. Step 4: Connecting back-up battery ...............................................................................................36 6.3.5. Step 5: Fitting the FloWEB .............................................................................................................37 6.3.6. Step 6: Making FloWEB external connection .................................................................................37 6.3.7. Step 7: Earthing FloWEB................................................................................................................38 6.3.8. Step 8: Connecting power supply ...................................................................................................38 6.4. FLOWEB MARKING AND SEALING .............................................................................................................39 7.

FLOWEB USER INTERFACE...............................................................................................................46

7.1. MAIN PAGE ..............................................................................................................................................46 7.1.1. Display Contrast Control.................................................................................................................50 7.2. FLOWEB CONFIGURATION MENU .............................................................................................................51 7.2.1. Date & Time Setting........................................................................................................................52 7.2.2. Engineering Units Configuration .....................................................................................................53 7.2.3. System Configuration .....................................................................................................................54 7.2.4. Analogue Inputs Configuration .......................................................................................................56 7.2.4.1. 7.2.4.2. 7.2.4.3.

7.2.5. 7.2.6. 7.2.7.

Analogues Outputs Configuration...................................................................................................63 Digital Outputs Configuration..........................................................................................................65 Calculations Settings Menu ............................................................................................................67

7.2.7.1. 7.2.7.2. 7.2.7.3. 7.2.7.4.

7.2.8. 7.2.9.

Hart Inputs Configuration ...................................................................................................................... 58 Flow Meter Inputs Configuration ........................................................................................................... 59 Flow Meter Error Correction Configuration............................................................................................ 61

GAS Composition Menu........................................................................................................................ 68 Totals Counters page ............................................................................................................................ 71 References Settings .............................................................................................................................. 76 Orifice Settings ...................................................................................................................................... 77

Alarms Configuration ......................................................................................................................79 Utility Page......................................................................................................................................81

7.2.9.1. 7.2.9.2. 7.2.9.3. 7.2.9.4. 7.2.9.5. 7.2.9.6. 7.2.9.7. 7.2.9.8.

Virtual printer......................................................................................................................................... 81 Communication Port Configuration ....................................................................................................... 82 Redundancy Configuration.................................................................................................................... 90 Date & Time Settings ............................................................................................................................ 91 Users Login Passwords......................................................................................................................... 92 TCP/IP Network Configuration .............................................................................................................. 94 Log and Report ..................................................................................................................................... 95 Compact Flash Configuration.............................................................................................................. 102

7.3. VISUALIZATIONS MENU ...........................................................................................................................103 7.3.1. Diagnostic and Alarms..................................................................................................................103 7.3.2. Alarms Acknowledgement ............................................................................................................105 7.3.3. Event Register ..............................................................................................................................105 7.3.4. Alarm Register ..............................................................................................................................105 7.3.5. Users Login Passwords ................................................................................................................107 7.3.6. Totalizers page .............................................................................................................................107 7.3.7. Current GAS Composition ............................................................................................................108 7.3.8. Compact Flash..............................................................................................................................109 7.3.9. Instant Print...................................................................................................................................110 FloWEB User Manual – Edition 2 Revision 4 Page 5/ 5 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 6 di 6

7.3.10. On/Off Date & Time...................................................................................................................111 7.3.11. Ultrasonic information ...............................................................................................................111 7.4. MAINTENANCE SETTINGS MENU .............................................................................................................112 7.4.1. Users Login Passwords ................................................................................................................112 7.4.2. Compact Flash servicing ..............................................................................................................113 7.4.3. Test...............................................................................................................................................114 8.

COMMUNICATION AND NETWORKING...........................................................................................115

8.1. FLOWEB COMMUNICATION CAPABILITIES ...............................................................................................115 8.1.1. Comm #1 & Comm #2 connections and settings (CPU Board) ....................................................115 8.1.2. Comm #3 connection and setting (Auxiliary Board)......................................................................118 8.1.3. Comm #4 & Comm #5 connections and settings (Analogue Input board)....................................120 8.1.4. Ethernet Connection .....................................................................................................................122 8.1.5. CAN Bus Connection....................................................................................................................123 8.2. FLOWEB MODBUS COMMUNICATION......................................................................................................124 8.2.1. Supported Commands..................................................................................................................124 8.2.2. Register Addressing .....................................................................................................................124 8.2.3. Floating Point Numbers ................................................................................................................125 8.2.4. Connecting FloWEB to a RS-232/RS-422/RS-485 Link ...............................................................125 8.2.5. Connecting FloWEB to a TCP/IP network ....................................................................................128 9.

HART, SMART AND FLOWEB ...........................................................................................................129

9.1.

INTRODUCTION TO HART, SMART AND THE FLOWEB..............................................................................129

10.

REDUNDANCY AND FLOWEB ..........................................................................................................131

10.1. 10.2. 11.

INTRODUCTION TO HOT/STANDBY, REDUNDANCY AND THE FLOWEB....................................................131 REDUNDANT FLOWEB WIRING REQUIREMENTS ...................................................................................132 PRINCIPLE OF OPERATION .............................................................................................................134

11.1. ACQUISITION ......................................................................................................................................134 11.1.1. Alarms .......................................................................................................................................134 11.1.2. Turbine - Fidelity check .............................................................................................................134 11.2. CALCULATION.....................................................................................................................................134 11.2.1. Compressibility Gas Factor (Z) .................................................................................................134 11.2.2. Calculation of Relative density (Specific gravity) ......................................................................136 11.2.3. TURBINE configuration.............................................................................................................137 11.2.3.1. 11.2.3.2. 11.2.3.3.

11.2.4. 11.2.4.1.

Correction of Flow Meter Error (Vc)..................................................................................................... 137 Volume at Base Conditions (Vb) ......................................................................................................... 138 Flow Rates (Q) .................................................................................................................................... 138

ORIFICE configuration ..............................................................................................................139 Flow Rate (Q) ...................................................................................................................................... 139

11.2.5. Mass Flow Rate (Qmb) .............................................................................................................139 11.2.6. Energy Flow Rate (Qeb) ...........................................................................................................140 11.3. COMMUNICATION ................................................................................................................................140 11.3.1. Serial Ports................................................................................................................................140 11.3.2. Ethernet Port .............................................................................................................................140 11.3.3. Communication Protocol ...........................................................................................................140 12.

FLOWEB HTTP...................................................................................................................................141

13.

TECHNICAL SPECIFICATIONS .........................................................................................................145

13.1. 13.2.

OVERALL SPECIFICATIONS ..................................................................................................................145 CPU BOARD .......................................................................................................................................145 FloWEB User Manual – Edition 2 Revision 4 Page 6/ 6 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 7 di 7

13.3. ANALOGUES INPUTS BOARD ................................................................................................................146 13.3.1. General Specification ................................................................................................................146 13.3.2. Analogues Inputs specifications................................................................................................146 13.3.3. Digital Inputs Specifications ......................................................................................................147 13.4. ANALOGUES OUTPUTS BOARD ............................................................................................................148 13.4.1. General Specifications ..............................................................................................................148 13.4.2. Outputs Specifications ..............................................................................................................148 13.5. DIGITAL OUTPUTS (ON POWER SUPPLY BOARD) ..................................................................................148 13.5.1. Mos-switches Digital outputs.....................................................................................................148 13.5.2. PSU Status Relays....................................................................................................................148 13.6. COMMUNICATION INTERFACES ............................................................................................................148 13.6.1. CPU Board ................................................................................................................................148 13.6.2. Auxiliary Board ..........................................................................................................................148 13.6.3. Analogues Inputs Board............................................................................................................149 13.6.4. Serial ports Specifications.........................................................................................................149 APPENDIX A

BLANK WIRING SCHEDULE .............................................................................................150

FloWEB User Manual – Edition 2 Revision 4 Page 7/ 7 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

1.

Page 8 di 8

Introduction 1.1.

Background

FloWEB is designed to meet the demand for a reliable, versatile, flexible, user friendly and cost effective instrument for natural gas metering. FloWEB can be used in addition to Turbine, Orifice and Ultrasonic flow meter with no hardware specialization. This document provides information on installation, principal of operation, user interface and hardware configuration, to support proper setup and use of FloWEB product.

1.1.1.

Product Range

FloWEB flow computer can be shipped in different hardware configurations, according to the type and number of boards installed in the system bus. The following table provides the list of available configurations and related product codes.

Part Number 3480410 3480411 3480412 3480413

Description Single stream configuration, no analogue outputs Single stream configuration with analogue outputs board Two stream configuration, no analogue outputs Two stream configuration with analogue outputs board

1.1.2.

Product identification

FloWEB configuration can be easily identified by checking the label on the right side of the instrument (see Fig. 1-1); please note that thanks to the modularity of hardware and firmware configuration initial set-up can be upgraded by simply adding the required board in any moment.

FloWEB User Manual – Edition 2 Revision 4 Page 8/ 8 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 9 di 9

FloWEB_UM_UNICIG_R24 ENG.doc

FloWEB Part Number

Figure 1-1: FloWEB part number identification (side view)

1.2.

References 1.2.1.

Hardware Configurations

The present document referee to FloWEB part numbers as reported at section 1.1.1; please contact your product dealer for information on different codes.

1.2.2.

Firmware Release

The software version dealt in this manual is given at Chapter 7.1 tells how to find out what software is installed in the instrument. Further software upgrades may affect the content of the present manual; please, be sure that the manual software version be aligned to the product software release installed.

FloWEB User Manual – Edition 2 Revision 4 Page 9/ 9 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 10 di 10

FloWEB_UM_UNICIG_R24 ENG.doc

2.

Operator’s Panel 2.1.

Front Panel

FloWEB front panel is shown in Fig. 2-1 below. It’s used to display and enter all data. All data can also be accessed via any of the serial ports or with a browser via Ethernet port.

2.1.1.

LCD Display

FloWEB use a graphic back-lit Liquid Crystal TFT Display with ¼ VGA 320 x 240 pixel resolution. Back lighting and viewing angle are adjustable using the ⇒ key to increase and ⇐ key to decrease when in the Main Page Menu.

Diagnostic LEDs

Ethernet Port

LCD Display

Programming Button

Keypad

Functional Keys

Memory Card

Figure 2-1: FloWEB Front Panel

FloWEB User Manual – Edition 2 Revision 4 Page 10/ 10 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

2.1.2.

Page 11 di 11

Diagnostic LEDs

Six diagnostic LEDs are located in the left side of the front panel. LEDs meaning are the followings: • • • • • •

ON WD ALM CFG ERX LNK

: glow green when Power Supply is present; : flash red when Watch Dog is triggered (FloWEB fault): during normal operation it’s off; : glow red when the real time diagnostic routine find a faulty condition; : glow yellow when the programming button is pressed to enter Configuration mode; : glow yellow whenever some activity is present on Ethernet port; : glow green when Ethernet connection is established.

2.1.3.

Operator Keypad

Control of the FloWEB is via a 24-button membrane keypad with tactile and audio feedback; on the right side are located 10 numeric and 8 fixed function keys, while on the bottom are located 6 menu driven function keys. Through the keypad the User has the capability to configure FloWEB, access and modify calibration data and view process data.

2.1.4.

Ethernet Port

On left side a RJ45 connector can be used for Ethernet connection of a Laptop PC with a Web Browser interface or to a TCP/IP ModBus network. FloWEB has an integrated WEB Server with HTML pages inside.

2.1.5.

Programming button

The programming button is located behind the metrological sealed screw on the front panel lower left corner. Whenever any parameter affecting custody transfer has to be changed, the operator has to remove the screw and push the programming button; otherwise no change will take place. When programming button is pressed, CFG LED will light; after 3 minutes without any change to parameters, programming button action will expire.

2.1.6.

Functional Keys

In the lower part of the front panel, under the display, there are 6 function keys, F1 to F6; the function of these keys changes according the menu. In the last row of the display a label indicate the function of each key.

2.1.7.

Memory Card

Removing the two screws of the panel in the right side of the front panel, it’s possible to access the memory card used to store all the logs, alarm and report generated by FloWEB.

FloWEB User Manual – Edition 2 Revision 4 Page 11/ 11 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 12 di 12

FloWEB_UM_UNICIG_R24 ENG.doc

3.

Back Panel 3.1.

Back Panel Terminations

Figure 3-1 below shows FloWEB Back Panel. Seven slots are available for various board types. Board positions in the back panel are fixed.

Slot ID 1 2 3 4 5 6 7

Board Type Power Supply Field Bus Analogue Output Analogue Input #2

Function Power supply board, provides internal supply and auxiliary output for sensors Field Bus Interface (future) Six output 4-20mA current Output Six analogue input: 4-20mA / 1-5V /Hart / RTD and Digital Counter/Comm Port #5 Input board – Stream #2 Analogue Input #1 Six analogue input: 4-20mA / 1-5V Hart / RTD and Digital Counter/Comm Port #4 Input board – Stream #1 CPU Main Processor Unit, Comm Port #1 & #2 Auxiliary Auxiliary board, Comm Port #3 and Ethernet front panel connection remainder

Each board has one or two removable terminals block for field signal and power connections.

Power Supply

FieldBus

Analogue Analogue Analogue Output Input #2 Input #1

CPU

Auxiliary

Board Fixing Screws Front panel Ethernet remainder

FUSE Earth Terminal

Ethernet Port

Figure 3-1: FloWEB Back Panel

FloWEB User Manual – Edition 2 Revision 4 Page 12/ 12 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 13 di 13

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.1.

Terminal Identification

Each terminal is identified with a letter and a number from A1 to A16 and B1 to B16. For board with single terminals block the terminal are identified with A1..A16. (FieldBus board has four terminals only).

Terminals Block Left Side A1

Terminals Block Right Side B1 A

A4/12/14/16

3.1.2.

B

B14/16

FloWEB configurations

FloWEB is a modular flow computer and according to the type of boards inserted in the system bus it’s possible to adapt to different metering needs. Some boards form FloWEB basis and are always present in the system; these boards are: Power Supply board, CPU board, Auxiliary board and Analogue Input #1 board. Other boards, Analogue Input #2, Analogue Output and FieldBus, are options.

FloWEB User Manual – Edition 2 Revision 4 Page 13/ 13 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 14 di 14

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.2.1. FloWEB single stream In Table 3-1 are shown back panel terminations for a single stream configuration instrument.

Pin

Power Supply A B Relay 2 Relay 1 Common Common

Analogue Input #1 A B Comm #4 Comm #4 Rs232 TX Rs485/422 TX+ Comm #4 Comm #4 Rs232 RX Rs485/422 TXComm #4 Comm #4 Common Common Comm #4 Comm #4 Rs232 RTS Rs485/422 RX+ Comm #4 Comm #4 Rs232 CTS Rs485/422 RXInput #1 Input #2 +24V Out +24V Out

CPU

Auxiliary A Comm #3 Rs232 TX

2

Relay 2 Open

Relay 1 Open

3

Relay 2 Close N/C

Relay 1 Close N/C

5

Digital Out1

Digital Out5

6

Digital Out2

Digital Out6

7

Digital Out3

Digital Out7

Input #1 In+

Input #2 In+

Comm #2 Rs232 RX

8

Digital Out4 Digital Out Common

Digital Out8 Digital Out Common

Input #1 In-

Input #2 In-

Input #3 +24V Out

Input #4 +24V Out

Comm #2 Common Comm #2 Rs232 RTS

10

+24V Out

N/C

Input #3 In+

Input #4 In+

Comm #2 Rs232 CTS

11

+24V Gnd

Input #3 In-

Input #4 In-

CanBus H

12

N/C

Input #5 +24V Out

CanBus L

N/C

Power Supply Input + Power Supply Gnd

Input #6 PT100 Supply + Input #6 PT100 In+

CanBus L

13

UPS Low Batt Input UPS Power Fail Input UPS Fail

B Comm #1 Rs485/422 TX+ Comm #1 Rs485/422 TXComm #1 Common Comm #1 Rs485/422 RX+ Comm #1 Rs485/422 RXComm #2 Rs485/422 TX+ Comm #2 Rs485/422 TXComm #2 Common Comm #2 Rs485/422 RX+ Comm #2 Rs485/422 RXCanBus H

CanBus Common

CanBus Common

N/C

UPS Common

Input #5 In-

Input #6 PT100 In-

N/C

N/C

N/C

15

Input #7 In+

16

Input #7 In-

Input #6 PT100 Supply +24V Out

1

4

9

14

Input #5 In+

A Comm #1 Rs232 TX Comm #1 Rs232 RX Comm #1 Common Comm #1 Rs232 RTS Comm #1 Rs232 CTS Comm #2 Rs232 TX

Comm #3 Rs232 RX Comm #3 Common Comm #3 Rs232 RTS Comm #3 Rs232 CTS Comm #3 Rs485/422 TX+ Comm #3 Rs485/422 TXComm #3 Common Comm #3 Rs485/422 RX+ Comm #3 Rs485/422 RXN/C

Table 3-1: Single stream configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 14/ 14 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 15 di 15

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.2.2. FloWEB two streams In Table 3-2 are shown back panel terminations for a two stream configuration instrument.

Pin

Power Supply A B Relay 2 Relay 1 Common Common

Analogue Input #2 A B Comm #5 Comm #5 Rs232 TX Rs485/422 TX+ Comm #5 Comm #5 Rs232 RX Rs485/422 TXComm #5 Comm #5 Common Common Comm #5 Comm #5 Rs232 RTS Rs485/422 RX+ Comm #5 Comm #5 Rs232 CTS Rs485/422 RXInput #1 Input #2 +24V Out +24V Out

Analogue Input #1 A B Comm #4 Comm #4 Rs232 TX Rs485/422 TX+ Comm #4 Comm #4 Rs232 RX Rs485/422 TXComm #4 Comm #4 Common Common Comm #4 Comm #4 Rs232 RTS Rs485/422 RX+ Comm #4 Comm #4 Rs232 CTS Rs485/422 RXInput #1 Input #2 +24V Out +24V Out

CPU

Auxiliary A Comm #3 Rs232 TX

2

Relay 2 Open

Relay 1 Open

3

Relay 2 Close N/C

Relay 1 Close N/C

5

Digital Out1

Digital Out5

6

Digital Out2

Digital Out6

7

Digital Out3

Digital Out7

Input #1 In+

Input #2 In+

Input #1 In+

Input #2 In+

Comm #2 Rs232 RX

8

Digital Out4 Digital Out Common

Digital Out8 Digital Out Common

Input #1 In-

Input #2 In-

Input #1 In-

Input #2 In-

Input #3 +24V Out

Input #4 +24V Out

Input #3 +24V Out

Input #4 +24V Out

Comm #2 Common Comm #2 Rs232 RTS

10

+24V Out

N/C

Input #3 In+

Input #4 In+

Input #3 In+

Input #4 In+

Comm #2 Rs232 CTS

11

+24V Gnd

Input #3 In-

Input #4 In-

Input #3 In-

Input #4 In-

CanBus H

12

N/C

Input #5 +24V Out

Input #6 PT100 Supply + Input #6 PT100 In+

CanBus L

CanBus L

N/C

Power Supply Input + Power Supply Gnd

Input #6 PT100 Supply + Input #6 PT100 In+

Input #5 +24V Out

13

UPS Low Batt Input UPS Power Fail Input UPS Fail

B Comm #1 Rs485/422 TX+ Comm #1 Rs485/422 TXComm #1 Common Comm #1 Rs485/422 RX+ Comm #1 Rs485/422 RXComm #2 Rs485/422 TX+ Comm #2 Rs485/422 TXComm #2 Common Comm #2 Rs485/422 RX+ Comm #2 Rs485/422 RXCanBus H

CanBus Common

CanBus Common

N/C

UPS Common

Input #5 In-

Input #6 PT100 In-

Input #5 In-

Input #6 PT100 In-

N/C

N/C

N/C

15

Input #7 In+

Input #7 In+

16

Input #7 In-

Input #6 PT100 Supply +24V Out

Input #6 PT100 Supply +24V Out

1

4

9

14

Input #5 In+

Input #5 In+

Input #7 In-

A Comm #1 Rs232 TX Comm #1 Rs232 RX Comm #1 Common Comm #1 Rs232 RTS Comm #1 Rs232 CTS Comm #2 Rs232 TX

Comm #3 Rs232 RX Comm #3 Common Comm #3 Rs232 RTS Comm #3 Rs232 CTS Comm #3 Rs485/422 TX+ Comm #3 Rs485/422 TXComm #3 Common Comm #3 Rs485/422 RX+ Comm #3 Rs485/422 RXN/C

Table 3-2: Two streams configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 15/ 15 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 16 di 16

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.2.3. FloWEB single stream with Analogue Output option board In Table 3-3 are shown back panel terminations for a single stream configuration with Analogue Output option board.

