ESA612 Electrical Safety Analyzer

Service Manual

PN 3477999 June 2010, Rev. 1 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty. This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

07/07

Contents (continued)

Notices All Rights Reserved  Copyright 2010, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.

Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.

Unpacking and Inspection Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.

Technical Support For application support or answers to technical questions, either email [email protected] or call 1-800- 850-4608 ext 2560 or 1-440-498-2560.

Claims Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.

Standard Terms and Conditions Refunds and Credits Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable condition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit. Restocking Charges Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and accessories will be applied to all returns. Return Procedure All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling. Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:   

Use a double-walled carton of sufficient strength for the weight being shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts. Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.

Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at 1-800- 850-4608 ext 2560 or 1-440-498-2560.

Repair and calibration: To find the nearest service center, go to www.flukebiomedical.com/service, or In the U.S.A.: Cleveland Calibration Lab Tel: 1-800-850-4606 Email: [email protected] Everett Calibration Lab Tel: 1-888-993-5853 Email: [email protected] In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-402-675300 Email: [email protected] In Asia: Everett Calibration Lab Tel: +425-446-6945 Email: mailto:[email protected]

Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.

WARNING Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.

Restrictions and Liabilities Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.

Manufacturing Location The ESA612 Electrical Safety Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.

Table of Contents

Chapter 1

Title

Introduction and Specifications......................................................... 1-1 Introduction........................................................................................................ Intended Use ...................................................................................................... Electrical Symbols ............................................................................................. Product Manuals ................................................................................................ About this Manual ............................................................................................. Safety Information ............................................................................................. Specifications..................................................................................................... Detailed Specifications ......................................................................................

2

Page

1-3 1-3 1-3 1-3 1-3 1-4 1-5 1-5

Theory of Operation ............................................................................ 2-1 Introduction........................................................................................................ Block Diagram Description ............................................................................... A1 Power PCA .............................................................................................. A2 Meter and User Interface PCA ................................................................ Block Diagram to Schematic Cross Reference .................................................. A1 Power PCA Schematic............................................................................. A2 Meter and User Interface PCA Schematic............................................... Detailed Circuit Description .............................................................................. A1 Power PCA .............................................................................................. Equipment Outlet Relays .......................................................................... Equipment Outlet Current Measurements................................................. Mains Measurement .................................................................................. Isolated Mains Voltage (Mains on Applied Parts or MAP) ...................... Connectors ................................................................................................ Isolation..................................................................................................... Relay Drivers ............................................................................................ Isolated Signals ......................................................................................... Power Supplies.......................................................................................... Resistance Current Source ........................................................................ Insulation Resistance Test Voltage ........................................................... A2 Meter and User Interface PCA Schematic............................................... Meter Inputs .............................................................................................. Patient Load .............................................................................................. i

2-3 2-3 2-3 2-4 2-5 2-9 2-10 2-11 2-11 2-11 2-11 2-11 2-12 2-12 2-12 2-12 2-12 2-12 2-12 2-13 2-13 2-13 2-13

ESA612 Service Manual

Overvoltage Protection.............................................................................. Mains on Applied Parts Compensation..................................................... AC Only .................................................................................................... DC Only .................................................................................................... Resistance Measurement ........................................................................... Relay Drivers ............................................................................................ Meter Power Supplies ............................................................................... Relay Disable ............................................................................................ Applied Parts............................................................................................. Meter Microprocessor ............................................................................... User Interface Microprocessor .................................................................. Keyboard and Display............................................................................... USB Port ................................................................................................... U/I Power Supplies ................................................................................... 3

Verification and Calibration................................................................ 3-1 Introduction........................................................................................................ Required Equipment .......................................................................................... Analyzer Verification ........................................................................................ Outlet Tension Tests...................................................................................... Setup .............................................................................................................. Verification Procedure................................................................................... Calibration Procedure ........................................................................................ Setup .............................................................................................................. Adjustments...................................................................................................

4

3-3 3-3 3-4 3-4 3-4 3-5 3-52 3-52 3-52

Remote Operations ............................................................................. 4-1 Introduction........................................................................................................ Serial Communication Connections .................................................................. Serial Communication ....................................................................................... Modes ................................................................................................................ Reading Formats ................................................................................................ Command Specifications ................................................................................... General Commands ....................................................................................... Remote Commands ....................................................................................... ECG Commands............................................................................................ Calibration Commands .................................................................................. Diagnostic Commands................................................................................... Error Responses ................................................................................................. Status Word Definitions..................................................................................... Firmware Updates..............................................................................................

5

2-13 2-13 2-14 2-14 2-14 2-14 2-14 2-14 2-14 2-15 2-15 2-15 2-15 2-15

4-3 4-3 4-3 4-3 4-5 4-5 4-5 4-6 4-11 4-12 4-18 4-21 4-22 4-24

Maintenance......................................................................................... 5-1 Introduction........................................................................................................ Testing and Replacing the Fuses........................................................................ Cleaning the Analyzer ....................................................................................... Disassembly/Assembly ...................................................................................... Troubleshooting ................................................................................................. Power on Failure............................................................................................ Output Failures .............................................................................................. Meter Failures................................................................................................ MAINS Power........................................................................................... Differential Current and Equipment Current ............................................ Meter Troubleshooting .................................................................................. ii

5-3 5-3 5-3 5-4 5-4 5-4 5-5 5-5 5-5 5-5 5-5

Contents (continued)

6

Replaceable Parts ............................................................................... 6-1 Introduction........................................................................................................ 6-3 How to Obtain Parts........................................................................................... 6-3 Parts List ............................................................................................................ 6-4

7

Schematics .......................................................................................... 7-1 Introduction........................................................................................................ 7-3

Appendices A Datasheets.................................................................................................... A-1

iii

ESA612 Service Manual

iv

List of Tables

Table 1-1. 2-1. 2-2. 3-1. 3-2. 3-3. 4-1. 4-2. 4-3. 6-1. 6-2. 6-3.

Title Symbols.................................................................................................................. A1 Power PCA to Schematic Cross Reference ...................................................... A2 Meter and User Interface PCA Schematic Cross Reference ............................ Required Equipment for Analyzer Verification ..................................................... Keypad Verification ............................................................................................... Nominal Values...................................................................................................... Reading Formats .................................................................................................... Error Responses ..................................................................................................... Status Word Definitions ......................................................................................... Final Assembly....................................................................................................... A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................

v

Page 1-3 2-9 2-10 3-3 3-6 3-52 4-5 4-21 4-22 6-4 6-7 6-12

ESA612 Service Manual

vi

List of Figures

Figure 2-1. 3-1. 3-2. 3-3. 3-4. 3-5. 3-6. 3-7. 3-8. 3-9. 3-10. 3-11. 3-12. 3-13. 3-14. 3-15. 3-16. 3-17. 3-18. 3-19. 3-20. 3-21. 3-22. 3-23. 3-24. 3-25. 3-26. 3-27. 3-28. 3-29. 3-30. 3-31. 3-32. 3-33. 3-34.

Title ESA612 Block Diagram......................................................................................... Electrical Equipment Outlet Polarity Test ............................................................. Electrical Outlet Ground to NULL Jack Resistance .............................................. GFI Verification Connections ................................................................................ Point to Point Voltage Verification Connections ................................................... Point to Point Resistance Verification Connections............................................... Insulation Resistance Verification Connections..................................................... Insulation Resistance with 110 kΩ Resistor Connections...................................... Mains to PE Verification Connections................................................................... AP to PE Verification Connections........................................................................ AP to NE Verification Connections ....................................................................... Mains to NE Verification Connections .................................................................. Mains to AP Verification Connections .................................................................. DC Leakage Verification Connections .................................................................. DC Leakage Current Verification Connection....................................................... AC Leakage Filter Frequency Response Connections ........................................... Differential Leakage Verification Connections ..................................................... ECG Leakage Functionality Connections .............................................................. Direct Applied Part Verification Ground to RA Connections ............................... Direct Applied Part Verification Red to RA Connections ..................................... Alternate Equipment Leakage: Red to Hot Connections ....................................... Alternative Applied Part Leakage: RA to Hot ....................................................... Alternative Applied Part Leakage: RA to Neutral Connections ............................ Patient Auxiliary Leakage: RA to RL Connections ............................................... Direct Equipment Leakage, Patient Leakage: Ground to RA ................................ Direct Equipment Leakage, Enclosure Leakage: Ground to Red Connections...... MAP Voltage and Current Limits Connections ..................................................... ECG Waveform Connections................................................................................. Equipment Current Calibration Connection........................................................... AC Meter Calibration Connections........................................................................ Leakage Calibration ............................................................................................... Point to Point Calibration Connections.................................................................. Insulation HV Calibration Connections ................................................................. Resistance Calibration Connections....................................................................... Differential Leakage Calibration Connections....................................................... vii

Page 2-7 3-7 3-9 3-10 3-13 3-15 3-16 3-17 3-18 3-19 3-20 3-21 3-22 3-23 3-24 3-26 3-28 3-29 3-31 3-33 3-35 3-36 3-38 3-39 3-41 3-43 3-46 3-48 3-50 3-54 3-56 3-57 3-58 3-59 3-60

ESA612 Service Manual

3-35. 3-36. 5-1. 6-1. 6-2. 6-3. 7-1. 7-2.

GFI Calibration Connections ................................................................................. Equipment Current Calibration Connection........................................................... Fuse Access............................................................................................................ Final Assembly....................................................................................................... A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................ A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................

viii

3-61 3-65 5-3 6-6 6-11 6-19 7-5 7-10

Chapter 1

Introduction and Specifications

Title Introduction.......................................................................................................... Intended Use ........................................................................................................ Electrical Symbols ............................................................................................... Product Manuals .................................................................................................. About this Manual ............................................................................................... Safety Information ............................................................................................... Specifications....................................................................................................... Detailed Specifications ........................................................................................

Page 1-3 1-3 1-3 1-3 1-3 1-4 1-5 1-5

1-1

ESA612 Service Manual

1-2

Introduction and Specifications Introduction

1

Introduction The Fluke Biomedical ESA612 Electrical Safety Analyzer (hereafter the Analyzer) is a full-featured, compact, portable analyzer, designed to verify the electrical safety of medical devices. The Analyzer tests to domestic (ANSI/AAMI ES1, NFPA 99) and international (IEC62353, AN/NZS 3551, and parts of IEC 60601-1) electrical-safety standards. The integrated ANSI/AAMI ES1 and IEC60601-1 patient loads are easily selectable.

Intended Use The Analyzer is intended for use by trained service technicians to perform periodic inspections on a wide range of medical equipment. The testing procedures are menudriven, and simple to operate.

Electrical Symbols Table 1-1 lists the symbols with their definitions found on the Analyzer and used in this manual. Table 1-1. Symbols Symbol

Description

W

Important information; refer to manual.

~

Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.

;

Conforms to relevant Australian EMC requirements

)

Conforms to relevant Canadian and US standards

X

Hazardous voltage

P

Conforms to European Union directives

CAT II

˜

IEC Measurement Category II – CAT II equipment designed to protect against transients from energy-consuming equipment supplied from fixed installations. Accessible Functional Earth Terminal

Product Manuals The Analyzer ships with a printed Getting Started Manual and a Users Manual on a CD. The ESA612 Getting Started Manual gives a brief overview of the Analyzers controls and connections. The ESA612 Users Manual provides details on operating the Analyzer.

About this Manual This is the Service Manual for the Analyzer. It contains all of the information a technician will need to verify the Analyzer’s performance to specifications and calibration adjustments to bring the Analyzer’s performance within specification. The manual is divided into the following chapters: Chapter 1 “Introduction and Specifications” provides information on how to safely use the Analyzer and specifications. Chapter 2 “Theory of Operation” provides information on how the Analyzer’s circuitry functions. Chapter 3 “Verification and Calibration” provides information on how to verify the 1-3

ESA612 Service Manual

Analyzer’s performance and bring its performance within specification. Chapter 4 “Remote Operation” contains the remote commands to operate the Analyzer through its computer port. Chapter 5 “List of Replaceable Parts” contains a complete list of part numbers with descriptions and associated illustrations to help locate each part. Chapter 6 “Schematics” contain the schematics of the Analyzer.

Safety Information In this manual, a Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data.

XW Warning To avoid possible electrical shock or personal injury, follow these guidelines:

1-4



Use this Analyzer only in the manner specified by the manufacturer or the protection provided may be impaired.



Read the Users Manual before operating the Analyzer.



Do not connect the Analyzer to a patient or equipment connected to a patient. The Analyzer is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the Analyzer would come in contact with a patient.



Do not use the product in wet or damp locations, around explosive gases or dust.



Inspect the Analyzer before using it. Do not use the Analyzer if abnormal conditions of any sort are noted (such as a faulty display, broken case, etc.)



Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Replace damaged leads before using the Analyzer.



When testing, always be sure to keep your fingers behind the safety barriers on the test leads.



Never open the Analyzer's case. Dangerous voltages are present. There are no user replaceable parts in the Analyzer.



Have the Analyzer serviced only by qualified personnel.



The Analyzer must be properly earthed. Only use a supply socket that has a protective earth contact. If there is any doubt as to the effectiveness of the supply socket earth, do not connect the Analyzer. Do not use a two-conductor adapter or extension cord; this will break the protective ground connection.



Do not use the 15-20 A adapter to power devices rated in excess of 15 A. Doing so may overload the installation.

Introduction and Specifications Specifications



Use extreme caution when working with voltages above 30 V.



Use the proper terminals, functions and ranges for the test being performed.



Do not touch metal parts of the device under test (DUT) during analysis. The DUT should be considered an electrical shock hazard when connected to the Analyzer as some tests involve high voltages, high currents, and/or the removal of DUT earth bond.

Specifications Temperature Operating ............................................................ 10 °C to 40 °C (50 °F to 104 °F) Storage ............................................................... -20 °C to 60 °C (-4 °F to 140 °F) Humidity................................................................. 10 % to 90 % non-condensing

1

Will be using ESA615 specs. So don't waste time marking this up.

Altitude 120 V ac mains supply voltage ........................... 5000 m 230 V ac mains supply voltage ........................... 2000 m Display ................................................................... LCD display Communications................................................... USB device port for computer control Modes of Operation .............................................. Manual and remote Power 120 Volt power outlet .......................................... 90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum 230 Volt power outlet .......................................... 180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum Weight .................................................................... 1.6 kg (3.5 lb) Size......................................................................... 28.5 cm x 17.6 cm x 8.4 cm (11.2 in x 6.9 in x 3.3 in) Safety Standards nd

CE ....................................................................... IEC/EN61010-1 2 Edition; Pollution degree 2 CSA..................................................................... CAN/CSA-C22.2 No 61010-1; UL61010-1 Electromagnetic Compatibility Standards (EMC) European EMC ................................................... EN61326-1

Detailed Specifications Test Standard Selections........................................ ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551 Voltage Ranges (Mains voltage) ...................................... 90.0 to 132.0 V ac rms 180.0 to 264.0 V ac rms Range (Point-to-point voltage)............................ 0.0 to 300.0 V ac rms Accuracy ............................................................. ±(2 % of reading + 0.2 V) Earth Resistance Modes ................................................................. Two terminal Test Current ........................................................ >200 mA ac Range ................................................................. 0.000 to 2.000 Ω Accuracy ............................................................. ±(2 % of reading + 0.015 Ω) Equipment Current Range ................................................................. 0.0 to 20.0 A ac rms Accuracy ............................................................. ±(5 % of reading + (2 counts or 0.2 A, whichever is greater)) Duty cycle ........................................................... 15 A to 20 A, 5 min. on/5 min. off 10 A to 15 A, 7 min. on/3 min. off 0 A to 10 A, continuous

1-5

ESA612 Service Manual

Leakage Current Modes*................................................................ AC+DC (True-rms) AC only DC only * Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP that are available in True-rms (shown as AC+DC) Patient Load Selection........................................ AAMI ES1-1993 Fig. 1 IEC 60601: Fig. 15 Crest factor ......................................................... ≤3 Ranges ............................................................... 0.0 to 199.9 μA 200 to 1999 μA 2.00 to 10.00 mA Accuracy DC to 1 kHz .................................................... ±(1 % of reading + (1 μA or 1 LSD, whichever is greater)) 1 to 100 kHz.................................................... ±(2 % of reading + (1 μA or 1 LSD, whichever is greater)) 1 to 5 kHz (current > 1.6 mA).......................... ±(4 % of reading + (1 μA or 1 LSD, whichever is greater)) 100 kHz to 1 MHz ........................................... ±(5 % of reading + (1 μA or 1 LSD, whichever is greater))

Note Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are +(2.5 μA or 1 LSD, whichever is greater). Mains on applied part test voltage ...................... 100 % ±7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI 100 % ±7 % of Mains for IEC 62353 current limited to 3.5 mA ±25 % per IEC 62353 100 % ±7 % of Mains for IEC 60601-1 current limited to 7.5 mA ±25 % per IEC 60601-1 Differential leakage Ranges ............................................................... 75 to 199 μA 200 to 2000 μA 2.00 to 20.00 mA Accuracy ............................................................. ±(10 % of reading + (2 counts or 20 μA, whichever is greater)) Insulation resistance Ranges ............................................................... 0.5 to 20.0 MΩ 20.0 to 100.0 MΩ Accuracy 20 MΩ Range ................................................. ±(2 % of reading + 0.2 MΩ) 100 MΩ Range ............................................... ±(7.5 % of reading + 0.2 MΩ) Source test voltage ............................................. 500 V dc (+20 %, -0 %) 2.0 ±0.25 mA short-circuit current or 250 V dc selectable Maximum load capacitance ................................ 1 μF ECG Performance Waveforms Accuracy Frequency....................................................... ±2 % Amplitude........................................................ ±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration Waveforms ECG Complex................................................. 30, 60, 120, 180, and 240 BPM Ventricular Fibrillation Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz Triangle wave ................................................. 2 Hz Pulse (63 ms pulse width)............................... 30 BPM and 60 BPM

1-6

Chapter 2

Theory of Operation

Title Introduction.......................................................................................................... Block Diagram Description ................................................................................. A1 Power PCA ................................................................................................ A2 Meter and User Interface PCA .................................................................. Block Diagram to Schematic Cross Reference .................................................... A1 Power PCA Schematic............................................................................... A2 Meter and User Interface PCA Schematic................................................. Detailed Circuit Description ................................................................................ A1 Power PCA ................................................................................................ Equipment Outlet Relays ............................................................................ Equipment Outlet Current Measurements................................................... Mains Measurement .................................................................................... Isolated Mains Voltage (Mains on Applied Parts or MAP) ........................ Connectors .................................................................................................. Isolation....................................................................................................... Relay Drivers .............................................................................................. Isolated Signals ........................................................................................... Power Supplies............................................................................................ Resistance Current Source .......................................................................... Insulation Resistance Test Voltage ............................................................. A2 Meter and User Interface PCA Schematic................................................. Meter Inputs ................................................................................................ Patient Load ................................................................................................ Overvoltage Protection................................................................................ Mains on Applied Parts Compensation....................................................... AC Only ...................................................................................................... DC Only ...................................................................................................... Resistance Measurement ............................................................................. Relay Drivers .............................................................................................. Meter Power Supplies ................................................................................. Relay Disable .............................................................................................. Applied Parts............................................................................................... Meter Microprocessor ................................................................................. User Interface Microprocessor .................................................................... Keyboard and Display................................................................................. USB Port ..................................................................................................... U/I Power Supplies .....................................................................................

Page 2-3 2-3 2-3 2-4 2-5 2-9 2-10 2-11 2-11 2-11 2-11 2-11 2-12 2-12 2-12 2-12 2-12 2-12 2-12 2-13 2-13 2-13 2-13 2-13 2-13 2-14 2-14 2-14 2-14 2-14 2-14 2-14 2-15 2-15 2-15 2-15 2-15 2-1

ESA612 Service Manual

2-2

Theory of Operation Introduction

2

Introduction WX Warning To avoid electric shock, personal injury, or death, use extreme caution when servicing the Analyzer. The Analyzer can generate high voltages and currents. The Analyzer tests electrical safety of medical equipment by making the following measurements: •

Resistance from the case of the instrument under test case to the power cord earth connection.



Leakage current to the case of the instrument under test. This and all other leakage currents are measured across one of two standardized patient loads that simulate a human body



Leakage current to earth



Leakage current to patient electrical connections



Leakage current to patient electrical connections when mains voltage is applied to the connections



Insulation resistance from mains supply to various instrument under test points



Although not a measure of electrical safety, mains voltage and equipment current are also measured



These tests are preformed in various configurations as specified by different regulatory agencies. See the ESA612 Users Manual for more information of the agency standards.

Block Diagram Description The Block Diagram shown in Figure 2-1 shows the two printed circuit assemblies (PCAs) and the interconnection between them and the Analyzer connections. A unique number in the upper-left corner of each block in the diagram references the Analyzer assemblies as follows: C = Chassis, P = A1 Power pca, M = A2 Meter and User Interface (Meter) pca. These numbers do not directly correlate to reference designators on the schematics. A1 Power PCA The mains connection is supplied through the Power Entry Module (C5), a 14-gauge cable (C4), a connector (P23), and the fuses (P16). The main power supply is a switching power supply module (P13). It is powered directly from mains power and operates from 90 to 264 V ac. The Analyzer has two circuit sections galvanically isolated from each other; one is earth referenced and is used mainly for the Equipment Outlet circuitry, and the other is used for the Meter circuits. The Earthed +12 V Supply [P13] drives the Isolated ±5 V Regulator (Reg) (P8), the Isolated +3.3 V Reg (P9), the Earthed +3.3 V Reg (P10) and is routed to the Meter PCA at connector (P1). The Isolated ±5 V Reg (P8) supplies power to the 250/500 V Reg (P3) and the 0.2 A Current Source (P4). The 250/500 V Reg is used for insulation resistance tests, which are measured by placing the voltage across the device and measuring the resultant current to calculate the resistance in the range of 0.5 - 100.0 MΩ. The 0.2 A is used for low resistance tests. The current is routed through the device to be measured and the resultant voltage is measured and used to calculate the resistance in the range of 0.0 – 2.0 Ω. The Mains on applied part transformer (P14, MAP XFORMER) has a 1:1 ratio to generate an isolated mains voltage. MAP Current Limit (P11) sets the MAP current limit to 1 mA, 3.5 mA, or 7.5 mA for the various regulatory standards. 2-3

ESA612 Service Manual

The Mains Meter (P15) attenuates mains voltage, measures it, and converts it to digital format. The digital signal is routed to the meter microprocessor (M19) on the A1 Meter pca through The Digital Isolator (P12) which provides galvanic isolation between the mains attenuator and the meter. The Test Setup Relays (P5) connect the various current sources and voltages for different test setups. The relays are controlled by relay drivers (P6), which are isolated by the digital isolators (P7). The Equipment Outlet (C7) is connected to the A1 Power pca through a 14-gauge cable (C6) and a connector (P24). Power relays are used to control the configuration of the Equipment Relay. Reverse Relays (P20) are for reversing hot and neutral. On/Off Relays (C19) are for opening hot, neutral or earth. There is a solid-state relay (P18), which is turned off before changing the state of the other equipment outlet relays and then turned on again to prevent arcing on loaded relays. The other set of reverse relays (P17) are used to reverse the polarity of incoming power if the Analyzer senses mains are reversed at the main input, thus providing extra relay protection by insuring that hot is routed through the solid state relay. Two feed-through transformers are connected in front of the equipment outlet to monitor current drawn by the device under test. The Differential Current Transformer (P22, DIFF CUR XFORMER) monitors the difference in current between hot and neutral for the differential current test and for ground fault current detection. These signals are routed to the test setup relays (P5). There are two cables between the two pcas, a 24-gauge cable (C2) between the connector (P1) and the connector (M34) for power signals and a Ribbon Cable (C3) connected between connector (P2) connector (M35) for digital signals. A2 Meter and User Interface PCA Power for the A2 Meter and User Interface (Meter and U/I, or Meter) PCA is supplied by U/I Supplies (M22) and Meter Supplies (M23). These are powered from the Earthed +12 V Supply (P13) on the A1 Power PCA. The U/I microprocessor (M14) operates the various user interface components. It writes to the display driver (M9) which controls the display (C1) through connector (M7). The display contrast is set by display contrast (M10). The U/I microprocessor reads the keys (M1) and sets the warning LED (M2). In addition, it controls the Audio Indicator (M8). It communicates with the meter microprocessor (M19) using an RS232 format through the digital isolators (M18). The U/I microprocessor also communicates with the port using an RS232 format through the digital isolators (M15) and the RS232 to USB converter (M11). The meter microprocessor (M19) sets up tests by controlling all the relay drivers on both pcas (M17, M20, P6). The measured signal comes from several places: the meter jacks (M5), applied parts (AP) jacks (M4), and the equipment outlet (C7). These are routed through the test setup relays (M21) to the patient load (M32) and then to the buffer amplifier (M30, BUFFER AMP). The signal is then separated by the AC Only Filter (M26) and the dc Only Filter (M29). The ac signal is converted to a dc voltage proportional to the rms value by rms converter (M25) and converted to digital format by the multi-channel analog-to-digital converter (ADC) (M24). The meter microprocessor (M19) reads the ADC (M24), formats it for the specific measurement being taken, and sends it to the U/I microprocessor (M14). The BUFFER AMP (M30) is protected by the over-voltage detect (M31) which can open all relays through relay driver disable (M28). The ground Fault Detect (M33, GFI) receives a signal from the DIFF CUR XFORMER (P22) on the A1 Power PCA, and if the signal is over the programmable threshold, it shuts down the relays through the relay driver disable (M28). 2-4

Theory of Operation Block Diagram to Schematic Cross Reference

2

The AP jacks (M4) are connected in various combinations to measurement circuit points depending on the test being performed. This is done by the relays (M13) and relay drivers (M17). The ECG waveform from the meter microprocessor (M19) is buffered and amplified by the ECG amplifier (M16, ECG AMP) and attenuated to the low voltages required for the ECG signals by the ECG Attenuator (M12, ECG ATTEN). The null jack (M6) is connected only during the ground resistance check and is used to zero out the resistance of the test leads.

Block Diagram to Schematic Cross Reference Tables 2-1 and 2-2 refer to one of the two schematics. The number in the first column references the number of a block on the block diagram (Figure 2-1). The second column is the page number for that pca’s particular schematic. The third column references the main parts on the schematic for that circuit. Some block diagram blocks do not have parts but are included for clarity only.

