ESA612 Electrical Safety Analyzer Service Manual PN 3477999 June 2010, Rev. 1 © 2010 Fluke Corporation. All rights res
Views 160 Downloads 4 File size 7MB
ESA612 Electrical Safety Analyzer
Service Manual
PN 3477999 June 2010, Rev. 1 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty. This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
07/07
Contents (continued)
Notices All Rights Reserved Copyright 2010, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.
Technical Support For application support or answers to technical questions, either email [email protected] or call 1-800- 850-4608 ext 2560 or 1-440-498-2560.
Claims Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
Standard Terms and Conditions Refunds and Credits Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable condition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit. Restocking Charges Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and accessories will be applied to all returns. Return Procedure All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling. Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double-walled carton of sufficient strength for the weight being shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts. Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at 1-800- 850-4608 ext 2560 or 1-440-498-2560.
Repair and calibration: To find the nearest service center, go to www.flukebiomedical.com/service, or In the U.S.A.: Cleveland Calibration Lab Tel: 1-800-850-4606 Email: [email protected] Everett Calibration Lab Tel: 1-888-993-5853 Email: [email protected] In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-402-675300 Email: [email protected] In Asia: Everett Calibration Lab Tel: +425-446-6945 Email: mailto:[email protected]
Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.
WARNING Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location The ESA612 Electrical Safety Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.
Table of Contents
Chapter 1
Title
Introduction and Specifications......................................................... 1-1 Introduction........................................................................................................ Intended Use ...................................................................................................... Electrical Symbols ............................................................................................. Product Manuals ................................................................................................ About this Manual ............................................................................................. Safety Information ............................................................................................. Specifications..................................................................................................... Detailed Specifications ......................................................................................
2
Page
1-3 1-3 1-3 1-3 1-3 1-4 1-5 1-5
Theory of Operation ............................................................................ 2-1 Introduction........................................................................................................ Block Diagram Description ............................................................................... A1 Power PCA .............................................................................................. A2 Meter and User Interface PCA ................................................................ Block Diagram to Schematic Cross Reference .................................................. A1 Power PCA Schematic............................................................................. A2 Meter and User Interface PCA Schematic............................................... Detailed Circuit Description .............................................................................. A1 Power PCA .............................................................................................. Equipment Outlet Relays .......................................................................... Equipment Outlet Current Measurements................................................. Mains Measurement .................................................................................. Isolated Mains Voltage (Mains on Applied Parts or MAP) ...................... Connectors ................................................................................................ Isolation..................................................................................................... Relay Drivers ............................................................................................ Isolated Signals ......................................................................................... Power Supplies.......................................................................................... Resistance Current Source ........................................................................ Insulation Resistance Test Voltage ........................................................... A2 Meter and User Interface PCA Schematic............................................... Meter Inputs .............................................................................................. Patient Load .............................................................................................. i
2-3 2-3 2-3 2-4 2-5 2-9 2-10 2-11 2-11 2-11 2-11 2-11 2-12 2-12 2-12 2-12 2-12 2-12 2-12 2-13 2-13 2-13 2-13
ESA612 Service Manual
Overvoltage Protection.............................................................................. Mains on Applied Parts Compensation..................................................... AC Only .................................................................................................... DC Only .................................................................................................... Resistance Measurement ........................................................................... Relay Drivers ............................................................................................ Meter Power Supplies ............................................................................... Relay Disable ............................................................................................ Applied Parts............................................................................................. Meter Microprocessor ............................................................................... User Interface Microprocessor .................................................................. Keyboard and Display............................................................................... USB Port ................................................................................................... U/I Power Supplies ................................................................................... 3
Verification and Calibration................................................................ 3-1 Introduction........................................................................................................ Required Equipment .......................................................................................... Analyzer Verification ........................................................................................ Outlet Tension Tests...................................................................................... Setup .............................................................................................................. Verification Procedure................................................................................... Calibration Procedure ........................................................................................ Setup .............................................................................................................. Adjustments...................................................................................................
4
3-3 3-3 3-4 3-4 3-4 3-5 3-52 3-52 3-52
Remote Operations ............................................................................. 4-1 Introduction........................................................................................................ Serial Communication Connections .................................................................. Serial Communication ....................................................................................... Modes ................................................................................................................ Reading Formats ................................................................................................ Command Specifications ................................................................................... General Commands ....................................................................................... Remote Commands ....................................................................................... ECG Commands............................................................................................ Calibration Commands .................................................................................. Diagnostic Commands................................................................................... Error Responses ................................................................................................. Status Word Definitions..................................................................................... Firmware Updates..............................................................................................
5
2-13 2-13 2-14 2-14 2-14 2-14 2-14 2-14 2-14 2-15 2-15 2-15 2-15 2-15
4-3 4-3 4-3 4-3 4-5 4-5 4-5 4-6 4-11 4-12 4-18 4-21 4-22 4-24
Maintenance......................................................................................... 5-1 Introduction........................................................................................................ Testing and Replacing the Fuses........................................................................ Cleaning the Analyzer ....................................................................................... Disassembly/Assembly ...................................................................................... Troubleshooting ................................................................................................. Power on Failure............................................................................................ Output Failures .............................................................................................. Meter Failures................................................................................................ MAINS Power........................................................................................... Differential Current and Equipment Current ............................................ Meter Troubleshooting .................................................................................. ii
5-3 5-3 5-3 5-4 5-4 5-4 5-5 5-5 5-5 5-5 5-5
Contents (continued)
6
Replaceable Parts ............................................................................... 6-1 Introduction........................................................................................................ 6-3 How to Obtain Parts........................................................................................... 6-3 Parts List ............................................................................................................ 6-4
7
Schematics .......................................................................................... 7-1 Introduction........................................................................................................ 7-3
Appendices A Datasheets.................................................................................................... A-1
iii
ESA612 Service Manual
iv
List of Tables
Table 1-1. 2-1. 2-2. 3-1. 3-2. 3-3. 4-1. 4-2. 4-3. 6-1. 6-2. 6-3.
Title Symbols.................................................................................................................. A1 Power PCA to Schematic Cross Reference ...................................................... A2 Meter and User Interface PCA Schematic Cross Reference ............................ Required Equipment for Analyzer Verification ..................................................... Keypad Verification ............................................................................................... Nominal Values...................................................................................................... Reading Formats .................................................................................................... Error Responses ..................................................................................................... Status Word Definitions ......................................................................................... Final Assembly....................................................................................................... A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................
v
Page 1-3 2-9 2-10 3-3 3-6 3-52 4-5 4-21 4-22 6-4 6-7 6-12
ESA612 Service Manual
vi
List of Figures
Figure 2-1. 3-1. 3-2. 3-3. 3-4. 3-5. 3-6. 3-7. 3-8. 3-9. 3-10. 3-11. 3-12. 3-13. 3-14. 3-15. 3-16. 3-17. 3-18. 3-19. 3-20. 3-21. 3-22. 3-23. 3-24. 3-25. 3-26. 3-27. 3-28. 3-29. 3-30. 3-31. 3-32. 3-33. 3-34.
Title ESA612 Block Diagram......................................................................................... Electrical Equipment Outlet Polarity Test ............................................................. Electrical Outlet Ground to NULL Jack Resistance .............................................. GFI Verification Connections ................................................................................ Point to Point Voltage Verification Connections ................................................... Point to Point Resistance Verification Connections............................................... Insulation Resistance Verification Connections..................................................... Insulation Resistance with 110 kΩ Resistor Connections...................................... Mains to PE Verification Connections................................................................... AP to PE Verification Connections........................................................................ AP to NE Verification Connections ....................................................................... Mains to NE Verification Connections .................................................................. Mains to AP Verification Connections .................................................................. DC Leakage Verification Connections .................................................................. DC Leakage Current Verification Connection....................................................... AC Leakage Filter Frequency Response Connections ........................................... Differential Leakage Verification Connections ..................................................... ECG Leakage Functionality Connections .............................................................. Direct Applied Part Verification Ground to RA Connections ............................... Direct Applied Part Verification Red to RA Connections ..................................... Alternate Equipment Leakage: Red to Hot Connections ....................................... Alternative Applied Part Leakage: RA to Hot ....................................................... Alternative Applied Part Leakage: RA to Neutral Connections ............................ Patient Auxiliary Leakage: RA to RL Connections ............................................... Direct Equipment Leakage, Patient Leakage: Ground to RA ................................ Direct Equipment Leakage, Enclosure Leakage: Ground to Red Connections...... MAP Voltage and Current Limits Connections ..................................................... ECG Waveform Connections................................................................................. Equipment Current Calibration Connection........................................................... AC Meter Calibration Connections........................................................................ Leakage Calibration ............................................................................................... Point to Point Calibration Connections.................................................................. Insulation HV Calibration Connections ................................................................. Resistance Calibration Connections....................................................................... Differential Leakage Calibration Connections....................................................... vii
Page 2-7 3-7 3-9 3-10 3-13 3-15 3-16 3-17 3-18 3-19 3-20 3-21 3-22 3-23 3-24 3-26 3-28 3-29 3-31 3-33 3-35 3-36 3-38 3-39 3-41 3-43 3-46 3-48 3-50 3-54 3-56 3-57 3-58 3-59 3-60
ESA612 Service Manual
3-35. 3-36. 5-1. 6-1. 6-2. 6-3. 7-1. 7-2.
GFI Calibration Connections ................................................................................. Equipment Current Calibration Connection........................................................... Fuse Access............................................................................................................ Final Assembly....................................................................................................... A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................ A1 Power PCA....................................................................................................... A2 Meter PCA........................................................................................................
viii
3-61 3-65 5-3 6-6 6-11 6-19 7-5 7-10
Chapter 1
Introduction and Specifications
Title Introduction.......................................................................................................... Intended Use ........................................................................................................ Electrical Symbols ............................................................................................... Product Manuals .................................................................................................. About this Manual ............................................................................................... Safety Information ............................................................................................... Specifications....................................................................................................... Detailed Specifications ........................................................................................
Page 1-3 1-3 1-3 1-3 1-3 1-4 1-5 1-5
1-1
ESA612 Service Manual
1-2
Introduction and Specifications Introduction
1
Introduction The Fluke Biomedical ESA612 Electrical Safety Analyzer (hereafter the Analyzer) is a full-featured, compact, portable analyzer, designed to verify the electrical safety of medical devices. The Analyzer tests to domestic (ANSI/AAMI ES1, NFPA 99) and international (IEC62353, AN/NZS 3551, and parts of IEC 60601-1) electrical-safety standards. The integrated ANSI/AAMI ES1 and IEC60601-1 patient loads are easily selectable.
Intended Use The Analyzer is intended for use by trained service technicians to perform periodic inspections on a wide range of medical equipment. The testing procedures are menudriven, and simple to operate.
Electrical Symbols Table 1-1 lists the symbols with their definitions found on the Analyzer and used in this manual. Table 1-1. Symbols Symbol
Description
W
Important information; refer to manual.
~
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
;
Conforms to relevant Australian EMC requirements
)
Conforms to relevant Canadian and US standards
X
Hazardous voltage
P
Conforms to European Union directives
CAT II
˜
IEC Measurement Category II – CAT II equipment designed to protect against transients from energy-consuming equipment supplied from fixed installations. Accessible Functional Earth Terminal
Product Manuals The Analyzer ships with a printed Getting Started Manual and a Users Manual on a CD. The ESA612 Getting Started Manual gives a brief overview of the Analyzers controls and connections. The ESA612 Users Manual provides details on operating the Analyzer.
About this Manual This is the Service Manual for the Analyzer. It contains all of the information a technician will need to verify the Analyzer’s performance to specifications and calibration adjustments to bring the Analyzer’s performance within specification. The manual is divided into the following chapters: Chapter 1 “Introduction and Specifications” provides information on how to safely use the Analyzer and specifications. Chapter 2 “Theory of Operation” provides information on how the Analyzer’s circuitry functions. Chapter 3 “Verification and Calibration” provides information on how to verify the 1-3
ESA612 Service Manual
Analyzer’s performance and bring its performance within specification. Chapter 4 “Remote Operation” contains the remote commands to operate the Analyzer through its computer port. Chapter 5 “List of Replaceable Parts” contains a complete list of part numbers with descriptions and associated illustrations to help locate each part. Chapter 6 “Schematics” contain the schematics of the Analyzer.
Safety Information In this manual, a Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data.
XW Warning To avoid possible electrical shock or personal injury, follow these guidelines:
1-4
•
Use this Analyzer only in the manner specified by the manufacturer or the protection provided may be impaired.
•
Read the Users Manual before operating the Analyzer.
•
Do not connect the Analyzer to a patient or equipment connected to a patient. The Analyzer is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the Analyzer would come in contact with a patient.
•
Do not use the product in wet or damp locations, around explosive gases or dust.
•
Inspect the Analyzer before using it. Do not use the Analyzer if abnormal conditions of any sort are noted (such as a faulty display, broken case, etc.)
•
Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Replace damaged leads before using the Analyzer.
•
When testing, always be sure to keep your fingers behind the safety barriers on the test leads.
•
Never open the Analyzer's case. Dangerous voltages are present. There are no user replaceable parts in the Analyzer.
•
Have the Analyzer serviced only by qualified personnel.