Pin

Power Supply A B Relay 2 Relay 1 Common Common

Analogue Output A Analog Out 1+

2

Relay 2 Open

Relay 1 Open

Analog Out 1-

3

Relay 2 Close N/C

Relay 1 Close N/C

Analog Out 2+

5

Digital Out1

Digital Out5

Analog Out 3+

6

Digital Out2

Digital Out6

Analog Out 3-

7

Digital Out3

Digital Out7

Analog Out 4+

Input #1 In+

Input #2 In+

Comm #2 Rs232 RX

8

Digital Out4 Digital Out Common

Digital Out8 Digital Out Common

Analog Out 4-

Input #1 In-

Input #2 In-

Analog Out 5+

Input #3 +24V Out

Input #4 +24V Out

Comm #2 Common Comm #2 Rs232 RTS

10

+24V Out

N/C

Analog Out 5-

Input #3 In+

Input #4 In+

Comm #2 Rs232 CTS

11

+24V Gnd

Analog Out 6+

Input #3 In-

Input #4 In-

CanBus H

12

N/C

Analog Out 6-

Input #5 +24V Out

CanBus L

N/C

Power Supply Input + Power Supply Gnd

+24V Out

Input #5 In+

Input #6 PT100 Supply + Input #6 PT100 In+

CanBus L

13

UPS Low Batt Input UPS Power Fail Input UPS Fail

B Comm #1 Rs485/422 TX+ Comm #1 Rs485/422 TXComm #1 Common Comm #1 Rs485/422 RX+ Comm #1 Rs485/422 RXComm #2 Rs485/422 TX+ Comm #2 Rs485/422 TXComm #2 Common Comm #2 Rs485/422 RX+ Comm #2 Rs485/422 RXCanBus H

CanBus Common

CanBus Common

N/C

UPS Common

+24V Gnd

Input #5 In-

Input #6 PT100 In-

N/C

N/C

N/C

15

Input #7 In+

16

Input #7 In-

Input #6 PT100 Supply +24V Out

1

4

9

14

Analog Out 2-

Analogue Input #1 A B Comm #4 Comm #4 Rs232 TX Rs485/422 TX+ Comm #4 Comm #4 Rs232 RX Rs485/422 TXComm #4 Comm #4 Common Common Comm #4 Comm #4 Rs232 RTS Rs485/422 RX+ Comm #4 Comm #4 Rs232 CTS Rs485/422 RXInput #1 Input #2 +24V Out +24V Out

CPU A Comm #1 Rs232 TX Comm #1 Rs232 RX Comm #1 Common Comm #1 Rs232 RTS Comm #1 Rs232 CTS Comm #2 Rs232 TX

Auxiliary A Comm #3 Rs232 TX Comm #3 Rs232 RX Comm #3 Common Comm #3 Rs232 RTS Comm #3 Rs232 CTS Comm #3 Rs485/422 TX+ Comm #3 Rs485/422 TXComm #3 Common Comm #3 Rs485/422 RX+ Comm #3 Rs485/422 RXN/C

Table 3-3: Single stream with Analogue Output option board configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 16/ 16 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 17 di 17

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.2.4. FloWEB two streams with Analogue Output option board In Table 3-4 are shown back panel terminations for a two stream configuration with Analogue Output option board.

Pin

Power Supply

Analogue Output A Analog Out 1+

Analogue Input #2

A Relay 2 Common

B Relay 1 Common

2

Relay 2 Open

Relay 1 Open

Analog Out 1-

Comm #5 Rs232 RX

3

Relay 2 Close N/C

Relay 1 Close N/C

Analog Out 2+

Comm #5 Common Comm #5 Rs232 RTS

5

Digital Out1

Digital Out5

Analog Out 3+

Comm #5 Rs232 CTS

6

Digital Out2

Digital Out6

Analog Out 3-

7

Digital Out3

Digital Out7

8

Digital Out4 Digital Out Common

10

Analogue Input #1 A Comm #4 Rs232 TX

Input #1 +24V Out

B Comm #5 Rs485/422 TX+ Comm #5 Rs485/422 TXComm #5 Common Comm #5 Rs485/422 RX+ Comm #5 Rs485/422 RXInput #2 +24V Out

Analog Out 4+

Input #1 In+

Digital Out8 Digital Out Common

Analog Out 4-

+24V Out

11

+24V Gnd

12

N/C

13

Power Supply Input + Power Supply Gnd

CPU A Comm #1 Rs232 TX

Input #1 +24V Out

B Comm #4 Rs485/422 TX+ Comm #4 Rs485/422 TXComm #4 Common Comm #4 Rs485/422 RX+ Comm #4 Rs485/422 RXInput #2 +24V Out

Input #2 In+

Input #1 In+

Input #2 In+

Comm #2 Rs232 RX

Input #1 In-

Input #2 In-

Input #1 In-

Input #2 In-

Analog Out 5+

Input #3 +24V Out

Input #4 +24V Out

Input #3 +24V Out

Input #4 +24V Out

Comm #2 Common Comm #2 Rs232 RTS

N/C

Analog Out 5-

Input #3 In+

Input #4 In+

Input #3 In+

Input #4 In+

Comm #2 Rs232 CTS

UPS Low Batt Input UPS Power Fail Input UPS Fail

Analog Out 6+

Input #3 In-

Input #4 In-

Input #3 In-

Input #4 In-

Analog Out 6-

Input #5 +24V Out

Input #5 +24V Out

+24V Out

Input #5 In+

Input #6 PT100 Supply + Input #6 PT100 In+

Input #6 PT100 Supply + Input #6 PT100 In+

UPS Common

+24V Gnd

Input #5 In-

Input #6 PT100 In-

Input #5 In-

Input #6 PT100 In-

15

Input #7 In+

Input #7 In+

16

Input #7 In-

Input #6 PT100 Supply +24V Out

Input #6 PT100 Supply +24V Out

1

4

9

14

Analog Out 2-

A Comm #5 Rs232 TX

Comm #4 Rs232 RX Comm #4 Common Comm #4 Rs232 RTS Comm #4 Rs232 CTS

Input #5 In+

Input #7 In-

Auxiliary A Comm #3 Rs232 TX

CanBus H

B Comm #1 Rs485/422 TX+ Comm #1 Rs485/422 TXComm #1 Common Comm #1 Rs485/422 RX+ Comm #1 Rs485/422 RXComm #2 Rs485/422 TX+ Comm #2 Rs485/422 TXComm #2 Common Comm #2 Rs485/422 RX+ Comm #2 Rs485/422 RXCanBus H

CanBus L

CanBus L

N/C

CanBus Common

CanBus Common

N/C

N/C

N/C

N/C

Comm #1 Rs232 RX Comm #1 Common Comm #1 Rs232 RTS Comm #1 Rs232 CTS Comm #2 Rs232 TX

Comm #3 Rs232 RX Comm #3 Common Comm #3 Rs232 RTS Comm #3 Rs232 CTS Comm #3 Rs485/422 TX+ Comm #3 Rs485/422 TXComm #3 Common Comm #3 Rs485/422 RX+ Comm #3 Rs485/422 RXN/C

Table 3-4: Two streams with Analogue Output option board configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 17/ 17 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 18 di 18

FloWEB_UM_UNICIG_R24 ENG.doc

3.1.2.5. FloWEB two streams with Analogue Output and FieldBus option boards In Table 3-5 are shown back panel terminations for a two stream configuration with Analogue Output and FieldBus option boards.

Pin

Power Supply

FieldBus

A Relay 2 Common

B Relay 1 Common

A FieldBus +

Analogue Output A Analog Out 1+

Analogue Input #2

2

Relay 2 Open

Relay 1 Open

FieldBus -

Analog Out 1-

Comm #5 Rs232 RX

3

Relay 2 Close N/C

Relay 1 Close N/C

FieldBus Sch N/C

Analog Out 2+

Comm #5 Common Comm #5 Rs232 RTS

5

Digital Out1

Digital Out5

Analog Out 3+

Comm #5 Rs232 CTS

6

Digital Out2

Digital Out6

Analog Out 3-

7

Digital Out3

Digital Out7

8

Digital Out4 Digital Out Common

10

A Comm #4 Rs232 TX

Input #1 +24V Out

B Comm #5 Rs485/422 TX+ Comm #5 Rs485/422 TXComm #5 Common Comm #5 Rs485/422 RX+ Comm #5 Rs485/422 RXInput #2 +24V Out

Analog Out 4+

Input #1 In+

Digital Out8 Digital Out Common

Analog Out 4-

+24V Out

11

+24V Gnd

12

N/C

13

Power Supply Input + Power Supply Gnd

CPU A Comm #1 Rs232 TX

Input #1 +24V Out

B Comm #4 Rs485/422 TX+ Comm #4 Rs485/422 TXComm #4 Common Comm #4 Rs485/422 RX+ Comm #4 Rs485/422 RXInput #2 +24V Out

Input #2 In+

Input #1 In+

Input #2 In+

Comm #2 Rs232 RX

Input #1 In-

Input #2 In-

Input #1 In-

Input #2 In-

Analog Out 5+

Input #3 +24V Out

Input #4 +24V Out

Input #3 +24V Out

Input #4 +24V Out

Comm #2 Common Comm #2 Rs232 RTS

N/C

Analog Out 5-

Input #3 In+

Input #4 In+

Input #3 In+

Input #4 In+

Comm #2 Rs232 CTS

UPS Low Batt Input UPS Power Fail Input UPS Fail

Analog Out 6+

Input #3 In-

Input #4 In-

Input #3 In-

Input #4 In-

Analog Out 6-

Input #5 +24V Out

Input #5 +24V Out

+24V Out

Input #5 In+

Input #6 PT100 Supply + Input #6 PT100 In+

Input #6 PT100 Supply + Input #6 PT100 In+

UPS Common

+24V Gnd

Input #5 In-

Input #6 PT100 In-

Input #5 In-

Input #6 PT100 In-

15

Input #7 In+

Input #7 In+

16

Input #7 In-

Input #6 PT100 Supply +24V Out

Input #6 PT100 Supply +24V Out

1

4

9

14

Analog Out 2-

A Comm #5 Rs232 TX

Analogue Input #1

Comm #4 Rs232 RX Comm #4 Common Comm #4 Rs232 RTS Comm #4 Rs232 CTS

Input #5 In+

Input #7 In-

Auxiliary A Comm #3 Rs232 TX

CanBus H

B Comm #1 Rs485/422 TX+ Comm #1 Rs485/422 TXComm #1 Common Comm #1 Rs485/422 RX+ Comm #1 Rs485/422 RXComm #2 Rs485/422 TX+ Comm #2 Rs485/422 TXComm #2 Common Comm #2 Rs485/422 RX+ Comm #2 Rs485/422 RXCanBus H

CanBus L

CanBus L

N/C

CanBus Common

CanBus Common

N/C

N/C

N/C

N/C

Comm #1 Rs232 RX Comm #1 Common Comm #1 Rs232 RTS Comm #1 Rs232 CTS Comm #2 Rs232 TX

Comm #3 Rs232 RX Comm #3 Common Comm #3 Rs232 RTS Comm #3 Rs232 CTS Comm #3 Rs485/422 TX+ Comm #3 Rs485/422 TXComm #3 Common Comm #3 Rs485/422 RX+ Comm #3 Rs485/422 RXN/C

Table 3-5: Two streams with Analogue Output and FieldBus option boards configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 18/ 18 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 19 di 19

FloWEB_UM_UNICIG_R24 ENG.doc

4.

FloWEB Input Connections 4.1.

Overview of FloWEB Inputs

This section features connections with external devices that provide only input signals to the FloWEB. Use the following list to quickly locate the section of interest: • • • • •

4.2 Channel configuration 4.3 Connection to Flow Meters 4.4 Connection to Temperature Transducers 4.5 Connection to Pressure Transducers 4.6 Digital Input connections

4.2.

: : : : :

page 19 page 22 page 24 page 25 page 26

Channels Configuration

Each Analogue Input board has 7 input channels; channel IN1 to lN5 need hardware configuration to set input mode type. Channel IN6 and IN7 are specialized inputs: Channel IN6 is for 4 wires RTD temperature sensor (PT100) while IN7 will be used for future purposes. Software configurations are also required to complete input configurations (see Chapter 7). There are three dip switches located on the Input Board for channel configuration: SW1, SW2 and SW3 (see Fig. 4-1). Set the switches according to Table 4-1. Input # IN1

IN2 IN3

Input Type Dip switch # 4-20 mA

1 ●

2 ○

SW1 3 4 ●

1-5V









4-20mA + Hart







● ●





1-5V







1



4-20mA

○ ○

Flow meter (Pulse)

IN5

6

4-20mA

1-5V IN4

5

2



● ○

SW2 3 4

5

6

1

2

SW3 3 4 ○

5



6











4-20mA









1-5V









Flow meter (Pulse)









4-20mA







1-5V







● = ON, ○ = OFF, (void) = don’t care Table 4-1 – Analogue Input type Dip-switches settings

FloWEB User Manual – Edition 2 Revision 4 Page 19/ 19 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 20 di 20

FloWEB_UM_UNICIG_R24 ENG.doc SW5

SW4

123456 123456

123456 123456 123456

3

SW1

SW2

SW3

Figure 4-1 – Analogue Input Board Dip-switches location

FloWEB User Manual – Edition 2 Revision 4 Page 20/ 20 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 21 di 21

FloWEB_UM_UNICIG_R24 ENG.doc

4.2.1.

Analogue Input Channels Assignments

Some analogue input channels (AI) are related with specific sensor type depending on FloWEB selected configuration. The table below shows channel number assignments for connection of specific transmitters: • •

Turbine Configuration: Pressure, Temperature (transmitter or sensor), Flow Meter Orifice Configuration: Pressure, Temperature (transmitter or sensor), Differential Pressure,

Configuration

Turbine

Orifice

(*) (**) (***)

Channel #

Input Type

Sensor Type

Pressure Transmitter (P) Temperature Transmitter (T)

1 2 (*)

Current Loop, Voltage, Hart Current Loop, Voltage

Flow Meter pick-up #1

3

Dry contact, Namur

Flow Meter pick-up #2

4 (**)

Dry contact, Namur

Temperature Sensor (T) Pressure Transmitter (P) Temperature Transmitter (T)

6 (*) 1 2 (*)

PRT (PT100) Current Loop, Voltage, Hart Current Loop, Voltage

Differential Pressure (***) DP Low (DP1)

# Cells Used 1 2 3 X

3

Current Loop, Voltage

DP Medium (DP2) DP High (DP3) Temperature Sensor (T)

X

X X

4 5

Current Loop, Voltage Current Loop, Voltage

6 (*)

PRT (PT100)

X X

Ch. #2 is for connection of Temperature Transmitter; Ch #6 is for connection of PT100 sensor. (see §4.4 for details) Refer to §7.2.2.2 for details on the use of double pick-ups. Refer to §7.2.2 for details on the number of Differential Pressure Cells used. Table 4-2: Analog Input Channels assignment

When digital (HART) transmitters are used, these all have to be connected to channel #1.

Warning!!

Connecting up “SMART” transducers has to be done with great care. Powering-up more than one point-to-point configured transmitter on a HART network loop can produce an electrical current (20mA per transmitter) that can damage the FloWEB Analogue Inputs board.

FloWEB User Manual – Edition 2 Revision 4 Page 21/ 21 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 22 di 22

FloWEB_UM_UNICIG_R24 ENG.doc

4.3.

Connections to Flow Meter

Support is provided for wiring up to 2 (pulse) outputs from Flow Meter to each Analogue board of FloWEB.

4.3.1.

Turbine (pulses) Flow Meter Inputs

FloWEB can accept either single or dual pulse signal trains from the flow meter. There are ± signal pins for each pulse trains. Input channel 3 and 4 of each analog input board can be independently jumpered to accept pulse flow meters (set SW2 on flow meter position). No hardware setting is necessary to interface low frequency (LF) or high frequency (HF) flow meters. Reed, open collector and Namur type are accepted. When Flow meters need power supply it can be powered by the on board isolated power supply pin (+24V).

FloWEB back panel terminal Output Common

Reed/Open Collector/Namur output Flow Meter

Figure 4-2: Turbine Meter wiring connection

Warning! When interfacing Transmitters/Instruments in Hazardous Area, ATEX Safety Barrier or explosion proof installation MUST be used

FloWEB User Manual – Edition 2 Revision 4 Page 22/ 22 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 23 di 23

FloWEB_UM_UNICIG_R24 ENG.doc

4.3.2.

Ultrasonic (AGA 9) Flow Meter Inputs

FloWEB can accept single pulse signal train from ultrasonic flow meter. FloWEB back panel terminal Output Common

Ultrasonic Pulse Output Flow Meter

Figure 4-3: Ultrasonic (pulse) Meter wiring connection

Warning! When interfacing Transmitters/Instruments in Hazardous Area, ATEX Safety Barrier or explosion proof installation MUST be used

4.3.3.

Orifice Flow Meter Inputs

Up to 3 loop powered differential pressure cells can be used to interface orifice plate flow meters. Refer to Table 4-2 for the input channel assignment in case of 1-3 differential pressure transducers use. Transmitter can be wired in loop mode with power supply feed by FloWEB or from an external source. When transmitters are powered by FloWEB the power supply voltage is 24V +/-5% isolated. Input channels can be configured for 4-20mA or 1-5V signal. If digital (HART) transmitters are used, these have to be connected to channel #1.

Figure 4-4: Passive Mode (Transmitter power supply feed by FloWEB)

FloWEB User Manual – Edition 2 Revision 4 Page 23/ 23 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 24 di 24

FloWEB_UM_UNICIG_R24 ENG.doc

Figure 4-5: Active Mode (External Transmitter power supply)

Warning! When interfacing Transmitters/Instruments in Hazardous Area, ATEX Safety Barrier or explosion proof installation MUST be used

4.4.

Connections to Temperature Transducers 4.4.1.

mA type Temperature Transmitter

Loop powered temperature transmitter has to be connected to Input Channel #2. Transmitter can be wired in loop mode with power supply feed by FloWEB or from an external source. When transmitters are powered by FloWEB the power supply voltage is 24V +/-5% isolated. Input channels can be configured for 4-20mA or 1-5V signal. For connection details refer to Fig. 4-4 and Fig. 4-5. If digital (HART) transmitter is used, this has to be connected to channel #1. According to the 2004/22/CE-MID (Measuring Instruments Directive) regarding measuring instruments, for areas where enforceable, the MID homologation foresees the employment of the following temperature transmitter: 

Temperature transmitter 3144P Emerson-Rosemount

In other countries, where MID is not enforceable, an equivalent temperature transmitter can be installed. 4.4.2.

RTD type Temperature Transducer

RTD type temperature transducer has to be connected to Input Channel #6. For connection details refer to Fig. 4-6. The energization current used is 1 mA (typical).

Figure 4-6: RTD Transducer wiring connection. FloWEB User Manual – Edition 2 Revision 4 Page 24/ 24 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 25 di 25

FloWEB_UM_UNICIG_R24 ENG.doc

Warning! When interfacing Transmitters/Instruments in Hazardous Area, ATEX Safety Barrier or explosion proof installation MUST be used

According to the 2004/22/CE-MID (Measuring Instruments Directive) regarding measuring instruments, for areas where enforceable, the MID homologation foresees the employment of one of the following Platinum Temperature Transducer (RTD PT100):   

Pt100 Master Pt100 Corai Pt100 Westeam

In other countries, where MID is not enforceable, an equivalent Platinum Temperature Transducer can be installed.

4.5.

Connections to Pressure Transducers 4.5.1.

mA type Pressure Transmitter

Loop powered pressure transmitter has to be connected to Input Channel #1. Transmitter can be wired in loop mode with power supply feed by FloWEB or from an external source. When transmitters are powered by FloWEB the power supply voltage is 24V +/-5% isolated. Input channels can be configured for 4-20mA or 1-5V signal. For connection details refer to Fig. 4-4 and Fig. 4-5. If digital (HART) transmitter is used, this has to be connected to channel #1. According to the 2004/22/CE-MID (Measuring Instruments Directive) regarding measuring instruments, for areas where enforceable, the MID homologation foresees the employment of one of the following Pressure Transmitter:  

Pressure Transmitter 3051 CA2/CA3/CA4 Emerson Rosemount Pressure Transmitter 3051 TA2/TA3/TA4 Emerson Rosemount

In other countries, where MID is not enforceable, an equivalent Platinum Temperature Transmitter can be installed.