2-5

ESA612 Service Manual

2-6

Electrical Safety Analyzer Schematic Diagrams

2

ESA612

A2 METER & USER INTERFACE PCA

M14

M8

AUDIO INDICATOR

M1

M2

U/I MICROPROCESSOR

KEYS

M22

M9

M18

M7

DISPLAY

C O N N

C O N N

P1 C2

24G CABLE

C O N N

P3

250/500V REG

METER SUPPIES

P4

0.2A CURRENT SOURCE

ADC

M10

USB CONN

RS232 TO USB CONV

M13

AP JACKS

RELAYS

M15

DIGITAL ISOLATORS

METER MICROPROCESSOR

DC ONLY FILTER

AC ONLY FILTER

TEST SETUP RELAYS

M30

BUFFER AMP

M16

POWER ENTRY

P17

REVERSE RELAYS

M17

RELAY DRIVERS

P11

P14

P12

P15

DIGITAL ISOLATORS

MAP XFORMER

MAINS METER

P18

P19

SOLID STATE RELAY

ON/OFF RELAYS

P6

RELAY DRIVERS

RELAY DRIVERS

M32

PATIENT LOAD

TEST SETUP RELAYS

P21

DIFF CUR XFORMER

P7 M35

NULL JACK

C5

REVERSE RELAYS

OVER VOLTAGE DETECT

M5

M6

14G CABLE

P20 M31

ECG AMP

M21

METER JACKS

C4

ISOLATED +3.3V REG

MAP CURRENT LIMIT

M29

M26

M20 M4

FUSES

C O N N

RMS CONV

MAP NULL

ECG ATTEN

P16

P9

EARTHED +3.3V REG

M27

M12

EARTHED +12V SUPPLY

P10

M25

DISPLAY CONTRAST

M11

ISOLATED +/-5V REG

P23

P13

M24

P5

M3

P8

M23

M19

DISPLAY DRIVER

M34

U/I SUPPLIES

DIGITAL ISOLATORS

WARNING LED

C1

A1 POWER PCA

M28

RELAY DRIVER DISABLE

M33

GFI DETECT

C O N N

P2 C3

RIBBON CABLE

C O N N

DIGITAL ISOLATORS

P24 P22

CUR XFORMER

C O N N

C6

14G CABLE

C7

EQUIP OUTLET

Figure 2-1. ESA612 Block Diagram

2-7

ESA612 Service Manual

2-8

Theory of Operation Block Diagram to Schematic Cross Reference

2

A1 Power PCA Schematic Table 2-1 lists the cross-references for the A1 Power PCA schematics. Table 2-1. A1 Power PCA to Schematic Cross Reference Block Number

Schematic Page

Reference Designators

P1

3

J2

P2

3

J4

P3

4

T2 U17 U24

P4

4

U23 Q4

1

K8 K9 U8 K12 K14

2

K17 K16

4

K13

P6

3

U14 U13 U16 U15 U25

P7

2

U2

P8

4

U18

P9

3

U28

P10

3

U12

P11

2

Q1 Q2

P12

3

U5

P13

4

U9

P14

2

T1

P15

1

R2-R7 R17 R18 U4 U3

P16

1

F1-F3

P17

1

K1 K2

P18

1

K3 K19

P19

1

K4 K5

P20

1

K6 K7

P21

1

L1 U1

P22

1

L3

P23

1

J5

P24

1

J1

P5

2-9

ESA612 Service Manual

A2 Meter and User Interface PCA Schematic Table 2-2 lists the cross-references for the A2 Meter and User Interface PCA Schematics. Table 2-2. A2 Meter and User Interface PCA Schematic Cross Reference Block Number

Reference Designators

M1

7

S3 S4

M2

6

DS1 Q1

M3

8

J3

M4

4

J60-J64

M5

1

P12 P13

M6

1

P24

M7

7

J1

M8

6

LS1 Q2

M9

7

U3

M10

7

U6 U5

M11

8

U7

M12

4

R53-R61

M13

4

K1-K10 K12 K22 K23

M14

6

U2

M15

8

U8

M16

4

U13

M17

4

U32 U33 U34 U35

M18

5

U42

M19

5

U40

M20

3

U35 U36 U55 U56 U38

1

K13 K15-K19 K21U27

2

U25 U39

M22

8

U11 U10

M23

3

U48-U52

M24

2

U24

M25

2

U23

M26

2

U28 U26

M27

1

U21 U22

M28

3

U47 U49 U14

M29

2

U29 U31

M30

2

U54

M21

2-10

Schematic Page

Theory of Operation Detailed Circuit Description

2

Table 2-2. A2 Meter and User Interface PCA Schematic Cross Reference (cont.) Block Number

Schematic Page

Reference Designators

M31

1

U18

M32

1

R78 R79 C100 R156

M33

3

U16 U17

M34

4

J20

M35

4

J4

Detailed Circuit Description The following section describes the operation of the Analyzer circuits. A1 Power PCA The A1 Power PCA contains the power supplies, transformers, power relays, Equipment Outlet, and some of the test selection relays. Equipment Outlet Relays The circuitry along the top of the schematic on sheet 1 of the A1 Power PCA schematic controls the Equipment Outlet (EO). Mains is connected to the pca at J5. The EO is protected by fuses F1 and F2. Relays K1 and K2 reverse the polarity of mains in the event it is reversed at the input. Following this is a solid state relay, K19. This is turned off when any of the mechanical relays controlling the EO are switched, and then it is turned on again. This protects the mechanical relays from damage caused by inductive or incandescent loads or shorts. In the case of an EO short, the solid state relay can pass up to 900 amps for the time it takes the fuses (F1 and F2) to open. Relay K3 shorts out the solid state relay after all relay switching is done to keep power dissipation low. Relays K4 and K5 turn on the EO. Relays K6 and K7 reverse the polarity. EO ground is opened by relay K11. Relay K12 connects the resistance measuring, current source return to the EO ground pin for earth resistance tests. Relay K14 connects the EO ground pin to the meter for earth leakage tests. Relay K8 connects the EO hot and neutral pins together for leakage and insulation resistance tests. Relays K9 and K10 connect the shorted EO hot and neutral pins to the meter, 500 V or the MAP voltage as required. Equipment Outlet Current Measurements Transformer L1 is a differential transformer with hot and neutral running through it. The difference in current between these is detected by the transformer. U1 converts the current to a voltage that is used for differential leakage measurements. This is selected for measurement by connecting it to PWR_GRP by closing U8 pins 14 and 15. The output of U1 is also connected to GFI_SENSE for measurement by the GFI Detect circuit on the A2 Meter PCA. Transformer L3 is a current transformer that monitors the EO current for the equipment current measurement. This is selected for measurement by U8 pins 2 and 3. Mains Measurement Hot is routed through fuse F3 to protect the Analyzer’s internal circuitry from excessive current. Mains voltages are measured and used to detect open ground, reverse polarity, and out of range conditions. Resistors R2, R3, R4, and R17 attenuate the hot to ground voltage and U4B buffers it. The neutral to ground voltage is similarly attenuated by resistors R5, R6, R7, and R18 then buffered by U4A. These analog signals are converted 2-11

ESA612 Service Manual

to digital signals by the two-channel analog to digital converter (U3, ADC). This has a two-wire serial port containing a clock and data line and a third line for chip selection. See page 3 of the A1 Power PCA schematic for further routing of the digital signals to and from the ADC. U7 monitors the hot to ground and neutral to ground signals to generate a mains zero-cross signal used by the meter microprocessor. Isolated Mains Voltage (Mains on Applied Parts or MAP) Transformer T1 is a 1:1 transformer that generates the isolated mains voltage used by many of the Analyzer tests. Transistors Q1 and Q2 form an active current limit circuit. The current limit can be set to 3 different values, 1 mA, 3.5 mA, and 7.5 mA depending on the selected international standard. U10 selects 3.5 mA, U29 selects 7.5 mA. As the current increases, the voltage across the selected resistor increases until it reaches about 7.5 V. At this point, Q2 begins to conduct preventing Q1 from being turned on any further. This clamps the current, but the voltage across Q1 can continue to increase. The bridge CR12-CR15 allows the circuit to operate with both polarities of current. Relay K17 connects the output to the test circuits and relay K16 reverses the polarity of the voltage. Connectors There are two connectors between the A1 Power PCA and the A2 Meter PCA. Jack J2 is a 6-pin 22-guage wire connector for high current signals. Jack J4 is a 20-pin ribbon cable for digital signals and low-current power supplies. Isolation To maintain galvanic isolation between the meter circuits and the earthed circuits, the digital signals are isolated by digital isolators U2 and U5. U2 isolates the relay driver signals and U5 isolates the Mains ADC signals. Optical isolators are used to isolate signals not needing the bandwidth of the digital isolators. U11 isolates the relay driver strobe and U6 isolates the mains zero cross signal. Power for the earthed side of the isolations is from the +3.3V supply U12. Power for the isolated side is from the +3.3V supply U28. Relay Drivers U14 and U16 are serial-in parallel-out shift registers. These have a three-wire serial interface with clock, data, and strobe lines. A fourth line, output enable (/OE), is used to enable or disable all the outputs and thus the relays. The outputs of the shift registers control the relay drivers U13 and U15. Isolated Signals Meter signal lines that control circuits on the isolated side are expanded by shift register U25, like the relay driver shift registers. Power Supplies The main power supply for the Analyzer is U9. This is a switching supply that can be driven from an ac source between 90 – 264 V ac and 50 - 60 Hz. The output is +12 V and is isolated from mains. U18 is an isolated dc to dc converter module with output switching ±5 V dc. It powers the resistance measuring current source and the insulation resistance voltage source. Resistance Current Source Resistance is measured by applying a current to the resistance to be measured and measuring the voltage developed across it. The Analyzer uses a 200 mA ac source. U23 and Q4 form a precision current source. The current is set to 400 mA by the 1 Ω resistance made up of R85, R86, R76, and R59. U23 sets the voltage across the resistors equal to the reference voltage at pin 3. The reference voltage is from U22, a 1.2 V 2-12

Theory of Operation Detailed Circuit Description

2

reference, divided down to 0.4 V by R77 and R78. Q4 provides the high output impedance needed for a true current source. Isolator U19 is switched on and off to turn the current source on and off to make the 200 mA ac current. Insulation Resistance Test Voltage The 500 V dc voltage source uses a transformer to boost the voltage. A pulse width modulator drives the primary of transformer T2 to set the secondary voltage. CR5 and CR10 rectify the secondary and C29 and C42 filter it, resulting in 500 V dc. The voltage is attenuated to 5 V by resistors R41, R56, R57, and R46. U17 compares this to its 5 V reference voltage and sets the width of the output pulses to maintain the divided voltage at 5 V. The voltage can be set to 250 V by turning off Q5 and turning on Q6. This changes the attenuator such that the output voltage is 250 V. U24 senses the return current to the transformer and shuts off the supply if it exceeds 1.5 mA. Relay K13 connects the supply for tests and when not selected, connects a resistance to ground to discharge capacitance in the device under test, R63 and R64. A2 Meter and User Interface PCA Schematic The A2 Meter and User Interface PCA contains the two microprocessors, the meter circuit, test jacks, keyboard USB port and display driver. Meter Inputs The meter input jacks are P12 and P13. These are connected for various tests by relays. Relay K13 connects the red jack to the patient load for leakage tests. Relay K15 connects the red jack to a voltage attenuator used for the Point to Point voltage. Relay K16 connects the red jack to earth through a current limiter. Relay K19 connects the red jack to the resistance current source. Relay K17 connects the black jack to the negative input of the meter, –Meter (GNDMA). Relay K21 connects the black jack to the resistance current source return. The null post, P24, is connected to –Meter (GNDMA) by relays K17 and K18 for nulling the test lead resistance for earth resistance measurements. Patient Load Leakage currents are measured across the patient load. Resistors R78 and R79 form the 1-kΩ resistance for the input of the AAMI and IEC60601 patient loads. For leakage currents greater than 2 mA, the meter input is switched from resistors R78 and R79 to divide the voltage by five. Resistor R68, capacitor C100, and resistor R156 form the rest of the patient load. Closing U27 pins 14 and 15 forms an AAMI load. Closing U27 pins 10 and 11 forms an IEC60601 load. Closing U27 pints 6 and 7 connects the load to the measurement circuits on page 2 of the meter schematic. With this open and U27 pins 2 and 3 shorted, the differential current and equipment current from the A1 Power PCA are selected. Overvoltage Protection Small over-voltages at the meter inputs are clamped by CR9, VR6, and VR7. Severe over-voltages over about 50 V are sensed through the attenuator U83, R86, and U18. This is routed to schematic page 2. Mains on Applied Parts Compensation The isolated mains voltage (MAP) used in many AP leakage tests causes a residual leakage that must be compensated for. Digital to analog converter (DAC) U21 and amplifier U22 generate an ac voltage waveform that accurately tracks the MAP voltage. This is summed with the meter input leakage current through C76. The digital value set in U21 sets the amplitude of this signal which is adjusted to cancel the residual leakage.

2-13

ESA612 Service Manual

AC Only U54 buffers the signal from the meter input circuitry. In the ac only mode, the signal is routed through U39 pins 10 and 11 to the ranging amplifiers U28 and U26. The gain is selectable to be 0.15, 1.5 or 15 by the state of analog switches U25. This signal is then AC coupled by C77 to the RMS converter U23. This generates a dc voltage equivalent to the rms value of the input signal. U31A forms a two-pole-low-pass filter to smooth out the signal from U23. Then the analog to digital converter (ADC) U24 converts this to digital format which is read serially by the meter microprocessor. DC Only The signal from buffer U54 is routed through U39 pins 14 and 15 for dc only mode. U29 and U31B form a five-pole low-pass filter with a cutoff frequency of 5 Hz. This removes all mains and higher frequencies. The resultant dc is then routed to the ADC. Resistance Measurement The current source used for resistance measurement is on the A1 Power PCA and is described above. U15 inverts the signal and U39 pins 14 and 15 and U39 pins 2 and 3 are toggled to rectify it. The signal at U46 pins 12 and 13 blanks both polarities for a period after the current is switched to eliminate inductive and capacitive effects. Relay Drivers U37, U36, U55, and U56 drive the relays on the A2 Meter PCA. The circuit is similar to the relay drivers section described in the A1 Power PCA section. Meter Power Supplies The meter is galvanically isolated from the earthed circuits. The power supplies are isolated by using isolated dc to dc converter modules. U48 is an isolated dc to dc converter modules that generates the +12 V isolated supply from the earthed +12 V supply for the relays. U51 is another dc to dc converter that outputs +5 V. U50 reregulates this down to +3.3 V for the meter microprocessor. U49 outputs ±12 V used by the amplifiers and analog switches on the meter input. U53 and U52 re-regulate these to ±5 V for the remainder of the amplifiers. Relay Disable The overvoltage signal from page 1 of the meter schematic, /FAULT_SENSE, sets the RS flip-flop composed of U47A and B. This signal causes the /RELAY_EN (not relay enable) signal at U19 pin 13 to go high, shutting off all relay drivers in the instrument. To clear this fault, the meter microprocessor sets pin U47 pin 2 low to clear the flip flop. U14 is a brown-out detector which will also set the flip-flop at power up and power down to prevent the relays from being in undefined states while the microprocessor is powering up and down. The /GFI_SENSE signal from the differential transformer on the power PCA is at mains frequency analog voltage. This is rectified and filtered by CR8 and C67. DAC U16 generates a digitally controlled voltage that is compared to the GFI signal at U17. When the voltage proportional to the GFI current exceeds the preset voltage, the output of U17 goes low setting the flip-flop, U47C and D, and, in turn, disabling the relays. Applied Parts The applied parts relay drivers, U32, U33, U34, and U35, operate as described previously. Each applied part post can be connected to several circuit points: Open, ECG Wave, Earth, +Meter, -Meter or 500 V. For the V1 post this is performed by relays K1, K2, K12, and K11. The other apples parts jacks operate similarly.

2-14

Theory of Operation Detailed Circuit Description

2

Meter Microprocessor U40 is the meter microprocessor. It is a single-chip controller with built-in Program memory (FLASH), RAM, serial peripheral interfaces, and timers. Calibration data and configuration data is stored in an external serial EEPROM (U44). The processor is clocked at 12 MHz by the crystal, Y2. The two microprocessors in the Analyzer communicate over a galvanically isolated serial port with U42 providing the isolation. The U/I processor is the master and can reset the meter microprocessor through U43. U41 is a brownout detector that resets the processor on power up and power down. The JTAG port (J21) is used for debugging and downloading firmware. User Interface Microprocessor U2 is the user interface microprocessor. It contains program memory (FLASH), RAM, timers, I/O ports, and communication ports. It is clocked at 12 MHz by the crystal, Y1. It runs off of the +3.3 V power supply and an internally generated +1.8 V supply. U1 is a serial FLASH memory for storing test data. U59 is an FRAM for temporary test data storage. Only entire test records are transferred to the FLASH memory. The warning LED DS1 is turned on by the U/I microprocessor to indicate when there are dangerous voltages or currents on any test jacks. The beeper, LS1, beeps as required. The internal temperature of the unit is monitored by RT1. At a preset temperature, the U/I microprocessor informs the meter microprocessor to shut down the relays and a message is placed on the display. The JTAG port (J2) is used for debugging and downloading code into FLASH memory. Keyboard and Display The keyboard and display are controlled by the U/I microprocessor. The Keyboard, S3 and S4, is arranged in rows and columns. It is read by consecutively setting each row line low and then reading the columns. The key pressed is at the intersection of the row and column. The display is controlled by the Display controller, U3, which is connected between the U/I microprocessor and the display at J1. Quad op-amp U6 generates the voltages needed to bias the LCD display. Switching voltage supply U5 generates the highest voltage in this string, and by varying this, the contrast is varied. This is varied by the CONTRAST signal at R22. The display can be shut off by the ENABLE_20 V signal which disconnects the +12 V supply from the +20 V supply. The display backlight can be turned on and off by Q3. USB Port U7 and U8 form the USB port. U8 provides galvanic isolation between the Analyzer and the communication port. U7 communicates with the U/I microprocessor in RS232 format. It converts incoming data from USB to RS232 and outgoing data from RS232 to USB format. U9 protects the port from ESD. The connector, U3, is a standard “type B” upstream pointing USB connector. The components on the USB side of the isolator, U8, are bus powered. U/I Power Supplies U11 is a switching regulator that generates the +3.3 V supply for the U/I microprocessor. U10 is a linear regulator that generates a 3.0 V supply for the display driver.

2-15

ESA612 Service Manual

2-16

Chapter 3

Verification and Calibration

Title Introduction.......................................................................................................... Required Equipment ............................................................................................ Analyzer Verification .......................................................................................... Outlet Tension Tests........................................................................................ Setup ................................................................................................................ Verification Procedure..................................................................................... Calibration Procedure .......................................................................................... Setup ................................................................................................................ Adjustments.....................................................................................................

Page 3-3 3-3 3-4 3-4 3-4 3-5 3-52 3-52 3-52

3-1

ESA612 Service Manual

3-2

Verification and Calibration Introduction

3

Introduction This section provides performance tests to verify the Analyzer is operating within published specifications and a calibration procedure for bringing the Analyzer within specifications. The verification procedure and, if necessary, the calibration procedure can be performed both periodically and after service or repair.

Required Equipment Table 3-1 lists the equipment required for performance testing and calibrating the Analyzer. Table 3-1. Required Equipment for Analyzer Verification Instrument Type

Recommended Model

Digital Multimeter

Fluke 8846A

Calibrator

Fluke 5520A

AC and DC Current Source

Keithley 6221

Programmable AC Power Supply

Elgar CW 1251

Personal Computer with Hyperterminal Receptacle Wiring Tester

Woodhead 1750

Oscilloscope

TEK TDS 3012B

Current Shunt

Fluke A40A, 20 A capable

Current Load Differential Amplifier

AM Systems 3000 AC-DC

Resistor Decade Box: 1.800 Ω, 1.400 Ω, and 0.450 Ω rated at 1 watt

General Resistance RDS-41A

Resistor: 110 kΩ rated at ½ watt Resistor Decade Box: 0.7 MΩ, 1.0 MΩ, 2.0 MΩ, 3.1 MΩ, 6.5 MΩ, 10 MΩ, IET Labs HRRS-B-7-1K 18 MΩ, 22 MΩ, 60 MΩ, and 100 MΩ rated at 500 V and 1.5 mA Resistor: 0.1 Ω rated at 75 watts Outlet Tension Tester

Extech 475040

USB Cable Test Leads

3-3

ESA612 Service Manual

Analyzer Verification The following sections cover the setup and procedural steps for verifying the Analyzer is operating within the product specifications. Outlet Tension Tests To test the equipment outlet for proper tension: 1. Insert the tension tester into the hot (L1) pin of the Analyzer test receptacle. Set the tension tester to measure peak ounces. 2. Zero the tension tester indicators. 3. Slowly pull the tension tester out of the receptacle. 4. Read and record the reading on the tension tester. 5. Repeat steps 1 through 4 on the neutral and ground pins of the test receptacle. Setup Perform the following steps to prepare the Analyzer for performance verification. 1. Power on all test equipment and allow a 1-hour warm up time. 2. Power on the PC and log in.

3. Connect the PC USB cable to the Analyzer. 4. Power on the Analyzer with the programmable ac power supply. 5. Establish serial communication with the Analyzer (Applicable COM port using USBView, 115200 Baud, N,8,1 bits, HW Flow control) Under File→ Properties and the Settings tab, click the ASCII Setup button, and ensure “Send line ends with line feeds” and “Echo typed characters locally” are CHECKED. 6. If the Analyzer is a non-US unit, the firmware must be switched to indicate a US version. a. Send command: REMOTE=F2810

b. Receive * c. Send command: NOMINAL?

d. Receive the current nominal voltage setting. Note this value for reconversion at the completion of verification. e. Send command: NOMINAL=115 f.

Receive *

g. Send command: LOCAL h. Receive *

3-4



Verification and Calibration Analyzer Verification

3

Note Many tests are preceded by the HIGH_RES-ON, and STD=XXX commands. These settings will not be “lost” when switching between tests, and are not required to be repeated if the Analyzer hasn’t been taken out of remote mode. They are intended to allow the procedure to be used in part or in whole for testing. Use them as needed throughout this procedure. When an MREAD is used to report a reading, the tenth reading is often specified. This is to allow settling time for the Analyzer. However, any reading after stabilization is sufficient. Verification Procedure The following steps make up the procedure for verifying the operation of the Analyzer and coincide with the verification datasheet found in Appendix A. A. Functional 1. Send command: REMOTE

2. Receive * 3. Send command: DIAG=F2810

4. Receive * 5.

Send command: IDENT

6. Verify that the Analyzer identifies itself with UI & MTR firmware revision level. Record UI & MER firmware revisions. 7. Send command: LED=ON

8. Receive * 9. Verify that the red warning LED is lit. 10. Send command: LED=OFF

11. Receive * and verify the red warning LED is off. Record PASS/FAIL. 12. Send command: BEEP

13. Receive * and verify the beeper beeps once. Record PASS/FAIL. 14. Send command: CONT=09

15. Receive * 16. Send command: CONT=07

17. Receive * 18. Send command: CONT=05

19. Receive * 20. Send command: CONT=03

21. Receive * 22. Send command: CONT=09

23. Receive * 24. Send command: CONT=11

25. Receive * 26. Send command: CONT=13

3-5

ESA612 Service Manual

27. Receive * 28. Send command: CONT=15

29. Receive * 30. Send command: CONT=09

31. Receive * 32. Verify that the contrast gets lighter for lower values of CONT and darker for higher values. Record PASS/FAIL. 33. Send command: KEY

34. Receive * a. Depress all keys on the keypad and check for the correct returned digit in Table 3-2. Table 3-2. Keypad Verification Key

Response



1



2



3



4



5



6



7



8



9



10



11



12



13



14



15



16



17



18



19



20



21

35. Press the ESC key, receive a carriage return only. Record PASS/FAIL 3-6

Verification and Calibration Analyzer Verification

3

36. Send command: PON 37. Receive * 38. Send command: POFF 39. Receive * 40. Verify that all pixels on the display came on and then went off. Record PASS/FAIL. 41. Send command: LITE=OFF 42. Receive * 43. Send command: LITE=ON 44. Receive * 45. Verify that the backlight goes off and comes back on. Record PASS/FAIL 46. Send command: EXIT 47. Receive * 48. Send command: LOCAL 49. Receive * 50. Send command: REMOTE 51. Receive * B. Electrical Outlet Polarity Testing 1. Connect the receptacle wiring tester to the equipment outlet (EO) as shown in Figure 3-1.

Fluke Biomedical ESA612

Receptacle Wiring Tester

Figure 3-1. Electrical Equipment Outlet Polarity Test

2. Send command: ENCL

gfi102.eps

3-7

ESA612 Service Manual

3. Receive * 4. Send command: POL=N

5. Receive * 6. Send command: NEUT=C

7. Receive * 8. Send command: EARTH=C

9. Receive * 10. Verify the yellow and white lights are lit. Record PASS/FAIL. 11. Send command: EARTH=O

12. Receive * 13. Verify the yellow light is lit. Record PASS/FAIL. 14. Send command: NEUT=O

15. Receive * 16. Send command: EARTH=C

17. Receive * 18. Verify the white light is lit. Record PASS/FAIL. 19. Send command: LOCAL

20. Receive * 21. Send command: REMOTE

22. Receive * 23. Send command: ENCL

24. Receive * 25. Send command: POL=R

26. Receive * 27. Verify the red light is lit. Record PASS/FAIL. 28. Send command: NEUT=C

29. Receive * 30. Verify the red and yellow lights are lit. Record PASS/FAIL. 31. Send command: EARTH=O

32. Receive * 33. Verify the yellow light is lit. Record PASS/FAIL. 34. Send command: LOCAL

35. Receive * 36. Send command: REMOTE

37. Receive * 38. Remove all testing connections from the UUT. 3-8

Verification and Calibration Analyzer Verification

3

C. Electrical Outlet Ground to NULL Jack Resistance 1. Connect the DMM Input (+) to the NULL Jack and the Input (-) to the EO Ground as shown in Figure 3-2. Select the 1 GΩ range on the DMM. Fluke 8846A INPUT

Fluke Biomedical ESA612

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

gfi103.eps

Figure 3-2. Electrical Outlet Ground to NULL Jack Resistance

2. Send command: CAL=F2810

3. Receive * 4. Send command: EOGNULL=O

5. Receive * 6. If the DMM reads OPEN, record PASS. Otherwise, record FAIL. 7. Change the DMM to the 10 Ω range. Send command: EOGNULL=C

8. Read and record the DMM resistance value. 9. Send command: EXIT

10. Receive * 11. Disconnect all testing connections to the UUT. D. Mains Voltage 1. Send command: HIGH_RES=ON

2. Receive * 3. Send command: STD=AAMI

4. Receive * 5. Send command: MAINS=L1-L2

6. Receive * 7. Send command: READ

8. Receive and record response. 9. Send command: MAINS=L1-GND

3-9

ESA612 Service Manual

10. Receive * 11. Send command: READ

12. Receive and record response. 13. Remove all testing connections from the UUT. E. GFI Verification 1. Connect the current source between EO Neutral and EO ground as shown in Figure 3-3. Set the current source for 4.5 mA (6.364 mA p-p), 60 Hz with a compliance voltage of 40 V.

Fluke Biomedical ESA612

Keithly 6221

Figure 3-3. GFI Verification Connections

2. command: PPV

gjc141.eps

3. Receive * 4. Send command: GFI=5MA

5. Receive * 6. Send command: EQCURR

7. Receive * 8. Turn on the current source output. 9. Send command: STAT3

10. Receive status word 3. The 4xxx bits will not be set if the Analyzer passes the test. 3-10

Verification and Calibration Analyzer Verification

3

Note If the 8xxx and 4xxx bits are set, the status word returned will be C000. This still indicates the 4xxx bit is set, and is a failing condition. 11. Record PASS/FAIL. 12. Turn off the current source output. 13. Send command: GFIR

14. Receive * 15. Set the current source to 5.5 mA (7.778 mA p-p), 60 Hz. 16. Send command: GFI=5MA

17. Receive * 18. Send command: EQCURR

19. Receive * 20. Turn on the current source output. 21. Send command: STAT3

22. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 23. Record PASS/FAIL. 24. Turn off the current source output. 25. Send command: GFIR

26. Receive * 27. Set the current source to 9 mA (12.727 mA p-p), 60 Hz, with a 40 V compliance voltage. 28. Send command: PPV

29. Receive * 30. Send command: GFI=10MA

31. Receive * 32. Send command: EQCURR

33. Receive * 34. Turn on the current source output. 35. Send command: STAT3

36. Receive status word 3. The bit xxx8 shall be set, indicating medium GFI level, but the 4xxx bit will not be set if the Analyzer passes this test. 37. Turn off the current source output. 38. Record PASS/FAIL. 39. Send command: GFIR

40. Receive * 3-11

ESA612 Service Manual

41. Set the current source to 11 mA (15.556 mA p-p), 60 Hz. 42. Send command: GFI=10MA

43. Receive * 44. Send command: EQCURR

45. Receive * 46. Turn on the current source output. 47. Send command: STAT3

48. Receive status word 3. The 4xx8 bit will be set if the Analyzer passes this test (GFI tripped). 49. Turn off the current source output. 50. Record PASS/FAIL. 51. Send command: GFIR

52. Receive * 53. Set the current source to 22.5 mA (31.82 mA p-p), 60 Hz. 54. Send command: PPV

55. Receive * 56. Send command: GFI=25MA

57. Receive * 58. Send command: EQCURR

59. Receive * 60. Turn on the current source output. 61. Send command: STAT3

62. Receive status word 3. The 4xxx bit will not be set if the Analyzer passes this test. 63. Turn off the current source output. 64. Record PASS/FAIL. 65. Send command: GFIR

66. Receive * 67. Set the current source to 27.5 mA (38.89 mA p-p), 60 Hz. 68. Send command: GFI=25MA

69. Receive * 70. Send command: EQCURR

71. Receive * 72. Turn on the current source output. 73. Send command: STAT3

3-12

Verification and Calibration Analyzer Verification

3

74. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 75. Turn off the current source output. 76. Record PASS/FAIL. 77. Remove all testing connections. 78. Send command: GFIR

79. Receive * F. Point to Point Voltage 1. Send command: HIGH_RES=ON

2. Receive * 3. Send command: STD=AAMI

4. Receive * 5. Send command: PPV

6. Receive * 7. Connect the Normal Outputs of the calibrator to the red and black jacks on the Analyzer as shown in Figure 3-4. Set the calibrator to 1 V, 120 Hz, and set the calibrator to OPR. Fluke 5520A 5520A CALIBRATOR

Fluke Biomedical ESA612 NORMAL V, , ,RTD

AUX

SCOPE

A, -SENSE, AUX V

OUT HI

LO

TRIG

20A

GUARD

20V PK MAX

TC

20V PK MAX

Figure 3-4. Point to Point Voltage Verification Connections

8. Send command: READ

gjc105.eps

9. Receive and record response. 10. Set the calibrator to 250 V, 120 Hz. 3-13

ESA612 Service Manual

11. Send command: READ

12. Receive and record response. 13. Set the calibrator to 100 V, 633 Hz. 14. Send command: READ

15. Receive and record response. Based upon the design of the patient load, this value is expected to have a SQRT(2)/2 drop from the source voltage. 16. Set the calibrator to 4 V, 120 Hz. 17. Send command: READ

18. Receive and record response. 19. Set the calibrator to 8 V, 120 Hz. 20. Send command: READ

21. Receive and record response. 22. Set the calibrator to 10 V, 120 Hz. 23. Send command: READ

24. Receive and record response. 25. Set the calibrator to 25 V, 120 Hz. 26. Send command: READ

27. Receive and record response. 28. Set the calibrator to 40 V, 120 Hz. 29. Send command: READ

30. Receive and record response. 31. Set the calibrator to 80 V, 120 Hz. 32. Send command: READ

33. Receive and record response. 34. Set the calibrator to 130 V, 120 Hz. 35. Send command: READ

36. Receive and record response. 37. Set the calibrator to 240 V, 120 Hz. 38. Send command: READ

39. Receive and record response. 40. Turn the calibrator output to STBY. Disconnect all testing connections to the UUT. G. Resistance 1. Connect two test leads to the red and black jacks. Short these two leads together as shown in Figure 3-5.