•
The Analyzer must be properly earthed. Only use a supply socket that has a protective earth contact. If there is any doubt as to the effectiveness of the supply socket earth, do not connect the Analyzer. Do not use a two-conductor adapter or extension cord; this will break the protective ground connection.
•
Do not use the 15-20 A adapter to power devices rated in excess of 15 A. Doing so may overload the installation.
Introduction and Specifications Specifications
•
Use extreme caution when working with voltages above 30 V.
•
Use the proper terminals, functions and ranges for the test being performed.
•
Do not touch metal parts of the device under test (DUT) during analysis. The DUT should be considered an electrical shock hazard when connected to the Analyzer as some tests involve high voltages, high currents, and/or the removal of DUT earth bond.
Specifications Temperature Operating ............................................................ 10 °C to 40 °C (50 °F to 104 °F) Storage ............................................................... -20 °C to 60 °C (-4 °F to 140 °F) Humidity................................................................. 10 % to 90 % non-condensing
1
Will be using ESA615 specs. So don't waste time marking this up.
Altitude 120 V ac mains supply voltage ........................... 5000 m 230 V ac mains supply voltage ........................... 2000 m Display ................................................................... LCD display Communications................................................... USB device port for computer control Modes of Operation .............................................. Manual and remote Power 120 Volt power outlet .......................................... 90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum 230 Volt power outlet .......................................... 180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum Weight .................................................................... 1.6 kg (3.5 lb) Size......................................................................... 28.5 cm x 17.6 cm x 8.4 cm (11.2 in x 6.9 in x 3.3 in) Safety Standards nd
CE ....................................................................... IEC/EN61010-1 2 Edition; Pollution degree 2 CSA..................................................................... CAN/CSA-C22.2 No 61010-1; UL61010-1 Electromagnetic Compatibility Standards (EMC) European EMC ................................................... EN61326-1
Detailed Specifications Test Standard Selections........................................ ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551 Voltage Ranges (Mains voltage) ...................................... 90.0 to 132.0 V ac rms 180.0 to 264.0 V ac rms Range (Point-to-point voltage)............................ 0.0 to 300.0 V ac rms Accuracy ............................................................. ±(2 % of reading + 0.2 V) Earth Resistance Modes ................................................................. Two terminal Test Current ........................................................ >200 mA ac Range ................................................................. 0.000 to 2.000 Ω Accuracy ............................................................. ±(2 % of reading + 0.015 Ω) Equipment Current Range ................................................................. 0.0 to 20.0 A ac rms Accuracy ............................................................. ±(5 % of reading + (2 counts or 0.2 A, whichever is greater)) Duty cycle ........................................................... 15 A to 20 A, 5 min. on/5 min. off 10 A to 15 A, 7 min. on/3 min. off 0 A to 10 A, continuous
1-5
ESA612 Service Manual
Leakage Current Modes*................................................................ AC+DC (True-rms) AC only DC only * Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP that are available in True-rms (shown as AC+DC) Patient Load Selection........................................ AAMI ES1-1993 Fig. 1 IEC 60601: Fig. 15 Crest factor ......................................................... ≤3 Ranges ............................................................... 0.0 to 199.9 μA 200 to 1999 μA 2.00 to 10.00 mA Accuracy DC to 1 kHz .................................................... ±(1 % of reading + (1 μA or 1 LSD, whichever is greater)) 1 to 100 kHz.................................................... ±(2 % of reading + (1 μA or 1 LSD, whichever is greater)) 1 to 5 kHz (current > 1.6 mA).......................... ±(4 % of reading + (1 μA or 1 LSD, whichever is greater)) 100 kHz to 1 MHz ........................................... ±(5 % of reading + (1 μA or 1 LSD, whichever is greater))
Note Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are +(2.5 μA or 1 LSD, whichever is greater). Mains on applied part test voltage ...................... 100 % ±7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI 100 % ±7 % of Mains for IEC 62353 current limited to 3.5 mA ±25 % per IEC 62353 100 % ±7 % of Mains for IEC 60601-1 current limited to 7.5 mA ±25 % per IEC 60601-1 Differential leakage Ranges ............................................................... 75 to 199 μA 200 to 2000 μA 2.00 to 20.00 mA Accuracy ............................................................. ±(10 % of reading + (2 counts or 20 μA, whichever is greater)) Insulation resistance Ranges ............................................................... 0.5 to 20.0 MΩ 20.0 to 100.0 MΩ Accuracy 20 MΩ Range ................................................. ±(2 % of reading + 0.2 MΩ) 100 MΩ Range ............................................... ±(7.5 % of reading + 0.2 MΩ) Source test voltage ............................................. 500 V dc (+20 %, -0 %) 2.0 ±0.25 mA short-circuit current or 250 V dc selectable Maximum load capacitance ................................ 1 μF ECG Performance Waveforms Accuracy Frequency....................................................... ±2 % Amplitude........................................................ ±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration Waveforms ECG Complex................................................. 30, 60, 120, 180, and 240 BPM Ventricular Fibrillation Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz Triangle wave ................................................. 2 Hz Pulse (63 ms pulse width)............................... 30 BPM and 60 BPM
1-6
Chapter 2
Theory of Operation
Title Introduction.......................................................................................................... Block Diagram Description ................................................................................. A1 Power PCA ................................................................................................ A2 Meter and User Interface PCA .................................................................. Block Diagram to Schematic Cross Reference .................................................... A1 Power PCA Schematic............................................................................... A2 Meter and User Interface PCA Schematic................................................. Detailed Circuit Description ................................................................................ A1 Power PCA ................................................................................................ Equipment Outlet Relays ............................................................................ Equipment Outlet Current Measurements................................................... Mains Measurement .................................................................................... Isolated Mains Voltage (Mains on Applied Parts or MAP) ........................ Connectors .................................................................................................. Isolation....................................................................................................... Relay Drivers .............................................................................................. Isolated Signals ........................................................................................... Power Supplies............................................................................................ Resistance Current Source .......................................................................... Insulation Resistance Test Voltage ............................................................. A2 Meter and User Interface PCA Schematic................................................. Meter Inputs ................................................................................................ Patient Load ................................................................................................ Overvoltage Protection................................................................................ Mains on Applied Parts Compensation....................................................... AC Only ...................................................................................................... DC Only ...................................................................................................... Resistance Measurement ............................................................................. Relay Drivers .............................................................................................. Meter Power Supplies ................................................................................. Relay Disable .............................................................................................. Applied Parts............................................................................................... Meter Microprocessor ................................................................................. User Interface Microprocessor .................................................................... Keyboard and Display................................................................................. USB Port ..................................................................................................... U/I Power Supplies .....................................................................................
Page 2-3 2-3 2-3 2-4 2-5 2-9 2-10 2-11 2-11 2-11 2-11 2-11 2-12 2-12 2-12 2-12 2-12 2-12 2-12 2-13 2-13 2-13 2-13 2-13 2-13 2-14 2-14 2-14 2-14 2-14 2-14 2-14 2-15 2-15 2-15 2-15 2-15 2-1
ESA612 Service Manual
2-2
Theory of Operation Introduction
2
Introduction WX Warning To avoid electric shock, personal injury, or death, use extreme caution when servicing the Analyzer. The Analyzer can generate high voltages and currents. The Analyzer tests electrical safety of medical equipment by making the following measurements: •
Resistance from the case of the instrument under test case to the power cord earth connection.
•
Leakage current to the case of the instrument under test. This and all other leakage currents are measured across one of two standardized patient loads that simulate a human body
•
Leakage current to earth
•
Leakage current to patient electrical connections
•
Leakage current to patient electrical connections when mains voltage is applied to the connections
•
Insulation resistance from mains supply to various instrument under test points
•
Although not a measure of electrical safety, mains voltage and equipment current are also measured
•
These tests are preformed in various configurations as specified by different regulatory agencies. See the ESA612 Users Manual for more information of the agency standards.
Block Diagram Description The Block Diagram shown in Figure 2-1 shows the two printed circuit assemblies (PCAs) and the interconnection between them and the Analyzer connections. A unique number in the upper-left corner of each block in the diagram references the Analyzer assemblies as follows: C = Chassis, P = A1 Power pca, M = A2 Meter and User Interface (Meter) pca. These numbers do not directly correlate to reference designators on the schematics. A1 Power PCA The mains connection is supplied through the Power Entry Module (C5), a 14-gauge cable (C4), a connector (P23), and the fuses (P16). The main power supply is a switching power supply module (P13). It is powered directly from mains power and operates from 90 to 264 V ac. The Analyzer has two circuit sections galvanically isolated from each other; one is earth referenced and is used mainly for the Equipment Outlet circuitry, and the other is used for the Meter circuits. The Earthed +12 V Supply [P13] drives the Isolated ±5 V Regulator (Reg) (P8), the Isolated +3.3 V Reg (P9), the Earthed +3.3 V Reg (P10) and is routed to the Meter PCA at connector (P1). The Isolated ±5 V Reg (P8) supplies power to the 250/500 V Reg (P3) and the 0.2 A Current Source (P4). The 250/500 V Reg is used for insulation resistance tests, which are measured by placing the voltage across the device and measuring the resultant current to calculate the resistance in the range of 0.5 - 100.0 MΩ. The 0.2 A is used for low resistance tests. The current is routed through the device to be measured and the resultant voltage is measured and used to calculate the resistance in the range of 0.0 – 2.0 Ω. The Mains on applied part transformer (P14, MAP XFORMER) has a 1:1 ratio to generate an isolated mains voltage. MAP Current Limit (P11) sets the MAP current limit to 1 mA, 3.5 mA, or 7.5 mA for the various regulatory standards. 2-3
ESA612 Service Manual
The Mains Meter (P15) attenuates mains voltage, measures it, and converts it to digital format. The digital signal is routed to the meter microprocessor (M19) on the A1 Meter pca through The Digital Isolator (P12) which provides galvanic isolation between the mains attenuator and the meter. The Test Setup Relays (P5) connect the various current sources and voltages for different test setups. The relays are controlled by relay drivers (P6), which are isolated by the digital isolators (P7). The Equipment Outlet (C7) is connected to the A1 Power pca through a 14-gauge cable (C6) and a connector (P24). Power relays are used to control the configuration of the Equipment Relay. Reverse Relays (P20) are for reversing hot and neutral. On/Off Relays (C19) are for opening hot, neutral or earth. There is a solid-state relay (P18), which is turned off before changing the state of the other equipment outlet relays and then turned on again to prevent arcing on loaded relays. The other set of reverse relays (P17) are used to reverse the polarity of incoming power if the Analyzer senses mains are reversed at the main input, thus providing extra relay protection by insuring that hot is routed through the solid state relay. Two feed-through transformers are connected in front of the equipment outlet to monitor current drawn by the device under test. The Differential Current Transformer (P22, DIFF CUR XFORMER) monitors the difference in current between hot and neutral for the differential current test and for ground fault current detection. These signals are routed to the test setup relays (P5). There are two cables between the two pcas, a 24-gauge cable (C2) between the connector (P1) and the connector (M34) for power signals and a Ribbon Cable (C3) connected between connector (P2) connector (M35) for digital signals. A2 Meter and User Interface PCA Power for the A2 Meter and User Interface (Meter and U/I, or Meter) PCA is supplied by U/I Supplies (M22) and Meter Supplies (M23). These are powered from the Earthed +12 V Supply (P13) on the A1 Power PCA. The U/I microprocessor (M14) operates the various user interface components. It writes to the display driver (M9) which controls the display (C1) through connector (M7). The display contrast is set by display contrast (M10). The U/I microprocessor reads the keys (M1) and sets the warning LED (M2). In addition, it controls the Audio Indicator (M8). It communicates with the meter microprocessor (M19) using an RS232 format through the digital isolators (M18). The U/I microprocessor also communicates with the port using an RS232 format through the digital isolators (M15) and the RS232 to USB converter (M11). The meter microprocessor (M19) sets up tests by controlling all the relay drivers on both pcas (M17, M20, P6). The measured signal comes from several places: the meter jacks (M5), applied parts (AP) jacks (M4), and the equipment outlet (C7). These are routed through the test setup relays (M21) to the patient load (M32) and then to the buffer amplifier (M30, BUFFER AMP). The signal is then separated by the AC Only Filter (M26) and the dc Only Filter (M29). The ac signal is converted to a dc voltage proportional to the rms value by rms converter (M25) and converted to digital format by the multi-channel analog-to-digital converter (ADC) (M24). The meter microprocessor (M19) reads the ADC (M24), formats it for the specific measurement being taken, and sends it to the U/I microprocessor (M14). The BUFFER AMP (M30) is protected by the over-voltage detect (M31) which can open all relays through relay driver disable (M28). The ground Fault Detect (M33, GFI) receives a signal from the DIFF CUR XFORMER (P22) on the A1 Power PCA, and if the signal is over the programmable threshold, it shuts down the relays through the relay driver disable (M28). 2-4
Theory of Operation Block Diagram to Schematic Cross Reference
2
The AP jacks (M4) are connected in various combinations to measurement circuit points depending on the test being performed. This is done by the relays (M13) and relay drivers (M17). The ECG waveform from the meter microprocessor (M19) is buffered and amplified by the ECG amplifier (M16, ECG AMP) and attenuated to the low voltages required for the ECG signals by the ECG Attenuator (M12, ECG ATTEN). The null jack (M6) is connected only during the ground resistance check and is used to zero out the resistance of the test leads.