FloWEB User Manual – Edition 2 Revision 4 Page 25/ 25 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 26 di 26

FloWEB_UM_UNICIG_R24 ENG.doc

4.6.

Analogue Inputs Board

Table 4-3 list all the connections to the Analogue Inputs Board (Part Number: SK0707T07M01). Term#

Name

Function

I/O

Term#

Name

Function

I/O

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16

TXD RXD COMS RTS CTS IN1P IN1+ IN1IN3P IN3+ IN3IN5P IN5+ IN5IN7+ IN7-

Comm #4 RS232 TX Comm #4 RS232 RX Comm #4 Common Comm #4 RS232 RTS Comm #4 RS232 CTS Input #1 +24V Out Input #1 In+ Input #1 InInput #3 +24V Out Input #3 In+ Input #3 InInput #5 +24V Out Input #5 In+ Input #5 InInput #7 In+ Input #7 In-

O I

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14 B15 B16

RTX+ RTXCOMS RX+ RXINP2 INP2+ INP2IN4P IN4+ IN4PTF+ PT+ PTPTFP24

Comm #4 RS485/422 TX+ Comm #4 RS485/422 TXComm #4 Common Comm #4 RS485/422 RX+ Comm #4 RS485/422 RXInput #2 +24V Out Input #2 In+ Input #2 InInput #4 +24V Out Input #4 In+ Input #4 InInput #6 PT100 Supply + Input #6 PT100 In+ Input #6 PT100 InInput #6 PT100 Supply +24V Out

I/O I/O

O I O I I O I I O I I I I

I I O I I O I I (O) I I (O) O

Table 4-3: Analog Input Board connections list.

4.7.

Digital Input Connections

The FloWEB Power Supply board has 3 digital inputs for optional UPS module status control. Such inputs can be connected to free voltage contact or open collector, for remote alarming and/or local action through Relay 1. When no external UPS is used, these digital inputs should be wired (“closed”) to common terminal in order to avoid false alarm indications in corresponding ModBus COIL registers and activation of Relay 1 relay. Should they remain open by chance, however, there is no influence on flow meter functions and/or calculations.

Term#

Name

Function

B11 B12 B13 B14l

Low Batt Pwr Fail UPS failure Common

normally closed, open when UPS battery is low normally closed, open when power to UPS fail normally closed, open when UPS fail 0V UPS common return terminal

FloWEB User Manual – Edition 2 Revision 4 Page 26/ 26 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 27 di 27

FloWEB_UM_UNICIG_R24 ENG.doc

5.

FloWEB Output Connections 5.1.

Overview of FloWEB Outputs

This section features connections with external devices that take signals from FloWEB. Use the following list to quickly locate the section of interest: • •

5.1.1 Digital/Pulse Outputs connections : 5.1.2 Analogue Outputs connections :

page 28 page 29

FloWEB User Manual – Edition 2 Revision 4 Page 27/ 27 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 28 di 28

FloWEB_UM_UNICIG_R24 ENG.doc

5.1.1.

Digital/Pulse Outputs Connections

On the Power Supply board are available 2 Digital Outputs, Relay type, with 2 contacts and 8 Digital Outputs, Solid State type, with single contact. The 2 relay outputs have maximum contact ratings of 120 Vac/30 Vdc @ 1 A. The 8 solid state outputs have a maximum contact rating of 60 V @ 0,5 A. Solid state Digital Outputs can be configured by the Operator from the menu; all such outputs can be configured as Pulse Outputs or Digital Outputs (see §7.2.2.4 for details on configuration). The Relays’ coils are kept energized in the normal (not alarmed) state, which is therefore called the INACTIVE state. As a consequence of that, any electric fault condition results in the contacts going in the “alarm” position, which is called the ACTIVE state even if in such state the relays’ coils are de-energized. This means that a board which is not powered has both Relays contacts in the so-called ACTIVE position! Relay Output #1 is activated if is active the UPS Low Battery digital input (when used, or LB input hasn’t been wired while not used) and/or the FloWEB power supply voltage is under 19 Volts. Relay Output #2 starts at the Inactive state, then goes in the Active state when the firmware has been loaded and has finished starting checks. It is inactivated again when FloWEB CPU board is in a fault condition leading to reset, preventing the CPU firmware from keeping the CPU OK signal present.

RLxC RLxNO

Relay

RL1: Active (NC) if Low Input Voltage OR UPS Low Batt detect (Inactive position shown)

RLxNC

RLxC RLxNO

RL2: Active (NC) if main processor and application are running (Active position shown)

RLxNC

Relay

Figure 5-1: Relay Outputs connection

DOCA/B DOX/Y

DOX=Output 1 to 4 DOCA = Common for DO1 to DO4 DOY=Output 5 to 8 DOCB = Common for DO5 to DO8 (Deactivate position shown)

Mos Relay

Figure 5-2: Solid State Outputs wiring connection (MOS)

FloWEB User Manual – Edition 2 Revision 4 Page 28/ 28 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 29 di 29

FloWEB_UM_UNICIG_R24 ENG.doc

5.1.2.

Analogue Outputs Connections

With the optional output board there are 6 Analogue Outputs of 4-20 mA type. All outputs can be configured by the Operator from the menu. The external load can be referred either to the internal, or an external, power supply or can be referred to internal 0 V reference. Fig. 5-3 shows the recommended method for wiring up one analogue output in both situations.

Load referred to Power Supply (+24V Out)

Load referred to Ground (+24V Gnd)

Figure 5-3: Analogues Outputs wiring connections

FloWEB User Manual – Edition 2 Revision 4 Page 29/ 29 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 30 di 30

FloWEB_UM_UNICIG_R24 ENG.doc

5.2.

Power Supply Board

Table 5-1 lists all the connections to the Power Supply Board (Part Number: SK0707T06M01).

Term # Name

Function

I/O

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14

Relays 2 Common Relays 2 Open Relays 2 Close N/C Digital Out 1 Digital Out 2 Digital Out 3 Digital Out 4 Digital Out Common 24V Out 24V Out N/C Power Supply Input+ Power Supply Gnd

O O O

RL2C RL2NO RL2NC N/C DO1 DO2 DO3 DO4 DOCA +24VOut +24VGnd N/C VS+ VS-

O O O O O O O I I

Term#

Name

Function

I/O

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14

RL1C RL1NO RL1NC N/C DO5 DO6 DO7 DO8 DOCB N/C LB PF UPSKO COM

Relays 1 Common Relays 1 Open Relays 1 Close N/C Digital Out 5 Digital Out 6 Digital Out 7 Digital Out 8 Digital Out Common N/C UPS Low Batt Input UPS Power Fail Input UPS Fail UPS Common

O O O O O O O O I I I I

Table 5-1: Power Supply Board connections list

The isolated 24V power supply available on terminals A10 and A11 has a maximum rating of 24V +/-5% @ 600mA; it can be used for external instruments power supply or digital outputs energization. Power Supply Input Voltage for FloWEB power supply is any DC external source with at least 18-36Vdc @ 2A. FloWEB is fuse protected (3.15AT / 250V glass 5x20mm); the fuse can be accessed from the back panel on the power supply board.

FloWEB User Manual – Edition 2 Revision 4 Page 30/ 30 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 31 di 31

FloWEB_UM_UNICIG_R24 ENG.doc

5.3.

Analogue Outputs Board

Table 5-2 lists all the connections to the Analogues Outputs Board (Part Number: SK0707T08M01).

Term # Name

Function

I/O

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14

Analogue Output 1+ Analogue Output 1Analogue Output 2Analogue Output 2+ Analogue Output 3Analogue Output 3Analogue Output 4+ Analogue Output 4Analogue Output 5+ Analogue Output 5Analogue Output 6+ Analogue Output 624V Out 24V Gnd

O

AO1+ AO1AO2+ AO2AO3+ AO3AO4+ AO4AO5+ AO5AO6+ AO6+24Vout +24VGnd

O O O O O O O O O O O O O

Table 5-2: Analogue Outputs Board connections list

FloWEB User Manual – Edition 2 Revision 4 Page 31/ 31 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

6.

Page 32 di 32

FloWEB installation 6.1.

Overview

This chapter will explain FloWEB installation best sequence; it doesn’t go into detail about how to install any peripheral devices (such as transducers, analysis instruments or computer), which are connected to the FloWEB. For this information please refer to the documentation supplied with these items.

6.2.

Safe and Hazardous Zones

FloWEB must be installed in a Safe zone; that means that it cannot be installed in an area where there is the risk of fire or explosion (hazardous zone, which is almost always the case when gases are involved). When FloWEB have to be connected to transducers or instruments that are installed in an hazardous area then safety barriers (ATEX approved) usually have to be wired into the circuit. However some instruments or installations are explosion-proof and barriers aren’t, therefore, needed. Always follow the manufacturers instructions and safety recommendations fully.

6.3.

Installation procedure

For FloWEB installation refer to the following procedure: Step 1: draw a wiring schedule Step 2: unpack the FloWEB Step 3: set the DIP switches Step 4: connect back-up battery Step 5: fit the FloWEB Step 6: make all external connections Step 7: earth the installation Step 8: connect power supply

The following sections will explain the steps in the procedure.

6.3.1.

Step 1: Drawing-up a wiring schedule

Before starting to make any connection to the FloWEB we suggest to draw-up a wiring schedule to help the wiring colors and terminal connections identification. In Appendix A a blank wiring schedule is supplied.

FloWEB User Manual – Edition 2 Revision 4 Page 32/ 32 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 33 di 33

FloWEB_UM_UNICIG_R24 ENG.doc

6.3.2.

Step 2: Unpacking the FloWEB

Remove the FloWEB from its packing and check, with the help of Table 6-1, that all items on the shipping list are present or to see if any items are loose or if it is been damaged in the shipment. If any items are missing or if the FloWEB is damaged, contact Pietro Fiorentini for further advice. Item FloWEB User Manual Panel mounting bracket 3.15AT fuse (this is a spare)

Quantity 1 1 4 1

Table 6-1: Shipping list

6.3.3.

Step 3: Setting the DIP switches

FloWEB has 3 blocks of DIP switches on each Analogue Input Board, as shown in Fig. 6-1, for Input type selection: • • •

SW1 switches: select whether Input #1 and Input #2 is 4-20 mA or 1-5 V SW2 switches: select whether Input #3 and Input #4 is 4-20 mA or 1-5 V SW3 switches: select whether Input #5 is 4-20 mA or 1-5 V, or Input #1 is HART, or Input #3 and Input #4 are Flow Meter

FloWEB has 2 block of DIP switches on each Analogue Input Board, as shown in Fig. 6-1, for serial Comm Port type selection (Comm #4 on Analogue Input board #1, Comm #5 on Analogue Input board #2): • •

SW4 switches: select whether Comm #4 or #5 is RS232/422/485 SW5 switches: select whether Comm #4 or #5 is terminated/biased in RS422/485 configuration

FloWEB is supplied with the DIP switches in these default setting: • • •

Input #1, #2, #4 and #5 Input #3 Comm #4 or #5

: 4-20 mA : Flow Meter : RS232

If any change to these DIP switches setting is necessary, please refer to §8 for Comm Ports settings and to §4.2 for channels configuration.

FloWEB User Manual – Edition 2 Revision 4 Page 33/ 33 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 34 di 34

FloWEB_UM_UNICIG_R24 ENG.doc SW5

SW4

123456 123456

123456 123456 123456

3

SW1

SW2

SW3

Figure 6-1: Analogue Input board Dip switches location FloWEB has 4 blocks of DIP switches on CPU Board, as shown in Fig. 6-2, for serial Comm Port type selection (Comm #1 and Comm #2): • • • •

SW1 switches: select whether Comm SW2 switches: select whether Comm SW3 switches: select whether Comm SW4 switches: select whether Comm

#1 is RS232/422/485 #1 is terminated/biased in RS422/485 configuration #2 is RS232/422/485 #2 is terminated/biased in RS422/485 configuration

FloWEB is supplied with the DIP switches in these default setting: •

Comm #1 and #2

: RS232

If any change to these DIP switches setting is necessary, please refer to §8.

FloWEB User Manual – Edition 2 Revision 4 Page 34/ 34 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 35 di 35

FloWEB_UM_UNICIG_R24 ENG.doc SW4

SW3

SW2

SW1

123456 123456 123456 123456

JP3

JP24

Figure 6-2: CPU board Dip switches location FloWEB has 2 blocks of DIP switches on Auxiliary Board, as shown in Fig. 6-3, for serial Comm Port type selection (Comm #3): • •

SW1 switches: select whether Comm #3 is RS232/422/485 SW2 switches: select whether Comm #3 is terminated/biased in RS422/485 configuration

FloWEB is supplied with the DIP switches in these default setting: •

Comm #3

: RS232

If any change to these DIP switches setting is necessary, please refer to §8. SW1

123456

123456

SW2

Figure 6-3: Auxiliary board Dip switches location

FloWEB User Manual – Edition 2 Revision 4 Page 35/ 35 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 36 di 36

FloWEB_UM_UNICIG_R24 ENG.doc

6.3.4.

Step 4: Connecting back-up battery

FloWEB is shipped with Backup Battery disconnected. Make sure to insert the jumper JP3 (see Fig. 6-4) at this step otherwise date & time and some other important data will be lost in case of power Off. SW4

SW3

SW2

SW1

123456 123456 123456 123456

JP3

JP24

Figure 6-4: CPU board back-up battery JP3 jumper location Back-up battery data maintains the following data: • Date & Hour • TCP/IP configuration • MAC address • User configured parameters in Main Page • COMM ports baud rate and configuration • User passwords • Totalizer values • Redundancy configuration • Display contrast • Reports

FloWEB User Manual – Edition 2 Revision 4 Page 36/ 36 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

6.3.5.

Page 37 di 37

Step 5: Fitting the FloWEB

Firstly, referring to Fig. 6-5 and Fig. 6-6, cut out an aperture in the front panel for each instrument which is to be mounted on it.

Figure 6-5: Minimum dimensions in mm for a panel with aperture to fit one FloWEB

Figure 6-6: Minimum dimensions in mm for a panel with aperture to fit four FloWEB’s Each instrument is mounted with four mounting brackets inserted in the holes on the side of the FloWEB and fixed on the rear of the front panel. 6.3.6.

Step 6: Making FloWEB external connection

1. Refer to the documentation supplied with the external equipments to see if any special procedures have to be carried out when connecting them to the FloWEB. Take special notice of any information about safety requirements in hazardous area and to comply with EMC regulations. 2. FloWEB back panel terminations are composed by a male connector on each boards and a female detachable connector; female connector has screw type terminations and can accept 1mmq wire. 3. Check the wirings thoroughly against the schedule and wiring diagram. 4. Attach all female connectors to the boards male connectors. Refer to Chapter 3 for a detailed table of all back panel terminations.

FloWEB User Manual – Edition 2 Revision 4 Page 37/ 37 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 38 di 38

FloWEB_UM_UNICIG_R24 ENG.doc

6.3.7.

Step 7: Earthing FloWEB

FloWEB chassis must be earthed in all cases; both for safety reasons and to ensure that the installations comply with EMC regulations. Do this by connecting an earth lead with the screw located in one of the four corners on the rear panel to a local good safety earth terminal (see Fig. 6-7).

Figure 6-7: Earthing the instrument chassis

Warning! Incorrect earthing can cause many unpredictable problems, so a good earthing of the chassis and the electronics is a MUST. However the way to do this correctly depends almost entirely on the type of installation done and the conditions under which the system operates. Therefore, because these instructions cannot cover every possible situation, Pietro Fiorentini recommends that earthing procedures should only be carried out by personnel who are skilled in such work.

6.3.8.

Step 8: Connecting power supply

FloWEB is DC powered from any external source in the range 18-36 V with a minimum current of 2,5 A. The external source can be an UPS too; in such case FloWEB offers 3 Digital Inputs to monitor the UPS status. The External power supply positive lead must be connected to terminals A13 of the Power Supply board, while the negative lead must be connected to terminal A14. FloWEB don’t have any power switch so, for maintenance reasons, an external breaker has to be used to power the instrument. After switching on the power, FloWEB loads the Operating System and starts a power on self-test.

FloWEB User Manual – Edition 2 Revision 4 Page 38/ 38 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 39 di 39

FloWEB_UM_UNICIG_R24 ENG.doc

6.4.

FloWEB Marking and Sealing

FloWEB is a metering device compliant with 2004/22/CE-MID (Measuring Instruments Directive). FloWEB is suitable for custody transfer applications in areas where the 2004/22/CE is enforceable.

Following there is a description of FloWEB marking and sealing for two kinds of housing; cabinet and rack. Nameplate: manufacturer, serial number, year of manufacture EC type approval

CE and supplementary metrology marking

Notified Body Identification Code

LEAD SEAL ON PROGRAMMING BUTTON

To prevent any change of the FloWEB parameters, a lead seal is apposed on the prog. button. Stickers seal are apposed between the inputs and CPU boards, as showed in the following picture.

FloWEB User Manual – Edition 2 Revision 4 Page 39/ 39 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 40 di 40

Following there are two examples of housings, cabinet and rack. In both of case the inside components are: 1. Meanwell AD-55B. It is a charger for the inside battery. The batteries are proportionate for power the FloWEB for a minimum of six hours. 2. UPS Module. It is a 24V power supply and circuit diagnostic. 3. Intrinsic Safety Barriers. For the safety isolation, all the input signals from the field are protect through the intrinsic safety barriers. Both connectors, FloWEB side and barriers side, are protect against intentional and unintentional manipulations by covers and customers seals. 4. In alternative three kind of modem is possible to install: ADSL modem, GSM modem and PSTN modem.

FloWEB User Manual – Edition 2 Revision 4 Page 40/ 40 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 41 di 41

FloWEB User Manual – Edition 2 Revision 4 Page 41/ 41 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 42 di 42

Customer Seals (input connectors)

FloWEB User Manual – Edition 2 Revision 4 Page 42/ 42 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 43 di 43

Customer seals (barriers and the internal door)

FloWEB User Manual – Edition 2 Revision 4 Page 43/ 43 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 44 di 44

FloWEB User Manual – Edition 2 Revision 4 Page 44/ 44 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 45 di 45

Customer Seals (input connectors)

FloWEB User Manual – Edition 2 Revision 4 Page 45/ 45 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 46 di 46

FloWEB_UM_UNICIG_R24 ENG.doc

7.

FloWEB User Interface 7.1.

Main Page

When FloWEB is switched on the first time after a reset or if no configuration is made (in this case the label “NO CFG” is visible at top of the page), the Setup Page appears. From This page, when the language label is selected, by pressing ↵ [Ret] key you will be able to change the FloWEB Man Machine Interface language. FloWEB has 2 different languages for the MMI interface: English and Italian. By pressing ↵ [Ret] key the MMI languages will change following this order: ↵ [Ret] English

Italian

↵ [Ret] At bottom of the page is visible the FloWEB CRC, the Serial Number and the Firmware Release. The FloWEB CRC is a CRC 32bit witch is calculated directly on the FloWEB firmware binary code and is reported at bottom of the binary file. When a new firmware release is downloaded on the FloWEB, the CRC is recalculated and is compared with the one reported on the file. If the two values are identical, the new release is saved in the flash memory and will be ready to run after a reboot. The new value of CRC will be displayed. The Serial Number is factory-set, through a boot-loader menu (accessible only when the device is un-sealed, since it requires access to the PROG push button), the Firmware Release is fixed and inextricably linked with the firmware itself, since it is a constant embedded in the binary code and his position is compiler-chosen (i.e. not known to be at a fixed address, and subject to the CRC control as described in extension D document “WELMEC_ext_D_SSWR01.doc”, so that it cannot be tempered without blocking the loading of the application itself). In the FloWEB application (i.e. outside of the boot-loader interface), they are both visible only. Just the Administrator can start the first configuration section.

