Citation preview

Safety Information

Product Overview Taking and Reading X-ray of a Subject Searching Patients and Referring Studies Verifying and Processing Images

(CR-IR 391CL)

Reference Guide 3rd Edition November 2010

Using Report Function Image Storage from Each Window Image Print from Each Window Editing Study Information

Verifying with the Indicators Changing Display Settings of the “Main” Window and “Image Viewer” Using “Study Utilities”

Configuring “Client Utilities”

Setting “User Utilities” This Reference Guide describes details on how to operate the FCR PRIMA Console (CR-IR 391CL) and cautions to be observed when operating it. Please read the

Troubleshooting

Reference Guide thoroughly before actually operating the FCR PRIMA Console (CR-IR 391CL) system. After reading this manual, store it nearby the FCR PRIMA Console (CR-IR 391CL) so that you can see it whenever necessary. Please also

Daily Inspection and Maintenance

read the “FCR PRIMA Console (CR-IR 391CL) Operation Manual” which provides detailed information on the operation of the FCR PRIMA Console (CR-IR 391CL).

897N101440B

Appendix Index

Introduction This Reference Guide applies to the following software. • FCR PRIMA Console Software V2.0(B) This Reference Guide offers the reader a summary insight into the FCR PRIMA Console (CR-IR 391CL) and proceeds to present information prerequisite to using the product, including operating instructions and precautions, and tips on daily inspections and care. FCR PRIMA Console is a CR workstation intended to provide diagnostic quality images and measuring function to aid the doctor in diagnosis, and also is a viewer for US referral images.

CAUTION 1 No part or all of this manual may be reproduced in any form without prior permission. 2 The information contained in this manual may be subject to change without prior notice. 3 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from installation, relocation, remodeling, maintenance, and repair performed by dealers other than those specified by FUJIFILM Corporation. 4 FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation products due to products of other manufacturers not supplied by FUJIFILM Corporation. 5 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling, maintenance, and repair using repair parts other than those specified by FUJIFILM Corporation. 6 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from negligence of the precautions and operating methods contained in this manual. 7 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use under environment conditions outside the range specified for this product, such as the power supply, installation environment, etc., as described in this manual. 8 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural disasters, such as fires, earthquakes, floods, lightning, etc. Caution: R  x Only in the United States (Federal law restricts this device to sale by or on the order of a physician)

This system is classified as a medical device under EC Directive 93/42/EEC. This equipment has been designed on the assumption that the patient would not come into direct contact with it or for operation by appropriately trained operator. Process waste correctly, as stipulated by local law or any regulations that apply. When discarding the CR-IR 391CL that incorporates the lithium battery, be sure to contact a licensed waste disposal contractor because it cannot be disposed of as a general waste. Trademarks

FCR is a trademark or a registered trademark of FUJIFILM Corporation. Other holders’ trademarks

Windows Vista and Windows 7 are the registered trademark of US Microsoft Corporation in the U. S. A. and other countries. Windows is the registered trademark of US Microsoft Corporation in the U. S. A. and other countries. Microsoft Internet Explorer logo is the registered trademark of US Microsoft Corporation in the U. S. A. and other countries. All other company names and product names described in this manual are the trademarks or registered trademarks of their respective holders. Copyright © 2009-2010 FUJIFILM Corporation. All rights reserved.

ii

FCR PRIMA Console Reference Guide

897N101440B

How to Read This Guide

■ Basic Page Organization The pages of this document is basically organized into the following elements: Lead

Section title

Describes information that is about to be explained in this section together with supplementary information.

A brief explanation of the operation that is covered in the section. Item Title A brief explanation of the operation that is covered under the item. Explanatory Text A summary introduction to the topic described under the item, with precautions that should deserve special notice, if any.

7.2

Displaying Dialogs for Saving

Saves images from the “Refer” window, “Image Viewer”, and “Study Utilities”.

7.2.1 Starting from the “Refer” Window

■ Operation Window

■ Displaying Dialogs for Saving

● “Main” Window - “Refer” Window

1.

Search the patient.

2.

From the “patient list” (A), select the patient in a “Verified” study status.

Select the patient and the image to be referred.

A list of headings that offers you a quick clue to opening chapters of interest.

“4.2 Searching for Patients”

7

→ The image is displayed in the “Refer” window (B).

Select the image (B) to be saved.

4.

Click the desired button (C) according to the storage media.

Image Storage from Each Window

3.

Functions of Each Button

Display Windows A window that opens during an operation.

Index

Searches the patient and saves the images shown in the “Refer” window to each media. This section describes how to display dialogs for saving from the “Refer” window according to the storage media.

Buttons

Functions

Operation procedure

Save to storage disks*1 (A)

“7.3 Saving to Storage Disks” Transmit to the DICOM Server*1 “7.4 Transmitting Studies to DICOM Server” Save to PDI Disks*1 “7.7 Saving to PDI Disks”

(B)

A sequence of numbered operational steps to guide the user through an operation.

*1 The image whose study status is “Verified“ cannot be saved or transmitted.

(C) *1 If the selected studies except for “Verified” are selected, a message displays to verify. *1 When the buttons (C) are not displayed, operation items must be added in the “Display Settings”.

To continue excluding the studies which are not able to process, click [OK].

“11.1.7 Display Settings of “Refer” window”

FCR PRIMA Console Reference Guide

897N101440B

7-3

Page number An indication of the order of a particular page in a chapter.

FCR PRIMA Console Reference Guide

897N101440B

iii

■ Marks Used in This Guide Marks are used in the descriptions to mark supplementary information and information that must be observed while using the FCR PRIMA Console. For your safety and correct use, always observe a Warning or Caution. Marks

WARNING CAUTION

Contents Indicates a dangerous situation which, if you ignored or could not follow for some reason, could result in death or serious injury. Indicates a hazardous situation which, if you ignored or could not follow for some reason, could result in a slight or moderate degree of personal injury or material damage. Indicates information that should be noted. Indicates information that may be helpful. Indicates an item that provides details of the procedure or related information.

■ Manners of Description In this Reference Guide, button names, menu actions, window names and messages appearing in the operational windows are designated in the following manners:

● Buttons (example) [Complete] Names of buttons in the operational windows are enclosed in brackets [ ], regardless of their shape.

● Menus (example) [Annotation] – [Rectangle] Menu names are enclosed in brackets [ ]. Sequences of menu choices are designated in the format [ ] – [ ]. In this example, users are supposed to choose [Annotation] first, then [Rectangle].

● Windows (example) “Read Image” window Window names are enclosed in quotation marks “ ”.

iv

FCR PRIMA Console Reference Guide

897N101440B

Contents Introduction.............................................................................................................................ii How to Read This Guide......................................................................................................iii

■ Basic Page Organization...............................................................................................................iii ■ Marks Used in This Guide.............................................................................................................iv ■ Manners of Description..................................................................................................................iv

Contents..................................................................................................................................v

Chapter 1

Safety Information

1.1

Precautions Before Operating This Equipment............................................. 1-2

1.2

Notes on Using Medical Electrical Equipment (for Safety Assurance and Risk Prevention)................................................... 1-3

1.3

Safety..................................................................................................................... 1-4

1.4

1.5

■ Preventing Explosion..................................................................................................................1-4 ■ Preventing Electrical Shock........................................................................................................1-4 ■ Connection Instruction................................................................................................................1-4 ■ Electromagnetic Compatibility Instruction..................................................................................1-5 Precautions in Using the FCR PRIMA Console.............................................. 1-7

■ Precautions when Using.............................................................................................................1-7 ■ Precautions when Installing........................................................................................................1-8 ■ Precautions when Reading Images...........................................................................................1-9 ■ Precautions when Managing Data...........................................................................................1-10 ■ Precautions when Processing Image......................................................................................1-10 ■ Precautions when Exposing.....................................................................................................1-11 ■ Precautions when Saving Data................................................................................................1-11 ■ Precaution when Outputting.....................................................................................................1-11 Labels.................................................................................................................. 1-12

■ FCR PRIMA Console (CR-IR 391CL) Labels..........................................................................1-12

Chapter 2 2.1

2.2 2.3

Product Overview

FCR PRIMA Console Highlights........................................................................ 2-2

■ Main Features of FCR PRIMA Console.....................................................................................2-2 ■ System Configuration Example..................................................................................................2-3 ■ Restrictions..................................................................................................................................2-4

Name and Window of Each Part....................................................................... 2-5

■ Display Window...........................................................................................................................2-5

Flow for Utilizing the System............................................................................. 2-6 FCR PRIMA Console Reference Guide

897N101440B

v

2.4

2.5

Version Information............................................................................................. 2-8

■ Operation Windows....................................................................................................................2-8 ■ Checking Version Information....................................................................................................2-8 Starting up and Exiting....................................................................................... 2-9 2.5.1

Starting up FCR PRIMA Console........................................................... 2-9

2.5.2

Exiting FCR PRIMA Console................................................................. 2-12

■ Operation Device / Operation Windows....................................................................................2-9 ■ Starting up.................................................................................................................................2-10 ■ Starting up the Server Application ...........................................................................................2-11 ■ Starting up the Client Application.............................................................................................2-11 ■ Starting up Each Function from the Server Application Icon..................................................2-11 ■ Operation Windows..................................................................................................................2-12 ■ Logging off the Current User....................................................................................................2-13 ■ Exiting the Client Application....................................................................................................2-13 ■ Exiting the Server Application .................................................................................................2-13 ■ Shutting down the PC...............................................................................................................2-14

2.6

2.7

2.8

Protecting the PC when Temporarily Leaving the Site................................ 2-15

■ Operation Windows..................................................................................................................2-15 ■ Locking the PC..........................................................................................................................2-15 ■ Unlocking the PC......................................................................................................................2-15 Before Using Disks........................................................................................... 2-16

■ Precautions...............................................................................................................................2-16 ■ Disk Types.................................................................................................................................2-17 ■ Drive Types................................................................................................................................2-18 Setting up Disks................................................................................................ 2-19

■ Operation Windows..................................................................................................................2-19 ■ Setting up Disks........................................................................................................................2-20 ■ Ejecting Disk..............................................................................................................................2-21 ■ Copying Disks...........................................................................................................................2-22 ■ When a Disk Becoming Full.....................................................................................................2-23

Chapter 3 3.1

Flow of Taking and Reading X-ray.................................................................... 3-2

3.2

Registering the Patient Information to be Exposed....................................... 3-3

3.3

vi

Taking and Reading X-ray of a Subject

■ Operation Window......................................................................................................................3-3 ■ Registering New Patient Information.........................................................................................3-3 ■ Registering through DICOM MWM Server (Optional)..............................................................3-4 Exposing Images................................................................................................. 3-5

■ Operation Window......................................................................................................................3-5 ■ Taking X-rays...............................................................................................................................3-5

FCR PRIMA Console Reference Guide

897N101440B

3.4

3.5

Reading and Viewing.......................................................................................... 3-7

■ Operation Window......................................................................................................................3-7 ■ Read Images in the Cassette.....................................................................................................3-7 ■ Viewing Read Images.................................................................................................................3-8 ■ Finishing Exposure / Reading....................................................................................................3-9

Import Various Images and Refer................................................................... 3-10

■ Operation Window....................................................................................................................3-10 ■ Importing DICOM Images from the Media or Folder (Import DICOM Off-line)......................3-10

Chapter 4

Searching Patients and Referring Studies

4.1

Flow for Searching / Referring ......................................................................... 4-2

4.2

Searching for Patients........................................................................................ 4-3

4.3

■ Operation Window......................................................................................................................4-3 ■ Searching for Patients.................................................................................................................4-3 Referring Images................................................................................................. 4-5

■ Operation Window......................................................................................................................4-5 ■ Manipulating Studies...................................................................................................................4-5

Chapter 5

Verifying and Processing Images

5.1

Flow of Verifying and Processing Images....................................................... 5-2

5.2

Outline of "Image Viewer" Window.................................................................. 5-3

5.3

Selecting and Arranging Images....................................................................... 5-4

5.4

■ Descriptions of Functions...........................................................................................................5-4 ■ Operation Windows....................................................................................................................5-4 ■ Switching Displayed Images......................................................................................................5-5 ■ Adding a Sub-window and Displaying the Image......................................................................5-5 ■ Displaying Multiple Sub-windows Lined up...............................................................................5-6 ■ Displaying Multiple Images Lined up in the Sub-window..........................................................5-6 ■ Maximizing the Sub-window.......................................................................................................5-6 ■ Switching the Display Position...................................................................................................5-7 ■ Displaying/Hiding Thumbnail Images.........................................................................................5-7 ■ Displaying/Hiding Various Information on the Image................................................................5-7 ■ Displaying as a Cine Display Mode...........................................................................................5-8 ■ Registering Displayed Images as a Key Image........................................................................5-8 ■ Clearing Images..........................................................................................................................5-8 Adjusting Image Display.................................................................................... 5-9

■ Descriptions of Functions...........................................................................................................5-9 ■ Operation Windows....................................................................................................................5-9 ■ Changing the Gradation (Contrast and Density).....................................................................5-10 FCR PRIMA Console Reference Guide

897N101440B

vii

■ Magnifying/Reducing................................................................................................................5-11 ■ Magnifying Images by Magnifying Glass.................................................................................5-12 ■ Moving Images..........................................................................................................................5-12 ■ Reversing Images.....................................................................................................................5-13 ■ Rotating and Flipping................................................................................................................5-13 ■ Editing FCR Image Processing Parameters............................................................................5-14 ■ Registering Preset Image Processing Settings.......................................................................5-16 ■ Applying Preset Image Processing Settings...........................................................................5-17 ■ Linking Image Processing Among Multiple Images................................................................5-17 ■ Resetting the Image Processings............................................................................................5-17 5.5

5.6

5.7

Adding Annotations on Images...................................................................... 5-18

■ Descriptions of Functions.........................................................................................................5-18 ■ Operation Windows..................................................................................................................5-18 ■ Adding Arrows/Shapes (Ellipses, Rectangles, and Polygons)...............................................5-19 ■ Adding Texts..............................................................................................................................5-20 ■ Adding Markers.........................................................................................................................5-20 ■ Adding Measured Lengths.......................................................................................................5-21 ■ Adding Measured Simplified Cardiothoracic Ratios................................................................5-21 ■ Adding Measured Cardiothoracic Ratios.................................................................................5-22 ■ Adding Measured Interline Angles...........................................................................................5-23 ■ Adding Measured Dual Angles.................................................................................................5-24 ■ Adding Measured Cobb Angles...............................................................................................5-25 ■ Adding Measured Pixel Values................................................................................................5-25 ■ Adding Measured Various ROIs...............................................................................................5-25 ■ Editing the Added Annotations.................................................................................................5-26 ■ Deleting Added Annotations.....................................................................................................5-27

Free Drawing on “Image Viewer” for Informed Consent............................ 5-28

■ Operation Windows..................................................................................................................5-28 ■ Using the Free Drawing Function.............................................................................................5-28

Saving/Verifying/Suspending Images............................................................ 5-30

■ Operation Window....................................................................................................................5-30 ■ Saving the Image Processing Result and Annotation.............................................................5-30 ■ Verifying/Suspending Images...................................................................................................5-30

Chapter 6

viii

Using Report Function

6.1

Outline of “Report” Window.............................................................................. 6-2

6.2

Creating/Editing the Report............................................................................... 6-3

■ Operation Windows....................................................................................................................6-3 ■ Creating/Editing the Report........................................................................................................6-4 ■ Saving/Verifying a Report...........................................................................................................6-5 ■ Deleting Reports.........................................................................................................................6-5

FCR PRIMA Console Reference Guide

897N101440B

6.3

Editing the Schema or Fixed Phrase................................................................ 6-6

■ Operation Windows....................................................................................................................6-6 ■ Editing the Schema.....................................................................................................................6-7 ■ Editing the Fixed Phrase.............................................................................................................6-7 ■ Registering the Fixed Phrase.....................................................................................................6-7

Chapter 7

Image Storage from Each Window

7.1

Relationships between Windows and Storage Destinations....................... 7-2

7.2

Displaying Dialogs for Saving........................................................................... 7-3 7.2.1

Starting from the “Refer” Window......................................................... 7-3

7.2.2

Starting from the “Image Viewer”.......................................................... 7-4

7.2.3

Starting from the “Study Utilities”......................................................... 7-5

■ Operation Window......................................................................................................................7-3 ■ Displaying Dialogs for Saving.....................................................................................................7-3 ■ Operation Window......................................................................................................................7-4 ■ Displaying Dialogs for Saving.....................................................................................................7-4 ■ Operation Window......................................................................................................................7-5 ■ Displaying Dialogs for Saving.....................................................................................................7-5

7.3

7.4

7.5

7.6

7.7

Saving to Storage Disks..................................................................................... 7-6

■ Operation Window......................................................................................................................7-6 ■ Saving the Screen.......................................................................................................................7-6 Transmitting Studies to DICOM Server (Optional)......................................... 7-7

■ Operation Window......................................................................................................................7-7 ■ Saving the Screen.......................................................................................................................7-7

Saving to Work Disks.......................................................................................... 7-8

■ Operation Windows....................................................................................................................7-8 ■ Saving the Screen.......................................................................................................................7-8

Saving in a General-Purpose File Format (Optional)................................... 7-10

■ Operation Windows..................................................................................................................7-11 ■ Saving the Screen.....................................................................................................................7-11 Saving to PDI Disks........................................................................................... 7-13

■ Operation Windows..................................................................................................................7-14 ■ Saving the Screen.....................................................................................................................7-14

Chapter 8 8.1

Image Print from Each Window

Printing Types and Flow..................................................................................... 8-2

■ Types of Printing.........................................................................................................................8-2 ■ Printing Flow................................................................................................................................8-3 FCR PRIMA Console Reference Guide

897N101440B

ix

8.2

Displaying Print Dialogs..................................................................................... 8-4 8.2.1

Starting from the “Refer” Window......................................................... 8-4

8.2.2

Starting from the “Image Viewer”.......................................................... 8-5

8.2.3

Starting from the “Study Utilities”......................................................... 8-6

■ Operation Window......................................................................................................................8-4 ■ Displaying Print Dialogs..............................................................................................................8-4 ■ Operation Window......................................................................................................................8-5 ■ Displaying Print Dialogs..............................................................................................................8-5 ■ Operation Window......................................................................................................................8-6 ■ Displaying Print Dialogs..............................................................................................................8-6

8.3

8.4

8.5

Printing on a Film (Optional).............................................................................. 8-7

■ Operation Windows....................................................................................................................8-7 ■ Printing on a Film........................................................................................................................8-7 Free Layout Printing........................................................................................... 8-9

■ Operation Windows....................................................................................................................8-9 ■ Free Layout Printing..................................................................................................................8-10 ■ Creating Additional Image Information Formats......................................................................8-12 ■ Editing Additional Image Information Formats.........................................................................8-12 ■ Deleting Formats.......................................................................................................................8-12 ■ Finishing....................................................................................................................................8-12 Printing Report................................................................................................... 8-13

■ Operation Window....................................................................................................................8-13 ■ Printing a Report.......................................................................................................................8-13 ■ Saving the Report as PDF File.................................................................................................8-13

Chapter 9 9.1

9.2

9.3

9.4

9.5

x

Editing Study Information

Starting up the Editing Function....................................................................... 9-2

■ Operation Windows....................................................................................................................9-2 ■ Starting up...................................................................................................................................9-2

Adding Exposure Menus.................................................................................... 9-4

■ Operation Window......................................................................................................................9-4 ■ Adding an Exposure Menu.........................................................................................................9-4 Changing Exposure Menus............................................................................... 9-5

■ Operation Window......................................................................................................................9-5 ■ Changing an Exposure Menu.....................................................................................................9-5 Deleting Exposure Menus/Canceling the Deletion........................................ 9-6

■ Operation Window......................................................................................................................9-6 ■ Deleting an Exposure Menu/ Canceling the Deletion...............................................................9-6

Inactivating Exposed Images/Activating the Inactivated Images................ 9-7

■ Operation Window......................................................................................................................9-7 ■ Inactivating/Activating Exposure Menus....................................................................................9-7

FCR PRIMA Console Reference Guide

897N101440B

9.6

9.7

9.8

9.9

9.10

Registering Re-exposure Menus...................................................................... 9-8

■ Operation Window......................................................................................................................9-8 ■ Registering a Re-exposure Menu..............................................................................................9-8 Exchanging Images............................................................................................ 9-9

■ Operation Window......................................................................................................................9-9 ■ Exchange Incorrectly Read Images to the Correct Exposure Menus......................................9-9 Setting for Printing, Saving, and Transmitting............................................. 9-10

■ Operation Window....................................................................................................................9-10 ■ Setting up the Destination and Timing for Outputting and Saving..........................................9-10 ■ Changing the Detailed Settings................................................................................................9-11 Editing Patient Information.............................................................................. 9-13

■ Operation Windows..................................................................................................................9-13 ■ Editing Patient Information........................................................................................................9-14 Changing the Order of Menus in the “Read Image” Window..................... 9-15

■ Operation Window....................................................................................................................9-15

Chapter 10 Verifying with the Indicators 10.1

Viewing Indicators and Activating Functions............................................... 10-2 10.1.1 Viewing Indicators.................................................................................. 10-2

■ Position of “Indicator” Window ................................................................................................10-2 ■ Indicator State and Corresponding Notice...............................................................................10-2

10.1.2 Activating Functions from Indicators.................................................. 10-3

■ Operation Window....................................................................................................................10-3 ■ Opening “Indicator” Windows...................................................................................................10-3

10.2

10.3

10.4

Verifying the Status of Image Reader and Cleaning Procedures.............. 10-4

■ Operation Windows..................................................................................................................10-4 ■ Verifying the Status of the Image Reader................................................................................10-4 ■ Cleaning the Scanner of the Image Reader............................................................................10-4 ■ Verifying the status of the Image Reader and Exiting Scanner Cleaning..............................10-4 Verifying Disk Status and Maintenance......................................................... 10-5

■ Operation Windows..................................................................................................................10-5 ■ Verifying Disk Status.................................................................................................................10-5 ■ Disk Maintenance.....................................................................................................................10-6 ■ Verifying Disk Status of the Image Reader and Exiting Scanner Cleaning............................10-6

Verifying NAS Status (Optional)...................................................................... 10-7

■ Operation Window....................................................................................................................10-7 ■ Verifying NAS Status.................................................................................................................10-7 ■ Exiting the Verification of NAS Status......................................................................................10-7

FCR PRIMA Console Reference Guide

897N101440B

xi

10.5

10.6

10.7

Verifying Film Printer Status (Optional)......................................................... 10-8

■ Operation Window....................................................................................................................10-8 ■ Verifying Film Printer Status.....................................................................................................10-8 ■ Verifying the Status of the Film Printer.....................................................................................10-8

Verifying Image Output Status ....................................................................... 10-9

■ Operation Window....................................................................................................................10-9 ■ Verifying Image Output Status..................................................................................................10-9 ■ Canceling Image Output...........................................................................................................10-9 ■ Outputting an Image Again that Has Once Failed to be Output.............................................10-9 ■ Exiting the Verification of Image Output Status.......................................................................10-9 Verifying Events (Errors, Warnings or Information).................................. 10-10

■ Operation Windows................................................................................................................10-10 ■ Verifying Events......................................................................................................................10-10 ■ Exiting Events Verification......................................................................................................10-10

Chapter 11 Changing Display Settings of the “Main” Window and “Image Viewer” 11.1

Display Settings of the “Main” Window......................................................... 11-2 11.1.1 Starting up and Shutting Down the Display Setting Function........ 11-2

■ Operation Windows..................................................................................................................11-2 ■ Starting up.................................................................................................................................11-2 ■ Shutting Down...........................................................................................................................11-2

11.1.2 Initial Display Settings of the “Main” Window................................... 11-3

■ Operation Window....................................................................................................................11-3 ■ Setting the Initial Display when Started up..............................................................................11-3

11.1.3 Display Settings of “ID input” Window............................................... 11-4

■ Operation Window....................................................................................................................11-4 ■ Setting Items and Their Order to be Displayed in the “ID input” Window...............................11-4

11.1.4 Display Settings of “Patient List” Window......................................... 11-5

■ Operation Window....................................................................................................................11-5 ■ Setting the Display Contents in the “Patient List” Window......................................................11-5

11.1.5 Display Settings of “FCR-Menus” Window........................................ 11-6

■ Operation Windows..................................................................................................................11-6 ■ Setting the Display Contents in the “FCR-Menus” Window....................................................11-6

11.1.6 Display Settings of Exposing and Reading........................................ 11-7

■ Operation Window....................................................................................................................11-7 ■ Setting Behavior on Selecting an Exposure Menu and on Reading Images with the “Image Viewer”..........................................................................11-7

11.1.7 Display Settings of “Refer” window.................................................... 11-8

xii

FCR PRIMA Console Reference Guide

897N101440B

■ Operation Windows..................................................................................................................11-8 ■ Setting Modalities to be Displayed in the Study List................................................................11-8 ■ Setting the Operation Icons in the Study List...........................................................................11-8 11.1.8 Setting the Font Size in the “Main” Window...................................... 11-9

■ Operation Window....................................................................................................................11-9 ■ Setting the Font Size for the “Main” Window...........................................................................11-9

11.2

Changing Settings of “Image Viewer” Window.......................................... 11-10 11.2.1 Starting the Customize Function....................................................... 11-10

■ Operation Window..................................................................................................................11-10 ■ Starting Up...............................................................................................................................11-10

11.2.2 Initial Settings of Annotation...............................................................11-11

■ Operation Window.................................................................................................................. 11-11 ■ Setting the Display/Tag Display of Annotations..................................................................... 11-11

11.2.3 Setting the Attribute Information........................................................ 11-12

■ Operation Windows................................................................................................................11-12 ■ Setting Image Attribute Information........................................................................................11-12

11.2.4 Setting Reading Protocol.................................................................... 11-13

■ Operation Window..................................................................................................................11-13 ■ Setting the Default Display Layout of each Modality.............................................................11-13

11.2.5 Display Settings of Toolbar................................................................. 11-14

■ Operation Window..................................................................................................................11-14 ■ Setting the Display of Toolbar.................................................................................................11-14

11.2.6 Display Settings of Report.................................................................. 11-15

■ Operation Window..................................................................................................................11-15 ■ Setting the Display of Report..................................................................................................11-15 ■ Creating a Format...................................................................................................................11-16 ■ Adding/Editing Template Tags................................................................................................11-17 ■ Editing Formats.......................................................................................................................11-17

11.2.7 Display Setting of “Image Viewer” Window..................................... 11-18

■ Operation Windows................................................................................................................11-18 ■ Setting Actions during Image Display or Storage..................................................................11-18

11.2.8 Returning to Standard Settings.......................................................... 11-20

■ Operation Window..................................................................................................................11-20 ■ Returning to Standard Settings..............................................................................................11-20

11.2.9 Importing/Exporting Settings............................................................. 11-21

■ Operation Window..................................................................................................................11-21 ■ Import Settings........................................................................................................................11-21 ■ Export Settings........................................................................................................................11-21

FCR PRIMA Console Reference Guide

897N101440B

xiii

Chapter 12 Using “Study Utilities” 12.1

Functions of “Study Utilities”.......................................................................... 12-2

12.2

Starting up and Exiting “Study Utilities”....................................................... 12-5

12.3

■ Operation windows...................................................................................................................12-5 ■ Starting up “Study Utilities”.......................................................................................................12-5 ■ Exiting........................................................................................................................................12-5

Searching for Studies and Creating New Queries....................................... 12-6 12.3.1 Searching for Studies............................................................................ 12-6

■ Operation Window....................................................................................................................12-6 ■ Searching for Studies................................................................................................................12-6 ■ Narrowing down Search...........................................................................................................12-7

12.3.2 Creating New Queries............................................................................ 12-8

■ Operation Windows..................................................................................................................12-8 ■ Creating New Queries..............................................................................................................12-8

12.3.3 Editing Queries..................................................................................... 12-11

■ Operation Windows................................................................................................................12-11 ■ Editing Queries........................................................................................................................12-11

12.4

Manipulating Studies...................................................................................... 12-12

■ Descriptions of Functions.......................................................................................................12-12 ■ Operation Window..................................................................................................................12-13 ■ Starting up the “Image Viewer”...............................................................................................12-13 ■ Reconducting a Study and Editing its Study Information......................................................12-14 ■ Locking and Unlocking Studies..............................................................................................12-14 ■ Completing and Verifying Studies..........................................................................................12-14 ■ Loading Images from a Disk...................................................................................................12-15 ■ Deleting Studies......................................................................................................................12-16 ■ Saving, Printing, and Transmitting Study Images..................................................................12-17

Chapter 13 Configuring “Client Utilities” 13.1

Setting Items of “Client Utilities”.................................................................... 13-2

13.2

Starting up and Exiting “Client Utilities”....................................................... 13-2

13.3

xiv

■ Operation Windows..................................................................................................................13-2 ■ Starting up the “Client Utilities”.................................................................................................13-3 ■ Exiting the “Client Utilities”........................................................................................................13-3

Maintaining Patient Information...................................................................... 13-4

■ Operation Windows..................................................................................................................13-4 ■ Registering New Patient Information.......................................................................................13-4 ■ Searching Patient Information..................................................................................................13-5 ■ Editing Patient Information........................................................................................................13-6

FCR PRIMA Console Reference Guide

897N101440B

■ Deleting Patient Information.....................................................................................................13-6 ■ Finishing Maintenance of Patient Information.........................................................................13-6 13.4

13.5

Verifying the IP Use Frequency....................................................................... 13-7

■ Operation Window....................................................................................................................13-7 ■ Verifying the IP Use Frequency................................................................................................13-7 ■ Finishing Verification of IP Use Frequency..............................................................................13-7

Setting User Information (For Administrator)............................................... 13-8 13.5.1 User Information Maintenance............................................................. 13-8

■ Operation Windows..................................................................................................................13-9 ■ Registering New User Information...........................................................................................13-9 ■ Editing User Information.........................................................................................................13-10 ■ Changing Display Order of Users..........................................................................................13-11 ■ Deleting User Information.......................................................................................................13-11 ■ Exiting User Information Maintenance...................................................................................13-11

13.5.2 Changing User Group Operational Authorities............................... 13-12

■ Descriptions of Functions.......................................................................................................13-12 ■ Operation Windows................................................................................................................13-12 ■ Changing User Group Operational Authorities......................................................................13-13 ■ Exiting the Change of User Group Operational Authorities..................................................13-13

13.5.3 Setting Images for Display in the “User Selection” Window......... 13-14

■ Operation Windows................................................................................................................13-14 ■ Adding Images for Display in the “User Selection” Window.................................................13-14 ■ Deleting an Displayed Image from the “User Selection” Window........................................13-15 ■ Exiting the Setup of Display Image........................................................................................13-15

13.6

13.7

13.8

Editing Information about the Logged-in User........................................... 13-16

■ Operation Windows................................................................................................................13-16 ■ Editing Information on the Logged-in User............................................................................13-16

Verifying History Information........................................................................ 13-17

■ Operation Windows................................................................................................................13-17 ■ Verifying the Login/Logoff History...........................................................................................13-18 ■ Verifying the Study Access History.........................................................................................13-18 ■ Verifying the Image Storage/Print History..............................................................................13-19 ■ Verifying the Backup/Restoration History..............................................................................13-19 ■ Exiting Verification of History..................................................................................................13-19 Disk Maintenance ........................................................................................... 13-20

■ Operation Windows................................................................................................................13-20 ■ Updating Display of Disk Information.....................................................................................13-21 ■ Replacing Disks......................................................................................................................13-21 ■ Setting Disks as “Write Enabled” or “Write Disabled”............................................................13-22 ■ Restoring Disks after Mismatch Occurred.............................................................................13-22 ■ Restoring Information in Disks to Database..........................................................................13-23 ■ Exiting Disk Maintenance.......................................................................................................13-23

FCR PRIMA Console Reference Guide

897N101440B

xv

Chapter 14 Setting “User Utilities” 14.1

Setting “User Utilities”...................................................................................... 14-2

14.2

Starting up and Exiting the “User Utilities”................................................... 14-3

14.3

■ Operation Windows..................................................................................................................14-3 ■ Starting up the “User Utilities”...................................................................................................14-3 ■ Exiting the “User Utilities”..........................................................................................................14-4

Setting the Exposure Menu and Study Menu............................................... 14-5 14.3.1 Setting the Exposure Menu................................................................... 14-6

■ Operation Windows..................................................................................................................14-6 ■ Creating New “Exposure Menu”...............................................................................................14-6 ■ Editing Exposure Menu..........................................................................................................14-13 ■ Deleting Exposure Menu........................................................................................................14-13

14.3.2 Setting Study Menu.............................................................................. 14-14

■ Operation Windows................................................................................................................14-14 ■ Creating New Study Menu.....................................................................................................14-14 ■ Editing Study Menus...............................................................................................................14-17 ■ Deleting Study Menus.............................................................................................................14-17

14.3.3 Setting the Display Menu Group........................................................ 14-18

■ Operation Windows................................................................................................................14-18 ■ Creating New Display Menu Group.......................................................................................14-18 ■ Editing Display Group Menu...................................................................................................14-19

14.4

14.5

14.6

14.7

14.8

14.9

xvi

Setting the Character String Annotated on a Film..................................... 14-21

■ Operation Window..................................................................................................................14-21 ■ Registering/Editing Character String Annotated on a Film...................................................14-21 ■ Deleting Character String to be Annotated on Film...............................................................14-22 ■ Changing Position of a Character String Button....................................................................14-22 ■ Applying Settings to Film Mark...............................................................................................14-22 Setting of Printing, Saving and Transmitting Images................................ 14-23