Block Diagram to Schematic Cross Reference Tables 2-1 and 2-2 refer to one of the two schematics. The number in the first column references the number of a block on the block diagram (Figure 2-1). The second column is the page number for that pca’s particular schematic. The third column references the main parts on the schematic for that circuit. Some block diagram blocks do not have parts but are included for clarity only.
2-5
ESA612 Service Manual
2-6
Electrical Safety Analyzer Schematic Diagrams
2
ESA612
A2 METER & USER INTERFACE PCA
M14
M8
AUDIO INDICATOR
M1
M2
U/I MICROPROCESSOR
KEYS
M22
M9
M18
M7
DISPLAY
C O N N
C O N N
P1 C2
24G CABLE
C O N N
P3
250/500V REG
METER SUPPIES
P4
0.2A CURRENT SOURCE
ADC
M10
USB CONN
RS232 TO USB CONV
M13
AP JACKS
RELAYS
M15
DIGITAL ISOLATORS
METER MICROPROCESSOR
DC ONLY FILTER
AC ONLY FILTER
TEST SETUP RELAYS
M30
BUFFER AMP
M16
POWER ENTRY
P17
REVERSE RELAYS
M17
RELAY DRIVERS
P11
P14
P12
P15
DIGITAL ISOLATORS
MAP XFORMER
MAINS METER
P18
P19
SOLID STATE RELAY
ON/OFF RELAYS
P6
RELAY DRIVERS
RELAY DRIVERS
M32
PATIENT LOAD
TEST SETUP RELAYS
P21
DIFF CUR XFORMER
P7 M35
NULL JACK
C5
REVERSE RELAYS
OVER VOLTAGE DETECT
M5
M6
14G CABLE
P20 M31
ECG AMP
M21
METER JACKS
C4
ISOLATED +3.3V REG
MAP CURRENT LIMIT
M29
M26
M20 M4
FUSES
C O N N
RMS CONV
MAP NULL
ECG ATTEN
P16
P9
EARTHED +3.3V REG
M27
M12
EARTHED +12V SUPPLY
P10
M25
DISPLAY CONTRAST
M11
ISOLATED +/-5V REG
P23
P13
M24
P5
M3
P8
M23
M19
DISPLAY DRIVER
M34
U/I SUPPLIES
DIGITAL ISOLATORS
WARNING LED
C1
A1 POWER PCA
M28
RELAY DRIVER DISABLE
M33
GFI DETECT
C O N N
P2 C3
RIBBON CABLE
C O N N
DIGITAL ISOLATORS
P24 P22
CUR XFORMER
C O N N
C6
14G CABLE
C7
EQUIP OUTLET
Figure 2-1. ESA612 Block Diagram
2-7
ESA612 Service Manual
2-8
Theory of Operation Block Diagram to Schematic Cross Reference
2
A1 Power PCA Schematic Table 2-1 lists the cross-references for the A1 Power PCA schematics. Table 2-1. A1 Power PCA to Schematic Cross Reference Block Number
Schematic Page
Reference Designators
P1
3
J2
P2
3
J4
P3
4
T2 U17 U24
P4
4
U23 Q4
1
K8 K9 U8 K12 K14
2
K17 K16
4
K13
P6
3
U14 U13 U16 U15 U25
P7
2
U2
P8
4
U18
P9
3
U28
P10
3
U12
P11
2
Q1 Q2
P12
3
U5
P13
4
U9
P14
2
T1
P15
1
R2-R7 R17 R18 U4 U3
P16
1
F1-F3
P17
1
K1 K2
P18
1
K3 K19
P19
1
K4 K5
P20
1
K6 K7
P21
1
L1 U1
P22
1
L3
P23
1
J5
P24
1
J1
P5
2-9
ESA612 Service Manual
A2 Meter and User Interface PCA Schematic Table 2-2 lists the cross-references for the A2 Meter and User Interface PCA Schematics. Table 2-2. A2 Meter and User Interface PCA Schematic Cross Reference Block Number
Reference Designators
M1
7
S3 S4
M2
6
DS1 Q1
M3
8
J3
M4
4
J60-J64
M5
1
P12 P13
M6
1
P24
M7
7
J1
M8
6
LS1 Q2
M9
7
U3
M10
7
U6 U5
M11
8
U7
M12
4
R53-R61
M13
4
K1-K10 K12 K22 K23
M14
6
U2
M15
8
U8
M16
4
U13
M17
4
U32 U33 U34 U35
M18
5
U42
M19
5
U40
M20
3
U35 U36 U55 U56 U38
1
K13 K15-K19 K21U27
2
U25 U39
M22
8
U11 U10
M23
3
U48-U52
M24
2
U24
M25
2
U23
M26
2
U28 U26
M27
1
U21 U22
M28
3
U47 U49 U14
M29
2
U29 U31
M30
2
U54
M21
2-10
Schematic Page
Theory of Operation Detailed Circuit Description
2
Table 2-2. A2 Meter and User Interface PCA Schematic Cross Reference (cont.) Block Number
Schematic Page
Reference Designators
M31
1
U18
M32
1
R78 R79 C100 R156
M33
3
U16 U17
M34
4
J20
M35
4
J4
Detailed Circuit Description The following section describes the operation of the Analyzer circuits. A1 Power PCA The A1 Power PCA contains the power supplies, transformers, power relays, Equipment Outlet, and some of the test selection relays. Equipment Outlet Relays The circuitry along the top of the schematic on sheet 1 of the A1 Power PCA schematic controls the Equipment Outlet (EO). Mains is connected to the pca at J5. The EO is protected by fuses F1 and F2. Relays K1 and K2 reverse the polarity of mains in the event it is reversed at the input. Following this is a solid state relay, K19. This is turned off when any of the mechanical relays controlling the EO are switched, and then it is turned on again. This protects the mechanical relays from damage caused by inductive or incandescent loads or shorts. In the case of an EO short, the solid state relay can pass up to 900 amps for the time it takes the fuses (F1 and F2) to open. Relay K3 shorts out the solid state relay after all relay switching is done to keep power dissipation low. Relays K4 and K5 turn on the EO. Relays K6 and K7 reverse the polarity. EO ground is opened by relay K11. Relay K12 connects the resistance measuring, current source return to the EO ground pin for earth resistance tests. Relay K14 connects the EO ground pin to the meter for earth leakage tests. Relay K8 connects the EO hot and neutral pins together for leakage and insulation resistance tests. Relays K9 and K10 connect the shorted EO hot and neutral pins to the meter, 500 V or the MAP voltage as required. Equipment Outlet Current Measurements Transformer L1 is a differential transformer with hot and neutral running through it. The difference in current between these is detected by the transformer. U1 converts the current to a voltage that is used for differential leakage measurements. This is selected for measurement by connecting it to PWR_GRP by closing U8 pins 14 and 15. The output of U1 is also connected to GFI_SENSE for measurement by the GFI Detect circuit on the A2 Meter PCA. Transformer L3 is a current transformer that monitors the EO current for the equipment current measurement. This is selected for measurement by U8 pins 2 and 3. Mains Measurement Hot is routed through fuse F3 to protect the Analyzer’s internal circuitry from excessive current. Mains voltages are measured and used to detect open ground, reverse polarity, and out of range conditions. Resistors R2, R3, R4, and R17 attenuate the hot to ground voltage and U4B buffers it. The neutral to ground voltage is similarly attenuated by resistors R5, R6, R7, and R18 then buffered by U4A. These analog signals are converted 2-11
ESA612 Service Manual
to digital signals by the two-channel analog to digital converter (U3, ADC). This has a two-wire serial port containing a clock and data line and a third line for chip selection. See page 3 of the A1 Power PCA schematic for further routing of the digital signals to and from the ADC. U7 monitors the hot to ground and neutral to ground signals to generate a mains zero-cross signal used by the meter microprocessor. Isolated Mains Voltage (Mains on Applied Parts or MAP) Transformer T1 is a 1:1 transformer that generates the isolated mains voltage used by many of the Analyzer tests. Transistors Q1 and Q2 form an active current limit circuit. The current limit can be set to 3 different values, 1 mA, 3.5 mA, and 7.5 mA depending on the selected international standard. U10 selects 3.5 mA, U29 selects 7.5 mA. As the current increases, the voltage across the selected resistor increases until it reaches about 7.5 V. At this point, Q2 begins to conduct preventing Q1 from being turned on any further. This clamps the current, but the voltage across Q1 can continue to increase. The bridge CR12-CR15 allows the circuit to operate with both polarities of current. Relay K17 connects the output to the test circuits and relay K16 reverses the polarity of the voltage. Connectors There are two connectors between the A1 Power PCA and the A2 Meter PCA. Jack J2 is a 6-pin 22-guage wire connector for high current signals. Jack J4 is a 20-pin ribbon cable for digital signals and low-current power supplies. Isolation To maintain galvanic isolation between the meter circuits and the earthed circuits, the digital signals are isolated by digital isolators U2 and U5. U2 isolates the relay driver signals and U5 isolates the Mains ADC signals. Optical isolators are used to isolate signals not needing the bandwidth of the digital isolators. U11 isolates the relay driver strobe and U6 isolates the mains zero cross signal. Power for the earthed side of the isolations is from the +3.3V supply U12. Power for the isolated side is from the +3.3V supply U28. Relay Drivers U14 and U16 are serial-in parallel-out shift registers. These have a three-wire serial interface with clock, data, and strobe lines. A fourth line, output enable (/OE), is used to enable or disable all the outputs and thus the relays. The outputs of the shift registers control the relay drivers U13 and U15. Isolated Signals Meter signal lines that control circuits on the isolated side are expanded by shift register U25, like the relay driver shift registers. Power Supplies The main power supply for the Analyzer is U9. This is a switching supply that can be driven from an ac source between 90 – 264 V ac and 50 - 60 Hz. The output is +12 V and is isolated from mains. U18 is an isolated dc to dc converter module with output switching ±5 V dc. It powers the resistance measuring current source and the insulation resistance voltage source. Resistance Current Source Resistance is measured by applying a current to the resistance to be measured and measuring the voltage developed across it. The Analyzer uses a 200 mA ac source. U23 and Q4 form a precision current source. The current is set to 400 mA by the 1 Ω resistance made up of R85, R86, R76, and R59. U23 sets the voltage across the resistors equal to the reference voltage at pin 3. The reference voltage is from U22, a 1.2 V 2-12
Theory of Operation Detailed Circuit Description
2
reference, divided down to 0.4 V by R77 and R78. Q4 provides the high output impedance needed for a true current source. Isolator U19 is switched on and off to turn the current source on and off to make the 200 mA ac current. Insulation Resistance Test Voltage The 500 V dc voltage source uses a transformer to boost the voltage. A pulse width modulator drives the primary of transformer T2 to set the secondary voltage. CR5 and CR10 rectify the secondary and C29 and C42 filter it, resulting in 500 V dc. The voltage is attenuated to 5 V by resistors R41, R56, R57, and R46. U17 compares this to its 5 V reference voltage and sets the width of the output pulses to maintain the divided voltage at 5 V. The voltage can be set to 250 V by turning off Q5 and turning on Q6. This changes the attenuator such that the output voltage is 250 V. U24 senses the return current to the transformer and shuts off the supply if it exceeds 1.5 mA. Relay K13 connects the supply for tests and when not selected, connects a resistance to ground to discharge capacitance in the device under test, R63 and R64. A2 Meter and User Interface PCA Schematic The A2 Meter and User Interface PCA contains the two microprocessors, the meter circuit, test jacks, keyboard USB port and display driver. Meter Inputs The meter input jacks are P12 and P13. These are connected for various tests by relays. Relay K13 connects the red jack to the patient load for leakage tests. Relay K15 connects the red jack to a voltage attenuator used for the Point to Point voltage. Relay K16 connects the red jack to earth through a current limiter. Relay K19 connects the red jack to the resistance current source. Relay K17 connects the black jack to the negative input of the meter, –Meter (GNDMA). Relay K21 connects the black jack to the resistance current source return. The null post, P24, is connected to –Meter (GNDMA) by relays K17 and K18 for nulling the test lead resistance for earth resistance measurements. Patient Load Leakage currents are measured across the patient load. Resistors R78 and R79 form the 1-kΩ resistance for the input of the AAMI and IEC60601 patient loads. For leakage currents greater than 2 mA, the meter input is switched from resistors R78 and R79 to divide the voltage by five. Resistor R68, capacitor C100, and resistor R156 form the rest of the patient load. Closing U27 pins 14 and 15 forms an AAMI load. Closing U27 pins 10 and 11 forms an IEC60601 load. Closing U27 pints 6 and 7 connects the load to the measurement circuits on page 2 of the meter schematic. With this open and U27 pins 2 and 3 shorted, the differential current and equipment current from the A1 Power PCA are selected. Overvoltage Protection Small over-voltages at the meter inputs are clamped by CR9, VR6, and VR7. Severe over-voltages over about 50 V are sensed through the attenuator U83, R86, and U18. This is routed to schematic page 2. Mains on Applied Parts Compensation The isolated mains voltage (MAP) used in many AP leakage tests causes a residual leakage that must be compensated for. Digital to analog converter (DAC) U21 and amplifier U22 generate an ac voltage waveform that accurately tracks the MAP voltage. This is summed with the meter input leakage current through C76. The digital value set in U21 sets the amplitude of this signal which is adjusted to cancel the residual leakage.