FloWEB User Manual – Edition 2 Revision 4 Page 46/ 46 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 47 di 47

When the configuration section is terminate (see §7.2), the page that appears is the Main Page; when this page contains more line than can fit in the display, it is possible to scroll up or down between the pages by using the function keys F5 [Prev.] and F6 [Next]. If the instrument has 2 Analogue Input boards, with the keys [1], [2] and [3] is possible to swap the Main Page on STREAM 1, STREAM 2 and STATION when more streams are configured. The label on the Main Page title shows to which stream the displayed data are referred. In the lower line of the display appears the action that can be done using the function keys. Each page is composed by a maximum of 4 row displaying, 4 values per page are shown in the Main Page. Inside each row it’s displayed the abbreviation with the meaning, the value and the engineering unit selected; if the value is outside the limit set in the Alarm Configuration Page, an exclamation mark (!) will be showed near the value. If a OVR is displayed, it means that the value isn’t a live value from the transmitter, but a user manual input value; if a DFT is displayed and the ALM LED glows red, it means that the value isn’t a live value from the transmitter, but a default value (if enabled). The first values displayed are the 4 mains Volume counters.

FloWEB User Manual – Edition 2 Revision 4 Page 47/ 47 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 48 di 48

Scrolling the Main Page the following list of data can be displayed: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. 33. 34. 35. 36.

Vb Base volume = Flow Meter Net Volume at Base conditions Vm Meas. Volume = Flow Meter Gross measured volume Vc Corr. Volume = Flow Meter Gross corrected volume (linearized for turbine error) Vem Meas. Error Volume = Flow Meter Gross measured volume during error conditions Veb base Error Volume = Flow Meter Gross Net volume @base conditions, during error conditions Vec corr. Error Volume = Flow Meter Gross corrected volume @base conditions, during error conditions P Meas. Pressure = Live pressure (gauge or absolute according to pressure transmitter used) T Meas. Temperature = Live Temperature M Mass Counter = measured mass (total) E Energy Counter = measured energy (total) f(Z) Z-Formula = compressibility calculation method used f(Q) Q-Formula = current method Zb base Z = Gas Compressibility at base conditions Z operating Z = Gas Compressibility at operating conditions RHOb Base Density = base density Grb Rel. Density = relative density HVb Heating Value = heating value C Conversion factor = Conversion factor Cf Correction factor = Correction factor (Turbine mode only) Pulse Weight = meter pulse weight (Turbine mode only) Qb Base FlowRate = base volume flow rate Qm Meas. FlowRate = measured volume flow rate M Mass FlowRate = measured mass flow rate E Energy FlowRate = measured energy flow rate DP Press. = Pressure DP value (Orifice mode only) DP1 Pres1 = Pressure DP of transmitter 1 (Orifice mode only) DP2 Pres2 = Pressure DP of transmitter 2 (Orifice mode only) DP3 Pres3 = Pressure DP of transmitter 3 (Orifice mode only) RD Reynolds Number = Reynolds Number (Orifice mode only) BETA Operating Beta = Operating Beta (Orifice mode only) VISC Absolute Viscosity = Absolute Viscosity (Orifice mode only) k Isentropic Exponent = Isentropic Exponent (Orifice mode only) C Discharge Coefficient = Discharge Coefficient (Orifice mode only) E Approach Velocity = Approach Velocity (Orifice mode only) Y Expansion Factor = Expansion Factor (Orifice mode only) ALPHA Alpha Coefficient = Alpha Coefficient (Orifice mode only)

In addition there is a blank page at the end; this last page is composed by four sections: the values to display in each section can be programmed by the user. By pressing the ↵ [Ret] key a scroll down menu will appear where is possible to select which group of information you want show between the following parameters: • • • •

Volumes Measurement (live values only) Calculations Vb Partial

FloWEB User Manual – Edition 2 Revision 4 Page 48/ 48 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 49 di 49

Use the [⇑ ⇑] up and [⇓ ⇓] down arrow keys to move between the menu items. Then press the ↵ [Ret] key to select one. To select the parameter to display, press the [⇓ ⇓] down arrow keys again to highlight the variable name, then press the ↵ [Ret] key to show a list of variables that is possible to select.

Use the [⇑ ⇑] up and [⇓ ⇓] down arrow keys to move within the list; press the ↵ [Ret] key to select one.

FloWEB User Manual – Edition 2 Revision 4 Page 49/ 49 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 50 di 50

Repeat the preceding steps if you want to program the other sections.

7.1.1.

Display Contrast Control

In the main pages the arrow keys [⇐ ⇐] [⇒ ⇒] allows the user to increase or decrease display contrast. Since LCD display contrast is temperature dependent, contrast is automatically compensated according to the working temperature of the instrument. Special care must be paid when installing the FloWEB stacked: allow at least 65,25 mm between two adjacent instruments.

FloWEB User Manual – Edition 2 Revision 4 Page 50/ 50 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.

Page 51 di 51

FloWEB Configuration Menu

From the Main Page, by pressing F2 [Config] key, the Setup Page will be reached; at this point is required an user Administrator password to start a Configuration Section. Push ↵ [Ret] key and select the user, then press ↵ [Ret] key again to confirm the choice. With the [⇓ ⇓] down key move to the Password field, then press ↵ [Ret] key and insert the four digit password and press ↵ [Ret] key to confirm; with the [⇓ ⇓] down key move to the OK field, then press ↵ [Ret] key to complete Login.

NOTE: Access is granted to modification of calculation parameters only if the “PROG” push button has been pressed and the proper level of user rights has been achieved. When the Program Mode is entered by pressing the “PROG” push button, the CFG LED glows yellow. Once the configuration section is terminate, the CFG LED is turned off. In the future, by using a modem for remote access to the FloWEB serial port configured for UNI_CIG protocol, will be possible to change the FloWEB configuration without removing the physical seals. The usage of the UNI_CIG protocol intrinsically fulfills access security requirements by means of user and password identification; upon access is granted all modifications performed are logged in the event register; information recorded are compliant to audit trail specification according to Welmec 7.2 definition. Refer to paragraph 7.3.3 of present document for further details. At the end of the configuration session, a confirmation pop-up will permit to save or discard all changes made during the current session. When asked if modifications should be saved or cancelled, choosing the “Save” button will immediately update the configuration database residing in battery back upped CPU internal RAM. If the system is configured as “Double Line” or Station” a pop-up message will permit to save or discard the same configuration that you made to the other stream except SNAM parameter protocol and the LOG configuration. N.B. This memory cannot be cleared by any means, except by un-sealing the device, extracting the CPU board, and opening the on-board dedicated jumper. FloWEB User Manual – Edition 2 Revision 4 Page 51/ 51 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.1.

Page 52 di 52

Date & Time Setting

When the Configuration Section is open, the first page that appears is the Date & Time page and the label “Config” is showed on top of the page. In the Date & Time page is possible to set FloWEB RTC (Real Time Clock) solar date & time and to program the DST (Daylight Saving Time) configuration. Just the user “Administrator” has the capability to make change to date & time. A remote adjustment of Date & Time (via serial Port or via LAN) can be performed within a maximum of ±10 minutes (± 600 sec). The amount of adjustment is not made in one step, but it is spread for two seconds every minute, so to be effective, the maximum time adjustment of 600 seconds will take 5 hours. The “Time Adjustment” indicates the number of seconds to adjust (± 600 sec) after the first time setting. When the Date & Time are changed, press F3 [OK] to confirm and ESC and then OK to save changes. The Contract Hour (Initial Day Hour) is used to set the contract starting hour (0 = midnight); take note that the contract hour is used to define the end of day time (that when different from zero doesn’t correspond to solar day), so many ModBus registers are updated at the contract hour (previous day, current day, previous month, current month) and the printing are generated at this time too. FloWEB RTC can be synchronized with a master reference clock; if you want to activate such feature, you must set the difference in time between your area and GMT (Greenwich Meridian Time), since usually time server give the time information with reference to GMT. Time synchronization is available through the TCP/IP port. Time server IP address has to be configured in the Net Configuration (see §7.2.9.6) page. Once you have entered the date & time, press F3 [OK] key to confirm the new settings.

If in your area is active the DST, you can enable this function. You can program the month and day that the daylight saving time begins and ends; if you don’t set the day, FloWEB will automatically use as Start Day the last Sunday of the Start Month and the last Sunday of the End Month as End Day. From the Date & Time page by pressing F2 [On/Off] key, you can see FloWEB last ON and/or OFF state date & time and date & time of the last configuration session.

FloWEB User Manual – Edition 2 Revision 4 Page 52/ 52 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 53 di 53

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.2. Engineering Units Configuration After the Date & Time settings you will reach the Engineering Units configuration. All the data can be displayed with different engineering units; three sets of engineering units are available, the user can make a choice between the International System (S.I.), Imperial System (U.S.) and Custom set. The Custom set can be modified, starting from the factory default, by scrolling the parameters using the [⇑ ⇑] up and [⇓ ⇓] down arrow keys and than pressing the ↵ [Ret] key on parameter than has to be changed and selecting the new engineering unit inside the popup list.

Warning! When engineering units are modified all totalizers are resetted

FloWEB User Manual – Edition 2 Revision 4 Page 53/ 53 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.3.

Page 54 di 54

System Configuration

From the Engineering Units configuration, you will reach the System Configuration page, by pressing F1 [OK] With this menu is possible to modify the following configurations: 1. The FloWEB system configuration. 2. The unit settings. 3. The analogue inputs. 4. The Flow Meter inputs. 5. The analogue outputs. 6. The digital outputs. 7. The calculation settings. 8. The alarm configuration. 9. The utility configuration. The utility configuration is composed by: 1. The com port settings. 2. The redundancy settings. 3. The Date & Time settings. 4. The users configuration. 5. The NET configuration. 6. The Log & Report configuration. Within System Configuration page, when the Product Type field is highlighted, by pressing the ↵ [Ret] key, you can select FloWEB configuration.

The configurations that you can select are: 1. Single Line (one Analogue Inputs board installed) 2. Dual Line (two Analogue Inputs board installed) 3. Station (two Analogue Inputs board installed)

FloWEB User Manual – Edition 2 Revision 4 Page 54/ 54 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 55 di 55

FloWEB_UM_UNICIG_R24 ENG.doc

Use the arrow keys ([⇑ ⇑] up and [⇓ ⇓] down), to move between the selections then, by pressing ↵ [Ret] key, confirm the selection. After the Product Type, the Measuring System has to be configured; the selection available are: 1. Turbine 2. Orifice 3. Ultrasonic

Warning! Any change of FloWEB Measuring System will reset all the Totalizers

3

FloWEB will display, by default, the Base Volume with m engineering unit; the user can change the unit prefix 3 3 3 with Nm (Normal cubic meter) or Sm (Standard cubic meter). When the Base Volume unit [N/S] m field is highlighted, by pressing the ↵ [Ret] key, a menu will be displayed, where is possible to choose the volume unit label to display. Use the arrow keys ([⇑ ⇑] up and [⇓ ⇓] down), to move between the selections then, by pressing ↵ [Ret] key, confirm the selection. When a fault condition related to Base Volume calculation is detected (pressure, temperature and/or calculations alarm), by default FloWEB updates all the Error Volume totalize (Vem, Vec & Veb) as long as it’s possible: this means that e.g. in Turbine configuration, Vem & Vec will be updated with counted pulses if present, while Veb will be updated only if also pressure and temperature are available for the calculation. In Orifice configuration, whenever the conditions for their calculation will exist, all of them are updated too. In the main page of FloWEB only Vem is shown; by pressing “Next” button and going in the second page the user can show also Veb and Vec with P and T. For each stream it’s also possible to select the strategy FloWEB will use when a Blade Failure Alarm is detected (turbine system only): 1. Normal Counting 2. Stop Counting When Normal Counting strategy is selected, in case of Blade Failure Alarm, all the counters will be updated. That means Vm and Vc will be calculated from pulses counted on first channel, while Vb is updated only if calculation is possible with valid parameters; otherwise the related Vem, Vec & Veb will be updated in case errors are such, that calculation is still possible but only with “default” parameters, as configured in “On error use Default” options. When Stop Counting strategy is selected, in case of Blade Failure, all calculation will be stopped: that means that pulses counted are ignored, instantaneous volumes and flows are zeroed.

FloWEB User Manual – Edition 2 Revision 4 Page 55/ 55 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.4.

Page 56 di 56

Analogue Inputs Configuration

To access Analogue Inputs Configuration page push F2 [Input] key.

The list of the input signals is compiled automatically according to the FloWEB configuration made in the System Configuration Page (Turbine or Orifice). The Analogue Inputs supported by FloWEB are: 4-20mA input channels: analogue inputs #1, #2, #3, #4, #5; 1-5V input channels: analogue inputs #1, #2, #3, #4, #5; RTD/PT100 input channel #6. Input channel #1 supports “SMART” transducers; such transducers use HART digital protocol to transmit digital values on the 4-20mA current loop. According to the type of transducers wired to the FloWEB some hardware settings can be required; in addition, for each signal, the channel type must be selected between analogue 4-20 mA, analogue 1-5V and PT100. To move between the input signals use the [⇓ ⇓] down key: when the field to configure is highlighted, press ↵ [Ret] key to enter a selection list or to insert the value with the numeric keyboard. When the input type field is highlighted, only the available selections for the selected signal will be showed: make the selection according to the type of transducer wired to the FloWEB. The input channel number will automatically change according to the type selected. When the selection has been made, or the value has been inserted, press ↵ [Ret] key to confirm and use the [⇓ ⇓] down key to move to next field. The minimum & maximum values of the transducer range has to be entered according to the transducer wired to that channel; the values have to be entered using the engineering unit configured in the Calculation Settings Menu (see §7.2.2).

FloWEB User Manual – Edition 2 Revision 4 Page 56/ 56 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 57 di 57

In the Orifice configuration FloWEB calculates the flow rate from a differential pressure measurement. To improve the dynamic in measuring the flow rate, it is possible to use more than one differential pressure transducer (up to three). Differential pressure is acquired by one or more probes (maximum three: low, medium and high). When two or three probes operation is selected, the differential pressure displayed by FloWEB will be the one of the probe active at the moment. The selection of the probe in use is done with the following method: -

When the differential pressure is above the 95% of DP Low Max, the DP Medium probe is used; When the differential pressure is above the 95% of DP Medium Max, the DP High probe is used; When the differential pressure return below the 90% of DP Medium Max, the DP Medium probe is used again; When the differential pressure return below the 90% of DP Low, the DP Low probe is used again.

If the value acquired by a probe is not valid (the measurement is out of range) the upper probe is used. Fig. 7-1 shows graphically how is made the selection of the differential pressure probe.

Figure 7-1: Differential Transducers selection method

FloWEB User Manual – Edition 2 Revision 4 Page 57/ 57 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 58 di 58

FloWEB_UM_UNICIG_R24 ENG.doc

Table 7-2 show the relationship between the number of differential pressure transducers used and the configuration that must be made. # Sensors Configuration 1 DP = DP1 DPL = DP1 2 DPH = DP2 DPL = DP1 DPM = DP2 3 DPH = DP3 Table 7-2: Differential Transducers configuration

7.2.4.1. Hart Inputs Configuration In case the physical input #1 has been configured with dip-switches for Hart protocol, and Hart input type has been selected for one or more input acquisition channel numbers in the Analogue Configuration menu page, it’s then necessary to go in the Hart Configuration Page by pressing the F2 [Hart] key, to set the Hart device address and the number of the register variable to read (up to 4 variables can be read for each SMART field transmitter). Each Hart variable is associated to the Analogue Configuration’s corresponding input channel: for example, if you want to read the Temperature (which is on input channel #2), Hart Device and Register have to be configured on channel number #2, irrespective of the physical wiring of Hart transmitters, which are all connected on the Hart bus on physical board input #1.

In the Device column, for each SMART field transmitter, the corresponding address number has to be inserted (the –1 value disables Hart protocol on an input channel, valid values are 0-15: if more than one SMART transmitter is used, don’t use 0 value), while the Register column number is set to a value between 0 and 3, to extract the corresponding variable value from the set of 4 process variables read from each device. Press the ESC[ESC] key to return to the Analogue Configuration page when finished.

FloWEB User Manual – Edition 2 Revision 4 Page 58/ 58 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 59 di 59

7.2.4.2. Flow Meter Inputs Configuration From the Input Configuration Page, when you press the F3 [Dig.] key the Flow Meter Configuration Page is entered. In this page the number of Flow Meter pick-up used has to be configured; if a single pick-up Flow Meter is connected to the FloWEB then input channel #3 (Ch 0) has to be enabled, while if a double pick-up Flow Meter is used input channel #4 (Ch 1) must be enabled too. For each input channel the Pulse Weight and the frequency type (Low = 2 Hz max., High = 5 KHz max) have to be configured. When a Low Frequency Flow Meter with a mechanical reed contact output is used and configured, then a filter has to be used, to avoid wrong pulse count; FloWEB implements a software filter algorithm with duty-cycle awareness, therefore TWO filter integration times have to be configured, one for input signal in High state and one for input signal in Low state. N.B. The filter parameters are not shown, unless at least one channel is put in Low Frequency mode first. Minimum Pulse Length represents the minimum duration that the incoming pulse must have to be considered valid (default value = 100 mSec). Any pulse shorter than Minimum Pulse Length will not be counted. The parameter Filter represents the time in milliseconds after a valid pulse receiving, during which other pulses will not be valid (default value = 50 mSec). Note: when a low frequency input is used, the Flow Rate calculation is made by integrating the pulses received on a 5 minutes sliding window (see §11.2.3.3.).

If a turbine with 2 pick-up is used (both counting inputs enabled), the parameters needed to enable and configure the Blade Failure detection are made visible (otherwise they are not shown); in such case the pulses from high frequency input are compared with the pulses coming from the reference input, also called monitor. The ratio between the high frequency pulse weight and low frequency pulse weight is defined as Blade Ratio: Blade Ratio BR = High frequency pulse weight (Ch 0) Low frequency pulse weight (Ch 1)

(calculated by FloWEB)

The number of pulses used in BR calculation is the total number of pulses counted in a sliding window comprising the last 10 (ten) seconds. FloWEB User Manual – Edition 2 Revision 4 Page 59/ 59 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 60 di 60

The algorithm is performed and updates the current alarm status only if at least 10 (ten) pulses have been counted in the last ten seconds’ window in the Low frequency channel, and at least (BR * 10) pulses in the High frequency channel. The percentage error to be considered as maximum difference acceptable between the two channels without rising an alarm, can also be set in the “Pulse Error %” field, with up to three decimal digits: the default value is 2.0 %. The Blade Failure alarm is triggered when: (LF x BR) - HF >= (ErrPerc x MaxF) Where: LF = Number of pulses from the low frequency pick-up in a given time. HF = Number of pulses from the high frequency pick-up in the same time. ErrPerc = value set in Pulse Error % field MaxF = higher value between HF and (LF x BR) Once the Blade Failure alarm has been triggered, the alarm condition will remain active until the following test succeeds: (LF x BR) - HF 95 – 105mBar

FloWEB User Manual – Edition 2 Revision 4 Page 77/ 77 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 78 di 78

Note: The Cut-Off low range value is 0.01% but is strictly recommended do not use values below 1%. A popup form is shoved if a value below 1% is inserted.

If the Orifice Diameter is greater than, or equal, to the Pipe Diameter a popup page will appear with the warning for inconsistent data.

FloWEB User Manual – Edition 2 Revision 4 Page 78/ 78 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.8.

Page 79 di 79

Alarms Configuration

By pressing F5 [AlmCfg] key in the System Configuration page a pop-up menu will appear where is possible to select which parameter to configure. In this page the limits on analogues inputs or calculated values can be set.

Use the [⇑ ⇑] up and [⇓ ⇓] down arrow keys to move within the list. Then, press the ↵ [Ret] key on the parameter to configure.

FloWEB User Manual – Edition 2 Revision 4 Page 79/ 79 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 80 di 80

For each parameter up to 7 different alarm limits can be configured; there are 4 working limits (very low, low, high, very high), 2 fault limits (low fault, high fault) and an excursion limit. In addition, it’s possible to configure FloWEB to get an alarm on the speed of change of the analog input (DeltaValue/DeltaTime); the excursion limit value is the maximum difference between the actual and the last value (DeltaValue) in the period (DeltaTime) of analysis (DeltaTime = 1 second for single stream FloWEB, DeltaTime = 2 seconds for double streams FloWEB). The hysteresis is programmed with the aid of 2 additional parameters: Full Range (= full scale value of the sensor in engineering units) and % Scale (=% of Full Range = hysteresis). To set a limit, a value has to be entered and the associated enabled (ON) or disabled (OFF) has to be checked; when in the box next to each limit a check is displayed, it means that the alarm limit has been passed. It’s possible to change the Status from Enabled to Disabled too. N.B. Alarm LED on front panel doesn’t glow red when alarm on measure are triggered.