■ Operation Window..................................................................................................................14-23 ■ Setting the Output Destination...............................................................................................14-23 Auto-Determine Settings................................................................................ 14-27

■ Operation Window..................................................................................................................14-27 ■ Set the Auto-Determine Time Period.....................................................................................14-27 Image Deletion Settings................................................................................. 14-28

■ Operation Window..................................................................................................................14-28 ■ Changing the Time Period for Data Storage.........................................................................14-28 Backup Notification Settings......................................................................... 14-29

■ Operation Window..................................................................................................................14-29 ■ Setting Backup Notification.....................................................................................................14-29

Automatically Startup Settings..................................................................... 14-30

FCR PRIMA Console Reference Guide

897N101440B

■ Operation Window..................................................................................................................14-30 ■ Client Application Startup Settings.........................................................................................14-30 14.10

Prefix Display Settings................................................................................... 14-31

■ Operation Window..................................................................................................................14-31 ■ Setting Patient Prefix..............................................................................................................14-31

14.11

Setting Image Incidental Information Display Contents........................... 14-32

14.12

Loading Patient Information.......................................................................... 14-33

14.13

14.14

14.15

14.16

14.17

■ Patient Information Input Format............................................................................................14-33 ■ Creating a File of Patient Information to Import.....................................................................14-34 ■ Importing Patient Information.................................................................................................14-36 Media Consistency Check............................................................................. 14-37

■ Operation Windows................................................................................................................14-37 ■ Media Consistency Check......................................................................................................14-37 ■ Result and Remedy................................................................................................................14-38 History Information Consistency Check..................................................... 14-40

■ Operation Window..................................................................................................................14-40 ■ History Information Consistency Check.................................................................................14-40 Automatically Logoff....................................................................................... 14-41

■ Operation Window..................................................................................................................14-41 ■ Setting Automatic Logoff.........................................................................................................14-41 Adding/Deleting User Marker........................................................................ 14-42

■ Operation Windows................................................................................................................14-42 ■ Registering User’s Markers....................................................................................................14-42 ■ Removing User’s Markers......................................................................................................14-42 Preset Setting................................................................................................... 14-43

■ Operation Window..................................................................................................................14-43 ■ Assigning Preset.....................................................................................................................14-43

Chapter 15 Troubleshooting 15.1

PC does Not Start.............................................................................................. 15-2

15.2

PC is Inoperative............................................................................................... 15-3

15.3

PC does Not Shut down................................................................................... 15-4

15.4

Event Indicator is Blinking............................................................................... 15-5

15.5

Error Message is Displayed............................................................................. 15-6

15.6

Image is Streaked.............................................................................................. 15-7

15.7

Message “Not Logged In” is Displayed when Trying to Operate the Screen................................................................ 15-8

15.8

Window Disappears.......................................................................................... 15-9

FCR PRIMA Console Reference Guide

897N101440B

xvii

15.9

Disk cannot be Removed............................................................................... 15-10

15.10

Toolbar is Not Displayed on “Image Viewer”.............................................. 15-11

15.11

Study was Scheduled or Registered but Patient List is Unchanged...... 15-12

15.12

Unintended Studies Appear on Patient List................................................ 15-13

15.13

Unable to Print or Save Images from “Image Viewer”.............................. 15-14

15.14

The “Webpage unavailable while offline” Dialog is Appeared................. 15-15

Chapter 16 Daily Inspection and Maintenance 16.1

Daily Inspections and Maintenances............................................................. 16-2

16.2

Backing up Information of Patient, Setting, Log, and Report.................... 16-3

■ Operation Windows..................................................................................................................16-3 ■ Starting up the “Backup Utilities”..............................................................................................16-4 ■ Backing up.................................................................................................................................16-5 ■ Restoring...................................................................................................................................16-6 ■ Collecting Maintenance Log.....................................................................................................16-7 ■ Shutting down the “Backup Utilities”........................................................................................16-8

16.3

Maintenance Record Sheet.............................................................................. 16-9

16.4

Maintenance and Inspection.......................................................................... 16-10

■ Maintenance and Inspection Items Assigned to Specified Dealer.......................................16-10

Appendix A Main Specifications A.1

A.2

xviii

Specifications......................................................................................................A-2

■ PC............................................................................................................................................... A-2 ■ Standard Monitor........................................................................................................................ A-3 ■ DIN Standard-Compliant Monitor.............................................................................................. A-4

External View and Weight..................................................................................A-5

■ PC............................................................................................................................................... A-5

FCR PRIMA Console Reference Guide

897N101440B

Appendix Z Precautions for Exposure Z.1

Precautions to Take Before Exposure............................................................. Z-2

Z.2

Precautions to Take in Each Mode................................................................... Z-3

Z.3

Z.2.1

Precautions for Exposure in the AUTO MODE.................................... Z-3

Z.2.2

Precautions for Exposure in the SEMI-AUTO MODE.......................... Z-6

Z.2.3

Precautions for Exposure in the FIX MODE......................................... Z-6

Other Precautions............................................................................................... Z-7

Index

FCR PRIMA Console Reference Guide

897N101440B

xix

xx

FCR PRIMA Console Reference Guide

897N101440B

1

1

Safety Information

Chapter

Safety Information

1.1

Precautions Before Operating This Equipment................ 1-2

1.2

Notes on Using Medical Electrical Equipment (for Safety Assurance and Risk Prevention)..................... 1-3

1.3

Safety....................................................................................1-4

1.4

Precautions in Using the FCR PRIMA Console................. 1-7

1.5

Labels..................................................................................1-12

FCR PRIMA Console Reference Guide

897N101440B

1-1

1.1

1 Safety Information

Precautions Before Operating This Equipment

Before using this equipment, please read “Precautions Before Operating This Equipment” carefully so that you can operate it correctly. Whenever you operate this equipment, be sure to observe those precautions. Failure to do so may cause injuries or property damages.

The institution where the equipment is installed is responsible for its use and maintenance. In addition, this equipment should not be used by persons other than doctors or suitably trained staff.

Process waste correctly, as stipulated by local law or any regulations that apply.

1-2

FCR PRIMA Console Reference Guide

897N101440B

1.2

Notes on Using Medical Electrical Equipment (for Safety Assurance and Risk Prevention)

1 Safety Information

Please read through these notes before using the equipment.

● Notes on Using Medical Electrical Equipment (for Safety Assurance and Risk Prevention) 1 Allow only skilled personnel to use the equipment. 2 Take notice of these precautions in installing the equipment: (1) Locate the equipment to guard against splashes of water. (2) Locate the equipment to protect against any ambient conditions that could adversely affect its performance, such as atmospheric pressure, temperature, humidity, ventilation, sunlight, dust and air containing salt or sulfur content. (3) Locate the equipment to keep it stable, protected from tilting, vibration, impact (including impact during transit) and so on. (4) Do not install the equipment where chemicals are stored or in or near sources of gases. (5) Observe the line frequency and voltage and the allowable current (or power consumption). (6) If the equipment is powered from batteries, check their status (standing, polarities and so on). (7) Connect the ground wire properly. 3 Before using the equipment, take notice of these precautions: (1) Check the switches for contact, polarities and dial settings, the meters and so on to ensure that the equipment works properly. (2) Check that the ground wire is fully connected. (3) Check that all cords are connected correctly and fully. (4) Remember fully that combined use of equipment could cause incorrect diagnoses, unsafety, or else. (5) Recheck the external circuits that are directly connected to the patient. (6) If the equipment is powered from batteries, check the batteries. 4 While the equipment is in operation, take notice of these precautions: (1) Be careful not to exceed the specified time and dosage limits for diagnostics or treatment. (2) Keep a constant eye on the equipment in general and the patient to ensure that they remain normal. (3) Whenever the equipment or the patient is found abnormal, take appropriate remedial action, such as shutting down the equipment with care to cause no harm to the patient. 5 After using the equipment, take notice of these precautions: (1) Reset the switches, dials and other controls to their initial status in specified routines before turning off the equipment. (2) Do not use undue force in disconnecting cords, as by pulling them out of position. (3) Take notice of these precautions in storing the equipment: I Store the equipment to guard against splashes of water. II Store the equipment to protect against any ambient conditions that could adversely affect its performance, such as atmospheric pressure, temperature, humidity, ventilation, sunlight, dust and air containing salt or sulfur content. III Store the equipment to keep it stable, protected from tilting, vibration, impact (including impact during transit) and so on. IV Do not store the equipment where chemicals are stored or in or near sources of gases. (4) Clean the components, cords, directors and so on and then sort and keep them in order. (5) Keep the equipment clean to prepare for the next sessions of operation. 6 If the product fails, do not tamper with the product but mark it properly and have it repaired by professionals. 7 Do not make unauthorized modifications to the equipment. 8 Maintenance and inspection (1) Inspect the equipment and parts periodically. (2) Before putting the equipment back to reuse after it has been left out of service for a certain length of time, check that the equipment works normally and safely. 9 Other Abide by instructions in the Operation Manual for proper equipment handling. FCR PRIMA Console Reference Guide

897N101440B

1-3

1.3

1

Safety

Safety Information

Read this section before using the equipment to ensure proper handling.

■ Preventing Explosion WARNING Because this equipment is not explosion-proof, do not use combustible and explosive gases near the equipment.

■ Preventing Electrical Shock WARNING To prevent electrical shock, observe these precautions by all means. ● Never open the equipment covers. Hands touching a high-voltage zone inside may cause electrical shock. ● Install the equipment where it will not be splashed with water. ● Check that the equipment is fully grounded. ● Check that all cords are connected correctly and safely. ● While this equipment is in operation, pay due attention not to have the patient come into direct contact with it.

■ Connection Instruction CAUTION Use external devices of specified types to connect to the equipment.

1-4

FCR PRIMA Console Reference Guide

897N101440B

■ Electromagnetic Compatibility Instruction

1

This equipment complies with the following electromagnetic compatibility (EMC) standard by comforting to the requirements of IEC 60601-1-2:2001+A1:2004/EN 60601-1-2:2001+A1:2006.

Safety Information



EMC Standard: CISPR 22:2005/EN 55022:2006



CISPR 24:1997 + A1:2001 + A2:2002/EN 55024:1998+A1:2001+A2:2003



IEC 61000-3-2:2005/EN 61000-3-2:2006







IEC 61000-3-3:1994 + A1:2001 +A2:2005

However, there is no guarantee that harmful electromagnetic interference will not occur in any installation environment. This equipment generates, uses and can radiate radio frequency energy. In a particular environment, e.g., the equipment is not installed and used in accordance with the instructions or any device that is not compliant with the EMC Standard is used in a surrounding area, harmful interference may result, causing the equipment or surrounding device to malfunction. If this equipment causes harmful interference to other equipment or receives harmful interference from other equipment, you are encouraged to try to correct the interference by one or more of the following measures: ● Reorient or relocate the equipment. ● Increase the separation between the equipment and the affected equipment. ● Connect the equipment into an outlet on a circuit different from that to which the affected equipment is connected. If the problem persists, discontinue the use of the equipment and consult the manufacturer of the affected equipment or our official dealer.

CAUTION Keep the equipment away from sources of strong electromagnetic radiation.

Guidance and manufacturer’s declaration - electromagnetic emissions The CR-IR 391CL is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-IR 391CL should assure that it is used in such an environment. Emissions test Noise terminal voltage CISPR22 Electric field noise strength CISPR22

Compliance Class B

Class B

Harmonic emissions EN61000-3-2 IEC61000-3-2 Voltage fluctuations/ flicker emissions EN61000-3-3 IEC61000-3-3

Class D

Complies

FCR PRIMA Console Reference Guide

897N101440B

1-5

Guidance and manufacturer’s declaration - electromagnetic immunity

1

The CR-IR 391CL is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-IR 391CL should assure that it is used in such an environment.

Safety Information

Immunity test

EN/IEC 60601 test level

Compliance level

±2kV contact ±4kV contact ±2kV air ±4kV air ±8kV air

±2kV contact ±4kV contact ±2kV air ±4kV air ±8kV air

±1kV for power supply lines

±1kV for power supply lines

EN61000-4-4 IEC61000-4-4

±0.5kV for input/output lines

±0.5kV for input/output lines

Surge

±0.5, ±1kV differential mode

±0.5, ±1kV differential mode

±0.5, ±1.0, ±2.0kV common mode

±0.5, ±1.0, ±2.0kV common mode

95% dip in UT) for 0.5 cycle

95% dip in UT) for 0.5 cycle

70% UT (30% dip in UT) for 25 cycles

70% UT (30% dip in UT) for 25 cycles

95% dip in UT) for 250 cycles

95% dip in UT) for 250 cycles

1 A/m

1 A/m

Electrostatic discharge (ESD) EN61000-4-2 IEC61000-4-2 Electrical fast transient/burst

EN61000-4-5 IEC61000-4-5

Voltage dips, short interruptions and voltage variations on power supply input lines EN61000-4-11 IEC61000-4-11

Power frequency (50/60Hz) magnetic field EN61000-4-8 IEC61000-4-8

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacturer’s declaration - electromagnetic immunity The CR-IR 391CL is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-IR 391CL should assure that it is used in such an environment.

1-6

Immunity test

EN/IEC 60601 test level

Compliance level

Conducted RF EN61000-4-6 IEC61000-4-6

3 Vrms 150 kHz to 80 MHz

3 Vrms

Radiated RF EN61000-4-3 IEC61000-4-3

3 V/m 80 MHz to 1 GHz

3 V/m

FCR PRIMA Console Reference Guide

897N101440B

1.4

Precautions in Using the FCR PRIMA Console

1 Safety Information

■ Precautions when Using ● Operating Precautions

• When working steadily on the FCR PRIMA Console (CR-IR 391CL) over an extended period of time, take breaks for 10 to 15 minutes after every hour to rest your eyes and hands. • Verify that time and date which displayed on the “Indicator” window is correct before using the FCR PRIMA Console (CR-IR 391CL). Time and date which displayed on the “Indicator” window are recorded as a time and date at the time of saving images to the disk. • Turn the power of this equipment or the computer installed this software off at least once a day with the shutting down operation. The FCR PRIMA Console (CR-IR 391CL) is not built for continuous running. • Use a screensaver when displaying the same image for extended period of time. If your screensaver fails to start up, consult our official dealer. If the same image is displayed for a long time, the burn-in or residual image of the previous image may result when switching the display on the monitor. • When selling this software alone, keep the CD cover of this software in a safe place because important matters such as production number are written on it.

● Hardware Precautions

• Be careful not to splash the FCR PRIMA Console (CR-IR 391CL) and its peripheral devices with water and other liquids. Do not handle the equipment with wet hands. • Before disposing of the FCR PRIMA Console (CR-IR 391CL), erase patients’ personal information stored on it completely. • Do not turn off the PC or DVD drive while FCR PRIMA Console (CR-IR 391CL) is in use. Images saved on the PC or DVD drive might be damaged. • Do not handle the FCR PRIMA Console (CR-IR 391CL) with Windows Explorer running. The FCR PRIMA Console (CR-IR 391CL) or PC may freeze. • When connecting FCR PRIMA Console and other study equipment, use the UTP (Unshield Twist Pair) cable. • Do not disconnect the network cable of the hub or the FCR PRIMA Console (CR-IR 391CL) from the port while image reading is in progress. If the network cable is accidentally disconnected, reinsert it into the port immediately. Do not shut down the FCR PRIMA Console (CR-IR 391CL) at this time. If you disconnect the network cable of the hub or the FCR PRIMA Console (CR-IR 391CL) from the port and insert it back into position while image reading is in progress, the image reading process, might take about 10 minutes to resume. If you shut down the FCR PRIMA Console (CR-IR 391CL) or Image Reader, the images being read might be lost. • The FCR PRIMA Console (CR-IR 391CL) might be accidentally powered off due, for example, to a power failure while image reading is in progress. When this happens, restart the PC and log in again to reload the images. Make sure that the correct images have been read in response to your exposure menu choice. After starting up and logging in the FCR PRIMA Console (CR-IR 391CL), when double-clicking and starting to expose the study which was reading images at the time of shutting down from the patient list on the “Main” window, images can be started to read again. If the images loaded do not match your exposure menu choice, exchange or otherwise manipulate the images to synchronize the exposure menu choice and the images with each other.

FCR PRIMA Console Reference Guide

897N101440B

1-7

• Before you use an uninterruptible power supply with the FCR PRIMA Console (CR-IR 391CL), verify its performance characteristics, such as the battery life. Replace the battery as required. If the FCR PRIMA Console (CR-IR 391CL) is forced to shut down while it is in operation due to electric power failure or other causes, data could be lost. You need to have a security, backup or any other relevant procedure programmed separately to save images safely. Use of an uninterruptible power supply is recommended to protect against emergencies. • Do not connect any additional hardware to the FCR PRIMA Console (CR-IR 391CL) in addition to the hardware already connected to it. It will result in malfunction. •Do not connect the FCR PRIMA Console (CR-IR 391CL) to a wide area network. It will result in malfunction. • If you need to install this software to the store-brought PC, contact our authorized dealer. Also, use our company’s recommended computer for installing this software. Contact our authorized dealer for the detailed specifications of the computer and the peripheral equipment. • Use a commercial personal computer and monitor that passed our compatibility test with FCR PRIMA Console (CR-IR 391CL). If non-approved ones are used, Fujifilm Corporation does not guarantee that the FCR PRIMA Console (CR-IR 391CL) will operate normally. A commercial personal computer and monitors verified for compatibility are listed in Appendix A.

1 Safety Information

● Software Precautions

• Do not modify the Windows setting of the FCR PRIMA Console (CR-IR 391CL). Image input/output might be disabled. • While using the FCR PRIMA Console (CR-IR 391CL), do not double-click any file associated with any other application software. If you attempt to open a file associated with any other application software by double-clicking it, the file would take a considerable time to open. If it is absolutely necessary to open that file, start up the associated application software first and then control the application software to open the target file. • Do not install any software in the FCR PRIMA Console (CR-IR 391CL) in addition to the preinstalled software. Do not uninstall any preinstalled software. Because the FCR PRIMA Console (CR-IR 391CL) comes with certain software preinstalled, installing additional software could interfere with the preinstalled software, resulting in malfunction. • Operate the disk drive with the FCR PRIMA Console (CR-IR 391CL) applications. • When selling this software alone, Fujifilm Corporation will not be responsible for any malfunction or damage of the personal computer or the monitor.

■ Precautions when Installing • Do not install the FCR PRIMA Console (CR-IR 391CL) monitor in a place that is exposed to direct or reflected intense light, such as sunlight or other bright illumination. If the monitor is installed at or near sources of intense light, images may not be read correctly. • Do not install the FCR PRIMA Console (CR-IR 391CL) in a dusty place, such as on a floor. If it is unavoidable to install the FCR PRIMA Console (CR-IR 391CL) close to a floor, keep it at least 5 cm above the floor. If the FCR PRIMA Console (CR-IR 391CL) is installed in a dusty environment, dust accumulating in the power supply unit or other components could result in smoke or fire. • Locate the FCR PRIMA Console (CR-IR 391CL) unit so noises emitted from the unit while in operation will not adversely affect the performance of medical examinations. • When relocating the FCR PRIMA Console (CR-IR 391CL) or the computer installed this software, contact our authorized dealer.

1-8

FCR PRIMA Console Reference Guide

897N101440B

■ Precautions when Reading Images

1 Safety Information

• Reading of images on the FCR PRIMA Console (CR-IR 391CL) monitor should be carried out at the responsibility of a user who is conscious of the intended use of the images and inherent differences from film-based diagnostics. • Verify the information that accompanies images and use it to aid in their reading. Pay particular attention to patient information. • When reading an image, be sure to verify the image flip information (front or back) and the actual exposure condition of patient. • Configure the image processing parameters for the FCR PRIMA Console (CR-IR 391CL) to suit display images. Improperly configured image processing parameters could make correct image reading difficult. • Adjust the FCR PRIMA Console (CR-IR 391CL) to display images in a proper size. The FCR PRIMA Console (CR-IR 391CL) may subject original images to interpolation processing (display processing) to enable you to view the images in an optional display size. As a result, images displayed in an improper size may not impart their intrinsic information. • Multi-frame images may not be displayed in the set frame rate in case of cine play mode. Confirm images after confirming the display of measured value of the frame rate.

● Monitor-related Precautions

• Luminance and contrast can be changed depending on the monitor types but never change them. If it changed, an image may not be verified properly. • Use the FCR PRIMA Console (CR-IR 391CL) with proper screen contrast (high/low) settings. Improper screen contrast (high/low) settings could make image reading difficult. • Before reading images on FCR PRIMA Console (CR-IR 391CL), make sure that the monitor is set at the right angle for viewing. If the monitor is not set at the right angle, adjust it. An improperly angled monitor could make it difficult to read images properly. • Do not change the display setting (for portrait or landscape) of the LCD monitor that was set at time of installation. If you need to change the display orientation setting, contact our authorized dealer.

FCR PRIMA Console Reference Guide

897N101440B

1-9

■ Precautions when Managing Data

1 Safety Information

• Print and retain original images on a film (called an “original film”). The hard disk or disk mounted on the FCR PRIMA Console (CR-IR 391CL) unit is not intended for permanent storage of diagnostic images.

● Media Management

• Do not leave disks for the FCR PRIMA Console (CR-IR 391CL) in dusty places, hot or damp places, places exposed to direct sunlight or in places that are subjected to sharp temperature differences. • Duplicate (copy) every disk to which images are stored. Check disks for consistency by following the instructions found in “13.13 Media Consistency Check”. Images stored to the FCR PRIMA Console (CR-IR 391CL) hard drive or disks could be lost as a result of equipment or media faults. Further, some characteristics of a disk could make images stored on the disk inaccessible for reading or writing if the disk has deteriorated over an extended period of usage, even though it has been stored properly. An improperly stored disk could make images stored on it inaccessible for reading or writing within months. Neither FUJIFILM Corporation nor the vendor assumes any liability or responsibility for any loss of images and data.

● Image Data Management

• Images that have been saved as “Processed” being selected can be reloaded into the FCR PRIMA Console (CR-IR 391CL), but cannot be printed on a film (optional). • Before deleting images on FCR PRIMA Console (CR-IR 391CL), make sure that there is no problem if the images are deleted. Although FCR PRIMA Console (CR-IR 391CL) allows you to delete images, if you delete required images by mistake, you would have to make a re-exposure of the subject and recapture the images.

■ Precautions when Processing Image • Evaluation of the results of measurement performed on the FCR PRIMA Console (CR-IR 391CL) is your responsibility. Images used for measurement on the FCR PRIMA Console (CR-IR 391CL) may contain errors in relation to the actual length of the subject, depending on how the images have been captured and other conditions. • When specifying coordinates on the FCR PRIMA Console (CR-IR 391CL) monitor, specify them correctly. Incorrectly specified coordinates will deliver incorrect measurement results.

1-10

FCR PRIMA Console Reference Guide

897N101440B

■ Precautions when Exposing

1 Safety Information

• When exposing an area where it is difficult to distinguish the exposed direction and the symmetric body part, mark it using markers for example to avoid the confusion of the exposed direction and body part. • Erase manually a remaining image on an imaging plate exposured at large X-ray dose. Incomplete erasing of an image on the imaging plate could make image reading difficult.

■ Precautions when Saving Data ● Common

• When importing image or study data into FCR PRIMA Console (CR-IR 391CL) (optional) or saving from FCR PRIMA Console (CR-IR 391CL) to media such as CD, be very careful to check the image or study data and patient information.

● Media

• Do not use the disks that are dedicated to image storage on the FCR PRIMA Console (CR-IR 391CL) for any other use. Images stored on those disks would be made inaccessible for browsing. • Use disks that meet the requirements listed in “2.7 Before Using Disks”. • Be careful not to confuse the DVD disk with the CD-R or DVD-R disk for PDI because the CD-R or DVD-R disk for PDI to be given to the patient is created using the DVD drive to create DVD disks. When creating a CD-R or DVD-R disk for PDI, write information (patient name, patient ID, birth date, disk creation date, study date, facility name) on the front of the CD-R or DVD-R disk in order to distinguish it from DVD+RW/-RW disk. When giving the created disk for PDI to the patient, advise the patient that the contents of the disk is for reference. • Do not handle the FCR PRIMA Console (CR-IR 391CL) in a Windows Explorer operation, unless otherwise specified in this Reference Guide. If you eject a disk in a Windows Explorer operation, the data stored on the disk could be corrupted as a consequence of the disk having been ejected while images were being written to or read from it, or your application might not be able to eject it. Further, when you browse through information stored on the disk, the information might take time before it shows up.

● NAS

• When the caution about lack of HDD space is displayed, add NAS. In such cases, consult our authorized dealer. • When an error of NAS is displayed in the indicator for Event, consult our authorized dealer.

■ Precaution when Outputting • Be careful not to confuse the patient when printing several images on a film for several patients and reading of images on the film should be carried out at the responsibility of a user who is conscious of the intended differences from the image quality by displayed image size. • Be careful that the height displays and outputs in meters, and the weight displays and outputs in kilograms.

FCR PRIMA Console Reference Guide

897N101440B

1-11

1.5

1

Labels

Safety Information

The FCR PRIMA Console (CR-IR 391CL) labels are shown below.

■ FCR PRIMA Console (CR-IR 391CL) Labels

2010-06

Sample year of manufacture

The following symbols are used on this equipment. Their meanings are described below. This symbol indicates compliance of the equipment with Directive 93/42/EEC. This symbol indicates that this product is not to be disposed of with your household waste, according to the WEEE Directive (2002/96/EC) and your national law. This product should be handed over to a designated collection point. Improper handling of this type of waste could have a possible negative impact on the environment and human health due to potentially hazardous substances that are generally associated with EEE. At the same time, your cooperation in the correct disposal of this product will contribute to the effective usage of natural resources. For more information about waste, please contact FUJIFILM dealers.

1-12

FCR PRIMA Console Reference Guide

897N101440B

2

Product Overview

Chapter

2

Product Overview

2.1

FCR PRIMA Console Highlights......................................... 2-2

2.2

Name and Window of Each Part......................................... 2-5

2.3

Flow for Utilizing the System.............................................. 2-6

2.4

Version Information............................................................. 2-8

2.5

Starting up and Exiting........................................................ 2-9 2.5.1

Starting up FCR PRIMA Console................................... 2-9

2.5.2

Exiting FCR PRIMA Console........................................ 2-12

2.6

Protecting the PC when Temporarily Leaving the Site... 2-15

2.7

Before Using Disks............................................................ 2-16

2.8

Setting up Disks................................................................. 2-19

FCR PRIMA Console Reference Guide

897N101440B

2-1

2.1

FCR PRIMA Console Highlights

This section describes the features of FCR PRIMA Console and an example of its system configuration.

2 Product Overview

■ Main Features of FCR PRIMA Console The FCR PRIMA Console is a diagnostic imaging workstation that not only simplifies the workflow ofdiagnosing X-ray images, but also offers a fully integrated solution to managing modality, as well as X-ray,study images. X-ray images captured by the image reader are transmitted to the FCR PRIMA Console (CR-IR 391CL). Transmitted or captured images are displayed on the FCR PRIMA Console (CR-IR 391CL) monitor screen, which provides information to aid in reading the images. A summary description of the key features of the system is as follows: 1. Digital image enhancements make X-ray images easier to view and better suited for diagnostic purposes. 2. The FCR PRIMA Console (CR-IR 391CL) lets you enter patient information and view images. For X-ray studies, you can enter exposure information (such as exposed body parts and exposure menus) , as well as patient information. 3 Image data can be backed up to DVD. 4. The FCR PRIMA Console (CR-IR 391CL) keeps images obtained from various studies under fully integrated management. 5. Study images can be saved on CD-R or DVD-R in the PDI format and made available to the patients for collaborative diagnostic purposes.

2-2

FCR PRIMA Console Reference Guide

897N101440B

■ System Configuration Example Network connections allow FCR PRIMA Console to connect to an Image Reader and read FCR study images in. FCR PRIMA Console is used from start of the FCR study until interpretation of the radiogram. FCR PRIMA Console can be configured on a single personal computer An example of a system configuration is shown below. Optional functions depend on whether the option is installed.

2 Product Overview

Image Reader

Imager

PC FCR PRIMA Console (CR-IR 391CL)

Hub

NAS

FCR PRIMA Console Reference Guide

897N101440B

2-3

■ Restrictions ● Restrictions of Study Images Handled by FCR PRIMA Console

Types of images that are handled by FCR PRIMA Console are listed below. Operation methods and usable functions may differ depends on types of study images.

2 Product Overview

Types of study image

2-4

Explanations

Unprocessed FCR study images

X-ray images created by FUJIFILM CR equipment have the FUJIFILM-unique image processing parameters added on the image data. Unprocessed FCR images are images on which only the parameters were adjusted to enable image editing such as changing gradation (contrast or density) and reversing the display for obtaining a better view on screen or film, while keeping the original image data as is. Therefore, the original images before editing can be restored. These images cannot be displayed in the DICOM viewer.

Processed FCR study images

X-ray images created by FUJIFILM CR equipment have the FUJIFILM-unique image processing parameters added on the image data. The Processed FCR images are images whose original data was processed along with the parameter change. Therefore the original images cannot be restored. Images can be viewed in the DICOM viewer.

Other than FCR study images

X-ray images loaded from a DICOM study disk created in another facility.

FCR PRIMA Console Reference Guide

897N101440B

2.2

Name and Window of Each Part

This section describes the name and display window of each parts for FCR PRIMA Console main unit.

2 Product Overview

■ Display Window By installation conditions for option or by system setting at the time of installation, there are buttons and menu which cannot be selected. Buttons and menus which cannot be selected are displayed in gray or are hidden. Image Viewer window

Main window

Patient information

Indicator window

Study summary Patient information edit button

Operation for Main window

Study pane Study status ([Suspend] [Verify])

Input and refer to patient information

Sub-window

Study operations FCR PRIMA Console (CR-IR 391CL) menu bar for operation

Image cart

Thumbnail images

FCR PRIMA Console Reference Guide

897N101440B

2-5

2.3

Flow for Utilizing the System

This section divides FCR PRIMA Console functions into “Regular Tasks”, “Semi-Regular Tasks”, and “Non-Regular Tasks” and describes the flow of utilizing these functions.

2 Product Overview

Regular Tasks Take and Read X-ray Images Chapter 3 Taking and Reading X-ray of a Subject Patient Registration

Exposure

Read

Search Patient and Refer Studies Chapter 4 Searching Patients and Referring Studies

Search

Refer

Semi-Regular Tasks Edit Study Information

Verifying by Viewing Indicators

Chapter 9 Editing Study Information

Chapter 10 Verifying with the Indicators

Re-Exposure

Print/Storage Settings

Edit Patient Information ...etc

2-6

FCR PRIMA Console Reference Guide

897N101440B

Chapter 11 Changing Display Settings of the “Main” Window and “Image Viewer”

Modify Study Menu Inactivate Images

Change Display Settings

Verify Unit Status, Output Errors, Alerts... etc

Buttons on “Main Menu” / “Image Viewer” and Change of Display Settings

2 Product Overview

Verify and Process Images

Using Report Function Chapter 6 Using Report Function

Chapter 5 Verifying and Processing Images

Schema

Process Images

Thumbnail Comment

Verify Images

Save Images from Each Windows Chapter 7 Image Storage from Each Window Storage Disks

DICOM Server

Work Disks .jpg

PDI Disks

.bmp

General-Purpose File Format

Print Images Chapter 8 Image Print from Each Window 富士フイルム病院

Film Print

04001676 A0000A020

富士フイルム病院

04001676 A0000A020

Report Print G 1.0G#0.7+0.50R5Q0.5 頭蓋概観

L 4.0S200C*1.2*1.0AP

01

外科 Surgi

G1.0G#0.7+0.50R5Q1.0

R->L 0000000001 フジ ハナコ

撮影日1992.

11. 27 [17:43]

[女] 23才 縮率:50%

RT-04

Free Layout Print G1.0G#0.7+0.50R5Q1.0

R->L

G 1.0G#0.7+0.50R5Q0.5 頭蓋概観

4.0S200C*1.2*1.0AP

外科 Surgi

0000000001 フジ ハナコ

撮影日1992.

11. 27 [17:43]

[女] 23才 縮率:50%

RT-04 L

01

Non-Regular Tasks Study Utilities Chapter 12 Study Utilities Delete Patient Information/Studies

Save Images... etc

Settings of Client Utilities Chapter 13 Configuring “Client Utilities” Patient Information Maintenance

Settings of User Utilities Chapter 14 Setting “User Utilities”

Edit/Add Exposure Menu and Study Menu

Set User Information

Confirm History Information Verify History Information

Environment Settings of FCR PRIMA Console

Handle Disks etc

FCR PRIMA Console Reference Guide

897N101440B

2-7

2.4

Version Information

Verifies the version of the FCR PRIMA Console (CR-IR 391CL) application software and the options installed.

2 Product Overview

■ Operation Windows

■ Checking Version Information

● “Main” Window

1. Select

- [Version] (A).

→→The “Version” window opens.

2. Check the version (B) and the options installed (C).

3. Click [OK] (D) after completion →→The “Version” window exits. (A)

● “Version” Window (B) (C)

(D)

The displayed information in the window differs depending on the version of the application software.

2-8

FCR PRIMA Console Reference Guide

897N101440B

2.5 2.

Starting up and Exiting

This section describes the workflow of starting up and exiting FCR PRIMA Console.