2-13
ESA612 Service Manual
AC Only U54 buffers the signal from the meter input circuitry. In the ac only mode, the signal is routed through U39 pins 10 and 11 to the ranging amplifiers U28 and U26. The gain is selectable to be 0.15, 1.5 or 15 by the state of analog switches U25. This signal is then AC coupled by C77 to the RMS converter U23. This generates a dc voltage equivalent to the rms value of the input signal. U31A forms a two-pole-low-pass filter to smooth out the signal from U23. Then the analog to digital converter (ADC) U24 converts this to digital format which is read serially by the meter microprocessor. DC Only The signal from buffer U54 is routed through U39 pins 14 and 15 for dc only mode. U29 and U31B form a five-pole low-pass filter with a cutoff frequency of 5 Hz. This removes all mains and higher frequencies. The resultant dc is then routed to the ADC. Resistance Measurement The current source used for resistance measurement is on the A1 Power PCA and is described above. U15 inverts the signal and U39 pins 14 and 15 and U39 pins 2 and 3 are toggled to rectify it. The signal at U46 pins 12 and 13 blanks both polarities for a period after the current is switched to eliminate inductive and capacitive effects. Relay Drivers U37, U36, U55, and U56 drive the relays on the A2 Meter PCA. The circuit is similar to the relay drivers section described in the A1 Power PCA section. Meter Power Supplies The meter is galvanically isolated from the earthed circuits. The power supplies are isolated by using isolated dc to dc converter modules. U48 is an isolated dc to dc converter modules that generates the +12 V isolated supply from the earthed +12 V supply for the relays. U51 is another dc to dc converter that outputs +5 V. U50 reregulates this down to +3.3 V for the meter microprocessor. U49 outputs ±12 V used by the amplifiers and analog switches on the meter input. U53 and U52 re-regulate these to ±5 V for the remainder of the amplifiers. Relay Disable The overvoltage signal from page 1 of the meter schematic, /FAULT_SENSE, sets the RS flip-flop composed of U47A and B. This signal causes the /RELAY_EN (not relay enable) signal at U19 pin 13 to go high, shutting off all relay drivers in the instrument. To clear this fault, the meter microprocessor sets pin U47 pin 2 low to clear the flip flop. U14 is a brown-out detector which will also set the flip-flop at power up and power down to prevent the relays from being in undefined states while the microprocessor is powering up and down. The /GFI_SENSE signal from the differential transformer on the power PCA is at mains frequency analog voltage. This is rectified and filtered by CR8 and C67. DAC U16 generates a digitally controlled voltage that is compared to the GFI signal at U17. When the voltage proportional to the GFI current exceeds the preset voltage, the output of U17 goes low setting the flip-flop, U47C and D, and, in turn, disabling the relays. Applied Parts The applied parts relay drivers, U32, U33, U34, and U35, operate as described previously. Each applied part post can be connected to several circuit points: Open, ECG Wave, Earth, +Meter, -Meter or 500 V. For the V1 post this is performed by relays K1, K2, K12, and K11. The other apples parts jacks operate similarly.
2-14
Theory of Operation Detailed Circuit Description
2
Meter Microprocessor U40 is the meter microprocessor. It is a single-chip controller with built-in Program memory (FLASH), RAM, serial peripheral interfaces, and timers. Calibration data and configuration data is stored in an external serial EEPROM (U44). The processor is clocked at 12 MHz by the crystal, Y2. The two microprocessors in the Analyzer communicate over a galvanically isolated serial port with U42 providing the isolation. The U/I processor is the master and can reset the meter microprocessor through U43. U41 is a brownout detector that resets the processor on power up and power down. The JTAG port (J21) is used for debugging and downloading firmware. User Interface Microprocessor U2 is the user interface microprocessor. It contains program memory (FLASH), RAM, timers, I/O ports, and communication ports. It is clocked at 12 MHz by the crystal, Y1. It runs off of the +3.3 V power supply and an internally generated +1.8 V supply. U1 is a serial FLASH memory for storing test data. U59 is an FRAM for temporary test data storage. Only entire test records are transferred to the FLASH memory. The warning LED DS1 is turned on by the U/I microprocessor to indicate when there are dangerous voltages or currents on any test jacks. The beeper, LS1, beeps as required. The internal temperature of the unit is monitored by RT1. At a preset temperature, the U/I microprocessor informs the meter microprocessor to shut down the relays and a message is placed on the display. The JTAG port (J2) is used for debugging and downloading code into FLASH memory. Keyboard and Display The keyboard and display are controlled by the U/I microprocessor. The Keyboard, S3 and S4, is arranged in rows and columns. It is read by consecutively setting each row line low and then reading the columns. The key pressed is at the intersection of the row and column. The display is controlled by the Display controller, U3, which is connected between the U/I microprocessor and the display at J1. Quad op-amp U6 generates the voltages needed to bias the LCD display. Switching voltage supply U5 generates the highest voltage in this string, and by varying this, the contrast is varied. This is varied by the CONTRAST signal at R22. The display can be shut off by the ENABLE_20 V signal which disconnects the +12 V supply from the +20 V supply. The display backlight can be turned on and off by Q3. USB Port U7 and U8 form the USB port. U8 provides galvanic isolation between the Analyzer and the communication port. U7 communicates with the U/I microprocessor in RS232 format. It converts incoming data from USB to RS232 and outgoing data from RS232 to USB format. U9 protects the port from ESD. The connector, U3, is a standard “type B” upstream pointing USB connector. The components on the USB side of the isolator, U8, are bus powered. U/I Power Supplies U11 is a switching regulator that generates the +3.3 V supply for the U/I microprocessor. U10 is a linear regulator that generates a 3.0 V supply for the display driver.
2-15
ESA612 Service Manual
2-16
Chapter 3
Verification and Calibration
Title Introduction.......................................................................................................... Required Equipment ............................................................................................ Analyzer Verification .......................................................................................... Outlet Tension Tests........................................................................................ Setup ................................................................................................................ Verification Procedure..................................................................................... Calibration Procedure .......................................................................................... Setup ................................................................................................................ Adjustments.....................................................................................................
Page 3-3 3-3 3-4 3-4 3-4 3-5 3-52 3-52 3-52
3-1
ESA612 Service Manual
3-2
Verification and Calibration Introduction
3
Introduction This section provides performance tests to verify the Analyzer is operating within published specifications and a calibration procedure for bringing the Analyzer within specifications. The verification procedure and, if necessary, the calibration procedure can be performed both periodically and after service or repair.
Required Equipment Table 3-1 lists the equipment required for performance testing and calibrating the Analyzer. Table 3-1. Required Equipment for Analyzer Verification Instrument Type
Recommended Model
Digital Multimeter
Fluke 8846A
Calibrator
Fluke 5520A
AC and DC Current Source
Keithley 6221
Programmable AC Power Supply
Elgar CW 1251
Personal Computer with Hyperterminal Receptacle Wiring Tester
Woodhead 1750
Oscilloscope
TEK TDS 3012B
Current Shunt
Fluke A40A, 20 A capable
Current Load Differential Amplifier
AM Systems 3000 AC-DC
Resistor Decade Box: 1.800 Ω, 1.400 Ω, and 0.450 Ω rated at 1 watt
General Resistance RDS-41A
Resistor: 110 kΩ rated at ½ watt Resistor Decade Box: 0.7 MΩ, 1.0 MΩ, 2.0 MΩ, 3.1 MΩ, 6.5 MΩ, 10 MΩ, IET Labs HRRS-B-7-1K 18 MΩ, 22 MΩ, 60 MΩ, and 100 MΩ rated at 500 V and 1.5 mA Resistor: 0.1 Ω rated at 75 watts Outlet Tension Tester
Extech 475040
USB Cable Test Leads
3-3
ESA612 Service Manual
Analyzer Verification The following sections cover the setup and procedural steps for verifying the Analyzer is operating within the product specifications. Outlet Tension Tests To test the equipment outlet for proper tension: 1. Insert the tension tester into the hot (L1) pin of the Analyzer test receptacle. Set the tension tester to measure peak ounces. 2. Zero the tension tester indicators. 3. Slowly pull the tension tester out of the receptacle. 4. Read and record the reading on the tension tester. 5. Repeat steps 1 through 4 on the neutral and ground pins of the test receptacle. Setup Perform the following steps to prepare the Analyzer for performance verification. 1. Power on all test equipment and allow a 1-hour warm up time. 2. Power on the PC and log in.
3. Connect the PC USB cable to the Analyzer. 4. Power on the Analyzer with the programmable ac power supply. 5. Establish serial communication with the Analyzer (Applicable COM port using USBView, 115200 Baud, N,8,1 bits, HW Flow control) Under File→ Properties and the Settings tab, click the ASCII Setup button, and ensure “Send line ends with line feeds” and “Echo typed characters locally” are CHECKED. 6. If the Analyzer is a non-US unit, the firmware must be switched to indicate a US version. a. Send command: REMOTE=F2810
b. Receive * c. Send command: NOMINAL?
d. Receive the current nominal voltage setting. Note this value for reconversion at the completion of verification. e. Send command: NOMINAL=115 f.
Receive *
g. Send command: LOCAL h. Receive *
3-4
Verification and Calibration Analyzer Verification
3
Note Many tests are preceded by the HIGH_RES-ON, and STD=XXX commands. These settings will not be “lost” when switching between tests, and are not required to be repeated if the Analyzer hasn’t been taken out of remote mode. They are intended to allow the procedure to be used in part or in whole for testing. Use them as needed throughout this procedure. When an MREAD is used to report a reading, the tenth reading is often specified. This is to allow settling time for the Analyzer. However, any reading after stabilization is sufficient. Verification Procedure The following steps make up the procedure for verifying the operation of the Analyzer and coincide with the verification datasheet found in Appendix A. A. Functional 1. Send command: REMOTE
2. Receive * 3. Send command: DIAG=F2810
4. Receive * 5.
Send command: IDENT
6. Verify that the Analyzer identifies itself with UI & MTR firmware revision level. Record UI & MER firmware revisions. 7. Send command: LED=ON
8. Receive * 9. Verify that the red warning LED is lit. 10. Send command: LED=OFF
11. Receive * and verify the red warning LED is off. Record PASS/FAIL. 12. Send command: BEEP
13. Receive * and verify the beeper beeps once. Record PASS/FAIL. 14. Send command: CONT=09
15. Receive * 16. Send command: CONT=07
17. Receive * 18. Send command: CONT=05
19. Receive * 20. Send command: CONT=03
21. Receive * 22. Send command: CONT=09
23. Receive * 24. Send command: CONT=11
25. Receive * 26. Send command: CONT=13
3-5
ESA612 Service Manual
27. Receive * 28. Send command: CONT=15
29. Receive * 30. Send command: CONT=09
31. Receive * 32. Verify that the contrast gets lighter for lower values of CONT and darker for higher values. Record PASS/FAIL. 33. Send command: KEY
34. Receive * a. Depress all keys on the keypad and check for the correct returned digit in Table 3-2. Table 3-2. Keypad Verification Key
Response
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
35. Press the ESC key, receive a carriage return only. Record PASS/FAIL 3-6
Verification and Calibration Analyzer Verification
3
36. Send command: PON 37. Receive * 38. Send command: POFF 39. Receive * 40. Verify that all pixels on the display came on and then went off. Record PASS/FAIL. 41. Send command: LITE=OFF 42. Receive * 43. Send command: LITE=ON 44. Receive * 45. Verify that the backlight goes off and comes back on. Record PASS/FAIL 46. Send command: EXIT 47. Receive * 48. Send command: LOCAL 49. Receive * 50. Send command: REMOTE 51. Receive * B. Electrical Outlet Polarity Testing 1. Connect the receptacle wiring tester to the equipment outlet (EO) as shown in Figure 3-1.