FloWEB User Manual – Edition 2 Revision 4 Page 80/ 80 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.2.9.

Page 81 di 81

Utility Page

From the System Configuration menu, by pressing F6 [Utilt.] key you can access to the Utility settings page. Here, you will be able to access the following setting pages: 1. 2. 3. 4. 5. 6. 7. 8.

Virtual Printer Communication Port Configuration Redundancy Configuration Date & Time Users Password TCP/IP Network configuration Log & Report Compact Flash Configuration

7.2.9.1. Virtual printer Use this selector to enable the virtual printer feature, it allows you to write in files instead of print it using a real printer. The files are named “VirtP.vp1” and “VirtP.vp1” and they are saved in internal flash in the “VIRTP” directory. The maximum file size is 40 Kbyte, the files are updated until the size become 40 Kbyte then the file is closed and at the same time the other file is opened, eventually cleared and updated.

FloWEB User Manual – Edition 2 Revision 4 Page 81/ 81 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 82 di 82

7.2.9.2. Communication Port Configuration Use the [⇑ ⇑] up and [⇓ ⇓] down keys, to select the COM #, then press ↵ [Ret] key on the selected communication port to enter the configuration. FloWEB can have 4 or 5 communication ports according to hardware configuration (1 or 2 Analog Input Board). Within the Comm Configuration menu, you can set the following communication parameters: • • • • •

Mode: ModBus, Print, GAS Chromatograph, US, SNAM; Baud Rate: from 300 to 38400; Bit: 7 or 8; Parity: None, Odd or Even; Stop Bit: 1 or 2.

Warning!!: For the changes to have effect, after exiting the configuration and confirming to save the changed configuration, the FloWEB must be powered off and then powered on again. The Mode parameter is used to select the device connected to any Comm port: the possibilities are the following: • • • • •

Mode = ModBus: FloWEB is a ModBus Slave device connected to a RTU or PC Mode = Print: when ASCII serial printer is connected to the Comm port Mode = GasCr.: FloWEB is a ModBus Master device connected to a Gas Chromatograph Mode = US: Ultra Sonic. Mode = SNAM: protocol for Italian market.

The communication parameters (Baud rate, number of bit, parity and stop bit) have to be configured to match the corresponding parameters on the connected device. When the Comm port is configured as “ModBus”, in the configuration page additional parameters are displayed, as in the following pictures:

FloWEB User Manual – Edition 2 Revision 4 Page 82/ 82 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

• •



Page 83 di 83

ModBus Address: enter the ModBus Slave ID number that this serial port will respond to (1 through 247 acceptable); Protocol type: default choice is “User”, which gives you the possibility of setting the number of bits per character and the stop bits freely, to accommodate for devices that use ModBus protocol but do not adhere to ModBus standard frame format. For full ModBus standard implementations, select ASCII or RTU options, in correspondence to the ModBus device connected to FloWEB; the “Bit#” and “StopBits” parameters will not be visible, because they will be forced to the ModBus standard frame values, as seen in the following image. Modicon Compatible: select NoModic or Modicon to configure the ModBus port to be compatible with Modicon PLC equipment and DCS systems.

ModBus Mode has to be used when a ModBus Master device, like RTU, PLC or PC, is connected to FloWEB to read and/or write data from the device memory. Typical applications are SCADA supervisory system, data and time synchronization, raw gas composition data update (“GAS Composition” selector has to be set on “Measured”: (see § 7.2.7.1.), etc.

FloWEB User Manual – Edition 2 Revision 4 Page 83/ 83 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 84 di 84

When the Comm port is configured as “Print” like in figure, in the configuration page no additional configuration parameters are displayed.

Attention: currently only one port can be configured as “Print”, whichever you want. The virtual printer is considered as a Comm; therefore it is not allowed to have the virtual printer enabled and a comm. Port configured as “Printer”. Should you want to change the printer port, you must follow the following procedure: • • •

Change the currently setted “Print” port Mode (e.g. “ModBus”); Save the configuration; Configure the new Comm as “Print”

FloWEB User Manual – Edition 2 Revision 4 Page 84/ 84 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 85 di 85

When the Comm port is configured as “GAS Chromatograph”, in the configuration page the following additional parameters are displayed: •

ModBus Address: enter the ModBus address of the Gas Chromatograph;



Brand: select between Daniel 2360, Yamatake or Varian 4900 according to the model of gas chromatograph connected to FloWEB;



Time Out: represents the maximum time in minutes between two gas analysis performed by the Gas Chromatograph; the FloWEB make a request for gas composition data each minute to the chromatograph (check on gas chromatograph user manual for the time necessary to complete the analysis). If there is a fault on the gas chromatograph response (a fault can be: incomplete or corrupted data), the FloWEB try to restore the communication. If the faults persist for the time specified by “Time Out”, there are two actions that the FloWEB can take. 1. The flag “On Error use dft” is set to true. All the data will be substituted by FloWEB with the default value programmed by the user in the Default column (refer to § 7.2.7) and the calculation will continue. 2. The flag “On Error use dft” is set to false. The calculation will be stopped. In both of the case, the “Zeta” alarm is raised, the red led (ALM) on the front panel is lighted and an event is generated.



GAS Comp. Table: this parameter has an effect just for double stream configuration, use “Single” if the GAS composition data are the same for each stream. If the GAS that flow in the two streams is different, the GAS chromatograph have to be configured for make two GAS analysis, in this case, use “Double” in the selector, the FloWEB will read alternatively the GAS composition for a stream 1 and for the stream 2. In this case the FloWEB is expected to receive the GAS composition alternatively for stream 1 and stream 2, if this do not happen, an alarm is risen.

FloWEB User Manual – Edition 2 Revision 4 Page 85/ 85 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 86 di 86

FloWEB_UM_UNICIG_R24 ENG.doc

Yamatake/Varian Flowchart Start

NW = Read 3058 New data flag

NW = 1 Yes AC = Read 3059 Ana/Cal flag

NW = 1

TimeOut = TimeOut -1 New data flag

No

Yes Bit 0° 6°and 8° are mascherate

Yes

Err = Read 3047 Error flag

TimeOut Error Raising

No Err = 0

No TimeOut = 0

GC Error Raising

Yes Read 3001 to 3016 Comp. Index

Read 7001 to 7016 Comp. Value

Read 7012 Heat Value Read 7013 Relative Density Read 7014 Wobbe Index Read 7035 He

Continue End

FloWEB User Manual – Edition 2 Revision 4 Page 86/ 86 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 87 di 87

FloWEB_UM_UNICIG_R24 ENG.doc

Daniel Flochart Start

NW = Read 3058 New data flag

NW = 1

No

Yes AC = Read 3059 Ana/Cal flag

NW = 1 Yes Bit 0° 6°and 8° are mascherate

Err = Read 3046 Error flag

Err = 0

TimeOut = TimeOut -1 New data flag

No

No TimeOut = 0 Yes TimeOut Error Raising

GC Error Raising

Read 3001 to 3016 Comp. Index

Read 7001 to 7016 Comp. Value

Read 7012 Heat Value Read 7013 Relative Density Read 7014 Wobbe Index

Continue End

When the Comm port is configured as “Ultrasonic”, in the configuration page the following additional parameters are displayed: FloWEB User Manual – Edition 2 Revision 4 Page 87/ 87 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 88 di 88



ModBus Address: enter the ModBus address of the Ultrasonic Meter;



Model: If different brand is supported, here is possible to choice the one that is physically connected to the FloWEB;



Ref. Time (min): insert the fixed interval in minutes between requests for data ultrasonic meter data.

FloWEB User Manual – Edition 2 Revision 4 Page 88/ 88 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 89 di 89

When the Comm port is configured as “SNAM”, in the configuration page the following additional parameters are displayed: • • • • •

REMI code: insert the numeric plant code assigned by SNAM authority (Italian Market); Qmax. Limit: Maximum Flow Limit for SNAM remote diagnostic ONLY (Italian Market); Qmin. Limit: Minimum Flow Limit for SNAM remote diagnostic ONLY (Italian Market); Password check: select between Disabled or Enabled to request the access password or not (Italian Market); Link type: select between Direct connection or Modem connection (Italian Market).

REMI code, Qmax Limit and Qmin Limit can by different between stream 1 and stream 2, than if the FloWEB is configured as double stream, the form change and allow the user to insert this parameter respectively for stream 1 and stream 2.

FloWEB User Manual – Edition 2 Revision 4 Page 89/ 89 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 90 di 90

7.2.9.3. Redundancy Configuration From the Utility Configuration page, by pressing F1 [Redund] key, you can access the Hot/Standby Redundancy configuration page. For details on this configuration refer to § 10.

From this page you can enable, or disable, the Hot/Standby, or Redundancy, configuration of two FloWEB interconnected between them. In the “Redund ID” parameter has to be configured the same value for the Master and the Slave FloWEB (default 1); if the Master and the Slave “Redund ID” aren’t the same, redundancy will not operate. When you have to connect the redundant FloWEB flow computers to a supervisory computer and you want that only the promoted Master will communicate, you have to program the “Redund ModBus ID” parameter (this parameter is common to all COMM ports); this is the dedicated ModBus address common to both flow computers; when the supervisory computer will send a ModBus read/write message, only the promoted Master will answer, on all COMM ports. When you configure the redundancy, you have to choose what FloWEB flow computer will be the preferred Master; set the “Redund Master Pref.” check box on the FloWEB that you want to be the preferred Master. Warning!!: When Redundancy configuration is activated one of the two FloWEB MUST be configured as preferred Master. If both or neither FloWEB are configured as preferred Master the Redundancy will not work correctly. The last two parameters in the page are a read only type; “Redund Status” will display the redundancy status of the two flow computers and the “Partner Status” of the other flow computer. Warning!!: To make active any change, FloWEB must be powered off and then powered on again.

FloWEB User Manual – Edition 2 Revision 4 Page 90/ 90 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 91 di 91

FloWEB_UM_UNICIG_R24 ENG.doc

Redundancy Symbols

Redund Disable Status Master

M

Status Slave

S

Initialization

I

Alignment

A

Slave Out Of Service

S

Configuration not aligned

C

After the redundancy initialization, the Master FloWEB will check the Slave configuration for alignment with its own. When Master and Slave flow computers are aligned, the Slave will not accept any configuration change. In fact, it will not be possible to make the Login on the Slave. The two redundant FloWEB have to be configured with the same ModBus ID, but only the promoted Master will answer to the ModBus messages. The Slave FloWEB will have the same inputs of the Master FloWEB; it performs all the calculation but the totals will be send by the Master. In this way, Master and Slave flow computers will be always aligned.

7.2.9.4. Date & Time Settings From the Utility Configuration page, by pressing F2 [Date] key, you can access the Date & Time configuration page. For details on this configuration refer to §7.2.1.

FloWEB User Manual – Edition 2 Revision 4 Page 91/ 91 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 92 di 92

7.2.9.5. Users Login Passwords From Utility Configuration page, by pressing F3 [Users] key, you will access the User Login Password Configuration.

The instrument implements a password system that restricts access to its facilities to those people with certain levels of authority. There are 5 different levels of authority, from the highest to the lowest: • • • • •

Administrator Installer Engineer Maintenance Operator

For each levels there are ten passwords that are possible to choice. The empty password and “0000” are not valid, for default there is just one password “1111” associate to the Administrator, The Administrator that make the login using the first password has the capability to change, his own password, the other Administrator password and the firsts passwords of the other users. Each user that makes the login using the first password has the capability to change, his own password and the other password of the same user.

FloWEB User Manual – Edition 2 Revision 4 Page 92/ 92 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 93 di 93

A user that make a login without using the first password, have just the capability to change his own password and in case the user is the Administrator, he can also change the firsts password of the other levels. Use the [⇑ ⇑] up and [⇓ ⇓] down keys to select the user than press ↵ [Ret] key to confirm the choice.

FloWEB User Manual – Edition 2 Revision 4 Page 93/ 93 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 94 di 94

7.2.9.6. TCP/IP Network Configuration FloWEB has an Ethernet port that supports ModBus and HTTP protocols. From Utility Configuration page, by pressing F4 [Net] key, you will reach Network Configuration Page.

The MAC address is the Ethernet port address and it’s unique for each FloWEB; it’s factory set and is displayed for information only. DHCP active is not yet implemented in the actual software version: must be left to OFF. The following two parameters are the IP address and the Network mask: they have to be configured when FloWEB is connected on an Ethernet network. Select parameter by using the [⇓ ⇓] down key, when the value is highlighted, pressing the ↵ [Ret] key, the field will be erased: using the keyboard insert the IP address or the Network mask composed by four numbers, composed by 3 digit each (i.e. to insert 1 you have to set 001), from 0 to 255 separated by periods, then press the ↵ [Ret] key to confirm. If you will hear a beep, it means that the format of the address is not correct, and the inserted value will be not accepted. When a valid IP address is programmed for the Time server, and the FloWEB is connected to a network, it will try to synchronize its RTC with the Time server every two minutes: when the synchronization has been established, the frequency of sync will be 60 minutes. Warning!!: To make active any change, FloWEB must be powered off and then powered on again. Warning!!: The DST option has to be disabled on the Time server, since FloWEB accept only time difference less than 10 minutes; the DST has to be enabled on FloWEB only (see § 7.2.1.).

FloWEB User Manual – Edition 2 Revision 4 Page 94/ 94 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 95 di 95

7.2.9.7. Log and Report FloWEB can perform data logging (according to API MPMS 21.1) to generate historical records (archives) of parameter data, and can print pre-defined or user-defined reports. The archived data can be exported through file download and retrieved by ModBus protocol devices. The archived data values in logs and reports are the mean values or mean, maximum and minimum values in the chosen log/report period. From the Utility Configuration page, by pressing F5 [LogRep] key a pop-up window will be displayed: use the [⇑ ⇑] up and [⇓ ⇓] down keys to select the Report page or Log page; to confirm the choice press ↵ [Ret] key.

FloWEB User Manual – Edition 2 Revision 4 Page 95/ 95 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 96 di 96

7.2.9.7.1. Report Configuration By selecting Report Select, you will enter the page where up to 3 reports can be configured, for each stream and also for station totals, according to system type as shown below. Each Report can be of any of the 3 different types: Extended (Ex.), Short (Sh.) or User defined (R1, R2 and R3).

According to the system configuration you can select report for the stream 1 (only L1 choice boxes are shown, if the system type is “Single Line”), stream 2 (L2 shown if the system type is “Double Line” or “Station”) or station (ST shown only when the system type is “Station”).

The time span interval of each report can be 15 minute, 1 hour, 1 day or 1 month. If the choice “none” is selected, the corresponding reports are effectively disabled. FloWEB User Manual – Edition 2 Revision 4 Page 96/ 96 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 97 di 97

FloWEB_UM_UNICIG_R24 ENG.doc

The following table lists all the parameters included in the extended and short report types and all the parameters that can be user selectable for Reports R1, R2 and R3.

Parameter Vb Base Volume Vm Gross Volume Vc Flow Meter Corrected Volume Veb Base Volume in Error Condition Vem Meter Volume in Error Condition Vec Meter Corrected Volume in Error Condition Es Energy Volume Ms Mass Volume Ee Energy Volume in Error Condition Me Mass Volume in Error Condition Zb Base Compressibility T Temperature P Pressure DP Differential Pressure Qb Base Flow Rate Qm Gross Flow Rate Qc Meter Corrected Flow Rate Mass Flow Rate Energy Flow Rate N2 CO2 H2S H2O He CH4 C2H6 C3H8 C4H10 N_C4H10 I_C4H10 C5H12 N_C5H12 I_C5H12 C6H14 C7H16 C8H18 C9H20 C10H22 O2 CO H2 Ar

Extended report √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √

Short report √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √ √

R1, R2, R3 User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select User select

FloWEB User Manual – Edition 2 Revision 4 Page 97/ 97 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 98 di 98

To include a parameter in the User defined Report, select the report you want to configure (F1 [Rep.1], F2 [Rep.2] and F3 [Rep.3]), and check the parameters that have to be included in the report.

By pressing F6 [Next] key in the Report Selection page, you will enter the second Report Selection page where up to 3 reports can be configured, for each stream and also for station totals, according to system type. These reports are compliant to the Italian market gas regulation (SNAM reports).

Daily and Monthly Reports (both printed and archived) will be generated at the Contract Hour (see § 7.2.1). Attention: the maximum number of active report is 16.

FloWEB User Manual – Edition 2 Revision 4 Page 98/ 98 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 99 di 99

7.2.9.7.2. Log Configuration FloWEB can perform data logging to generate historical records in log files, which adhere to the MID and Welmec requirements for validity checking against corruption and/or tampering. The archived data can be retrieved and copied locally through the built-in HTTP protocol and web interface using any PC with a web browser.

For each stream and each of the six time intervals (30 seconds, 1 minute, 15 minutes, 1 hour, 1 day and 1 month) the page gives information on the space occupied in the internal on-board disk. Attention: the total space occupied “Total Space” shown should never be allowed to be greater than the reserved space for this function “Reserved Space”. The log for which the “Num. of Files” fields is zero are disabled; to enable such logs, use the [⇑ ⇑] up and [⇓ ⇓] down keys to highlight the corresponding time box and push the ↵ [Ret] button. In the configuration page that opens you have to select a value for the following four choices: • • • • •

Policy Record time Unit Exec. Status Delete Files

FloWEB User Manual – Edition 2 Revision 4 Page 99/ 99 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 100 di 100

In the “Policy” field you can choose between “Circular” and “Fill”: the first choice will cause the older files to be deleted to make space for the new ones, when the space limit has been reached. The second choice will stop logging when full. The space limit for each log is given in the “Space (bytes)” fields of the preceding log select page. The number of samples “Samp. x File” multiplied for the number of files “Num. of Files” gives the total number of records hold in the onboard disk for the selected log. Those information are accessible only for local visualization.

The “Record time” and “Unit” fields together give the total time span of the log: choose in “Unit” fields between Hour, Day and Month, and in the “Record time” how many of that must be kept in the log. Check the “Delete Files” check box to erase the old log files. Push the F1 [LogVar] button to access the variables configuration page. FloWEB User Manual – Edition 2 Revision 4 Page 100/ 100 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 101 di 101

The corresponding number of necessary samples per file and total files needed to keep all that data will be automatically calculated and updated, when the measures to be logged will be selected in the following page.

When first configured, all the choices will be on “DIS” that stands for “Disabled”; therefore no data will be logged in the report, even if configured, until you choose in this page which fields make up the record to be logged for each time period. For each measure or value shown that must be logged, you must select the “ENA” or “STA” option. “ENA” means “Enabled”, and enables the logging of the measure at his right, given as average value in the log time period. For a one-second period only, the average is the same as the instantaneous value, since the FloWEB has a minimum one-second cycle period for calculations. “STA” means “Statistics”, and also logs minimum and maximum values in the period. For the intervals 30 sec. and 1 min a set of variable that can be logged is restricted as show below.

FloWEB User Manual – Edition 2 Revision 4 Page 101/ 101 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 102 di 102

7.2.9.8. Compact Flash Configuration FloWEB has a removable Compact Flash memory. From Utility Configuration page, by pressing F6 [CFlash] key, you will reach the Compact Flash configuration page. In this page is possible to format the compact flash. This operation is allowed only if the compact flash is inserted but inactivated. It is also possible enable the above command in the maintenance mode and than perform the command without broken the program seal.

Attention!!: Format the Compact Flash exclusively using FloWEB.

FloWEB User Manual – Edition 2 Revision 4 Page 102/ 102 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.3.

Page 103 di 103

Visualizations Menu

From the default pages, by pressing F3 [Vis], after the login, you will reach the Visualization Menu. With this menu is possible to view the following information: 1. The FloWEB current configuration. 2. The Diagnostic and Alarms situation. 3. View the event register. 4. View/Modify the current user password. 5. View the current totalizers. 6. View the current GAS composition. 7. View the Ultrasonic information.