2

2.5.1 Starting up FCR PRIMA Console

Product Overview

This section also describes the normal startup procedures, each procedure for starting up the server application and client application after backing up from the backup notification message, and procedures for starting up each function from the server application icon.

■ Operation Device / Operation Windows

● “Select Server Application Behavior” Dialog

● Monitor and Main Unit

(H)

(A)

(I) (B)

(C)

● “Login” Window (User Selection) (D)

(E)

● “Login” Window (Enter a Password)

(F)

(G)

FCR PRIMA Console Reference Guide

897N101440B

2-9

■ Starting up

4. Select a user (D) to log in to FCR PRIMA Console.

Starts up FCR PRIMA Console.

1. Turn on the monitor.

2

• Guidance that leads you through the operating process is displayed in the “User Selection” window and “Password Entry” window.

Product Overview

2. Turn on the server PC power (B). →→The startup window of FCR PRIMA Console opens, and then the server application starts up. displays in the task The server application icon tray at the lower right-hand corner of the screen. Counter-measure when the application does not start: “■ Starting up the Server Application” →→After the server application starts up, the client application starts up. Counter-measure when the application does not start: “■ Starting up the Client Application” →→When the client application starts up, “Login” window (user selection) opens.

3. Turn on the Image Reader and printer (optional). • To connect with the Image Reader, start up FCR PRIMA Console before turning on the Image Reader. • For information about starting up the peripheral devices connected to the FCR PRIMA Console system, refer to the Operation Manual pertaining to the peripheral devices.

• FCR PRIMA

• Users displayed in the “Login” window (user selection) can be newly added, edited and deleted by using the “Client Utilities”. New registration, deletion, editing of the user : ”13.5 Setting User Information (For Administrator)” →→The “Login” window (enter a password) opens.

5. Enter a password (F). 6. Click [Login] (G). →→FCR PRIMA Console starts up.

• If you have not backed up the disk for a certain period of time or longer, FCR PRIMA Console may display a backup prompt message before it starts up. Click [OK] in response to the backup prompt message and the “Backup Utility” will launch to perform the requested backup operation. Back up notification: “14.8 Backup Notification Settings” Backup: “16.2 Backing up Information of Patient, Setting, Log, and Report” When the backup operation is completed, start up the server application and client application again. “■ Starting up the Server Application” “■ Starting up the Client Application” • With the server application running, you can shut down, restart Windows OS and log off from Windows OS.

Power switch

I. Turn on the main power switch of Image Reader. → Image Reader starts up. • DRYPIX 2000

Power button



I. Press the power button. → Printer starts up.

2-10

FCR PRIMA Console Reference Guide

897N101440B

■ Starting up the Server Application ■ Starting up Each Function from Starts up each server application separately after performing backup in response to backup notification message or when FCR PRIMA Console is terminated while the PC is running. on the desktop.

When the FCR PRIMA Console server application starts up, it displays server application icon in the task tray in the lower right-hand corner of the screen.

2

→→The startup window of FCR PRIMA Console opens, and then the server application starts up.

1. Double-click the server application icon

• On the PC, the auto startup for the client application can be set.

2. Select the operation to be performed from the

“14.9 Automatically Startup Settings” • You can do the same by selecting [Start] - [All Programs] (P) - [Fuji Film] - [FCR Application].

Product Overview

1. Double-click

the Server Application Icon

.

→→The “Select Server Application Behavior” dialog appears.

pull-down menu ( I ), and then click [OK] (K). Selection Items Items

Contents

Shutdown the PC Shuts down the PC.

■ Starting up the Client Application If no operation window (“User Selection” window) is displayed while the server application icon appears in the PC’s task tray, starts up the client application.

1. Double-click

on the desktop.

→→The client application starts up. When the client application starts up, the “User Selection” window opens.

“2.5.2 Exiting FCR PRIMA Console” Exit

Exits the “Server Application”. “2.5.2 Exiting FCR PRIMA Console”

Start up Client Application

Starts up the “Client Application”.

User Utilities

Starts up the “User Utilities”. “Chapter 14 Setting “User Utilities””

Backup Utilities

“16.2 Backing up Information of Patient, Setting, Log, and Report”

• On the PC, the auto startup for the client application can be set. “14.9 Automatically Startup Settings” • You can do the same by the following steps: 1) Double-clicking server application icon in the task tray at the lower right-hand corner of the screen.

Disk Copy

Starts up the “disk copy tool”. “2.8 Setting up Disks”

Event Display

3) Clicking [OK].

Displays the “Event Log” window. “10.7 Verifying Events (Errors, Warnings or Information)”

→The “Select Server Application Behavior” dialog appears. 2) Selecting “Start up Client Application” from the pull-down menu.

Starts up the “Backup Utilities”.

Version Display

Displays the version. “2.4 Version Information”

FCR PRIMA Console Reference Guide

897N101440B

2-11

2.5.2 Exiting FCR PRIMA Console Exits FCR PRIMA Console with the following methods: • Log off the current user Exits FCR PRIMA Console to change its users.

2 Product Overview

• Exit the client application Exits FCR PRIMA Console to close the “Main” window temporarily and using a utility tool. • Exit the server application Exits FCR PRIMA Console completely and switches to another application. • Shut down the PC Shuts down the PC after exiting the Client Application of FCR PRIMA Console.

Before exiting FCR PRIMA Console, terminate the study in progress. If you are browsing an image in the “Image Viewer”, exit the “Image Viewer”.

■ Operation Windows

● “Select Server Application Behavior” Dialog

● “Main” Window

(E)

(F)

(A)

(B)

● “Login” Window (C)

(D)

2-12

FCR PRIMA Console Reference Guide

897N101440B

■ Logging off the Current User 1. Click

(B) on the lower left of “Main”

■ Exiting the Server Application When exiting the server application, FCR PRIMA Console exits completely.

window.

→→A message displays to verify.

→→The “Main” window closes and the “Login” window (user selection window) opens.

Product Overview

2. Click [OK] in the confirmation dialog.

2

Exit the client application before exiting the server application.

1. Double-click the server application icon placed in the task tray.

• To start up another client application, select a user (C) from the “Login” window (user selection window) and log in. • To exit the current client application, click [Finish] (D) in the “Login” window (user selection).

→→The “Select Server Application Behavior” dialog appears.

2. Select [Exit] from the pull-down menu and click [OK].

→→The server application exits and the server application icon disappears from the task tray.

■ Exiting the Client Application The server application of PC keeps running, even when exiting the client application.

1. Click

The message displays when there is a study whose study status is Not Verified. To exit the server application with the current study status unchanged, click [OK].

(A).

→→“Main” window closes, and the client application exits. When the client application exits, the “Image Viewer” window also closes. →→The “Select Server Application Behavior” dialog appears. To use various utility tools: “2.5.1 ■ Starting up Each Function from the Server Application Icon” To exit the server application: “■ Exiting the Server Application”-2.

FCR PRIMA Console Reference Guide

897N101440B

2-13

■ Shutting down the PC Shuts down the PC after exiting the client application of FCR PRIMA Console.

2

1. Click

(A).

Product Overview

→→The “Select Server Application Behavior” dialog appears.

2. Select [Shutdown the PC] from the pull-down menu (E) and click [OK] (F).

→→The PC shuts down.

In case of Windows Vista, the following window opens if the PC is shut down in invalid procedure.

In this case, click [Cancel] and then double-click the placed in the task tray to server application icon display the “Select Server Application Behavior” dialog. Close the PC from the “Select Server Application Behavior” dialog. Also if “Shut down now” is accidentally clicked when DRYPIX PRIMA is connected, the power of DRYPIX PRIMA will not be linked and turned off. Turn off the power by holding the stand-by switch on the operation panel.

2-14

FCR PRIMA Console Reference Guide

897N101440B

2.6

Protecting the PC when Temporarily Leaving the Site 2

Protects the PC with password to prevent unauthorized use when temporarily leaving the site while working. FCR PRIMA Console is protected with the current operation status retained.



Product Overview

■ Operation Windows

■ Locking the PC Protect the PC by password.

Windows Task Tray

You can do the same when the PC runs without any operation in a certain time. (A)

“14.15 Automatically Logoff”

● Login Window for Windows

1. Click

(A) displayed in the task tray.

→→The login window for Windows opens.

■ Unlocking the PC 1. Enter a login password (B) for Windows and click

(B)

(C).

(C) The default password for user of fujicr is “system”. →→FCR PRIMA Console becomes ready.

FCR PRIMA Console Reference Guide

897N101440B

2-15

2.7

Before Using Disks

This section describes precautions in handling disks in FCR PRIMA Console, compatible disk types, drive types, and how to manage.

2 Product Overview

■ Precautions

● Disk Management

● Precaution in Handling Disks

Before using disks, read instructions on the disk package.

● Precautions in Handling Files Stored to Disks

• Do not save new files to a disk on which other files are already stored. • Do not delete, move or modify files stored on the disks.

Failure to do so could result in data unknown to the FCR PRIMA Console (CR-IR 391CL) or result in a failure to save data or loss of saved data.

● Compatible Disk Type

Non-cartridge disks can be used with the FCR PRIMA Console (CR-IR 391CL). • Single-sided 4.7GB • UDF 1.5 format or UDF 2.01 format

• 8-cm disks cannot be used. • Unformatted disks cannot be used. To use an unformatted disk with the FCR PRIMA Console (CR-IR 391CL), format the disk in the “Disk” window of “Indicator”. Switch disks : “13.8 ■Replacing Disks”

● Ejecting Disks

Do not force the ejection of a disk by pressing the disk drive EJECT button. When a disk needs to be ejected, perform in the “Disk” window of “Indicator”. Remove disks : “2.8 ■Setting up Disks”

If you use an application other than the FCR PRIMA Console to eject the disk, the PC may freeze or the disk may become invalid.

2-16

FCR PRIMA Console Reference Guide

897N101440B

Before using a new disk, you are advised to put the save start date on the label seal. For disks that are not used for further writing, enter the label name (media identification number) in the label seal. You should also put the save end date on the label seal. Write protection : “13.8 ■Setting Disks as “Write Enabled” or “Write Disabled””

● Windows Explorer

Do not attempt to access the disk drive from Windows Explorer. The drive might take a considerable time to access.

● FCR PRIMA Console Shutdown

Do not turn off the FCR PRIMA Console while writing to or reading from the disk is in progress (with the drive lamp illuminating).

● Disk Identification

When a disk is inserted in the disk drive, it will take about 30 seconds before it is identified to the FCR PRIMA Console.

■ Disk Types FCR PRIMA Console uses different types of disks for use. Disk type can be classified into “Image storage disks”, “Work disks”, “Other disks” and “PDI disks” depending on its use.

Disks intended for image storage are called storage disks. These disks are used to manage images. FCR PRIMA Console remembers which images are stored on which storage disks. Even though images are deleted from the FCR PRIMA Console database, FCR PRIMA Console can still import the copies of the same images from the storage disks.

• As the free space of the PC falls below a certain limit, stored images are automatically deleted from the FCR PRIMA Console database in sequence in ascending time order of the browsed studies. • With images imported from disks, you cannot: - Rotate or reverse the images, - Save FCR image processing parameters, - Edit markers, - Update study information, and - Verify or suspend studies.

Studies that are deleted automatically when the number of studies stored in the FCR PRIMA Console database exceeds a certain limit set on FCR PRIMA Console, studies stored on a storage disk are automatically deleted from the oldest study browsed. Once studies are removed from the FCR PRIMA Console database, they no longer appear in the “Main” window.

These disks are used to temporarily save images from the FCR PRIMA Console system for such purposes, such as the provision of images to other facilities and collection of images depicting various diseases. It saves DICOM-compatible format disks. FCR PRIMA Console also treats disks storing images provided by other facilities and DICOMcompatible format disks storing images captured by the FCR system as work disks.

2 Product Overview

● Image Storage Disks

● Work Disks

FCR PRIMA Console does not remember which images are stored on which work disks. FCR PRIMA Console treats the images imported from the work disk as unrelated images with those in the FCR PRIMA Console system.

● Other Disks

These disks are used to store images in the JPEG, bitmap or DICOM format for presentations and purposes alike. FCR PRIMA Console cannot display the names of such disks and the amount of available space on them.

● PDI Disks

CD-R or DVD-R disk for providing study result images to the patient for diagnostic cooperation with other facilities. Save to PDI disks: “7.7 Saving to PDI Disks”

Studies which will not be deleted are: • To be verified, • Being displayed in the “Image Viewer”, • Printing or being saved, or • Locked. Image data and study information associated with the images are stored on a storage disk, thereby studies can be backed up in the FCR PRIMA Console database by “Disk Restoration”. Restore Disks: “13.9 ■Restoring Information in Disks to Database” If too many studies have been registered and there are no studies that can be deleted, a message displays. Delete unnecessary studies from the “Main” window. “12.4 ■Deleting Studies”

FCR PRIMA Console Reference Guide

897N101440B

2-17

■ Drive Types There are two types of DVD disks: for control drive and normal drive. The PC consists of two types of equipment configuration, one is a control drive only or with a control drive and a normal drive.

2 Product Overview

Images in the disk can be loaded from both the control drive and normal drive.

● Control Drive

Control drive can write data to storage disk, work disk, other disks, and PDI disk. Type of disk on the control drive is displayed in the “Disk” window of the “Indicator” and the “Disk Operation” window of the “Client Utilities”. “Start Disk Handling” is performed on the control drive.

● Normal Drive

Normal drive can write data to other disks and PDI disk, but cannot write data to storage disk or work disk. Normal drive is not displayed in the “Disk” window of the “Indicator” nor the “Disk Operation” window of the “Client Utilities”.

2-18

FCR PRIMA Console Reference Guide

897N101440B

2.8

Setting up Disks

Before proceeding with a study, inserts a disk into the drive and set up. Disk handling is started on the control drive of the PC. If the last written disk remains inserted in the disk drive, disk handling will start automatically (without the operation described here).

2 Product Overview

This section describes the following operations: • Setting up disks • Ejecting a saved disk • Copying a disk • When the disk becomes full

The PC may freeze or the disk may become invalid if you use an application other than the FCR PRIMA Console. Verify precautions for using disks before setting up disks. “2.7 Before Using Disks”

■ Operation Windows

● “Replacing Disks” Dialog

● “Indicator” Window (A)

(F)

● “Indicator” Window - “Disk” Window (B) (G)

(H)

(C)

(D)

(E)

(I)

(J)

(K)

FCR PRIMA Console Reference Guide

897N101440B

2-19

● “Select Server Application Behavior” Dialog

■ Setting up Disks If the disk has not been inserted into the drive, inserts the disk and then set up.

2

(L)

Product Overview

Put necessary information on the disk label such as usage, name of disk, and recording start date. (M)

1. Click the “Disk” indicator (A) in the

● “Login” Window

“Indicator” window.

→→The “Disk” window of the “Indicator” window opens.

(N)

2. Select a disk drive in “Device Name” (B).

(O)

→→The status of the disk inserted into the selected drive displays. →→Nothing appears until the disk operation is performed.

(P)

● “Disk Copy Tool”

3. Click [Replace] (D).

(window on which copies are made)

→→The “Replacing Disks” dialog appears and the disk drive tray opens.

(Q) (R)

4. Place a disk in the disk drive tray and close the tray.

5. When the disk drive pilot lamp illuminates

(S)

green, click [Next] (F).

● Disk Copying Process Window

→→If the replaced disk is an image storage or work disk that has once been used with FCR PRIMA Console, the disk replacement operation has completed.

(T)

● “Indicator” Window - “Event” Window (U)

(V)

2-20

FCR PRIMA Console Reference Guide

897N101440B

→→If the replaced disk is unformatted, the confirmation message displays to format the disk. Click [OK] (G) to start up the software to format the disk. Proceed to 6. after formatting completed. For using method of softwares to perform formatting, see the instruction manual attached to disk drive. →→If the replaced disk is a new formatted disk, storage disk or disks other than for work, “Set Disk Usage” dialog appears. Proceed to 6..

6. Select the way you want to handle the disk inserted in the disk drive (H).

■ Ejecting Disk

Selection Items Items

Contents

For Image Storage

Storage disk

DICOMCompatible Format

Work disk

Other

Other disk

Disk Types: “2.7 Before Using Disks”

7. Click [Next] ( I ). →→If you select “For storage” or “DICOM-Compatible Format”, a “Disk Name Entry” dialog appears. Go to 7.. →→Selecting “Other” returns to the “Disk” window of the “Indicator” window. The disk replacement operation has completed.

8. Enter a disk name (J).

2 Product Overview

Not For Image Storage

Do not use the Open/Close (Eject) button on the drive mechanically because the PC may freeze or the disk may become invalid if an application other than the FCR PRIMA Console has been used. The disk in the normal drive of the client PC can be ejected by Windows operation or by using the drive’s Open/Close (Eject) button mechanically.

1. Click the “Disk” indicator (A) in the “Indicator” window.

→→The “Disk” window of the “Indicator” window opens.

2. Select a disk drive from Device Name (B) if the disk drive is added.

3. Click [Eject] (E). →→The disk drive tray opens.

4. Remove the disk from the disk drive tray and close the tray.

9. Click [Next] (K). →→The disk replacement operation has completed.

When the disk replacement operation completes, be sure to click [Update] to check that the disk has been replaced. Simply replacing a disk will not refresh the disk information.

10. Click [Update] (C) and verify that the contents of disk which has been set up is displayed. Display Items Items

Contents

Device Status

Disk name and space (%)

Use*1

Purpose of using the disk • For Image Storage • For Work

Writing Setting

Writing setting for the disk • Enable • Disable

*1 Nothing appears for other disks.

FCR PRIMA Console Reference Guide

897N101440B

2-21

■ Copying Disks

● Copying Disks

Duplicates image storage disks against possible failures to read them as a result of their data breakage or deterioration by using the “disk copy tool”.

2 Product Overview

• FCR PRIMA Console will exit when you attempt to start up the “disk copy tool”. Before starting up the copy tool, check that all the windows have closed. Check also that if any different user is operating the FCR PRIMA Console system. • Store backup disks in a secure place that is not accessible to third parties. • Disks may take up to 210 minutes at the longest to be copied. Avoid copying disks during the periods of the day in which a study needs to be started immediately.

● Starting up the Copy Tool 1. Double-click the server application icon

Inserts a disk into the disk drive and starts to copy it.

If the drive in the PC is used, the source and target drives are the same. Data on the source disk is temporarily saved to the hard drive in the PC before it is finally written to a target disk that is replaced with the source disk.

1. Select Source drive name for copying (Q) and then click [Eject] (R).

→→The source disk drive tray opens.

2. Place a source disk in the disk drive tray and close the tray.

3. Click [Execute] (S). →→Data on the source disk is temporarily saved to the hard drive in the PC. When the saving to the hard drive completes, a message is displayed

placed in the task tray.

→→The “Select Server Application Behavior” dialog appears.

2. Select [Disk Copy] (L) and click [OK] (M). →→The “Login” window opens after a message displayed.

3. Select the User Name (N) to log in. 4. Enter the Password (O) and click [OK] (P). →→If the user is authenticated, the “Disk Copy Tool” window opens.

To cancel the disk copying process, follow these steps: 1) Click [Cancel] (T). All data that has been recorded on the target disk is deleted. 2) When the message “Copying Incomplete” is displayed, click [OK]. The target disk drive tray opens. 3) The message “Place a target disk and click [OK]” is displayed. To resume the disk copying process, place a disk in the target disk drive and close the tray. When the disk drive pilot lamp illuminates green, click [OK].

When the disk access lamp is ON, never press the [Eject] button on the disk drive.

2-22

FCR PRIMA Console Reference Guide

897N101440B

4. Place a target disk in the disk drive tray and close the tray.

5. When the disk drive pilot lamp illuminates green, click [OK].

6. Click [OK].

When the disk in the control drive of the PC becomes full, the drive tray opens. The disk indicator turns off if the disk is an image storage disk.

2

Goes off

Product Overview

→→The disk copying process starts. When the copying process completes, the message displays.

■ When a Disk Becoming Full

(for an image storage disk)

→→The target drive tray opens and a message displays.

To continue copying data to another disk, replace the current target disk with another one. When the disk drive pilot lamp illuminates green, click [OK].

Save the data to the storage disk when the disk copying process completes. Switching procedure: “■Setting up Disks”

● Verifying that the Disk is Full

First, views the “Event” window to verify that the disk is full. Then, ejects the full disk and starts the handling of a new disk. Whenever a disk becomes full, takes relevant action as instructed on the right. The following steps describe how to handle new disks.

1. Click the “Event” indicator to display the “Event” window.

2. Select the latest warning on display disk storing failed (c) and click [Details] (d).

3. Check that the message reporting a full disk is displayed in the “Error Details” window.

4. Remove the full disk from the disk drive tray. 5. Start the handling of a new disk. “■Setting up Disks”

Simply placing a disk in the disk drive and closing the tray will not start the handling of the disk. Set up the disk.

FCR PRIMA Console Reference Guide

897N101440B

2-23

● Handling Full Disks Store backup disks in a controlled place that is not accessible to third parties.

2 Product Overview

1. Start up the “User Utilities” to perform a media consistency check.

“14.13 Media Consistency Check”

2. Verify the message displayed after the check and take suggested action.

3. Start up the “disk copy tool” to create a backup copy of the disk. “■Copying Disks”

2-24

FCR PRIMA Console Reference Guide

897N101440B

Chapter

3

3 Taking and Reading X-ray of a Subject

Taking and Reading X-ray of a Subject 3.1

Flow of Taking and Reading X-ray...................................... 3-2

3.2

Registering the Patient Information to be Exposed.......... 3-3

3.3

Exposing Images................................................................. 3-5

3.4

Reading and Viewing........................................................... 3-7

3.5

Import Various Images and Refer..................................... 3-10

FCR PRIMA Console Reference Guide

897N101440B

3-1

3.1

Flow of Taking and Reading X-ray

Starting the Exposure

Reading Images

3

4) Select the exposure menu

Taking and Reading X-ray of a Subject

1) Register or select the patient information

“3.2 Register the Patient Information to Expose”

5) Insert the cassette

“3.3 ■Read Images in the Cassette” 2) Select the exposure menu

6) Confirm images

3) Start the exposure

“3.3 ■Viewing Read Images”

“3.3 ■Taking X-rays”

3-2

FCR PRIMA Console Reference Guide

897N101440B

3.2

Registering the Patient Information to be Exposed

Patient information consists of the patient name, sex, date of birth and other related information. Each set of patient information is managed with reference to a unique ID (Patient ID). FCR PRIMA Console searches through the database for registered patient information using its patient ID as a keyword and, if the information is found, it is displayed in the patient list.

■ Registering New Patient Information

● “Main” Window - “Input ID” Window

1. Input the patient information in the patient

Taking and Reading X-ray of a Subject

■ Operation Window

3

information input column (A). Input Items Items

(A)

(B)

• In the “Display Settings” of Utilities, whether or not to display the window at the time of starting the “Main” window can be set.

Contents

Input ID*1*2

Enter a patient ID

Patient name

Enter a name of patient

Sex

Select a sex

Date of birth

Enter a date of birth

Height

Enter a height

Weight

Enter a weight

Patient comment

Enter a comment

*1 Enter the patient ID with single-byte alphanumeric characters. The allowable length is determined by a system setting. *2 The following characters cannot be used. "#%&`+?/\*@< However, the space can be used but the patient ID which consists of solely space is not allowed.

“11.1 Display Settings of the “Main” Window” • In the “Display Settings” of Utilities, whether or not to display the input item in the patient information input column can be set. “11.1.3 Display Settings of “ID input” Window”

When the patient ID which already registered in the FCR PRIMA Console database is entered in the patient ID column, an applicable patient information displays. In such cases, verify the patient information and start the exposure.

2. Check the contents which have been input, and click

(B) .

→→ The new patient information is registered.

FCR PRIMA Console Reference Guide

897N101440B

3-3

■ Registering through DICOM MWM Server (Optional)

Searches the patient information through MWM server and, if the information is found, it is displayed in the "Input ID" window. When searching the patient information through MWM server, use the patient ID which has been registered in MWM server. To search patient information, the following two methods can be used. • Search the MWM server prior to others • Search the FCR PRIMA Console database prior to others The method to be used depends on the system settings at the time of installation.

3 Taking and Reading X-ray of a Subject

Here describes the procedures of priorly searching the MWM server as an example.

Changes to the studies executing at FCR PRIMA Console or the patient information displayed in the patient list at the MWM server side will have no effect on the patient list. In these cases, change the mentioned patient information as well. “13.3 Patient Information Maintenance”

1. Input the patient ID in the ID input column of the patient information input column (A).

→→The MWM server is searched with the input patient ID. The patient information whose ID is input is displayed in the patient information column.

If failed to connect to DICOM MWM server, a message displays. Check whether or not the network cable is connected or whether or not the unit works normally. Please contact the maintenance personnel if the situation did not improve after taking the abovementioned actions. To search the MWM server again, perform procedure 1. again.

2. Check the patient information which is displayed, and click

.

→→The new patient information is registered.

3-4

FCR PRIMA Console Reference Guide

897N101440B

3.3

Exposing Images

FCR PRIMA Console offers groups of exposure menus on which the specific exposure conditions and method required for a body part are preset. Selects an exposure menu for X-ray studies to start taking X-ray. Exposure menu also can be scheduled.

■ Taking X-rays

● “Main” Window - “FCR” Window

1. Select the “Exposure Menu” (A).

3 Taking and Reading X-ray of a Subject

■ Operation Window

→→The “Read Image” window opens.

(A)

To expose the scheduled study, select the study in a “Scheduled” study status from the “Patient List” window and double-click it.

2. Select an exposure menu for the image to be read.

“3.4 Reading and Viewing” • The frequently used exposure menu can be registered as a shortcut menu in the “FCR” window. “11.1.5 Display Settings of “FCR-Menus” Window” • The contents of “Others” which are displayed in the “FCR” window can be created, edited and deleted.

3. Using a cassette containing a blank IP, take an X-ray of the subject.

Before starting exposures, be sure to read through the precautions. “Appendix Z Precautions for Exposure”

“14.3.1 Setting the Exposure Menu”

FCR PRIMA Console Reference Guide

897N101440B

3-5

● Scheduling an Exposure Operation To select an exposure menu and schedule an exposure, you need to set up. In the “FCROperations” window of “Display Settings”, set up an operation at the time of selecting the exposure menu to schedule a study.

3

“11.1.6 Display Settings of Exposing and Reading”

● Exposing a Study which has not Completed the Exposure

Performs the following operations from the “Patient List” window in the “Main” window.

1. Select a study in a “Scheduled” or “Started” study status from the patient list, and then double-click it.

Taking and Reading X-ray of a Subject

1. Select the “Exposure Menu” (A). →→The message displays to verify whether or not to schedule a study.

2. Click [OK]. • To proceed to schedule another study, input the patient ID in the patient ID column of “Input ID” window. • To verify the status of scheduled studies, view the “Patient List” window for a list of studies in the "Scheduled" study status. • Once a study is scheduled, an exposure operation can be started from the “Patient List” window.

3-6

FCR PRIMA Console Reference Guide

897N101440B

→→The “Read Image” window of the selected study opens.

Click [Cancel] to cancel waiting the image reading.

3.4

Reading and Viewing

Reads X-ray images corresponding to an exposure menu. Takes an X-ray of a subject prior to reading. When inserts the cassette used for the exposure into the Image Reader, the X-ray image of the subject opens in the FCR PRIMA Console monitor for viewing.

■ Read Images in the Cassette

● “Read Image” Window (Detail View)

1. Select an “Exposure Menu” (A) for the image

Before reading an image

(A)

3 Taking and Reading X-ray of a Subject

■ Operation Window

to be read.

The exposure menu for the image to be read is identified by image read icon

.

2. Check that the Cassette Ready Lamp is lit on the operation panel.

The Cassette Ready Lamp is on.

After reading an image

(A’) (B)

3. Insert a cassette into the equipment in (C)

When [Simple] is clicked, “Read Image” window switches from Detail view to Simple view.

alignment with the guide on the right, with the cassette’s barcode window facing as illustrated below.

• Remove the exposure marker from the cassette before inserting the cassette in the Image Reader. • Insert the cassette in the Image Reader after confirming that the top surface of the cassette is faced up and in the right direction. Moreover, insert the cassette straight and slowly. If a cassette is inserted upside down, back to front or at a slant, the Image Reader can be damaged. • Do not shake the Image Reader or the cassette while reading the image. Uneven image, which will prevent image reading, might be generated. • After inserting a cassette into the Image Reader, do not touch the cassette until the Cassette Removal Lamp lights. Touching an inserted cassette during image reading may cause the Image Reader failure.

FCR PRIMA Console Reference Guide

897N101440B

3-7

When inserting a cassette, check the patient name displayed on the operation panel of the Image Reader.

3 Taking and Reading X-ray of a Subject

Cassette guide Barcode window Insert straight carefully

4. Check that the Cassette Removal Lamp is blinking on the operation panel and then remove the cassette.

■ Viewing Read Images When inserts a cassette into the Image Reader and the image is read, it is displayed in the FCR PRIMA Console monitor. When an image is read, it is thumbnailed on the lefthand side of the exposure menu. When selects an exposure menu, the selected image opens on the lefthand side of the window. Read images are saved on the FCR PRIMA Console system.

1. Select the “Exposure Menu” (A’) after reading an image.

2. View an displayed image (B).

The following functions can be used for read images or exposure menu.

The Cassette Removal Lamp is blinking. Functions of Each Button Buttons *1

*1

Functions Rotate images 90 degrees to the left. Rotate images 90 degrees to the right.

*1

Rotate images 180 degrees.

*1

Flip images horizontally. Reverse rotated images to the default.

Modify*2

(previous page) (next page)*3 Simple

3-8

FCR PRIMA Console Reference Guide

897N101440B

Display the “Main Menu Modification” window and modify the study information. Display the previous or next page on the current page. • Progress bar A progress of image reading can be verified. • [Detail] “Read Image” window switches to “Detail” view. • [Image Viewer] Start up the “Image Viewer”. • [Complete] Finish the exposure of this study.

Buttons Suspend*4

Image Viewer

Functions Close the “Read Image” window and return to the “Main” window. The current study is left in the “Started” study status. Close the “Read Image” window and display the “Image Viewer” window

*1 The rotation or reversal of an image is reflected in the image when it is printed on a film (optional), when it is saved on the disk, or when it is transmitted to the DICOM server (optional). *2 “Main Menu Modification” window: “Chapter 9 Editing Study Information” *3 When manipulating hidden registered exposure menus into view. *4 The “Read Image” window can be displayed to resume the exposure.

If there are multiple exposure menus, proceed to take X-rays, insert cassettes into the “Image Reader” and view the resultant images with regard to the rest of the exposure menus.

1. Click [Completed] (C). →→The “Read Image” window closes and returns to the “Main” window.

3

Before ending exposure on the “Read Image” window, check the patient information, exposure menu, and image association.

Taking and Reading X-ray of a Subject

"Chapter 5 Verifying and Processing Images"

■ Finishing Exposure / Reading

If exposure menus for which images have not been exposed are existed, the message displays. To delete exposure menus for which images have not been read and start up the “Image Viewer”, click [OK].

• In this case, the study status sets to “Completed”. If FCR PRIMA Console is configured to print the image on a film with the film print (optional) upon completion of the exposure, the image is printed or saved automatically. • To reconduct a study, add additional menus or complete the re-exposure procedure and display the “Read Image” window. “Chapter 9 Editing Study Information”

FCR PRIMA Console Reference Guide

897N101440B

3-9

3.5

Import Various Images and Refer

In the “Import” window of “Main” window, DICOM images can be imported. Imported data and images can be referred in the “Refer” window. • Specify the media or folder and import DICOM images.

■ Operation Window

3 Taking and Reading X-ray of a Subject



■ Importing DICOM Images from the Media or Folder (Import DICOM Off-line)

“Main" Window - "Import" Window

(A)

1. Click

(A).

→→The "Import DICOM off-line" window opens.

2. Select the drive (C) which DICOM images are saved and click

(D).

→→The study information displays in the "Study list" (E). Display Items Displays

● ”Import DICOM off-line” Window

Contents

Study date

The date on which the exposure was started

Study time

The time at which the exposure was started

Modality

The modality of study

(E)

Number of images

Number of images which are exposed

(F)

Multi-frame

Existence of multi-frame images

(C) (D)

(G)

3. Select the study to import from "Study list" (E). (H) • To select all of studies which displayed, click • To cancel all of selections, click (I)

3-10

FCR PRIMA Console Reference Guide

897N101440B

(J)

(F).

(F).

(F). Select • To see contents of files, click the file to open when the "Open" window opens, and then click [Open].

4. Edit the patient information (G) if necessary. Input/Display Items Items

Contents Put a check mark if the patient ID and name of DICOM image to import is to be overwritten to the patient ID and name of FCR PRIMA Console database and to be imported.

(Left side) Patient ID Patient Name

The information of DICOM image to be imported is displayed.

(Right side) Patient ID Patient Name

Input the information in the system.

For others (sex etc.), overwrite with the information in the system

Put a check mark if overwriting the sex of the DICOM image to be imported to the sex registered in the system and importing it.

3 Taking and Reading X-ray of a Subject

Overwrite the patient information and import

• If the information input in the right side column is different with the information in the system, the confirmation window to overwrite the information in the system with the information input in the right side column opens. • If "Import DICOM off-line" window is opened while the patient is selected in "Main" window, the patient information selected in the "Main" window inserted automatically into the right side column. • Contents in the left can be pasted in the right by clicking

.