3-14

Verification and Calibration Analyzer Verification

3

Stackable Banana Plugs

Fluke Biomedical ESA612

Figure 3-5. Point to Point Resistance Verification Connections

2. Send command: HIGH_RES=ON

gfi108.eps

3. Receive * 4. Send command: STD=AAMI

5. Receive * 6. Send command: PPR

7. Receive * 8. Send command: ZERO

9. Receive * 10. Send command: READ

11. Receive and record response. 12. Connect the two leads across a 1.800 Ω resistor. 13. Send command: READ

14. Receive and record response. Compare with the actual value of the resistor. 15. Move the lead from the black jack to the equipment outlet ground. Short the two leads together again. 16. Send command: ERES

17. Receive * 18. Send command: ZERO

19. Receive * 20. Send command: READ

3-15

ESA612 Service Manual

21. Receive and record response. 22. Connect these same two leads across a 1.800 Ω resistor. 23. Send command: READ

24. Receive and record response. Compare to the actual value of the 1.800 Ω resistor. 25. Remove all testing connections to the UUT. H. Insulation Resistance 1. Connect the EO ground to the Input (-) of the DMM and the EO Neutral to the Input (+) as shown in Figure 3-6. Set the DMM for 1000 V ac mode. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

Fluke Biomedical ESA612

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

Figure 3-6. Insulation Resistance Verification Connections

2. Send command: HIGH_RES=ON

FRONT

gjc109.eps

3. Receive * 4. Send command: STD=AAMI

5. Receive * 6. Send command: MINS

7. Receive * 8. Send command: INS=LOW

9. Receive * 10. Send command: MREAD

11. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 12. Press the ESC key. The Analyzer will stop reporting readings. 13. Send command: INS=HIGH

14. Receive * 15. Send command: MREAD 3-16

Verification and Calibration Analyzer Verification

3

16. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 17. Press the ESC key. The UUT will stop reporting readings. 18. Connect the EO ground to one side of a 110 kΩ resistor. Connect the other side to the Input (-) of the DMM. Connect the 400 mA jack of the DMM to the EO Neutral. See Figure 3-7. Set the DMM to 10 mA ac mode. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

Fluke Biomedical ESA612

HI

HI

1000 V CAT I 600V CAT II

110 kΩ Resistor

300V

LO

LO 1V FUSED

100 mA

10A REAR

Figure 3-7. Insulation Resistance with 110 kΩ Resistor Connections

19. Send command: MREAD

FRONT

gfi110.eps

20. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Remove all testing connections to the Analyzer.

3-17

ESA612 Service Manual

Mains to PE 23. Connect the EO Neutral and the EO ground across a 10.0-MΩ resistor as shown in Figure 3-8.

10 MΩ Resistor

Fluke Biomedical ESA612

Figure 3-8. Mains to PE Verification Connections

24. Send command: HIGH_RES=ON

25. Receive * 26. Send command: STD=AAMI

27. Receive * 28. Send command: MINS

29. Receive * 30. Send command: INS=HIGH

31. Receive * 32. Send command: MREAD

33. Following the tenth reading, press the ESC key. Record the tenth reading.

3-18

gfi111.eps

Verification and Calibration Analyzer Verification

3

AP to PE 34. Disconnect the lead from the EO Neutral and connect it to the RA Applied Part. The other lead should remain connected to the EO ground. Both leads should still be attached to the 10-MΩ resistor. See Figure 3-9.

Fluke Biomedical ESA612

RA

10 MΩ Resistor

Figure 3-9. AP to PE Verification Connections

35. Send command: HIGH_RES=ON

gjc135.eps

36. Receive * 37. Send command: STD=AAMI

38. Receive * 39. Send command: APINS

40. Receive * 41. Send command: INS=HIGH

42. Receive * 43. Send command: MREAD

44. Following the tenth reading, press the ESC key. Record the tenth reading.

3-19

ESA612 Service Manual

AP to NE 45. With one side of the 10 MΩ resistor still connected to RA, connect the other side to the red jack as shown in Figure 3-10.

10 MΩ Resistor

Fluke Biomedical ESA612

RA

Figure 3-10. AP to NE Verification Connections

46. Send command: HIGH_RES=ON

47. Receive * 48. Send command: STD=AAMI

49. Receive * 50. Send command: INSD

51. Receive * 52. Send command: INS=HIGH

53. Receive * 54. Send command: MREAD

55. Following the tenth reading, press the ESC key. Record the tenth reading.

3-20

gjc136.eps

Verification and Calibration Analyzer Verification

3

Mains to NE 56. Connect the red jack to one side of a 10-MΩ resistor. Connect EO Hot to the other side of the resistor as shown in Figure 3-11.

10 MΩ Resistor

Fluke Biomedical ESA612

Figure 3-11. Mains to NE Verification Connections

57. Send command: HIGH_RES=ON

gjc114.eps

58. Receive * 59. Send command: STD=AAMI

60. Receive * 61. Send command: INSB

62. Receive * 63. Send command: INS=HIGH

64. Receive * 65. Send command: MREAD

66. Following the tenth reading, press the ESC key. Record the tenth reading.

3-21

ESA612 Service Manual

Mains to AP 67. With the EO Hot still connected to one side of a 10 MΩ resistor, connect the other side of the resistor to the RA Applied Part as shown in Figure 3-12.

10 MΩ Resistor

Fluke Biomedical ESA612

RA

Figure 3-12. Mains to AP Verification Connections

68. Send command: HIGH_RES=ON

gjc115.eps

69. Receive * 70. Send command: STD=AAMI

71. Receive * 72. Send command: INSE

73. Receive * 74. Send command: INS=HIGH

75. Receive * 76. Send command: MREAD

77. Following the tenth reading, press the ESC key. Record the tenth reading. 78. Repeat steps 76-77 for each of the other Applied Parts: LL, LA, RL, and V1. 79. Change the 10 MΩ resistor for a 0.7 M resistor, while still connected to V1 and EO hot. 80. Send command: INS=LOW

81. Receive * 82. Send command: MREAD

83. Following the tenth reading, press the ESC key. Record the tenth reading. 84. Send command: INS=HIGH 85. Receive * 3-22

Verification and Calibration Analyzer Verification

86. Send command: MREAD

3

87. Following the tenth reading, press the ESC key. Record the tenth reading. 88. Repeat steps 80-87 for each resistance value of 1.0 MΩ, 6.5 MΩ, 18 MΩ, 22 MΩ, 60 MΩ, and 100 MΩ. 89. Disconnect all testing connections from the Analyzer. I.

DC Leakage 1. Connect the red and black jacks to the Input (+) and Input (-) jacks of the DMM and set the DMM for 1-kΩ range as shown in Figure 3-13.

Fluke 8846A INPUT

Fluke Biomedical ESA612

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Figure 3-13. DC Leakage Verification Connections

2. Send command: HIGH_RES=ON

gjc137.eps

3. Receive * 4. Send command: STD=AAMI

5. Receive * 6. Send command: PPL

7. Receive * 8. Send command: MODE=ACDC

9. Receive * 10. Read and record the load value from the DMM. 11. Remove all testing connections from the Analyzer.

3-23

ESA612 Service Manual

12. Connect the current source to the red and black jacks as shown in Figure 3-14. Set the current source for 10 μA dc. Keithly 6221

Fluke Biomedical ESA612

Figure 3-14. DC Leakage Current Verification Connection

13. Send command: HIGH_RES=ON

14. Receive * 15. Send command: STD=AAMI

16. Receive * 17. Send command: PPL

18. Receive * 19. Send command: MODE=DC

20. Receive * 21. Turn on the current source output. 22. Send command: MREAD

23. Receive multiple readings. Record the tenth reading. 24. Press the ESC key. The UUT will stop reporting readings. 25. Adjust the current source for 50 μA dc. 26. Send command: MREAD

27. Receive multiple readings. Record the tenth reading. 28. Press the ESC key. The UUT will stop reporting readings. 3-24

gjc117.eps

Verification and Calibration Analyzer Verification

3

29. Adjust the current source for 100 μA dc. 30. Send command: MREAD

31. Receive multiple readings. Record the tenth reading. 32. Press the ESC key. The UUT will stop reporting readings. 33. Adjust the current source for 160 μA dc. 34. Send command: MREAD

35. Receive multiple readings. Record the tenth reading. 36. Press the ESC key. The UUT will stop reporting readings. 37. Adjust the current source for 0.340 mA dc. 38. Send command: MREAD

39. Receive multiple readings. Record the tenth reading. 40. Press the ESC key. The UUT will stop reporting readings. 41. Adjust the current source for 0.500 mA dc. 42. Send command: MREAD

43. Receive multiple readings. Record the tenth reading. 44. Press the ESC key. The UUT will stop reporting readings. 45. Adjust the current source for 1.000 mA dc. 46. Send command: MREAD

47. Receive multiple readings. Record the tenth reading. 48. Press the ESC key. The UUT will stop reporting readings. 49. Adjust the current source for 1.600 mA dc. 50. Send command: MREAD

51. Receive multiple readings. Record the tenth reading. 52. Press the ESC key. The UUT will stop reporting readings. 53. Adjust the current source for 3.400 mA dc. 54. Send command: MREAD

55. Receive multiple readings. Record the tenth reading. 56. Press the ESC key. The UUT will stop reporting readings. 57. Adjust the current source for 5.000 mA dc. 58. Send command: MREAD

59. Receive multiple readings. Record the tenth reading. 60. Press the ESC key. The UUT will stop reporting readings. 61. Adjust the current source for 7.000 mA dc. 62. Send command: MREAD

63. Receive multiple readings. Record the tenth reading. 3-25

ESA612 Service Manual

64. Press the ESC key. The UUT will stop reporting readings. 65. Turn off the current source output and disconnect all testing connections from the Analyzer. J. AC Leakage Filter Frequency Response 1. Connect the current source to the red jack and the DMM 400 mA current input jack. Connect the DMM Input (-) to the black jack. See Figure 3-15. Set the current source to 1.000 mA (1.4142 mA p-p), 60 Hz, with a 25 V compliance voltage, and the DMM to the 1 mA ac range. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

Figure 3-15. AC Leakage Filter Frequency Response Connections

2. Send command: PPL

gjc126.eps

3. Receive * 4. Send command: LOAD=AAMI

5. Receive * 6. Turn on the current source output. 7. Send command: READ

8. Receive and record response of the Analyzer as the Analyzer current. Record the DMM current as the measured current.

3-26

Verification and Calibration Analyzer Verification

3

9. Calculate the measured transfer ratio: divide the measured current by the Analyzer current. Record this transfer ratio, and compare to the design transfer ratio. TRmeas =

I meas Analyzer Current

For this frequency, the design transfer ratio is 0.9980 10. Turn off the current source. 11. Calculate the transfer ratio limits: multiply the design transfer ratio by the tolerance. Add this value to the result of the offset current divided by the measured current. Record this calculated transfer ratio limit. Add and subtract this limit to the design transfer ratio to obtain the maximum and minimum, respectively.  Offset   TRLimit = TRdesign ∗ %TOL +   I meas 

(

)

For this frequency, the tolerance is 0.5 % and the offset is 0.005 mA. 12. Verify that: TRdesign – TRLimit ≤ TRmeas ≤ TRdesign + TRlimit 13. Set the current source to 1.000 mA, 1,000 Hz, and turn on the output. 14. Send command: READ

15. Receive and record response. Turn off the current source output. 16. Perform the same calculations that were completed in steps 9 through 12 above. The design transfer ratio for this frequency is 0.6910, the tolerance is 2.0 %, and the Offset current is 0.005 mA. 17. Set the current source to 1.000 mA, 10,000 Hz and turn on the output. 18. Send command: READ 19. Receive and record response. Turn off the current source. 20. Perform the same calculations that were completed in steps 9 through 12 above. The design transfer ratio for this frequency is 0.09558, the tolerance is 2.0 %, and the Offset current is 0.005 mA. 21. Turn off the current source. Disconnect all test connections.

3-27

ESA612 Service Manual

K. Differential Leakage 1. Connect the current source to the EO Neutral and the DMM 400 mA current input jack. Connect the DMM Input (-) to the EO ground. See Figure 3-16. Set the current source to 76.00 μA (107.48 μA p-p), 100 Hz, with a 25 V compliance voltage, and the DMM to the 100 μA ac range. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

Figure 3-16. Differential Leakage Verification Connections

2. Send command: DIFF

3. Receive * 4. Send command: EARTH=C

5. Receive * 6. Send command: POL=N

7. Receive * 8. Send command: NOMINAL=OFF

9. Receive * 10. Turn on the current source output. 11. Send command: READ 3-28

gjc118.eps

Verification and Calibration Analyzer Verification

3

12. Receive and record response; compare it to the actual current received by the DMM. 13. Adjust the DMM to the 1 mA range and repeat steps 11 and 12 with the current source set to 0.16mA (0.2263 mA p-p), 0.24 mA (0.3394 mA p-p), 0.50 mA (0.7071 mA p-p), 0.76 mA (1.0748 mA p-p), 100 Hz. 14. Adjust the DMM to the 10 mA range and repeat steps 11 and 12 with the current source set to 1.6 mA (2.263 mA p-p), 2.4 mA (3.394 mA p-p), 5 mA (7.071 mA p-p), and 7.6 mA (10.748 mA p-p), 100 Hz. 15. Adjust the DMM to the 100 mA range and repeat steps 11 and 12 with the current source set to 16 mA (22.63 mA p-p), 100 Hz. 16. Turn off the current source output and disconnect all testing connections from the Analyzer. L. ECG Leakage Functionality 1. Connect the DMM Input (+) to the RA applied part. Connect the DMM Input (-) to the EO ground. See Figure 3-17. Set the DMM for 1 KΩ resistance range. During this test, the DMM will function as both a current source and an ohmmeter. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

RA

Figure 3-17. ECG Leakage Functionality Connections

2. Send command: HIGH_RES=ON

gjc139.eps

3. Receive * 4. Send command: STD=601

3-29

ESA612 Service Manual

5. Receive * 6. Send command: PAT

7. Receive * 8. Send command: NOMINAL=OFF

9. Receive * 10. Send command: MODE=DC

11. Receive * 12. Send command: AP=RA//GND

13. Receive * 14. Send command: READ

15. Receive response. The DMM is acting as a current source, and should be supplying approximately 100 μA. If the Analyzer responds accordingly, record PASS. If it responds with 0 μA, record FAIL. 16. Send command: AP=LL//OPEN

17. Receive * 18. Note the reading from the DMM. If it reads OL (open), this indicates that the RA applied part is connected to the Open Bus. Record PASS. If the DMM indicates a significant resistance or a short (0 Ω), record FAIL. 19. Send command: AP=LL//GND

20. Receive * 21. Note the reading from the DMM. If it reads 0 Ω, this indicates that the RA applied part is connected to ground. Record PASS. If the DMM indicates a significant resistance or OL (open), record FAIL. 22. Move the lead connected to Input (+) from RA to LL. Repeat steps 14 – 21 for each applied part (LL, LA, RL, and V1). 23. Remove all test connections from the DMM.

3-30

Verification and Calibration Analyzer Verification

3

M. Direct Applied Part Leakage, Alternative Applied Part Leakage: Ground to RA 1. Connect the RA Applied Part to one side of a 110 kΩ resistor. Connect the 400 mA jack of the DMM to the other side of the 110 kΩ resistor. Connect the DMM Input (-) to the EO ground. See Figure 3-18. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

Fluke Biomedical ESA612

FRONT

110 kΩ Resistor

RA

Figure 3-18. Direct Applied Part Verification Ground to RA Connections

2. Send command: NOMINAL=OFF

gjc121.eps

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: DMAP

9. Receive * 10. Send command: MAP=3.5MA

11. Receive * 12. Send command: POL=N

13. Receive * 14. Send command: AP=RA/RL/GND

15. Receive * 3-31

ESA612 Service Manual

16. Send command: MODE=AC

17. Receive * 18. Send command: MREAD

19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF

23. Receive * 24. Send command: HIGH_RES=ON

25. Receive * 26. Send command: STD=353

27. Receive * 28. Send command: SPAT

29. Receive * 30. Send command: MAP=3.5MA

31. Receive * 32. Send command: AP=ALL//

33. Receive * 34. Send command: MODE=AC

35. Receive * 36. Send command: MREAD

37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE 41. Receive *

3-32

Verification and Calibration Analyzer Verification

3

N. Direct Applied Part Leakage, Alternative Applied Part Leakage: Red to RA 1. Move the lead connected to the EO ground to the red jack; the other end should be connected to the DMM Input (-). The RA Applied Part should still be connected to one end of the 110-kΩ resistor. See Figure 3-19. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612 110 kΩ Resistor

RA

Figure 3-19. Direct Applied Part Verification Red to RA Connections

2. Send command: NOMINAL=OFF

gjc125.eps

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: DMAP

9. Receive * 10. Send command: MAP=3.5MA

11. Receive * 12. Send command: POL=N

13. Receive * 14. Send command: AP=RA/RL/GND

15. Receive * 16. Send command: MODE=AC

3-33

ESA612 Service Manual

17. Receive * 18. Send command: MREAD

19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF

23. Receive * 24. Send command: HIGH_RES=ON

25. Receive * 26. Send command: STD=353

27. Receive * 28. Send command: SPAT

29. Receive * 30. Send command: MAP=3.5MA

31. Receive * 32. Send command: AP=ALL//

33. Receive * 34. Send command: MODE=AC

35. Receive * 36. Send command: MREAD

37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE 41. Receive *

3-34

Verification and Calibration Analyzer Verification

3

O. Alternative Equipment Leakage: Red to Hot 1. Move the lead from the RA applied part to the EO hot; the other end should still be connected to the 110-kΩ resistor. The red jack should be connected to the DMM Input (-). See Figure 3-20. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO

Fluke Biomedical ESA612

1V FUSED

100 mA

10A REAR

FRONT

110 kΩ Resistor

Figure 3-20. Alternate Equipment Leakage: Red to Hot Connections

2. Send command: NOMINAL=OFF

gjc142.eps

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: SAF

9. Receive * 10. Send command: MAP=3.5MA

11. Receive * 12. Send command: AP=ALL//

13. Receive * 14. Send command: MODE=AC

15. Receive * 16. Send command: ALTEARTH=C

17. Receive * 18. Send command: MREAD

3-35

ESA612 Service Manual

19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: IDLE

23. Receive * P. Alternative Equipment Leakage, Alternative Applied Part Leakage: RA to Hot 1. Move the lead from the Red jack to the RA applied part, the other end should still be connected to the Input (-) of the DMM. The EO hot should still be connected to the 110-kΩ resistor. See Figure 3-21. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

Fluke Biomedical ESA612

FRONT

110 kΩ Resistor

RA

Figure 3-21. Alternative Applied Part Leakage: RA to Hot

2. Send command: NOMINAL=OFF

3. Receive * 4. Send command: HIGH_RES=ON 5. Receive * 6. Send command: STD=353 7. Receive * 8. Send command: SAF 3-36





gjc123.eps

Verification and Calibration Analyzer Verification

3

9. Receive * 10. Send command: MAP=3.5MA

11. Receive * 12. Send command: MODE=AC

13. Receive * 14. Send command: AP=ALL//

15. Receive * 16. Send command: ALTEARTH=C

17. Receive * 18. Send command: MREAD

19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF

23. Receive * 24. Send command: HIGH_RES=ON

25. Receive * 26. Send command: STD=353

27. Receive * 28. Send command: SPAT

29. Receive * 30. Send command: MAP=3.5MA

31. Receive * 32. Send command: AP=ALL//

33. Receive * 34. Send command: MODE=AC

35. Receive * 36. Send command: MREAD

37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE

41. Receive * 3-37

ESA612 Service Manual

Q. Alternative Applied Part Leakage: RA to Neutral 1. Move the lead from the EO hot to the EO neutral, the other end should still be connected to the 110-kΩ resistor. The RA applied part lead should still be connected to the DMM Input (-). See Figure 3-22. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

Fluke Biomedical ESA612

FRONT

110 kΩ Resistor

RA

Figure 3-22. Alternative Applied Part Leakage: RA to Neutral Connections

2. Send command: NOMINAL=OFF

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: SPAT

9. Receive * 10. Send command: AP=ALL//

11. Receive * 12. Send command: MAP=3.5MA

13. Receive * 14. Send command: MODE=AC 15. Receive * 3-38

gjc132.eps

Verification and Calibration Analyzer Verification

16. Send command: MREAD

3

17. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 18. Receive and record response; compare it to the actual current received by the DMM. 19. Press the ESC key. The Analyzer will stop reporting readings. 20. Send command: IDLE

21. Receive * 22. Disconnect all test leads from the 100 kΩ resistor. R. Patient Auxiliary Leakage: RA to RL 1. Connect the current source to the 400 mA jack of the DMM and to the RL applied part. The DMM Input (-) should still be connected to the RA applied part. See Figure 3-23. Set the DMM to the 100 μA ac range and the current source to 100.000 μA (141.42 μA p-p), 100 Hz, with a 25 V compliance voltage. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

RA

RL

Figure 3-23. Patient Auxiliary Leakage: RA to RL Connections

2. Send command: NOMINAL=OFF

gjc127.eps

3. Receive * 3-39

ESA612 Service Manual

4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=601

7. Receive * 8. Send command: AUX

9. Receive * 10. Send command: AP=RA/RL/GND

11. Receive * 12. Send command: POL=N

13. Receive * 14. Send command: MODE=AC

15. Receive * 16. Send command: NEUT=C

17. Receive * 18. Send command: EARTH=C

19. Receive * 20. Turn on the current source output. 21. Send command: READ

22. Receive and record response; compare it to the actual current received by the DMM. 23. Turn off the current source output. 24. Send command: IDLE 25. Receive *

3-40

Verification and Calibration Analyzer Verification

3

S. Direct Equipment Leakage, Patient Leakage: Ground to RA 1. Move the lead from the RL applied part to the EO ground, the other end should still be connected to the current source. The RA applied part lead should still be connected to the DMM Input (-). See Figure 3-24. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

RA

Figure 3-24. Direct Equipment Leakage, Patient Leakage: Ground to RA

2. Send command: NOMINAL=OFF

gjc131.eps

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: DIRL

9. Receive * 10. Send command: AP=RA/RL/GND

11. Receive * 12. Send command: POL=N

13. Receive * 3-41

ESA612 Service Manual

14. Send command: MODE=AC

15. Receive * 16. Send command: EARTH=C

17. Receive * 18. Turn on the current source output. 19. Send command: READ

20. Receive and record response; compare it to the actual current received by the DMM. 21. Turn off the current source output. 22. Send command: IDLE

23. Receive * 24. Send command: NOMINAL=OFF

25. Receive * 26. Send command: HIGH_RES=ON

27. Receive * 28. Send command: STD=601

29. Receive * 30. Send command: PAT

31. Receive * 32. Send command: AP=RA/RL/GND

33. Receive * 34. Send command: POL=N

35. Receive * 36. Send command: MODE=AC

37. Receive * 38. Send command: NEUT=C

39. Receive * 40. Send command: EARTH=C

41. Receive * 42. Turn on the current source output. 43. Send command: READ

44. Receive and record response; compare it to the actual current received by the DMM. 45. Turn off the current source output. 46. Send command: IDLE 47. Receive * 3-42

Verification and Calibration Analyzer Verification

3

T. Direct Equipment Leakage, Enclosure Leakage: Ground to Red 1. Move the lead from the RA applied part to the Red jack, the other end should still be connected to the DMM Input (-). The EO ground lead should still be connected to the current source. See Figure 3-25. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

Figure 3-25. Direct Equipment Leakage, Enclosure Leakage: Ground to Red Connections

2. Send command: NOMINAL=OFF

gjc129.eps

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=353

7. Receive * 8. Send command: DIRL

9. Receive * 10. Send command: AP=RA/RL/GND

11. Receive * 12. Send command: POL=N

13. Receive * 3-43

ESA612 Service Manual

14. Send command: MODE=AC

15. Receive * 16. Send command: EARTH=C

17. Receive * 18. Turn on the current source output. 19. Send command: READ

20. Receive and record response; compare it to the actual current received by the DMM. 21. Turn off the current source output. 22. Send command: IDLE

23. Receive * 24. Send command: NOMINAL=OFF

25. Receive * 26. Send command: HIGH_RES=ON

27. Receive * 28. Send command: STD=601

29. Receive * 30. Send command: ENCL

31. Receive * 32. Send command: POL=N

33. Receive * 34. Send command: MODE=ACDC

35. Receive * 36. Send command: NEUT=C

37. Receive * 38. Send command: EARTH=C

39. Receive * 40. Turn on the current source output. 41. Send command: READ

42. Receive and record response; compare it to the actual current received by the DMM. 43. Turn off the current source output. 44. Send command: IDLE

45. Receive * 46. Do not disconnect any connections.

3-44

Verification and Calibration Analyzer Verification

3

U. Filter Tests 1. Attempt to measure ac current with the dc filter: set the current source to generate 100.00 μA (141.42 μA p-p), 100 Hz, with a 25 V compliance voltage. 2. Send command: NOMINAL=OFF

3. Receive * 4. Send command: HIGH_RES=ON

5. Receive * 6. Send command: STD=601

7. Receive * 8. Send command: ENCL

9. Receive * 10. Send command: POL=N

11. Receive * 12. Send command: MODE=DC

13. Receive * 14. Send command: NEUT=C

15. Receive * 16. Send command: EARTH=C

17. Receive * 18. Turn on the current source output. 19. Send command: READ

20. Receive and record response. The result should be approximately zero. 21. Turn off the current source output. 22. Set the current source to generate 100 μA dc and turn the output on. 23. Send command: MODE=AC

24. Receive * 25. Send command: READ

26. Disconnect all testing connections from the Analyzer.

3-45

ESA612 Service Manual

V. MAP Voltage and Current Limits 1. Connect the DMM Input (+) to the RA applied part and the DMM Input (-) to the EO ground as shown in Figure 3-26. Set the DMM to the 1000 V ac range. Fluke 8846A INPUT

SENSE 4W

V

Fluke Biomedical ESA612

2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

RA

Figure 3-26. MAP Voltage and Current Limits Connections

2. Send command: DMAP

gjc140.eps

3. Receive * 4. Send command: MAP=NORM

5. Receive * 6. Send command: AP=RA/RL/GND

7. Receive * 8. Send command: MREAD

9. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 10. Press the ESC key. The Analyzer will stop reporting readings. 11. Turn off the output of the programmable ac power supply. Adjust it for 230 V, 50 Hz, and turn on the output. 12. Once the ESA612 powers on again, it will report an error code of “!53” and the display will show “Fault detected…Mains out of range,” Send command: REMOTE=F2810

13. Receive *. The WARNING LED will remain lit. 14. Send command: DMAP

15. Receive * 16. Send command: MAP=3.5MA

17. Receive * 18. Send command: AP=RA/RL/GND 3-46

Verification and Calibration Analyzer Verification

3

19. Receive * 20. Send command: MREAD

21. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 22. Press the ESC key. The Analyzer will stop reporting readings. 23. Move the lead connected to the RA applied part from the DMM Input (+) to the 400 mA jack. The DMM Input (-) should still be connected to EO ground. Set the DMM to the 10 mA range. 24. Send command: DMAP

25. Receive * 26. Send command: MAP=3.5MA

27. Receive * 28. Send command: AP=RA/RL/GND

29. Receive * 30. Send command: MREAD

31. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 32. Press the ESC key. The Analyzer will stop reporting readings. 33. Send command: DMAP

34. Receive * 35. Send command: MAP=7.5MA

36. Receive * 37. Send command: AP=RA/RL/GND

38. Receive * 39. Send command: MREAD

40. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 41. Press the ESC key. The Analyzer will stop reporting readings. 42. Turn off the output of the programmable ac power supply. Readjust it for 115 V, 60 Hz, and turn on the output. 43. Once the ESA612 powers on again, it will report the model number and firmware version. Send command: REMOTE 44. Receive * 45. Send command: DMAP

46. Receive * 47. Send command: MAP=1MA

48. Receive * 49. Send command: AP=RA/RL/GND

3-47

ESA612 Service Manual

50. Receive * 51. Send command: MREAD

52. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 53. Press the ESC key. The Analyzer will stop reporting readings. .

54. Disconnect all testing connections from the Analyzer. W. ECG Waveform 1. Connect the Input (+) of the differential amplifier to the RA applied part and the Input (-) of the differential amplifier to the RL applied part. Connect the differential amplifier output to the oscilloscope. See Figure 3-27. Oscilliscope

Differential Amplifier Chan 1

Out

Chan 2

Fluke Biomedical ESA612

RA RL

Figure 3-27. ECG Waveform Connections

2. Establish the following settings for the differential amplifier: a. GAIN = 1K b. Divide by 100 = Out c. (+) Input = DC d. (-) Input = DC 3-48

gjc128.eps

Verification and Calibration Analyzer Verification

3

e. HF -3dB = 0.1 f. LF -3db = DC OFFSET 3. Send command: STD=AAMI

4. Receive * 5. Send command: ECG

6. Receive * 7. Send command: SQ2

8. Receive * 9. Verify the waveform has a 2.000 ±0.04 Hz frequency. Record PASS/FAIL. 10. Connect the RA applied part to the DMM Input (+) and the RL applied part to the DMM Input (-) and set the DMM for 100 mV dc range. 11. Using DMM MIN MAX function, record the waveform peak to peak amplitude. 12. Move the DMM Input (+) from the RA applied part to the LL applied part. Repeat step 11 for each of the following applied parts: LL, LA, and V1. 13. Press ESC. Receive only a carriage return. 14. Send commend: EXIT

15. Receive * 16. Disconnect all testing connections.

3-49

ESA612 Service Manual

X. Equipment Current Verification 1. Connect the current shunt to the DMM using the voltage inputs. Connect the EO hot to the DMM Input (+) side of the current shunt and the EO ground to one side of the variable current load. Connect the other side of the variable current load to the current shunt jack corresponding to the Input (-) of the DMM. Set the DMM to the 10 V dc range.