The information related to the current configuration that is possible to see, depends on witch user has made the login. Some information could be not visible for certain user profile. The information are logically grouped in the rows, you can use the [⇐ ⇐] left and [⇒ ⇒] right keys to select the information grouped in the row, then you can expand it simply by pressing ↵ [Ret] key when the row is selected. For each configuration parameter is possible to see, the current value and the default value. 7.3.1. Diagnostic and Alarms FloWEB runs real time diagnostic routines to verify proper behavior of all functionalities; whenever a faulty condition is verified the red LED “ALM” on the front panel glows and a specific alarm condition is raised. Detailed diagnostic and alarms information can be reached from the Main Page or from the Visualization menu Main Page with F1 [Alarms] key; according to the stream displayed in the Main Page (Stream 1, Stream 2 or Station) the corresponding alarm page will be displayed:

FloWEB User Manual – Edition 2 Revision 4 Page 103/ 103 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 104 di 104

The type of alarm that are detected and recorded are: Hardware alarm, caused by: •

Analogue Input board failure

Measure limit alarms, caused by one or more of: • • • •

Pressure transducer alarm: value out of range (20mA) Temperature transducer alarm Differential Pressure transducer alarm (Orifice configuration) Differential Pressure Cut-Off (Orifice configuration)

Calculation alarm, caused by: • •

Zeta calculation parameter error (invalid value for gas composition and/or pressure, temperature out of range) Blade failure (Turbine configuration)

Event alarm, caused by: • •

When the event register is filled up for the 90% of its size. When the event register is full

Alarm alarm, caused by: • •

When the alarm register is filled up for the 90% of its size. When the alarm register is full

Compact Flash alarm, caused by: •

When the compact flash card is full

FloWEB User Manual – Edition 2 Revision 4 Page 104/ 104 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 105 di 105

FloWEB_UM_UNICIG_R24 ENG.doc

Whenever an alarm condition is detected the related check box is ticked and in the Main Page on the upper line of the display, after the Time, one or more alarm icons (see Table 7-1) are displayed. The column “Current” provides information on actual alarm conditions, and the column “Historical” logs alarms conditions which may not be active at the time the alarm page is displayed.

Hardware Measure Zeta Table 7-1: Alarm Icons

7.3.2.

Alarms Acknowledgement

Alarms can be acknowledged and reset by means of the F6 [Reset] key, provided that the proper level of access is being granted to the operator, upon login recognition. Alarm conditions that are active when the “Reset” key is pressed cannot be cleared. 7.3.3.

Event Register

From the Visualization Menu, by pressing F2 [Events] key you can access the events register and the alarms register. In this page you can browse for the events of your interest, by insert the “date from” and “date to”, is possible to select all the events in that period of time. You can also choose the kind of event and show all the events of a specific kind that occurs in that specific period of time. Use the F1 [Find] key to confirm the selection (see Figure 7-3) By using the arrow keys [⇑ ⇑] up and [⇓ ⇓] down is possible to select the event grid, one’s the grid is selected using the arrow keys [⇐ ⇐] [⇒ ⇒] allows you to select the particular event, by press the ↵ [Ret] key you can visualize all the information for that specific event. With using the F2 [Del] and F3 [Print], you can respectively delete the all events or print the selected events. N.B. you can delete the events just if the current user is the Administrator. Before clear the event register a backup is made and the new event register start with a registration of the delete event. The percentage on top of the page indicates the amount of space that the events take. When that value reach the 90% an alarm is raised, when the events take a 100% of the space an other alarm is raised and the eventually new event is overwrite to the last one. 7.3.4.

Alarm Register

From the Visualization Menu, by pressing F2 [Events] key you can access the events register and the alarms register.

FloWEB User Manual – Edition 2 Revision 4 Page 105/ 105 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 106 di 106

In this page you can browse for the alarms of your interest, by insert the date from and date to, is possible to select all the alarms in that period of time. You can also choose the kind of alarm and show all the alarms of a specific kind that occurs in that specific period of time. Use the F1 [Find] key to confirm the selection (see Figure 7-3). By using the arrow keys [⇑ ⇑] up and [⇓ ⇓] down is possible to select the alarm grid, one’s the grid is selected using the arrow keys [⇐ ⇐] [⇒ ⇒] allows you to select the particular alarm, by press the ↵ [Ret] key you can visualize all the information for that specific alarm. With using the F2 [Del] and F3 [Print], you can respectively delete the all alarms or print the selected alarms. N.B. you can delete the alarms just if the current user is the Administrator. Before clear the alarm register a backup is made. The percentage on top of the page indicates the amount of space that the alarms take. When that value reach the 90% an alarm is raised, when the alarms take a 100% of the space an other alarm is raised and the new alarm is overwrite to the last one. Reference structure of event/alarm record is reported in the followings: Field Date and Time User Event type Vb – Base Volume Modifications log

Meaning Time stamp of logged event/alarm User who performed the logged action Type of event (Field Alarms, Parameters Modification, Date and time update, Event deletion, Totalizers preset) Base Volume Totalizer on event occurrence List of parameters affected by modification, previous and updated value.

Contents of event registers can be accessed through user interface and in the future, collected via serial port using UNI-CIG protocol. Event/alarm register display is showed in the following example.

Figure 7-3: Events (or Alarm) register display FloWEB User Manual – Edition 2 Revision 4 Page 106/ 106 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.3.5.

Page 107 di 107

Users Login Passwords

From the Visualization Menu, by pressing F3 [Users] key, you will access the User Login Password Configuration. For details on this configuration refer to § 7.2.9.5. 7.3.6.

Totalizers page

From the Visualization Menu, by pressing F4 [Count] key you can access the “Counters Menu”. In the main page all the following totalizers are displayed: • • • • • •

Vb Base volume = Net Volume at Base conditions Vm Meas. volume = Flow Meter Gross Measured volume Vc Corr. volume = Flow Meter corrected volume (linearized for turbine error) Vem Error Meas. volume = Flow Meter Gross Measured volume during error conditions Veb Error Base volume = Flow Meter Net volume during error conditions Vec Error Corr. volume = Flow Meter corrected volume during error conditions

From here pressing F6 [Next] another page with the following totals is displayed: • • • •

Ms Mass Flow = measured mass flow Es Energy Flow = measured energy flow Me Error Mass Flow = measured mass flow during error conditions Ee Error Energy Flow = measured energy flow during error conditions

From the these pages, by pressing F1 [Hour], F2 [Day] or F3 [Month] keys, the related Counters pages will be displayed; for all the counters, the Current and the Previous hour totalizers will be displayed.

FloWEB User Manual – Edition 2 Revision 4 Page 107/ 107 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.3.7.

Page 108 di 108

Current GAS Composition

From the Visualization Menu, by pressing F5 [ComGas] key you can access the gas composition page. This page shows the values of GAS Composition used by FloWEB for calculations. When the “GAS Composition” selector is set on “Local”, the current values correspond to the GAS Composition values set by the user. When FloWEB is connected to a gas chromatography device (or to a RTU, PLC or PC that can supply gas composition data) the “GAS Composition” selector have to be set on “Measured”; FloWEB will regularly request raw gas composition information, the gas calorific value (HV) and the gas specific gravity (defined also as relative density: GRb) value directly from the gas chromatography device. In such case, requests are first made to the chromatograph for some status flag (MODBUS registers) so that some preliminary checks can be performed. In the “Measured” configuration, current values correspond to the GAS Composition values received from the chromatograph, if they are valid; otherwise they correspond to the default values set by the user when the “On Error use dft” selector is enabled. The label between the component value and his unit, tells you if that value is a local value “O”, the default value “D” or measured value “” (empty). The “GC Time Out” is set to true if no GAS composition data is read within a time out time. The “GC Error” is set to true if there is no response from GC. While in normal condition, both the flags are set to false, as represented in the following figure.

FloWEB User Manual – Edition 2 Revision 4 Page 108/ 108 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.3.8.

Page 109 di 109

Compact Flash

From the second page of the Visualization Menu, by pressing F1 [CFlash] key you can access the Compact Flash data. In this page is possible to access at the compact flash status and information; all the information are read only data.

FloWEB User Manual – Edition 2 Revision 4 Page 109/ 109 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.3.9.

Page 110 di 110

Instant Print

From the second page of the Visualization Menu, by pressing F2 [Print] key you can access the Print Selection page. In this page is possible to select and print the current configuration and the current data. The entire system configuration is divided in the followings sections:          

System Configuration Inputs Configuration Calculation Parameters Default Parameters User Parameters Turbine Calibration Curve Analog Out Configuration Digital Out Configuration COM Configuration Redund Configuration

Select one or both the following items print the current volume values:  

Volume Print Print Partial Volume Vb

Select the section you want print and then press “Print” button.

FloWEB User Manual – Edition 2 Revision 4 Page 110/ 110 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 111 di 111

7.3.10. On/Off Date & Time From the second page of the Visualization Menu, by pressing F3 [On/Off] key, you can see FloWEB last ON and/or OFF state date & time and date & time of the last configuration session.

7.3.11. Ultrasonic information From the second page of the Visualization Menu, by pressing F4 [Ultra] key you can access the Ultrasonic Data. When the Measure System is selected as “Ultrasonic”, you can see in this pages all the information related to the ultrasonic gas flow meter.

The first field indicates the status of the ModBus communication with the ultrasonic system. FloWEB User Manual – Edition 2 Revision 4 Page 111/ 111 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.4.

Page 112 di 112

Maintenance Settings Menu

From the default pages, by pressing F4 [Maint], after the login, you will reach the Maintenance Settings Menu. Only the “Maint.” user has the capability to do that, when the FloWEB is in the maintenance state, it can be visible by the label “MAINT” on top of the page. In this state: 1. The alarms are inhibited. 2. There is no update of the events register. 3. The FloWEB continue to make is calculations but the results are saved with maintenance mark state.

Press F1 [LogOut] to close a Maintenance section, F2 [Date], F3 [User], F4 [Lang.] respectively for change date, user password, and language.

7.4.1.

Users Login Passwords

From the Maintenance Menu, by pressing F3 [Users] key, you will access the User Login Password Configuration. For details on this configuration refer to § 7.2.9.5.

FloWEB User Manual – Edition 2 Revision 4 Page 112/ 112 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.4.2.

Page 113 di 113

Compact Flash servicing

From Maintenance Settings menu, by pressing F5 [CFlash] key the Compact Flash servicing page will be displayed; from this page operations required for the extraction and/or insertion of a Compact Flash Card in the dedicated Card Slot can be performed. On the Compact Flash Card, when properly formatted and inserted using the procedures shown in the following pages, two separate trees of sub-directories are maintained, one for long-time archiving of log files and the other for archiving of the “virtual print rolls” generated by the virtual printer.

When a Flash Card is already inserted in the Slot, it is necessary to inform the system of the intention of extracting it in order to guarantee that all operations eventually active, either for log files backup or virtual printer rolls backup are properly terminated. Warning: failure to properly extract the CF-Card can lead to data corruption, file-system corruption and in extreme cases damage to the Compact Card itself. To remove the CF Card you must make a “Removal” request, this command will stop all the activity on the CF Card and than put the Operating state in “Inactivated”. Reestablishing the activity of the CF Card, after physical insertion you must press the “Insertion” button, the operating state will begun “Active”. Finally, the “Removal” request may be used to disable the long-time archiving, in fact the “Removal” request stops all the activity on the CF Card. N.B. When the “Removal” request has been fulfilled, even if the CF Card is left in the slot, no archiving will be done. To enable long-time archiving, an “Insertion” operation must be done. With the last command “CF Format”, the CF Card can be formatted. This command is available only it has been enabled during configuration (see § 7.2.9.8.). By pressing F1 [CFInfo] key the CF Dataset page will be displayed.

FloWEB User Manual – Edition 2 Revision 4 Page 113/ 113 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

7.4.3.

Page 114 di 114

Test

From Maintenance Settings menu, by pressing F6 [Test] key the Test page will be displayed; from this page some diagnostic test of FloWEB can be performed: • • • •

Video test: checks display regular operation; Kbd test: checks all keyboard keys regular operation; Revert: allows to restore factory configuration; GoBoot: starts FloWEB in Floboot mode;

FloWEB User Manual – Edition 2 Revision 4 Page 114/ 114 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 115 di 115

FloWEB_UM_UNICIG_R24 ENG.doc

8.

Communication and Networking

This chapter will guide the user to FloWEB serial communications and networking. Since this subject area is vast, only FloWEB point of view will be covered. Therefore, it is assumed that the reader has a reasonable working knowledge of data communication and networking.

8.1.

FloWEB Communication Capabilities

FloWEB has extensive facilities that allow it to communicate with almost any device that has support for ModBus protocol. According to FloWEB hardware, configuration of 4 or 5 serial communication ports will be available, other than an high speed Ethernet port and a dedicated CAN bus port. Table 8-1 shows there are 3 physical layer interfaces supported by FloWEB: RS232C, RS422 and RS485 (2 or 4 wires). Comm #1 (CPU board)

Comm #2 (CPU board)

Comm #3 (Aux board)

RS-232C or RS-422 or RS-485

RS-232C or RS-422 or RS-485

RS-232C or RS-422 or RS-485

Comm #4 (Analog Input board #1) RS-232C or RS-422 or RS485

Comm #5 (Analog Input board #2) RS-232C or RS-422 or RS-485

Table 8-1: Serial Ports and supported interfaces

8.1.1.

Comm #1 & Comm #2 connections and settings (CPU Board)

Table 8-2 lists all the connections to the CPU Board (Part Number: SK0707T03M01). Term#

Name

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14

TXD1 RXD1 COMS1 RTS1 CTS1 TXD2 RXD2 COMS2 RTS2 CTS2 CANH CANL COMC N/C

Function Comm #1 RS232 TX Comm #1 RS232 RX Comm #1 Common Comm #1 RS232 RTS Comm #1 RS232 CTS Comm #2 RS232 TX Comm #2 RS232 RX Comm #2 Common Comm #2 RS232 RTS Comm #2 RS232 CTS CanBus H CanBus L CanBus Common N/C

I/O

Term#

Name

Function

I/O

O I

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14

RTX1+ RTX1COMS1 RX1+ RX1RTX2+ RTX2COMS2 RX2+ RX2CANH CANL COMC N/C

Comm #1 RS485/422 TX+ Comm #1 RS485/422 TXComm #1 Common Comm #1 RS485/422 RX+ Comm #1 RS485/422 RXComm #2 RS485/422 TX+ Comm #2 RS485/422 TXComm #2 Common Comm #2 RS485/422 RX+ Comm #2 RS485/422 RXCanBus H CanBus L CanBus Common

I/O I/O

O I O I O I I/O I/O

N/C

Table 8-2: CPU board connections list

FloWEB User Manual – Edition 2 Revision 4 Page 115/ 115 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

I I I/O I/O I I I/O I/O

FloWEB User Manual

Page 116 di 116

FloWEB_UM_UNICIG_R24 ENG.doc

Table 8-3 gives the connections according to the physical interface: Serial Type

RS-232

RS-422 RS-485 (4 wires)

RS-485 (2 wires)

Terminals Comm#1 Comm#2 A1 A6 A2 A7 A3 A8 A4 A9 A5 A10 B1 B6 B2 B7 B3 B8 B4 B9 B5 B10 B1 B6 B2 B7 B3 B8

Signal TX RX COM RTS CTS TX+ TXCOM RX+ RXTX+ TXCOM

Function TX (O) RX (I) Common Request To Send (O) Clear To Send (I) TX not inverting (O) (A) TX inverting (O) (B) Common RX not inverting (I) (A) RX inverting (I) (B) RX/TX not inverting (A) RX/TX inverting (B) Common

Table 8-3: CPU board RS-232, RS-422 and RS-485 connections list

Fig. 8-1 shows the 4 DIP switches banks position on the CPU board; the settings of these DIP switches defines the physical interfaces of the Comm #1 and Comm #2 serial ports. SW4

SW3

SW2

SW1

123456 123456 123456 123456

JP3

JP24

Figure 8-1: CPU board Dip switches location

FloWEB User Manual – Edition 2 Revision 4 Page 116/ 116 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 117 di 117

FloWEB_UM_UNICIG_R24 ENG.doc

Table 8-4 gives the DIP switches settings for the four different physical interfaces of each serial port:

Serial Type

SW1 (Comm #1) or SW3 (Comm #2) 1 2 3 4 5 6 ● ○ ○ ○ ○ ○ ○ ○ ● ○ ● ○ ○ ● ○ ○ ○ ● ○ ○ ● ● ○ ○

RS232 RS422 RS485 (2 wires) RS485 (4 wires)

●= ON, ○=OFF Table 8-4: CPU board RS-232, RS-422 and RS-485 DIP switches settings

Table 8-5 gives the DIP switches settings for the insertion of the termination resistor for RS-422 and RS-485 and for the insertion of the polarization for RS-485:

Serial Type RS422 RS485 (2 wires)

RS485 (4 wires)

Termination / Polarization Termination Termination Biasing Termination Biasing

Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted

SW2 (serial #1) or SW4 (serial #2) 1 2 3 4 5 6 ● ● ○ ○ ● ○ ● ● ○ ○ ● ● ○ ○ ● ● ● ● ○ ○ ○ ○

●=ON, ○=OFF, (void)=don’t care Table 8-5: CPU board RS-422 and RS-485 termination and polarization DIP switches settings

Termination resistors have to be inserted on the first and on the last devices on the network; terminations reduce voltage reflections that can cause the RS-485 receiver to misread logic levels. The Biasing resistor is useful to hold the RS-485 line in a known state when all the drivers in the FloWEB are disabled; the Biasing resistor has to be inserted on the first device of the network only.

FloWEB User Manual – Edition 2 Revision 4 Page 117/ 117 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 118 di 118

FloWEB_UM_UNICIG_R24 ENG.doc

8.1.2.

Comm #3 connection and setting (Auxiliary Board)

Table 8-6 lists all the connections to the Auxiliary Board (Part Number: SK0707T08M01). Term # Name

Function

I/O

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14

Comm #3 RS232 TX Comm #3 RS232 RX Comm #3 Common Comm #3 RS232 RTS Comm # 3 RS232 CTS Comm #3 RS485/422 TX+ Comm #3 RS485/422 TXComm #3 Common Comm #3 RS485/422 RX+ Comm #3 RS485/422 RXN/C N/C N/C N/C

O I

TXD3 RXD3 COMS3 RTS3 CTS3 RTX3+ RTX3COMS3 RX3+ RX3N/C N/C N/C N/C

O I I/O I/O I I

Table 8-6: Auxiliary board connections list Table 8-7 gives the connections according to the physical interface: Serial Type

RS232

RS422 RS485 (4 wires)

RS485 (2 wires)

Terminal A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A6 A7 A8

Signal TX RX COM RTS CTS TX+ TXCOM RX+ RXTX+ TXCOM

Function TX (O) RX (I) Common Request To Send (O) Clear To Send (I) TX not inverting (O) (A) TX inverting (O) (B) Common RX not inverting (I) (A) RX inverting (I) (B) RX/TX not inverting (A) RX/TX inverting (B) Common

Table 8-7: Auxiliary board RS-232, RS-422 and RS-485 connections list

Fig. 8-2 shows the 2 DIP switches banks position on the Auxiliary board; the settings of these DIP switches define the physical interfaces of the Comm #3 serial port.

FloWEB User Manual – Edition 2 Revision 4 Page 118/ 118 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 119 di 119

FloWEB_UM_UNICIG_R24 ENG.doc

123456

SW1

123456

SW2

Figure 8-2: Auxiliary board Dip switches location

Table 8-8 gives the DIP switches settings for the four different physical interfaces of the serial port: Serial Type

SW1 1 ● ○ ○ ○

RS232 RS422 RS485 (2 wires) RS485 (4 wires)

2 ○ ○ ● ○

3 ○ ● ○ ●

4 ○ ○ ○ ●

5 ○ ● ○ ○

6 ○ ○ ● ○

●=ON, ○=OFF Table 8-8: Auxiliary board RS-232, RS-422 and RS-485 DIP switches settings Table 8-9 gives the DIP switches settings for the insertion of the termination resistor for RS-422 and RS-485 and for the insertion of the polarization for RS-485: Interface RS422 RS485 (2 Wires)

RS485 (4 wires)

Termination / Polarization Termination Termination Polarization Termination Polarization

Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted

SW2 1 ● ○ ● ○

2

3

● ○

● ○

● ○

4 ● ○

5

6

● ○

● ○

● ○ ● ○

● ○

●=ON, ○=OFF, (void)=don’t care Table 8-9: Auxiliary board RS-422 and RS-485 termination and polarization DIP switches settings Termination resistors have to be inserted on the first and on the last devices on the network; terminations reduce voltage reflections that can cause the RS-485 receiver to misread logic levels. The Biasing resistor is useful to hold the RS-485 line in a known state when all the drivers in the FloWEB are disabled; the Biasing resistor has to be inserted on the first device of the network only.