5. Select "Intended use for import" (H). • To import for storing, select "It will be managed by storing on the hard disk". • To import as a reference, select "It will be retrieved temporality for reference and then deleted afterday(s).". Input the date to save in the column.

6. Click [Start retrieving] ( I ). The image data can be imported and can be referred to in the Image Viewer at the same time by clicking [Viewer startup] (J).

FCR PRIMA Console Reference Guide

897N101440B

3-11

3 Taking and Reading X-ray of a Subject

3-12

FCR PRIMA Console Reference Guide

897N101440B

Chapter

4

4

4.1

Flow for Searching / Referring . .........................................4-2

4.2

Searching for Patients......................................................... 4-3

4.3

Referring Images.................................................................. 4-5

FCR PRIMA Console Reference Guide

897N101440B

Searching Patients and Referring Studies

Searching Patients and Referring Studies

4-1

4.1

Flow for Searching / Referring

The study images imported in FCR PRIMA Console after being exposed can be referred if searching by a study status or modality in the “Image Viewer” window. This section describes the flow of referring the images for a study after searching for patients in the “Image Viewer” window.

Referring Images

Searching for Patients

4 Searching Patients and Referring Studies

1) E  nter the search queries 2) Search 4) Select a patient

3) Display the search results

5) Select a study 6) Refer images

“4.2 Searching for Patients”

“4.3 Referring Studies”

To the Image Viewer Window

“Chapter 5 Verifying and Processing Images”

4-2

FCR PRIMA Console Reference Guide

897N101440B

4.2

Searching for Patients

Specifies the requirements such as the study status and modality to search a patient.

■ Operation Window

■ Searching for Patients

● “Main” Window - “Patient List” Window

1. Select “Search Requirement display column”

4

(A) to display “Search panel” (B).

2. Enter or select each search query in “Search

Searching Patients and Referring Studies

The study which matches with the selected search queries displays as a Patient List. Further, width of “Main” window may differ from the following figure depends on the system settings at the time of installation.

If a search query is entered, results are displayed in “Patient List” (C).

panel” (B).

Input and Selection Items (A)

Items Study Status

*1 In default, the study that the registration date is set to today is displayed in the list.

When move the mouse pointer on the desired button, the button name and description of functions are displayed as a pop-up message.

Progress of the study

Scheduled

An exposure operation is scheduled with patient information and exposure menus are registered.

Started

An exposure operation is in progress or suspended.

Completed

All the images are read and [Upon Completion] is selected.

Verified

The image quality of images read is verified and [Upon Verification] is selected.

(B)

(C) *1

Contents

Modality

Study modality • CR

Requesting Physician

The exposure requesting physician.

Registration date*1 *2

The date on which the study was registered (scheduled).

Registration time*1 *2

The date on which the exposure was started.

Patient ID or Patient Name

Patient ID or Patient Name of the patient studied.

*1 When the starting date and finishing date are specified, the specified period will be the search query. *2 When specifying the registration date and study date, the calendar displays. When clicking another area on the window without selecting the date in the calendar, display of calendar may remain. To hide the calendar, select the date in the calendar.

FCR PRIMA Console Reference Guide

897N101440B

4-3

• Click

The ascending order (▲) and descending order (▼) are changed if select a heading based on the display order.

• Click defaults.

• The orders for items in the patient list can be changed.

to reset the search query to its

Drag the items to change the order with selecting it, and drop it into the rows of destination to change.

.

Searching Patients and Referring Studies

→→The study which matches with the selected search query displayed in “Patient List” (C). Display Items Items Study Status

Contents The status of progress of the study : Scheduled : Started : Completed : Verified

Sex

Sex of a patient

Patient Name

Name of a patient

Patient ID

ID number of a patient studied

Registration date The date on which the study was registered (scheduled) Registration time The time at which the study was registered (scheduled)

4-4

• Display order of patient list can be reset.

• When the search panel closes automatically at the time of searching, put a check mark to the “Close the panel automatically” check box.

3. Click

4

to save the search query.

Study date

The date on which the exposure was started

Study time

The time at which the exposure was started

Modality

The modality of study

Requesting Physician

The exposure requesting physician

Date of Birth

The date of birth of patient studied

FCR PRIMA Console Reference Guide

897N101440B

• Items or orders to be displayed in the patient list can be set in the customize. “11.1.4 Display Settings of “Patient List” Window” • The following functions are available in “Patient List”. : Protect the privacy When this button is clicked with the condition which the patient is selected, the patient information except the selected patient will change to gray. : Reset sort Display order of patient list will be reset. When the sort is reset, the patient list will be sorted by descending of registration time and date. : Manual update of list The patient list is updated manually with the search query displayed in the column which displays the search queries.

4.3

Referring Images

Selects a study of the patient found to refer the study image in “Image Viewer” window.

■ Operation Window

■ Manipulating Studies

● “Main” Window - “Patient List” Window,

1. Search a patient to refer the image.

“Refer” Window

“4.2 Searching for Patients”

Select a patient to refer the image.

4

2. From the “Patient List” (A), select the patient

Searching Patients and Referring Studies

in a “Completed” or “Verified” study status.

→→The study of the selected patient is displayed in “Refer” window (B).

3. Select a study (B) to be referred. 4. Click

(C).

→→The selected study is displayed in the “Image Viewer” window. “5.1 Flow of Verifying and Processing Images” (A) You can do the same by: • Double-clicking the study to refer in “Patient List” (A). • Dragging and dropping the thumbnail image displayed in the “Image Viewer”. (B) *1

(C)

• Starting up the “Image Viewer” from the “Study Utilities”. However when the “Image Viewer” is started up from the “Study Utilities”, the simplified “Image Viewer” which limits the output function of image and so on opens. “12.4 ■Starting up the “Image Viewer””

*1 displays on the image which the report is registered. By clicking

, input contents can be confirmed.

“6.2 Creating/Editing the Report”

If a “Scheduled” or “Started” study status is in the selected studies, a window opens to verify whether to continue or not. Click [OK] if a study status is other than “Scheduled” or “Started” displayed in the “Image Viewer” window.

When (C) is not displayed at the bottom of the “Refer” window, add operation items in the “Display Settings”. “11.1.7 Display Settings of “Refer” window”

FCR PRIMA Console Reference Guide

897N101440B

4-5

The following functions are also available for a study. : Reconduct a Study and Editing Information A restudy can be set to perform additional and re-exposure operations. Also, the study information such as the menu change of the exposure images can be modified. “9.1 Starting up the Editing Function” : Saving to a PDI Disk

4

Images for a study are saved in a CD-R disk or a DVD-R disk in PDI format.

Searching Patients and Referring Studies

“7.7 Saving to PDI Disks” : Printing on a Film (optional) Print the study images on a film. “8.3 Printing on a Film (Optional)” : Saving to a Storage Disk Images for a study are saved in a storage disk which is set in the DVD drive of a PC. “7.3 Saving to Storage Disks” : Transmitting Studies to DICOM Server (optional) Transmit images for a study to DICOM server. “7.4 Transmitting Studies to DICOM Server (Optional)”

Click , when the item is not displayed in the bottom of “Refer” window. “11.1.7 Display Settings of “Refer” window”

4-6

FCR PRIMA Console Reference Guide

897N101440B

Chapter

5

Verifying and Processing Images Flow of Verifying and Processing Images......................... 5-2

5.2

Outline of "Image Viewer" Window.................................... 5-3

5.3

Selecting and Arranging Images........................................ 5-4

5.4

Adjusting Image Display..................................................... 5-9

5.5

Adding Annotations on Images........................................5-18

5.6

Free Drawing on “Image Viewer” for Informed Consent... 5-28

5.7

Saving/Verifying/Suspending Images.............................. 5-30

FCR PRIMA Console Reference Guide

897N101440B

Verifying and Processing Images

5.1

5

5-1

5.1

Flow of Verifying and Processing Images

This section describes the flow of verifying and processing images that are exposed/read or searched/ referred.

Expose/Read

Search/Refer

“Chapter 3 Taking and Reading X-ray of a Subject”

“Chapter 4 Searching Patients and Referring Studies”

Select and arrange images

5

Adjust image display

Verifying and Processing Images Select images Adjust place and display pattern “5.3 Selecting and Arranging Images”

Adding annotations on images

Arrow, marker, measured value of length, CTR, and angle between two line segments

Adjust gradation, rotate/flip Magnify/reduce, apply presets “5.4 Adjusting Image Display”

Verify/suspend images

Verify/suspend “5.7 Saving/Verifying/Suspending Images"

Free Drawing for Informed Consent “5.5 Adding Annotations on Images” “5.6 Free Drawing on“ Image Viewer” for Informed Consent”

Print/Save “Chapter 6 Using Report Function” “Chapter 7 Image Storage from Each Window” “Chapter 8 Printing Images”

5-2

FCR PRIMA Console Reference Guide

897N101440B

5.2

Outline of "Image Viewer" Window

This section describes the outline of names of each part and functions for "Image Viewer" window. 1) Display Check Box 2) Patient Information

7) [Edit Patient Information] 8) [Customize]

3) Tool bar

9) [Restore]/[Retry]

4) Study Summary

5

10) Report

Verifying and Processing Images

5) Image Cart

6) Sub-window

1)

Display Check Box

2)

Patient Information

Display/Hide each window of toolbar, study summary, image cart, and report.

The patient ID, name, sex, date of birth and age of the displayed image displays.

3)

6)

Sub-window to display images. Multiple sub-windows can be displayed, and multiple images can be displayed in the subwindow. "5.3 ■Displaying Multiple Sub-windows Lined up" "5.3 ■Displaying Multiple Images Lined up in the Sub-window"

Toolbar

Each function of button to place a image into sub-window, to adjust display of image, and so on displays. "5.3 Selecting and Arranging Images"

7)

Study Summary

The study image of the patient selected in the "Main" window displays as a list.

"9.9 Editing Patient Information"

8)

5)

Image Cart

A window to put images to save or print. Images in the image cart can be saved/printed. "Chapter 7 Image Storage from Each Window"

[Customize]

Display settings of "Image Viewer" window can be changed. "11.2 Changing Settings of “Image Viewer” Window"

"5.3 Selecting and Arranging Images"

By selecting the modality in the pull-down menu, the study which displays in the image summary can be narrowed.

[Edit Patient Information]

The patient information of the displayed image can be modified.

"5.4 Adjusting Image Display"

4)

Sub-window

9)

[Restore]/[Retry]

Image adjustment, such as changing gradation and so on, can be undone/redone. "5.4 Adjusting Image Display"

10) Report

Report can be created by adding comments to study images, putting annotations to schema images. Only one report can be created for each study image. "Chapter 6 Using Report Function" FCR PRIMA Console Reference Guide

897N101440B

5-3

5.3

Selecting and Arranging Images

Displays images that are exposed/read, searched/referred in the “Image Viewer” window. Switches image, changes the display pattern of images, or displays past study images side by side with the current image to adjust image display, process images or add annotations on images.

■ Descriptions of Functions

■ Operation Windows

● Switching images

● “Basic Operation” Tool Bar

Images which display on the sub-window can be switched by the image unit inside the study or study unit. Also, images can be displayed by adding a window.

5

"■Switching Displayed Images" Verifying and Processing Images

"■Adding a Sub-window and Displaying the Image"

● Displaying Sub-windows and Images Lined up

When comparing multiple studies and so on, images can be displayed and lined up in columns by lining up multiple sub-windows or dividing the sub-window.

(A)

"■Displaying Multiple Sub-windows Lined up" "■Displaying Multiple Images Lined up in the Sub-window"

● Switching and Maximizing Sub-windows,

● “Other” Tool Bar

Displaying information, etc.

Images can be displayed easier to see by switching the display position of sub-window, maximizing sub-window itself, and switching the display of images and sub-windows. "■Maximizing the Sub-window" "■Switching the Display Position" "■Displaying/Hiding Thumbnail Images" "■Displaying/Hiding Various Information on the Image" "■Displaying as a Cine Display Mode" "■Registering Displayed Images as a Key Image"

5-4

FCR PRIMA Console Reference Guide

897N101440B

(B)

■ Switching Displayed Images

● “Study summary” Window (E) (D)

Switches images which display in the subwindow by the image unit inside the study or study unit.

(C) (F)

1. Select the study or image (G) in the "Study Summary".

(G) By selecting the modality in the pull-down menu of "Study Summary" (F), the study of "Study Summary" can be narrowed with the selected modality.

→→Display on sub-window switches on the selected study image.

(H) (I)

Verifying and Processing Images

● “Sub-window”

5

2. Click [Switch] (C) in the "Study Summary".

• Displayed images can be switched by the following operations. - Double-click the study of "Study Summary" or image - Drag and Drop the study of "Study Summary" or image into the sub-window • If one study has multiple images (study view/series view etc.), displayed images can be switched by clicking [Previous page]/[Next page] or rotating mouse wheel on the image.

■ Adding a Sub-window and Displaying the Image

Displays an image in a new sub-window. An image displayed previously moves onto the previous page.

1. Select the study or image in the "Study Summary".

By selecting the modality in the pull-down menu of "Study Summary" (F), the study of "Study Summary" can be narrowed with the selected modality.

2. Click [Add] (E) in the "Study Summary". →→The new sub-window added and the image displays.

Sub-window can be switched by [Previous page]/ [Next page] in the toolbar.

FCR PRIMA Console Reference Guide

897N101440B

5-5

■ Displaying Multiple Sub-windows ■ Displaying Multiple Images Lined up

Lined up in the Sub-window

When multiple studies are compared, multiple sub-windows displayed lined up in columns.

If one study has multiple images (series images etc.), the sub-window is divided and multiple images are displayed lined up in columns.

Maximum of 4x4 (16 windows) sub-windows can be displayed lined up in columns.

1. Click [Layout] (A) on the "Basic operation" tool bar.

→→The display pattern displays under the button.

5

2. Move the mouse on the display pattern and

Verifying and Processing Images

click the display pattern of the sub-window to arrange.

→→The sub-windows display lined up.

• If only one sub-window is displayed, the subwindow reduces for fitting with the selected display pattern. To display images lined up, add a subwindow. "■Adding a Sub-window and Displaying the Image" • Also, the displayed sub-window and selected image can be displayed lined up by selecting the study (image) in the "Study Summary" and clicking [Compare] (D) when the sub-window is displayed.

• Maximum of 6x6 (36 windows) images can be displayed lined up in columns on one sub-window. • To display how many images lined up as a default display can be set by each modality. "11.2.4 ■Setting the Display of each Modality"

1. Click

( I ) in the sub-window.

→→The display pattern displays under the button.

2. Move the mouse on the display pattern and

click the display pattern of the sub-window to arrange.

→→The sub-window is being divided and images are displayed lined up.

■ Maximizing the Sub-window When sub-windows are displayed smaller, for example multiple sub-windows are displayed lined up in columns, maximize one sub-window by hiding the buttons on toolbar except tabs.

By maximizing the sub-window, magnification/ reduction of the image can be reset.

1. Click

(H) on the sub-window.

→→Buttons on toolbar hides and the sub-window maximizes.

• The sub-window can maximize by double-clicking the image. (H) once again, the window returns • By clicking to layout which displays sub-windows lined up.

5-6

FCR PRIMA Console Reference Guide

897N101440B

■ Switching the Display Position When multiple sub-windows are displayed lined up and when the sub-window is divided and multiple images are displayed lined up, position of each sub-window (image) can switch. "■Displaying Multiple Sub-windows Lined up" "■Displaying Multiple Images Lined up in the Sub-window"

The switched display position will not save. If the image is displayed again, it will display as a previous condition.

dragged.

→→The sub-window is being selected (In case of images, the frame turns to green.).

2. Drag and drop the sub-window onto the

desired position of other sub-window (or image).

→→The position of two images switches.

Information on the Image

Set the following information to display/hide. • Attribute Information • Annotation • Ruler • Patient Positioning • Reference Line

• Attribute information does not display when the layout of an image is small. • The ruler, patient positioning, and reference line may not display due to the DICOM information of an image.

1. Click [Overlay] (B) on the "Other" tool bar. →→The item to display/hide displays.

2. Put a check mark to the information to display.

Selection Items Items

Contents

Attribute Information

Image incidental information of exposure menu name, magnification rate, and etc.

Annotation

Arrows, figures, texts, measuring values, etc.

Ruler

A ruler

Images

Patient Positioning

Position of the patient on the image

Switches the display/hide of the thumbnail of study images.

Reference line

Reference line on the image

The display order of thumbnail images will not change from the previous condition.

■ Displaying/Hiding Thumbnail

5 Verifying and Processing Images

1. Select the sub-window (or image) to be

■ Displaying/Hiding Various

1. Click [Thumbnail display] (A) on the "Basic operation" tool bar.

→→The thumbnail image displays under the subwindow.

Each time of clicking, display/hide of thumbnail images can be switched.

Reference line →→The information which put a check mark on displays on the image.

FCR PRIMA Console Reference Guide

897N101440B

5-7

■ Displaying as a Cine Display Mode

Cine display mode can display and switch images automatically in the order when one study has multiple images. Display images after setting the switching speed and repeating.

■ Registering Displayed Images as a Key Image

Registers the displayed image as a key image.

By registering the key image, only the key image can be easily displayed from the pull-down menu of the sub-window.

1. Click [Cine display] (B) on the "Other" tool bar.

→→The control bar displays under the sub-window and switches to the cine display mode.

5

2. Set the display method.

1. Select the image.

Verifying and Processing Images

fps measured value

→→The frame of the selected image turns to green.

Reverse

2. Click [Register key image] (A) on the "Basic operation" tool bar.

Cine progress bar

Repeat play

Number of images

Play speed

Images may not play in a set speed even fps has been set. Confirm images after confirming the measured value of fps.

The range of frame of images displays in the subwindow can be changed by dragging the start position of the frame of cine progress bar (▲ on the left side) and last position of the frame cine progress bar (▲ on the right side) to the left or right.

3. Click

5-8

For the key image, an icon displays on the upper left of the image.

■ Clearing Images 1. Click [Clear image] (A) on the "Basic operation" tool bar.

(play).

→→The sub-window closes, and images displayed will be cleared.



When

is ON, images are playing reversely.



When

is ON, images are playing repeatedly.

4. To stop, click

→→The image registers as a key image.

.

FCR PRIMA Console Reference Guide

897N101440B

5.4

Adjusting Image Display

Adjusts gradation, rotates/flips, magnifies/reduces selected or arranged images.

■ Descriptions of Functions Images that are displayed in the “Image Viewer” window can be subjected to the image processing listed below.

■ Operation Windows ● “Basic operation” Tool Bar

● Changing the Gradation

(Contrast and Density) of Images

Changes the contrast and density temporarily.

5

"■Changing the Gradation (Contrast and Density)"

Verifying and Processing Images

● Magnifying/Reducing Images and Moving the Display Position

Magnifies and reduces images temporarily. The display position can be moved by dragging while the image is magnified. Also, one part of the image can be magnified by magnifying glass. "■Magnifying/Reducing" "■Moving Images" "■Magnifying Images by Magnifying Glass"

(A)

● “Image rotation/reversal settings” Window

● Reverse/Rotate and Flip

(B)

Reverses and rotates/flips the image. "■Reversing Images" "■Moving Images"

(C)

● Changing the FCR Image Processing Parameters

For the image of FCR study (unprocessed), FCR image processing parameters can be changed. Also, the changed parameters can be registered as a preset settings and images which are displaying by applying the registered preset settings can be processed at once. "■Editing FCR Image Processing Parameters"

(D)

● “Changing FCR Image Processing Parameters” Window

"■Registering Preset Image Processing Settings" "■Applying Preset Image Processing Settings"

● Linking the Image Processing Settings

(E)

Among Multiple Images

Applies the image processing, which have been done to one image, to other images at once by linking multiple images. "■Linking Image Processing Among Multiple Images"

● Resetting the Image Processings

(H)

(F) (I)

(G)

(J)

Resets the image processings and reverses the image status. "■Resetting the Image Processings" FCR PRIMA Console Reference Guide

897N101440B

5-9

● “Organize preset list”

3. Drag the mouse pointer up, down, to left or to right.

(K)

Operation Items

(M)

Items

(N)

(L)

Move up

Increase contrast (GA)

Move down

Reduce contrast (GA)

(Q)

Move to left

Reduce density (GS)

(R)

Move to right

Increase density (GS)

(O) (P)

● “Register Preset Name”

5

Contents

Up: Increased GA

(S)

Left: Reduced GS

Right: Increased GS

Verifying and Processing Images

(T) Down: Reduced GA

■ Changing the Gradation

→→The image changes contrast (GA) and density (GS).

(Contrast and Density)

Changes the image contrast temporarily.

• In case of monochrome images other than FCR images, the gradation changes as follows by dragging up, down, to left, or to right.

Only the monochrome image can change the gradation.

Operation Items Items

Changes of contrast and density are temporary. To save the changing result of gradation, change it by FCR image processings.

Drag up

Window center (WC) increases

Drag down

Window center (WC) decreases

Drag to left

Window width (WW) becomes narrower

Drag to right

Window width (WW) becomes wider

"■Editing FCR Image Processing Parameters"

1. Click [Gradation] (A) on the "Basic operation" tool bar.

→→The system enters the gradation change mode.

You can do the same by right-clicking image pane and selecting "Gradation".

• The mode changes in the following order by right-clicking with holding the left mouse button on the image. Gradation Change Mode → Magnification/ Reduction Mode → Panning Mode → Gradation Change Mode → ...

4. To exit the gradation change mode, click

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

Contents

[Gradation] (A) once again.

. You can do the same by right-clicking the image pane and selecting “Gradation” once again. →→The shape of the mouse pointer will return to

5-10

FCR PRIMA Console Reference Guide

897N101440B

.

■ Magnifying/Reducing

• When magnifying, the display position can be moved by dragging the image. "■Moving Images"

Magnifying/Reducing images do not reflect for printing or saving.

Images will come up having their original status next time they are displayed.

Images can be magnified or reduced in 2 methods: gradually magnifies or reduces, and magnifies or reduces with set conditions (fit to the window, in full size, or in pixel scale).

With the magnification/reduction mode, images can be magnified/reduced by dragging up or down on the image.

"Basic operation" tool bar.

→→ The system enters the magnification/reduction mode.

You can do the same by right-clicking image pane and selecting "Magnification/reduction".

.

3. Drag the image and move the mouse pointer up or down.

Operation Items Contents

Move up

Magnifies

Move down

Reduces

Size of images cannot reduce any smaller than its original size. →→The image magnifies or reduces.

5

You can do the same by right-clicking the image pane and selecting "Magnification/reduction" once again. .

● Magnifying or Reducing with the Set Conditions

Magnifies or reduces with the following settings. • AutoFit to Window :

Displays images in a width fit to the image pane. • 100%-Size Image :

2. Move the mouse pointer over the image.

Items

click [Magnification/reduction] (A) once again.

→→The shape of the mouse pointer will return to

1. Click [Magnification/reduction] (A) on the

→→The shape of the mouse pointer will turn to

4. To exit the magnification/reduction mode,

Verifying and Processing Images

● Gradually Magnifying or Reducing

• The mode changes in the following order by right-clicking with holding the left mouse button on the image. Gradation Change Mode → Magnification/ Reduction Mode → Panning Mode → Gradation Change Mode → ...

Displays images in actual life size (100%) magnification based on the aspect ratio of an image. • Pixel Scale :

Displays images with their picture elements being represented each as one pixel of monitor resolution.

1. Select a image you want to magnify or reduce.

→→The selected image pane will appear in bold frame.

2. Select [▼] of [Magnification/reduction] (A) on the "Basic operation" tool bar, select "Fit to window", "100%-Size Image", or "Pixel Scale".

→→The image is magnified or reduced with the selected setting.

You can do the same by right-clicking image pane and selecting "Fit to window", "100%-Size Image", or "Pixel Scale".

FCR PRIMA Console Reference Guide

897N101440B

5-11

■ Magnifying Images by Magnifying Glass

The size of magnifying glass can be changed. "11.2.2 Initial Settings of Annotation"

1. Click [Magnifying glass] (A) on the "Basic operation" tool bar.

5

■ Moving Images You can move the image only when the image is magnified. "■Magnifying/Reducing"

1. Click [Panning] (A) on the "Basic operation" tool bar.

→→The system enters the magnifying glass mode.

→→The system enters the image move mode.

• You can do the same by right-clicking image pane and selecting "Magnifying glass".

You can do the same by right-clicking the image pane and selecting “Panning”.

Verifying and Processing Images

• To change the magnification of magnifying glass, click [▼] under the [Magnifying glass] and select the magnification.

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

2. Click (hold) the image. →→The magnifying glass displays and the position where clicked (held) displays and magnifies. By releasing the mouse button, magnifying glass exits.

By dragging the image up, down, to left, or to right, the position of mouse pointer on the image magnifies.

3. To exit the magnifying glass mode, click [Magnifying glass] (A) once again.

You can do the same by right-clicking image pane and selecting "Magnifying glass" once again.

.

3. With the left-hand button of the mouse held

pressed, drag the hand-shaped pointer over the image to an optional position.

→→The image moves to the way it is dragged.

The mode changes in the following order by rightclicking with holding the left mouse button on the image. Gradation Change Mode → Magnification/Reduction Mode → Panning Mode → Gradation Change Mode → ...

4. To exit the image move mode, click [Panning] (A) once again.

You can do the same by right-clicking the image pane and selecting “Panning” once again. →→The shape of the mouse pointer will return to

5-12

FCR PRIMA Console Reference Guide

897N101440B

.

■ Reversing Images Reversing is a temporal function. Images will come up in their original status next time they are displayed.

1. Select the image to be reversed. →→ The selected image pane will appear in a bold frame.

2. Click [Reverse] (A) on the "Basic operation" tool bar.

→→The target image is reversed.

1. Select the image to rotate or flip. →→The selected image pane will appear in bold frame.

2. Click [Rotate] (A) on the "Basic operation" tool bar.

→→The “Image rotation/reversal settings” window opens.

You can do the same by right-clicking the image pane and selecting “Processing” - “Rotate/Flip”.

5

3. Select the direction to be rotated or flipped

Verifying and Processing Images

You can do the same by right-clicking image pane and selecting “Processing” - “Reverse”.

■ Rotating and Flipping

(B) (C).

Selection Items Items

Contents

No Horiz Flip Original Image

Images not horizontally flipped The original image in the current direction shown in the “Image Viewer” window

Rotate Right Rotated 90 degrees to the right 90° Rotate Left 90°

Rotated 90 degrees to the left

Rotate 180°

Rotated 180 degrees

Horiz Flip

Horizontally flipped image

Original Image

The original image in the current direction shown in the "Image Viewer" window

Horizontal Flip

Horizontally flipped

Vert Flip

Vertically flipped

H-Flip + Right 90°

Horizontally flipped and then rotated 90 degrees to the right

H-Flip + Left 90°

Horizontally flipped and then rotated 90 degrees to the left

The frame of the selected image turns to green.

4. Click [Selection] (D). →→The image rotates/flips.

FCR PRIMA Console Reference Guide

897N101440B

5-13

■ Editing FCR Image Processing

GP (Gradation Processing) Parameter

Parameters

Items GA

Value range

GT

processing parameters.

2. Click [FCR image processing] (A) on the "Basic operation" tool bar.

Verifying and Processing Images

→→The "Changing FCR Image Processing Parameters" window opens.

You can do the same by right-clicking the image pane and selecting "Processing" - "FCR image processing".

Shape of the gray level curve

Value range GC

Center density of gray level curve slope

Value range GS

Parallel shift of the gray level curve

Value range

GP/MFP

TAS

GP/RP

Items RN

Characteristics The frequency band specified here is subjected to enhanced processing Value range

• By clicking [Default] (G), the parameter values can be changed to the default image processing parameters according to exposure menus.

2) Click [Apply] (Q).

5-14

FCR PRIMA Console Reference Guide

Shape of the enhanced curve corresponding to the density of the image

Value range RE

897N101440B

F, P to Z (12 types) Enhancement of frequency processing

○: Exclusive control Characteristics Change the enhancement level of processing in accordance with the specified value Value range

→→Image processing parameters change to preset settings and the previewing image of "Changing FCR Image Processing Parameters" renews.

0 to 9 (10 types)

Characteristics By changing this parameter, the enhancement level can be adjusted within a certain density region



1) Select [Anatomical Region] (K) and [Preset Name] (L).

Contents Enhanced frequency band for frequency processing

RT

• By clicking [Preset Setting] (F), if an image is an unprocessed FCR image, the "Organize preset list" window opens and image processing parameters can be changed to preset settings which have been registered.

-1.44 to 1.44 (2 decimal places)

RP (Spatial Frequency Processing) Parameter

• To edit DRC processing parameters, select [DRC] and then click [+] or [-].

DRC

0.30 to 2.64 (2 decimal places)

Characteristics Change the display density of the entire image

parameter (E).

• Certain image processing sequences have no effect when carried out concurrently. Each image processing sequence takes the course shown below.

A to Z (26 types)

Characteristics Change GA according to the density specified here

3. Click [+] or [-] and change the value of each

• When [+] or [-] are held, the parameter value changes continuously.

-4.0 to -0.1 0.1 to 4.0 (1 decimal place, excluding 0.0)

Characteristics Yield the same result as changing the γ-curve of X-ray films

→→ The selected image pane will appear in a bold frame.

5

Slope of the gray level curve Characteristics Adjust the image contrast

If you select an image other than an unprocessed FCR image, the FCR image processing parameters cannot be modified.

1. Select the image to edit FCR image

Contents

0.0 to 9.9 (1 decimal place) 10 to 16 (no decimal places)

DRC (Dynamic Range Control Processing) Parameter Items DRN

Contents Mask size of DR compression

Characteristics Set the processing region to match DRT Value range DRT

0 to 9 (10 sizes) DR compression filter type

Characteristics Set the density region for processing Value range DRE

A to T (20 types) Enhancement of DR compression filter processing

Value range

0.0 to 2.0 (1 decimal place)

TAS (Tomographic Artifacts Suppression Processing) Parameter Items ORN

Contents One-dimensional unsharpness mask

Characteristics Set the most effective size for the tomographic angle Value range ORE

0 to 9 (10 sizes) Enhancement level of onedimensional unsharpness mask processing

Characteristics Change the enhancement level of processing in accordance with the specified value Value range ORD

0.0 to 9.9 (1 decimal place) 10 to 16 (no decimal places) Processing direction of onedimensional unsharpness mask

Characteristics Set in accordance with the reading direction and orbital direction of tomography Value range

Items MDB

Contents DR compression frequency type

Characteristics Set the processing region to match MDT Value range MDT

A to Z (26 types) DR compression filter type

Characteristics Set the density region for processing Value range MDE

A to T (20 types)

5

Enhancement of DR compression filter processing

Characteristics Change the enhancement level of processing in accordance with the specified value Value range MRB

Verifying and Processing Images

Characteristics Change the enhancement level of processing in accordance with the specified value

MFP (Multi-objective Frequency Processing) Parameter (Optional)

0.0 to 2.0 (1 decimal place) Enhanced frequency band for frequency processing

Characteristics This frequency band is subjected to enhanced processing Value range MRT

A to Z (26 types) Shape of the enhanced curve corresponding to the density of the image

Characteristics By changing this parameter, the enhancement level can be adjusted within a certain density region Value range MRE

F, P to Z (12 types) Enhancement of frequency processing

Characteristics Change the enhancement level of processing in accordance with the specified value Value range

0.0 to 9.9 (1 decimal place) 10 to 16 (no decimal places)

0: The orbit is at right angles to the green line of the cassette. 1: The orbit is parallel to the green line of the cassette

FCR PRIMA Console Reference Guide

897N101440B

5-15

Sensitivity, Latitude Items Sensitivity (S-value)

Contents Central point of the image reading sensitivity established by the EDR (automatic sensitivity adjustment feature)

Characteristics The higher the incident X-ray level, the lower the S-value becomes; the lower the incident X-ray level, the higher the S-value becomes Value range Latitude (L-value)

5

4 to 20047 Reading latitude point of the image reading sensitivity established by the EDR (automatic sensitivity adjustment feature)

Verifying and Processing Images

Characteristics Give a logarithmic reading of the range of incident X-ray levels (minimum to maximum) Value range

0.50 to 4.00

■ Registering Preset Image Processing Settings

The image processing settings for the selected image are registered as preset settings.

1. Select the image. 2. Click [FCR image processing] (A) on the "Basic operation" tool bar.

→→The "Changing FCR Image Processing Parameters" window opens.

3. Click [+] or [-] and set the parameter values register as a preset settings (E).

"■Editing FCR Image Processing Parameters"

4. Click [Preset Setting] (F). →→The "Organize preset list" window opens.

4. Click [Refresh] (H). →→The image processes with modified condition and the previewing image reproduces.

5. Check the previewing image and click [Apply] (J).

→→The image in the sub-window changes and displays according to the parameters.

5. Click [New Registration] (M). →→The "Register Preset Name" window opens.

6. Input the preset name in [Preset Name] (S). 7. Click [OK] (T). →→The preset name registers.

8. Check the preset name. • By clicking [Show/Hide] (O), whether to display as a preset can be set. • By clicking [Change Name] (N), the preset name can be changed. • By clicking [Delete] (P), the preset can be deleted. • By dragging and dropping, the display order of preset can be changed.

9. Click [Close] (R). →→The system returns to "Changing FCR Image Processing Parameters" window window.

10.

5-16

FCR PRIMA Console Reference Guide

897N101440B

Click [Cancel] ( I ).