Fluke Biomedical ESA612
Receptacle Wiring Tester
Figure 3-1. Electrical Equipment Outlet Polarity Test
2. Send command: ENCL
gfi102.eps
3-7
ESA612 Service Manual
3. Receive * 4. Send command: POL=N
5. Receive * 6. Send command: NEUT=C
7. Receive * 8. Send command: EARTH=C
9. Receive * 10. Verify the yellow and white lights are lit. Record PASS/FAIL. 11. Send command: EARTH=O
12. Receive * 13. Verify the yellow light is lit. Record PASS/FAIL. 14. Send command: NEUT=O
15. Receive * 16. Send command: EARTH=C
17. Receive * 18. Verify the white light is lit. Record PASS/FAIL. 19. Send command: LOCAL
20. Receive * 21. Send command: REMOTE
22. Receive * 23. Send command: ENCL
24. Receive * 25. Send command: POL=R
26. Receive * 27. Verify the red light is lit. Record PASS/FAIL. 28. Send command: NEUT=C
29. Receive * 30. Verify the red and yellow lights are lit. Record PASS/FAIL. 31. Send command: EARTH=O
32. Receive * 33. Verify the yellow light is lit. Record PASS/FAIL. 34. Send command: LOCAL
35. Receive * 36. Send command: REMOTE
37. Receive * 38. Remove all testing connections from the UUT. 3-8
Verification and Calibration Analyzer Verification
3
C. Electrical Outlet Ground to NULL Jack Resistance 1. Connect the DMM Input (+) to the NULL Jack and the Input (-) to the EO Ground as shown in Figure 3-2. Select the 1 GΩ range on the DMM. Fluke 8846A INPUT
Fluke Biomedical ESA612
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
gfi103.eps
Figure 3-2. Electrical Outlet Ground to NULL Jack Resistance
2. Send command: CAL=F2810
3. Receive * 4. Send command: EOGNULL=O
5. Receive * 6. If the DMM reads OPEN, record PASS. Otherwise, record FAIL. 7. Change the DMM to the 10 Ω range. Send command: EOGNULL=C
8. Read and record the DMM resistance value. 9. Send command: EXIT
10. Receive * 11. Disconnect all testing connections to the UUT. D. Mains Voltage 1. Send command: HIGH_RES=ON
2. Receive * 3. Send command: STD=AAMI
4. Receive * 5. Send command: MAINS=L1-L2
6. Receive * 7. Send command: READ
8. Receive and record response. 9. Send command: MAINS=L1-GND
3-9
ESA612 Service Manual
10. Receive * 11. Send command: READ
12. Receive and record response. 13. Remove all testing connections from the UUT. E. GFI Verification 1. Connect the current source between EO Neutral and EO ground as shown in Figure 3-3. Set the current source for 4.5 mA (6.364 mA p-p), 60 Hz with a compliance voltage of 40 V.
Fluke Biomedical ESA612
Keithly 6221
Figure 3-3. GFI Verification Connections
2. command: PPV
gjc141.eps
3. Receive * 4. Send command: GFI=5MA
5. Receive * 6. Send command: EQCURR
7. Receive * 8. Turn on the current source output. 9. Send command: STAT3
10. Receive status word 3. The 4xxx bits will not be set if the Analyzer passes the test. 3-10
Verification and Calibration Analyzer Verification
3
Note If the 8xxx and 4xxx bits are set, the status word returned will be C000. This still indicates the 4xxx bit is set, and is a failing condition. 11. Record PASS/FAIL. 12. Turn off the current source output. 13. Send command: GFIR
14. Receive * 15. Set the current source to 5.5 mA (7.778 mA p-p), 60 Hz. 16. Send command: GFI=5MA
17. Receive * 18. Send command: EQCURR
19. Receive * 20. Turn on the current source output. 21. Send command: STAT3
22. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 23. Record PASS/FAIL. 24. Turn off the current source output. 25. Send command: GFIR
26. Receive * 27. Set the current source to 9 mA (12.727 mA p-p), 60 Hz, with a 40 V compliance voltage. 28. Send command: PPV
29. Receive * 30. Send command: GFI=10MA
31. Receive * 32. Send command: EQCURR
33. Receive * 34. Turn on the current source output. 35. Send command: STAT3
36. Receive status word 3. The bit xxx8 shall be set, indicating medium GFI level, but the 4xxx bit will not be set if the Analyzer passes this test. 37. Turn off the current source output. 38. Record PASS/FAIL. 39. Send command: GFIR
40. Receive * 3-11
ESA612 Service Manual
41. Set the current source to 11 mA (15.556 mA p-p), 60 Hz. 42. Send command: GFI=10MA
43. Receive * 44. Send command: EQCURR
45. Receive * 46. Turn on the current source output. 47. Send command: STAT3
48. Receive status word 3. The 4xx8 bit will be set if the Analyzer passes this test (GFI tripped). 49. Turn off the current source output. 50. Record PASS/FAIL. 51. Send command: GFIR
52. Receive * 53. Set the current source to 22.5 mA (31.82 mA p-p), 60 Hz. 54. Send command: PPV
55. Receive * 56. Send command: GFI=25MA
57. Receive * 58. Send command: EQCURR
59. Receive * 60. Turn on the current source output. 61. Send command: STAT3
62. Receive status word 3. The 4xxx bit will not be set if the Analyzer passes this test. 63. Turn off the current source output. 64. Record PASS/FAIL. 65. Send command: GFIR
66. Receive * 67. Set the current source to 27.5 mA (38.89 mA p-p), 60 Hz. 68. Send command: GFI=25MA
69. Receive * 70. Send command: EQCURR
71. Receive * 72. Turn on the current source output. 73. Send command: STAT3
3-12
Verification and Calibration Analyzer Verification
3
74. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 75. Turn off the current source output. 76. Record PASS/FAIL. 77. Remove all testing connections. 78. Send command: GFIR
79. Receive * F. Point to Point Voltage 1. Send command: HIGH_RES=ON
2. Receive * 3. Send command: STD=AAMI
4. Receive * 5. Send command: PPV
6. Receive * 7. Connect the Normal Outputs of the calibrator to the red and black jacks on the Analyzer as shown in Figure 3-4. Set the calibrator to 1 V, 120 Hz, and set the calibrator to OPR. Fluke 5520A 5520A CALIBRATOR
Fluke Biomedical ESA612 NORMAL V, , ,RTD
AUX
SCOPE
A, -SENSE, AUX V
OUT HI
LO
TRIG
20A
GUARD
20V PK MAX
TC
20V PK MAX
Figure 3-4. Point to Point Voltage Verification Connections
8. Send command: READ
gjc105.eps
9. Receive and record response. 10. Set the calibrator to 250 V, 120 Hz. 3-13
ESA612 Service Manual
11. Send command: READ
12. Receive and record response. 13. Set the calibrator to 100 V, 633 Hz. 14. Send command: READ
15. Receive and record response. Based upon the design of the patient load, this value is expected to have a SQRT(2)/2 drop from the source voltage. 16. Set the calibrator to 4 V, 120 Hz. 17. Send command: READ
18. Receive and record response. 19. Set the calibrator to 8 V, 120 Hz. 20. Send command: READ
21. Receive and record response. 22. Set the calibrator to 10 V, 120 Hz. 23. Send command: READ
24. Receive and record response. 25. Set the calibrator to 25 V, 120 Hz. 26. Send command: READ
27. Receive and record response. 28. Set the calibrator to 40 V, 120 Hz. 29. Send command: READ
30. Receive and record response. 31. Set the calibrator to 80 V, 120 Hz. 32. Send command: READ
33. Receive and record response. 34. Set the calibrator to 130 V, 120 Hz. 35. Send command: READ
36. Receive and record response. 37. Set the calibrator to 240 V, 120 Hz. 38. Send command: READ
39. Receive and record response. 40. Turn the calibrator output to STBY. Disconnect all testing connections to the UUT. G. Resistance 1. Connect two test leads to the red and black jacks. Short these two leads together as shown in Figure 3-5.
3-14
Verification and Calibration Analyzer Verification
3
Stackable Banana Plugs
Fluke Biomedical ESA612
Figure 3-5. Point to Point Resistance Verification Connections
2. Send command: HIGH_RES=ON
gfi108.eps
3. Receive * 4. Send command: STD=AAMI
5. Receive * 6. Send command: PPR
7. Receive * 8. Send command: ZERO
9. Receive * 10. Send command: READ
11. Receive and record response. 12. Connect the two leads across a 1.800 Ω resistor. 13. Send command: READ
14. Receive and record response. Compare with the actual value of the resistor. 15. Move the lead from the black jack to the equipment outlet ground. Short the two leads together again. 16. Send command: ERES
17. Receive * 18. Send command: ZERO
19. Receive * 20. Send command: READ
3-15
ESA612 Service Manual
21. Receive and record response. 22. Connect these same two leads across a 1.800 Ω resistor. 23. Send command: READ
24. Receive and record response. Compare to the actual value of the 1.800 Ω resistor. 25. Remove all testing connections to the UUT. H. Insulation Resistance 1. Connect the EO ground to the Input (-) of the DMM and the EO Neutral to the Input (+) as shown in Figure 3-6. Set the DMM for 1000 V ac mode. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
Fluke Biomedical ESA612
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
Figure 3-6. Insulation Resistance Verification Connections
2. Send command: HIGH_RES=ON
FRONT
gjc109.eps
3. Receive * 4. Send command: STD=AAMI
5. Receive * 6. Send command: MINS
7. Receive * 8. Send command: INS=LOW
9. Receive * 10. Send command: MREAD
11. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 12. Press the ESC key. The Analyzer will stop reporting readings. 13. Send command: INS=HIGH
14. Receive * 15. Send command: MREAD 3-16
Verification and Calibration Analyzer Verification
3
16. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 17. Press the ESC key. The UUT will stop reporting readings. 18. Connect the EO ground to one side of a 110 kΩ resistor. Connect the other side to the Input (-) of the DMM. Connect the 400 mA jack of the DMM to the EO Neutral. See Figure 3-7. Set the DMM to 10 mA ac mode. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
Fluke Biomedical ESA612
HI
HI
1000 V CAT I 600V CAT II
110 kΩ Resistor
300V
LO
LO 1V FUSED
100 mA
10A REAR
Figure 3-7. Insulation Resistance with 110 kΩ Resistor Connections
19. Send command: MREAD
FRONT
gfi110.eps
20. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Remove all testing connections to the Analyzer.
3-17
ESA612 Service Manual
Mains to PE 23. Connect the EO Neutral and the EO ground across a 10.0-MΩ resistor as shown in Figure 3-8.
10 MΩ Resistor
Fluke Biomedical ESA612
Figure 3-8. Mains to PE Verification Connections
24. Send command: HIGH_RES=ON
25. Receive * 26. Send command: STD=AAMI
27. Receive * 28. Send command: MINS
29. Receive * 30. Send command: INS=HIGH
31. Receive * 32. Send command: MREAD
33. Following the tenth reading, press the ESC key. Record the tenth reading.
3-18
gfi111.eps
Verification and Calibration Analyzer Verification
3
AP to PE 34. Disconnect the lead from the EO Neutral and connect it to the RA Applied Part. The other lead should remain connected to the EO ground. Both leads should still be attached to the 10-MΩ resistor. See Figure 3-9.
Fluke Biomedical ESA612
RA
10 MΩ Resistor
Figure 3-9. AP to PE Verification Connections
35. Send command: HIGH_RES=ON
gjc135.eps
36. Receive * 37. Send command: STD=AAMI
38. Receive * 39. Send command: APINS
40. Receive * 41. Send command: INS=HIGH
42. Receive * 43. Send command: MREAD
44. Following the tenth reading, press the ESC key. Record the tenth reading.
3-19
ESA612 Service Manual
AP to NE 45. With one side of the 10 MΩ resistor still connected to RA, connect the other side to the red jack as shown in Figure 3-10.
10 MΩ Resistor
Fluke Biomedical ESA612
RA
Figure 3-10. AP to NE Verification Connections
46. Send command: HIGH_RES=ON
47. Receive * 48. Send command: STD=AAMI
49. Receive * 50. Send command: INSD
51. Receive * 52. Send command: INS=HIGH
53. Receive * 54. Send command: MREAD
55. Following the tenth reading, press the ESC key. Record the tenth reading.
3-20
gjc136.eps
Verification and Calibration Analyzer Verification
3
Mains to NE 56. Connect the red jack to one side of a 10-MΩ resistor. Connect EO Hot to the other side of the resistor as shown in Figure 3-11.
10 MΩ Resistor
Fluke Biomedical ESA612
Figure 3-11. Mains to NE Verification Connections
57. Send command: HIGH_RES=ON
gjc114.eps
58. Receive * 59. Send command: STD=AAMI
60. Receive * 61. Send command: INSB
62. Receive * 63. Send command: INS=HIGH
64. Receive * 65. Send command: MREAD
66. Following the tenth reading, press the ESC key. Record the tenth reading.
3-21
ESA612 Service Manual
Mains to AP 67. With the EO Hot still connected to one side of a 10 MΩ resistor, connect the other side of the resistor to the RA Applied Part as shown in Figure 3-12.
10 MΩ Resistor
Fluke Biomedical ESA612
RA
Figure 3-12. Mains to AP Verification Connections
68. Send command: HIGH_RES=ON
gjc115.eps
69. Receive * 70. Send command: STD=AAMI
71. Receive * 72. Send command: INSE
73. Receive * 74. Send command: INS=HIGH
75. Receive * 76. Send command: MREAD
77. Following the tenth reading, press the ESC key. Record the tenth reading. 78. Repeat steps 76-77 for each of the other Applied Parts: LL, LA, RL, and V1. 79. Change the 10 MΩ resistor for a 0.7 M resistor, while still connected to V1 and EO hot. 80. Send command: INS=LOW
81. Receive * 82. Send command: MREAD
83. Following the tenth reading, press the ESC key. Record the tenth reading. 84. Send command: INS=HIGH 85. Receive * 3-22
Verification and Calibration Analyzer Verification
86. Send command: MREAD
3
87. Following the tenth reading, press the ESC key. Record the tenth reading. 88. Repeat steps 80-87 for each resistance value of 1.0 MΩ, 6.5 MΩ, 18 MΩ, 22 MΩ, 60 MΩ, and 100 MΩ. 89. Disconnect all testing connections from the Analyzer. I.
DC Leakage 1. Connect the red and black jacks to the Input (+) and Input (-) jacks of the DMM and set the DMM for 1-kΩ range as shown in Figure 3-13.