Fluke Biomedical ESA612

Fluke A40A Current Shunt

Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Variable Load

Figure 3-28. Equipment Current Calibration Connection

2. Send command: STD=AAMI

gjc146.eps

3. Receive * 4. Send command: EQCURR

5. Receive * 6. Establish an 8 A draw through the current load, with the shunt resistor in series with the load. 7. Send command: READ

8. Receive * 9. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 10. Establish a 3 A draw through the current load. 11. Send command: READ

12. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 13. Establish a 1 A draw through the current load. 14. Send command: READ 3-50

Verification and Calibration Analyzer Verification

3

15. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 16. Turn off the equipment outlet. Send command: PPL

17. Receive * 18. Remove all testing connections. Y. Save Calibration Information If this is AS LEFT verification and the device has passed verification, record the new calibration date and technician information. 1. Send command: CAL=F2810

2. Receive * 3. Write new calibration date in the format of MxxDyyYzzzz, where xx stands for the two digits of the current month, yy stands for the two digits of the current day, and zzzz stands for the four digits of the current year. 4. Send command: DATE=MxxDxxYxxxx

5. Receive * 6. Write calibration technician in the format of xxxxx, where xxxxx stands for the five digits of the technician’s employee number. 7. Send command: TECH=xxxxx

8. Receive * 9. Check Analyzer serial number, cal date, and cal technician are correct. 10. Send command: INFO=

11. Receive serial number, cal date, and cal technician in response. Record response. 12. Send command: SAVE

13. Send command: EXIT

14. Receive * 15. If the unit did not come in with a nominal voltage of 115, readjust the nominal voltage according to the destination country. a. Send command: NOMINAL=xxx where xxx is the value recorded when the unit was originally converted for verification. See Table 3-3 for standard nominal values.

3-51

ESA612 Service Manual

Table 3-3. Nominal Values Country

Value

Europe

230

UK

230

Australia

240

Israel

230

France/Belgium

230

US

120

Swiss

230

Thailand

220

Japan

100

b. Receive * 16. Turn the Analyzer power off. 17. After the Analyzer has restarted, verify the device has retained its calibration date by pressing SETUP. Next press the softkey labeled More twice, and then the softkey labeled Instrument Information. Verify the calibration date, firmware version, and serial number are correct. 18. Turn the Analyzer off and remove all connections.

Calibration Procedure The following steps will bring the Analyzer into published specifications. The required equipment for this procedure is shown in Table 3-1. Setup Perform the following steps to prepare the Analyzer for calibration. 1. 2. 3. 4.

Power on all test equipment and allow a 1-hour warm up time. Power on the PC and log in. Connect the PC USB cable to the UUT. Power on the UUT. Establish serial communication with the Analyzer (Applicable COM port using USBView, 115200 Baud, N,8,1 bits, HW Flow control) Under File->Properties and the Settings tab, click the ASCII Setup button, and ensure “Send line ends with line feeds” and “Echo typed characters locally” are CHECKED. 5. Many of the calibrations require ac or dc meter calibration to be completed first. If the ac or dc meter calibration is changed, all other calibrations should at least be verified.

Adjustments The following procedure calibrates the Analyzer to the published specifications and coincides with the calibration datasheet found in Appendix A: A. AC Meter Calibration 1. Send command: REMOTE

2. Receive * 3. Send command: CAL=F2810 3-52

Verification and Calibration Calibration Procedure

3

4. Receive * 5. Send command: IDENT

6. Receive and record UI & MTR firmware revision level. 7. Send command: EXIT

8. Receive * 9. Send command: NOMINAL?

10. Receive and record the current nominal voltage. 11. If nominal is not set to 115 V, reset nominal and restart the Analyzer. a. Send command: NOMINAL=115

b. Receive * c. Send command: RSTUI

d. When the Analyzer restarts, receive the identification information. e. Send command: REMOTE

f. Receive * 12. Send command: DIAG=F2810

13. Receive * 14. Send command: XFLASHT

15. Receive and record PASS/FAIL 16. Send command: EXIT

17. Receive * 18. Send command: CAL=F2810

19. Receive * 20. Send command: DEFAULTS

21. Receive * 22. Connect the calibrator to the red and black jacks as shown in Figure 3-29. Set the calibrator for 2.000 V, 120 Hz (±200 mV).

3-53

ESA612 Service Manual

Fluke 5520A 5520A CALIBRATOR

Fluke Biomedical ESA612 NORMAL

AUX

V, , ,RTD

A, -SENSE, AUX V

SCOPE OUT

HI

LO

TRIG

20A

GUARD

20V PK MAX

TC

20V PK MAX

Figure 3-29. AC Meter Calibration Connections

23. Send command: METER_AC

24. Receive * 25. Send command: RANGE=1,0,0

26. Receive * 27. Turn on the calibrator output. 28. Send command: GAIN=2000.00

29. Receive * 30. Adjust the calibrator output to 1.000 V, 120 Hz (±100 mV). 31. Send command: GAIN=1000.00

32. Receive * 33. Adjust the calibrator output to 200 mV, 120 Hz (±20 mV). 34. Send command: RANGE=2,0,0

35. Receive * 36. Send command: GAIN=200.00

37. Receive * 38. Adjust the calibrator output to 20.00 mV, 120 Hz (±2 mV). 39. Send command: RANGE=3,0,0

40. Receive * 41. Send command: GAIN=200.00 42. Receive * 3-54

gjc105.eps

Verification and Calibration Calibration Procedure

3

43. Short the Red and Black jacks together using a test lead. 44. Send command: OFFSET

45. After a significant delay, receive * 46. Reconnect the normal outputs of the calibrator to the red and black jacks. Adjust the calibrator output to 1.500 V, 120 Hz. 47. Send command: RANGE=1,0,0

48. Receive * 49. Send command: READ

50. Receive and record response. 51. Adjust the calibrator output to 500.0 mV, 120 Hz. 52. Send command: READ

53. Receive and record response. 54. Adjust the calibrator output to 150.0 mV, 120 Hz. 55. Send command: RANGE=2,0,0

56. Receive * 57. Send command: READ

58. Receive and record response. 59. Adjust the calibrator output to 15.0 mV, 120 Hz. 60. Send command: RANGE=3,0,0

61. Receive * 62. Send command: READ

63. Receive and record response. 64. Disconnect the calibrator. B. DC Meter Calibration 1. Short the red and black jacks together using a test lead. 2. Send command: METER_DC

3. Receive * 4. Send command: OFFSET

5. Receive * 6. Connect the calibrator the red and black jacks, and adjust the calibrator output for 2000 mV dc. 7. Send command: GAIN=2000.00

8. Receive * 9. Adjust the calibrator output to 1500 mV dc. 10. Send command: READ

11. Receive and record response. 12. Disconnect the calibrator. 3-55

ESA612 Service Manual

C. Leakage Calibration (Dependent upon AC Meter) 1. Connect the current source to the red jack and the 400 mA jack of the DMM. Connect the DMM Input (-) to the black jack. See Figure 3-30. Set the current source to 1.00 mA (1.4142 mA p-p), 60 Hz with a 10 V compliance voltage and the DMM to 1 mA ac mode. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

Figure 3-30. Leakage Calibration

2. Send command: PPL

gjc126.eps

3. Receive * 4. Send command: LOAD=601

5. Receive * 6. Send command: RANGE=0,0,2

7. Receive * 8. Turn on the current source. 9. Send command: GAIN=X.XXXX where x.xxxx is the actual value of the current in mA measured by the DMM. 10. Receive * 11. Adjust the current source to 0.5 mA (0.7071 mA p-p). 12. Send command: READ

13. Receive and record response. Also record the DMM reading. 3-56

Verification and Calibration Calibration Procedure

3

14. Turn off the current source output. Adjust the current source to 5.00 mA (7.071 mA p-p), 60 Hz. Set the DMM to the 10 mA range. 15. Send command: RANGE=0,0,1

16. Receive * 17. Turn on the current source output. 18. Send command: GAIN=X.XXXX where X.XXXX is the actual value of the current in mA measured by the DMM. 19. Receive * 20. Adjust the current source to 3.0 mA (4.243 mA p-p). 21. Send command: READ

22. Receive and record response. Also record the DMM reading. 23. Turn off the current source and disconnect all testing connections. D. Point to Point Calibration (Dependent on AC Meter) 1. Connect the calibrator to the red jack and black jacks as shown in Figure 3-31. Set the calibrator to 300.0 V, 120 Hz. Fluke 5520A 5520A CALIBRATOR

Fluke Biomedical ESA612 NORMAL

AUX

V, , ,RTD

A, -SENSE, AUX V

SCOPE OUT

HI

LO

TRIG

20A

GUARD

20V PK MAX

TC

20V PK MAX

Figure 3-31. Point to Point Calibration Connections

2. Send command: PPV

gjc105.eps

3. Receive * 4. Turn on the calibrator output. 5. Send command: GAIN=300.00

6. Receive * 7. Adjust the calibrator output to 250.0 V, 120 Hz. 3-57

ESA612 Service Manual

8. Send command: READ

9. Receive and record response. 10. Turn off the calibrator output and disconnect the calibrator. E. MAP Compensation Calibration (Dependent on AC Meter) 1. Remove all connections except mains power and the USB cable. 2. Send command: COMP 3. After significant delay, receive * if the null was found. 4. If no null was found, !33 will be returned, and the device needs repair prior to calibration. F. Insulation HV Calibration (Dependent on AC Meter) 1. Connect the DMM to the RA applied part and the red jack as shown in Figure 332 and set it for the 1000 V dc range. Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

Fluke Biomedical ESA612

100 mA

10A REAR

FRONT

Figure 3-32. Insulation HV Calibration Connections

2. Send command: INSD

3. Receive * 4. Send command INS=HIGH

5. Receive * 6. Send command: MREAD

7. Receive multiple readings, all approximately 10 MΩ. While the Analyzer is conducting is read, note the stabilized voltage reading from the DMM. 8. Press ESC to stop MREAD. 9. Send command: GAIN=XXX.XXX where XXX.XXX is the actual value of the voltage measured by the DMM. 10. Receive * 11. Connect the applied part and the red jack across a 10 MΩ resistor. 3-58

gjc144.eps

Verification and Calibration Calibration Procedure

12. Send command: READ

3

13. Receive and record response. 14. Disconnect all testing connections. G. Resistance Calibration (Dependent on AC Meter) 1. Connect two leads to the red and black jacks, then short the other ends together as shown in Figure 3-33. Stackable Banana Plugs

Fluke Biomedical ESA612

Figure 3-33. Resistance Calibration Connections

2. Send command: PPR

gjc108.eps

3. Receive * 4. Send command: OFFSET

5. Receive * 6. Connect the test leads across a 4.0 Ω (3.6 – 4.4 Ω) resistor. 7. Send command: GAIN=X.XXX

where X.XXX is the actual value of the resistor.

8. Receive * 9. Exchange the 4.0 Ω resistor for a 3.0 Ω resistor. 10. Send command: READ

11. Receive and record response. 12. Disconnect all testing connections.

3-59

ESA612 Service Manual

H. Differential Leakage Calibration (Dependent on AC Meter) 1. Connect the current source to the EO neutral and the 400 mA jack of the DMM. Connect the DMM Input (-) to the EO ground. See Figure 3-34. Set the current source to 15.0 mA (21.21 mA p-p), 120 Hz with a 10 V compliance voltage and the DMM to 100 mA ac range. Fluke 8846A INPUT

Keithly 6221

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Fluke Biomedical ESA612

Figure 3-34. Differential Leakage Calibration Connections

2. Send command: DIFF

3. Receive * 4. Send command: EARTH=C

5. Receive * 6. Send command: POL=N

7. Receive * 8. Turn on the current source output. Note the reading on the DMM. 9. Send command: GAIN=XX.XXXX where XX.XXXX is the actual value of the current in mA measured by the DMM. 10. Receive * 11. Set the current source to 11 mA (15.556 mA p-p). 3-60

gjc118.eps

Verification and Calibration Calibration Procedure

12. Send command: READ

3

13. Receive and record response. Compare it to the actual value of the current. 14. Turn off the current source and disconnect all testing connections. I.

GFI Calibration 1. Connect the current source to the EO neutral and EO ground as shown in Figure 3-35. Set the current source to 5.0 mA (7.071 mA p-p), 60 Hz with a 40 V compliance voltage.

Fluke Biomedical ESA612

Keithly 6221

Figure 3-35. GFI Calibration Connections

2. Send command: PPV

gjc141.eps

3. Receive * 4. Send command: GFI=5MA

5. Receive * 6. Turn on the current source output. 7. Send command: CAL_GFI

8. Receive * 9. Turn off the current source output. 10. Set the current source to 4.5 mA (6.364 mA p-p), 60 Hz. 11. Send command: GFI=5MA

12. Receive * 3-61

ESA612 Service Manual

13. Send command: EQCURR

14. Receive * 15. Turn on the current source output. 16. Send command: STAT3

17. Receive status word 3. The 4xxx bits will not be set if the Analyzer passes the test. Note If the 1xxx and 4xxx bits are set, the status word returned will be 5000. This still indicates the 4xxx bit is set, and is a failing condition. 18. Record PASS/FAIL. 19. Turn off the current source output. 20. Send command: GFIR

21. Receive * 22. Set the current source to 5.5 mA (7.778 mA p-p), 60 Hz. 23. Send command: EQCURR

24. Receive * 25. Turn on the current source output. 26. Send command: STAT3

27. Receive status word 3. The 4xxx bits will be set if the Analyzer passes this test (GFI tripped). 28. Verify the device is in a GFI tripped state: send command: PPV

29. Receive error code !50 (GFI error). 30. Record PASS/FAIL; pass if !50 is received. 31. Turn off the current source output. 32. Send command: GFIR

33. Receive * 34. Set the current source to 10.0 mA (14.142 mA p-p), 60 Hz, with a 40 V Compliance voltage. 35. Send command: PPV

36. Receive * 37. Send command: GFI=10MA

38. Receive * 39. Turn on the current source output. 40. Send command: CAL_GFI

41. Receive * 42. Turn off the current source output. 43. Set the current source to 9 mA (12.728 mA p-p), 60 Hz. 3-62

Verification and Calibration Calibration Procedure

44. Send command: GFI=10MA

3

45. Receive * 46. Send command: EQCURR

47. Receive * 48. Turn on the current source output. 49. Send command: STAT3

50. Receive status word 3. The status bit xxx8 shall be set, indicating Medium GFI level, but the 4xxx bit will NOT be set if the Analyzer passes this test. 51. Turn off the current source output. 52. Record PASS/FAIL 53. Send command: GFIR

54. Receive * 55. Set the current source to 11 mA (15.556 mA p-p), 60 Hz. 56. Send command: GFI=10MA

57. Receive * 58. Send command: EQCURR

59. Receive * 60. Turn on the current source output. 61. Send command: STAT3

62. Receive status word 3. The 4xx8 bit will be set if the Analyzer passes this test (GFI tripped). 63. Turn off the current source output. 64. Record PASS/FAIL 65. Send command: GFIR

66. Receive * 67. Set the current source to 25.0 mA (35.36 mA p-p), 60 Hz. 68. Send command: PPV

69. Receive * 70. Send command: GFI=25MA

71. Receive * 72. Turn on the current source output. 73. Send command: CAL_GFI

74. Receive * 75. Turn off the current source output. 76. Set the current source to 22.5 mA (31.82 mA p-p), 60 Hz. 77. Send command: GFI=25MA

78. Receive * 3-63

ESA612 Service Manual

79. Send command: EQCURR

80. Receive * 81. Turn on the current source output. 82. Send command: STAT3

83. Receive status word 3. The 4xxx bit will NOT be set if the Analyzer passes this test. 84. Turn off the current source output. 85. Record PASS/FAIL 86. Send command: GFIR

87. Receive * 88. Set the current source to 27.5 mA (38.89 mA p-p), 60 Hz. 89. Send command: GFI=25MA

90. Receive * 91. Send command: EQCURR

92. Receive * 93. Turn on the current source output. 94. Send command: STAT3

95. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 96. Turn off the current source output. 97. Record PASS/FAIL 98. Remove all testing connections. 99. Send command: GFIR 100. Receive *

3-64

Verification and Calibration Calibration Procedure

3

J. Equipment Current Calibration (Dependent on AC Meter) 1. Connect the current shunt to the DMM using the voltage inputs. Connect the EO hot to the DMM Input (+) side of the current shunt and the EO ground to one side of the variable current load. Connect the other side of the variable current load to the current shunt jack corresponding to the (-) DMM Input.

Fluke Biomedical ESA612

Fluke A40A Current Shunt

Fluke 8846A INPUT

SENSE 4W

V 2W/4W

HI

HI

1000 V CAT I 600V CAT II

300V

LO

LO 1V FUSED

100 mA

10A REAR

FRONT

Variable Load

Figure 3-36. Equipment Current Calibration Connection

2. Send command: EQCURR

gjc146.eps

3. Receive * 4. Establish an 8 A draw through the current load, with the shunt resistor in series with the load. 5. Note the actual current flowing in the circuit using the current shunt and DMM. 6. Send command: GAIN=XX.XXXX in amperes.

where XX.XXXX is the actual current calculated

7. Receive * 8. Establish an 3 A draw through the current load. 9. Send command: READ

10. Receive and record response. Compare it with the actual value of the current flowing through the current shunt. 11. Turn off the equipment outlet. Send command: PPL

12. Receive * 13. Remove all testing connections. 14. Clear GFI faults. Send command: GFIR

15. Receive * 3-65

ESA612 Service Manual

K. Mains Calibration Part I Note During the mains calibration, the differential current offset is also calculated and stored. Ensure nothing is connected to the Analyzer except mains power and the USB cable. 1. Send command: MAINS=L1-GND

2. Receive * 3. Send command: GAIN=XXX.X voltage (90 – 264 V ac).

where XXX.X is the actual value of the hot line

4. Receive * 5. Adjust the Elgar power supply to 100 V ac, 60 Hz. 6. Send command: READ

7. Receive and record response. L. Mains Calibration Part II 1. Send command: SAVE

2. Receive * 3. Send command: EXIT

4. Receive * 5. Send command: LOCAL

6. Receive * 7. Turn off the Analyzer. 8. Chang the power supply configuration such that: a. L2 → hot b. L1 → neutral c. GND → ground d. Adjust voltage back to 115 V, 60 Hz. 9. Turn the Analyzer on. The error code !55 will be received. 10. Send command: REMOTE=F2810 11. Receive * 12. Send command: CAL=F2810 13. Receive * 14. Send command: MAINS=L2-GND 15. Receive * 16. Send command: GAIN=XXX.X voltage (90 – 264 V ac).

where XXX.X is the actual value of the hot line

17. Receive * 18. Adjust the Elgar power supply to 100 V ac, 60 Hz. 19. Send command: READ 3-66

Verification and Calibration Calibration Procedure

3

20. Receive and record response. 21. Return the Elgar power supply to 115 V ac, 60 Hz. Do not turn off power to the Analyzer until section M is completed. M. Save Calibration 1. Write new calibration date in the format of MxxDyyYzzzz, where xx stands for the two digits of the current month, yy stands for the two digits of the current day, and zzzz stands for the four digits of the current year. 2. Send command: DATE=MxxDxxYxxxx 3. Receive * 4. Write calibration technician in the format of xxxxx, where xxxxx stands for the five digits of the technician’s employee number. 5. Send command: TECH=xxxxx

6. Receive * 7. Check Analyzer serial number, cal date, and cal technician are correct. 8. Send command: INFO=

9. Receive serial number, cal date, and cal technician in response. Record response. 10. Save the primary cal data. Send command: SAVE 11. Exit cal mode. Send command: EXIT



12. Receive * 13. Return to local mode. Send command: LOCAL

14. Receive * N. Conduct Verification

3-67

ESA612 Service Manual

3-68

Chapter 4

Remote Operations

Title Introduction.......................................................................................................... Serial Communication Connections .................................................................... Serial Communication ......................................................................................... Modes .................................................................................................................. Reading Formats .................................................................................................. Command Specifications ..................................................................................... General Commands ......................................................................................... Remote Commands ......................................................................................... ECG Commands.............................................................................................. Calibration Commands .................................................................................... Diagnostic Commands..................................................................................... Error Responses ................................................................................................... Status Word Definitions....................................................................................... Firmware Updates................................................................................................

Page 4-3 4-3 4-3 4-3 4-5 4-5 4-5 4-6 4-11 4-12 4-18 4-21 4-22 4-24

4-1

ESA612 Service Manual

4-2

Remote Operations Introduction

4

Introduction The Analyzer is designed for remote operation by sending it commands through its computer port. Measurement data and command responses are returned through the computer port for computer or user analysis. This remote capability can be employed to automate the verification, calibration, and troubleshooting procedures described in this manual. Extensive automation can be realized by writing a computer program that sends commands to the Analyzer and evaluates the responses to generate a user report. Or, individual commands can be typed and responses read using a program such as Hyperterminal. The degree of automation is up to the users needs. This chapter describes how to connect and setup the Analyzer for remote operation. Also included is a complete list of valid commands, each with a functional description, proper command syntax, Analyzer response, associated command parameters, and examples. A set of diagnostic commands are also included to help isolate problems in some functions and circuits while troubleshooting the Analyzer.

Serial Communication Connections When unprogrammed devices are connected to a PC, they look different to Windows. Without intervention, Windows will enumerate each device separately. To force Windows to ignore the serial numbers of FT232R devices, add a registry setting. •

Edit the registry to add an entry at: "HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\UsbFlags\".



Add a REG_BINARY entry "IgnoreHWSerNum04036001".



Set the binary value to "01".



Refer to FTDI Application Note AN232B-10 "Advanced Driver Options".

Note This only works if you add the registry entry before connecting any FTDI devices. You might have to uninstall FTDI device drivers with their "FTCLEAN" utility, then reload the drivers. You can force Windows (2000, XP, and above) to use the LocId of the USB port so all FTDI devices on a port are assigned to the same COM port number. Refer to FTDI Application Note AN232B-07 "Configuring FTDI's VCP Drivers to use Location IDs". Note If you properly set IgnoreHWSerNum04036001 as described above, you shouldn't have to use LocId.

Serial Communication These are common specifications for all serial ports: Baud rate: 115,200 baud Format: 8 data bits, no parity, 1 stop bit

Modes The following is a list of Analyzer modes with a brief description of each: 1. Power Up •

The Analyzer performs a number of self-diagnotic tests during this temporary 4-3

ESA612 Service Manual

start-up mode. •

The display shows a “splash” screen during this period.

2. Local •

This is the default mode entered after power-up.



The user controls the Analyzer by pressing buttons and viewing the display.



The computer interface is active at any of the user menus, but the UI can only execute a few commands: those that put the Analyzer into remote mode or request instrument status. All other commands are illegal.

3. Remote •

The display indicates that the Analyzer is under remote control.



The UI receives commands from the USB interface to control the Analyzer.



The Analyzer can be put back to local mode by a command or by pressing a key.



The UI will directly execute some commands that it can without contacting the Meter.



Other commands not executable by the UI will be forwarded to the Meter exactly as received. Then, the UI will get the response from the Meter and pass it back to the computer as received.

4. Cremote •

Same as Remote Mode, except that responses from the Analyzer are enclosed in ‘packets.’ Commands from the PC to the Analyzer must also follow this format.



This protocol also applies to sub-modes within remote control.



When returning to Local Mode, the communication protocol is restored to normal operation.



The packet format is defined below: STX

Byte Count

Data

ETX

Checksum

NULL

1 byte

4 bytes

N bytes

I byte

2 bytes

1 byte

STX = ‘0x02’ character Byte count =

Number of bytes in the Data field only (N bytes)

Data =

Command to or response from the Analyzer

ETX =

‘0x03’ character

Checksum =

Sum of bytes in the byte count and data fields only (4 + N bytes)

NULL =

‘0x00’ character

Note Special characters that do not need to be followed by carriage return/line feed, such as ESC and BACKSPACE characters, must be sent outside of packets. 5. Error

4-4



The Analyzer enters the error mode when it detects an internal fault.



The display shows an error screen where the user is prompted to take a given

Remote Operations Reading Formats

4

action to correct the error. 6. Test •

This is a special mode for Engineering testing purposes only.



The Analyzer is placed in an infinite loop, where it continuously takes measurements that are shown on the display.

7. Sub-modes •

The Analyzer can be put in certain sub-modes within remote control.



Sub-modes can allow only certain commands.



UI sub-modes for the Analyzer are Diagnostic and Calibration.

Reading Formats Readings returned by the meter microprocessor have a fixed number of decimal places, which may very from test to test. Table 4-1 lists the tests and their associated returned reading formats. Note Readings do not have leading zeroes, except when preceding the decimal point. Table 4-1. Reading Formats Character Preceding Reading Value

Maximum Number of Digits before Decimal Point

Number of Decimal Places

Example [1]

Voltage

V

3

1

“V221.2”

Resistance

O

1

3

“O1.001”

Insulation Resistance

M

2

1

“M5.3”

Current

A

2

1

“A10.4”

Leakage (mA)

L

Depends on range

Depends on range

Depends on range

Leakage (μA)

U

Depends on range

Depends on range

Depends on range

Test

[1]

All reading values are preceded by a letter that specifies the type of reading.

Command Specifications The following sections list the available commands which remotely control the Anaylzer. The commands are divided into the following functional analyzer categories: General, Remote, ECG, Calibration, Diagnostic, and Error. General and global setup commands ready the Analyzer for remote operation from a computer. The calibration commands are used to calibrate the Analyzer remotely. The diagnostic category of commands are for isolating problems in certain Analyzer functions and circuits. General Commands General commands set the Analyzer up for remote operation and are used to extract Analyzer information such as its calibration and current mode of operation.

4-5

ESA612 Service Manual

CREMOTE Description Puts the Analyzer in computer remote control. Parameters

Examples

P = Packet mode

CREMOTE=P

Note To escape from packet mode, return the Analyzer to local mode and reenter remote mode. IDENT Description Queries the Analyzer for identification. Query

IDENT

Returns the Analyzer model, UI firmware version, and meter processor firmware version.

Note When in calibration or diagnostic mode, the UI firmware and meter microprocessor firmware version are listed separately . LOCAL Description Exits remote control and returns to local control. REMOTE Description Sets the Analyzer to remote control mode. RESEND Description Resends the last response to the PC. RSTUI Description Resets the Analyzer. RSTM Description Resets the meter microprocessor. STAT Description Returns the UI status word. Remote Commands The remote commands are used to select or retrieve the tests and perform the necessary setup conditions for the various tests. Some control is also available for retrieving readings from the Analyzer through these commands. ALTEARTH Description Opens or closes the ground to the equipment outlet during Alternative Equipment Leakage function. Parameters

C closes equipment outlet ground and O opens it.

Examples

Closes equipment outlet ground.

ALTEARTH=C

AP Description Selects applied parts to connect to meter (+), applied parts to and connect to meter (-), and where to connect the remaining applied parts. Parameters

4-6

parts+ = list of parts to connect to meter (+) parts- = list of parts to connect to meter (–)

Remote Operations Command Specifications

4

parts remaining = where to connect remaining parts. Examples

AP=RL,LL/RA,V1/GND Connects RL and LL to meter (+), RA and V1 to

meter (-), and all others to ground (GND).

APINS Description Selects applied parts insulation test. AUX Description Select patient auxiliary leakage test. CAL Description Enters Analyzer calibration mode. Parameters

F2810 = Password for the Analyzer.

Examples

Puts Analyzer in calibration mode.

CAL=F2810

DIAG Description Enters Analyzer diagnostic mode. Parameters

F2810 = Password for the Analyzer.

Examples

Puts Analyzer in diagnostic mode.

DIAG=F2810

DIFF Description Selects differential current test. DIRL Description Selects direct leakage test. DMAP Description Selects direct applied part leakage test. Uses MAP voltage. EARTH Description Opens or closes the ground to the equipment outlet. Parameters

C closes equipment outlet ground and O opens it.

Examples

Closes equipment outlet ground.

EARTH=C

EARTHL Description Selects earth leakage test. ECG Description Enters ECG simulation mode and connects all applied parts to ECG. ENCL Description Selects Enclosure Leakage measurement. EOGNULL Description Opens or closes the connection between the NULL jack and the the equipment outlet ground. Parameters

C closes connection between NULL jack and equipment outlet ground and O opens it.

Examples

Closes connection between NULL jack and equipment outlet ground.

EARTH=C

4-7

ESA612 Service Manual

EQCURR Description Selects equipment current test. ERES Description Selects earth resistance 200 mA current. FN Description Returns the current function number. Examples

FN

0 = No function selected 1 = Mains voltage 2 = Equipment current 3 = Earth resistance 4 = Mains to earth insulation 5 = Applied parts to earth insulation 6 = Earth leakage 7 = Enclosure leakage 8 = Patient leakage 9 = Patient auxiliary leakage 10 = Direct equipment leakage 11 = Direct applied parts leakage 12 = MAP leakage 13 = Alternative applied parts leakage 14 = Alternative equipment leakage 15 = Differential leakage 16 = Not used 17 = Point to point leakage 18 = Not used 19 = Point to point voltage 20 = Point to point resistance 21 = Mains to neutral insulation 22 = Applied parts to neutral insulation 23 = Mains to applied parts insulation 24 = Lead isolation leakage

GFI Description Selects ground fault interrupt (GFI) trip level. Parameters

nn = 5, 10, or 25

Examples

Sets ground fault interrupt to 5 mA.

GFI=5MA

GFIR Description Resets ground fault interrupt ATTENTION. HIGH_RES Description Controls the number of digits in the displayed reading. Parameters Examples

HIGH_RES=ON

Increases the number of digits in the displayed reading.

IDLE Description Turns off all relays and ASWs, clears faults and status words. INS Description Controls the insulation resistance test voltage. 4-8

Remote Operations Command Specifications

Parameters

xxxx = Insulation voltage variable. LOW = Selects 250 V insulation voltage HIGH = Selects 500 V insulation voltage (default)

Examples

Sets insulation voltage to 250 V.

INS=LOW

4

INSB Description Selects insulation test B. INSD Description Selects insulation test D. INSE Description Selects insulation test E. LEAD_ISO Description Selects lead isolation function. LOAD Description Selects the load for the meter input. Parameters

xxxx = Meter load variable. 601 = IEC 60601 patient load AAMI = ANSI/AAMI patient load NONE = IEC 60601 patient load

Examples

Selects the IEC 60601 patient load.

LOAD=601

LOADDSP Description Puts the meter microprocessor in boot loader mode. LOCAL Description Puts the Analyzer in local control mode. MAINS Description Selects the mains voltage measurement test. Parameters

L1-L2 = Measures ac, hot to neutral. L1-GND = Measures ac, hot to ground. L2-GND = Measures ac, neutral to ground.