■ Applying Preset Image Processing ■ Linking Image Processing Settings

Among Multiple Images

Batch processing of displayed images is possible by selecting saved preset image processing.

When multiple images are displayed, applies the image processing, which have been done to one image, to other images at once by linking multiple images. • Apply the image processing of one subwindow to other sub-window • Apply same image processing to all of images in the sub-window

Only unpressed FCR study image can apply the preset settings.

1. Select the image to be manipulated. →→The selected image pane will appear in a bold frame.

tool bar (or [Preset all] on the "Other" tool bar).

→→The preset displays in the list.

3. Select the preset settings. When the [Preset all] is clicked, settings to apply from the preset settings of all of the body parts can be selected, regardless of the displayed body part of the image. →→Image processing is done according to the selected preset settings.

When preset settings are assigned to keypad keys, image processing can be done by pressing a keypad key.

5

2. Link images and sub-windows.

• When an image processing of one sub-window is

Verifying and Processing Images

2. Click [Preset] (A) on the "Basic operation"

1. Display the image to be linked.

on applying to other sub-window, click all of sub-windows to be linked. • When a same image processing is applying to all of images in the sub-window, click window.

on the sub-

3. Perform the image process. "■Magnifying/Reducing" to "■Applying Preset Image Processing Settings" →→Applies the image processing, which have been done to one image, to other images at once by linking multiple images.

■ Resetting the Image Processings Resets the image processings and reverses the image status.

"14.17 Preset Setting" By resetting, the data returns to the state it was in prior to being saved. Once results are saved, they would not be undone if [Reset] is clicked. "5.7 ■Saving the Image Processing Result and Annotation"

The image processing can be undone or redone by clicking [Restore]/[Retry] on the toolbar.

1. Click [Reset] (A) on the "Basic operation" tool bar.

→→The image process resets and the data returns to the state it was in prior to being saved.

FCR PRIMA Console Reference Guide

897N101440B

5-17

5.5

Adding Annotations on Images

Arrows, shapes, text (character strings), and markers can be added to an image.

■ Descriptions of Functions

■ Operation Windows

● Adding Arrows, Shape, and Text

● “Measurement function” Tool Bar

Arrows and shapes (ellipses, rectangles, or polygons) can be drawn on images and text can be added. Also, system markers registered in advance and user markers registered by a user can be added.

5

(A)

Verifying and Processing Images

"■Adding Arrows/Shapes (Ellipses, Rectangles, and Polygons)" (B)

"■Adding Texts" "■Adding Markers"

● Adding Measured Values

The following kinds of measured values can be added. • Measure line segment • Simplified cardiothoracic ratio • Cardiothoracic ratio • Interline angle • Dual angle • Cobb angle • VHS

(C)

● “Property change” Window (D)

"■Adding Measured Lengths" to "■Adding Measured Hip Joint Angles"

● Adding Measured Pixel Values

The pixel value of one point on the image can be displayed, and pixel values (average, standard deviation, maximum value, minimum value, and etc.) of range surrounded by the diagram (ellipses, rectangles, and polygon) can be measured and added.

(E)

"■Adding Measured Pixel Values"

● Editing/Deleting Added Annotations Annotations added on the image can be edited/deleted. "■Editing the Added Annotations"

(D)

"■Deleting Added Annotations"

As in case of image editing, the data can be returned to the state where it was in prior to being saved by resetting the annotation before saving. "5.4 ■Resetting the Image Processings"

5-18

FCR PRIMA Console Reference Guide

897N101440B

(E)

■ Adding Arrows/Shapes (Ellipses, Rectangles, and Polygons)

1. Click an icon ([Arrow], [Ellipse], [Rectangle],

[Polygon]) (A) on the "Measurement function" tool bar.

(D)

You can do the same by right-clicking the image and selecting "Annotation” - “Arrow” / “Ellipses” / “Rectangles” / “Polygon”. (E)

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

Verifying and Processing Images

● “Marker Setting” Window

5

3. Specify a range of drawing with the mouse. Operation Items Items

Contents

Arrows

Dragged range from the start point to the end point

(F) Shape Ellipses

Dragged ellipse’s range

Rectangles

Dragged rectangle’s range

Polygon

Click to specify each apex of polygon, and double-click on the last apex

(G)

• Arrows, ellipses, and rectangles can be drawn by clicking the start point and end point (same as start point and end point of range to surround). • It is possible to drag a mouse only inside the image pane. • When only two points or fewer are specified as apex and the mouse is double-clicked, the display returns to the status before apex specification. →→An annotation is drawn over the range specified with the mouse. The shape of the mouse pointer will return to

.

FCR PRIMA Console Reference Guide

897N101440B

5-19

■ Adding Texts

■ Adding Markers Adds system markers registered in advance or user markers registered by a user.

1. Click [Text] (A) on the "Measurement function" tool bar.

1. Click [Marker] (A) on the "Measurement function" tool bar.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” - “Text”.

→→The "Marker Setting" window opens.

2. Move the mouse pointer over the image.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” - “Marker”.

→→The shape of the mouse pointer will turn to

5

.

3. Click the mouse to specify the location to

2. Select the marker to use (F).

Verifying and Processing Images

input texts.

• Click [System] to use system markers or [User] to use user markers and switch the displayed markers.

→→The text input area and the insertion point are displayed.

4. Input text.

• The user markers can be registered or deleted. System markers cannot be registered or deleted.

→→The entered text is displayed.

"14.16 Adding/Deleting User Marker"

3. Click [Selection] (G).

Inputting texts up to a maximum of 120 characters are possible.

5. Click the mouse at a position outside the text input area.

→→This operation will verify the text. The shape of the mouse pointer will return to

.

4. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

5. Click the mouse at a location where a marker is to be added.

→→A marker is displayed on the clicked point, and The shape of the mouse pointer will turn to

5-20

FCR PRIMA Console Reference Guide

897N101440B

.

■ Adding Measured Lengths

■ Adding Measured Simplified Cardiothoracic Ratios

Measurement results yielded by this processing are presented only for reference purposes without any warranties of their accuracy. It is the user’s responsibility to evaluate the measurement results.

Measurement results yielded by this processing are presented only for reference purposes without any warranties of their accuracy. It is the user's responsibility to evaluate the measurement results.

The measured results are displayed to two digit of decimal point by round to the thousandth digit.

The measured results are displayed to thousandth digit of decimal point by round to the ten-thousandth digit.

1. Click [Measure line segment] (B) on the "Measurement function" tool bar.

5 Verifying and Processing Images

1. Click [Simplified cardiothoracic ratio] (B) on the "Measurement function" tool bar.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” “Measure line segment”.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” “Simplified cardiothoracic ratio".

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

3. Drag the mouse pointer from the start point

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

to the end point of a straight line.

,

3. Click both edges of the longest span of the Measurement can be performed by clicking the start point and end point.

medial borders of the thoracic cage.

• By clicking the image, the vertical ruled line guide displays. * The picture below shows an operational procedure.

1) Click

2) Click

→→A linear length displays. The shape of the mouse pointer will return to

.

• The longest span of the medial borders can be specified by dragging both edges of it. →→The maximum inside diameter of the thoracic cage is indicated.

FCR PRIMA Console Reference Guide

897N101440B

5-21

4. Click both edges of the longest span of the cardiac shadow.

Cardiothoracic Ratios

• By clicking the image, the vertical ruled line guide displays. * The picture below shows an operational procedure.

1) Click

5

■ Adding Measured

2) Click

Measurement results yielded by this processing are presented only for reference purposes without any warranties of their accuracy. It is the user's responsibility to evaluate the measurement results.

The measured results are displayed to thousandth digit of decimal point by round to the ten-thousandth digit.

Verifying and Processing Images

1. Click [Cardiothoracic ratio] (B) on the "Measurement function" tool bar.

• The longest span of the medial borders can be specified by dragging both edges of it.

→→The selected image pane appears in bold frames.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” “Cardiothoracic ratio”.

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

3. Click the start point and end point of the →→The maximum transverse diameter of the cardiac shadow and CTR are indicated. CTR mode exits and the shape of the mouse pointer will return to

.

midline.

4. Click both edges of the longest span of the medial borders of the thoracic cage.

The vertical ruled line guide displays adjusting with the angle of midline. →→The maximum inside diameter of the thoracic cage is indicated.

5-22

FCR PRIMA Console Reference Guide

897N101440B

5. Click both edges of the longest span of the cardiac shadow.

The longest span of the medial borders can be specified by dragging both edges of it.

■ Adding Measured Interline Angles Two lines are drawn on the image in the "Image Viewer" window and the angle between those two lines is measured and added on the image. Also, if there is no intersection on the screen, the angle of the intersection of the lines extended off screen is displayed.

Measurement results yielded by this processing are presented for reference purposes without any warranties of their accuracy. It is the user’s responsibility to evaluate the measurement results.

the mouse pointer will return to

.

Verifying and Processing Images

→→The maximum transverse diameter of the cardiac shadow and CTR are indicated. Cardiothoracic ratio mode exits, and the shape of

5

• Measurement result is rounded to first decimal place and displayed to second decimal place. • The angles which are to be measured can be edited from within the lines drawn. "■Editing the Added Annotations"

1. Click [Interline angle] (B) on the

"Measurement function" tool bar.

You can do the same by right-clicking anywhere in the image pane and selecting “Annotation” “Interline angle”.

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

3. Draw a 1st line by dragging the mouse

pointer from the start point to the end point of the line.

FCR PRIMA Console Reference Guide

897N101440B

5-23

4. Draw a 2nd line by dragging the mouse

pointer from the start point to the end point of the line.

Lines can be drawn by clicking the start and end point for 1st or 2nd line.

■ Adding Measured Dual Angles Adds on the image by drawing optional three lines on the image of "Image Viewer" window and measuring angles within 1st and 3rd line and within 2nd and 3rd line. Also, when the intersection does not exist on the image, the angle of the intersection existed outside of the image can be displayed.

Measurement results yielded by this processing are presented for reference purposes without any warranties of their accuracy. It is the user’s responsibility to evaluate the measurement results.

5 Verifying and Processing Images

→→The angle within two lines (measured result) displays. The shape of the mouse pointer will return to

.

• Measurement result is rounded to first decimal place and displayed to second decimal place. • The angles which are to be measured can be edited from within the lines drawn. "■Editing the Added Annotations"

1. Click [Dual angle] (B) on the "Measurement function" tool bar.

You can do the same by right-clicking anywhere on the image pane and selecting "General measurement" - "Dual angle".

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

3. Draw a 1st line by dragging the mouse

pointer from the start point to the end point of the line.

4. Draw a 2nd line by dragging the mouse

pointer from the start point to the end point of the line.

5. Draw a 3rd line by dragging the mouse

pointer from the start point to the end point of the line.

Lines can be drawn by clicking the start and end point for 1st, 2nd or 3rd line. →→The angles within 1st and 3rd line and within 2nd and 3rd line (measured result) display. The shape of the mouse pointer will return to

5-24

FCR PRIMA Console Reference Guide

897N101440B

.

■ Adding Measured Cobb Angles ■ Adding Measured Pixel Values 1. Click [Pixel value] (C) on the "Measure Pixel value" too bar.

Measurement results yielded by this processing are presented for reference purposes without any warranties of their accuracy. It is the user’s responsibility to evaluate the measurement results.

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

.

3. Click the position to measure the pixel value. • Measurement result is rounded to first decimal place and displayed to second decimal place. • The angles which are to be measured can be edited from within the lines drawn.

5

"■Editing the Added Annotations"

function" tool bar.

■ Adding Measured Various ROIs

You can do the same by right-clicking anywhere on the image pane and selecting "General measurement" - "Cobb angle".

Adds on the image by measuring pixel values (average value, standard deviation, maximum value, minimum value, etc.) of region surrounded by diagrams (ellipse, rectangle, polygon).

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

1. Click the button ([Ellipse ROI], [Rectangular .

3. Draw a 1st line by dragging the mouse

pointer from the start point to the end point of the line.

4. Draw a 2nd line by dragging the mouse

pointer from the start point to the end point of the line.

ROI], [Polygonal ROI]) (C) on the "Measurement Pixel value" tool bar.

2. Move the mouse pointer over the image. →→The shape of the mouse pointer will turn to

→→The angle within two lines (measured result) displays. The shape of the mouse pointer will return to

.

3. Specify the range to measure ROI. Operation Items Items Ellipse ROI

Lines can be drawn by clicking the start and end point for 1st or 2nd line.

Verifying and Processing Images

→→While clicking, the pixel value displays on the upper part of mouse cursor.

1. Click [Cobb angle] (B) "Measurement

Contents Drag the range surround by ellipse

Rectangular ROI Drag the range surround by rectangle Polygonal ROI

Drag the range surround by polygon

. The range of ellipse ROI and rectangular ROI can be specified by clicking the start and end point of the range to be surrounded. →→The shape of the mouse pointer will return to

FCR PRIMA Console Reference Guide

.

897N101440B

5-25

■ Editing the Added Annotations Edits the position and angle of annotations. Also, properties (thickness of lines, fonts, etc.) of annotations can be edited.

● Editing Properties of Annotations Edits the width of lines, colors of lines, background/font/style of texts.

1. Right-click the annotation and select the "Properties".

If a new annotation is added surrounding the original annotation, the original annotation cannot be selected. By moving or editing the added annotation, the original annotation can be selected again.

● Editing the Position and Angle of

→→The "Property change" window opens. Select the tab of the object to be edited, and set the width of line/color of line or font type/ style/size.

2. Edit the properties (D).

Annotations

5

Depends on the annotation, items are different.

Verifying and Processing Images

1. Select the annotation.

Annotations

2. Edit the annotation. Selection Items Move position

Contents Drag the mouse pointer to the destination Drag

Move measuring position

1) Select the apex of the annotation

Select

Editable properties

Arrow, Ellipse, Rectangle, Polygon

Width and colors of lines

Text

Font (character, style, size), font color, background

Measure line segment, Simplified cardiothoracic ratio, Cardiothoracic ratio, Interline angle, Dual angle, Cobb angle

Width and colors of lines

Marker

Direction of marker (horizontal flipping, rotation), font color, background

Ellipse ROI, Rectangular ROI, Polygonal ROI

Display the line style, line color, measurement angle, and supplementary angle

Selection Items

2) Drag the mouse pointer to the destination → The linear length is remeasured while moving and displayed

Items

Contents

Width of line

Select the thickness of lines of each annotation *1

Color of line

Select the colors of lines of each annotation *2

Font Drag

Character

Select the type font

Style

Select from normal, italic, bold, bold italic Select from 5 to 96 points

Size Font color and background Font color

Select the font color *2

Background

Select the background *2

Flip horizontal

5-26

FCR PRIMA Console Reference Guide

897N101440B

Items

■ Deleting Added Annotations

Contents

Rotate

Original Image 90 right rotation 180 rotation 90 left rotation

Font color and background (A)

(C)

1. Select the annotation to be deleted.

(D)

A:W  hite characters on black background B:W  hite characters on transparent background C:B  lack characters on white background D : Black characters on transparent background Select the area of angle to be measured • Reference angle Base of an angle at the time of measurement • Adjacent angle 1/Adjacent angle 2 An angle next to the reference angle • Vertically opposite angle An angle on the diagonal line of the reference angle Vertically opposite angle

Adjacent angle 1

Adjacent angle 2

2. Right-click anywhere in the image pane and click [Delete].

→→The selected annotation will be deleted.

5 Verifying and Processing Images

Measurement angle

(B)

If a new annotation is added surrounding the original annotation, the original annotation cannot be selected. By moving or editing the added annotation, the original annotation can be selected again.

To delete all of annotations on the image, right-click the image pane and select "Delete all annotations on image".

Vertically opposite angle Adjacent angle 2

Reference angle

Supplementary Put a check mark when displaying angle will be the supplementary angle of an displayed angle measured *1 In case of an arrow, the apex of an arrow becomes bigger or smaller in proportion to the setting of width of line. *2 Color which has been set displays on the box. By clicking the box, the window to select colors opens and the color can be changed.

3. Click [Apply] (E). →→Properties changes according to the settings.

FCR PRIMA Console Reference Guide

897N101440B

5-27

5.6

Free Drawing on “Image Viewer” for Informed Consent

Lines can be drawn freely on the “Image Viewer” window with a pen.

■ Operation Windows ● “Other” Tool Bar

■ Using the Free Drawing Function During use of the free drawing function, other operations cannot be used. To use operations other than free drawing function, click

5

drawing mode off, or click drawing mode.

to turn the free

to cancel the free

Verifying and Processing Images

1. Click [Free Drawing] (A) on the "Other" tool bar.

You can do the same by right-clicking the image and selecting "Free Drawing". →→"Function" window appears.

(A)

2. Click each button (B) on the "Function"

● “Function” Window

window to free draw.

Functions of Each Button Buttons

Functions Pen size Click and select from the following pen size • Thin pen (3 points) • Medium pen (9 points) • Thick pen (27 points) Pen color Click and select the color Drawing mode ON/OFF Switching the function ON/OFF each time of clicking

(B) *1 *2

By clicking, capture the drew line and screen Delete all the drawn lines

5-28

FCR PRIMA Console Reference Guide

897N101440B

*1 Captured image data displays in the "Paint". *2 When images are displayed on multiple monitors, the image data of both monitors are copied to the clipboard as one file. In case of different screen sizes, the images copied to the clipboard with their upper edges aligned. Image save arrangement: Monitor configuration Monitor 1

Monitor 2

Image arrangement Monitor 1

Monitor 2 Monitor 1

Monitor 2

5

Monitor 2 Monitor 1

Verifying and Processing Images

3. Drag the mouse on the “Image Viewer” window to draw a line.

4. To exit the free drawing function, click

(B).

→→The “Function” window closes.

• When the free drawing function is ended, the drawn lines are all deleted. • When exiting the Free Drawing function, the selected pen size and pen color are stored, and they are set again when the Free Drawing function is started again.

FCR PRIMA Console Reference Guide

897N101440B

5-29

5.7

Saving/Verifying/Suspending Images

Saves the image processing result and annotation, makes the study status of images "Suspend" or "Verified".

The studies whose status can be suspended or verified are the completed studies.

■ Operation Window

■ Verifying/Suspending Images

● “Sub-window”

5

If the study has been verified, the be clicked.

Verifying and Processing Images

1. Select the study and click (A)

(B) (C)

or

(B) or

cannot

(C).

Functions of Each Button Buttons

Functions

[Suspend]*1 Leaves the study status as "Completed" [Verify]*2

■ Saving the Image Processing

The quality of images in the study is verified. The study is updated from the "Completed" to the "Verified" study status

Result and Annotation

Once the study is saved, it cannot be undone by [Reset], [Restore]/[Retry].

The following contents can be saved. • Rotated or flipped images • FCR Image Processing Parameter • Assigned annotation (Arrows, ellipses, rectangles, polygons, texts, markers) • General measurement (Measure line segment, simplify cardiothoracic ratio, cardiothoracic ratio, interline angle, hip joint angle, VHS) • Pixel measurement (Ellipse ROI, rectangular ROI, polygonal ROI)

1. Click

(A) in the sub-window.

→→The image processing result and annotation are saved.

5-30

FCR PRIMA Console Reference Guide

897N101440B

*1 Suspended studies are set aside from the scope of automatic verification. “14.6 Auto-Determine Settings" *2 Images are automatically printed, saved, or transmitted at the time of verification if settings are configured to do so. • Saving to a storage disk • Printing images in a study on a film (optional) • Transmitting to a DICOM server(optional) “14.5 Setting of Printing, Saving and Transmitting Images"

Chapter

6

Using Report Function 6

Outline of “Report” Window............................................... 6-2

6.2

Creating/Editing the Report................................................ 6-3

6.3

Editing the Schema or Fixed Phrase.................................. 6-6

FCR PRIMA Console Reference Guide

897N101440B

Using Report Function

6.1

6-1

6.1

Outline of “Report” Window (1)

(2) (3)

(1) Print/Save as PDF file The report format can be specified, and the report can be printed out or saved as PDF file. “8.5 Printing a Report”

(4)

(2) Save/Verify/Re-edit The created or edited report can be saved or verified. The saved/verified report also can be edited again. “6.2 ■Saving/Verifying a Report”

(3) Delete The report can be deleted. “6.2 ■Deleting a Report”

6 (5) Using Report Function

(4) Schema • Select the same schema as the exposure region and assign it. “6.2 ■Creating/Editing the Report”

• The annotation function and the free drawing function can be used on the assigned schema. “6.2 ■Creating/Editing the Report”

• The schema can be registered or deleted. “6.3 ■Editing the Schema” (6)

(5) Thumbnail Assign thumbnail images of the study. “6.2 ■Creating/Editing the Report”

(6) Comment • Comment can be input to the target patient or study images. “6.2 ■Creating/Editing the Report”

• The registered fixed phrase can be inserted. “6.2 ■Creating/Editing the Report”

• The fixed phrase can be edited/added. “6.3 ■Editing the Fixed Phrase”

6-2

FCR PRIMA Console Reference Guide

897N101440B

6.2

Creating/Editing the Report

Creates or edits a report to save or verify.

■ Operation Windows

● “Select schema” Window

● “Basic Operation Tool bar” (T)

(A)

● “Report” Window

6

(B)(C)(D) (G)

(H)

(U)

(I) (J)

Using Report Function

(F)

(E)

● “Select fixed phrase” Window (V)

(K)

(L)

(M) (N)

(X) (W)

● “Register” Window (Y)

(R) (Z) (O)

(P)

(Q) (S)

The display settings of a report can be performed in the [Customize] - [Report] in “Image Viewer” window. “11.2.6 Display Settings of Report”

FCR PRIMA Console Reference Guide

897N101440B

6-3

■ Creating/Editing the Report 1. Display the image in the sub-window. “5.3 Selecting and Arranging Images” →→The “Report” window opens if the image is to be created in the report.

2. To create a new report, click [Open report] (A) of the tool bar.

→→The “Report” window opens.

• You can do the same by right-clicking anywhere on the image pane and selecting [Open report]. • If the default schema is set, it will be assigned automatically.

6

5. Attach an annotation to the schema or draw freely(J).

Functions of Each Button Buttons

Select an annotation on the schema *1

*1

Drag from the start point to the end point

*1

Enter a text by clicking the position to enter a text

*2

*1

“6.3 ■Editing the Schema”

Using Report Function

*1*3

3. To set a schema, click

(G).

Functions

Draw a line on the dragged track by dragging After an arrow is drawn by dragging the start and end point of an arrow, enter a text on the end point After an arrow is drawn by dragging the start and end point of an arrow, the number displays automatically on the end point

→→The “Select schema” window opens.

• Click

(F) to assign a favorite schema.

“6.3 ■Editing the Schema” • Click (H) to open the “Schema setting edit” window opens. “6.3 ■Editing the Schema” • Click

( I ) to clear the assigned schema.

*1 The operation of an annotation on a schema is the same as the operation on the “Image Viewer” window. ”5.5 ■Adding Arrows/Shapes (Ellipses, Rectangles, and Polygons)”, ”5.5 ■Adding Texts” “5.5 ■Editing the Added Annotations” “5.5 ■Deleting Added Annotations” *2 Lines drawn can be edited same as annotations.

4. Select a schema to be used(T) and click [Select] (U).

→→The selected schema is assigned.

You can do the same by double-clicking the schema to be used.

“5.5 ■Editing the Added Annotations” *3 The number can be set in the column at the right side of the button.

6. Drag and drop the image to be displayed in the report to the thumbnail (N).

→→The thumbnail images are assigned.

• Set to ON by

(K), reprint including

annotations at the time of printing report. • Click (L) to register the key image as a thumbnail image. “5.3 ■Registering Displayed Images as a Key Image” • Click (M) to clear the thumbnail image in the thumbnail column.

6-4

FCR PRIMA Console Reference Guide

897N101440B

7. Input comments in each column (R). Click the desired button to use the following functions. Functions of Each Button Buttons

Functions

(O)*1*2

Insert a registered fixed phrase. 1. Select an fixed phrase (V) in the “Select fixed phrase” window. 2. Click [Select] (X). → The fixed phrase is inserted.

(P)

Clear the currently input text.

(Q) (S)*3

Insert the text “Nothing abnormal” in all comment columns.

*1 Click the [Select](X) and the counts in “The number of use” of the inserted fixed phrase increases. *2 Put a check mark to “Close the screen after inserting”(W) to close the window and return to the “Report” window after clicking [Select](X). *3 It is possible to edit the text other than “Nothing abnormal”. “11.2.6 Display Settings of Report”

• Save: Save the report temporarily. The report can be edited. • Verify: Verify the image together. The report cannot be edited.

1. Click

(B) or

(C).

Functions of Each Button Buttons

Functions Save the report Verify the report

6 is displayed in the thumbnail of “Refer” window. Also, “ ” is attached to the target tree of the study in study summary.

→→

Using Report Function

Register the currently input text as a fixed phrase. 1. Input in the [Heading] (Y) in the “Register” window. 2. Click [Register] (Z). → The fixed phrase is registered.

■ Saving/Verifying a Report

• The study can be verified automatically again at the time of clicking

(B)

“11.2.6 Display Settings of Report” • Verified report contents can be edited by clicking (D). Editing report:“■Creating/Editing the Report”

■ Deleting Reports 1. Open a report. 2. Click

(E).

→→The message to confirm the deletion appears.

3. Click [OK].

FCR PRIMA Console Reference Guide

897N101440B

6-5

6.3

Editing the Schema or Fixed Phrase

The schema can be registered or deleted, and the fixed phrase can be added or edited.

■ Operation Windows

● “Select fixed phrase” Window

● “Report” Window

(I) (J) (K) (L)

(A)

● “Register” Window

6 Using Report Function

(M)

(N) (B)

● “Schema setting edit” Window

(C)

6-6

(D)

(E) (F)

FCR PRIMA Console Reference Guide

(G)

897N101440B

(H)

■ Editing the Schema Types of schema can be added/deleted, and the default schema or favorite schema can be set.

1. Click the

(A) in the schema column.

■ Registering the Fixed Phrase 1. Click

(B) in the comment column in which the fixed phrase is to be registered.

→→The “Select fixed phrase” window opens.

→→The “Schema setting edit” window opens.

2. Edit the schema.

Click when the comment column is input with text to register it as the fixed phrase.

Functions of Each Button Buttons

Functions

[New item] (E) 1) Select “User setting” tab (C). 2) Click [New item] (E). → The “Open” window opens. 3) Select the file of schema. 4) Click [Open]. [Delete] (D)

Delete the selected schema. 1) Select the schema (C) *1 (G) 2) Click [Delete] (D).

1) Click (F) of default schema or favorite schema according to the destination to register.

→→The “Register” window opens.

3. Input the fixed phrase (M).

6

Input Items Items

Contents

Heading

Input the heading of the fixed phrase

Content

Input the contents of the fixed phrase

Using Report Function

Register the selected schema as a default schema or favorite schema. Select the schema (C).

2. Click [Add] (K).

4. Click [Register] (N).

*1 Only the schema in the “User setting” tab can be deleted.

3. Click [Setting is completed] (H).

■ Editing the Fixed Phrase 1. Click

(B) in the comment column in which the fixed phrase to be edited is registered.

→→The “Select fixed phrase” window opens.

2. Select the fixed phrase ( I ). 3. Click [Edit] (J). →→The “Register” window opens.

Clicking [Delete] (L) to delete the selected fixed phrase.

4. Edit the fixed phrase (M). Input Items Items

Contents

Heading

Edit the heading of the fixed phrase

Content

Edit contents of the fixed phrase

5. Click [Register] (N). FCR PRIMA Console Reference Guide

897N101440B

6-7

6 Using Report Function

6-8

FCR PRIMA Console Reference Guide

897N101440B

Chapter

7

Image Storage from Each Window 7 Relationships between Windows and Storage Destinations... 7-2

7.2

Displaying Dialogs for Saving............................................ 7-3 7.2.1

Starting from the “Refer” Window................................ 7-3

7.2.2

Starting from the “Image Viewer”................................. 7-4

7.2.3

Starting from the “Study Utilities”................................ 7-5

7.3

Saving to Storage Disks...................................................... 7-6

7.4

Transmitting Studies to DICOM Server (Optional)............ 7-7

7.5

Saving to Work Disks.......................................................... 7-8

7.6

Saving in a General-Purpose File Format (Optional)...... 7-10

7.7

Saving to PDI Disks........................................................... 7-13

FCR PRIMA Console Reference Guide

897N101440B

Image Storage from Each Window

7.1

7-1

7.1

Relationships between Windows and Storage Destinations

Saves images from the “Refer” window, “Image Viewer”, and “Study Utilities”. Displays dialogs for saving in accordance with media type in each window, and then specify the storage method such as priority and density before saving. Set Storage Method

Display the Storage Setting Window

Save Images to Destination

“Main” Window - “Refer” Window

Click General-Purpose Image

7

Storage Setting Dialog of each window (The following dialog is an example when saving to storage disks.)

“Image Cart”

Image Storage from Each Window

DICOM Server

Click

Work Disks Storage Disks

PDI Disks (CD-R, DVD-R)

“Study Utilities”

Click

NAS

● List of Destinations that can be Used from Each Window

7-2

Storage Disk (NAS)

DICOM Server

Work Disk

General-Purpose File Format

PDI

“Refer” Window





×

×



Image Viewer

×

×

×





Study Utilities







×



FCR PRIMA Console Reference Guide

897N101440B

7.2

Displaying Dialogs for Saving

Saves images from the “Refer” window, “Image Viewer”, and “Study Utilities”.

7.2.1 Starting from the “Refer” Window Searches the patient and saves the images shown in the “Refer” window to each media. This section describes how to display dialogs for saving from the “Refer” window according to the storage media.

■ Operation Window

■ Displaying Dialogs for Saving

● “Main” Window - “Refer” Window

1. Search the patient.

Select the patient and the image to be referred.

“4.2 Searching for Patients”

2. From the “patient list” (A), select the patient in a “Verified” study status.

7

→→The image is displayed in the “Refer” window (B).

Image Storage from Each Window

3. Select the image (B) to be saved. 4. Click the desired button (C) according to the storage media.

Functions of Each Button Buttons

Functions Save to storage disks*1

(A)

“7.3 Saving to Storage Disks” Transmit to the DICOM Server*1 “7.4 Transmitting Studies to DICOM Server” Save to PDI Disks*1 “7.7 Saving to PDI Disks”

(B)

*1 The image whose study status is “Verified“ cannot be saved or transmitted.

(C) *1 If the selected studies except for “Verified” are selected, a message displays to verify. *1 When the buttons (C) are not displayed, operation items must be added in the “Display Settings”.

To continue excluding the studies which are not able to process, click [OK].

“11.1.7 Display Settings of “Refer” window”

FCR PRIMA Console Reference Guide

897N101440B

7-3

7.2.2 Starting from the “Image Viewer” Saves images registered in the “Image Cart” in the “Image Viewer” window. This section describes how to display dialogs for saving from the “Image Cart” according to the storage media.

■ Operation Window

■ Displaying Dialogs for Saving

● “Image Cart”

1. Drag and Drop the images to the “Image

(A)

Cart”.

• The images can be dragged and dropped from the “study summary”, thumbnail images in the subwindow, and the “Refer” window. • You can do the same by right-clicking the images in the sub-window and selecting “Put in an image cart”. • Click

7

to clear the images in the “Image Cart”.

Image Storage from Each Window

2. Click the desired button (A) according to the storage media.

Functions of Each Button Buttons

Functions Save in a DICOM format “7.6 Saving in a General-Purpose File Format (Optional)” Save in a BMP/JPEG file format “7.6 Saving in a General-Purpose File Format (Optional)” Save to PDI Disks “7.7 Saving to PDI Disks”

7-4

FCR PRIMA Console Reference Guide

897N101440B

7.2.3 Starting from the “Study Utilities” Searches the study and saves the images shown in the list to each storage media. This section describes how to display dialogs for saving from the “Study Utilities” according to the storage media.

■ Operation Window

■ Displaying Dialogs for Saving

● “Study Utilities”

1. Search the study. “12.3 Searching for Studies and Creating New Queries” (A)

2. From the “patient list” (A), select the patient in a “Verified” study status.

(B)

3. Click the desired menu (B) in [Storage/Print] according to the storage media. Selection Items Items

7

Contents

Image Storage from Each Window

Storage Disk Save to storage disks*1 “7.3 Saving to Storage Disks” DICOM Transmit to the DICOM Server*1 Transmission “7.4 Transmitting Studies to DICOM Server” Work Disk

Save to work disks*1 “7.5 Saving to Work Disks”

Create PDI

Save to PDI Disks*1 “7.7 Saving to PDI Disks”

*1 The image whose study status is other than “Verified“ cannot be saved or transmitted.

If the selected studies include those other than “Verified“, a message displays to verify. To continue excluding the studies which are not able to process, click [OK].

FCR PRIMA Console Reference Guide

897N101440B

7-5

7.3

Saving to Storage Disks

Saves images automatically to the storage disk set in the PC’s control drive when the study is verified. Automatic storage: “14.5 Setting of Printing, Saving and Transmitting Images”.

This section describes how to save or save again when [Automate Saving] is not activated.

If the same study already exists on the destination storage disk, it is overwritten.

■ Operation Window

■ Saving the Screen

● “Storage Disk Settings” Window

1. Select the Priority (A).

7

LOW: Give higher priority to other save operations. MEDIUM: Save with the ordinary priority. HIGH: Start to save immediately.

(A)

Image Storage from Each Window

(B)

“2.7 Before Using Disks“.