Fluke 8846A INPUT
Fluke Biomedical ESA612
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Figure 3-13. DC Leakage Verification Connections
2. Send command: HIGH_RES=ON
gjc137.eps
3. Receive * 4. Send command: STD=AAMI
5. Receive * 6. Send command: PPL
7. Receive * 8. Send command: MODE=ACDC
9. Receive * 10. Read and record the load value from the DMM. 11. Remove all testing connections from the Analyzer.
3-23
ESA612 Service Manual
12. Connect the current source to the red and black jacks as shown in Figure 3-14. Set the current source for 10 μA dc. Keithly 6221
Fluke Biomedical ESA612
Figure 3-14. DC Leakage Current Verification Connection
13. Send command: HIGH_RES=ON
14. Receive * 15. Send command: STD=AAMI
16. Receive * 17. Send command: PPL
18. Receive * 19. Send command: MODE=DC
20. Receive * 21. Turn on the current source output. 22. Send command: MREAD
23. Receive multiple readings. Record the tenth reading. 24. Press the ESC key. The UUT will stop reporting readings. 25. Adjust the current source for 50 μA dc. 26. Send command: MREAD
27. Receive multiple readings. Record the tenth reading. 28. Press the ESC key. The UUT will stop reporting readings. 3-24
gjc117.eps
Verification and Calibration Analyzer Verification
3
29. Adjust the current source for 100 μA dc. 30. Send command: MREAD
31. Receive multiple readings. Record the tenth reading. 32. Press the ESC key. The UUT will stop reporting readings. 33. Adjust the current source for 160 μA dc. 34. Send command: MREAD
35. Receive multiple readings. Record the tenth reading. 36. Press the ESC key. The UUT will stop reporting readings. 37. Adjust the current source for 0.340 mA dc. 38. Send command: MREAD
39. Receive multiple readings. Record the tenth reading. 40. Press the ESC key. The UUT will stop reporting readings. 41. Adjust the current source for 0.500 mA dc. 42. Send command: MREAD
43. Receive multiple readings. Record the tenth reading. 44. Press the ESC key. The UUT will stop reporting readings. 45. Adjust the current source for 1.000 mA dc. 46. Send command: MREAD
47. Receive multiple readings. Record the tenth reading. 48. Press the ESC key. The UUT will stop reporting readings. 49. Adjust the current source for 1.600 mA dc. 50. Send command: MREAD
51. Receive multiple readings. Record the tenth reading. 52. Press the ESC key. The UUT will stop reporting readings. 53. Adjust the current source for 3.400 mA dc. 54. Send command: MREAD
55. Receive multiple readings. Record the tenth reading. 56. Press the ESC key. The UUT will stop reporting readings. 57. Adjust the current source for 5.000 mA dc. 58. Send command: MREAD
59. Receive multiple readings. Record the tenth reading. 60. Press the ESC key. The UUT will stop reporting readings. 61. Adjust the current source for 7.000 mA dc. 62. Send command: MREAD
63. Receive multiple readings. Record the tenth reading. 3-25
ESA612 Service Manual
64. Press the ESC key. The UUT will stop reporting readings. 65. Turn off the current source output and disconnect all testing connections from the Analyzer. J. AC Leakage Filter Frequency Response 1. Connect the current source to the red jack and the DMM 400 mA current input jack. Connect the DMM Input (-) to the black jack. See Figure 3-15. Set the current source to 1.000 mA (1.4142 mA p-p), 60 Hz, with a 25 V compliance voltage, and the DMM to the 1 mA ac range. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
Figure 3-15. AC Leakage Filter Frequency Response Connections
2. Send command: PPL
gjc126.eps
3. Receive * 4. Send command: LOAD=AAMI
5. Receive * 6. Turn on the current source output. 7. Send command: READ
8. Receive and record response of the Analyzer as the Analyzer current. Record the DMM current as the measured current.
3-26
Verification and Calibration Analyzer Verification
3
9. Calculate the measured transfer ratio: divide the measured current by the Analyzer current. Record this transfer ratio, and compare to the design transfer ratio. TRmeas =
I meas Analyzer Current
For this frequency, the design transfer ratio is 0.9980 10. Turn off the current source. 11. Calculate the transfer ratio limits: multiply the design transfer ratio by the tolerance. Add this value to the result of the offset current divided by the measured current. Record this calculated transfer ratio limit. Add and subtract this limit to the design transfer ratio to obtain the maximum and minimum, respectively. Offset TRLimit = TRdesign ∗ %TOL + I meas
(
)
For this frequency, the tolerance is 0.5 % and the offset is 0.005 mA. 12. Verify that: TRdesign – TRLimit ≤ TRmeas ≤ TRdesign + TRlimit 13. Set the current source to 1.000 mA, 1,000 Hz, and turn on the output. 14. Send command: READ
15. Receive and record response. Turn off the current source output. 16. Perform the same calculations that were completed in steps 9 through 12 above. The design transfer ratio for this frequency is 0.6910, the tolerance is 2.0 %, and the Offset current is 0.005 mA. 17. Set the current source to 1.000 mA, 10,000 Hz and turn on the output. 18. Send command: READ 19. Receive and record response. Turn off the current source. 20. Perform the same calculations that were completed in steps 9 through 12 above. The design transfer ratio for this frequency is 0.09558, the tolerance is 2.0 %, and the Offset current is 0.005 mA. 21. Turn off the current source. Disconnect all test connections.
3-27
ESA612 Service Manual
K. Differential Leakage 1. Connect the current source to the EO Neutral and the DMM 400 mA current input jack. Connect the DMM Input (-) to the EO ground. See Figure 3-16. Set the current source to 76.00 μA (107.48 μA p-p), 100 Hz, with a 25 V compliance voltage, and the DMM to the 100 μA ac range. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
Figure 3-16. Differential Leakage Verification Connections
2. Send command: DIFF
3. Receive * 4. Send command: EARTH=C
5. Receive * 6. Send command: POL=N
7. Receive * 8. Send command: NOMINAL=OFF
9. Receive * 10. Turn on the current source output. 11. Send command: READ 3-28
gjc118.eps
Verification and Calibration Analyzer Verification
3
12. Receive and record response; compare it to the actual current received by the DMM. 13. Adjust the DMM to the 1 mA range and repeat steps 11 and 12 with the current source set to 0.16mA (0.2263 mA p-p), 0.24 mA (0.3394 mA p-p), 0.50 mA (0.7071 mA p-p), 0.76 mA (1.0748 mA p-p), 100 Hz. 14. Adjust the DMM to the 10 mA range and repeat steps 11 and 12 with the current source set to 1.6 mA (2.263 mA p-p), 2.4 mA (3.394 mA p-p), 5 mA (7.071 mA p-p), and 7.6 mA (10.748 mA p-p), 100 Hz. 15. Adjust the DMM to the 100 mA range and repeat steps 11 and 12 with the current source set to 16 mA (22.63 mA p-p), 100 Hz. 16. Turn off the current source output and disconnect all testing connections from the Analyzer. L. ECG Leakage Functionality 1. Connect the DMM Input (+) to the RA applied part. Connect the DMM Input (-) to the EO ground. See Figure 3-17. Set the DMM for 1 KΩ resistance range. During this test, the DMM will function as both a current source and an ohmmeter. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
RA
Figure 3-17. ECG Leakage Functionality Connections
2. Send command: HIGH_RES=ON
gjc139.eps
3. Receive * 4. Send command: STD=601
3-29
ESA612 Service Manual
5. Receive * 6. Send command: PAT
7. Receive * 8. Send command: NOMINAL=OFF
9. Receive * 10. Send command: MODE=DC
11. Receive * 12. Send command: AP=RA//GND
13. Receive * 14. Send command: READ
15. Receive response. The DMM is acting as a current source, and should be supplying approximately 100 μA. If the Analyzer responds accordingly, record PASS. If it responds with 0 μA, record FAIL. 16. Send command: AP=LL//OPEN
17. Receive * 18. Note the reading from the DMM. If it reads OL (open), this indicates that the RA applied part is connected to the Open Bus. Record PASS. If the DMM indicates a significant resistance or a short (0 Ω), record FAIL. 19. Send command: AP=LL//GND
20. Receive * 21. Note the reading from the DMM. If it reads 0 Ω, this indicates that the RA applied part is connected to ground. Record PASS. If the DMM indicates a significant resistance or OL (open), record FAIL. 22. Move the lead connected to Input (+) from RA to LL. Repeat steps 14 – 21 for each applied part (LL, LA, RL, and V1). 23. Remove all test connections from the DMM.
3-30
Verification and Calibration Analyzer Verification
3
M. Direct Applied Part Leakage, Alternative Applied Part Leakage: Ground to RA 1. Connect the RA Applied Part to one side of a 110 kΩ resistor. Connect the 400 mA jack of the DMM to the other side of the 110 kΩ resistor. Connect the DMM Input (-) to the EO ground. See Figure 3-18. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
Fluke Biomedical ESA612
FRONT
110 kΩ Resistor
RA
Figure 3-18. Direct Applied Part Verification Ground to RA Connections
2. Send command: NOMINAL=OFF
gjc121.eps
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: DMAP
9. Receive * 10. Send command: MAP=3.5MA
11. Receive * 12. Send command: POL=N
13. Receive * 14. Send command: AP=RA/RL/GND
15. Receive * 3-31
ESA612 Service Manual
16. Send command: MODE=AC
17. Receive * 18. Send command: MREAD
19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF
23. Receive * 24. Send command: HIGH_RES=ON
25. Receive * 26. Send command: STD=353
27. Receive * 28. Send command: SPAT
29. Receive * 30. Send command: MAP=3.5MA
31. Receive * 32. Send command: AP=ALL//
33. Receive * 34. Send command: MODE=AC
35. Receive * 36. Send command: MREAD
37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE 41. Receive *
3-32
Verification and Calibration Analyzer Verification
3
N. Direct Applied Part Leakage, Alternative Applied Part Leakage: Red to RA 1. Move the lead connected to the EO ground to the red jack; the other end should be connected to the DMM Input (-). The RA Applied Part should still be connected to one end of the 110-kΩ resistor. See Figure 3-19. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612 110 kΩ Resistor
RA
Figure 3-19. Direct Applied Part Verification Red to RA Connections
2. Send command: NOMINAL=OFF
gjc125.eps
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: DMAP
9. Receive * 10. Send command: MAP=3.5MA
11. Receive * 12. Send command: POL=N
13. Receive * 14. Send command: AP=RA/RL/GND
15. Receive * 16. Send command: MODE=AC
3-33
ESA612 Service Manual
17. Receive * 18. Send command: MREAD
19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF
23. Receive * 24. Send command: HIGH_RES=ON
25. Receive * 26. Send command: STD=353
27. Receive * 28. Send command: SPAT
29. Receive * 30. Send command: MAP=3.5MA
31. Receive * 32. Send command: AP=ALL//
33. Receive * 34. Send command: MODE=AC
35. Receive * 36. Send command: MREAD
37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE 41. Receive *
3-34
Verification and Calibration Analyzer Verification
3
O. Alternative Equipment Leakage: Red to Hot 1. Move the lead from the RA applied part to the EO hot; the other end should still be connected to the 110-kΩ resistor. The red jack should be connected to the DMM Input (-). See Figure 3-20. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO
Fluke Biomedical ESA612
1V FUSED
100 mA
10A REAR
FRONT
110 kΩ Resistor
Figure 3-20. Alternate Equipment Leakage: Red to Hot Connections
2. Send command: NOMINAL=OFF
gjc142.eps
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: SAF
9. Receive * 10. Send command: MAP=3.5MA
11. Receive * 12. Send command: AP=ALL//
13. Receive * 14. Send command: MODE=AC
15. Receive * 16. Send command: ALTEARTH=C
17. Receive * 18. Send command: MREAD
3-35
ESA612 Service Manual
19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: IDLE
23. Receive * P. Alternative Equipment Leakage, Alternative Applied Part Leakage: RA to Hot 1. Move the lead from the Red jack to the RA applied part, the other end should still be connected to the Input (-) of the DMM. The EO hot should still be connected to the 110-kΩ resistor. See Figure 3-21. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
Fluke Biomedical ESA612
FRONT
110 kΩ Resistor
RA
Figure 3-21. Alternative Applied Part Leakage: RA to Hot
2. Send command: NOMINAL=OFF
3. Receive * 4. Send command: HIGH_RES=ON 5. Receive * 6. Send command: STD=353 7. Receive * 8. Send command: SAF 3-36
gjc123.eps
Verification and Calibration Analyzer Verification
3
9. Receive * 10. Send command: MAP=3.5MA
11. Receive * 12. Send command: MODE=AC
13. Receive * 14. Send command: AP=ALL//
15. Receive * 16. Send command: ALTEARTH=C
17. Receive * 18. Send command: MREAD
19. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 20. Receive and record response; compare it to the actual current received by the DMM. 21. Press the ESC key. The Analyzer will stop reporting readings. 22. Send command: NOMINAL=OFF
23. Receive * 24. Send command: HIGH_RES=ON
25. Receive * 26. Send command: STD=353
27. Receive * 28. Send command: SPAT
29. Receive * 30. Send command: MAP=3.5MA
31. Receive * 32. Send command: AP=ALL//
33. Receive * 34. Send command: MODE=AC
35. Receive * 36. Send command: MREAD
37. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 38. Receive and record response; compare it to the actual current received by the DMM. 39. Press the ESC key. The Analyzer will stop reporting readings. 40. Send command: IDLE
41. Receive * 3-37