FloWEB User Manual – Edition 2 Revision 4 Page 119/ 119 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 120 di 120

FloWEB_UM_UNICIG_R24 ENG.doc

8.1.3.

Comm #4 & Comm #5 connections and settings (Analogue Input board)

Table 8-10 lists all the connections to the Analogue Inputs Board (Part Number: SK0707T07M01). Term#

Name

Function

I/O

Term#

Name

Function

I/O

A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16

TXD RXD COMS RTS CTS IN1P IN1+ IN1IN3P IN3+ IN3IN5P IN5+ IN5IN7+ IN7-

Comm #4/5 RS232 TX Comm #4/5 RS232 RX Comm #4/5 Common Comm #4/5 RS232 RTS Comm #4/5 RS232 CTS Input #1 +24V out Input #1 In+ Input #1 InInput #3 +24V out Input #3 In+ Input #3 InInput #5 +24V out Input #5 In+ Input #5 InInput #7 In+ Input #7 In-

O I

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14 B15 B16

RTX+ RTXCOMS RX+ RXINP2 INP2+ INP2IN4P IN4+ IN4PTF+ PT+ PTPTFP24

Comm #4/5 RS485/422 RTX+ Comm #4/5 RS485/422 RTXComm #4/5 Common Comm #4/5 RS485/422 RX+ Comm #4/5 RS485/422 RXInput #2 +24V out Input #2 In+ Input #2 InInput #4 +24V out Input #4 In+ Input #4 InInput #6 PT100 Supply + Input #6 PT100 In+ Input #6 PT100 InInput #6 PT100 Supply +24V Out

I/O I/O

O I O I I O I I O I I I I

I I O I I O I I (O) I I (O) O

Table 8-10: Analogue Input Board connections list. Table 8-11 gives the connections according to the physical interface: Serial Type

RS232

RS422 RS485 (4 wires)

RS485 (2 wires)

Terminal A1 A2 A3 A4 A5 B1 B2 B3 B4 B5 B1 B2 B3

Signal TXD RXD COM RTS CTS TX+ TXCOM RX+ RXRTX+ RTXCOM

Function TX (O) RX (I) Common Request To Send (O) Clear To Send (I) not inverting TX (O) (A) inverting TX (O) (B) Common not inverting RX (I) (A) inverting RX (I) (B) RX/TX not inverting (A) RX/TX inverting (B) Common

Table 8-11: Analogue Input board RS-232, RS-422 and RS-485 connections list

FloWEB User Manual – Edition 2 Revision 4 Page 120/ 120 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 121 di 121

FloWEB_UM_UNICIG_R24 ENG.doc

Fig. 8-3 shows the 2 DIP switches banks position on the Analogue Input boards; the settings of these DIP switches define the physical interfaces of the Comm #4 (for Analogue Input board #1) and of Comm #5 (for Analogue Input board #2) serial ports. SW5

SW4

123456 123456

123456 123456 123456

3

SW1

SW2

SW3

Figure 8-3 – Analogue Input Board Dip switches location

Table 8-12 gives the connections according to the physical interface: Interface type RS232 RS422 RS485 (2 wires) RS485 (4 wires)

SW4 1 ● ○ ○ ○

2 ○ ○ ● ○

3 ○ ● ○ ●

4 ○ ○ ○ ●

5 ○ ● ○ ○

6 ○ ○ ● ○

● = ON, ○ = OFF, (void)=don’t care Table 8-12: Analogue Input board RS-232, RS-422 and RS-485 DIP switches settings

FloWEB User Manual – Edition 2 Revision 4 Page 121/ 121 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 122 di 122

FloWEB_UM_UNICIG_R24 ENG.doc

Table 8-13 gives the DIP switches settings for the insertion of the termination resistor for RS-422 and RS-485 and for the insertion of the polarization for RS-485: Serial Type RS422

Termination / Polarization Termination Termination

RS485 (2 wires)

Biasing Termination

RS485 (4 wires)

Biasing

Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted Inserted Not Inserted

SW5 1 ● ○ ● ○

2

3

● ○

● ○

● ○

4 ● ○

5

6

● ○

● ○

● ○ ● ○

● ○

●=ON, ○=OFF, (void)= don’t care Table 8-13: Analogue Input board RS-422/RS-485 termination and polarization DIP switches settings Termination resistors have to be inserted on the first and on the last devices on the network; terminations reduce voltage reflections that can cause the RS-485 receiver to misread logic levels. The Biasing resistor is useful to hold the RS-485 line in a known state when all the drivers in the FloWEB are disabled; the Biasing resistor has to be inserted on the first device of the network only.

8.1.4.

Ethernet Connection

FloWEB has an high speed Ethernet port on the CPU board; such port can be connected on the rear panel (directly on CPU board Ethernet receptacle) or on the front panel (by using a short Ethernet cable that bridges CPU board and Auxiliary board receptacle). Fig. 8-4 shows both connections. The Ethernet cable to use for the connection to this port is a crossed type for point-to-point connection and standard type for network connection.

Rear Ethernet Connection

Bridge for Front Panel Ethernet Connection

Figure 8-4: Ethernet port rear panel connection and front panel bridge cable

FloWEB User Manual – Edition 2 Revision 4 Page 122/ 122 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 123 di 123

FloWEB_UM_UNICIG_R24 ENG.doc

8.1.5.

CAN Bus Connection

FloWEB has a high speed CAN (Controller Area Network) communication port on the CPU board; this communication port is used when redundant FloWEB configuration is required. Table 8-14 gives the connection of CAN interface. All connections are duplicated on terminal A and B line to allow an easy FloWEB chain build-up.

Port

Terminal A11/B11 A12/B12 A13/B13

CAN

Signal CAN H CAN L CAN COM

Function H signal (I/O) L signal (I/O) Common

Table 8-14: CPU board CAN connection list

The CAN network requires a bus termination placed on the first and last FloWEB on the chain; on the CPU board JP24, when closed, insert bus termination (see Fig. 8-5).

SW4

SW3

SW2

SW1

123456 123456 123456 123456

JP3

JP24

Figure 8-5: CAN termination jumper

FloWEB User Manual – Edition 2 Revision 4 Page 123/ 123 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

8.2.

Page 124 di 124

FloWEB ModBus Communication

The FloWEB has extensive facilities that allow it to communicate with almost any device that has support for Gould Modicon ModBus protocol. FloWEB supports ModBus protocol on four physical layer interfaces: • • • •

RS232C RS422 RS485 TCP/IP

FloWEB implements the two standard modes of transmission: ASCII (7 data bit, parity odd, 2 stop bit) or RTU (Remote Terminal Unit, 8 data bit, parity none, 1 stop bit). Additionally, a non-standard mode (called “User Mode”) is implemented, that permits to choose each physical transmission parameter independently from the others (user-defined byte-frame). Usually, FloWEB acts as a Slave device on a ModBus network; the Slave address is user programmable. When connected directly to a gas chromatograph or ultrasonic meter, FloWEB operates as a Master device in the ModBus network. The Gould Modicon ModBus specification is designed to transfer data in 1-bit (coil) or 16-bit (register) blocks. This protocol has not been designed to deal with data, which require a minimum of 32-bit blocks, such as floating point numbers. For this reason, every manufacturer of computer equipments, which deals with this type of data, must decide in which way the protocol should be extended. As a result, many different implementations exist for the transfer of 32-bit floating-point data. In the following the main features of the ModBus communication will be described; refer to the document FloWEB ModBus – ModBus TCP Protocol manual for more detailed information. 8.2.1. Supported Commands The FloWEB supports the following ModBus protocol commands: 1. 2. 3. 4. 5. 6. 7. 8.

Command 01 – Read multiple Boolean points Command 02 – Read single Boolean point Command 03 – Read multiple registers Command 04 – Read single register Command 05 – Write single Boolean point Command 06 – Write single register Command 15 – Write multiple Boolean point Command 16 – Write multiple registers

All data stored within the FloWEB is represented by one or more 16-bit registers. There may be registers which contain a collection of bits.

8.2.2. Register Addressing The database in the FloWEB uses a unique index, called location number, for every software parameter. ModBus register numbers are always expressed as database location number minus 1. Therefore, a master device will ask for ModBus register 1519 in order to read the data in FloWEB database location 1520.

FloWEB User Manual – Edition 2 Revision 4 Page 124/ 124 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 125 di 125

8.2.3. Floating Point Numbers The FloWEB uses the 32-bit IEEE format for floating point numbers. Since Gould Modicon doesn’t define 32bit transfers, the order of transmission for (16-bit) words of 32-bit value is also not defined. When FloWEB send 32-bit data, the first word is the most significant.

8.2.4. Connecting FloWEB to a RS-232/RS-422/RS-485 Link A simple ModBus network can consist of just two devices. They could be a Personal Computer (PC), a Gas Chromatograph or an ultrasonic meter, and a FloWEB connected by a RS-232C “straight through” cable.

PC Com Port connection to FloWEB COMM #1, COMM #3, COMM #4 and COMM #5 RS-232 wiring with no RTS/CTS handshaking

PC Com Port connection to FloWEB COMM #2 RS-232 wiring with no RTS/CTS handshaking

FloWEB User Manual – Edition 2 Revision 4 Page 125/ 125 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 126 di 126

PC Com Port connection to FloWEB COMM #1, COMM #3, COMM #4 and COMM #5 RS-232 wiring with RTS/CTS handshaking

PC Com Port connection to FloWEB COMM #2 RS-232 wiring with RTS/CTS handshaking

Larger and more intricate networks are possible. For example a SCADA system may want to get flow rate from each FloWEB dedicated to a stream. In such cases a RS-422/RS-485 connection is the best choice.

Gas Chromatograph connection to two FloWEB COMM #1, COMM #4, and COMM #5 RS-485 wiring

FloWEB User Manual – Edition 2 Revision 4 Page 126/ 126 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 127 di 127

Gas Chromatograph connection to two FloWEB COMM #2 and COMM #3 RS-485 wiring

Gas Chromatograph connection to two FloWEB COMM #1, COMM #4, and COMM #5 RS-422 wiring

Gas Chromatograph connection to two FloWEB COMM #2 and COMM #3 RS-422 wiring After all the wiring to external communication devices is completed, follow the instructions on §7.2.9.2 to configure the communication port.

FloWEB User Manual – Edition 2 Revision 4 Page 127/ 127 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

8.2.5.

Page 128 di 128

Connecting FloWEB to a TCP/IP network

FloWEB implements ModBus protocol on the TCP/IP protocol too, so it can be connected to any TCP/IP network. A ModBus messaging service provides a Client/Server communication between devices connected on the network. A ModBus TCP/IP network usually consist of PLC (Programmable Logic Controller), RTU (Remote Terminal Unit), Gas chromatograph and SCADA supervisory System than can exchange or retrieve data from FloWEB. To configure the TCP/IP port follows the instructions on § 7.2.9.6. The ModBus TCP implementation follows the guidelines of the Modicon/Schneider ModBus TCP/IP specifications, except for the “Gateway” service, which isn’t applicable to the FloWEB since it is not meant to be used as a gateway device. A maximum of 5 (five) simultaneous ModBus TCP connections are possible; the FloWEB acts as a server on fixed TCP port 502 for this protocol. N.B. The ModBus TCP protocol is not to be confused with the ModBus Ethernet protocol: both protocols use Ethernet interfaces as a physical transport medium, but they are different and incompatible. The so-called “Industrial Ethernet” protocol is still another, different and incompatible too.

FloWEB User Manual – Edition 2 Revision 4 Page 128/ 128 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

9.

Page 129 di 129

HART, Smart and FloWEB

FloWEB has a native facility to use digital communications with “Smart” field transmitters; no special add-on board is required. A “SMART” transmitter is said to be intelligent because it contains a microprocessor that provides extra functionality. This may take various forms, such as on-board calculations, handling multiple sensors, combining types of measurement, measurement integrity indicators, and so on. “SMART” is also used for the ability to re-use existing field wiring.

9.1.

Introduction to HART, Smart and the FloWEB

FloWEB is able to communicate as a Current Input Device (Primary Master) on a network (two-wire 4-20mA loop) of “SMART” field transmitters (Slaves) for each Analogue Inputs board.

Warning!!

Each network loop must have no more than five “SMART” field transmitters connected at any one time. Exceeding this number will damage Analogue Inputs boards.

Fig. 9-1 shows a “safe area only” diagram with all the HART network loops with the maximum number of “Smart” field transmitters. In practice, far fewer transmitters are used.

Figure 9-1: Smart field transmitters

Warning!!

Connecting up “SMART” transducers has to be done with great care. Powering-up more than one point-to-point configured transmitter on a HART network loop can produce an electrical current (20mA per transmitter) that can damage the FloWEB Analogue Inputs board.

The communications standard for each network loop is the HART (Highway Addressable Remote Transducer) Protocol. A full technical discussion of this standard is outside the scope of this User Manual. FloWEB application software is able to request data from dynamic variables that are kept and maintained by a “SMART” transmitter. These dynamic variables can be thought of as being very much like virtual analogue inputs for FloWEB. Four dynamic variables per “SMART” field transmitter can be requested. A total of twenty dynamic variables can be input to the FloWEB. FloWEB User Manual – Edition 2 Revision 4 Page 129/ 129 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 130 di 130

FloWEB_UM_UNICIG_R24 ENG.doc

Fig. 9-2 gives an example of an HART network loop with four field transmitters; from the HART transmitter with address 1 two variables are retrieved, primary and secondary, and assigned to FloWEB inputs 1 and 2 respectively. From the HART transmitter with address 5 the fourth variable is retrieved and assigned to FloWEB input 5, from the HART transmitter with address 3 the secondary variable is retrieved and assigned to FloWEB input 4 and from the HART transmitter with address 8 the third variable is retrieved and assigned to FloWEB input 9.

HART INPUTS HART 1

HART 2

HART 3

HART Transmitter

HART Transmitter

Variables

Variables

Primary

Primary

Secondary

Secondary

Third

Third

Fourth

Fourth

HART 4 Address = 1 HART 5 HART loop HART 6

HART Transmitter

HART Transmitter

Variables

Variables

Primary

Primary

Secondary

Secondary

Third

Third

Fourth

Fourth

Figure 9-2: Example of HART network loop with four transmitters HART 7

HART 8

HART 9

HART !0

Address = 3

Address = 8

FloWEB User Manual – Edition 2 Revision 4 Page 130/ 130 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

10.

Page 131 di 131

Redundancy and FloWEB

Redundancy involves using two identically equipped FloWEB flow computers and connecting them in such a way to ensure uninterrupted measurement in the event of failure of one of the units. This requires that all input signals are connected to both computers. During normal operation, one computer is designated the Master (or Hot) and the other computer the Slave (or Standby). To ensure synchronization between both devices, important variables such as totals, date and time must be transmitted from the Master flow computer via the peer-to-peer link to the Slave flow computer. Should a failure of the Master flow computer occur, the Slave flow computer is automatically promoted to Master and assumes all measurement functions. In this case the data flow on the peer-to-peer link reverses automatically and the new master begins to transmit critical data to the slave, assuming that it is functioning.

10.1. Introduction to Hot/Standby, Redundancy and the FloWEB Redundancy (or Hot/Standby) controls the selection of a flow computer that is to be the main fiscal accounting device (‘Master FloWEB’) and a flow computer that is to be the backup device (‘Slave FloWEB’). In the event of any system alarm, such as those in Table 10-1, a hand-over will take place. On completion of the hand-over, the ‘Slave FloWEB’ becomes the ‘Master FloWEB’.

Power fail Blade failure RTD out of range Z calculation error mA sensors out of range Master CPU error Table 10-1: System alarms for flow computers hand-over

The main feature of Redundancy is to provide synchronization of fiscal totals (Vb, Vm, Vc, Ve, Es and Ms) between the two FloWEB. Communications between the two flow computers is accomplished using the peer-to-peer function. This function is available only on the dedicated CAN communication bus. The Master FloWEB sends periodically, every 10 seconds, a “I am the Master” message on the peer-to-peer communication bus; should a failure of the Master flow computer or any of the alarms of Table 10-1 occur, after a 10 seconds timeout, the Slave flow computer is automatically promoted to Master and assumes all measurement functions. The totals from the ‘Master FloWEB’ are copied to the ‘Standby FloWEB’ with a periodic time interval of 5 minutes. It’s important that the two flow computer have exactly the same configuration; in case of small differences in the configuration, the Slave flow computer will go in a “Not aligned configuration” status and the totals will not be updated from the Master. In this situation, and only in this case, it will be possible to Login in the Slave and change its configuration to be aligned with the Master; when the Salve configuration will be aligned with the Master, the latter will start again to align the totals to the Slave. When everything work correctly, the Login will be possible only in the promoted Master. The two Redundant FloWEB flow computers have a dedicate ModBus address common to both flow computers. This feature is particularly useful in metering systems applications whereby only the ‘Master’ flow computer is required to communicate with the supervisory computer. When the supervisory computer will send a read/write message, only the promoted Master will answer.

FloWEB User Manual – Edition 2 Revision 4 Page 131/ 131 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 132 di 132

10.2. Redundant FloWEB wiring requirements Fig. 10-1 shows the wiring needed when flow computers are applied in a redundancy scheme via the peer-topeer feature and using the proprietary CAN communication bus.

Figure 10-1: CAN communication bus wiring for redundant FloWEB connection

There are two typical connections of two FloWEB for Hot/Standby, or Redundancy, architecture; Fig. 10-2 show the base architecture with shared transmitters between the two flow computers; in a redundant system all input signals must be connected to both Master and Slave flow computers. Voltage pulse signals such as flow meters devices must be connected in parallel to the appropriate inputs of both flow Master and Slave computers. Analogue 4-20 mA signals should be converted to 1-5 V; as a general rule, follow the wiring recommendations shown for a normal single flow computer installation (see § 4.) and then simply wire the second flow computer terminals in parallel with the first computer.

Warning!!

With shared input signals, only the power fail or Master CPU fault situations can be handled; any fault in the input transducers cannot be managed.

Figure 10-2: Hot/Standby scheme with shared transmitters

FloWEB User Manual – Edition 2 Revision 4 Page 132/ 132 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 133 di 133

Fig. 10-3 shows the preferred architecture that use redundant transmitters connected to each flow computer; with such wiring arrangement any fault in the input transducers can be managed correctly.

Figure 10-3: Hot/Standby scheme with redundant transmitters

FloWEB User Manual – Edition 2 Revision 4 Page 133/ 133 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

11.

Page 134 di 134

Principle of Operation

The FloWEB Gas Flow Computer is designed to calculate the total energy, volume and instantaneous flow rates of natural gas. Calculation is carried out using inputs from pulse generating turbine meters or from differential pressure measurement across orifice plates together with temperature sensors and transmitters for line pressure. Turbine and Orifice configurations are both supported by the same FloWEB configuration, thus allowing supporting both measurement setups with the same unit; measurement configuration is selected by means of the user interface. Single stream, Double stream and Station configurations are also selectable by the user, providing enhanced application flexibility with a unique device. Since acquisitions and calculations are repeated at each cycle, together with re-reading, update and memorization in non-volatile storage of all new values, no transitional disturbance can have lasting effect; the fault recovery from such disturbances is therefore automatic, one or two seconds after the disturbance has ceased (for single and double stream, respectively). The following sections provide description of FloWEB operation and overall functionalities.