7-6

FCR PRIMA Console Reference Guide

897N101440B

2. Click [Output] (B). When the disk access lamp is ON, do not press the Eject button on the disk drive.

7.4

Transmitting Studies to DICOM Server (Optional)

Transmits the unprocessed FCR study images to the DICOM server when the study is verified. “14.5 Setting of Printing, Saving and Transmitting Images”

This section describes how to manually save or save again when [Automate Saving] is not activated.

■ Operation Window

■ Saving the Screen

● “Settings for DICOM Transmission”

1. Set the transmission settings (A).

Window

Selection Items Items

Contents

Destination Specify a destination Example: DICOM Server (A)

Density

(B)

[Default Pixel Count]*1: Transmit images with a reduced pixel density and hence in a smaller file size

7

Priority

Image Storage from Each Window

[Max Pixel Count]: Transmit images with their pixel density unchanged [LOW]: Give higher priority to other save operations [MEDIUM]: Save with the ordinary priority [HIGH]: Start to save immediately *1 The transmitted image size decreases.

2. Click [Output] (B).

FCR PRIMA Console Reference Guide

897N101440B

7-7

7.5

Saving to Work Disks

Saves all the images in a selected study to work disks.

• If the same study already exists on the destination work disk, it is overwritten. • The Magnify/Reduce of image has no effect on storage to work disk.

■ Operation Windows

■ Saving the Screen

● “Work Disk Storage Settings” Window

1. Insert a work disk in the control drive of the

(Start from the “Study Utilities”)

PC.

2. Set the transmission settings (A). Selection Items

7

(A)

Items Destination

Contents Specify a drive to be used Example: DVD

Image Storage from Each Window

Compression [Uncompressed]: Save images without compression (B)



“2.7 Before Using Disks“.



“10.3 Verifying Disk Status and Maintenance“.

[JPEG Lossless]: Save images with compression The image file size decreases Compressed images can fully be restored Density

[Default Pixel Count]: Save images with a reduced pixel density and hence in a smaller file size [Max Pixel Count]: Save images with their pixel density unchanged

Image Proc.

[Processed]: Perform FCR image processing and display gradation correction for the pixel data in the images to be saved*1 [Unprocessed]: The FCR image processing and display gradation correction for the pixel data are not performed to the images to be saved*2

Annotation Output

[No]: Annotations are not saved to work disks [Yes]: Annotations are save to work disks

7-8

FCR PRIMA Console Reference Guide

897N101440B

*1 Image data thus processed is accessible to the DICOM server for viewing the same way as FCR PRIMA Console views them when the DICOM server does not support FCR image processing and display gradation correction processing. *2 Choose [Unprocessed] to handle images on equipment that supports FCR image processing.

Although images that have been saved with [Processed] being selected can be reloaded into the FCR PRIMA Console, the images thus loaded cannot be printed on film (optional). If you plan to perform these operations, select [Unprocessed] when saving.

3. Click [Output] (B). When the disk access lamp is ON, never press the Eject button on the disk drive.

7 Image Storage from Each Window

FCR PRIMA Console Reference Guide

897N101440B

7-9

7.6

Saving in a General-Purpose File Format (Optional)

Images for a study can be saved in a general-purpose file format, such as “.jpg” or “.bmp”. Images saved in these formats can be used for purposes such as presentation.

Patient information file

Image files

xxx.txt

xxx.jpg

7

xxx.bmp

Image Storage from Each Window

(The file format has been set at installation.)

No patient information file is created for images saved in the DICOM server.

• The destination is determined by the installation setting. • If modifications have been made to the status of image display, some modifications have effects on film print but some do not. - FCR Image Processing : Applicable - Annotation : Applicable - Rotate/Flip : Applicable - Magnify/Reduce : Not applicable - Gradation Correction : Applicable - Reverse : Applicable • The image displayed in the sub-window can be saved as a thumbnail image by clicking the “Thumbnail image transfer” of “Other” tool bar.

7-10

FCR PRIMA Console Reference Guide

897N101440B

■ Operation Windows

■ Saving the Screen

● “General-Purpose File” Window (as a

● Saving in a BMP or JPEG File Format

BMP or JPEG File Format)

All the images for a selected study can be saved in a general-purpose file format.

1. Set the transmission settings (C). (C)

Input / Selection Items Items

Contents

Save format

(D)



Patient information files are always saved in the text format The file extension is “.txt” Image quality*1

“DICOM file output” Window

JPEG: JPEG format The file extension is “.jpg” BMP: BITMAP format. The file extension is “.bmp”

Specify the image quality when the image is saved in the JPEG format

7

[Standard]: Save images with standard image quality and standard compression

(E)

[High]: Save images with high image quality and low compression (F)

Annotation Output

[Yes]: Annotation is saved with the image [No]: Annotation is not saved with the image

Destination folder for storage

Click , and the “Browse for Folder” window opens. Specify a destination folder to save

File name

Input a file name

*1 Set only if saving in JPEG file format.

2. Click [Output] (B). • Saving progress can be confirm in the “Progress of saving” column • When the same file is already existed, the message to confirm whether to overwrite or not appears. →→The “General-Purpose File” window closes after saving process completed.

3. Check if the files have been created in the specified folder.

FCR PRIMA Console Reference Guide

897N101440B

7-11

Image Storage from Each Window

[Low]: Save images with low image quality and high compression

● Saving in a DICOM Format

All the images for a selected study can be saved in a DICOM format.

1. Set the transmission settings (E). Selection Items Items

Contents

Output images Put a check mark to output images with annotations embedded with annotations embedded Put a check mark to save the Save information information with multi-byte characters (Kanji, Kana) Folder targeted for data storage

Input a storage destination folder name Click to open the “Browse For Folder” window, and the folder in which files is saved can be specified.

File name

Input a file name

7 Image Storage from Each Window

2. Click [Next] (F). →→The saved file name and percentage progress of the storage can be verified.

When the same file is already existed, the message to confirm whether or not to overwrite appears.

3. Click [Close] after the storage completes. →→The “Save multi-purpose file” window close.

4. Check if the files have been created in the specified folder.

7-12

FCR PRIMA Console Reference Guide

897N101440B

7.7

Saving to PDI Disks

The study images can be saved to CD-R disks or DVD-R disks in PDI format.

• Studies that can be saved as a PDI file are only those with a “Verified” study status. • Unformatted or unused formatted 640MB or 700MB CD-R disks can be used. • Unformatted or unused formatted 4.7GB DVD-R disks can be used. • Before saving study image in PDI disk, write the following information on the surface of the PDI disk to prevent confusing with study image for other patient ID. - Patient name - Birth date - Disk creation date - Study date (write all if more than one) - Facility name • The process takes time when a large quantity of PDI data is saved to a DVD-R disk. As a guide value for the processing time, it takes about 15 minutes to save all measuring data.

7 Image Storage from Each Window

FCR PRIMA Console Reference Guide

897N101440B

7-13

■ Operation Windows

■ Saving the Screen

● ”PDI creation wizard”

1. Insert a CD-R or DVD-R into the disk drive tray.

(A)

2. To output images with annotations

embedded, put a check mark to “Output images with annotations embedded” (A).

3. Click [Next] (B). →→Writing to the CD-R or DVD-R disk starts. (B) Progress is displayed when write to CD-R or DVD-R disk starts. Do not press the Eject button or use any other program while writing. Otherwise, write may not complete properly.

● “Disk writing finished.”

Note that once write starts, it cannot be canceled. (C)

7 Image Storage from Each Window

The following messages might be displayed. To continue the PDI file creation, perform the following procedures. • “No disk is inserted.”, or “Unable to write to this disk.” Reset the correct CD-R or DVD-R disk and then click [Next] in the “PDI Creation Wizard” window.

(D)

• “Unable to write to this disk because the capacity is exceeded.” Remove the CD-R or DVD-R disk from the disk drive tray and click [OK]. If multiple studies are selected, reduce the number of studies and redo with a new CD-R or DVD-R disk. →→The dialog “Disk writing finished” (C) opens when write to CD-R or DVD-R disk completes, and the drive tray opens.

4. Remove the CD-R or DVD-R disk from the disk drive tray.

5. Click [Close] (D). If a PDI file is created in the storage disk drive, set the storage disk in the drive.

7-14

FCR PRIMA Console Reference Guide

897N101440B

Chapter

8

Image Print from Each Window 8.1

Printing Types and Flow...................................................... 8-2

8.2

Displaying Print Dialogs...................................................... 8-4 Starting from the “Refer” Window................................ 8-4

8.2.2

Starting from the “Image Viewer”................................. 8-5

8.2.3

Starting from the “Study Utilities”................................ 8-6

8.3

Printing on a Film (Optional)............................................... 8-7

8.4

Free Layout Printing............................................................ 8-9

8.5

Printing Report................................................................... 8-13

FCR PRIMA Console Reference Guide

897N101440B

Image Print from Each Window

8.2.1

8

8-1

8.1

Printing Types and Flow

This section describes the types of printing applicable to FCR PRIMA Console and the flow of printing.

■ Types of Printing The following printing methods are applicable to FCR PRIMA Console.

● Film Print (Optional)

Prints images on a film with the destination and print count set. An unprocessed FCR study image prints automatically on the film when completing an exposure or verifying a study. Automatic output: “14.5 Setting of Printing, Saving and Transmitting Images” When automatic output is not activated or print again, manually printing is possible.

When [Output after Re-verified] in the “User Utilities” printer settings has been unchecked (setting for no film printing at the time of inspection re-verification), print output has to be performed manually in image units at the time of adding study menu.

Check the indicator to check the printer status.

8

“10.5 Verifying Film Printer Status (Optional)”

Image Print from Each Window

● Free Layout Print

Prints multiple images to one piece of film with a free layout. The image size can be adjusted, and the additional image information or annotation can also be added to the output.

● Print Report

Prints reports created in the “Image Viewer” window. ”6.2 Creating/Editing the Report”

8-2

FCR PRIMA Console Reference Guide

897N101440B

■ Printing Flow FCR PRIMA Console can print a study from the “Refer” window, “Image Viewer” window, and “Study Utilities”. Prints studies by displaying the printing window according to the printing method on each window and setting contents for printing such as selecting the object to be printed or numbers of printings.

Set Printing Method

Display Print Dialog

Print

“Main” Window ー”Refer” Window

Film Print 04001676 A 0000 A 020

FUJI HOSPITAL

G 1.1G#0.9+0.30 R 5TO.3 D 2MO.2

SKULL, GENERAL

L 2.0 S 187 C *1.0*1.0 AP

0000000001 John Smith 1992. 11. 27 [17:43]

R-> L

[M] 23 SCALE: 50% RT-04

G 1.0G#0.9+0.30 R 5T0.31. D 2MO.2

Free Layout Print “Image Viewer”

8 Image Print from Each Window

Report Print “Study Utilities”

● List of Print Destination that can be Used from Each Window Film Print

Free Layout Print

Report Print

“Refer“ window





x

Image Viewer







Study Utilities





x

FCR PRIMA Console Reference Guide

897N101440B

8-3

8.2

Displaying Print Dialogs

Images can be printed from the “Refer” window, “Image Viewer”, and “Study Utilities”.

8.2.1 Starting from the “Refer” Window Searches the patient and prints the images shown in the “Refer” window to each media. This section describes how to display print dialogs from “Refer” window according to the printing method.

■ Operation Window

■ Displaying Print Dialogs

● “Main“ Window - “Refer“ Window

1. Search the patient.

Select the patient and the print dialog appears.

“4.2 Searching for Patients”

2. From the “patient list” (A), select the patient in a “Verified” study status.

→→The image is displayed in the “Refer” window (B).

3. Select the image (B) to be printed.

8

4. Click the desired button (C) according to the

Image Print from Each Window

printing method.

Functions of Each Button (A)

Buttons

Functions Print images on a film “8.3 Printing on a Film (Optional)”

(B)

(C) *1

Print image with a free layout print “8.4 “Free Layout Printing“

If the selected studies include those other than “Verified“, a message displays to verify. To continue excluding the studies which are not able to process, click [OK].

*1 When the buttons (C) are not displayed, operation items must be added in the Display Settings. “11.1.7 Display Settings of “Refer” window”

8-4

FCR PRIMA Console Reference Guide

897N101440B

8.2.2 Starting from the “Image Viewer” Prints images displayed in the “Image Cart” of “Image Viewer”. This section describes how to display print dialogs from the “Image Viewer” according to the printing method.

■ Operation Window

■ Displaying Print Dialogs

● “Image Cart”

1. Drag and Drop the images to the “Image

(A)

Cart”.

• The images can be dragged and dropped from the “study summary”, thumbnail images in the subwindow, and the “Refer” window. • You can do the same by right-clicking the images in the sub-window and selecting “Put in an image cart”. • Click

to clear the images in the “Image Cart”.

2. Click the desired button (A) according to the

8

printing method.

Functions of Each Button

Image Print from Each Window

Buttons

Functions Print images on a film “8.3 Printing on a Film (Optional)” Print image with a free layout print “8.4 Free Layout Printing”

FCR PRIMA Console Reference Guide

897N101440B

8-5

8.2.3 Starting from the “Study Utilities” Searches studies and prints the images shown in the list. This section describes how to display print dialogs from the “Study Utilities” according to the printing method.

■ Operation Window

■ Displaying Print Dialogs

● “Study Utilities”

1. Search the study. “12.3 Searching for Studies and Creating New Queries” (A)

2. From the “patient list” (A), select the patient in a “Verified” study status.

3. Click the menu (B) in [Storage/Print] according to the printing method.

(B)

Selection Items Items

Contents Print images on a film

8

Film Print

Image Print from Each Window

Free Layout

“8.3 Printing on a Film (Optional)” Print image with a free layout print “8.4 “Free Layout Printing“

*1 The images whose study status is other than “Verified“ or “Completed“ cannot be printed.

If the selected studies include those except for “Verified“ or “Completed“, a message displays to verify. To continue excluding the studies which are not able to process, click [OK].

8-6

FCR PRIMA Console Reference Guide

897N101440B

8.3

Printing on a Film (Optional)

Prints images on a film with the destination and print count set. The setting items shown in the print dialog are different between starting from the “Refer” window / ”Study Utilities” and starting from the “Image Viewer”.

Image magnification and reduction will have no effect on the printing of images in the case of [Film Print].

■ Operation Windows

■ Printing on a Film

● “Film Print Settings” Window

(Start from the “Main” Window - “Refer” Window or “Study Utilities”)

● Starting from the “Main Window” -

“Refer” Window or “Study Utilities”

1. Set the film print settings (A). Selection Items Items (A)

Contents

Destination

Print destination

Target

Study imgs yet-to-be output

● “Film Print” Window

(Start from the “Image Viewer”)

Print Count

Image Print from Each Window

(B)

8

All imgs Print all the images in the study on included in a a film. Images already printed on a study film are reprinted Print only those images in a film that have not been printed on a film 1 to 9 copies can be printed

Priority

(C)

LOW

Give higher priority to other save operations

MEDIUM

Save with the ordinary priority

HIGH

Start to save immediately

2. Click [Output] (B).

(D)

FCR PRIMA Console Reference Guide

897N101440B

8-7

● Starting from the “Image Viewer” 1. Set the film print settings (C). Input and Selection Items Items

Contents

Print Destination

Print Destination

Print Count

1 to 9

Print Format Single Format

2 ON 1 Format*1

4 ON 1 Format*2

8 Image Print from Each Window

Prioritized Output

Print images in this study on a priority basis

Film Count

The film requirement is calculated from the “Print Count” and “Print Format” and is shown in the “Film Count” field

Annotation Output Yes

Output annotations

No

Not output annotations

*1 P  rint two different images on a single film. Printed images are arranged from left to right in the study window. *2 P  rints four different images on a single film. Printed images are arranged from upper left to upper right, lower left to lower right in the study window.

2. Click [Output] (D).

8-8

FCR PRIMA Console Reference Guide

897N101440B

8.4

Free Layout Printing

Prints multiple images to one piece of film with a free layout. The image size can be adjusted, and the additional image information or annotation can also be added to the output.

■ Operation Windows

● “Format information setting“

Create/Set additional image information format.

● “Free layout print”

(H)

Set the layout of images, display/hide the additional image information and annotation. (A)

(B)

(I) (C) (D)

(J)

● “Input format name“ (E)

● “Footer information setting“

Sets the additional image information format name.

8

(K)

Image Print from Each Window

(L)

Set to display/hide footer and the footer information.

(F)

(G)

FCR PRIMA Console Reference Guide

897N101440B

8-9

■ Free Layout Printing

Images can be moved inside the cell. Move the mouse pointer over the image and it will turn to .

1. Select Printer and Film Size (C). 2. Select layout orientation. (Landscape:

, Portrait:

Specify the range by mouse dragging to trim and display the specified range. The trimmed range will be displayed with the frame fitted to the cell window. If moving the mouse pointer over the image, the shape of the mouse pointer will turn to .

)

→→A message displays to confirm, click [OK].

3. Select print format (C). 1) Click . 2) Move the pointer on the grid and click to specify rows and columns.

When the cursor is moved onto the border of the

Change the window level to a monochrome image. If moving the mouse pointer over the image, the shape of the mouse pointer will turn to .

thumbnail display section (A).

The value of WC (Window Center) and WW (Window Wide) can be changed according to the dragging direction of the mouse pointer. • Up: WC value increases. • Down: WC value decreases. • Left: WW value increases. • Right: WW value decreases.

Images can be switched by following buttons.

Switch the position of the image. Move the mouse pointer over the image and it will turn to .

cell, it changes to

or

and the cell width or

height can be changed.

4. Select the images for the layout from the

8

*1

Image Print from Each Window

: The images of the previous patient appears : The images of the next patient appears : The previous images of the same patient appears : The next images of the same patient appears

5. Drag an image and drop it at the desired layout position (B).

→→A check mark is put on the thumbnail of the image.

6. Adjust image size and position with the desired buttons (D).

Functions of Each Button Buttons

Functions The image zooms in with every click. Move the mouse pointer over the image and it will turn to . The image zooms out with every click. Move the mouse pointer over the image and it will turn to .

a. Drag the selected source image to be switched. b. Drop it onto the target image to be replaced. Undo the change. Delete a selected image. A message displays to verify deletion. Delete all images in the image window. A message displays to verify deletion. *1 Edited WC and WW values can be set to display/ hide as an image attribute information.

The deletion of image cannot be undone. Select an image again on the thumbnail (A) and the image can be added to the image window.

• In magnifying or reducing operation, images can be moved without clicking

.

• The image display positions are switched by drag-and-drop of the image. • The Display/Hide settings for layout format, additional image information and footer are not deleted.

8-10

FCR PRIMA Console Reference Guide

897N101440B

7. Set the footer information. 1) Click (D). 2) Set the footer information (F).

10. Click

(E).

→→Click [OK] if a message displays to verify. The film print starts.

Selection Items Items

Contents

Display footer Select from [Yes] or [No] Footer position

Select from [Left], [Center], or [Right]

When is clicked, Patient ID, patient name, sex, date, and facility name are added to the text box. The added information can be edited. (G).

3) Click

In case of landscape layout, the footer information is displayed on the right side.

8. Select the annotation, additional image

8

information, and ruler to display/hide.

Buttons

Image Print from Each Window

Selection Items Functions All the annotations appear All the additional image information appear The scale of the selected cell displays All the scale of the selected cell displays

9. Select the additional image information format when displaying them.

1) Click (D). The “Format information setting” window opens. 2) Select the format in “Current Format“ (H). 3) Click [OK].

The additional image information format can be created or set in the “Format information setting” window. “■Creating Additional Image Information Format” “■Editing Additional Image Information Format”

FCR PRIMA Console Reference Guide

897N101440B

8-11

■ Creating Additional Image

■ Editing Additional Image

1. Click

1. Click

Information Formats (D).

→→The “Format information setting“ window opens.

2. Click

Information Formats (D).

→→The “Format information setting“ window opens.

2. Select the format in “Current Format“ (H).

(H).

→→The “Input format name“ window opens.

3. Enter format name (K). 4. Click [OK] (L).

→→The displaying position and contents of additional image information are shown.

3. Modify items to be displayed as creation. “■ Creating Additional Image Information Formats”

→→Return to the “Format information setting“ window. The format name input in 3. shows in the “Current Format“

5. Select the position and items to be displayed

4. Click [OK] (J). →→The format is created. Return to the “Format information setting“ window.

in the additional image information ( I ).

1) Select the position of the additional image information on [Top-left], [Top-right], [Bottom-left], or [Bottom-right]. 2) Select the items to be displayed from the “List of information items”.

8 Image Print from Each Window

. 3) Click The selected items are added.

The height displays and outputs in meters, and the weight displays and outputs in kilograms.

The display contents for the additional image information of each image can be set even after image layout.

■ Deleting Formats 1. Click

(D).

→→The “Format information setting“ window opens.

2. Select the format in “Current Format“ (H). For deletion of an item, select the additional image information displayed in the box at [Top-left], [Topright], [Bottom-left], or [Bottom-right] and then click .

→→The displaying position and contents of additional image information are shown.

3. Click

(H).

→→The format is deleted.

6. Click [OK] (J). →→The format is created. Return to the “Format information setting“ window.

■ Finishing 1. Click [Close] in the “Free layout print” window.

→→A message displays to verify.

2. Click [OK]. →→The ”Free layout print” window closes.

8-12

FCR PRIMA Console Reference Guide

897N101440B

8.5

Printing Report

The created report can be printed out or saved as a PDF file.

■ Operation Window

■ Printing a Report

● “Report” Window

1. Select the report format from the pull-down

(A)

(B)

menu (A).

2. Click

(B).

→→The AdobeReader starts and the preview of the report displays.

Click “No, thank you” if “Improvement Program” appears.

3. Print the report.

■ Saving the Report as PDF File

8 Image Print from Each Window

1. Select the report format from the pull-down menu (A).

2. Click

(B).

→→The “Select Folder” opens.

3. Select the folder to be saved and click [OK].

FCR PRIMA Console Reference Guide

897N101440B

8-13

8 Image Print from Each Window

8-14

FCR PRIMA Console Reference Guide

897N101440B

Chapter

9

Editing Study Information

Starting up the Editing Function........................................ 9-2

9.2

Adding Exposure Menus..................................................... 9-4

9.3

Changing Exposure Menus................................................. 9-5

9.4

Deleting Exposure Menus/Canceling the Deletion........... 9-6

9.5

Inactivating Exposed Images/ Activating the Inactivated Images......................................9-7

9.6

Registering Re-exposure Menus........................................ 9-8

9.7

Exchanging Images............................................................. 9-9

9.8

Setting for Printing, Saving, and Transmitting................ 9-10

9.9

Editing Patient Information............................................... 9-13

9 Editing Study Information

9.1

9.10 Changing the Order of Menus in the “Read Image” Window............................................ 9-15

FCR PRIMA Console Reference Guide

897N101440B

9-1

9.1

Starting up the Editing Function

This section describes procedures for starting up the various operation windows to edit studies.

■ Operation Windows

■ Starting up

● “Read Image” Window

● Starting from the ”Read Image “ Window 1. Click

(A).

→→The “Main Menu Modification” window opens.

2. Click the desired button (E). Functions of Each Button (A)

Buttons Add Menu

Functions Exposure menus will be added to the current study “9.2 Adding Exposure Menus”

● “Main” Window - “Refer” Window

Change Menu

Change executed menus and process images for selected part “9.3 Changing Exposure Menus”

Delete/Inactivate Certain menus will be deleted or Menu inactivated for re-exposure

9

“9.4 Deleting Exposure Menus/Canceling the Deletion”

Editing Study Information

(B)

“9.5 Inactivating Exposed Images/Activating the Inactivated Images”

● “Image Viewer” Window

“9.6 Registering Re-exposure Menus” (C)

Relocate Display Exposure menus / images will be Order relocated “9.10 Changing the Order of Menus in the “Read Image “ Window”

(D) Exchange Images

Menus will be associated correctly with images exposed “9.7 Exchanging Images”

● “Main Menu Modification” Window

Output/Storage

Film printout or data storage settings will be changed “9.8 Setting for Printing, Saving, and Transmitting”

(E)

9-2

FCR PRIMA Console Reference Guide

897N101440B

Edit Patient Information

Patient data such as DOB and sex will be corrected “9.9 Editing Patient Information”

● Starting from the “Refer” Window 1. Select the study to be edited from the study list in the “Patient List” window.

2. Click

Window

1. Display the study to be edited in the subwindow.

(B).

“5.3 Selecting and Arranging Images”

→→The “Main Menu Modification” window opens.

3. Click the desired button (E). Functions of Each Button Buttons Add Menu

Change Menu

Exposure menus will be added to the current study

Change executed menus and process images for selected part “9.2 Changing Exposure Menus”

Delete/Inactivate Certain menus will be deleted or Menu inactivated for re-exposure “9.4 Deleting Exposure Menus/Canceling the Deletion” “9.5 Inactivating Exposed Images/Activating the Inactivated Images”

Exposure menus / images will be relocated “9.10 Changing the Order of Appearance of Menus in the Read Image Window”

Exchange Images

Menus will be associated correctly with images exposed “9.7 Exchanging Images”

Output/Storage

3. Click

(C) or the desired button (D).

Film printout or data storage settings will be changed “9.8 Setting for Printing, Saving, and Transmitting”

When editing patient information, display only one sub-window. If two sub-windows are arranged, the editing result of patient information cannot be verified correctly. Functions of Each Button Buttons

Functions Editing Patient Information “9.9 Editing Patient Information” Change menus “9.3 Changing Exposure Menus” Mis-exposure/Re-exposure

9

“9.4 Deleting Exposure Menus/ Canceling the Deletion”

Editing Study Information

“9.6 Registering Re-exposure Menus” Relocate Display Order

2. Select images from the sub-window.

Functions

“9.2 Adding Exposure Menus”

Edit Patient Information

● Starting from the “Image Viewer”

“9.5 Inactivating Exposed Images/Activating the Inactivated Images” “9.6 Registering Re-exposure Menus” Additional exposures “9.2 Adding Exposure Menus” Exchange images “9.7 Exchanging Images” Set the output destination “9.8 Setting for Printing, Saving, and Transmitting”

Patient data such as DOB and sex will be corrected “9.9 Editing Patient Information”

FCR PRIMA Console Reference Guide

897N101440B

9-3

9.2

Adding Exposure Menus

Registers additional exposure menus for exposure.

■ Operation Window

■ Adding an Exposure Menu

● “Adding Exposure Menu” Window

1. Click a menu group (A). (A)

2. Click a exposure menu (B). 3. Click

[Finish] (C).

→→The “Read Image” window or the “Image Viewer” (B)

(C)

window opens.

When a verified study is changed, a message displays to verify the change. If you click [OK], the study would return from the “Verified” to “Started” study status. After completing the study, re-verify it.

9 Editing Study Information

9-4

FCR PRIMA Console Reference Guide

897N101440B

9.3

Changing Exposure Menus

Changes menus on the current “Read Image” window on display. When the menu for a read image is changed, the image is subjected to image processing according to the new menu. If the image has been rotated or reversed, its status is updated according to the new menu.

■ Operation Window ● “Exposure Menu Change” Window

■ Changing an Exposure Menu (B)

1. Choose an exposure menu to be changed (A). 2. Select a new menu group (B).

(A)

3. Select a new exposure menu (C). 4. Click [Finish] (D).

(C)

→→The “Read Image” window or the “Image Viewer” window opens.

(D)

When a verified study is changed, a message displays to verify the change. If you click [OK], the study would return from the “Verified” to “Completed” study status. After completing the study, re-verify it.

9 Editing Study Information

The exposure menu is displayed as image information. If no image information is shown in the image pane, show it. “11.2.3 Setting the Attribute Information”

FCR PRIMA Console Reference Guide

897N101440B

9-5

9.4

Deleting Exposure Menus/Canceling the Deletion

Deletes unnecessary menus and schedules re-exposures.

Completed or verified images started up from the “Image Viewer” cannot be deleted.

■ Operation Window ●

■ Deleting an Exposure Menu/

“Deleting Menus/Inactivating Images” Window

Canceling the Deletion

1. Select an exposure menu prior to reading images (A).

(B) (A)

2. Click [Delete] (B). →→A deletion mark ( menu.

) is attached to the exposure

To cancel deleting the exposure menu, click [Undo (C)

9

Delete].

3. Click

mark to indicate the deletion hides.

[Finish] (C).

Editing Study Information

→→The “Read Image” window opens.

9-6

FCR PRIMA Console Reference Guide

897N101440B

9.5

Inactivating Exposed Images/ Activating the Inactivated Images

Inactivates a read image. Or, activates the inactivated image.

• Once Images are verified and saved to a disk, they cannot be inactivated/activated. • Inactivated images are not printed or saved even if “Film Print“(optional) or “Work Disk” is selected.

■ Operation Window

■ Inactivating/Activating

● “Deleting Menus/Inactivating Images” Window

Exposure Menus

1. Select the exposure menu for a read image (A).

(A)

(B)

2. Click [Inactivate] (B). →→An image inactivation mark ( the image.

) is attached to

To cancel inactivating the exposure menu, click (C)

[Activate].

9

[Finish] (C).

Editing Study Information

3. Click

mark to indicate the inactivation hides.

→→The “Read Image” window or the “Image Viewer” window opens.

FCR PRIMA Console Reference Guide

897N101440B

9-7

9.6

Registering Re-exposure Menus

Inactivates a read image and registers a re-exposure menu.

Once Images are verified and saved to a disk, they cannot be re-exposed.

■ Operation Window

■ Registering a Re-exposure

● “Deleting Menus/Inactivating Images” Window

Menu

1. Select the exposure menu for a read image (A).

(A)

2. Click [Re-exposure] (B). →→A re-exposure mark ( exposure menu.

(B) (C)

9

) is attached to the

This image is inactive and the same exposure menu as that for this image is added to the “Read Image” window.

3. Click

[Finish] (C).

Editing Study Information

→→The “Read Image” window or the “Image Viewer” window opens.

When a verified study is changed, a message displays to verify the change. If you click [OK], the study would return from the “Verified” to “Started” study status. After completing the study, re-verify it.

9-8

FCR PRIMA Console Reference Guide

897N101440B

9.7

Exchanging Images

Exchanges images to correct the order in which they are read. When images are exchanged, they are subjected to image processing according to the destination menu. If the images have been rotated or reversed, their status is updated according to the destination menu.

■ Operation Window ●

■ Exchange Incorrectly Read

Images to the Correct Exposure Menus

“Image Exchange” Window

(B) (A)

1. Select the two exposure menus (A) to be exchanged.

Selected exposure menus are displayed with checks. When a selected exposure menu is reselected, it is unchecked and thus deselected. (C)

2. Click [Exchange] (B). →→The images in the two selected exposure menus will be exchanged.

9

3. Click

Editing Study Information

Once images are exchanged, the operation cannot be undone. To reverse the image status, select the two exposure menus again and exchange the images.

[Finish] (C).

→→The “Read Image” window or the “Image Viewer” window opens.

When a verified study is changed, a message displays to verify the change. If you click [OK], the study would return from the “Verified” to “Completed” study status. After completing the study, re-verify it.

FCR PRIMA Console Reference Guide

897N101440B

9-9

9.8

Setting for Printing, Saving, and Transmitting

Verifies or modifies the following settings: • Saving images to a storage disk • Printing images on a film (optional) • Transmitting images to DICOM server (optional)

■ Operation Window

■ Setting up the Destination and

● “Setting Image Print/Storage” Window

Timing for Outputting and Saving

1. Perform various settings (A). Selection Items Items

(A)

(B)

9 Editing Study Information

The window comes up with the settings configured in the “User Utilities”. The “User Utilities” can be configured for users to apply the same set of settings for all the studies they conduct using FCR PRIMA Console. “14.5 Setting of Printing, Saving and Transmitting Images”

Contents

N/A*1

Output or not

Status*2

Status of the output

Destination*3

Current destination

Timing*3

Study status at which the image is to be output

Upon Completion

When the study status becomes “Completed”

Upon Verification

When the study status becomes “Verified”

*1 Put a check mark in order to output the image. *3 To change these settings, click [Details]. Depending on the functions, some settings are already specified and cannot be changed. “■Changing the Detailed Settings”

*2 If images have been output, the “Unoutputted” check box under Status is selected. This check box is not accessible for handling.

2. Click [Finish] (B). • To print or save images for this study on a priority basis, select the “Execute prioritized printout” check box. →→The “Read Image” window or the “Image Viewer” window opens.

9-10

FCR PRIMA Console Reference Guide

897N101440B

■ Changing the Detailed Settings ● Editing the Settings for Saving Images to Disks

Changes the pixel density with which images are saved to a storage disk.

● Editing Film Print (Optional) Settings Modifies the following settings: • Type of film printer used • Film print timing • Print count

“Print Destination Detail” Window

“Save Destination Detail” Window (A) (B) (C) (A) (D) (B)

1. Select an imager to be used in the “Destination” field (A).

The timing at which images are saved to a storage disk (output timing) is when their study status changes to “Verified”. The timing at which images are saved to a storage disk cannot be changed.

1. Click “Output Density” (A).

[Save by Max Pixel Count] Save images with the maximum pixel density that allows them to be saved or printed on the system.

Once images have their pixel density set to “Save by Default Pixel Count” on FCR PRIMA Console, the pixel density is kept as is even if “Save by Max Pixel Count” is specified after reloading the images into FCR PRIMA Console and saving them to the disk.

[Upon Completion]: Print images on a film when their study status changes to “Completed”.

9

[Upon Verification]: Print images on a film when their study status changes to “Verified”.