ESA612 Service Manual
Q. Alternative Applied Part Leakage: RA to Neutral 1. Move the lead from the EO hot to the EO neutral, the other end should still be connected to the 110-kΩ resistor. The RA applied part lead should still be connected to the DMM Input (-). See Figure 3-22. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
Fluke Biomedical ESA612
FRONT
110 kΩ Resistor
RA
Figure 3-22. Alternative Applied Part Leakage: RA to Neutral Connections
2. Send command: NOMINAL=OFF
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: SPAT
9. Receive * 10. Send command: AP=ALL//
11. Receive * 12. Send command: MAP=3.5MA
13. Receive * 14. Send command: MODE=AC 15. Receive * 3-38
gjc132.eps
Verification and Calibration Analyzer Verification
16. Send command: MREAD
3
17. Receive multiple readings, all approximately 1 mA ac. While the Analyzer is conducting its read, record the value of the DMM current. 18. Receive and record response; compare it to the actual current received by the DMM. 19. Press the ESC key. The Analyzer will stop reporting readings. 20. Send command: IDLE
21. Receive * 22. Disconnect all test leads from the 100 kΩ resistor. R. Patient Auxiliary Leakage: RA to RL 1. Connect the current source to the 400 mA jack of the DMM and to the RL applied part. The DMM Input (-) should still be connected to the RA applied part. See Figure 3-23. Set the DMM to the 100 μA ac range and the current source to 100.000 μA (141.42 μA p-p), 100 Hz, with a 25 V compliance voltage. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
RA
RL
Figure 3-23. Patient Auxiliary Leakage: RA to RL Connections
2. Send command: NOMINAL=OFF
gjc127.eps
3. Receive * 3-39
ESA612 Service Manual
4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=601
7. Receive * 8. Send command: AUX
9. Receive * 10. Send command: AP=RA/RL/GND
11. Receive * 12. Send command: POL=N
13. Receive * 14. Send command: MODE=AC
15. Receive * 16. Send command: NEUT=C
17. Receive * 18. Send command: EARTH=C
19. Receive * 20. Turn on the current source output. 21. Send command: READ
22. Receive and record response; compare it to the actual current received by the DMM. 23. Turn off the current source output. 24. Send command: IDLE 25. Receive *
3-40
Verification and Calibration Analyzer Verification
3
S. Direct Equipment Leakage, Patient Leakage: Ground to RA 1. Move the lead from the RL applied part to the EO ground, the other end should still be connected to the current source. The RA applied part lead should still be connected to the DMM Input (-). See Figure 3-24. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
RA
Figure 3-24. Direct Equipment Leakage, Patient Leakage: Ground to RA
2. Send command: NOMINAL=OFF
gjc131.eps
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: DIRL
9. Receive * 10. Send command: AP=RA/RL/GND
11. Receive * 12. Send command: POL=N
13. Receive * 3-41
ESA612 Service Manual
14. Send command: MODE=AC
15. Receive * 16. Send command: EARTH=C
17. Receive * 18. Turn on the current source output. 19. Send command: READ
20. Receive and record response; compare it to the actual current received by the DMM. 21. Turn off the current source output. 22. Send command: IDLE
23. Receive * 24. Send command: NOMINAL=OFF
25. Receive * 26. Send command: HIGH_RES=ON
27. Receive * 28. Send command: STD=601
29. Receive * 30. Send command: PAT
31. Receive * 32. Send command: AP=RA/RL/GND
33. Receive * 34. Send command: POL=N
35. Receive * 36. Send command: MODE=AC
37. Receive * 38. Send command: NEUT=C
39. Receive * 40. Send command: EARTH=C
41. Receive * 42. Turn on the current source output. 43. Send command: READ
44. Receive and record response; compare it to the actual current received by the DMM. 45. Turn off the current source output. 46. Send command: IDLE 47. Receive * 3-42
Verification and Calibration Analyzer Verification
3
T. Direct Equipment Leakage, Enclosure Leakage: Ground to Red 1. Move the lead from the RA applied part to the Red jack, the other end should still be connected to the DMM Input (-). The EO ground lead should still be connected to the current source. See Figure 3-25. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
Figure 3-25. Direct Equipment Leakage, Enclosure Leakage: Ground to Red Connections
2. Send command: NOMINAL=OFF
gjc129.eps
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=353
7. Receive * 8. Send command: DIRL
9. Receive * 10. Send command: AP=RA/RL/GND
11. Receive * 12. Send command: POL=N
13. Receive * 3-43
ESA612 Service Manual
14. Send command: MODE=AC
15. Receive * 16. Send command: EARTH=C
17. Receive * 18. Turn on the current source output. 19. Send command: READ
20. Receive and record response; compare it to the actual current received by the DMM. 21. Turn off the current source output. 22. Send command: IDLE
23. Receive * 24. Send command: NOMINAL=OFF
25. Receive * 26. Send command: HIGH_RES=ON
27. Receive * 28. Send command: STD=601
29. Receive * 30. Send command: ENCL
31. Receive * 32. Send command: POL=N
33. Receive * 34. Send command: MODE=ACDC
35. Receive * 36. Send command: NEUT=C
37. Receive * 38. Send command: EARTH=C
39. Receive * 40. Turn on the current source output. 41. Send command: READ
42. Receive and record response; compare it to the actual current received by the DMM. 43. Turn off the current source output. 44. Send command: IDLE
45. Receive * 46. Do not disconnect any connections.
3-44
Verification and Calibration Analyzer Verification
3
U. Filter Tests 1. Attempt to measure ac current with the dc filter: set the current source to generate 100.00 μA (141.42 μA p-p), 100 Hz, with a 25 V compliance voltage. 2. Send command: NOMINAL=OFF
3. Receive * 4. Send command: HIGH_RES=ON
5. Receive * 6. Send command: STD=601
7. Receive * 8. Send command: ENCL
9. Receive * 10. Send command: POL=N
11. Receive * 12. Send command: MODE=DC
13. Receive * 14. Send command: NEUT=C
15. Receive * 16. Send command: EARTH=C
17. Receive * 18. Turn on the current source output. 19. Send command: READ
20. Receive and record response. The result should be approximately zero. 21. Turn off the current source output. 22. Set the current source to generate 100 μA dc and turn the output on. 23. Send command: MODE=AC
24. Receive * 25. Send command: READ
26. Disconnect all testing connections from the Analyzer.
3-45
ESA612 Service Manual
V. MAP Voltage and Current Limits 1. Connect the DMM Input (+) to the RA applied part and the DMM Input (-) to the EO ground as shown in Figure 3-26. Set the DMM to the 1000 V ac range. Fluke 8846A INPUT
SENSE 4W
V
Fluke Biomedical ESA612
2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
RA
Figure 3-26. MAP Voltage and Current Limits Connections
2. Send command: DMAP
gjc140.eps
3. Receive * 4. Send command: MAP=NORM
5. Receive * 6. Send command: AP=RA/RL/GND
7. Receive * 8. Send command: MREAD
9. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 10. Press the ESC key. The Analyzer will stop reporting readings. 11. Turn off the output of the programmable ac power supply. Adjust it for 230 V, 50 Hz, and turn on the output. 12. Once the ESA612 powers on again, it will report an error code of “!53” and the display will show “Fault detected…Mains out of range,” Send command: REMOTE=F2810
13. Receive *. The WARNING LED will remain lit. 14. Send command: DMAP
15. Receive * 16. Send command: MAP=3.5MA
17. Receive * 18. Send command: AP=RA/RL/GND 3-46
Verification and Calibration Analyzer Verification
3
19. Receive * 20. Send command: MREAD
21. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM voltage. 22. Press the ESC key. The Analyzer will stop reporting readings. 23. Move the lead connected to the RA applied part from the DMM Input (+) to the 400 mA jack. The DMM Input (-) should still be connected to EO ground. Set the DMM to the 10 mA range. 24. Send command: DMAP
25. Receive * 26. Send command: MAP=3.5MA
27. Receive * 28. Send command: AP=RA/RL/GND
29. Receive * 30. Send command: MREAD
31. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 32. Press the ESC key. The Analyzer will stop reporting readings. 33. Send command: DMAP
34. Receive * 35. Send command: MAP=7.5MA
36. Receive * 37. Send command: AP=RA/RL/GND
38. Receive * 39. Send command: MREAD
40. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 41. Press the ESC key. The Analyzer will stop reporting readings. 42. Turn off the output of the programmable ac power supply. Readjust it for 115 V, 60 Hz, and turn on the output. 43. Once the ESA612 powers on again, it will report the model number and firmware version. Send command: REMOTE 44. Receive * 45. Send command: DMAP
46. Receive * 47. Send command: MAP=1MA
48. Receive * 49. Send command: AP=RA/RL/GND
3-47
ESA612 Service Manual
50. Receive * 51. Send command: MREAD
52. Receive multiple readings. While the Analyzer is conducting its read, record the value of the DMM current. 53. Press the ESC key. The Analyzer will stop reporting readings. .
54. Disconnect all testing connections from the Analyzer. W. ECG Waveform 1. Connect the Input (+) of the differential amplifier to the RA applied part and the Input (-) of the differential amplifier to the RL applied part. Connect the differential amplifier output to the oscilloscope. See Figure 3-27. Oscilliscope
Differential Amplifier Chan 1
Out
Chan 2
Fluke Biomedical ESA612
RA RL
Figure 3-27. ECG Waveform Connections
2. Establish the following settings for the differential amplifier: a. GAIN = 1K b. Divide by 100 = Out c. (+) Input = DC d. (-) Input = DC 3-48
gjc128.eps
Verification and Calibration Analyzer Verification
3
e. HF -3dB = 0.1 f. LF -3db = DC OFFSET 3. Send command: STD=AAMI
4. Receive * 5. Send command: ECG
6. Receive * 7. Send command: SQ2
8. Receive * 9. Verify the waveform has a 2.000 ±0.04 Hz frequency. Record PASS/FAIL. 10. Connect the RA applied part to the DMM Input (+) and the RL applied part to the DMM Input (-) and set the DMM for 100 mV dc range. 11. Using DMM MIN MAX function, record the waveform peak to peak amplitude. 12. Move the DMM Input (+) from the RA applied part to the LL applied part. Repeat step 11 for each of the following applied parts: LL, LA, and V1. 13. Press ESC. Receive only a carriage return. 14. Send commend: EXIT
15. Receive * 16. Disconnect all testing connections.
3-49
ESA612 Service Manual
X. Equipment Current Verification 1. Connect the current shunt to the DMM using the voltage inputs. Connect the EO hot to the DMM Input (+) side of the current shunt and the EO ground to one side of the variable current load. Connect the other side of the variable current load to the current shunt jack corresponding to the Input (-) of the DMM. Set the DMM to the 10 V dc range.
Fluke Biomedical ESA612
Fluke A40A Current Shunt
Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Variable Load
Figure 3-28. Equipment Current Calibration Connection
2. Send command: STD=AAMI
gjc146.eps
3. Receive * 4. Send command: EQCURR
5. Receive * 6. Establish an 8 A draw through the current load, with the shunt resistor in series with the load. 7. Send command: READ
8. Receive * 9. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 10. Establish a 3 A draw through the current load. 11. Send command: READ
12. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 13. Establish a 1 A draw through the current load. 14. Send command: READ 3-50
Verification and Calibration Analyzer Verification
3
15. Receive and record response. Compare to the actual current flowing through the load as read on the DMM. 16. Turn off the equipment outlet. Send command: PPL
17. Receive * 18. Remove all testing connections. Y. Save Calibration Information If this is AS LEFT verification and the device has passed verification, record the new calibration date and technician information. 1. Send command: CAL=F2810
2. Receive * 3. Write new calibration date in the format of MxxDyyYzzzz, where xx stands for the two digits of the current month, yy stands for the two digits of the current day, and zzzz stands for the four digits of the current year. 4. Send command: DATE=MxxDxxYxxxx
5. Receive * 6. Write calibration technician in the format of xxxxx, where xxxxx stands for the five digits of the technician’s employee number. 7. Send command: TECH=xxxxx
8. Receive * 9. Check Analyzer serial number, cal date, and cal technician are correct. 10. Send command: INFO=
11. Receive serial number, cal date, and cal technician in response. Record response. 12. Send command: SAVE
13. Send command: EXIT
14. Receive * 15. If the unit did not come in with a nominal voltage of 115, readjust the nominal voltage according to the destination country. a. Send command: NOMINAL=xxx where xxx is the value recorded when the unit was originally converted for verification. See Table 3-3 for standard nominal values.
3-51
ESA612 Service Manual
Table 3-3. Nominal Values Country
Value
Europe
230
UK
230
Australia
240
Israel
230
France/Belgium
230
US
120
Swiss
230
Thailand
220
Japan
100
b. Receive * 16. Turn the Analyzer power off. 17. After the Analyzer has restarted, verify the device has retained its calibration date by pressing SETUP. Next press the softkey labeled More twice, and then the softkey labeled Instrument Information. Verify the calibration date, firmware version, and serial number are correct. 18. Turn the Analyzer off and remove all connections.