Examples

Sets measurement for ac, hot to neutral.

MAINS=L1-L2

MAP Description Selects the mains on applied parts test. Parameters

LOW = Switches to MAP level of 100 %. NORM = Switches to normal polarity MAP voltage. REV = Switches to reverse polarity MAP voltage. 1MA = Sets MAP limit to 1 mA. 3.5MA = Sets MAP limit to 3.5 mA. 7.5MA = Sets MAP limit to 7.5 mA.

Examples

Sets MAP voltage to 100 %.

MAP=LOW

MINS Description Selects mains insulation test. 4-9

ESA612 Service Manual

MODE Description Selects voltage measurement mode. Parameters

DC = Selects dc measurement mode. AC = Selects ac measurement mode. ACDC = Selects ac+dc measurement mode.

Examples

Selects ac measurement mode.

MODE=AC

MREAD Description Returns meter readings continuously (~400 ms on a single range) until an escape character is received. NEUT Description Opens or closes the neutral to the equipment outlet. Parameters

C = Closes equipment outlet neutral and O opens it.

Examples

Closes neutral to equipment outlet.

NEUT=C

NOMINAL Description Causes all leakage except differential to be multiplied by the ratio of Nominal Mains/Actual Mains. Parameters

ON = leakage multiplied by ratio Nominal Mains/Actual Mains. OFF = leakage without multiplier. xxx = Nominal value to be stored by user.

Examples

Leakage value multiplied by Nominal Mains/Actual Mains.

NOMINAL=ON

NOMINAL? Description Returns the current stored value for nominal. NOSHOW Description Return reading only during MREAD command (default at power-on). OVR Description Resets over voltage ATTENTION. PAT Description Selects patient leakage test. POL Description Sets equipment outlet polarity. Parameters

OFF = Sets equipment outlet polarity off. N = Sets equipment outlet polarity to normal. R = Sets equipment outlet polarity to reverse.

Examples

Sets equipment outlet polarity to reverse.

POL=R

PPL Description Selects point to point leakage test. PPR Description Selects point to point resistance test and set the test current value. Parameters < LOW > 4-10

LOW = Selects low current (200 mA).

Remote Operations Command Specifications

4

Note A PPR command without a parameter just selects the test. Examples

PPR=LOW

Selects the point to point resistance test.

PPV Description Selects point to point voltage test. READ Description Returns a single meter reading. RPTIME Description Sets the equipment outlet (EO) polarity switch time between zero and five seconds. Parameters

n = Time in seconds (0 – 5).

Examples

Sets EO polarity switch time to 3 seconds.

RPTIME=3

SAF Description Selects substitute appliance fault leakage test. SHOWALL Description Returns range, ADC count, and reading during MREAD command. SN Description Returns ESA612 serial number. SPAT Description Selects substitute patient leakage test. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3. STD Description Selects the test standard for tests. Based on the standard, the Analyzer automatically sets the load, GFI trip, and MAP voltage and current values. Parameters

xxxx = Desired test standard. 601 = IEC 60601 standard AAMI = ANSI/AAMI standard ASNZ = AS/NZS 3551 standard 353 = IEC 62353 standard

Examples

Selects the ANSI/AAMI standard.

STD=AAMI

ZERO Description Zero the resistance meter. $ Description Puts the UI microprocessor in boot loader mode. ECG Commands The following commands control the ECG function of the Analyzer. 4-11

ESA612 Service Manual

CPL30 Description Runs ECG complex wave @ 30 Hz. CPL60 Description Runs ECG complex wave @ 60 Hz. CPL120 Description Runs ECG complex wave @ 120 Hz. CPL180 Description Runs ECG complex wave @ 180 Hz. CPL240 Description Runs ECG complex wave @ 240 Hz. EXIT Description Exits ECG simulation mode and disconnects all applied parts. PLS30 Description Runs ECG 63 ms pulse @ 30 Hz. PLS60 Description Runs ECG 63 ms pulse @ 60 Hz. RESEND Description Resends the last response to the PC. SN10 Description Runs ECG sine wave @ 10 Hz. SN40 Description Runs ECG sine wave @ 40 Hz. SN50 Description Runs ECG sine wave @ 50 Hz. SN60 Description Runs ECG sine wave @ 60 Hz. SN100 Description Runs ECG sine wave @ 100 Hz. SQ125 Description Runs ECG square wave @ 0.125 Hz. SQ2 Description Runs ECG square wave @ 2.0 Hz. TR2 Description Runs ECG triangle wave @ 2 Hz. VFIB Description Runs ECG ventricular fibrillation. Calibration Commands The following commands are used to remotely calibrate the Analyzer. APINS 4-12

Remote Operations Command Specifications

4

Description Selects applied parts insulation test. CAL_GFI Description The GFI level must be selected before running this command. Current must be supplied from AC Hot to GND, and CAL_GFI must be executed. This will calibrate the selected GFI (5 mA, 10 mA, or 25 mA) to trip at the current value supplied. CLEAREE Description Clears all data stored in EEPROM (both UI and meter data). COMP Description Generates MAP comp DAC offset value. DATE Description Programs calibration date into Analyzer EEPROM. Parameters

Mxx = Month number Dnn = Day number Ynnnn = Year number

Examples

Sets Calibration date to 11/23/2009.

M11D23Y2009

DEFAULTS Description Sets calibration constants to defaults. DIFF Description Selects differential current test. DIS_FAULT Description Enables or disables GFI and OV faults. Parameters

0 = Enables GFI and OV faults 1 = Disables GFI and OV faults

Examples

Enables GFI and OV faults.

DIS_FAULT=0

EARTH Description Opens or closes the ground to the equipment outlet. Parameters

C = Closes equipment outlet ground and O opens it.

Examples

Closes ground to equipment outlet.

EARTH=C

EQCURR Description Selects equipment current test. ERES Description Selects earth resistance test. Parameters

LOW = 200 mA test current

Examples

Sets earth resistance test current to 200 mA.

ERES=LOW

EXIT Description Exits the calibration mode. FN Description Returns the current function number. 4-13

ESA612 Service Manual

Examples

FN

0 = No function selected 1 = Mains voltage 2 = Equipment current 3 = Earth resistance 4 = Mains to earth insulation 5 = Applied parts to earth insulation 6 = Earth leakage 7 = Enclosure leakage 8 = Patient leakage 9 = Patient auxiliary leakage 10 = Direct equipment leakage 11 = Direct applied parts leakage 12 = MAP leakage 13 = Alternative applied parts leakage 14 = Alternative equipment leakage 15 = Differential leakage 16 = Not used 17 = Point to point leakage 18 = Not used 19 = Point to point voltage 20 = Point to point resistance 21 = Mains to neutral insulation 22 = Applied parts to neutral insulation 23 = Mains to applied parts insulation 24 = Lead isolation leakage 25 = Meter ac 26 = Meter dc

GAIN Description Allows the Analyzer to record current meter value and calculate counts per unit gain, which is then stored in EEPROM. Parameters

nnnn = Actual value of applied signal.

Examples

Command to calibrate current meter value to an applied signal (in this case, mains voltage).

GAIN=115.0

GFI Description Selects ground fault interrupt (GFI) trip level. Parameters

nn = 5, 10, or 25

Examples

Sets ground fault interrupt to 5 mA.

GFI=5MA

GFIR Description Resets ground fault interrupt ATTENTION.

HIGH_RES Description Increases the number of digits displayed in the reading. This command is only used by automated calibration.

4-14

Parameters

ON = increase the number of display digits. OFF = normal number of display digits..

Examples

Increase the number of display digits.

HIGH_RES=ON

Remote Operations Command Specifications

4

INFO Description Returns serial number, calibration date, and technician number. INS Description Controls the insulation resistance test voltage. Parameters

xxxx = Insulation voltage variable. LOW = Selects 250 V insulation voltage HIGH = Selects 500 V insulation voltage (default)

Examples

Sets insulation voltage to 250 V.

INS=LOW

INSWAP Description Swaps MAINS input neutral and hot. Parameters

0 = MAINS neutral and hot to normal polarity 1 = MAINS neutral and hot to reverse polarity.

Examples

Sets MAINS neutral and hot to normal polarity.

INSWAP=0

LEAD_ISO Description Selects the lead isolation function. LOAD Description Selects the load for the meter input. Parameters

xxxx = Meter load variable. 601 = IEC 60601 patient load AAMI = ANSI/AAMI patient load NONE = IEC 60601 patient load

Examples

Selects the IEC 60601 patient load.

LOAD=601

MAINS Description Selects the mains voltage measurement test. Parameters

L1-L2 = Measures ac, hot to neutral. L1-GND = Measures ac, hot to ground. L2-GND = Measures ac, neutral to ground.

Examples

Sets measurement for ac, hot to neutral.

MAINS=L1-L2

MAP Description Selects the mains on applied parts test. Parameters

LOW = Switches to MAP level of 100 %. NORM = Switches to normal polarity MAP voltage. REV = Switches to reverse polarity MAP voltage. 1MA = Sets MAP limit to 1 mA. 3.5MA = Sets MAP limit to 3.5 mA. 7.5MA = Sets MAP limit to 7.5 mA.

Examples

Sets MAP voltage to 100 %.

MAP=LOW

METER_AC Description Selects a test that measures ac voltage on red and black jacks without the potential divider, i.e. connects directly to the meter. 4-15

ESA612 Service Manual

Note Like the PPV test, this command forces IEC filtering. METER_DC Description Selects a test that measures dc voltage on red and black jacks without the potential divider, i.e. connects directly to the meter. Note Like the PPV test, this command forces IEC filtering. MODE Description Selects voltage measurement mode. Parameters

DC = Selects dc measurement mode. AC = Selects ac measurement mode. ACDC = Selects ac+dc measurement mode.

Examples

Selects ac measurement mode.

MODE=AC

MREAD Description Returns meter readings continuously (on a single range) until an escape character is received. NOSHOW Description Return reading only during MREAD command (default at power-on). OFFSET Description Sets offsets to zero all ranges for the current function. OVR Description Resets over voltage ATTENTION. POL Description Sets equipment outlet polarity. Parameters

OFF = Sets equipment outlet polarity off. N = Sets equipment outlet polarity to normal. R = Sets equipment outlet polarity to reverse.

Examples

Sets equipment outlet polarity to reverse.

POL=R

PPL Description Selects point to point leakage test. PPR Description Selects point to point resistance test and set the test current value. Parameters

LOW = Selects low current (200 mA).

Note A PPR command without a parameter just selects the test. Examples

PPR=LOW

Selects the point to point resistance test.

PPV Description Selects point to point voltage test. PSN Description Sets serial number into EEPROM. Parameters 4-16

nnnnnnn = serial number up to seven digits.

Remote Operations Command Specifications

Examples

4

PSN=1234567

RANGE Description Sets the meter range. Parameters

n = range and function numbers (0 – 4).

Query

RANGE?

Returns the number of the meter range.

Examples

RANGE=0,0,1

Sets the ac meter range to range 1.

READ Description Returns a single meter reading with no autoranging. RPTIME Description Sets the equipment outlet (EO) polarity switch time between zero and five seconds. Parameters

n = Time in seconds (0 – 5).

Examples

Sets EO polarity switch time to 3 seconds.

RPTIME=3

RSTM Description Resets the meter processor. RSTUI Description Resets the Analyzer. RSTEECAL Description Clears the meter calibration constants stored in EEPROM. SAVE Description Saves the current calibration constants into EEPROM. SHOWALL Description Returns range, ADC count, and reading during MREAD command. STAT Description Returns the UI ASCII hex status word. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3. STD Description Selects the test standard for tests. Based on the standard, the Analyzer automatically sets the load, GFI trip, and MAP voltage and current values. Parameters

xxxx = Desired test standard. 601 = IEC 60601 standard AAMI = ANSI/AAMI standard ASNZ = AS/NZS 3551 standard 353 = IEC 62353 standard

Examples

Selects the ANSI/AAMI standard.

STD=AAMI

4-17

ESA612 Service Manual

TECH Description Programs technician code into EEPROM. Twenty alphanumeric characters allowed. TIMEOUT Description Turns the UI to DSP communications timeout on and off. Parameters Examples

ON = Turns on timeout. OFF = Turns off timeout

TIMEOUT=ON

UPLOAD Description DSP sends formatted EEPROM data to USB port. ZERO Description Zero the resistance meter. Diagnostic Commands The following commands set specific conditions within the Analyzer and are used strictly for troubleshooting Analyzer problems. These commands allow isolation of, and access to, functional blocks of the Analyzer. These commands can also be used during Analyzer verification and calibration. BEEP Description

Sounds the beeper once.

CONT Description Sets the contrast to the specified value (in ASCII hex). Range is 00 to 11. Examples

CONT=0F

Sets contrast.

DAC Description Sets ECG DAC register to a specified hex value Parameters

xxx = Hex value

Examples

Sets ECG DAC to 0F0.

DAC=0F0

DSPECHO Description Runs continuous DSP communication test. (“ESC” to stop). Characters received from USB are shown on the display and sent to the DSP. The DSP echoes the character back to the UI, which then sends it to the PC. DUPDATE Description Turns the display update on or off. Parameters Examples

DUPDATE=ON

Display update on.

EEPROMV Description Returns the UI EEPROM values. ENCL Description Selects Enclosure Leakage measurement. EXIT Description Exits diagnostic mode. 4-18

Remote Operations Command Specifications

4

FLASH Description Returns the UI flash checksum [XXXX]. FRAMT Description Runs an FRAM test and returns “PASS” or “FAIL”. GFIR Description Resets ground fault interrupt ATTENTION. IDENT Description Queries the Analyzer for identification. Query

IDENT

Returns the Analyzer model, UI firmware version, and meter processor firmware version.

Note When in calibration or diagnostic mode, the UI firmware and meter microprocessor firmware version are listed separately . KEY Description Runs continuous keypad test (“ESC” to stop). Pressed key number is shown in the display. KRL Description Disables all relays by asserting the RELAY_DISABLE line. LED Description Turns LED on or off. Parameters Examples

LED=ON

Turns LED on.

LITE Description Turns backlight on or off. Parameters Examples

LITE=ON

Turns backlight on.

MCS Description Returns the meter microprocessor flash checksum [xxxx]. OTEMP Description Returns the over temperature status bit. OVR Description Resets over voltage ATTENTION. POFF Description Turns off all display pixels. PON Description Turns on all display pixels. PXL Description Write pixel column (8-bit) to the display, given the address and value (all in ASCII hex). Parameters 4-19

ESA612 Service Manual

Examples

Writes FF to row 5 and column 0A7.

PXL=05,0A7,FF

R Description Repeat last command. READ_AD Description Returns A/D value. Takes up to 100 ms. RESEND Description Resends the last response to the PC. RFRAM Description Reads the value (8-bit) from the specified FRAM location. Parameters Examples

RFRAM=0023

xxxx = Memory address (0000 – 7FFF). Returns the 8-bit word at address 0023.

RXFLASH Description Reads the value (8-bit) from the specified external FLASH location. Parameters

xxxx = Memory address (000000 – 20FFFF).

Examples

Returns the 8-bit word at address 000023.

RXFLASH=000023

SET_GFI Description Sets GFI DAC register to a specified hex value Parameters

xxx = Hex value

Examples

Sets GFI DAC register to 00A.

SET_GFI=00A

SET_MAP Description Sets MAP DAC register to a specified hex value Parameters

xxx = Hex value

Examples

Sets MAP DAC register to 002.

SET_MAP=002

SPI Description Writes single SPI byte. Parameters

x = SPI byte (0 – 9) y = value (0x00 – 0xFF) z = loop (Z=1 means loop until ESC).

Examples

Write 01 to SPI byte 5 repeatedly until escape character is received.

SPI=5,01,1

SPIR Description Returns SPI 0-9 values. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3.

4-20

Remote Operations Error Responses

4

UIECHO Description Runs continuous USB test (“ESC” to stop). Characters received from USB are shown on the display and echoed back to the PC. WFRAM Description Writes the value (8-bit) to the specified external FRAM location. Parameters ,

xxxx = Memory address (0000 – 7FFF) dd = data.

Examples

Writes A3 to address 0023.

WFRAM=0023, A3

WXFLASH Description Writes the value (8-bit) from the specified external FLASH location. Parameters Examples

xxxxxx = Memory address (000000 – 20FFFF) dd = data.

WXFLASH=000023, A3 Writes A3 to address 000023.

XFLASHT Description Runs a test on the external FLASH chip and returns “PASS” or “FAIL”. ZOMBIE Description Places the Analyzer in test mode. Parameters Examples

Turns test mode on.

ZOMBIE=ON

Error Responses Table 4-2 list the error codes and their descriptions, that may be returned from the Analyzer. Table 4-2. Error Responses Code

Description

00

No commands allowed now

01

Unknown command

02

Illegal command, not allowed in current mode

03

Illegal parameter

04

Receive buffer overrun

05

General failure

06

Option not installed

21

Analog to Digital Converter out of range

30

Test passed

31

Test failed

32

No current

33

Can not null

37

No reading available

4-21

ESA612 Service Manual

Table 4-2. Error Responses (cont.) Code

Description

38

Load discharge timeout

40

Over temperature

41

CREMOTE protocol error

42

Initialization error

50

GFI

51

Overvoltage

52

Analyzer out of calibration

53

Mains out of range

54

Open ground

55

Reverse voltage

56

Polarity time wait

58

External memory error

Status Word Definitions Table 4-3 lists the status word bit definitions. Table 4-3. Status Word Definitions Description

Status Word

Definition

UI Status Word Bit Definitions

4-22

POWER_UP

0x0001

Device is in power-up mode

LOCAL

0x0002

Device is in local mode

REMOTE

0x0004

Device is in remote mode

CREMOTE

0x0008

Device is in cremote mode

DIAG

0x0010

Device is in diagnostic mode

CAL

0x0020

Device is in cal mode

ERROR

0x0040

Device is in error mode

TEST

0x0080

Device is in test mode

OVER_TEMP

0x0100

Overtemp condition

SPARE

0x0200

Spare

SPARE

0x0400

Spare

SPARE

0x0800

Spare

SPARE

0x1000

Spare

SPARE

0x2000

Spare

SPARE

0x4000

Spare

Remote Operations Status Word Definitions

4

Table 4-3. Status Word Definitions (cont.) Description SPARE

Status Word

Definition

0x8000

Spare

REMOTE

0x0001

Device is in remote mode

DIAG

0x0002

Device is in diagnostic mode

CAL

0x0004

Device is in cal mode

ECG

0x0008

Device is in ECG mode

Spare

0x0010

Spare

SVOLTS

0x0020

Measure from 0 to 300 volts

SLEAK

0x0040

Measure from 0 to 10,000 μA

SOHMS

0x0080

Measure from 0 to 2 Ω @ 200 mA

Spare

0x0100

Spare

SMEG

0x0200

Measure from 0 to 100 MΩ

SEQUIP

0x0400

Measure from 0 to 20 A ac

SDIFF

0x0800

Measure from 0 to 10 mA ac

AC_ONLY

0x1000

Measure ac only

DC_ONLY

0x2000

Measure dc only

ACDC

0x4000

Measure ac+dc

SPARE

0x8000

Spare

LDAAMI

0x0001

AAMI load selected

Spare

0x0002

Spare

LD601

0x0004

601 load selected

EO

0x0008

Equipment outlet on

Spare

0x0010

Spare

MAPR

0x0020

MAP reverse selected

MAPON

0x0040

MAP voltage on

L2OPEN

0x0080

Neutral open

EOPEN

0x0100

Earth open

POLR

0x0200

EO polarity reversed

GFIL

0x0400

GFI low selected

GFIH

0x0800

GFI high selected

INS_ON

0x1000

Insulation voltage (500 V) on

RCURON

0x2000

Resistance current on

Status Word 1 Bit Definitions

Status Word 2 Bit Definitions

4-23

ESA612 Service Manual

Table 4-3. Status Word Definitions (cont.) Description

Status Word

Definition

MAINS0

0x4000

MAINS0-MAINS1 = Mains parameter selection.

MAINS1

0x8000

MAINS0-MAINS1: 00 = Not Used 01 = L2-GND 10 = L1-GND 11 = L1-L2

RPT0

0x0001

RPT0 – RPT2 reverse EO polarity switch time = 0 – 3.5 sec

RPT1

0x0002

RPT0 – RPT2 is a count of 1 sec increments

RPT2

0x0004

GFIM

0x0008

GFI medium level (10 mA) selected

SHOWALL

0x0010

SHOWALL selected

NOMINAL

0x0020

NOMINAL selected

INS_LOW

0x0040

1 – insulation volts = 250 V; 0 – insulation volts = 500 V

MAP3MA

0x0080

MAP test – current limit = 3.5 mA if on or 1 mA if off

Spare

0x0100

Spare

MAINS

0x0200

0 = 115 V ac, 1 = 230 V ac Mains

EEP_CS_ERR

0x0400

EEPROM checksum error

VOLT_BAD

0x0800

MAINS – voltage L1-L2 out of range

BAD_GND

0x1000

MAINS – AC GND bad

REV_PWR

0x2000

MAINS – L1 and L2 reversed

GFITRIP

0x4000

GFI interrupt has occurred

FAULT

0x8000

FAULT interrupt has occurred

Status Word 3 Bit Definitions

Firmware Updates To update the firmware in the Analyzer, two firmware updates must be done: one for the meter and one for the UI.

4-24

Remote Operations Firmware Updates

4

W Caution To prevent damage to the Analyzer, do not interrupt the firmware download process. If this should happen, then both boards will have to be returned to the factory for reinitialization. Do not attempt to return the Analyzer to an earlier firmware version. This could result in a corrupt firmware load, and both boards will have to be returned to the factory for reinitialization. 1. With the Analyzer powered off, connect your PC to the Analyzer and run AnsurMLC.exe. 2. Click Tools->Upload Firmware. 3. Select ESA612, select the appropriate COM port, and click NEXT. 4. Browse to the appropriate DSP and UI firmware HEX files, and click NEXT. 5. Power the Analyzer on. 6. Click UPLOAD. 7. The LCD will blank but AnsurMLC will begin downloading the new firmware. When both downloads are complete, cycle power to the Analyzer and verify the new firmware has been correctly uploaded.

4-25

ESA612 Service Manual

4-26

Chapter 5

Maintenance

Title Introduction.......................................................................................................... Testing and Replacing the Fuses.......................................................................... Cleaning the Analyzer ......................................................................................... Disassembly/Assembly ........................................................................................ Troubleshooting ................................................................................................... Power on Failure.............................................................................................. Output Failures ................................................................................................ Meter Failures.................................................................................................. MAINS Power............................................................................................. Differential Current and Equipment Current .............................................. Meter Troubleshooting ....................................................................................

Page 5-3 5-3 5-3 5-4 5-4 5-4 5-5 5-5 5-5 5-5 5-5

5-1

ESA612 Service Manual

5-2

Maintenance Introduction

5

Introduction This chapter covers the general maintenance as well as troubleshooting information for isolating failed Analyzer components. The Analyzer needs little maintenance or special care. However, treat it as a calibrated measuring instrument. Avoid dropping or other mechanical abuse that could cause a shift in the calibrated settings.

Testing and Replacing the Fuses XW Warning To prevent electric shock, remove all power cords and test leads from the Analyzer before opening the fuse door. For electrical protection of the equipment outlet, the Analyzer uses two fuses, one in the live (L1) line and one in the neutral (L2) line. To test the fuses, do the following while referring to Figure 5-1: 1. Turn the Analyzer so the case bottom is facing up. 2. Flip up the tilt stand. 3. Remove the fuse door from the Analyzer by removing the screw holding the fuse door with a #2 Phillips head screwdriver and lifting the fuse door from the Analyzer. 4. Remove the fuses from the Analyzer.

F1 - F2

Figure 5-1. Fuse Access

fis111.eps

5. Using a multimeter, measure the continuity of each fuse. If one or both fuses do not show continuity, replace the fuse(s) with fuses that have the same current and voltage rating. Appropriate fuse ratings are posted on the case bottom label of the Analyzer. The final assembly table in the Replaceable Parts chapter lists available fuses with Fluke Biomedical part numbers. 6. Reinstall the fuse door and secure it with the screw.

Cleaning the Analyzer XW Warning To avoid electric shock, do not clean the Analyzer plugged into mains or attached to a DUT. W Caution Do not pour fluid onto the Analyzer surface; fluid seepage into the electrical circuitry may cause the Analyzer to fail. 5-3

ESA612 Service Manual

W Caution Do not use spray cleaners on the Analyzer; such action may force cleaning fluid into the Analyzer and damage electronic components. Clean the Analyzer occasionally utilizing a damp cloth and mild detergent. Take care to prevent the entrance of liquids. Wipe down the adapter cables with the same care. Inspect them for damage to and deterioration of the insulation. Check the connections for integrity before each use.

Disassembly/Assembly To disassemble the Analyzer, do the following: 1. Turn the Analyzer so the case bottom is facing up. 2. Remove the six screws in the case bottom with a Philips screwdriver 3. Lift the case bottom off of the case top and disconnect the cables that connect the two circuit boards together. The A1 Power PCA will remain attached to the case bottom and the A2 Meter PCA will remain with the case top. To reassemble the Analyzer, follow the disassembly steps in reverse order.

Troubleshooting The verification procedure contained in this manual is an effective troubleshooting tool; based upon the section of the test that has failed, or on the failure mode, the component that has caused the malfunction can be isolated. Power on Failure If the Analyzer fails to power up, verify fuses F1 and F2 inside the rear fuse-panel door have not blown. If they have not blown, verify power is applied to the power inlet assembly and that the wiring on the power switch internal to the device is correct. Failures during power up are usually related to mismatched firmware or to power supply problems. To verify firmware, if the device will power up and gives an “Initialization Error,” it may be possible to reach the remote mode through the USB port. Establish hyperterminal communication with the device and send the following command: REMOTE=F2810

If the device displays “Remote Mode” on the front panel and responds with a “*,” determine the firmware status. CAL=F2810 Receive * IDENT

Receive the meter and UI firmware builds. Verify that the two are matched; if they are not, upload the latest firmware to the device and try again. If the firmware matches and is up to date, verify the power and data cables connecting the A1 Power PCA and A2 Meter PCA are connected at J4 and J7 on the power board and at J4 and J20, respectively, on the meter board; also inspect the soldering of the connectors to the pcas. If these connections are not at fault, verify the power supplies for the processors and for the device as a whole are functional. First verify U9, the power supply module at the bottom of the power board, is functional and that CR4, the red LED, is on. Then verify the outputs of U12 and U28 are correct. On the meter board, check that Q9 and U48 – U53 are all operating correctly. Finally, check the power inputs and resets to the microprocessors.

5-4

Maintenance Troubleshooting

5

Output Failures If sourcing a current or voltage through the electrical outlet, such as during an Earth Resistance, or powering a device with normal or reversed polarity, such as during a leakage test, and no signal is applied to the electrical outlet, failures could reside in the physical construction of the electrical outlet, in the reversing relays, or in the signal generation circuit. Verify the electrical outlet wiring is in accordance with the J1 annotation on the pca. Also, if troubleshooting an Australian Analyzer, measure the impedance of the ground connector of the electrical outlet; some outlets have been found to have more than 0.5 Ω. These should be replaced. For US electrical outlets, ensure the backing plate is fully installed; this forces the insulation displacement crimp connection to engage. If this backing plate has been separated, replace the entire electrical outlet. If sourcing a mains power source through the electrical outlet, measure the voltage at the electrical outlet in normal and reversed polarity. If the power configuration is incorrect, inspect the reversing relays on the power PCB: K1-K2, K4-K7. If GFCI trips are also a symptom, verify proper operation of the solid-state relay. K9 and K10 deliver the insulation resistance and MAP signals to the equipment outlet, respectively, and are a good place to half-split the source circuit. The protective earth resistance signal is selected via K12. For leakage measurements, the return path from the electrical outlet is selected by K14. The MAP signal is generated by transformer T1 and regulated by Q1, Q2, and the supporting resistors and is routed through K16 and K17. The insulation resistance voltage is created by transformer T2 and rectified by the nearby circuitry, including Q7; the resultant signal is sensed by U24. Meter Failures While most measurements made by the Analyzer are made by the A2 Meter PCA directly, several are detected or analyzed directly on the A1 Power PCA. MAINS Power Most Mains voltage measurements that are erroneous are the result of power supply failures in U9, isolated supply U18, or in calibration failures. If a calibration error is suspected, calibrate the Mains voltage to determine if this can correct the problem. Otherwise, evaluate F3, the resistor network R2 through R7, U3, and U4 for failures. Differential Current and Equipment Current Both differential and equipment currents are measured at the outlet of the reversing relays, just prior to the Equipment Outlet using L1 and L3, respectively, installed on the backside of the A1 Power PCA beneath the SSR. If all the solder connections on the inductors are good, verify U1, U8, and the supporting circuitry. Meter Troubleshooting If the source signal is functional for a particular measurement, but no measurement is read by the Analyzer, then the meter inputs are suspect. K13 and K15-K18 connect the meter jacks, Red, Black, and Null, to the positive and negative sides of the meter for measurement or resistance nulling. The patient load is the circuit by which a human body is simulated for patient leakage measurements, and is comprised of R78 and R79 (800 and 200 Ω, respectively). For greater leakage currents of 2 mA or more, the meter input is shifted from R78 to R79. To troubleshoot leakage measurements, for ac measurements, half split the circuit at the output of the rms converter, U23 pin 5 or the input of the low pass filter, U31 pin 3. The input to the analog-to-digital converter for ac measurements is at TP95. For dc measurements, use the output of the low pass filter, TP 91. 5-5

ESA612 Service Manual

Resistance measurements are driven by the power supplies on the A1 Power PCA, but is inverted by U15. Applied parts problems should be troubleshot separately than leakage measurements, if possible. First, determine if a single ECG post is affected, or the entire bank. If a single ECG has a failure, inspect the diodes CR24, 29, and 46, L9, or relays K9, K10, and K22 (or respective components with other ECG posts). If all applied parts are affected equally, verify K11 or inspect the signal source (MAP, 500 V dc, etc). For ECG waveform problems, the buffer U13 or resistor network of R5361 drive the entire circuit; the waveform itself is generated by the meter microprocessor. U/I and meter microprocessor failures will normally result in complete nonfunctionality of the Analyzer. Calibration data table and configuration information failures indicate a problem in U44; a loss or corruption of stored test data indicates a failure in U1 or U59. Display board failures are investigated through the driver, which consists of U3 and the power supply of U5, U6, R22 and Q3. Otherwise, the display is replaced as a unit.