Editing Study Information

[Save by Default Pixel Count] Save images with a reduced pixel density and hence in a smaller file size.

2. Select “Timing” (B).

3. Select a film output count (C). 1 to 9 copies can be printed.

4. Click [Setup](D). →→The system returns to the ““Setting Image Print/ Storage” window. A check is put into the check box for outputting.

2. Click [Setup] (B). →→The system returns to the “Setting Image Print/ Storage” window. A check is put into the check box for outputting.

FCR PRIMA Console Reference Guide

897N101440B

9-11

● Editing Settings for Transmitting Images to DICOM Server (Optional)

Modifies the transmission destination (output destination) in DICOM server to which images are transmitted to DICOM server and the pixel density (output density) with which the images are transmitted. “Print Destination Detail” Window

(A)

(B)

(C)

The timing at which images are transmitted to DICOM server (output timing) is when their study status changes to “Verified”. The timing at which images are transmitted to DICOM server cannot be changed.

9 Editing Study Information

1. Select a destination DICOM server in the “Destination” field (A).

2. Select an “Output Density” (B).

[Save by Default Pixel Count]: Transmit images with a reduced pixel density and hence in a smaller file size. [Save by Max Pixel Count]: Transmit images with the maximum pixel density that allows them to be saved or printed on the system.

Once images have their pixel density set to “Save by Default Pixel Count” on FCR PRIMA Console, the pixel density is kept as is even if “Save by Max Pixel Count” is specified after reloading the images into FCR PRIMA Console and saving them in the disk.

3. Click [Setup] (C). →→The system returns to the “Setting Image Print/ Storage” window.

9-12

FCR PRIMA Console Reference Guide

897N101440B

9.9

Editing Patient Information

Edits made to the patient information can be reflected in the corresponding information stored in the FCR PRIMA Console database.

■ Operation Windows

● “Search for ID Number” Window

● “Modifying Patient Information” Window (G)

(A) (H) (B)

● “Edit Searched Patient Information” Window

● “Modify ID Number” Window (I)

9

(C) Editing Study Information

(J)

(D)

● “Input Patient Information” Window (E)

(D)

FCR PRIMA Console Reference Guide

897N101440B

9-13

■ Editing Patient Information ● Editing Patient ID or Detailed Patient Information

1. Click [Edit] (A) next to the ID field or the detailed patient information fields.

→→“Modify ID Number” window or the “Input Patient Information” window opens.

2. Make edits to the patient ID (C) or the detailed patient information (E).

• Entering patient ID Enter the patient ID with single-byte alphanumeric characters. The allowable character length is determined by a system setting. The following characters cannot be used. “ # % & `+ ? / \ * @ < However, the space can be used but the patient ID which consists of solely space is not allowed. “3.2 Registering the Patient Information to be Exposed”

9 Editing Study Information

A message displays whether to apply the change also to the database. If you click [OK], the change is also applied to the database. →→The “Read Image” window or the “Image Viewer” window opens.

4. Check that the patient ID or the detailed patient information has been modified.

When a verified study is changed, a message displays to verify the change.

If you click [OK], the study would return from the “Verified” to “Completed” study status. After completing the study, re-verify it.

FCR PRIMA Console Reference Guide

Study as a New Study for Another Set of Patient Information

1. Click [Another] (B). →→The “Search for ID Number” window opens.

2. Enter the patient ID (G) of the target patient information and click [Search] (H).

→→The “Edit Patient Information” window opens.

3. Verify or edit the patient information ( I ). If the patient information matching the patient ID entered in the “Search for ID Number” window is not registered in the FCR PRIMA Console database, no detailed patient information is displayed in the “Search for and Edit Patient Information” window. To register new patient information in the FCR PRIMA Console database, proceed to enter detailed patient information. “3.2 Registering the Patient Information to be Exposed

4. Click [Finish] (J).

3. Click [Finish] (D).

9-14

● Reregistering the Displayed Current

897N101440B

A message displays whether to apply the change also to the database. If you click [OK], the change is also applied to the database. →→The system returns to the “Read Image” window or the “Image Viewer” window.

5. Check that the patient information has been modified.

When a verified study is changed, a message displays to verify the change.

If you click [OK], the study would return from the “Verified” to “Completed” study status. After completing the study, re-verify it.

9.10 Changing the Order of Menus in the “Read Image” Window

■ Operation Window

1. Select the exposure menu (A) to be relocated.

● “Exposure Menu Relocate” Window

2. Click the direction buttons (B) to change the (A)

order of the appearance. Functions of Each Button Buttons

(B)

(C)

Functions

To the TOP

Moves the selected exposure menu to the top

UP

Moves up the selected exposure menu

DOWN

Moves down the selected exposure menu

To the END

Moves down the selected exposure menu

3. Click

[Finish] (C).

→→The “Read Image” window opens.

9 Editing Study Information

FCR PRIMA Console Reference Guide

897N101440B

9-15

9 Editing Study Information

9-16

FCR PRIMA Console Reference Guide

897N101440B

Chapter

10

Verifying with the Indicators 10.1 Viewing Indicators and Activating Functions.................. 10-2 10.1.1

Viewing Indicators........................................................ 10-2

10.1.2

Activating Functions from Indicators......................... 10-3

10.2 Verifying the Status of Image Reader and Cleaning Procedures......................................................... 10-4 10.3 Verifying Disk Status and Maintenance........................... 10-5 10.4 Verifying NAS Status (Optional).......................................10-7

10 Verifying with the Indicators

10.5 Verifying Film Printer Status (Optional)........................... 10-8 10.6 Verifying Image Output Status . .......................................10-9 10.7 Verifying Events (Errors, Warnings or Information)......... 10-10

FCR PRIMA Console Reference Guide

897N101440B

10-1

10.1 Viewing Indicators and Activating Functions Indicators provide visual indications of the status of the Image Reader, disk, NAS, printer (optional), data transfer and events (errors, warnings and information). By installing conditions for options or by the system settings at the time of installing, some icons are not displayed.

10.1.1 Viewing Indicators

■ Position of “Indicator” Window Indicators are separated from the “Main” window or “Image Viewer” window, so it can be placed in any spots. (1) (2) (3) (4) (5) (6)

Icons on Indicators: (1) Image Reader (2) Disk (3) NAS (optional) (4) Film printer (optional) (5) Output Status (6) Events (errors, warnings, information)

10 Verifying with the Indicators

■ Indicator State and Corresponding Notice Indicators turn on, turn off or blink, to notify necessary information listed in the following tables. State

(1)

(2)

(3) (4) (5) (6)

Contents Off (Unavailable) No Image Reader is connected to the FCR PRIMA Console. Images cannot be read. Connect an Image Reader. Either a storage disk is not inserted in the PC’s control drive or a read-only disk is inserted in the PC’s control drive. Images cannot be saved to the disk. NAS is not available. Images cannot be saved to NAS. The default film printer is not available. Images cannot be printed on a film. There is no image being output. FCR PRIMA Console is running normally.

On (Available)

Blinking (Error occurs)

An Image Reader is connected to the FCR PRIMA Console. Images can be read.

An error has occurred with the Image Reader connected to the FCR PRIMA Console. *1

A write-enabled storage disk is inserted in the PC’s control drive. Images can be saved to the disk.

-

NAS is available. No free space in NAS. Images can be saved to NAS. The default film printer is available. An error has occurred with the Images can be printed on a film. default film printer. *1 An image is being output.

-

-

An event (error, warning or information) has occurred. *1

*1 Click the indicator to view detailed information about the event. Verify the event and take appropriate action.

10-2

FCR PRIMA Console Reference Guide

897N101440B

10.1.2 Activating Functions from Indicators Activates desired functions by clicking indicators.

■ Operation Window

■ Opening “Indicator” Windows

● ”Indicators”

1. Click an desired “Indicator” button to verify

(1) (2) (3) (4) (5) (6)

(From (1) to (6)).

Functions of Each Button Button

Functions Status of “Read Image” part “10.2 Verifying the Status of Image Reader and Cleaning Procedures” Disk status “10.3 Verifying Disk Status and Maintenance” NAS status “10.4 Verifying NAS Status (Optional)” Film printer status “10.5 Verifying Film Printer Status (Optional)” Output status “10.6 Verifying Image Output Status” Verifying Events (Errors, Warnings or Information)

10 Verifying with the Indicators

“10.7 Verifying Events (Errors, Warnings or Information)”

• You can do the same by clicking tabs in the top of each “Indicator” window. • Events (Errors, Warnings or Information) can be also performed by right-clicking the server application icon docked in the task tray and clicking [Event indicator].

FCR PRIMA Console Reference Guide

897N101440B

10-3

10.2 Verifying the Status of Image Reader and Cleaning Procedures Verifies the status of the image reader connected to the FCR PRIMA Console and performs scanner cleaning.

■ Operation Windows ●

■ Verifying the Status of the Image Reader

The “Reader” Window - “Status”

1. Click [Status] (A). (A)

(B)

(C)

3. Verify the status and the detailed information (B).

■ Cleaning the Scanner of the

(D)

Image Reader

● The “Reader” Window - “Operation” 10

2. Click [Update] (C).

Perform scanner cleaning to remove dust from the scanner head which you are reminded to carry out periodically.

(E)

1. Click [Operation] (E)

(F)

2. Click [Operation] (F).

Verifying with the Indicators

Before performing scanner cleaning, check that the Image Reader is available. (D) Scanner cleaning takes about 15 seconds to complete. Click [Status] to return to the status view of the “Reader” window.

■ Verifying the status of the

Image Reader and Exiting Scanner Cleaning.

1. Click [OK] (D). →→The “Reader” window closes.

10-4

FCR PRIMA Console Reference Guide

897N101440B

10.3 Verifying Disk Status and Maintenance Checks the status of the disk set in the PC’s control drive and performs maintenance.

■ Operation Windows

■ Verifying Disk Status

● “Disk” Window - “Status/Operation”

Click [Update] (F) to update information.

(A) (B) (C) (G) (F)

● Status of Each Drive 1. Click [Status/Operation] (A). 2. Select a drive in “Device Name” (B).

(H)

3. Verify the detailed information.

(I)

Display Items

● “Disk” Window - “Store-Disk List”

Items

Contents

Status

Disk name and the amount of free space (%)

(D)

Usage*1

Usage of disks • storage disk • work disk

(E)

Disk Write

Disk write settings • write-enabled • read-only

*1 Nothing appears for any other type of disk. ( I’ )

“2.7 Before Using Disks”

FCR PRIMA Console Reference Guide

Verifying with the Indicators

(F’)

10

897N101440B

10-5

● Status of the Registered Storage Disk 1. Click [Store-Disk List] (D). 2. Click [Update] (F’). 3. Verifying the storage disk list (E). Display Items Items

Contents

Disk Name

Storage Disk name

Write Start

The date and time at which writing to a storage disk started

Write End

The date and time at which writing to a storage disk ended

Status

Disk write display • write-enabled • read-only

■ Disk Maintenance ● Replacing Disks 1. Click [Replace] (G).

10

→→The “Replacing Disks” window opens, and then the disk drive tray opens “13.8 Disk Maintenance”

Verifying with the Indicators

● Perform from “Client Utilities” 1. Click [Utility] (H). →→The “Disk Operation” window opens, and then the disk drive tray opens. “13.8 Disk Maintenance”

■ Verifying Disk Status of the Image Reader and Exiting Scanner Cleaning

1. Click [OK] (( I ), ( I’ )). →→The “Disk” window closes.

10-6

FCR PRIMA Console Reference Guide

897N101440B

10.4 Verifying NAS Status (Optional) Verifies the status of the NAS connected to the FCR PRIMA Console.

■ Operation Window

■ Verifying NAS Status

● “Disk” Window - “Status/Operation”

1. Click [Status/Operation] (A).

(A) (B) (C)

2. Select a NAS in “Device Name” (B). 3. Verify the detailed information (C). Display Items Items

(E)

(D)

• When the HDD does not have enough space, install another NAS. In such cases, contact our official dealer. • When an error of NAS is displayed in the indicator for Event, contact our official dealer to fix it.

Contents

Status

Disk name and the amount of free space (%)

Disk Write

Display Disk Write

Click [Update] (E) to update information.

■ Exiting the Verification of NAS Status

10

1. Click [OK] (D).

FCR PRIMA Console Reference Guide

Verifying with the Indicators

→→The “Disk” window closes.

897N101440B

10-7

10.5 Verifying Film Printer Status (Optional) Verifies the status of the film printer connected to the FCR PRIMA Console.

■ Operation Window

■ Verifying Film Printer Status

● “Printer” Window

1. Select a printer in “Device Name” (A). 2. Verify the status and the detailed information

(C)

(D)

(A)

(B).

(B)

Click [Update] (C) to update information.

■ Verifying the Status of the Film Printer

1. Click [OK] (D). →→The “Printer” window closes.

10 Verifying with the Indicators

10-8

FCR PRIMA Console Reference Guide

897N101440B

10.6 Verifying Image Output Status Verifies the output status of images as they are printed on a film (optional), saved to a disk or transmitted to DICOM server (optional). Canceled or failed output of images can be retried.

■ Operation Window

■ Verifying Image Output Status

● “Output” Window

1. Select an output destination of interest in Filter Conditions (A).

(A)

(E)

(B)

(C)

→→The status of the selected output destination is displayed (E).

■ Canceling Image Output 1. Select an image to cancel output (E).

(D)

2. Click [Delete] (B).

■ Outputting an Image Again that Has Once Failed to be Output

1. Select an image to be output again (E). 2. Click [Force Output] (C).

10 Verifying with the Indicators

■ Exiting the Verification of Image Output Status

1. Click [OK] (D). →→The “Output” window closes.

FCR PRIMA Console Reference Guide

897N101440B

10-9

10.7 Verifying Events (Errors, Warnings or Information) Verifies events (errors, warnings or information) occurring on the FCR PRIMA Console system.

■ Operation Windows

■ Verifying Events

● “Events” Window

1. Select [Source] (A) and [Status] (B).

(A)

Selection Items

(B)

Items

Contents

Source

• • • • •

Status

All Unconfirmed Only Confirmed Only

(F)

(C)

All Printer Disk DICOM Storage Image Reader

• Study Registration • Image Browse • Common • Other

→→The events of the selected Source and Status is displayed.

(D)

2. Select an event of interest (F). The events are shown in the descending order of Date/Time.

10

● “Error Details” Window

3. Click [Details] (C). →→The “Error Details “ window opens. The following functions are also available. Functions of Each Button

Verifying with the Indicators

Buttons

Functions Deletes an event Delete all the events Display the previous page of the event list

(E)

Display the next page of the event list *1

Mark the selected events as being verified

*1

Check all the events in the current page as being verified

*1

Check all the events as being verified

*1 Verified events are hidden when “Unconfirmed Only” is selected in Status.

■ Exiting Events Verification 1. Click [Close] (E) →→The “Error Details” window closes.

2. Click [OK] (D). →→The “Event” window closes.

10-10

FCR PRIMA Console Reference Guide

897N101440B

Chapter

11

Changing Display Settings of the “Main” Window and “Image Viewer” 11.1 Display Settings of the “Main” Window........................... 11-2 11.1.1

Starting up and Shutting Down the Display Setting Function......................................... 11-2

11.1.2

Initial Display Settings of the “Main” Window........... 11-3

11.1.3 Display Settings of “ID input” Window....................... 11-4 11.1.4 Display Settings of “Patient List” Window................. 11-5 11.1.5 Display Settings of “FCR-Menus” Window................ 11-6 11.1.6 Display Settings of Exposing and Reading................ 11-7

11 Changing Display Settings of the "Main" Window and "Image Viewer"

11.1.7 Display Settings of “Refer” window........................... 11-8 11.1.8 Setting the Font Size in the “Main” Window.............. 11-9

11.2 Changing Settings of “Image Viewer” Window............. 11-10 11.2.1

Starting the Customize Function.............................. 11-10

11.2.2

Initial Settings of Annotation..................................... 11-11

11.2.3

Setting the Attribute Information.............................. 11-12

11.2.4

Setting Reading Protocol........................................... 11-13

11.2.5

Display Settings of Toolbar....................................... 11-14

11.2.6

Display Settings of Report......................................... 11-15

11.2.7

Display Setting of “Image Viewer” Window............. 11-18

11.2.8

Returning to Standard Settings................................. 11-20

11.2.9

Importing/Exporting Settings.................................... 11-21

FCR PRIMA Console Reference Guide

897N101440B

11-1

11.1 Display Settings of the “Main” Window Configures settings of the “Main” window, such as which items are to be displayed in each window in the “Main” window, or which window is to be displayed when started up.

11.1.1 Starting up and Shutting Down the Display Setting Function

■ Operation Windows

Items

● “Main” Window

ID input

Contents Set the items to display in the “Input ID” window and its order. Also name of items can be set to display/hide “11.1.3 Display Settings of “ID input” Window”

(A)

● “Display setting” Window

Patient List The items to be displayed in the patient list and their order can be set. Auto-update of the patient list can also be set “11.1.4 Display Settings of “Patient List” Window” FCRMenus

The shortcuts for the exposure menu can be set “11.1.5 Display Settings of “FCRMenus” Window”

(B)

FCRBehavior on registering a study and Operations behavior after reading images with the “Image Viewer” can be set “11.1.6 Display Settings of Exposing and Reading” Modalities

11

(C)

Changing Display Settings of the "Main" Window and "Image Viewer"

■ Starting up 1. Select

“11.1.7 Display Settings of “Refer” window” Toolbar

2. Click a desired tab (B) and change display settings.

Operations which can be performed for studies in the study list can be added or removed “11.1.7 Display Settings of “Refer” window”

- [Customize] (A).

→→The “Display setting” window opens.

The modalities to be displayed in the study list and their order can be set

Font

The font size can be set for the each screen in the “Main” window “11.1.8 Setting the Font Size in the “Main” Window”

Selection Items Items General

Contents The initial display for the “Main” window can be set “11.1.2 Initial Display Settings of the “Main” Window”

■ Shutting Down 1. Click [OK] (C). →→Ths system returns to the “Main” window.

11-2

FCR PRIMA Console Reference Guide

897N101440B

11.1.2 Initial Display Settings of the “Main” Window Sets the window to be displayed first when the “Main” window is started.

■ Operation Window ●

■ Setting the Initial Display when

“Display setting” Window - “General” Window (A) (B)

Started up

1. From the pull-down menu (A), select the

window to be displayed first in the upper side of the “Main” window when the FCR PRIMA Console is started.

2. From the pull-down menu (B), select the

window to be displayed in the lower side of the window when a patient is selected in the “Input ID” window or in the “Patient List” window,

(C)

3. Click [OK] (C). Select another tab to set another display settings.

11 Changing Display Settings of the "Main" Window and "Image Viewer"

FCR PRIMA Console Reference Guide

897N101440B

11-3

11.1.3 Display Settings of “ID input” Window Sets the items and its order to be displayed in the “Input ID” window. Also, the names of items can be set to display/hide.

■ Operation Window

■ Setting Items and Their Order to

● “Display setting” Window - “ID input” Window

be Displayed in the “ID input” Window

1. Check the check box of items to be displayed

(A)

as input items (A).

(B)

(C)

Uncheck the check box of items not to be displayed as input items.

2. Select an item to change the display order in the list, and click [Up] or [Down] (B).

(D)

By clicking [Up] or [Down], the item is moved up or down respectively.

3. Repeat steps 1. and 2. to set the display items.

4. Select display/do not display the item name (C). 5. Click [OK] (D).

11 Changing Display Settings of the "Main" Window and "Image Viewer"

Select another tab to set another display settings.

11-4

FCR PRIMA Console Reference Guide

897N101440B

11.1.4 Display Settings of “Patient List” Window Sets the items to be displayed in the patient list and the maximum number of the list to be displayed. Also sets whether the patient list is to be updated automatically or not and the update interval when autoupdate is enabled.

■ Operation Window

■ Setting the Display Contents in the “Patient List” Window

● “Display setting” Window - “Patient List” Window

1. Check the check box of items to be displayed

(A)

(A) in the patient list.

(B)

(C) (D) (E) (F)

• Uncheck the check box of items not to be displayed in the patient list. • The items are displayed left aligned in the patient list according to the order of items on the customize window.

2. Select an item to change the display order in the patient list, and click [Up] or [Down] (B).

By clicking [Up] or [Down], the item is moved up or down respectively.

3. Repeat steps 1. and 2. to set the display items.

4. Set the maximum number of the list to be displayed (C) in the patient list.

11 Changing Display Settings of the "Main" Window and "Image Viewer"

5. Put a check mark to “Enable automatic updating” (D).

To disable the autoupdate, uncheck the check box of “Enable automatic updating”.

6. Set the update interval (E). Change the value by clicking the arrow buttons.

7. Click [OK] (F). Select another tab to set another display settings.

FCR PRIMA Console Reference Guide

897N101440B

11-5

11.1.5 Display Settings of “FCR-Menus” Window Sets frequently used exposure menus as shortcuts in the “FCR” window of the“ Main” window. Exposure menus which is not set as shortcuts can be selected from [Others].

This item will be applied to the entire system.

■ Operation Windows ●

■ Setting the Display Contents in

“Display setting” Window - “FCR-Menus” Window

the “FCR-Menus” Window

1. Click [Select] (A). →→The “Exposure Menu (Detail)” window opens.

2. Select a menu group (C).

(A)

(B)

→→The exposure menus and/or the study menus in the selected menu group are displayed in the lower side of the window.

3. Select a study menu or an exposure menu (D). →→The menu is displayed in the list of selected menus on the right side of the window.

4. Click [OK] (E) at the bottom right of the

● “Exposure Menu (Detail)” Window

window.

(C)

→→The selected menu is added.

5. Click [OK] (B).

11

Select another tab to set another display settings.

Changing Display Settings of the "Main" Window and "Image Viewer"

(E)

(D)

11-6

FCR PRIMA Console Reference Guide

897N101440B

11.1.6 Display Settings of Exposing and Reading Sets whether to start an exposure immediately or to register only the study after selecting the exposure menu. Also sets whether or not the window is to be closed after reading an image to the simple read image window.

■ Operation Window

■ Setting Behavior on Selecting an Exposure Menu and on Reading Images with the “Image Viewer”

● “Display setting” Window - “FCROperations” Window

1. Select the “Behavior on registering a study” (A). (A)

(B)

2. Select the “Behavior after reading images” (B). 3. Click [OK] (C). Select another tab to set another display settings.

(C)

11 Changing Display Settings of the "Main" Window and "Image Viewer"

FCR PRIMA Console Reference Guide

897N101440B

11-7

11.1.7 Display Settings of “Refer” window The following settings can be performed. • Setting the modalities to be displayed in the study list of the “Refer” window. • Adding and deleting operation items to be performed for studies in the study list. Each operation item is displayed in an icon in the lower side of the “Refer” window.

■ Operation Windows

2. Click [Up] or [Down] (B) while selecting an

● “Display setting” Window - “Modalities”

item for which the display order is to be changed.

Window

(A)

By clicking [Up] or [Down], the item is moved up or down respectively. (B)

3. Repeat steps 1. and 2. to set the display items.

4. Click [OK] (C). Select another tab to set another display settings. (C)

● “Display setting” Window - “Toolbar” Window

the Study List

(D) (E)

11

(F) (G)

Changing Display Settings of the "Main" Window and "Image Viewer"

(H)

■ Setting Modalities to be

Displayed in the Study List

1. Put the check box of items to be displayed

11-8

■ Setting the Operation Icons in 1. Select an operation item from the operation item list on the left side of the window (D).

2. Click [>>>] (E) on the left side of items to be added.

• Deleting an operation item from a block. To be removed operation items, click [Remove] (G) • To change the display order, click [Up] or [Down] while selecting an item for which the display order is to be changed.

3. Click [OK] (H).

(A) in the study list.

• When operation items have been added or their order has been changed, these settings are also applied to the “Main” window.

Remove the check box of items not to be displayed in the study list.

• Select another tab to set another display settings.

FCR PRIMA Console Reference Guide

897N101440B

11.1.8 Setting the Font Size in the “Main” Window Font size can be set for each window in the “Main” window.

■ Operation Window ●

■ Setting the Font Size for the “Main” Window

“Display setting” Window - “Font” Window

1. Select the font size for the “Main” window from the pull-down menu (A).

2. Click [OK] (B). (A) Select another tab to set another display settings.

(B)

11 Changing Display Settings of the "Main" Window and "Image Viewer"

FCR PRIMA Console Reference Guide

897N101440B

11-9

11.2 Changing Settings of “Image Viewer” Window The following settings can be performed in the “Customize” menu of “Image Viewer” window. • Annotation initial settings • Attribute info • Reading protocol • Tool bar • Report • Display Settings These settings can be returned to the initial settings all together or can be imported/exported.

11.2.1 Starting the Customize Function

■ Operation Window

Items

● “Image Viewer” Window

Contents

Reading protocol The “Display setting for series view” window opens. Change the display settings when referring series images. The display settings can be set with each modality. “11.2.4 ■Setting the Display of each Modality” (A)

Toolbar

The “Set viewer tool bar” window opens. Set the tool bar to be displayed in the “Image Viewer” window. “11.2.5 ■Setting the Display of Toolbar”

11 Changing Display Settings of the "Main" Window and "Image Viewer"

11-10

Report

■ Starting Up 1. Click [Customize] and select the setting menu

The “Report tag settings” window opens. Set the format for the comment column, display of schema, and during printing. “11.2.6 ■Setting the Display of Report”

to be changed (A).

Items

Contents

Annotation initial settings

The “Annotation setting” window opens. Set to display the annotation.

The “Style setting” window opens. Set the image display in subwindow or the initial display of study summary/image cart. Also, set the actions during storage of FCR image processing parameters.

“11.2.2 ■Setting the Display/ Tag Display of Annotations”

“11.2.7 ■Setting Actions during Image Display or Storage”

→→The selected display setting window opens. Selection Items

Attribute info

The “Settings for image incidental information” window opens. Set the image information. “11.2.3 ■Setting Image Attribute Information

FCR PRIMA Console Reference Guide

897N101440B

Display Setting

You can do the same by right-clicking the images displayed in the “Image Viewer” and select the settings from the [Customize] menu.

11.2.2 Initial Settings of Annotation Sets the initial display of basic graphic annotation and measurement annotation.

■ Operation Window ● “Annotation setting” Window

The items to be set are different from the selected annotation. (A) (B)

(C)

Annotation Magnifying glass

Settable contents • Width of Magnifying glass (pixel) • Height of Magnifying glass (pixel)

Arrow, Ellipse, • Line thickness Rectangle, Polygon • Line color Text

• • • • •

(D)

■ Setting the Display/Tag Display of Annotations

• Line thickness Measure line • Line color segment, Simplified cardiothoracic ratio, Cardiothoracic ratio, Interline angle, Dual angle, Cobb angle

1. Select annotation (A). 2. Set the initial value (B). Input/Selection Items Items

Font Style Size Text color Background color (By putting a check mark on the “Transparent color” check box, the background hides.)

Contents

Width of Input the width of magnifying Magnifying glass glass in pixel (pixel)

annotation (C).

Input/Selection Items Items

Contents

Font

Select font type

Style

Select the style of measurement label (standard, bold, italic, bold italic)

Select the font

Size

Input text size

Style

Select the text style (standard, bold, italic, bold italic)

Text color

Select text color

Input text size

Background color

Select text background color

Size

Line thickness

Select the line thickness of annotation

Line color

Select the line color of annotation

Font

Text color Select text color Background color Background color Select text background color Transparent color:

By putting a check mark on the “Transparent color” check box, the background hides

Transparent color:

Put a check mark to make the background color transparent

4. Click [OK] (D). →→The system returns to the “Image Viewer” window.

FCR PRIMA Console Reference Guide

897N101440B

11 Changing Display Settings of the "Main" Window and "Image Viewer"

Height of Input the height of magnifying Magnifying glass glass in pixel (pixel)

3. Set the text style for the measurement

11-11

11.2.3 Setting the Attribute Information Sets which attribute information on the image is to be displayed with each modality.

■ Operation Windows

2. Click the display location of the attribute information (B).

● “Settings for image incidental

→→The “Set display item” window opens.

information” Window

3. Select the attribute information from the list (E). (A) (B)

Select the [Category], [Group], [Element], [Tag Name] and click attribute information.

to narrow down the

4. Choose from “Image incidental information (C)

● “Set display item” Window

will be displayed” or “Image incidental information will not be displayed” in (D).

5. Set the display method (F) (G). (D)

Input/Selection Items Items

(E)

(H)

11 Changing Display Settings of the "Main" Window and "Image Viewer"

11-12

(F)

(G)

■ Setting Image Attribute Information

1. Select the modality (A). Click the [Add/Delete] to add/delete the modality.

FCR PRIMA Console Reference Guide

897N101440B

Contents

Prefix

Input the prefix to be displayed

Suffix

Input the suffix to be displayed

Display elements

Input the display elements of attribute information

Display magnification

Input the display magnification of attribute information

Valid number of digits

Input the valid number of digits of attribute value

Comment

Input the comment

6. Click [OK] (H). →→The system returns to the “Settings for image incidental information” window.

7. Click [OK] (C). →→The system returns to the “Image Viewer” window.

11.2.4 Setting Reading Protocol Sets the initial arrangement of sub-window, the image arrangement in sub-window with each modality.

■ Operation Window

Items

● “Display setting for series view” Window

Contents

Tile division

Set number of images and layout to display in one sub-window • Maximum number of lines (1 to 6 rows) • Maximum number of columns (1 to 6 columns) • Sort sequence (Horizontal direction, Vertical direction)

Comparative Split

Set whether to add how many subwindows when clicking [Compare] in the study summary • Maximum number of lines (1 to 6 rows) • Maximum number of columns (1 to 6 columns) • Sort sequence (Horizontal direction, Vertical direction)

Use series integration

Put a check mark to use series integration

Use external link

Put a check mark to use external link

Report will be automatically displayed

Put a check mark to display automatically

Report will be automatically created

Put a check mark to display automatically. Specify whether to automatically create the report from - days ago

(A)

(B)

(C)

■ Setting the Default Display Layout of each Modality

1. Select the modality (A). Click [Add / Delete] to add or delete the modality.

11

3. Click [OK] (C). →→The system returns to the “Image Viewer” window.

Changing Display Settings of the "Main" Window and "Image Viewer"

2. Set the sub-window and image display (B) of each modality. Selection Items Items Series division

Contents Set the layout to display by dividing a sub-window in each series • Maximum number of lines (1 to 4 rows) • Maximum number of columns (1 to 4 columns) • Sort sequence (Horizontal direction, Vertical direction)

FCR PRIMA Console Reference Guide

897N101440B

11-13

11.2.5 Display Settings of Toolbar Customizes the buttons to be assigned to the toolbar in “Image Viewer” window.

■ Operation Window

■ Setting the Display of Toolbar

● “Set viewer tool bar” Window

1. Select the tab name from “Tab” (A).

(A) (B)

Click [Edit] to add/delete tabs, or change tab name. (C)

(D) (E)

2. Select the group name from “Group” (B). Click [Edit] to add/delete groups, or change group name.

3. Drag and drop the tool buttons to be used to the “Tool button” column to add it.

Select the tool button from the “Tool button” column and click [Delete] to delete it.

4. Select the display style of tool button (D). Selection Items Items

Group name will Put a check mark to display the be displayed. group name in toolbar.

11

Display style of tool button

Changing Display Settings of the "Main" Window and "Image Viewer"

11-14

Contents

: Icon + Text (Vertical) : Only icon : Icon + Text (Horizontal) : Only text

5. Click [OK] (E). →→The system returns to the “Image Viewer” window.

FCR PRIMA Console Reference Guide

897N101440B

11.2.6 Display Settings of Report Sets the initial display of schema, title and text size of the comment column in “Report:” window.

■ Operation Window

■ Setting the Display of Report

● “Report tag settings” Window

1. Set comment column (A). (A)

Input/Selection Items Items

(B) (C) (D) (E) (F) (G) (H) (I)

● “Tag setting for Report” Window

Font Size

Select from “Large”, “Medium”, “Small”

The number of comment lines

Specify the numbers of rows of comment

Title of comment Input the title of Comment 1 to 3 1 to 3 Nothing abnormal Input the text to be displayed in message in Comment 1 to 3 Comment box 1 to 3

2. Set the initial display of schema (B). (J)

(M) (K)

Selection Items Items

Contents

Display default schema

Display the default schema when creating new reports

Display Favorites

Display the schema registered as favorite schema when creating new reports

(L)

● “Format list” Window

Contents

No initial display Not display the schema when creating new reports

(O)

(P) (Q) (R)

3. Click [Edit Template] (C) to add or delete the format file list.

“■ Editing Formats”

4. To reverify the study at the time of saving a report, put a check mark to “It will be automatically re-verified when report is saved.” check box (D).

5. To add or edit the template tag (items to be displayed in the format), click [Add] (E) or [Edit] (F). “■ Adding/Editing Template Tag”

To delete the template tag, select the template from the list (H) and click [Delete] (G).

6. Click [OK] ( I ). →→The system returns to the “Image Viewer” window.

FCR PRIMA Console Reference Guide

897N101440B

Changing Display Settings of the "Main" Window and "Image Viewer"

(N)

11

11-15

■ Creating a Format

3. Input the information which is inserted into the report to the cell of the created layout.*1

To create a report template, the Excel2007 (the environment which can create .xlsx file) is necessary.