11.1. Acquisition Input variables, such as Pressure, Temperature, Status and Flow Meter inputs are being updated every 500 milliseconds, according to acquisition cycle defined by CPU and forwarded to Analogue Inputs boards. Transmitters can be both interfaced as traditional analogue variables, such as 4-20 mA or RTD inputs, and by means of HART digital interface, to overcome periodic recalibration needs. 11.1.1. Alarms All input and output signals are tested to ensure that they are within the designed operating limits; an alarm display records the time when alarms occur and clear. During an alarm condition the total flow is displayed on a separate counter. Specific functional module provides FloWEB with the capability of generating alarms upon user defined conditions, such as threshold trespassing for analog variables, variations of status contacts, etc. Each variable can be independently configured to generate an alarm. 11.1.2. Turbine - Fidelity check Whenever Turbine metering is being selected fidelity check on double pulse emitters can be enabled, providing real time verification of proper emitter connection, according to defined blade ratio. If a mismatch is detected an alarm is raised and acquisition can be configured to either stop or keep on going.

11.2. Calculation 11.2.1. Compressibility Gas Factor (Z) Compressibility Gas factor (Z) is calculated according to selected algorithm, as follows: − − − − − −

AGA-NX19mod ISO12213-3: 1997 (SGERG91) MGERG AGA8 gross 1 AGA8 gross 2 AGA8 Detailed

FloWEB User Manual – Edition 2 Revision 4 Page 134/ 134 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 135 di 135

Compressibility algorithm is performed only if gas analysis parameters and input sensors (pressure and temperature) are in a valid range. Outside the valid range a faulty condition is detected and Volumes at base condition are not computed. Alternatively use can select to use default values instead of live values, so that calculation is kept ongoing; also in this case volume at measurement conditions are computed in a separate error counter (Veb). Range of input gas analysis parameters and input sensors (pressure and temperature) for each selected formula are reported in the following: AGA-NX19mod: Relative density Carbon dioxide Nitrogen Meas. Pressure Meas. Temperature

= 0.554 - 1.00 = 0 - 15% = 0 - 15% = 0..75 Bar = -20..+50 °C

ISO12213-3 (SGERG91): Relative density = 0.55 - 0.9 Heating value = 20 - 48Mj/m^3 Carbon dioxide = 0 - 30% Hydrogen = 0 - 10% Meas. Pressure = 0..75 Bar Meas. Temperature = -20..+50 °C MGERG: Relative density Methane Nitrogen Carbon dioxide Ethane Hydrogen Propane Carbon Monoxide Butanes Helium Pentanes Hexanes Heptane Octanes Meas. Pressure Meas. Temperature

= 0.55 - 0.9 = 50 – 100% = 0 - 50% = 0 - 30% = 0 – 20% = 0 - 10% = 0- 5% = 0 - 3% = 0 – 1,5% = 0 – 0,5% = 0 – 0,5% = 0 –0,1% = 0 – 0,1% = 0 – 0,1% = 0..100 Bar = -23..+60 °C

AGA8 gross 1 and gross 2: Relative density = 0.55 - 0.87 Heating value = 16.65 - 45.12 MJ/m^3 Carbon dioxide = 0 - 30% Hydrogen = 0 - 10% Nitrogen = 0 - 15% Carbon monoxide = 0 - 30% Meas. Pressure = 0..75 Bar Meas. Temperature = -10..+50 °C FloWEB User Manual – Edition 2 Revision 4 Page 135/ 135 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 136 di 136

FloWEB_UM_UNICIG_R24 ENG.doc

AGA8 Detailed: Nitrogen Carbon Dioxide Ethane Propane Water H2S Hydrogen Carbon Monoxide Oxygen Butanes Pentanes Hexanes Helium Argon Meas. Pressure Meas. Temperature

= 0 –100% = 0 – 100% = 0 – 100% = 0- 12% = 0 – 10% = 0 – 100% = 0 – 100% = 0 - 3% = 0 – 21% = 0 – 6% = 0 – 4% = 0 – 10% = 0 – 3% = 0 - 100% = 0..100 Bar = -30..+70 °C

Acquisition of new measures and calculation is performed every 1 second. 11.2.2. Calculation of Relative density (Specific gravity) The relative density is the ratio between the density of the specific gas and the density of air at the same stated conditions (reference temperature and pressure). Density = RHOgas / RHOair Table 9-1 shows the air density (RHOair) considered by FloWEB according to reference pressure and temperature conditions: Reference Pressure 1,01325 bar 1,01325 bar 1,01325 bar 14,73 psi

Reference RHO Temperature air(*) 0°C 1,292923 15°C 1,22541 20°C 1,204449 60°F 1,225878

(*) Tabled values according to ISO and AGA standards. Table 9-1: Air Density at different reference conditions

FloWEB User Manual – Edition 2 Revision 4 Page 136/ 136 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 137 di 137

11.2.3. TURBINE configuration Calculations performed depend on type of measurement system being selected; the following sections provide description of supported algorithms whenever Turbine is selected.

11.2.3.1.Correction of Flow Meter Error (Vc) FloWEB can perform automatic correction of Flow Meter error, by applying the following formula:

Vc = Vm ×

100 100 + Err %

Where: Vc = Flow Meter Error Corrected Volume at Operating Conditions Vm = Flow Meter Measured Volume at Operating Conditions Err% = Flow Meter Error Percentage at working frequency As a result a Correction Factor Cf is defined as follows:

Cf = Vc / Vm =

100 100 + Err %

Whenever Flow Meter error correction shall apply, operator is being asked to enter Flow Meter calibration certificate values, as follows: Qmin = minimum gas flow rate Qmax = maximum gas flow rate Err% / Q (10 couples max.) = Flow Meter error percentage (Err%) at the defined flow rate (Q), as percentage of Qmax (Qmax%). Correction applies within the defined range only, while actual Flow Meter error is calculated according to actual flow rate Q, as a result of linear interpolation upon the entered curve. When no setup is performed Err% is by default set to 0, so that Vc = Vm. No correction applies whenever input frequency < 10 Hz and Q ≤ Qmin, while correction factor remains at the value obtained for Qmax when Q ≥ Qmax.

FloWEB User Manual – Edition 2 Revision 4 Page 137/ 137 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 138 di 138

11.2.3.2.Volume at Base Conditions (Vb) FloWEB performs Measured Volume (Vm) conversion to Volume at reference conditions (base conditions, Vb) by applying AGA7 formula, as follows:

Vb = Vc ×

p tb + To Zb × × pb t + To Z

where: Vb = Volume at reference conditions Vc = Flow Meter Volume corrected at operating conditions p = Operating Gas pressure pb = Reference Pressure t = Operating Gas temperature tb = Reference Temperature To = Absolute Temperature conversion factor (273,15°K) Z = Compressibility factor at operating conditions Zb = Compressibility factor at reference conditions As a result of calculation above, conversion factor C is defined as follows:

C = Vb / Vc =

p tb + To Zb × × pb t + To Z 11.2.3.3.Flow Rates (Q)

FloWEB performs calculation of Hourly Gas flow rates (Q) as a result of either volume acquisition (turbine input) or calculation (orifice), as follows: Qm = f (Vm) Qc = f (Vc) Qb = f (Vb) In order to provide effective flow rate measurement, a different calculation algorithm is used according to the type of input signal, as follows: •

Low Frequency (Max 2 Hz): While Flow rate calculation is performed every 1 second, in order to provide a stable measurement even for very low frequency inputs, pulses are integrated over a “running-window” width of 5 minutes, with totalizers updated at each 30-seconds sub-step. As a result hourly flow rate Q is computed multiplying the pulses computed in the last 30 seconds for 120, as follows: Q = Σn

(n-120)

Vn(1) * 120

Where Vn(1) are the volume computed in the considered 30” period. The formula above applies to Vm, Vc and Vb for the calculation of Qm, Qc and Qb respectively.

FloWEB User Manual – Edition 2 Revision 4 Page 138/ 138 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc



Page 139 di 139

High Frequency (Max 5 kHz) Since input frequency is supposed to allow better resolutions, hourly flow rate is simply calculated every second as follows: Q = Vn(1) * 3600 Where Vn(1) are the volume computed in the last 1” period. The formula above applies to Vm, Vc and Vb for the calculation of Qm, Qc and Qb respectively.

11.2.4. ORIFICE configuration Calculations performed depend on type of measurement system being selected; the following sections provide description of supported algorithms whenever Orifice is selected. 11.2.4.1.Flow Rate (Q) FloWEB performs Measured Flow (Qm) conversion to Flow at reference conditions (base conditions, Qb) by applying the selected formula, as follows: • • •

AGA3 ISO5167-1 amd1 - 1998 ISO5167 - 2 – 2003

Calculated quantities are directly dependent on measured differential pressure, as a result of Orifice flow measurement principle; to avoid calculation errors whenever the cut off (flow rate = 0) condition is being approached, FloWEB manages a so defined “cut off” threshold, below which flow rate is set to 0. Cut off threshold value is set by the user.

11.2.5. Mass Flow Rate (Qmb) Mass Flow Rate calculation is performed on the basis of Volume Flow Rate, considering gas density at base conditions, as follows: Qmb = Qb x rhob Where: Qmb = Mass Flow Rate at Base conditions Qb = Volume Flow at Base conditions rhob = Gas density at base conditions Calculation algorithm above applies for both Turbine and Orifice setups.

FloWEB User Manual – Edition 2 Revision 4 Page 139/ 139 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 140 di 140

11.2.6. Energy Flow Rate (Qeb) Energy Flow Rate calculation is performed on the basis of Volume Flow Rate, considering Heating Value at base conditions, as follows: Qeb = Qb x HVb Where: Qeb = Mass Flow Rate at Base conditions Qb = Volume Flow at Base conditions HVb = Gas Heating Value at base conditions Calculation algorithm above applies for both Turbine and Orifice setups.

11.3. Communication 11.3.1. Serial Ports FloWEB provides up to five RS232/RS422/RS485 serial data lines which support ModBus RTU/ASCII Slave communication protocol to allow remote reading and SCADA applications; furthermore each port can be alternatively setup for Printer, Ultrasonic meter or Gas Chromatograph operation, namely supporting ASCII text output and ModBus RTU/ASCII Master protocols, or for SNAM gas data collection protocol.

11.3.2. Ethernet Port FloWEB provides an Ethernet Port on which both ModBus TCP and WEB browsing operations are simultaneously supported.

11.3.3. Communication Protocol For details on supported communication protocols and related data structure please refer to “FloWEB ModBus / ModBus TCP protocol” Ed.1, Rev.1, related to FMW release referred herein.

FloWEB User Manual – Edition 2 Revision 4 Page 140/ 140 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

12.

Page 141 di 141

FloWEB HTTP

FloWEB has a built-in Web server, accessible on the machine’s 10 Mbps Ethernet 802.3 compliant connection at standard Port 80 with standard HTTP protocol. Built-in Web pages provide a way to upload firmware upgrades to the machine, as well as access to local files for download. Built-in IP address when FloWEB is switched on, if not programmed, is set to 10.3.5.201 with network mask set to 255.255.0.0. A laptop PC can be directly connected to the Ethernet port through use of a crossed Ethernet patch cable for direct connection. The PC Ethernet board/internet protocol settings must be configured with the same network mask and local IP address in the range 10.3.x.x (excluding FloWEB own address). When correct electrical connection is achieved, led “LNK” on FloWEB front panel lights on: logical connection can then be tested by pinging address 10.3.5.201 before opening a web browser to said address as in the following picture:

Warning!!: for regular operation when using Microsoft Internet Explorer 7, go to Tools -> Internet Options -> Advanced and under Browsing un-check the 'Show friendly HTTP error messages' check box.

FloWEB User Manual – Edition 2 Revision 4 Page 141/ 141 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 142 di 142

Click on the arrow next to “File Upload” description: this will open the following page where you can enter files information to upload new firmware revisions.

To upload a new firmware version, in the “DESTINATION filename” box write the result name of the code file to upload (currently “Floweb.bin”). Click the “Browse” button next to the “SOURCE filename” box to open a Windows explorer file selection window, and select the path and name of the file, containing the new firmware application code, to upload. Then click the “Send File” button and wait for the file to upload. Warning!!: to upload a file in the FloWEB, the programming button has to be pressed, otherwise an error message will be displayed. When a message appears stating that the upload was successfully completed, switch Off and then On the FloWEB to start with the new application firmware just uploaded.

FloWEB User Manual – Edition 2 Revision 4 Page 142/ 142 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 143 di 143

Select “FloWEB directory listing” at the bottom of the page to list files currently residing in the Flash file system of the FloWEB.

In factory configuration, a few files and directories will be displayed in FloWEB main directory. The most important file is “Floweb.bin”: this is the FloWEB application software file. The second most important file is “vars.sav”, the file containing the entire FloWEB configuration made by the user. After FloWEB configuration has been completed, we suggest to back-up “vars.sav” file for future use (i.e., to configure another FloWEB in case of multiple streams). Other files in the root directory are part of the Factory Built-In Web Interface. There are also some subdirectories, each dedicated to the files connected with a specific function: the “DBSNAM” sub-dir for instance holds the data files necessary to the proprietary SNAM protocol required for Italian market, while the “DBLOG” sub-dir holds logs (produced when they’re configured and enabled in the related Config. Menu). The “VIRTP” sub-dir will contain virtual print rolls, if the “Virtual Printer” function has been enabled and report configured for it.

FloWEB User Manual – Edition 2 Revision 4 Page 143/ 143 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 144 di 144

The “WUI” sub-dir holds the WEB USER INTERFACE files, distinct from the Factory Built-In Interface: the WUI sub-dir files can optionally be absent or even personalized so as to give a different Web User Interface from the Standard WUI provided, in which case it’s advisable to download the factory provided files for backup before trying to do any change.

When WUI sub-dir is present with all the files, writing FloWEB IP address on the address bar of the browser (e.g. 10.3.5.201), the browser will be automatically redirected to the WUI main page. To access the WUI interface from Factory Built-In Interface, go to the WUI sub-dir by clicking on it in the web browser file root dir file list, then click the “index.html” file to load the WUI Main Page (N.B. This is a different file from the “Index.htm” in the root directory, which is the Factory Interface main page instead).

FloWEB User Manual – Edition 2 Revision 4 Page 144/ 144 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 145 di 145

FloWEB_UM_UNICIG_R24 ENG.doc

13.

Technical Specifications 13.1. Overall Specifications

Dimension 19” Rack Mounting Weight Case Material Environmental rating Operating Temperature Storage Temperature PowerSupply Fuse

: 213 x 232 x 128.4 mm (see figure) : possible (fixing screw supplied) : 2Kg : Aluminum : IP20 : -10°C to +50°C (condensing humidity) : -20°C to +60°C : 18 to 36VDC / 2A max : 5x20mm glass 3.15AT / 250V slow blow

Pietro Fiorentini

7

8

4

5

6

1

2

3

9

ON BL ALM CFG

0

J1

PROG

I/O Connectors Max wire section Pitch Display Type Backlight Keyboard

F1

F2

F3

F4

F5

F6

+ ESC -

: Multiple (14 or 16 pole) removable connectors 2 : 1mm : 3.81mm

: Monochrome (blue background) ¼ VGA size graphic TFT display : White led type backlight (always active) : Numeric, selection and six Soft-Function keys

13.2. CPU board CPU Architecture CPU RAM On board RAM On board Flash Supported external memory Clock & Calendar Clock accuracy Reference condition (+25°C) Thermal drift(-10°C to +50°C) Backup battery

: 32bit Cold Fire processor : 64Kbyte (battery back-up) : 16Mbyte : 4Mbyte : Compact Flash Type I : Real-time clock calendar with DST management : +/- 3 minutes/year : -50ppm/°C : Type: CR2032 (lithium) / Estimated backup time: 4.5 years

FloWEB User Manual – Edition 2 Revision 4 Page 145/ 145 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 146 di 146

FloWEB_UM_UNICIG_R24 ENG.doc

13.3. Analogues Inputs board Analogue Input Board: SK0707T07M01 13.3.1. General Specification Analogue Input channels : 5 multifunction input channels: voltage, current loop : 1 PT100 input channel Digital Input channels : 1 ultrasonic flow meter input : 2 Flow Meter inputs (part of the multifunction input channels) Hart : 1 HART input A/D converter resolution Conversion Time Isolation Isolation Voltage

: 16bit : 500mS (all channels) : all inputs channels are isolated from earth logic : 500Vdc

13.3.2. Analogues Inputs specifications Current loop (4-20mA) Input Range : 0 - 21mA (nominal 4-20mA) Input Load : 250Ω On board Energization Supply Voltage : 24V +/-5% External Energization : 30V max Accuracy Reference condition (+25°C) : +/- 0.02% FS Thermal drift (-10°C to +50°C) : +/- 6 ppm/°C DC Voltage input Input range Input impedance Accuracy Reference conditions (+25°C) Thermal drift (-10°C to +50°C)

: 0 - 5.25V (nominal: 1 - 5V) : ≥1 MΩ : +/-0.01% FS : +/-4 ppm/°C

PT100 Input Input type : 4 wires RDT 100 Energization current : 1mA typical Input range : -30°C to +70°C Accuracy Reference condition (+25°C) : +/-0.08°C Thermal Drift 8-10°C to +50°C) : +/-600 µ°C/°C

FloWEB User Manual – Edition 2 Revision 4 Page 146/ 146 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 147 di 147

FloWEB_UM_UNICIG_R24 ENG.doc

13.3.3. Digital Inputs Specifications Flow Meter Inputs Input Type Frequency range BF HF Ultrasonic Input Frequency accuracy

: voltage free contact / Namur (8V/1.2mA) : up to 2Hz : up to 5KHz : up to 10Khz : +/-50ppm

NOTE: The manufacturer recommends a periodical (max two years) metrological verification. Such verification can be performed on the site.

FloWEB User Manual – Edition 2 Revision 4 Page 147/ 147 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

Page 148 di 148

FloWEB_UM_UNICIG_R24 ENG.doc

13.4. Analogues Outputs board Analogue output board: SK0707T08M01 13.4.1. General Specifications Output channels : 6 4-20mA current D/A converter resolution : 12bit Refresh rate : 100mS Isolation : all output channels are singularly isolated from earth logic Isolation Voltage : 500Vdc 13.4.2.

Outputs Specifications Output type : current sink (passive mode) Current range : 4-20mA External Supply Voltage : 5-30V Max load : 1KΩ Accuracy Reference condition (+25°C) : +/- 0. 1% FS Thermal Drift (-10 to +50°C) : +/-0.01%/°C 13.5.

Digital Outputs (on Power Supply board)

Available on PSU Board SK0707T06M01 13.5.1. Mos-switches Digital outputs Output type : 4+4 Mos Relays not polarized Voltage/Current ratings : 60Vdc / 500mA Isolation : group of 4 to other group and earth logic Isolation voltage : 500Vdc 13.5.2. PSU Status Relays Contact Voltage/Current ratings : 130Vac –30Vdc / 1A Isolation (contacts /coil) : 1500Vac 50Hz (1 minute) Expected electrical life : 100000 operations

13.6. Communication Interfaces 13.6.1. CPU Board Ethernet Connector Serial port CAN port

: Ethernet 10 Base T : RJ45 (twisted pair cable) : two selectable mode serial ports : up to 512Kbit/s

13.6.2. Auxiliary Board Ethernet Serial port

: front panel connector (RJ45) for CPU Ethernet port : 1 selectable mode serial port

FloWEB User Manual – Edition 2 Revision 4 Page 148/ 148 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 149 di 149

13.6.3. Analogues Inputs Board Serial port : 1 selectable mode serial port 13.6.4. Serial ports Specifications Specifications applied to all serial ports Mode RS232 signal : Tx, Rx, RTS, CTS Output Level : ≥ +/- 5V RS485 connections : 2 and 4 wires termination : selectable (120Ω) polarization : selectable (680Ω) RS422 Speed (all mode) : 300 to 38400baud Selection : by DIP switches Connections : terminals block

FloWEB User Manual – Edition 2 Revision 4 Page 149/ 149 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Blank wiring schedule

Name

External connections

Wiring schedule

Type

Connector & pin no.

Pin

Pin

Connector & pin no.

Wiring Signal Colour Barrier

FloWEB

Sheet

of

Comments

APPENDIX A

Page 150 di 150

FloWEB User Manual – Edition 2 Revision 4 Page 150/ 150 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice

FloWEB User Manual

FloWEB_UM_UNICIG_R24 ENG.doc

Page 151 di 151

FloWEB User Manual – Edition 2 Revision 4 Page 151/ 151 The information contained in this document is confidential and property of Pietro Fiorentini S.p.A. Technical data may be subject to change without notice