5-6

Chapter 6

Replaceable Parts

Title

Page

Introduction.......................................................................................................... 6-3 How to Obtain Parts............................................................................................. 6-3 Parts List .............................................................................................................. 6-4

6-1

ESA612 Service Manual

6-2

Replaceable Parts Introduction

6

Introduction This section contains an illustrated list of replaceable parts for the Analyzer. Parts are listed by assembly, alphabetized by reference designator. Each assembly is accompanied by an illustration showing the location of each part and its reference designator. Refer to Table 6-1. The parts lists give the following information: • • • •

Reference designator (for example, “R52”) An indication if the part is subject to damage by static discharge (* near the part description) Description Fluke stock number WCaution A * symbol indicates a device that may be damaged by static discharge.

How to Obtain Parts Electronic components may be ordered directly from Fluke Biomedical and its authorized representatives by using the part number under the heading Fluke Stock No. Parts price information is available from Fluke Biomedical or its representatives. To contact Fluke Biomedical, call one of the following telephone numbers: • • •

North America 1-800-850-4608 (Email: [email protected]) Europe, Middle East, and Africa +31-402-675-300 (Email: [email protected]) All other countries 1-440-248-9300 (Email: [email protected])

In the event that the part ordered has been replaced by a new or improved part, the replacement will be accompanied by an explanatory note and installation instructions, if necessary. To ensure prompt delivery of the correct part, include the following information when you place an order: • • • • • •

Instrument model and serial number Part number and revision level of the pca (printed circuit assembly) containing the part Reference designator Fluke stock number Description (as given under the Description heading) Quantity

6-3

ESA612 Service Manual

Parts List Tables 6-1 though 6-3 lists the replaceable parts of the Analyzer and their location identified in Figures 6-1 though 6-3. Table 6-1. Final Assembly Item

Description

Fluke Stock No.

A1

ESA612-4002,PCA, POWER

3316505

A2

ESA612-4001,PCA, METER

3316497

C1

ESA612-4401,CABLE ASSY, AC INPUT

3316689

C2

ESA612-4402,CABLE ASSY, POWER -METER

3316692

C3

ESA612-4403,CABLE ASSY, POWER TO METER SUPPLY

3316704

US Japan

WFUSE,20A,250V,3AB,TIME LAG,.25X1.25,BULK

2183691

WFUSE,16A,250V,TIME LAG,6.3X32,CERAMIC BODY,BULK

3321245

WFUSE ,FUSE,.25X1.25,10A,250V,SLOW

109298

France Europe F1, F2

Israel United Kingdom Thailand Australia Switzerland

6-4

H1

SCREW,5-14,.750,PAN,PHILLIPS,STEEL,BLACK CHROMATE,THD FORMING

832246

H2

SCREW,4-14,.375,PAN,PHILLIPS,STEEL,ZINC-ROHS CLEAR,THREAD FORM

448456

H3

SCREW,M3X0.5,6MM,PAN,PHILLIPS,STEEL,ZINC-BLACK CHROMATE

2032792

H4

SCREW,M3X0.5,5MM,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK

2032811

H5

SCREW,6-32,.312,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK

152157

H6

SCREW,8-32,.375,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK

114124

MP1

ESA612-2008,CASE TOP ASSY

3274779

MP2

ESA612-2010,CASE BOTTOM ASSY

3276744

MP3

ESA612-2007-02,FEET SET, LEFT, RIGHT

3359523

MP4

ESA612-2004,DOOR, FUSE ACCESS

3236200

MP5

ESA612-2009,PLATE, TOP END

3236274

MP6

ESA612-2006,BRACKET, DISPLAY

3236239

MP7

ESA612-2019,INSULATOR, INPUT RECEPTACLE

3383167

MP8

ESA612-8002,KEYPAD, UPPER

3258971

MP9

ESA612-8002-01,KEYPAD, LOWER

3258980

MP10

CONTRACT MFG ITEM, RELAY SOLID STATE 75A VDECSA

2077927

Replaceable Parts Parts List

6

Table 6-1. Final Assembly (cont.) Item

Description

Fluke Stock No.

MP11

LCD MODULE,320X240,FSTN,3.8 IN MONO GRAPHIC,TRANSFLCTV,WHITE LED,0.22MM PITCH

3299802

MP12

POWER ENTRY MODULE,A/C INLET,ON/OFF SWITCH,2POLE,20A,250VAC,SNAP-IN,BULK

2791306

MP13

LABEL,SERIAL NUMBER

3097856

US

ESA612-8009,DECAL, BOTTOM CASE, 20 AMP

3343933

ESA612-8009-01,DECAL, BOTTOM CASE, 16 AMP

3343940

ESA612-8009-02,DECAL, BOTTOM CASE, 10 AMP

3343957

ESA612-4405,WIRE ASSY, EQUIPMENT OUTLET

3316728

France

ESA612-4411,WIRE ASSY, EQUIPMENT OUTLET

3343887

Europe

ESA612-4412,WIRE ASSY, EQUIPMENT OUTLET

3343893

Israel

ESA612-4413,WIRE ASSY, EQUIPMENT OUTLET

3343902

Australia

ESA612-4415,WIRE ASSY, EQUIPMENT OUTLET

3343916

United Kingdom

ESA612-4410,WIRE ASSY, EQUIPMENT OUTLET

3316762

Switzerland

ESA612-4416,WIRE ASSY, EQUIPMENT OUTLET

3343925

Thailand

ESA612-4418,WIRE ASSY, EQUIPMENT OUTLET

3474220

LINE CORD,N AMER,20A/125V,5-20/C19,12-3 SJT,2.5M,BLK

2238680

75026,PWRCORDSET EUROPE W/IEC320

2238615

LINE CORD,ISRAEL,16A/250V,SI 32 STRAIGHT OR RT ANG/C19,BLACK,2.5 METERS

2434122

75025,PWRCORD SET AUST W/IEC320

2238603

United Kingdom

75024,PWRCORD SET UK W/IEC320

2238596

Switzerland

FBC-ESA620-4420,SWISS 10A POWER CORD ASSY WITH C-19 CONNECTOR

3379149

Thailand

75033,POWERCORD US WITH C19 PLUG

2238644

Japan France Europe MP14

Israel United Kingdom Thailand Australia Switzerland US Japan

MP15

US Japan France Europe Not Israel Shown Australia

Not ESA612-2016,BJ2ECG ADAPTER ASSY Shown

3359538

6-5

ESA612 Service Manual

H1 (6x) MP3 MP15

MP14 MP13

MP4 H3

MP10, H5 (2x), H6 (2x) A2 C1 MP2

MP5

F2 F1

MP12

C2 MP7 (5x)

H2 (7x) C3 H3 (3x)

H2 (2x)

H2

A1

MP6

MP11 MP8 MP9

MP1

Figure 6-1. Final Assembly

6-6

gjc145.eps

Replaceable Parts Parts List

6

Table 6-2. A1 Power PCA

Item

Description

Fluke Stock No.

C2 C4 C6 C7 C11 C14 C17 C19-26 CAPACITOR,CERAMIC,0.1UF,+-10%,50V,X7R,0805,TAPE C32 C35 C36 C41 C43

690500

C3 C12 C15 C18 C28 C53 C55 C60

CAPACITOR,ELECTROLYTIC,TANTALUM,10UF,+-20%,16V,3216,TAPE

3093654

C5 C38-40

CAPACITOR,CERAMIC,1UF,+-20%,25V,X5R,0805,TAPE

2530275

C8 C9 C37

CAPACITOR,CERAMIC,0.047UF,+-10%,50V,X7R,0805,TAPE

1708909

C10

CAPACITOR SMR,CAP,CER,4700PF,+-10%,50V,X7R,0805

604231

C13 C54

CAPACITOR,CERAMIC,0.01UF,+-10%,50V,X7R,0805,TAPE

106146

C27 C74

CAPACITOR,CERAMIC,10PF,+-5%,100V,C0G,0805,TAPE

3331432

C29 C42

CAPACITOR,CERAMIC,0.047UF,\+-10%,500V,X7R,1812,TAPE

644077

C30 C33 C44 C57 C73

CAPACITOR SMR,CAP,CER,1000PF,+-10%,50V,X7R,0805

484378

C31

CAPACITOR SMR,CAP,CER,470PF,+-1%,50V,C0G,0805

929476

C34

560180002224,CAP-SM,0805,16V,10%,X7R,.22UF

2331299

C45-52 C56 C58 C59 C61-71

CAPACITOR SMR,CAP,CER,100PF,+-1%,50V,C0G,0805

867650

C72

CAPACITOR,ELECTROLYTIC,TANTALUM,47UF,+-10%,16V,200MOHM ESR,7343,TAPE

3352434

CR1

ZENER,UNCOMP,MMBZ5235B,6.8V,5%,20MA,225MW,SOT-23,TAPE

837195

CR2 CR3 CR11 CR19

ZENER,UNCOMP,MMBZ5242B,12V,5%,20MA,225MW,SOT-23,TAPE

2044054

CR4

LED,RED,AM2520,280MCD,20MA,2.5V,20 DEG VIEW ANGLE,GULL WING PREP,T3/4,TAPE

2281407

CR5 CR10

DIODE,SI,PN,S1M,1KV,1A,75NS,SMA,TAPE

2087962

CR6 CR1618 CR20

DIODE,SI,BAS16,75V,200MA,6NS,SOT-23,TAPE

830489

CR7 CR1215

DIODE,SI,PN,S1M,1KV,1A,1.8US,DO-214AC,TAPE

2646221

F3

FUSE,2.5A,250V,TIME LAG,8X8.5MM,RADIAL

3331539

H1 H2

SPACER,ROUND,4.2MM ID,2.0MM HEIGHT,8.74MM OD,STEEL,TIN,SMD,TAPE

3350760

H3 H4

SPACER,ROUND,3.6MM ID,2.0MM HEIGHT,7.14MM OD,STEEL,TIN,SMD,TAPE

3352172

J1

CONNECTOR,TERMINAL BLOCK,6.35MMCTR,VERT PWB,30A,300V,3 POS

3342141

J2

CONNECTOR,SOCKET,2 ROW,3.0MM CTR,VERT SMD,6 POS,TAPE

3343846

6-7

ESA612 Service Manual

Table 6-2. A1 Power PCA (cont.)

Item

Description

J4

CONNECTOR,HEADER,2 ROW,0.1 CTR,VERT SMT,SHROUDED,20 POS,TAPE

2791599

J5

CONNECTOR,HEADER,1 ROW,.25 CTR,RT ANG PWB,LOCKING,W/EARS,5 POS,BULK

3351138

K1-7 K11

RELAY,ARMATURE,1 FORM C,12V,1 C/O CONTACT.16A,29X12.7MM,BULK

2152365

K8 K9 K13 K17

RELAY,ARMATURE,2 FORM C,12VDC,SINGLE STABLE,15X7.4MM,TAPE

2082903

K10 K12 K14 RELAY,ARMATURE,2 FORM C,12V,15X9MM,SMT,TAPE K16

6-8

Fluke Stock No.

3301088

L1

TRANSFORMER,SIGNAL,200:1,DIFFERENTIAL,SHIELDED,VERTICAL TOROID,26X27X11MM,4-PIN,BULK

3405268

L3

TRANSFORMER,SIGNAL,CURRENT SENSOR,2-20A,100 OHM,50/60HZ,3PIN,BULK

2750107

Q1

MOSFET,SI,N,IRFBF20S,900V,1.7A,8 OHMS,3.1W,D2-PAK,TAPE

2742520

Q2

TRANSISTOR,SI,NPN,MMBT5089,30V,50MA,50MHZ,200MW,SOT-23,TAPE

820902

Q3

MOSFET,SI,N,IRLR3410,100V,17A,105 MOHMS,3W,D-PAK,TAPE

1579273

Q4

MOSFET,SI,N,FQB5N90,900V,5.4A,2.3 OHMS,3.13W,D2-PAK,TAPE

3343629

Q5 Q6

MOSFET,SI,N,2N7002E,60V,240MA,3 OHMS,350MW,SOT-23,TAPE

1756473

Q7-9

TRANSISTOR,SI,NPN,MUN2216,50V,100MA,DIGITAL,R1=4.7K,R2=INF,230M 2002486 W,SC-59,TAPE

R2-7 R23 R24 R27

RESISTOR,METAL FILM,330K,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE

3326205

R9 R16 R20 R43 R58

RESISTOR,CERMET,100,+-1%,0.125W,100PPM,0805,TAPE

928937

R12

RESISTOR,CERMET,10,+-1%,0.125W,100PPM,0805,TAPE

928924

R13

RESISTOR,CERMET,16.2K,+-1%,0.125W,100PPM,0805,TAPE

2005953

R14 R15 R25 R30 R31 R33 R34 R39 R44 R54 R62 R65 R69-74 R7983 R88-91 R96-99 R101-113

RESISTOR,CERMET,1K,+-1%,0.1W,100PPM,0805,TAPE

928713

R17 R18 R21 R22

RESISTOR,CERMET,6.2K,+-1%,0.125W,100PPM,0805,TAPE

3336652

R19 R52 R67 R68 R128

RESISTOR,CERMET,100K,+-1%,0.125W,100PPM,0805,TAPE

928866

R26

RESISTOR,CERMET,7.5K,+-1%,0.125W,100PPM,0805,TAPE

2278492

R10

RESISTOR SMR,RES,CERM,332,+-1%,0.1W,100PPM,0805

604926

R11 R100

RESISTOR,CERMET,33,+-1%,0.125W,100PPM,0805,TAPE

2803420

Replaceable Parts Parts List

6

Table 6-2. A1 Power PCA (cont.)

Item

Description

Fluke Stock No.

R29

RESISTOR SMR,RES,CERM,1.82K,+-1%,0.1W,100PPM,0805

930172

R32 R37 R49 R51 R55 R61

RESISTOR,CERMET,10K,+-1%,0.125W,100PPM,0805,TAPE

928791

R41 R56 R57

RESISTOR,METAL FILM,1M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE

3326231

R42 R53 R59 R76 R85 R86

RESISTOR,CERMET,1,+-1%,0.125W,400PPM,0805,TAPE

2005982

R45

RESISTOR,CERMET,56K,+-1%,0.125W,100PPM,0805,TAPE

3331360

R46

RESISTOR,CERMET,11.5K,+-1%,0.1W,100PPM,0805,TAPE

1997330

R47

RESISTOR,CERMET,1M,\+-1%,0.125W,100PPM,0805,TAPE

928945

R48

RESISTOR,CERMET,4.7K,+-1%,0.125W,100PPM,0805,TAPE

3326333

R50

RESISTOR SMR,RES,CERM,332K,+-1%,0.1W,100PPM,0805

602695

R60

RESISTOR,CERMET,470,+-1%,0.125W,100PPM,0805,TAPE

2803447

R63 R64

RESISTOR,CERMET,68K,+-5%,1W,200PPM,2512,TAPE

3331385

R66

RESISTOR,CERMET,23.2K,+-1%,0.125W,100PPM,0805,TAPE

3331397

R77

RESISTOR,CERMET,6.81K,+-1%,0.1W,100PPM,0805,TAPE

2087241

R78

RESISTOR,CERMET,3.6K,+-1%,0.125W,100PPM,0805,TAPE

3331415

R92-95

RESISTOR,CERMET,200,+-1%,0.125W,100PPM,0805,TAPE

3372234

R114-118

RESISTOR,CERMET,2M,+-5%,1W,200PPM,2512,TAPE

3310288

R119-127

RESISTOR,CERMET,JUMPER,0,+0.05 MAX,0.125W,1206,TAPE

810747

RV1 RV2

VARISTOR

706838

RV40 RV75

THERMISTOR,POSITIVE,1.1K,+-20%,COATED,RADIAL,TAPE

1277360

T1

TRANSFORMER,POWER,115/230V,50/60HZ,[email protected],0.51VA,EI30/1 2.5,PC MOUNT,BULK

3342134

T2

TRANSFORMER,FLY-BACK,6V,20KHZ,1:10:10,1.5W,HV,FLUKE165X,RM6,SMT8,TAPE

2032204

TP12 TP29 TP33-37

CONNECTOR,TERMINAL,TEST POINT,SMD,510 PH BRONZE,TAPE

602125

U1

IC,OP AMP,AD8510,+-4.5V TO +-18V,0.9MV OFFSET,8MHZ,PRECISION,JFET,S08,TAPE

2434633

U2 U5

IC,DIGITAL ISOLATOR,ADUM1401,QUAD,3/1,10MHZ,2500V,3V/5V,SOICW16,TAPE

3331444

U3

IC,ADC,MCP3202,2.7-5.5V,12BIT,2-CHANNEL,SPI,S08,TAPE

3331459

U4

IC,OP AMP,TLV2372,2.7V TO 16V,4.5MV OFFSET,3MHZ,DUAL,R/R,S/S,SO8,TAPE

3329934

U6 U10 U11 U19-21 U29

OPTICAL,OPTOCOUPLER,PHOTO TRANSISTOR,SMD DIP4,TUBE

3328473

U7

IC,COMPARATOR,LMV7271,1.8V TO 5V,4MV OFFSET,R/R INPUT,LO PW,800NS PD,SOT-23-5,TAPE

3332950

U8

IC,ANALOG SWITCH,ADG431,5-44V,24 OHMS,QUAD,SPST,NO,SO16,TAPE

3367437

R05R,VARISTOR,430V,+-10%,1MA

6-9

ESA612 Service Manual

Table 6-2. A1 Power PCA (cont.)

Item

6-10

Description

Fluke Stock No.

U9

POWER SUPPLY,SW,15W,85-264VAC,[email protected],32X64MM,PCB MOUNT,BULK

3331480

U12 U28

IC,VOLTAGE REGULATOR,LINEAR,LP2985,3.3V,150MA,LDO,LOW POWER,SOT-23-5,TAPE

2559443

U13 U15

TRANSISTOR,SI,NPN,ULN2003,50V,500MA,HEPTADARLINGTON,3.5W,SO16,TAPE

2113871

U14 U16 U25

IC,LOGIC,74HC595,2.0V-6.0V,8-BIT SHIFT REGISTER,TRISTATE,SOIC16,TAPE

904388

U17

IC,PULSE WIDTH MODULATOR,UCC3805,4V,35MA,1MHZ,10V SUPPLY,SO8,TAPE

1541378

U18

POWER SUPPLY,DC-DC,2W,12VDC,+-5V@200MA,3KV ISO,SIP5,TUBE

3338710

U22

603022-008,VOLTAGE REF 1.2V LM4041CM3

1790839

U23 U24

IC,OP AMP,TLV2371,2.7V TO 16V,4.5MV OFFSET,3MHZ,R/R,S/S,SOT-235,TAPE

3331500

U26 U27

IC,LOGIC,NC7SZ32,1.65V-5.5V,SINGLE OR GATE,NON-INV,SOT23-5,TAPE

1604581

VR2 VR3

TVS DIODE,SMA6J6.5A,6.5V,5%,UNIPOLAR,SMA,TAPE

3385450

W1

WIRE,PVC,UL1015,600V,105C,14AWG,STRANDED,BLU

2198323

W2

WIRE,PVC,UL1015,14AWG,STRANDED,INDIVIDUAL TINNED,BRN

2148907

XAF1 XAF2 XBF1 XBF2

FUSE CLIP,FUSE CLIP PCB

756460

Replaceable Parts Parts List

6

Assembly Top View (CKT 1)

Assembly Bottom View (CKT 4)

ESA612-4002

Figure 6-2. A1 Power PCA

gjc134.eps

6-11

ESA612 Service Manual

Table 6-3. A2 Meter PCA

Item

6-12

Description

Fluke Stock No.

C1 C73 C103 C116 C132 C157 C194

CAPACITOR SMR,CAP,CER,1000PF,+-10%,50V,X7R,0805

484378

C2

CAPACITOR,CERAMIC,0.33UF,+-10%,16V,X7R,0805,TAPE

1745562

C3-8 C10 C11 C14-21 C35-45 C47 C48 C63 C64 C69 C70 C72 C74 C75 C79-83 C93 C94 C98 C99 C105 C107-112 C114 C119 C123 C124 C126 C127 C129-131 C133 C135137 C141145 C155 C163-165 C167 C168 C170 C174 C177-179

CAPACITOR,CERAMIC,0.1UF,+-10%,50V,X7R,0805,TAPE

690500

C9

CAPACITOR,CERAMIC,4.7UF,+-10%,10V,X7R,0805,TAPE

3092853

C12

CAPACITOR,CERAMIC,2.2UF,+-10%,10V,X5R,0805,TAPE

3093102

C13

CAPACITOR SMR,CAP,CER,470PF,+-1%,50V,C0G,0805

929476

C22 C46 C50 C57 C58 C60 C61 C104 C117 C122 C146-150 C171

CAPACITOR,ELECTROLYTIC,TANTALUM,10UF,+-20%,16V,3216,TAPE

3093654

C24 C25

CAPACITOR SMR,CAP,CER,22PF,+-5%,200V,C0G,0805

643459

C26 C27 C29-34 C51 C62 C67 C106 C113 C115 C120 C121 C169 C189-191

CAPACITOR,CERAMIC,1UF,+-20%,25V,X5R,0805,TAPE

2530275

C28 C96 C97

CAPACITOR SMR,CAP,CER,4.7PF,+-0.25PF,50V,C0G,0805

806760

C49 C52 C118 C139

CAPACITOR,CERAMIC,0.01UF,+-10%,50V,X7R,0805,TAPE

106146

Replaceable Parts Parts List

6

Table 6-3. A2 Meter PCA (cont.)

Item

Description

Fluke Stock No.

C53-56 C134 C140 C151154 C156 CAPACITOR SMR,CAP,CER,100PF,+-1%,50V,C0G,0805 C159-162 C166

867650

C65

560380008223,CAP-SM,0805,50V,10%,X7R,.022UF

2332024

C68

560380008333,CAP-SM,.033UF,50V,10%,X7R,0805

2332036

C71 C76

CAPACITOR SMR,CAP,CER,4700PF,+-10%,50V,X7R,0805

604231

C77 C78 C87 C88 C90 C91

CAPACITOR,CERAMIC,0.47UF,+-5%,16V,X7R,0805,TAPE

3341246

C86 C92

CAPACITOR,CERAMIC,0.047UF,+-5%,50V,X7R,0805,TAPE

3341222

C89

CAPACITOR,CERAMIC,0.1UF,+-5%,50V,X7R,0805,TAPE

3349793

C95 C125

CAPACITOR,CERAMIC,15PF,+-5%,50V,C0G,0805,TAPE

514174

C101 C102

CAPACITOR,CERAMIC,33PF,+-5%,50V,C0G,0805,TAPE

603172

C175 C193

CAPACITOR,CERAMIC,0.027UF,+-10%,1000V,X7R,1812,TAPE

1579931

C180 C181 C187 C188

CAPACITOR,CERAMIC,100PF,+-5%,50V,C0G,0402,TAPE

2813217

C182-186

CAPACITOR,CERAMIC,0.1UF,+-20%,16V,X7R,0402,TAPE

2812858

CR1 CR8 CR15 CR17

DIODE,BAT54A SMR,DIODE,SI,SCHOTT,DUAL,30V,200MA,SOT-23

942594

CR2 CR5 CR10 CR11

DIODE,SI,BAS16,75V,200MA,6NS,SOT-23,TAPE

830489

CR3 CR4

DIODE,MBR0530 SMR,DIODE,SI,SCHOTTKY,30V,0.5A,SOD-123

691717

CR6 CR2029 CR33-37

TVS DIODE,SMAJ440CA,440V,5%,BIPOLAR,SMA,TAPE

2636264

CR9 CR41

DIODE,SI,PN,MMBD1503A,150V,200MA,DUAL,SERIES,LOW LEAKAGE,SOT-23,TAPE

928143

CR18

TVS DIODE,SMAJ5.0CA,5V,5%,BIPOLAR,SMA,TAPE

2073208

CR19 CR3032 CR38

VARISTOR,0603 3.6 V VARISTOR (CLAMP @ 10 V), VC060303A100

2764916

CR42-46

DIODE,SI,PN,BAV199,70V,215MA,3US,DUAL,SERIES,SOT-23,TAPE

605805

DS1

LED,RED,AM2520,280MCD,20MA,2.5V,20 DEG VIEW ANGLE,GULL WING PREP,T3/4,TAPE

2281407

F1

THERMISTOR,PTC,50,+-20%,300V,3A MAX,8MM RADIAL,BULK

3339126

H1

CABLE ACCESSORY ,CABLE ACCESS,TIE,5.50L,.10W,1.25 DIA

530360

J1

CONNECTOR,FLAT-FLEX,RECEPTACLE,SINGLE SIDE,0.5MM PITCH,VERTICAL SMD,20 POS,TAPE

2739499

J2 J21

CONNECTOR,FLAT FLEX,1MMCTR,VERTICAL SMD,ZIF,LOCKING,10 POS,TAPE

3408166

J3

CONNECTOR,MICRO-RIBBON, USB,SERIES B, RECEPTACLE,RT ANG PWB,4 POS,BULK

1541856

6-13

ESA612 Service Manual

Table 6-3. A2 Meter PCA (cont.)

Item

6-14

Description

Fluke Stock No.

J4

CONNECTOR,HEADER,2 ROW,0.1 CTR,VERT SMT,SHROUDED,20 POS,TAPE

2791599

J20

CONNECTOR,SOCKET,2 ROW,3.0MM CTR,VERT SMD,6 POS,TAPE

3343846

J60-64

CONNECTOR,HOLDER,FUSE,5MM DIA,RT ANG SURFACE MOUNT,TAPE

1620687

L1 L2

INDUCTOR,10UH,20%,1.8ADC,140MOHMS,6X6X3MM,SMT,TAPE

3338177

L3-10

INDUCTOR,BEAD,95 OHMS@100MHZ,1ADC,1MOHM,3612,TAPE

867734

LS1

AUDIO TRANSDUCER,MAGNETIC,SOUNDER,2.4KHZ,5V,TOP PORT,13MM SQ,TAPE

690271

P12 P13 P24 ESA612-8003,STRAP, INPUT

3258998

Q1

MOSFET,SI,N,2N7002,60V,115MA,7.5 OHMS,225MW,SOT-23,TAPE

927538

Q2-4

NPN,MMBT3904 SMR,TRANSISTOR,SI,NPN,60V,350MW,SOT-23

742676

Q7

MOSFET,SI,P,FDN338P,20V,1.6A,115 MOHMS,500MW,SOT-23,TAPE

2538867

Q8

TRANSISTOR,SI,NPN,MMUN2211,50V,100MA,DIGITAL,R1=10K,R2=10K,246 3331929 MW,SOT-23,TAPE

R2

RESISTOR,CERMET,130,+-1%,0.125W,100PPM,0805,TAPE

690644

R3 R5 R1115 R33 R4043 R50 R7072 R74 R82 R90 R91 R102 R119 R133 R135140 R146 R147 R168171 R194 R195 R243245 R249257 R260 R404 R405 R408 R412 R413

RESISTOR,CERMET,10K,+-1%,0.125W,100PPM,0805,TAPE

928791

R4 R21 R23 R36-39 R4548 R77 R95 R96 R118 R129 R131 R132 R134 R141 R148 R149 R172 R190-193 R196-199 R203 R215 R218 R225 R230-232 R236-239 R246-248 R270 R271 R345 R346

RESISTOR,CERMET,1K,+-1%,0.1W,100PPM,0805,TAPE

928713

Replaceable Parts Parts List

6

Table 6-3. A2 Meter PCA (cont.)

Item

Description

Fluke Stock No.

R7 R29-32 R73 R144 R150 R157167 R206214 R217 R233-235 R240 R407

RESISTOR,CERMET,100,+-1%,0.125W,100PPM,0805,TAPE

928937

R16 R20 R76 R80 R81 R86 R92 R117 R121 R142 R143 R145 R179 R180 R223

RESISTOR,CERMET,100K,+-1%,0.125W,100PPM,0805,TAPE

928866

R19

RESISTOR,CERMET,2M,+-1%,0.125W,100PPM,0805,TAPE

3326041

R22

RESISTOR,CERMET,24K,\+-1%,0.125W,100PPM,0805,TAPE

2499314

R24 R25 R27 R28

RESISTOR,CERMET,121K,+-1%,0.125W,100PPM,0805,TAPE

801560

R26 R53

RESISTOR,CERMET,1.5M,+-1%,0.125W,100PPM,0805,TAPE

3326090

R34 R35

RESISTOR SMR,RES,CERM,150,+-1%,0.1W,100PPM,0805

930086

R44

RESISTOR,CERMET,16.2K,+-1%,0.125W,100PPM,0805,TAPE

2005953

R49 R226229

RESISTOR,CERMET,3.01K,+-1%,0.125W,100PPM,0805,TAPE

1591278

R51

40034,RES,9.53K,1/10W,1%,0805

2499039

R52

RESISTOR SMR INAC,RES,CERM,13.7K,+-1%,0.1W,100PPM,0805

928812

R54

576186043120,RES-SM,0805,1/10W,1%,TF,604K

2338933

R55

RESISTOR,CERMET,732K,+-1%,0.125W,100PPM,0805,TAPE

1727162

R56

RESISTOR,CERMET,487K,+-1%,0.1W,100PPM,0805,TAPE

1665340

R57-61

RESISTOR,CERMET,499,+-1%,0.125W,100PPM,0805,TAPE

928978

R62 R104

RESISTOR,CERMET,30.1K,+-1%,0.125W,100PPM,0805,TAPE

2005994

R63 R182

RESISTOR SMR,RES,CERM,11K,+-1%,0.1W,100PPM,0805

928796

R64-66 R98100

RESISTOR,METAL FILM,330K,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE

3326205

R67

RESISTOR,CERMET,6.81K,+-1%,0.1W,100PPM,0805,TAPE

2087241

R68

RESISTOR,WIREWOUND,10K,+-1%,3W,20PPM,AXIAL,TAPE

2076480

R75

RESISTOR,CERMET,4.99K,+-1%,0.125W,100PPM,0805,TAPE

928767

R78

RESISTOR,WIREWOUND,800,+-1%,5W,20PPM,AXIAL,TAPE

2741174

R79

RESISTOR,WIREWOUND,200,+-1%,3W,20PPM,AXIAL,TAPE

2741195

R83-85

RESISTOR,METAL FILM,1M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE

3326231

R87-89 R127 RESISTOR,CERMET,150K,+-1%,0.125W,100PPM,0805,TAPE

1612759

6-15

ESA612 Service Manual

Table 6-3. A2 Meter PCA (cont.)