1. Open the “Template.xlsx” file in “\FujiFilm\ Template” of My Documents.

Input Items Items !Schema,0,a,b

Schema image a: Length of the cell (Numbers of cells to be combined) b: Width of the cell (Numbers of cells to be combined)

!Image,c,d,e

Thumbnail image c: Index number of thumbnail from 0*2 d: Length of the cell (Numbers of cells to be combined) e: Width of the cell (Numbers of cells to be combined)

!CommentTitle,f

Title of the “Comment” f: Index number of “Comment” 0 to 2*3

!Comment,g

Contents of the “Comment” g: Index number of “Comment” 0 to 2*3

!SeriesCount,h

Image No. of the thumbnail image (Image order in the series) h: Index number of thumbnail from 0*2

!PatientID

Patient ID

!PatientNameSBCS

Patient name (singlebyte)

!PatientSex

Patient sex

!StudyDate

Study date

!BirthDate

Birth date of the patient

!Age

Patient age

!StudyTime

Study time

!Modality

Modality

2. Perform the following settings to create a report layout.

11 Changing Display Settings of the "Main" Window and "Image Viewer"

11-16

• Frame line of the report • Header • Footer • Title • Image pasting area • Comment area

To paste/input information to multiple lines of the image pasting area or the comment area, it is necessary to combine the cells of the concerning area.

FCR PRIMA Console Reference Guide

897N101440B

Contents

Items

Contents

!PatientSize

Height

!PatientWeight

Weight

!PatientAddress

Patient address

!PatientMotherBirthName Mother’s name of the SBCS patient (single-byte) !CountryOfResidence

Country of Residence

!PatientTelephoneNumber Telephone numbers of the patient !EthnicGroup

Ethnic group

!SmokingStatus

Smoking status

!LastMenstrualDate

Late menstrual date

!PatientComments

Patient comments

!PregnancyStatus

Pregnancy status

!FacilityNameSBCS

Facility name (single-byte)

!ReportAuthorSBCS

Author of the report (single-byte)

If the specified index number does not exist, the thumbnail image cannot be inserted into the report. (When there are only 2 thumbnail images registered in “Thumbnail”, the images cannot be inserted into the Report even specifying the index number as “2”.) *1 The input information is displayed on the report according to the settings of each cell (font size, font, displaying position, etc).

*3 Index number 0: “Comment 1” Index number 1: “Comment 2” Index number 2: “Comment 3”

4. Save the report template file. • The “Template.xlsx” is a read-only file. It is indicated to save as another name when trying to save the file. Input any name and save the file into any destination. • Do not cancel the read-only setting of the “Template.xlsx” file.

The settings of template tag is performed with the unit of system.

1. Click [Add] (E) or [Edit] (F). →→The “Tag setting for Report” window opens.

2. Select the attribute information from the list (J).

Tags can be narrowed down by selecting “Category”, “Group”, “Element”, and “Tag name” and clicking .

3. Set the display method (K) (L). Input/Selection Items Items

Contents

Prefix

Input a Prefix for displaying

Surfix

Input a Surfix for displaying

Display elements

Input the display elements of tag

Display magnification Input the display magnification of tag Valid number of digits Input a valid number of digits of the value of tag Comment

Input a comment

11

4. Click [OK] (M). →→The system returns to the “Report tag settings” window.

■ Editing Formats

1. Click [Edit Template] (C). →→The “Format list” window opens.

2. Click [Add] (P). • To delete the format, select the format in the list (N) and click [Delete] (Q). • The order of the format can be changed by clicking [Move up]/[Move down] (O). →→The “Open” window opens.

3. Select the format file and click [Open]. 4. Click [Close] (R). →→The system returns to the “Report tag settings” window. FCR PRIMA Console Reference Guide

897N101440B

Changing Display Settings of the "Main" Window and "Image Viewer"

*2 Specify which image in the “Thumbnail” is inserted. Since the index number starts from 0, input 0 to specify the 1st thumbnail image.

■ Adding/Editing Template Tags

11-17

11.2.7 Display Setting of “Image Viewer” Window Sets the font size or display size of the sub-window, study summary, image cart, and sets the actions during study storage.

■ Operation Windows ●

■ Setting Actions during Image Display or Storage

“Style setting” Window - “Display settings” Window

● Setting Color or Size of the “Image Viewer” Window

1. Change style settings. (A)

Selection Items Items

Contents

Image display

(B)

● “Style setting” Window - “Save settings” Window

(C)

Standard*1 font size of attribute information

Select font size from 5 to 96pt

Minimum*1 font size of attribute information Width of Input the width and height of thumbnail image thumbnail image in pixel (Pixel) Height of thumbnail image (Pixel) Color of attribute 1. Click . information 2. Select the color.

11

(D)

Color of ruler

Changing Display Settings of the "Main" Window and "Image Viewer"

11-18

Color of reference line

Color of patient positioning Color of image selection frame Study summary Font size

Select from “Large”, “Medium”, “Small”

Opening method of a study

• Replace • Add • Compare

Width of Input the width and height of thumbnail image thumbnail image in pixel (Pixel) Height of thumbnail image (Pixel)

FCR PRIMA Console Reference Guide

897N101440B

Items

Contents

Image cart Font size

Select from “Large”, “Medium”, “Small”

Width of Input the width and height of thumbnail image thumbnail image in pixel (Pixel) Height of thumbnail image (Pixel) *1 The text size of the attribute information switches to standard/minimum/hidden according to the display layout size of images.

2. Click [OK] (B).

● Setting Actions During Study Storage 1. Set the actions during storage (C). Selection Items Items

Contents

Confirmation dialog whether or not changes are saved will be displayed.

Put a check mark to display the confirmation dialog.

FCR image processing parameter

Select whether or not to save the FCR image processing parameter during study storage.

11

2. Click [OK] (D).

Changing Display Settings of the "Main" Window and "Image Viewer"

→→The system returns to the “Image Viewer” window.

FCR PRIMA Console Reference Guide

897N101440B

11-19

11.2.8 Returning to Standard Settings Returns the changed settings of “Image Viewer” window to the condition when installing.

■ Operation Window

■ Returning to Standard Settings

● Setting Windows (Example: “Annotation

1. Click [Other] (A).

setting” Window)

2. Click [Return to the standard settings]. →→The settings return to the standard settings.

3. Click [OK] (B).

(A)

(B)

11 Changing Display Settings of the "Main" Window and "Image Viewer"

11-20

FCR PRIMA Console Reference Guide

897N101440B

11.2.9 Importing/Exporting Settings Imports/Exports the settings of “Image Viewer” window.

■ Operation Window

■ Import Settings

● Setting Windows (Example: “Annotation

1. Click [Other] (A).

setting” Window)

2. Select [Import of setting information]. →→The “Open” dialog appears.

3. Select the xml file of the config information, and click [Open].

→→The settings are updated according to the imported file.

(A)

■ Export Settings 1. Click [Other] (A). 2. Select [Export of setting information]. →→The “Save As” dialog appears.

3. Select the folder of export destination, input a file name and click [Save].

→→The settings are exported.

11 Changing Display Settings of the "Main" Window and "Image Viewer"

FCR PRIMA Console Reference Guide

897N101440B

11-21

11

Changing Display Settings of the "Main" Window and "Image Viewer"

897N101440B

FCR PRIMA Console Reference Guide

11-22

Chapter

12

Using “Study Utilities”

12.1 Functions of “Study Utilities”...........................................12-2 12.2 Starting up and Exiting “Study Utilities”......................... 12-5 12.3 Searching for Studies and Creating New Queries.......... 12-6 12.3.1

Searching for Studies................................................... 12-6

12.3.2

Creating New Queries.................................................. 12-8

12.3.3

Editing Queries........................................................... 12-11

12.4 Manipulating Studies....................................................... 12-12

12 Using “Study Utilities”

FCR PRIMA Console Reference Guide

897N101440B

12-1

12.1 Functions of “Study Utilities” Each part of the study utilities is described below.

• Buttons and menus which cannot be selected or clicked exist by installation condition for option or by system settings at the time of installation. Buttons and menus which cannot be selected or clicked will be displayed in gray or hidden. • Usable functions for study images are different from studies in the PC and disk. The following functions cannot be utilized for studies in the disk. - 3) Thumbnail images - 6) Storage/Print - 7) List operations • Only the storage disk and work disk can be used in the “Study Utilities”. Disk Types: “2.7 ■Disk Types”

2) Studies

Keyword entry field

5) Study operations 1) Search queries

6) Storage/Print 7) List operations 8) Update, “Image Viewer” 3) Thumbnail images

12

4) Page navigation

1) Search queries Shows registered search queries. Only those studies matching the selected set of search queries appear in Study list.

Using “Study Utilities”

“12.3 Searching for Studies and Creating New Queries”

• To select a single study, the following are shown below. When a study is selected, check mark is put in the Selection check box, with its row displayed in green. - Select a row of study. - Put check marks to “Selection” check boxes.

“PC” that appeared as a search query is a name that was set at time of installation and may vary from one facility to another. To change this name, contact our official dealer.

2) Studies A list of the studies that matches the set of conditions selected in “Search queries”, is displayed. To narrow down the search, enter a keyword in the “Keyword entry field”. “12.3.1 Searching for Studies”

12-2

FCR PRIMA Console Reference Guide

897N101440B

• To select multiple studies, the following methods are shown below. - Put check marks to “Selection” check boxes. - To select multiple discontinuous studies, select their lines while holding the Ctrl key. - To select a continuous sequence of studies (range selection), drag the mouse up or down over the list with the row of a given study on the list selected. Alternatively, select the row of each study while holding down the Shift key. • To select all the studies in the current page, click [Select All] in the Study List on the right-hand side of the window.

Display Items Items No.

*1

Selection

Items Contents Serial numbers 1, 2, 3 ... of the studies in the page given from the top Put a check mark when selecting studies

Patient ID

ID number of patient

Patient Name

Name of patient

Sex

Sex of patient

Date of Birth

Date of birth of patient

Study Instance UID

Numbers uniquely assigned to each study

Date of Registration

Date on which the study was registered (scheduled)

Time of Registration

Time at which the study was registered (scheduled)

Study Date

Date on which the exposure was started

Study Time

Time at which the exposure was started

Study Description Name of the study menu that has been created with the “User Utilities” Study Status

Progress of the study

Scheduled

An exposure operation is scheduled with patient information and exposure menus having been registered

Started*2

An exposure is in progress or suspended The reading of all images is completed and [Upon Completion] is clicked

Verified

The image quality of images read is verified and [Upon Verification] is clicked

Storage/Print

The status of image storage/ printing

Undelivered

Images are ready to be stored and delivered

Delivering

Images are being delivered from storage/printing

Delivered

Storage/printing has ended

Whether the study is locked or not

-

The image is not locked

Locked

The image is locked

# of Img

Number of images that have been read or that are scheduled to be read

Image

Whether images are stored on the FCR PRIMA Console system or not

Nothing*3

No images are stored

Existing

The state in which an exposed image is stored or image is to be read

Storage Disk

Name of the storage disk on which the image is saved

DB

Whether the study i s managed by FCR PRIMA Console or not

-*4

The image is not managed

DB managing

The image is managed

History

Whether study information has been modified after the study was verified

-

Study information has not been modified

Modified

Study information has been modified

DICOM transmission

Name of the DICOM server to where images were transmitted

Modality

Modality of study

CR

12

A CR study

*1 When a number is selected, the image in that study is displayed in the “Image Viewer” window.

Using “Study Utilities”

Completed

Lock

Contents

*2 There are exposure menus in the study from which images are to be read. *3 To browse the image for this study, read it from a storage disk. *4 This is the image which has been read from a work disk.

FCR PRIMA Console Reference Guide

897N101440B

12-3

3) Thumbnail images A thumbnail images of the selected study displays. 4) Page navigation Display other pages without changing the current status of study selection. : First page : Previous page : Next page : Last page 5) Study operations Enable each operation on selected studies. “12.4 Manipulating Studies”

Selecting “Study Operations” hides the operational menu.

6) Storage/Print Enable to save or print selected studies. “12.4 Manipulating Studies”

Selecting “Storage/Print” hides the operational menu.

7) List operations Select and deselect all studies in the current page, and update the display of studies in the study list. Selection Items

12

Items

Contents

Using “Study Utilities”

Select All

Select all the studies in the current page

Unselect All

Deselect all the studies in the current page

Update

Update the display of the study list to the latest status

Selecting “List Operations“ hides the operational menu.

12-4

FCR PRIMA Console Reference Guide

897N101440B

8) Update, “Image Viewer” Updating a study list and displaying images in the “Image Viewer” window can be performed. Functions of Each Button Buttons

Functions

[Update]

Update the display of the study list to the latest status

[Image Viewer]

Display the image of the selected study in the “Image Viewer” window

12.2 Starting up and Exiting “Study Utilities” This section describes the procedures for starting up and exiting “Study Utilities” in the “Main” window.

■ Operation windows

■ Starting up “Study Utilities”

● “Main” Window

1. Select

- “Study Utilities” (A) in the “Main” window.

→→The “Study Utilities” window opens.

■ Exiting

(A)



1. Click [Close] (B).

“Study Utilities” Window

→→Returns to the “Main” window.

(B)

12 Using “Study Utilities”

FCR PRIMA Console Reference Guide

897N101440B

12-5

12.3 Searching for Studies and Creating New Queries Searches with predefined queries, creates new queries/search, and narrows down the search results by patient ID/name.

12.3.1 Searching for Studies Searches studies with predefined queries and displays results in the list.

■ Operation Window

Selection Items Items

● “Study Utilities” Window (B)

(A)

■ Searching for Studies 1. Select search queries (A) for “PC” or “Disk”. Search queries for “Disk” is only “All”.

12

Click “▼” to hide the search queries.

Contents

All

Display all studies

Today*1

Display only studies which have been started the exposure with today’s study date

Scheduled

Display only studies in the “Scheduled” study status

Exposing

Display only studies in the “Started” study status

Unverified

Display only studies in the “Completed” study status

Undelivered

Display only studies in the “Undelivered” Storage/Print condition

Temporal

Where the FCR PRIMA Console system is connected to an EMR (optional), displays only those studies on patients specified by the EMR in the study list, among all studies that are saved on the PC.

Using “Study Utilities”

*1 The studies that are displayed with [Today] clicked as a search query are those that have started an exposure operation. The studies that have been scheduled by clicking [Register] in the “Schedule Exposure window” are not displayed in the study list. The studies that are in the “Scheduled” study status, including those scheduled by clicking [Register] in the “Schedule Exposure” window, can be manipulated into view in the study list in one of the following methods: • [Scheduled] • [All] →→The study which matches with selected search queries displays in the study list.

12-6

FCR PRIMA Console Reference Guide

897N101440B

■ Narrowing down Search Study results can be narrowed down after search queries are selected and search results are displayed.

1. Set search queries to narrow search in the keyword entry field (B).

“Search Queries” are entered in different ways according to the input items. See the following table.

• If you enter multiple keywords in the keyword entry field, the studies that match all of these input keywords are listed. • If keywords are entered as search queries, the keywords entered in the keyword entry field will be given priority. →→The search for studies is narrowed down to meet the specified set of search queries.

• Click [Update] to update the status of the list. Input Items Items Patient ID*1 Study Instance UID # of Img Storage Disk

Contents Enter a keyword in the entry field and press the “Enter” key.

• To change the display order, select the header of the field to be used as a reference. Each time you select the header, the display order toggles between ascending (▲) and descending (▼).

Studies which match completely with the keyword searches.

Patient Name

Enter a keyword in the entry field and press the “Enter” Study Description key. DICOM Transmission

Studies which partly match with the keyword searches.

Sex

Click at the rightmost end of each entry field to open a pulldown menu and select a query.

Study Status Storage/Print Lock Images DB Management

12

History

Date of Birth Study Date Date of Registration

Using “Study Utilities”

Modality at the rightmost end Click of each entry field. When the “Specifying the Date“ window opens, specify a date and click [Setup].

*1 Depending on how FCR PRIMA Console was configured for the ID digit setting at installation, each input patient ID is registered with 0s in the beginning accordingly. For example, if a patient ID of 1 is entered where FCR PRIMA Console is configured for 10 digits, the ID is displayed as 0000000001.

FCR PRIMA Console Reference Guide

897N101440B

12-7

12.3.2 Creating New Queries Creates new search queries with matching patient ID, sex, and study date etc. to search PC or disks. Also, items to be displayed in the search results can be set.

■ Operation Windows

■ Creating New Queries

● “Study Utilities” Window

1. Click [Add New Queries] (A). (A)

→→The “Setting List Search Conditions” window opens.

2. Click each window (B) and set search queries.

• Set search queries. When you open the “Set Search Queries” window, four windows are displayed: “Search Items”, “Items/ Refresh”, “Order” and “Display Items”. In each window, you can set search conditions.

● “Setting List Search Conditions”

• Since the study list comes up with the current choice of search queries, edit them to create new search queries. (B)

“Search queries” window: Set the items to be used as search conditions.

(C)

12 Using “Study Utilities”

Input and Selection Items Items Query Name

Contents Enter the search query name.

Search Target

12-8

FCR PRIMA Console Reference Guide

897N101440B

PC

Search through FCR PRIMA Console system for studies. The search query created is displayed under the PC on the left-hand side of the window.

Disk

Search the disk for studies. The search query created is displayed under the Disk on the left-hand side of the window.

Items

Contents

Query Keyword

Set a unique keyword for each search query. The entry field is already filled with “Query ...”.

Search Query Search Items*1

Set items to be used as a search query. When using as a search query, put a check mark to the check box and enter “Search Queries“.

Patient ID

Enter a keyword in the entry field and press the “Enter” key.

Patient Name

“Items/Refresh” window: Set the number of studies to be viewed in the “Study Utilities” and the interval at which the list is refreshed automatically.

Study Instance UID Study Description # of Img

Input and Selection Items

Storage Disk

Items

DICOM Transmission Sex Study Status Storage/Print

Click at the rightmost end of each entry field to open a pulldown menu and select a condition.

Lock

Contents

Number of Select the number of studies you studies displayed want to view in a single page of within a window the “Study List” window. Auto-refresh upon query selection

Select whether to refresh the list automatically when search queries are selected.

Periodical refresh Select whether to refresh the list automatically upon expiry of a certain period of time.

Images DB Management History

Refresh interval*1 Enter an interval of time auto refresh interval in minutes between 1 and 60 minutes.

Modality

Font Size

Date of Birth Study Date

at the right end of each

entry field. When the “Date

*1 Set when “Yes” is selected in the “Periodical refresh”.

12

Settings “ window opens, specify Using “Study Utilities”

Date of Registration

Click

Select a font size to use for the study list.

a date and click [Setup].

FCR PRIMA Console Reference Guide

897N101440B

12-9

“Order” window:

“Display Items” window:

Set the order of list display.

Set the items that are displayed in the “Study Utilities”.

Selection Items

Put check marks to the check box for each item to be viewed in the “Study Utilities”.

Items First-Priority Item

Contents Select the item that requires first priority upon relocation. Select ascending or descending order.

Second-Priority Item

Set Second-Priority Item as a First-Priority Item

Third-Priority Item

Set Third-Priority Item as a FirstPriority Item

• Remove the check mark from the check box for each item to be hidden in the “Study Utilities”. • Selected items appeared in the “Study Utilities” from top to bottom in sequence, left-aligned. • To change the display position of an item, choose (Move Down) or (Move UP) with the selected display item to be repositioned.

3. Click [Save] (C).

12 Using “Study Utilities”

12-10

FCR PRIMA Console Reference Guide

897N101440B

12.3.3 Editing Queries The following operations can be performed to registered search queries. • Editing search queries • Deleting search queries • Relocating search queries • Setting the search queries under which a search starts up right after the startup of FCR PRIMA Console

■ Operation Windows

■ Editing Queries

● “Study Utilities” Window

1. Click [Arrange Queries] (A). →→The “Search Queries List” window opens.

(A)

2. Click the desired button (B) and edit study queries.

Functions of Each Button Buttons

● “Search Queries List” Window

Functions

Edit*3 *4

Open the “Setting List Search Conditions” window and edit the setting for selected search queries.

Delete*1 *2 *5

Delete created search queries.

Move Up Move Down

Relocate the display position of each search query in the “Study Utilities”.

Set to Initial Condition

Set the initial condition for displaying “Study Utilities”.

(B) *1 To delete the search query set to initial condition, set another search query to initial condition; then, the search query can be deleted. *2 Predefined search queries cannot be deleted.

(C)

*4 The “Setting List Search Conditions” window opens with settings of the selected search query. “12.3.2 Creating New Queries” *5 A message displays to verify. Click [OK] to delete.

12 Using “Study Utilities”

● “Setting List Search Conditions” Window

*3 You may not be able to edit some windows or items in the predefined set of search queries depending on the selected search queries.

3. Click [Save] (C). • To delete the settings which have been set, click [Delete]. • You can do the same by clicking [Save as is] after selecting the search query, entering the keyword, and then narrowing down the search result.

FCR PRIMA Console Reference Guide

897N101440B

12-11

12.4 Manipulating Studies The following operations can be performed. - Viewing studies displayed in the “Study Utilities” of “Image Viewer” - Saving images to disks - Loading images from disks - Deleting studies and so on.

■ Descriptions of Functions For the studies in the PC or disk, usable functions for images of studies are different. Each usable function is shown below.

● Images in the PC

• Starting up the “Image Viewer” to view images “■Starting up the “Image Viewer””

• Reconducting a study / editing study information “■Reconducting a Study and Editing its Study Information”

• Locking / unlocking a study “■Locking and Unlocking Studies”

• Changing a study status to “Completed” and “Verified” “■Completing and Verifying Studies”

12 Using “Study Utilities”

12-12

• Starting up the “Image Viewer” to view images “■Starting up the “Image Viewer””

• Loading an image from a disk into the PC “■Loading Images from a Disk”

When the study is imported from the disk, the verification window opens. • If study importing has completed: Studies that have been imported from any source other than the storage disk, such as those saved at another facility, have ID numbers and other patient information based on different means used for FCR studies. When reading the study image stored in another facility, be sure to modify the patient information for the local facility. After completion of modification, click [Update] to check whether the patient information has been modified properly. • If the image could not be imported: - Check the disk and drive status. - To view study images stored on the PDI disk, browse them in the PDI viewer.

• Deleting a study “■Deleting Studies”

• If certain study could not be imported:

• Saving / Printing images of studies “■Saving, Printing, and Transmitting Study Images”

FCR PRIMA Console Reference Guide

● Images in the Disk

897N101440B

- After a study has been successfully imported, start up the “Image Viewer” window to correct the relevant patient information. - To view study images stored on the PDI disk, browse them in the PDI viewer.

■ Operation Window

■ Starting up the “Image Viewer”

● “Study Utilities” Window

Starts up the “Image Viewer” to view images in the “Image Viewer” window.

1. Select the images that are in the “Completed” or “Verified” study status from the list and click [Image Viewer] (A).

(B) (C) (D) (E) (F) (G)

(A)

• When the image in the disk is selected, the “Requesting for Disk Insert dialog“ appears and the image starts to be imported from the disk. “■Loading Images from a Disk” • Images that have been loaded for viewing in the “Image Viewer” window are automatically deleted from the FCR PRIMA Console system when a specified retention period expires. A retention period can be set with the “User Utilities”. “14.7 ■Changing the Time Period for Data Storage”

• Multiple studies can be selected. • You can do the same by: - Selecting a study in the study list and clicking [Image Viewer] on the right-hand side of the window. - Entering the number of the study to be viewed in the “Image Viewer” window (item at the left end of the list). →→The “Image Viewer” window opens.

Click [OK] to continue excluding the studies which cannot be processed.

FCR PRIMA Console Reference Guide

897N101440B

12 Using “Study Utilities”

When the study except in the “Verified” or “Completed” study status is included in the selected studies, the verification window opens to continue the process.

12-13

■ Reconducting a Study and

Editing its Study Information

In conducting a restudy, sets exposure menus for performing additional and re-exposure operations. Study information can be edited by modifying or relocating menus in the display order, replacing images and making modifications to the Storage/Print settings. Also, the patient information can be modified.

1. Select a study from the list and click [Edit/ Re-Study] (B).

■ Completing and Verifying Studies Updates the study status to “Completed”, and verifying the study updates the study status from “Completed” to “Verified” (the status appears in the list as “Verified”).

● Completing an Exposure Operation

Completes the exposure operation for a study after verifying that images have been properly read in all studies. When a study is completed, it is updated from the “Started” to the “Completed” study status.

→→The “Main Menu Modification” window opens. “Chapter 9 Editing Study Information”

■ Locking and Unlocking Studies If a study is locked, it is no longer accessible for deletion. If a study is unlocked, it becomes accessible for deletion.

1. Select a study from the list and click [Lock] or

If film printing (optional) is programmed by the “User Utilities”, the images in the study are automatically printed or saved when the study is verified. “14.5 Setting of Printing, Saving and Transmitting Images”

● Verifying a Study

Verifies the image quality of read images in a study. When a study is verified, the study is updated from the “Completed” to the “Verified” study status.

[Unlock] (C).

• Multiple studies can be selected. • The selected studies are locked and are marked “Locked” in the Lock column of the list. • The selected studies are unlocked and are marked “-” in the Lock column of the list.

If film printing (optional) is programmed, the images in the study are automatically printed or saved when the study is verified. “14.5 Setting of Printing, Saving and Transmitting Images”

12 Using “Study Utilities”

12-14

If automatic saving of the images in a study to a storage disk is programmed, the images are saved to the disk when the study is verified.

FCR PRIMA Console Reference Guide

897N101440B

1. Select the study to change the study status from the study list, and click “Complete Study” or “Verify Study” (D).

Multiple studies can be selected.

If multiple studies have been selected and verified, they might take considerable time before being updated to the “Verified” study status depending on the number of selected studies. If the update of the study status does not complete minutes after the start of [Verify Study], contact our official dealer leaving the FCR PRIMA Console system untouched. Do not force FCR PRIMA Console to shutdown. To do so may result in failure to execute studies next time FCR PRIMA Console starts up. →→The verification dialog appears.

2. Click [OK]. When the status of study that an unexposed menu exists is changing to “Completed”, the verification window opens. To delete the exposure menus from which images are to be read and change study status to “Completed”, click [OK].

■ Loading Images from a Disk Unless images in a study have been saved to the FCR PRIMA Console system, loads images from a disk. The loaded images can be displayed in the “Image Viewer” window.

• The following operations cannot be performed on images loaded from a disk: - Rotating and reversing images - Saving FCR image processing parameters - Editing markers - Editing study information - Verifying and suspending a study - Disabling and enabling images, re-exposure • If the size of image data to be loaded from a disk is larger than the size of available space on the FCR PRIMA Console system, existing images stored in the FCR PRIMA Console database are automatically deleted in sequence in ascending time order of browsed studies.

Studies that do not have images saved on the FCR PRIMA Console system are marked with “Nothing” in the Image column.

1. Select a study from the list and click [Load Disk Images] (E).

Multiple studies can be selected. →→The “Retrieve Images from Disk” dialog appears.

2. Select the purpose for which you are loading • For System: Load images to save to the FCR PRIMA Console system. • Temporarily: Load images for temporary browsing on the FCR PRIMA Console system. Specify the number of days in “Store for“.

Using “Study Utilities”

images from a disk and click [Retrieve].

12

Loaded images are automatically deleted when the specified storage period expires. →→The “Requesting for disk insert” dialog appears.

FCR PRIMA Console Reference Guide

897N101440B

12-15

3. Select the drive used for loading images and click [OK].

→→The “Ready to retrieve Images” dialog appears and the disk drive tray opens.

■ Deleting Studies The deleted study cannot be restored.

4. Place the requested disk in the drive tray, and close the tray.

5. Click [OK] in the “Ready to retrieve images” dialog.

Studies in the following study status cannot be deleted: • Study in progress • Studies shown in the “Image Viewer” window • Studies being printed or being saved

• If no disk is mounted, the message “Preparing to retrieve images” displays. When the disk mounting process completes, the “Retrieve Images from Disk” dialog appears, and allowing you to verify the status of image loading. When the image loading process completes, the message “Loading of the study completed” displays. Click [OK]. • To manipulate the imported studies, search through the “Study Utilities” for target studies stored in the PC. “12.3 Searching for Studies and Creating New Queries”

• Locked studies

1. Select the study from a list and click [Delete] (F).

Multiple studies can be selected. →→The verification dialog appears.

2. Click [OK]. If a study which cannot be deleted is existed, the verification window opens. To delete only those studies that can be deleted among all selected studies, click [OK].

12 Using “Study Utilities”

12-16

FCR PRIMA Console Reference Guide

897N101440B

■ Saving, Printing, and

Transmitting Study Images

1. Select the study from a list and click each item of [Store/Print] (G).

Multiple studies can be selected.

The following functions can be used for study images: Selection Items Items Create PDI

Contents Save into the PDI disk “7.7 Saving to PDI Disks”

Film Print*1

Print on film “8.3 Printing on a Film (Optional)”

Freelayout

Print images in Freelayout “8.4 Free Layout Printing”

Work Disk

Save into the work disk “7.5 Saving to Work Disks”

Storage Disk

Save into the storage disk “7.3 Saving to Storage Disks”

DICOM Transmission*1 Transmit to DICOM server “7.4 Transmitting Studies to DICOM Server (Optional)”

12

*1 Optional function

Using “Study Utilities”

FCR PRIMA Console Reference Guide

897N101440B

12-17

12 Using “Study Utilities”

12-18

FCR PRIMA Console Reference Guide

897N101440B

Chapter

13

Configuring “Client Utilities”

13.1 Setting Items of “Client Utilities”...................................... 13-2 13.2 Starting up and Exiting “Client Utilities”......................... 13-2 13.3 Maintaining Patient Information.......................................13-4 13.4 Verifying the IP Use Frequency........................................13-7 13.5 Setting User Information (For Administrator)................. 13-8 13.5.1

User Information Maintenance.................................... 13-8

13.5.2

Changing User Group Operational Authorities....... 13-12

13.5.3

Setting Images for Display in the “User Selection” Window................................ 13-14

13.6 Editing Information about the Logged-in User.............. 13-16 13.7 Verifying History Information..........................................13-17 13.8 Disk Maintenance . .......................................................... 13-20

13 Configuring "Client Utilities"

FCR PRIMA Console Reference Guide

897N101440B

13-1

13.1 Setting Items of “Client Utilities” The following operations can be performed in “Client Utilities”: • Registering, editing and deleting patient information • Verifying the IP use frequency • Setting user information • Verifying history information • Maintenance of disks

13.2 Starting up and Exiting “Client Utilities” Starts the “Client Utilities” from the “Main” window. This section describes how to use each function and how to exit the “Client Utilities”.

■ Operation Windows

● “Main Menu”

Starts the functions of “Client Utilities” from the desired button.

● “Main” Window

Starts up the “Client Utilities“.

(B)

(C)

13 Configuring "Client Utilities"

(A)

13-2

FCR PRIMA Console Reference Guide

897N101440B

■ Starting up the “Client Utilities” ■ Exiting the “Client Utilities” 1. Click

(A) in the “Main” window.

1. Click [Close] (C) in the “Main Menu” window. →→The “Main Menu” window closes and exit the “Client Utilities”.

2. Click [Client Utilities]. →→The “Main Menu” window opens.

3. Click the desired buttons (B) according to the functions to be used.

Functions of Each Button Buttons Patient Info Maintenance

Functions Register, edit or delete patient information. “13.3 Maintaining Patient Information”

IP Use Frequency

Verify the IP use frequency for image reading. Verify IP use frequency as a guideline for IP replacement. “13.4 Verifying the IP Use Frequency”

User Settings (for Administrator)

Set various information and the authority about users of FCR PRIMA Console. “13.5 Setting User Information (For Administrator)”

User Settings (for General Users)

Edit the information (display name, image... etc) about the currently logged-in user. “13.6 Editing Information about the Logged-in User”

History Info

Verify the histories of login to FCR PRIMA Console, access to studies... etc. “13.7 Verifying History Information” Maintain the disks set in the PC drive and restore the database.

Configuring "Client Utilities"

Disk Maintenance

13

“13.8 Disk Maintenance”

Guidance is displayed in the “Client Utilities” window to lead the user through operating procedures.

FCR PRIMA Console Reference Guide

897N101440B

13-3

13.3 Maintaining Patient Information Registers new patient information in FCR PRIMA Console database. Also, edits or deletes registered patient information.

■ Registering New Patient

■ Operation Windows ●

Information

“Patient Information List”

Searches the patient to be maintained and the result will be shown in the list. (A) (C)

(B)

(D)

Inputs patient ID, names... etc, and the patient information will be registered in database.

1. Click [New] (H). →→The “Patient Information Registration“ window opens.

2. Enter the patient information ( I ) to be registered in the FCR PRIMA Console database. Input and Selection Items Items

(E)

(F)

(G)

(H)



“Patient Information Registration” (I)

13

(J)

ID*1*2

An ID is a key used to identify patient information to FCR PRIMA Console

Name*3

Enter a patient name within 64 characters

Sex

Click from [Male], [Female], and [Other]

DOB

Enter a date of birth according to the sample entry

Height

Patient’s height Input with half-byte numbers

Weight

Patient’s weight Input with half-byte numbers

Patient comment

Input comment

*1 The following characters cannot be used. “#%&`+?/\*@< However, the space can be used but a patient ID which consists of solely space is not allowed.

(L)

(K)

Contents

Configuring "Client Utilities"

● “Edit Patient Information”

*2 IDs of patient information already registered in the FCR PRIMA Console database cannot be used. (M)

*3 The following characters cannot be used: *, @, “, #, %, &, ‘, +, ?, \, /,