Item

6-16

Description

Fluke Stock No.

R93 R94 R101 R103

RESISTOR,CERMET,3.92K,+-1%,0.125W,100PPM,0805,TAPE

1591284

R97

RESISTOR,CERMET,8.06K,+-1%,0.125W,100PPM,0805,TAPE

928788

R105

RESISTOR,CERMET,360K,+-1%,0.125W,100PPM,0805,TAPE

3326268

R106

RESISTOR,CERMET,43.2K,+-1%,0.125W,100PPM,0805,TAPE

2499017

R107

RESISTOR,CERMET,432K,+-1%,0.125W,100PPM,0805,TAPE

1641918

R108 R110 R112 R122 R124 R154

RESISTOR,CERMET,2.43K,+-1%,0.125W,100PPM,0805,TAPE

928754

R109 R152

RESISTOR,CERMET,931,+-1%,0.125W,100PPM,0805,TAPE

3339167

R111 R113 R123 R153

RESISTOR,CERMET,1.1K,+-1%,0.125W,100PPM,0805,TAPE

691014

R114 R115 R125 R126

RESISTOR,CERMET,51.1K,+-1%,0.125W,100PPM,0805,TAPE

2011360

R116

RESISTOR,CERMET,4.7K,+-1%,0.125W,100PPM,0805,TAPE

3326333

R120

RESISTOR,CERMET,23.2K,+-1%,0.125W,100PPM,0805,TAPE

3331397

R128 R347349

RESISTOR,CERMET,10M,+-1%,0.125W,400PPM,0805,TAPE

943659

R156

RESISTOR,CERMET,100,+-1%,0.1W,100PPM,0603,TAPE

642196

R173-175 R409-411

RESISTOR,METAL FILM,3.32M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE

3408749

R176

RESISTOR,CERMET,39.2K,+-1%,0.1W,100PPM,0805,TAPE

943092

R177 R178 R185 R186 R200 R201 R204 R205 R219 R242 R352

RESISTOR,CERMET,33,+-1%,0.125W,100PPM,0805,TAPE

2803420

R18 R151 R187-189 R220 R221

RESISTOR,CERMET,10,+-1%,0.125W,100PPM,0805,TAPE

928924

R181 R258

RESISTOR,CERMET,1M,\+-1%,0.125W,100PPM,0805,TAPE

928945

R183 R184

RESISTOR,CERAMIC COMPOSITION,27,+-10%,2W,-1300 +300PPM,AXIAL,TAPE

3454161

R216

RESISTOR,CERMET,51.1,+-1%,0.125W,100PPM,0805,TAPE

2005930

R350 R351

RESISTOR SMR,RES,CERM,249,+-1%,0.1W,100PPM,0805

606092

R353 R354 R362 R363 R366-368 R373 R374 R379 R388 R389 R392 R395-399

122421,RES - 102R, 1%, 1/16W, 0402, PB-FREE

3093309

R355-358 R401

RESISTOR,CERMET,JUMPER,0,+0.05 MAX,0.125W,0805,TAPE

928705

Replaceable Parts Parts List

6

Table 6-3. A2 Meter PCA (cont.)

Item

Description

Fluke Stock No.

R359-361 R364 R365

RESISTOR,CERMET,10K,+-1%,0.063W,100PPM,0402,TAPE

1706323

R370-372 R406

RESISTOR SMR,RES,CERM,261K,+-1%,.125W,100PPM,1206

877014

R375-378 R380-387 R390 R391 R393 R394 R402

RESISTOR,CERMET,499,+-1%,0.063W,100PPM,0402,TAPE

3254825

RT1

THERMISTOR,NTC,10K,+-5%,NON-LINEAR,0805,TAPE

2030909

RV1

VARISTOR,VARISTOR,430V,+-10%,1.0MA

105871

TP86 TP87 TP89-92 TP94 TP95 CONNECTOR,TERMINAL,TEST POINT,SMD,510 PH BRONZE,TAPE TP97 TP124127

602125

U1

IC,MEMORY,FLASH,45DB161,16MB,2MX8,2.73.6V,20MHZ,SERIAL,SPI,SO8,TAPE

3328276

U2

IC,MICROPROCESSOR,AT32UC3A,1.65-1.95V,32-BIT,66MZH,512K FLASH,64K SRAM,LQFP144,TAPE

3337006

U3

IC,LCD CONTROLLER,3.3V,32K SRAM,8-BIT,TQFP64

3328311

U4

OSCILLATOR,HCMOS CLOCK,40MHZ,3.3V,100PPM,15PF LOAD,TRISTATE,3.2X2.5MM SMT,TAPE

3328327

U5

IC,VOLTAGE REGULATOR,SWITCHING,1615-1,ADJ,100MA,STEP UP,SOT23-5,TAPE

1791133

U6

IC, OP AMP,LP324,+-3V TO +-32V,+-2MV OFFSET,100KHZ,QUAD,BIPOLAR,SOIC14,TAPE

2841280

U7

IC,COMM,FT232RL,3.3V-5.25V,USB TO RS232 UART,SSOP28,TAPE

2670615

U8 U42

IC,DIGITAL ISOLATOR,ADUM1402,QUAD,2/2,2MHZ,2500V,3V/5V,SOICW16,TAPE

2528074

U9

IC,COMM,SN75240,USB TRANSIENT SUPPRESSOR 15KV,7VBR,TSSOP8,TAPE

2044695

U10

IC,VOLTAGE REGULATOR,LINEAR,LP2980,3V,50MA,LDO,LO PWR,W/SHUT 3328412 DOWN,SOT-23-5,TAPE

U11

IC,VOLTAGE REGULATOR,SWITCHING,LT1616,ADJ,600MA,1.4MHZ,STEP DOWN,SOT-23-6,TAPE

3328447

U12 U43

OPTICAL,OPTOCOUPLER,PHOTO TRANSISTOR,SMD DIP4,TUBE

3328473

U13 U15 U54

IC,OP AMP,AD8510,+-4.5V TO +-18V,0.9MV OFFSET,8MHZ,PRECISION,JFET,S08,TAPE

2434633

U14

IC,DIGITAL,SUPERVISORY RESET IC,ACTIVE HIGH,SOT23-3,TAPE

3328499

U16 U21

IC,DAC,TLV5623,2.7V-5.5V,8-BIT,SPI,VOUT,SOIC8,TAPE

2761956

U17

589085015001,IC-SM,LMV331,SGL COMPTR,SOT235

2343218

U18

IC,COMPARATOR,LM393,+-1V TO +-18V,5MV OFFSET,DUAL,LOW PWR,SO8,TAPE

837211

6-17

ESA612 Service Manual

Table 6-3. A2 Meter PCA (cont.)

Item

6-18

Description SMR,IC,CMOS,QUAD INPUT NOR GATE,SOIC

Fluke Stock No.

U19

CMOS 74HC02

U22

IC,OP AMP,TLV2372,2.7V TO 16V,4.5MV OFFSET,3MHZ,DUAL,R/R,S/S,SO8,TAPE

3329934

U23

IC,RMS CONVERTER,LTC1967,5V,LOW POWER,R/R INPUT,MS0P8,TAPE

3385912

U24

IC,ADC,AD7799,2.7V-5.25V,24-BIT,3-CH,500HZ,SIGMADELTA,SERIAL,TSSOP16,TAPE

2630817

U25 U27 U39

IC,ANALOG SWITCH,ADG431,5-44V,24 OHMS,QUAD,SPST,NO,SO16,TAPE

3367437

U26 U28

IC,OP AMP,AD8034,5V TO 24V,2MV OFFSET,80 MHZ,DUAL,FET INPUT,SO8,TAPE

2558152

U29 U31

IC,OP AMP,AD8539,2.7 TO 5.5V,15UV OFFSET,430KHZ,R/R,S/S,DUAL,S08,TAPE

3385920

U30

CONTRACT MFG ITEM, IC 2.5V REF LM4040CIM3-2.5

2073591

U32 U34 U37 U38 U55

IC,LOGIC,74HC595,2.0V-6.0V,8-BIT SHIFT REGISTER,TRISTATE,SOIC16,TAPE

904388

U33 U35 U36 U56

TRANSISTOR,SI,NPN,ULN2003,50V,500MA,HEPTADARLINGTON,3.5W,SO16,TAPE

2113871

U40

IC,MICROCONTROLLER,LPC2138,3.3V,16/32 BIT,512KB FLASH,ARM7,25MHZ,LQFP64,TAPE

3329792

U41

IC,VOLTAGE MONITOR,MAX803,VTH=2.93V,MICROPOCESSOR RESET CIRCUIT,OPEN-DRAIN OUTPUT,SOT23-3,TAPE

3350940

U44

IC,MEMORY,EEPROM,93LC66A,4KB,256X16,2.55.5V,2MHZ,SERIAL,SO8,TAPE

3344058

U45 U46

CMOS 74HC00

830703

U47

CMOS 74HC132 SMR,IC,CMOS,QUAD 2 IN NAND W/SCHMT,SOIC

837245

U48

POWER SUPPLY,DC-DC,2W,12VDC,12V@167MA,3KV ISO,SIP4,TUBE

3338731

U49

POWER SUPPLY,DC-DC,2W,12VDC,+-12V@83MA,3KV ISO,SIP5,TUBE

3338746

U50

IC,VOLTAGE REGULATOR,LINEAR,LP2985,3.3V,150MA,LDO,LOW POWER,SOT-23-5,TAPE

2559443

U51

POWER SUPPLY,DC-DC,2W,12VDC,5V@400MA,3KV ISO,SIP4,TUBE

3338754

U52

589480084001,IC-SM,MC79L05A,-5V REG,8-SOIC

2343646

U53

BIPOLAR 78L05 SMR,IC,VOLT REG,FIXED,5 VOLTS,0.1 A,SOIC8

604105

U59

IC,MEMORY,FRAM,FM25L256B,256KB,32KX8,2.7-3.6V,SERIAL,SO8,TAPE

3374127

U60

IC,VOLTAGE CONVERTER,LTC1144,SWITCH CAP,WIDE INPUT RANGE,SO8,TAPE

2609823

VR3 VR4

TVS DIODE,SMBJ15A,15V,5%,UNIPOLAR,SMB,TAPE

2813982

VR6 VR7

ZENER,UNCOMP,MMBZ5236B,7.5V,5%,20MA,225MW,SOT-23,TAPE

837138

VR12-14 VR16

ZENER,UNCOMP,MMBZ5242B,12V,5%,20MA,225MW,SOT-23,TAPE

2044054

W2

WIRE,PVC,UL1015,600V/105C,14AWG,STRAND,INDIVIDUAL TIN,GRN/YEL

2279979

Y1 Y2

CRYSTAL,12MHZ,50/100PPM,20PF,PLASTIC ENCAPSULATED,SMD,TAPE

2041235

SMR,IC,CMOS,QUAD INPUT NAND GATE,SOIC

830711

Replaceable Parts Parts List

6

ESA612-4001 Figure 6-3. A2 Meter PCA

gjc133.eps

6-19

ESA612 Service Manual

6-20

Chapter 7

Schematics

Title

Page

Introduction.......................................................................................................... 7-3

7-1

ESA612 Service Manual

7-2

Schematics Introduction

7

Introduction The schematic and pca layout diagrams for the ESA612 are shown in Figures 7-1 through 7-3.

7-3

ESA612 Service Manual

7-4

Electrical Safety Analyzer Schematic Diagrams

Assembly Top View (CKT 1)

6

Assembly Bottom View (CKT 4)

ESA612-4002

Figure 7-1. A1 Power PCA

7-5

ESA612 Service Manual

RV2

706838 1 2 1MA

XAF2

XBF2

12VDC

XAF1

8

2152365

E1

GND

2.5A 250V

CHGND

3C6>

CRL1+2

VSS

7 2 5 4

1

3331539

CRL3

K3

3 6 12VDC

1

IN

IN

3C6>

K2

430V 2

+12V

2

F3

8

2152365

12VDC

1

8

LOAD+

LOAD3326205

R2

1% 330K 3326205

9

3

+12V CRL8

IN

3326205

3326205

MELF0204

6.2K

6.2K

5

X7R

R18

6.2K

3336652

C9

6

0.047U

+ -

1708909

U4

X7R

V4

GND 3A6>

7

SO8

3

R17

C8

0.047U

1708909

2

+ -

SO8

V4

IN

GND

12VDC

1

DIP8

2152365

IN

8

12

+12V CRL12

TP19

10

9

8 3

4

1 TX2SA 2082903

IN

1

CLK 7 CS 1

ADC_DOUT ADC_DCK

OUT

ADC_/CS

IN

IN

V+

5

SOT23-5

+12V

K10

MAP_RTN 500V

CRL10

IN

2C1>

R12

3C2> 4D2>

GNDMS

10.0

+12VM

3

1U

R128

100K

2

+

GND

2

GND

C6

C19

0.1U

-

7 V+

U1

V4

AD8510AR

GNDMS

LMV7271

3332950 C10

0.0047U

R19

100K

928866

OUT

1

1C4>

+12V

+3.3VMD

VDD

VL

SO16

0.1U

3

GNDMS

1

/EQU_CUR_ASW

IN

VSS GND

4

5

-12VM

GNDMS

PWR GRP TP29

GNDMS

C7

0.1U

-12VM

690500 X7R

13

R9

GNDMS

100

10P

3B6>

2

C21

690500

2434633

C27

3B6>

690500

GNDMS

6

SO8

928866

3A6


IN

3D1>

OUT

3B2>

5

GFI_SENSE

SOLID STATE RELAY SPACERS UNNAMED_1_RLY1C1COIL_I12_W1

U7 OUT

8

13 12

U8 ADG431B 3367437

0.1U

UNNAMED_1_RLY1C1COIL_I12_NO2

MAINS_ZC

METER

10

0.1U

OUT

SO8

4

TP38

+12VM

C20

GNDMS 6

3 5

12

METER CIRCUITS

TP20

5

+12V

K14

CRL9

3C6>

CRL14

IN

1

12VDC

12VDC

+12V

3D8>

3B2> 4A3


IN

TP21

12VDC

8

4

5

12

+12V

IN

9

K10

OUT

TP17

400MA_RTN

10

9

8

3 5

9 METER

+3.3VPD

928791

K12 4

3 6

3D6>

MAINS ZERO CROSSINGS

3

CRL11

GNDMS

GND

928791

R32

IN

CHGND

+12V

C5

DOUT

R37

1

3D6>

928937

3

4

10.0K

3329934

100

7 2 5 4

X7R

VSS

10.0K

GND

3C6>

3331459

TLV2372

BAS16

TERM-BLOCK

3342141

K11

2

R16

830489

CR18 SOT23

P

5304

690500

3 CH1

1

BAS16

1

830489

8 U3 VDD/VREF

2 CH0

U4

CR17 SOT23

1

GND

MCP3202

3329934

GND

8 V+

DIP8

1

200:1 3

1

332

5 DIN

ADC_DIN

TLV2372

+3.3VPD

GND

4

8

0.1U

GND 8 V+

12VDC

1

D

O

R10

690500

0.1U

+3.3VPD

3 6

WP4 W2 L3 2750107

L1

GNDMS

C2

C17

330K 1%

CRL6+7

BROWN

+3.3VPD

+3.3VPD

R22

6.2K

AP_METER

IN

K9

TP18

12VDC

TX2SA 2082903

MAINS VOLTAGE A TO D

R5

K7

2152365

1

3326205

R21

3336652

5

12

330K 1%

330K 1%

10 8 3

4

R7

X7R

+3.3VPD

K8

3C2>

R6

IN

+12V

J1

3 EO_N 2 EO_PE 1 EO_L

14AWG

SH1 SH2

MAINS CIRCUITS

SSR_DRV

X7R

1% 330K

MELF0204

R3

MELF0204

3326205

MELF0204

1% 330K

+12V

NOLOAD

E2 R4

3D6>

D

EQUIPMENT OUTLET

WP2

4

MECCA

L2_IN

OUT

DIP8

2152365

3C6>

MELF0204

4A7< 2C7


2152365

12VDC

2152365

TP16

DIP8

+12V

RV1

3 6

1

CRL4

IN

3C6>

7 2 5 4

3 6

DIP8

1

10% 1MA

DIP8

XBF1

K4

7 2 5 4

3 6

1

F1

HEADER

3351138

K1

7 2 5 4

X7R

CHGND EOGS PE N L

430V 10%

F2

1 2 3 4 5

X7R

J5

X7R

MAINS INLET

3B3< 13

11 9 4

5

R14

H5 H6

1.00K

R112

604231 X7R

13

12

7

U8 ADG431B 3367437

6 8 4

5

PAD-375C-WSP PAD-375C-WSP PAD-300C-WSP PAD-300C-WSP

252-HOLE 252-HOLE 215-HOLE 215-HOLE

3350760 3350760 3352172 3352172

COMMENTS COMMENTS COMMENTS COMMENTS

POWER INLET CONN SPACERS

12

10

U8 ADG431B 3367437

H1 H2 H3 H4

PAD-325C-WSP PAD-325C-WSP

215-HOLE 215-HOLE

3376173 3376173

NOLOAD=TRUE NOLOAD=TRUE

1.00K

R113

ESA612-1002 (1 of 4)

1.00K

GNDMS

Figure 7-1. A1 Power PCA (cont.)

7-6

3D1>

Electrical Safety Analyzer Schematic Diagrams

6

MAP CURRENT LIMITS CR12

CR13

S1M 1KV 1A

S1M 1KV 1A

SMA

T1

4A7< 1C8>

IN

L2_IN

5

S1 O

O

102V

115V

P2

S2 O

O

R24

4

3326205

1

NOLOAD

7 9

C16

1000P 20% 2083392 300V Y5P

R27

10P

Q2

3 SOT23

1

2 3 1

TP22 TP23

3331432

820902

S

2742520 3

3326205

10

MMBT5089

K16

900V D2PAK

330K 1%

C74

3342134

Q1

IRFBF20S

C0G

4

P1

2646221

3326205

6 MELF0204

2

102V

115V

330K 1%

SOT23

1

T394

IN

L1_IN_F

THRU-HOLE

4A7< 1C8>

B0710099

MELF0204

MAP TRANSFORMER

R23

330K 1%

MELF0204

2646221

SMA

R25 1.00K

4

CR1

6.8V 5% 5235 837195

R26

7.5K

2278492

930172

3

U10

DIP4S

HCPL-817

3328473

+12V

1 2

R31 1.00K

4

MAP_3MA

IN

3C6>

3

9

3 5

4

12

8

1

IN

OUT

1C4>

5

CRL16 12

1

12VDC

K16

+12V

MAP_RTN

3

4 3C6>

10

+12V

GNDMS

+12V

U29

1

2 R104

DIP4S

1.00K

HCPL-817

3328473

CR14 SMA

CR15

S1M 1KV 1A

S1M 1KV 1A

2646221

10 8

TX2SA 2082903

1.82K

3.01K

1591278

K17

12VDC

R29

R28

928713

CRL17

IN

3C6>

9

Q8

2002486

4.7K 1

INF

SMA

2646221

MUN2216

3

2

MAP_7.5MA

IN

3C7>

SC59

GND

ESA612-1002 (2 of 4)

Figure 7-1. A1 Power PCA (cont.)

7-7

ESA612 Service Manual

CR19 SOT23

CR20 SOT23

MMBZ5242B 2044054 12V 5%

75V 200MA

CRL14

Q9

3

+12V

BAS16

+3.3VMD

830489

R100

4.7K 1 INF

33 16 VCC

SC59

MUN2216

2002486

R107

GND

928713

C57

R105

3B7


R110 C53

16V 10U 3093654

1.00K

3

IN

OUT

LP2985

3.3V

SOT23-5

ON

BYP GND 2

5 4

GNDMS

GNDMD

GNDMD

Figure 7-1. A1 Power PCA (cont.)

7-8

RLY_SDI_M RLY_SCK_M RLY_RCK_M RLY_SDO_M

ADC_DCK_M ADC_CS_M ADC_DIN_M 400MA_FREQ RELAY_EN_M MAINS_ZC_M ADC_DOUT_M GFI_SENSE

PWR_GRP

GNDMA

2791599

+3.3VMD

C54

0.01U

106146

C55

10U 16V 3093654

E4 MECCA

C60

10U 16V 3093654

500V_RTN

+12VM -12VM GNDM GNDM

E12

C46

100P

400MA_RTN

+3.3VMD U28

2559443

928713

867650

HEADER

GNDMD

+12VM 1

928713

GNDMD

C4

ADC_DCK_M

928713

VI C 5

ADUM1401

GNDMD

GNDMD

3343846

VSS

400MA_RTN 500V_RTN

OUT OUT

928713

VI A 3

GND 10U 16V 3093654

867650

VDD1 1

U5

3331444

C68

C47

100P

200

3372234

16 VDD2

1.00K

TANT 3216

+12V

OUT

ADC_/CS

R95

3372234

C0G

R94

200

C0G

ADC_DCK

C67

100P

867650

928713

1.00K

MAINS_ZC

E13

R33

2

DIP4S

GND

C0G

OUT

C66

100P

867650

TANT 3216

1B5


3328473

GND

R39

AP_METER

OUT

J2

HEADER

1 400MA 2 METER 3 +12V 4 AP_METER 5 0V 6 GND

GNDMS

GND

C12

U11

HCPL-817

X7R

4

4 3

928713

C0G

RLY_SCK_G

RLY_RCLK

GNDMD

C0G

OUT

OUT

METER

TP35

METER

OUT

+3.3VMD

R30

1.00K

HCPL-817

3C6>

C0G

3D8


R15

C0G

+3.3VPD

C23

3D8< 3B7


C28

GNDMD

U26 1

5

4

GNDMD

GND

X7R

X7R

/RELAY_EN_M

TP34

0.1U

C43

690500

4C7


SC59

2

GNDMD

R55

10.0K

R50

928791

332K

-5VCM

3

4

2N7002E

R46

1 2

R69

HV_MEN_500V

1.00K

IN

3D1>

IN

3D1>

928713

11.5K

2

S

3328473

GNDMD

3331397

3

60V 240MA

1756473

DIP4S

23.2K

1997330

Q5

U20

HCPL-817

R66

Q6 3

2N7002E

1

60V 240MA

1756473

R68

R86

2

S

4

U21

DIP4S

1 2

R70

HV_MEN_250V

1.00K

928713

HCPL-817

R67

3328473

100K

928866

-5VCM

3

1

100K

2005982

2005982

-

3

1 400PPM

1 400PPM

+

-5VCM

R47

2005982

R59

V2

3

484378

R85

MUN2216

3

928791

X7R 50V 10%

1 400PPM

3B2>

+12VM

0.001U

-5VCM

2005982

1

CR6

BAS16

C33

SOT23-5

3331500

830489

928945

1 400PPM

-5VCM

R48

1.00M

R76

1

2

929476

OUT

10.0K

5 V+

TLV2371

3326333

C31

484378

1M 1%

3326231

4.7K

SO8

C30

R57

4

-5VCM

470

X7R

X7R

R121

484378

0.047U

1708909

S

C73

S

R44

Q4

4

5 V+

-5VCM

3331415

3

S1M 1KV 1A

D2PAK

6.81K

3

X7R

10.0K

928713

X7R

R77

1579273

2646221

1.00K

R61

IRLR3410

100V 17A .1 OHMS 3W

56K

MELF0204

1.1K 20%

400MA

+5VCM

R62

RV75

THERM-THRU

+t

7

-5VCM

SOT23

GNDMD

TP36

Q3 4

500V_RTN

R49

U24

SOT23

HCPL-817

3

CR16

BAS16

C35

0.1U 50V 690500

SOT23

DIP4S

400MA

R45

GNDCM

C0G

2

4

X7R

1.00K

928713

1

690500

DPAK

R65

C32

0.1U

+5VCM

3

830489

0.22U

3326231

1277360

Q7

5% 12V MMBZ5242B 2044054

10.0K

690500

1C4


3 5

SMA

0.0

-5VCM

U19

9

1

MELF0204

1812

C42

0.047U

10%

10 8

SOT23

47U 16V

.2OHM ESR

3352434

X7R

7343

GNDCM

0.0

6

Z

500V X7R 644077

R123

2

O

C41

0.1U

2087962

0.047U

RV40

THERM-THRU

K13

TP26

MELF0204

O

X

8

C72

Y

C29

1812

4

US1M 1KV 1A

X7R

O

68K 5%

3331385

GNDMS

SMA

0.0

1

2512 1W

68K 5%

CR5

R122

T2

2512 1W

3331385

SOT23

INSULATION INVERTER

TP37

R63

928866

GNDMD

-5VCM

+5VCM E16

+12V

MAINS POWER SUPPLY 85-264VAC,15W,12VDC

2C7< 1C8> 2C7< 1C8>

IN IN

L2_IN

2

L1_IN_F

1

U9

ACL +VO 32X64MM ACN -VO

ECL15US12

3331480

3 4

R120

0.0

R38

3.01K

0.0

+12V

20%

C40

RED

VSS

0.0

VOUT+

2 VIN-

VOUT-

SIP7-5P

GND

D215EI 3338710

5

+5VCM

C39

VR2

1U 25V 20% 2530275

7

COM 6

CR4

AM2520 2281407 LED-SMD

VIN+

25V 1U 2530275

1591278

VSS

U18 1

R119

R125

GNDCM

TP12

6.5V

E5

R126 0.0

R127

C38

MECCA

VR3

1U 25V 20%

2530275

6.5V

GNDCM -5VCM

0.0

400MA_RTN

IN

3B2> 1D1
METER

3

IN

METER INPUTS

METER JACKS RED 1

2

3

4

3

TP168

L4

BLACK 1

2

P13

3

CR6

2

1

440V

CR36

10

440V

L10

4

95_OHM

1

NULL 1

2

CR34

2

440V

P24

3

1

CR37

10

TP100

2

4

5

R85

1M

R65

R66

330K

330K

330K

3

TP107

5 8

6.81K

1

2

1

440V NULL WIRE FROM POWER PCA GREEN

C175

CR35

2

TP105

440V

TP169

R370

261K

3 10

R184

2

27

F1 C

3A

0.027U

C135 0.1U

CHGND WIRE FROM POWER PCA GREEN/YELLOW

GHGND WIRE TO PAGE 4 GREEN/YELLOW

4B7> 400MA

E27

261K

5 8

W2 D

14AWG

4

3 5

9

10 8

E28

K18

2 4 9

GNDMA

R88

IN

IN

1 2

TP101 TP102

10

4

5 8

TP103

3

TP104

10

5 8

C139

0.01U

6

R86

100K

K21 4

GND

R89

GNDMA GNDMA

V+

8

4

150K

9

-5VMA

VDD

C145 0.1U

GNDMA

VL

R182

3

-12VMA

PWR_GRP

4A6>

C138

VSS GND

4

IN

11.0K

1

0.015U

5

GNDMA

GNDMA

IN

GNDMA

PWR_GRP_SEL

3B6>

-12VMA

1

OUT

/FAULT_SENSE

3B8


3

3

8

3C6>

R71

C165

+5VMA

150K

TP106

+12VMA

ADG431B U27

+3.3VMA

VR7

7.5V

AAMI_LOAD

13 12

7.5V

+5VMA

+5VMA

R87

R406

IN

5

2

VR6

CR9

FAULT SENSE

27

R372

0.027U

16 4

10.0K

R183

3C6>

12

+12VMA

MMBD1503 1

IEC_LOAD

IN

5

14

U27 ADG431B

GNDMA

0.0047U

261K

C193

9 4

15

C71

R371

261K

11

13

9

GNDMA

9

8

12

100

4

4

5

2C8


5

R70

K16

NO_LOAD

OUT

GNDMA

R67

METER_AMP_IP

OUT

R156

K20

95_OHM

CR33

4

U27 ADG431B

200

R64

8

9

8

440V

L3

4

3

TP99

6

U27 ADG431B

10

R79

10

K15

C100

0.015U

12

7

13

1M

9

8

2

R84

10K

9

8

800

4

5

95_OHM

1

K13

10

R78

1M

13

R68

4

5

R83

P12

K25

6

3

TP108

U18 7

LM393DT

R100

R99

TP109 R98

330K

330K

330K

MAP NULL

K24 4

5 10

+5VMA

R219

9

8

0.1U

-5VMA

GNDMA

-5VMA

R97

R103

8.06K

3.92K

C74

0.1U

GNDMA

C75

0.1U

-5VMA

U22 7

TLV2372

8 V+ V4

-5VMA

+

R94

6

3.92K

5

R96

1.00K

GNDMA

7

R93 3.92K

+5VMA

+5VMA

9

33

C141

U22 1

TLV2372

4

+5VMA 8 V+ V4

-5VMA

3 +

2 3

R101 3.92K

R95

1.00K

8

R102

VDD OUT FS

U21

VREF DIN

TLV5623

CS-

SCLK

6 1 2

10.0K

R117

100K

IN IN

ADC16_SDO ADC16_SCK

5C6> 2B2< 3A8< 5C6> 2B2< 3A8


5C6> 3A8
METER_AMP_IP

3

IN

NC

+

NC

5

U54

V4

NC

U39

13

6

11

AD8510AR

4

3

R111

+12VMA 1.10K +3.3VMA

5

ADG431B

-12VMA

2

2.43K

9

8

2A2< 5C6> CONT_GATE

R108

8 V+

+

2

VDD

U25 ADG431B

-12VMA

R154

1

6

2.43K

5

R153

V4

+

1.10K

8 V+

U28

R110

7

14

1

U25 ADG431B

16

U25 ADG431B

5

1

GNDMA 13

9 5

+

V4

TP87

U26

8 V+

U23

+5VMA

+5VMA

1.10K

AD8034

-5VMA

C77

R129

0.47U

1.00K

7

AD8034

-5VMA

R127

LTC1967

VDD

2

IN1

RTN 6

IN2

8

3

150K

GND 1

6

U25 ADG431B

8 4

IN

E118

E225

MECCA

MECCA

7

VDD

U39 ADG431B

-12VMA

VL

6

0.47U

C143

VSS GND

13

0.1U

5

12

15

14 16

GNDMA

GNDMA

4

R114

2

51.1K

3

+

R126

5

ADG431B

2

1

2

-12VMA

3

NC

+

NC

5

GNDMA

V4

NC

-12VMA

R105 360K

C89

IN

0.47U

+5VMA

6 5

+

8 V+

U29

V4

AD8539

0.1U

5D5< 2C2< 2B6< 2A2> 2V5REF

6

8

5

ADG431B

7

R106

R107

43.2K

432K

C86

6 5

+

8 V+ V4

3

U46 74HC00

AD8510AR

2

74HC00 6

U46

U46 74HC00

AD8539

4.7K

11

4 +3.3VMD

1

2

U46 74HC00

12 13

U30 LM4040 2.5V

CR19

2

3.3V

+3.3VMD

14 DVDD

U24

R119

DOUT/RDY

AD7799

3 AIN3+/P1

R116

GNDMA

10

1

8 AIN2-

+5VMA

GNDMA

9

5

0.1U

2

7 AIN2+

7

0.047U

GNDMA

CR18

6 AIN1-

GNDMA

8

2B5< 2B6< 5D5


GNDMA

249

10 REFIN-

DC INPUT

TP91

+5VMA

0.1U

VCC=+3.3VMD;GND=GNDMD

R351

249

13 AVDD 9 REFIN+

2.5V REF

3

4

U15

7 V+

R104

0.1U

5V

TP90

C87

C88 0.47U

30.1K

C82

0.1U

R350

1

HV MEAS

1

+12VMA

0.1U

GNDMA

1

1

+3.3VMD

C81

0.1U

12

10.0K

0.1U

AD8539

AD8539

GNDMA

C177

GNDMA

U39

13

R168

C72

V4

GNDMA

10.0K

C70

U29

C92

U31

V4

0.047U

+5VMA

TP89

51.1K

R171

+12VMA

8 V+

+

8 V+

GNDMA

AC INPUT

51.1K

+5VMA

U39

-

GNDMA

4

TP95

GNDMA

3

1.00K

NC

GNDMA

0.1U

R115

+3.3VMA

8 4

R125

51.1K

13 12

100K

R118

GNDMS

GNDMD

C80 +3.3VMA

23.2K

2

METER ADC

+5VMA

+12VMA

0.47U

R121

IN

0.1U

METER DC ONLY

C78

C90

R120

GNDMA

GNDMA

IN

-EN

0.47U

10.0M

5

C99

5

IN

5D5< 2C2< 2B5< 2A2> 2V5REF

VOUT

C91

R128

7

12

7

GNDMA 3C6> RMS_GAIN_X1P5

5

R123

12

4

6

2.43K

V4

11

3C6> RMS_GAIN_X10

0.1U

GNDMA

+

GNDMA 13

C142

-

R124

+5VMA

GNDMA

R349

10.0M

2.43K

U26

0.1U

10.0M

R122

8 V+

C83

R348

4.7P

+5VMA

1.10K

-5VMA

12

4

3

R113

3

+3.3VMA

2

2.43K

V4

AD8034

GNDMA

-5VMA

-5VMA

10.0M

GNDMA

C97

+5VMA

R347

0.1U

R112

10

5

C98

0.1U

GNDMA

2.43K

+5VMA

15

VSS GND

4

U28

13

VL

4.7P

931

AD8034

GNDMA

13 12

IN

15P

+5VMA

12

10

C96

R152

931

7 V+

C125

+5VMA

+5VMA

0.1U

-5VMA

15P

-12VMA

C126

0.1U

GNDMA

C95

+5VMA

4 AIN3-/P2 GND

15

DIN 16

SCLK 1 CS 2 PSW 11

10.0K

IN IN

ADC16_SDI

5C6
1B1< 3A8
1B1< 3A8


12

GNDMA

IN OUT

HV_MEAS 4B6> 2V5REF 2B5


IN

CONT_GATE

5C6> 2C7
ADC16_SDO

IN

2B2< 1B1< 5C6> ADC16_SCK

IN

1 2

+5VMD

VDD

DIN

10.0K

1A1
GFI_SENSE

IN

IN

FS

3

V+

5

R77

U17

1.00K

4

GND

2

LMV331

GNDMD

Q9

R401

SI2312B

0

S

VOUT-

2

GNDMD

0.1U

GNDMD 2

CR8

3

C68

0.033U

BAT54A

R76

100K

10U

+12VM

E18

GNDMA

5

C120

COM 6

GND GND

E20

3

1U

+5VMA

5

NC

10.0

-12VMA

C148

C123 0.1U

6

10U

R220 10.0

GNDMA

NOLOAD

6

10U

E11

GNDMA

1

E12

C67 1U

1

3

NC NC GND

1 8

C124

2

10.0

5

10U

U47

100 10

8

R235 C132

0.001U

74HC132

GNDMD

12 13

R407

U47 74HC132

100

U47

4 5

6

GNDMD

1 2

U47

GNDMD

100K

VCC=+3.3VMD;GND=GNDMD

2

74HC02

U19

5

1

6

+3.3VMD

C155

74HC132

0.1U

VCC=+3.3V_U47;GND=GNDMD

11

+3.3VMD

3

VCC=+3.3V_U47;GND=GNDMD

3

GNDMD

74HC02

U19

4

8 9

IN

74HC02

U19

11

10

12

GNDMD

RELAY_DISABLE

5C2>

74HC02

U19

13

R170

OUT IN

/RELAY_EN 4A6> RELAY_ENABLE

3D8< 4D8< 5C1>

10.0K

GNDMD

OUT

/FAULT 5C6
/CLR_FAULT

100

3

CS-

R75

IN

+5VMD

R72

1

RST MCP101T GND 3

3

R172

100K

DAC_FS 5C6>

4

R73

GND 5

R80

5C6> /GFI_DAC_CS

7

OUT

TLV5623

SCLK

R195

BAS16

2 VCC

R74

10.0K

CR10

+5VMD

+3.3VMD

+12VMA

C146

MC79L05A

10.0K

+5VMD

GNDMD

U16

CR40

1.00K

4

2

U14

8

VREF

C121

IN

R91

0.1U

6

8

7

5C1


GFI DAC

33

10U

R151

+3.3VMD

+5VMD

+12VMA

10U

C122

0.01U

VSS

R90

10.0K

VIN+

2 VIN-

NOLOAD

R403

IN

1C1
ECG_SDO

+12VMRLY

33

1.00K

IN

4D8< 5C2> METER_RCK

R186

R192

IN

4D8< 5C2> 4A6> RELAY_SCK

U48

D213EI

+12V

6

/GFI_TRIP

5C6
4A6> RELAY_SCK

3D8< 5C2> METER_RCK

5C1> 4A6> RELAY_SDO

3D8< 4A6> 3B1> /RELAY_EN

33

1.00K 1.00K

11 SHCP

R196

IN

12 STCP

1.00K

14 DS

R198

IN

ULN2003AD

U32

1.00K

74HC595

C156

13 OE 10 MR

100P

C134

15 1 2 3 4 5 6 7

Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7

TP12

GND

8

GND 8

R237

16 15 14 13 12 11 10

2 3 4 5 6 7

R239

GNDMD

1.00K

1.00K

100P

12V

9

TP10 1

9

Q7S

+12VMRLY

VR12

U33

16 VCC

R199

IN

+12VMRLY

R178

R197

IN

GNDMD

C151 100P

+12VMRLY

C140

GNDMD

11 SHCP

ULN2003AD

U34

12 STCP 14 DS

C170 0.1U

GNDMD

74HC595

13 OE 10 MR

R238

VR13

U35

16 VCC

100P

15 1 2 3 4 5 6 7

Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7

TP11 1

16 15 14 13 12 11 10

2 3 4 5 6 7

TP13

GND

9

Q7S

12V

9

8

GND 8

GNDMD

1.00K 3D8< ECG_SDO

12

1

K11

3

K10

10

12

1

12

K8 1

TP172

3

12

K6 1

TP173

10

12

K4 1

12

K2 1

3 5 10

K22

12

1

12

1 K23

GNDMA

1

12

1 K9

12

1 K7

12

1 K5

12

1 K3

12

1 K1

8

E149 MECCA

ECG WAVEFORM AMP

E177

R176

MECCA

39.2K

OUT

9 3

TP174

K23

1B8
4D8
3D8< 4D8


MAINS_SDI

5C2


OUT

/ADC12_CS

5C1>

ADC12_SDO

5C6>

OUT

400MA_FREQ

OUT OUT OUT OUT

/RELAY_EN

OUT

ADC12_SDI

5C6


+12V

VR4

15V

E7

GHGND WIRE TO PAGE 1 GREEN/YELLOW

U60

GNDMA

1U

GNDPD

LTC1144

8

V+

TP166

1

BOOST

TP167

7

+12V

C190

VOUT 5

OSC -SD 6

5C1> 2A2
METER_RESET

R141

IN

U43

1

1.00K

2

OUT IN

OUT OUT

5A4> UI_RXD 3A1> /GFI_TRIP

3A1> /FAULT

OUT

IN

OUT

33P

61 XTAL2

R374

102

R375 R376 R377 R378

499 499 499 499

102

2B2< 1B1< ADC16_SCK 3A8< 2B2< 1B1< ADC16_SDO 3A8
METER_TXD

IN

6C6> /METER_ISP

IN

GNDPD

10.0K

R206

R140

10.0K

100

100

11 VI D

R208 100

C159 100P

U42

ADUM1402

R209 100

16 VDD2 14 VO A 13 VO B 12 VI C

R207

C160

TP119 TP120 TP121

R382 R383

499 499

R384 R385

499 499

R386 R387

499 499

OUT

TP118

OUT

TP124 TP122

OUT OUT

OUT

TP123

IN

U41

RST MAX803 GND 1

RELAY_SCK 4A6> 3D8
ADC12_SDI

IN

12MHZ

C102

0

C112

VDD

R203

Y2

GNDMD

GNDMD

0

TP22

VDD

33P

0

C111

VSS VSS VSS VSS VSS

1.00K

GNDMA

0

C110

R358 0.1U

GNDMD

VREF

R148

3

C101

8

+3.3VPD

10U

VDD

1 2 3 4

+3.3VPD

TP19

100K

C104

2

R357

VSSA

HEADER

1

R356

59

R145

J22

100

CR17

BAT54A

2

49

NOLOAD

R355

1U

63

DEBUG RS232

R144

IN

2C2< 2B6< 2B5< 2A2> 2V5REF

CR30 3.3V

C115

VBAT

+3.3VMD

1

6 18 25 42 50

METER CIRCUITS

MAINS CIRCUITS

+3.3VMD

METER PROCESSOR

6

10 VE2 9 15 GND2 GND2

R137

VDD1 1

C164 0.1U

R147

10.0K

10.0K

VI A 3 VI B 4 VO C 5

R210

100

R211

100

R212

100

GNDMD

VE1 7 GND1 2 GND1 8

R138

10.0K

100P

GNDMD

C161 100P

C162 100 100P

1

UI_TXD 5C6>

OUT

UI_RXD 5C6


OUT

/ISP 5C6
METER_ATTN

GNDPD

1

7D6>

KEYCOL_0

7D6>

KEYCOL_1

7D6>

KEYCOL_2

7D6>

KEYCOL_3

7D6>

GNDPA

TP131 TP132

5A7< /METER_ISP

OUT

IN

5C8< METER_RESET

OUT

OUT

R388

102

TP133 TP134 100P TP135 TP136 TP137 GNDPD TP138

C188

R5 10.0K

OUT

TP129 TP130

499 499 102 499 499

IN IN IN

IN IN

KEYCOL4

+3.3VPD

DS1

RED

AM2520

R2

Q1

0.1U

3C4< MPSO 7C3


7D3< /DISP_CLK_OFF

1.00K SI

IN

FM25L256

+3.3VPD

ISP

SO

R346

U59

USB_RXD

8C7> /USB_PRESENT

OUT

8C7< USB_RTS

C174

8D7>

OUT

VDDOUT

GNDPA

GNDANA

0.1U

1.8V 1

VBUS N/C

85 86 124 125 132 133

TP152 TP153

TP73

1 2 4 10 12 24 26 31 33 35 38 40 42 44 46 59 61 63 R399 65 67 87 89 91 95 97 99 101 103 105 107 110 112 114 118 120 135 137 140 142 144

TP156 TP157 TP158

23

12MHZ

7 6 5 4 3 2 1 0

TP127 102

TP160 TP161 TP162 TP163 TP164

GNDPD

15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0

/RD 7C5


/DISP_CLK_OFF

GNDPD

150

R260

IN

100K

TP77

29

BL_ENABLE

6B6>

150

6B6>

DISPLAY CONTRAST

E24

R181

S

1.00M

Q7

1

E25

U5

10K

5

VIN

SW

1

10K

A

CR3

MBR0530

R19

4 SHDN 10U

24K

1.00K

1U

R23 1.00K

121K

20V

LP324

1

R29

1

100

R20

+

4 V+

U6

V11

LP324

7

R30

4.7P

C26 1U

1U

R26

9 10

+

4 V+

U6

V11

LP324

8

C42

C45

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

C36

C32

C35

R31

3

100

+3.0V

C38 C37

C27 1U

+3.3VPD 2

100

10.0

C28

1.5M

100K

6 5

E17 C29

R21

R22

3

GND 2

C46

GNDPD

FB

R25

+20V

R18

C

2M

2

V11

1U

1U

LT1615

3

+

U6

C30

C31

20%

Q8

-

4 V+

L1 10UH

MMUN2211

2 3

FDN338P

DISP_V

0

MECCA

R24

121K

+12V

100K

IN

R81

TP76

31

1

R34

NOLOAD

IN

-ST

DISP_V

6C3> DISP_OFF

34

40MHZ GND 2

GNDPD

GNDPD

30 26 23

OSC

U4

B1 B2

R180

6B6> CONTRAST

D0 D1 D2 D3 D4 D5 D6 D7

+3.3VPD

R253

10.0K

10.0K

TEST

+12V

6C3> ENABLE_20V

R252

CLKI CNF0 CNF1 CNF2 CNF3 CNF4

6B6>

S4

LEFT

C1 C2

6C7
USB_TXD

OUT OUT IN IN

100

10U

C53

100P

R47

16 VDD2

GNDPD

1.00K

14 VO A 13 VO B 12 VI C

R45 1.00K

R46

C55

10 VE2 9 15 GND2 GND2

C56

100P

100P

1.00K

VI A 3

R38

1.00K

R37

1.00K

VO C 5

R39

1.00K

VE1 7 GND1 2 GND1 8

C58 10U

R42

TP79 TP80

10.0K

10.0K

OUT

23 22 13 14 12

U12

1

3

2

USBDM

0.1U

21

10U

17

7

0.01U

16

USBDP

15

RESET

19

J3

C47

0.1U

OSCI 27

TP84

OSCO 28

TP85

USB

USB

2

+

GND

1

GND

3

U9 75240

TEST 26

GND

C49

1 2 3 4

FT232RL

CBUS0 CBUS1 CBUS2 CBUS3 CBUS4

GNDUSB

4

3V3OUT

U7

C60

1.00K 0.1U

VCC

TXD RXD RTS CTS DTR DSR DCD RI

C136

+3.3VPD

R50

6C5< /USB_PRESENT

TP113

10.0K

10.0K

1 5 3 11 2 9 10 6

51.1

R215 C48

20

VCCIO

R41

R40

GNDPD GNDPD

4

VO D 6

ADUM1402

R48

R36

VI B 4

11 VI D

1.00K 1.00K

VDD1 1

U8

R216

+5V_USB

MT1

C54

100P

MT2

C57

-

AGND

4

25

18

R49

NOLOAD

L11

95_OHM

OUT

NOLOAD

1

CR39

2

440V

POS_ESD 4B1
24 oz

B

>24 oz

C

>8 oz

Functional A.11

PASS

FAIL

A.13

PASS

FAIL

A,32

PASS

FAIL

A.36

PASS

FAIL

A.41

PASS

FAIL

A.46

PASS

FAIL

Electrical Outlet Polarity B.10

PASS

FAIL

B.13

PASS

FAIL

B.18

PASS

FAIL

B.27

PASS

FAIL

B.30

PASS

FAIL

B.33

PASS

FAIL

EO Ground to NULL Jack C.6

PASS

FAIL 0 to 2.5 Ω

0

+2.5 Ω

+2.5 Ω

D.8

112.5 – 117.5 V

115.0 V

±2.5 V

±2.0 % +0.2 V

D.12

112.5 – 117.5 V

115.0 V

±2.5 V

±2.0 % +0.2 V

C.8 Mains Voltage

GFI Verification

A-2

E.11

PASS

FAIL

E.23

PASS

FAIL

E.38

PASS

FAIL

E.50

PASS

FAIL

E.64

PASS

FAIL

E.76

PASS

FAIL

Datasheets Introduction

A

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

Point to Point Voltage F.9

0.78 to 1.22 V

1.00 V (120 Hz)

±0.22 V

±2.0 % +0.2 V

F.12

244.8 to 255.2 V

250 V (60 Hz)

±5.20 V

±2.0 % +0.2 V

F.15

60.10 to 81.32 V

70.71 V (633 Hz)

±10.61 V

±15.0 %

F.21

3.72 to 4.28 V

4.00 V

±0.28 V

±2.0 % +0.2 V

F.24

7.64 to 8.36 V

8.00 V

±0.36 V

±2.0 % +0.2 V

F.27

9.60 to 10.40 V

10.00 V

±0.40 V

±2.0 % +0.2 V

F.30

24.30 to 25.70 V

25.00 V

±0.70 V

±2.0 % +0.2 V

F.33

39.00 to 41.00 V

40.00 V

±1.00 V

±2.0 % +0.2 V

F.36

78.20 to 81.80 V

80.00 V

±1.80 V

±2.0 % +0.2 V

F.39

127.2 to 132.8 V

130.00 V

±2.80 V

±2.0 % +0.2 V

F.42

235.00 to 245.00 V

240.00 V

±5.00 V

±2.0 % +0.2 V

-0.015 to +0.015 Ω

0.0000 Ω

±0.015 Ω

±2.0 % +0.015 Ω

Resistance G.11

1.800 Ω

G.14 [1]

G.21

-0.015 to +0.015 Ω

0.0000 Ω

±2.0 % +0.015 Ω ±0.015 Ω

1.800 Ω

G.24 [1]

±2.0 % +0.015 Ω ±2.0 % +0.015 Ω

Insulation Resistance H.11

250.0 to 300.0 V

250.0 V

+50.0 V

+20.0 %

H.16

500.0 to 600.0 V

500 V

+100.0 V

+20.0 %

H.20

1.750 to 2.250 mA

2.000 mA

±0.250 mA

±12.5 %

A-3

ESA612 Service Manual

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

Mains to PE 9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

H.79.RA

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

LL

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

LA

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

RL

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

V1

9.6 to 10.4 MΩ

10.000 MΩ

±0.4 MΩ

±2.0 % +0.2 MΩ

250

0.486 to 0.914 MΩ

0.700 MΩ

±0.214 MΩ

±2.0 % +0.2 MΩ

500

0.486 to 0.914 MΩ

0.700 MΩ

±0.214 MΩ

±2.0 % +0.2 MΩ

250

0.780 to 1.220 MΩ

1.000 MΩ

±0.220 MΩ

±2.0 % +0.2 MΩ

500

0.780 to 1.220 MΩ

1.000 MΩ

±0.220 MΩ

±2.0 % +0.2 MΩ

250

6.170 to 6.830 MΩ

6.500 MΩ

±0.330 MΩ

±2.0 % +0.2 MΩ

500

6.170 to 6.830 MΩ

6.500 MΩ

±0.330 MΩ

±2.0 % +0.2 MΩ

250

17.440 to 18.560 MΩ

18.000 MΩ

±0.560 MΩ

±2.0 % +0.2 MΩ

500

17.440 to 18.560 MΩ

18.000 MΩ

±0.560 MΩ

±2.0 % +0.2 MΩ

AP to PE H.44 AP to NE H.57 Mains to NE H.68 Mains to AP

A-4

Datasheets Introduction

A

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

250

20.150 to 23.850 MΩ

22.000 MΩ

± 1.850 MΩ

± 7.5 % +0.2 MΩ

500

20.150 to 23.850 MΩ

22.000 MΩ

± 1.850 MΩ

± 7.5 % +0.2 MΩ

250

55.300 to 64.700 MΩ

60.00 MΩ

± 4.700 MΩ

± 7.5 % +0.2 MΩ

500

55.300 to 64.700 MΩ

60.00 MΩ

± 4.700 MΩ

± 7.5 % +0.2 MΩ

250

92.30 to 107.70 MΩ

100.00 MΩ

± 7.70 MΩ

± 7.5 % +0.2 MΩ

500

92.30 to 107.70 MΩ

100.00 MΩ

± 7.70 MΩ

± 7.5 % +0.2 MΩ

DC Leakage I.10

0.980 to 1.020 KΩ

1.000 KΩ

±0.02 KΩ

±2.0 %

I.23

8.900 to 11.100 μA

10.00 μA

±1.1 μA

±1.00 % +1.0 μA

I.27

48.500 to 51.500 μA

50.00 μA

±1.5 μA

±1.00 % +1.0 μA

I.31

98.000 to 102.000 μA

100.00 μA

±2.0 μA

±1.00 % +1.0 μA

I.35

157.40 to 162.600 μA

160.00 μA

±2.6 μA

±1.00 % +1.0 μA

I.39

335.60 to 344.40 μA

340.00 μA

±4.4 μA

±1.00 % +1.0 μA

I.43

494.00 to 506.00 μA

500.00 μA

±6.0 μA

±1.00 % +1.0 μA

I.47

0.989 to 1.011 mA

1.000 mA

±0.011 mA

±1.00 % +0.001 mA

I.51

1.583 to 1.617 mA

1.600 mA

±0.017 mA

±1.00 % +0.001 mA

I.55

3.356 to 3.444 mA

3.40 mA

±0.044 mA

±1.00 % +0.01 mA

I.59

4.940 to 5.060 mA

5.00 mA

±0.060 mA

±1.00 % +0.01 mA

I.63

6.020 to 7.080 mA

7.00 mA

±0.080 mA

±1.00 % +0.01 mA

A-5

ESA612 Service Manual

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

AC Filter Frequency Response J.8

J.9

J.15

Analyzer Current Measured Current 0.98801 – 1.00799

±0.00999

Analyzer Current Measured Current 0.67218 – 0.70982

J.19

0.99800

0.69100

±0.01882

Analyzer Current Measured Current 0.088669 – 0.102492

0.09558

±0.0069116

Differential Leakage

A-6

K.12 [1]

76.00 μA

±10.0 % +20.0 μA

K.13 [1]

160.00 μA

±10.0 % +20.0 μA

K.13 [1]

240.00 μA

±10.0 % +20.0 μA

K.13 [1]

500.00 μA

±10.0 % +20.0 μA

K.13 [1]

760.00 μA

±10.0 % +20.0 μA

K.14 [1]

1.600 mA

±10.0 % +0.02 mA

K.14 [1]

2.40 mA

±10.0 % +0.02 mA

K.14 [1]

5.00 mA

K.14 [1]

7.60 mA

±10.0 % +0.02 mA

K.15 [1]

16.00 mA

±10.0 % +0.02 mA

±10.0 % +0.02 mA

Datasheets Introduction

A

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

Leakage Functionality ECG Leakage Functionality RA L.15

PASS

FAIL

L.18

PASS

FAIL

L.21

PASS

FAIL

L.15

PASS

FAIL

L.18

PASS

FAIL

L.21

PASS

FAIL

L.15

PASS

FAIL

L.18

PASS

FAIL

L.21

PASS

FAIL

L.15

PASS

FAIL

L.18

PASS

FAIL

L.21

PASS

FAIL

L.15

PASS

FAIL

L.18

PASS

FAIL

L.21

PASS

FAIL

LL

LA

RL

V1

Direct Applied Part, Alternative Applied Part: Ground to RA M.20 [1]

M.38 [1]

1.000 mA

±1.00 % +1.0 μA

1.000 mA

±1.00 % +1.0 μA

1.000 mA

±1.00 % +1.0 μA

1.000 mA

±1.00 % +1.0 μA

Direct Applied Part, Alternative Applied Part: Red to RA N.20 [1]

N.38 [1]

A-7

ESA612 Service Manual

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

Alternative Equipment: Red to Hot O.20 [1]

1.000 mA

±1.00 % +1.0 μA

Alternate Equipment Leakage, Alternative Applied Part: RA to Hot P.20 [1]

P.38 [1]

1.000 mA

±1.00 % +1.0 μA

1.000 mA

±1.00 % +1.0 μA

1.000 mA

±1.00 % +1.0 μA

100.00 μA

±1.00 % +1.0 μA

100.00 μA

±1.00 % +1.0 μA

100.00 μA

±1.00 % +1.0 μA

100.00 μA

±1.00 % +1.0 μA

100.00 μA

±1.00 % +1.0 μA

Alternative Applied Part Patient: RA to Neutral Q.18 [1] Patient Auxiliary: RA to RL R.22 [1]

Direct Equipment, Patient: Ground to RA S.20 [1]

S.44 [1] Direct Equipment, Enclosure: Ground to Red T.20 [1]

T.42 [1] Filter Tests

A-8

U.20 [1]

0.000 μA

±1.00 % +1.0 μA

U.26 [1]

0.000 μA

±1.00 % +1.0 μA

Datasheets Introduction

A

Table A-1. ESA612 Verification Datasheet (cont.) Step No.

Measured Value

Limits

Nominal (Actual) [1]

Range

Percentage

MAP Voltage and Current Limits V.9

115.0 V

±5.0 %

V.21

230.0 V

±5.0 %

V.31

3.500 mA

±25.0 %

V.40

7.500 mA

±25.0 %

V.52

1.000 mA

±25.0 %

ECG Waveforms W.9

PASS

FAIL

W.11

0.674 mV

±5.0 %

LL

1.673 mV

±5.0 %

LA

1.384 mV

±5.0 %

V1

2.074 mV

±5.0 %

X.9 [1]

8.000 A

±5.0 % +0.2 A

X.12 [1]

3.000 A

±5.0 % +0.2 A

X.15 [1]

1.000 A

±5.0 % +0.2 A

Equipment Current

[1]

These steps require measurement of the standard prior to testing the UUT. The value of the standard is entered into the “Actual” block, and the range and limits are filled in based on this value.

Serial Number on Analyzer: UI Firmware Revision: Meter Firmware Revision: Nominal Voltage: Final Data Serial Number: Cal Date: Technician Number: Technician Signature:

A-9

ESA612 Service Manual

Table A-2. Calibration Datasheet Step No. A.15

Measured Value PASS

Limits

(Actual) [1]

Range

Percentage

FAIL

A.50

1470.00 – 1530.00 mV

1.500 V

±30 mV

±2 %

A.53

490.00 – 510.00 mV

0.5000 V

±10 mV

±2 %

A.58

147.00 – 153.00 mV

150.00 mV

±3 mV

±2 %

A.63

14.70 – 15.30 mV

15.00 mV

±0.3 mV

±2 %

B.11

1470.00 – 1530.00 mV

1.500 V

±30 mV

±2 %

0.5000 mA

C.13 [1]

D.9

244.8 – 255.2 V

F.13 [1]

G.11 [1]

H.13 [1] I.18

PASS

FAIL

I.30

PASS

FAIL

I.52

PASS

FAIL

I.64

PASS

FAIL

I.85

PASS

FAIL

PASS

FAIL

I.97 J.12

[1]

250.00 V

±1 % +0.1 μA ±5.2 V

97.8 – 102.2 v

L.20

97.8 – 102.2 v

±2 % +0.2 V

10.000 MΩ

±2 % +0.2 MΩ

3.000 Ω

±2 % +0.015 Ω

11.000 mA

±10 % +0.02 mA

3.000 A

K.7

[1]

±1 % +0.1 μA

3.0000 mA

C.22 [1]

A-10

Nominal

±5 % +0.2 A

100 V

±2.2 V

±2.0 % +0.2 V

100 V

±2.2 V

±2.0 % +0.2 V

These steps require measurement of the standard prior to testing the UUT. The value of the standard is entered into the “Actual” block, and the range and limits are filled in based on this value.

Datasheets Introduction

A

Serial Number on Analyzer: UI Firmware Revision: Meter Firmware Revision: Nominal Voltage: Final Data Serial Number: Cal Date: Technician Number: Technician Signature:

A-11

ESA612 Service Manual

A-12