Calibration Procedure The following steps will bring the Analyzer into published specifications. The required equipment for this procedure is shown in Table 3-1. Setup Perform the following steps to prepare the Analyzer for calibration. 1. 2. 3. 4.
Power on all test equipment and allow a 1-hour warm up time. Power on the PC and log in. Connect the PC USB cable to the UUT. Power on the UUT. Establish serial communication with the Analyzer (Applicable COM port using USBView, 115200 Baud, N,8,1 bits, HW Flow control) Under File->Properties and the Settings tab, click the ASCII Setup button, and ensure “Send line ends with line feeds” and “Echo typed characters locally” are CHECKED. 5. Many of the calibrations require ac or dc meter calibration to be completed first. If the ac or dc meter calibration is changed, all other calibrations should at least be verified.
Adjustments The following procedure calibrates the Analyzer to the published specifications and coincides with the calibration datasheet found in Appendix A: A. AC Meter Calibration 1. Send command: REMOTE
2. Receive * 3. Send command: CAL=F2810 3-52
Verification and Calibration Calibration Procedure
3
4. Receive * 5. Send command: IDENT
6. Receive and record UI & MTR firmware revision level. 7. Send command: EXIT
8. Receive * 9. Send command: NOMINAL?
10. Receive and record the current nominal voltage. 11. If nominal is not set to 115 V, reset nominal and restart the Analyzer. a. Send command: NOMINAL=115
b. Receive * c. Send command: RSTUI
d. When the Analyzer restarts, receive the identification information. e. Send command: REMOTE
f. Receive * 12. Send command: DIAG=F2810
13. Receive * 14. Send command: XFLASHT
15. Receive and record PASS/FAIL 16. Send command: EXIT
17. Receive * 18. Send command: CAL=F2810
19. Receive * 20. Send command: DEFAULTS
21. Receive * 22. Connect the calibrator to the red and black jacks as shown in Figure 3-29. Set the calibrator for 2.000 V, 120 Hz (±200 mV).
3-53
ESA612 Service Manual
Fluke 5520A 5520A CALIBRATOR
Fluke Biomedical ESA612 NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
SCOPE OUT
HI
LO
TRIG
20A
GUARD
20V PK MAX
TC
20V PK MAX
Figure 3-29. AC Meter Calibration Connections
23. Send command: METER_AC
24. Receive * 25. Send command: RANGE=1,0,0
26. Receive * 27. Turn on the calibrator output. 28. Send command: GAIN=2000.00
29. Receive * 30. Adjust the calibrator output to 1.000 V, 120 Hz (±100 mV). 31. Send command: GAIN=1000.00
32. Receive * 33. Adjust the calibrator output to 200 mV, 120 Hz (±20 mV). 34. Send command: RANGE=2,0,0
35. Receive * 36. Send command: GAIN=200.00
37. Receive * 38. Adjust the calibrator output to 20.00 mV, 120 Hz (±2 mV). 39. Send command: RANGE=3,0,0
40. Receive * 41. Send command: GAIN=200.00 42. Receive * 3-54
gjc105.eps
Verification and Calibration Calibration Procedure
3
43. Short the Red and Black jacks together using a test lead. 44. Send command: OFFSET
45. After a significant delay, receive * 46. Reconnect the normal outputs of the calibrator to the red and black jacks. Adjust the calibrator output to 1.500 V, 120 Hz. 47. Send command: RANGE=1,0,0
48. Receive * 49. Send command: READ
50. Receive and record response. 51. Adjust the calibrator output to 500.0 mV, 120 Hz. 52. Send command: READ
53. Receive and record response. 54. Adjust the calibrator output to 150.0 mV, 120 Hz. 55. Send command: RANGE=2,0,0
56. Receive * 57. Send command: READ
58. Receive and record response. 59. Adjust the calibrator output to 15.0 mV, 120 Hz. 60. Send command: RANGE=3,0,0
61. Receive * 62. Send command: READ
63. Receive and record response. 64. Disconnect the calibrator. B. DC Meter Calibration 1. Short the red and black jacks together using a test lead. 2. Send command: METER_DC
3. Receive * 4. Send command: OFFSET
5. Receive * 6. Connect the calibrator the red and black jacks, and adjust the calibrator output for 2000 mV dc. 7. Send command: GAIN=2000.00
8. Receive * 9. Adjust the calibrator output to 1500 mV dc. 10. Send command: READ
11. Receive and record response. 12. Disconnect the calibrator. 3-55
ESA612 Service Manual
C. Leakage Calibration (Dependent upon AC Meter) 1. Connect the current source to the red jack and the 400 mA jack of the DMM. Connect the DMM Input (-) to the black jack. See Figure 3-30. Set the current source to 1.00 mA (1.4142 mA p-p), 60 Hz with a 10 V compliance voltage and the DMM to 1 mA ac mode. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
Figure 3-30. Leakage Calibration
2. Send command: PPL
gjc126.eps
3. Receive * 4. Send command: LOAD=601
5. Receive * 6. Send command: RANGE=0,0,2
7. Receive * 8. Turn on the current source. 9. Send command: GAIN=X.XXXX where x.xxxx is the actual value of the current in mA measured by the DMM. 10. Receive * 11. Adjust the current source to 0.5 mA (0.7071 mA p-p). 12. Send command: READ
13. Receive and record response. Also record the DMM reading. 3-56
Verification and Calibration Calibration Procedure
3
14. Turn off the current source output. Adjust the current source to 5.00 mA (7.071 mA p-p), 60 Hz. Set the DMM to the 10 mA range. 15. Send command: RANGE=0,0,1
16. Receive * 17. Turn on the current source output. 18. Send command: GAIN=X.XXXX where X.XXXX is the actual value of the current in mA measured by the DMM. 19. Receive * 20. Adjust the current source to 3.0 mA (4.243 mA p-p). 21. Send command: READ
22. Receive and record response. Also record the DMM reading. 23. Turn off the current source and disconnect all testing connections. D. Point to Point Calibration (Dependent on AC Meter) 1. Connect the calibrator to the red jack and black jacks as shown in Figure 3-31. Set the calibrator to 300.0 V, 120 Hz. Fluke 5520A 5520A CALIBRATOR
Fluke Biomedical ESA612 NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
SCOPE OUT
HI
LO
TRIG
20A
GUARD
20V PK MAX
TC
20V PK MAX
Figure 3-31. Point to Point Calibration Connections
2. Send command: PPV
gjc105.eps
3. Receive * 4. Turn on the calibrator output. 5. Send command: GAIN=300.00
6. Receive * 7. Adjust the calibrator output to 250.0 V, 120 Hz. 3-57
ESA612 Service Manual
8. Send command: READ
9. Receive and record response. 10. Turn off the calibrator output and disconnect the calibrator. E. MAP Compensation Calibration (Dependent on AC Meter) 1. Remove all connections except mains power and the USB cable. 2. Send command: COMP 3. After significant delay, receive * if the null was found. 4. If no null was found, !33 will be returned, and the device needs repair prior to calibration. F. Insulation HV Calibration (Dependent on AC Meter) 1. Connect the DMM to the RA applied part and the red jack as shown in Figure 332 and set it for the 1000 V dc range. Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
Fluke Biomedical ESA612
100 mA
10A REAR
FRONT
Figure 3-32. Insulation HV Calibration Connections
2. Send command: INSD
3. Receive * 4. Send command INS=HIGH
5. Receive * 6. Send command: MREAD
7. Receive multiple readings, all approximately 10 MΩ. While the Analyzer is conducting is read, note the stabilized voltage reading from the DMM. 8. Press ESC to stop MREAD. 9. Send command: GAIN=XXX.XXX where XXX.XXX is the actual value of the voltage measured by the DMM. 10. Receive * 11. Connect the applied part and the red jack across a 10 MΩ resistor. 3-58
gjc144.eps
Verification and Calibration Calibration Procedure
12. Send command: READ
3
13. Receive and record response. 14. Disconnect all testing connections. G. Resistance Calibration (Dependent on AC Meter) 1. Connect two leads to the red and black jacks, then short the other ends together as shown in Figure 3-33. Stackable Banana Plugs
Fluke Biomedical ESA612
Figure 3-33. Resistance Calibration Connections
2. Send command: PPR
gjc108.eps
3. Receive * 4. Send command: OFFSET
5. Receive * 6. Connect the test leads across a 4.0 Ω (3.6 – 4.4 Ω) resistor. 7. Send command: GAIN=X.XXX
where X.XXX is the actual value of the resistor.
8. Receive * 9. Exchange the 4.0 Ω resistor for a 3.0 Ω resistor. 10. Send command: READ
11. Receive and record response. 12. Disconnect all testing connections.
3-59
ESA612 Service Manual
H. Differential Leakage Calibration (Dependent on AC Meter) 1. Connect the current source to the EO neutral and the 400 mA jack of the DMM. Connect the DMM Input (-) to the EO ground. See Figure 3-34. Set the current source to 15.0 mA (21.21 mA p-p), 120 Hz with a 10 V compliance voltage and the DMM to 100 mA ac range. Fluke 8846A INPUT
Keithly 6221
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Fluke Biomedical ESA612
Figure 3-34. Differential Leakage Calibration Connections
2. Send command: DIFF
3. Receive * 4. Send command: EARTH=C
5. Receive * 6. Send command: POL=N
7. Receive * 8. Turn on the current source output. Note the reading on the DMM. 9. Send command: GAIN=XX.XXXX where XX.XXXX is the actual value of the current in mA measured by the DMM. 10. Receive * 11. Set the current source to 11 mA (15.556 mA p-p). 3-60
gjc118.eps
Verification and Calibration Calibration Procedure
12. Send command: READ
3
13. Receive and record response. Compare it to the actual value of the current. 14. Turn off the current source and disconnect all testing connections. I.
GFI Calibration 1. Connect the current source to the EO neutral and EO ground as shown in Figure 3-35. Set the current source to 5.0 mA (7.071 mA p-p), 60 Hz with a 40 V compliance voltage.
Fluke Biomedical ESA612
Keithly 6221
Figure 3-35. GFI Calibration Connections
2. Send command: PPV
gjc141.eps
3. Receive * 4. Send command: GFI=5MA
5. Receive * 6. Turn on the current source output. 7. Send command: CAL_GFI
8. Receive * 9. Turn off the current source output. 10. Set the current source to 4.5 mA (6.364 mA p-p), 60 Hz. 11. Send command: GFI=5MA
12. Receive * 3-61
ESA612 Service Manual
13. Send command: EQCURR
14. Receive * 15. Turn on the current source output. 16. Send command: STAT3
17. Receive status word 3. The 4xxx bits will not be set if the Analyzer passes the test. Note If the 1xxx and 4xxx bits are set, the status word returned will be 5000. This still indicates the 4xxx bit is set, and is a failing condition. 18. Record PASS/FAIL. 19. Turn off the current source output. 20. Send command: GFIR
21. Receive * 22. Set the current source to 5.5 mA (7.778 mA p-p), 60 Hz. 23. Send command: EQCURR
24. Receive * 25. Turn on the current source output. 26. Send command: STAT3
27. Receive status word 3. The 4xxx bits will be set if the Analyzer passes this test (GFI tripped). 28. Verify the device is in a GFI tripped state: send command: PPV
29. Receive error code !50 (GFI error). 30. Record PASS/FAIL; pass if !50 is received. 31. Turn off the current source output. 32. Send command: GFIR
33. Receive * 34. Set the current source to 10.0 mA (14.142 mA p-p), 60 Hz, with a 40 V Compliance voltage. 35. Send command: PPV
36. Receive * 37. Send command: GFI=10MA
38. Receive * 39. Turn on the current source output. 40. Send command: CAL_GFI
41. Receive * 42. Turn off the current source output. 43. Set the current source to 9 mA (12.728 mA p-p), 60 Hz. 3-62
Verification and Calibration Calibration Procedure
44. Send command: GFI=10MA
3
45. Receive * 46. Send command: EQCURR
47. Receive * 48. Turn on the current source output. 49. Send command: STAT3
50. Receive status word 3. The status bit xxx8 shall be set, indicating Medium GFI level, but the 4xxx bit will NOT be set if the Analyzer passes this test. 51. Turn off the current source output. 52. Record PASS/FAIL 53. Send command: GFIR
54. Receive * 55. Set the current source to 11 mA (15.556 mA p-p), 60 Hz. 56. Send command: GFI=10MA
57. Receive * 58. Send command: EQCURR
59. Receive * 60. Turn on the current source output. 61. Send command: STAT3
62. Receive status word 3. The 4xx8 bit will be set if the Analyzer passes this test (GFI tripped). 63. Turn off the current source output. 64. Record PASS/FAIL 65. Send command: GFIR
66. Receive * 67. Set the current source to 25.0 mA (35.36 mA p-p), 60 Hz. 68. Send command: PPV
69. Receive * 70. Send command: GFI=25MA
71. Receive * 72. Turn on the current source output. 73. Send command: CAL_GFI
74. Receive * 75. Turn off the current source output. 76. Set the current source to 22.5 mA (31.82 mA p-p), 60 Hz. 77. Send command: GFI=25MA
78. Receive * 3-63
ESA612 Service Manual
79. Send command: EQCURR
80. Receive * 81. Turn on the current source output. 82. Send command: STAT3
83. Receive status word 3. The 4xxx bit will NOT be set if the Analyzer passes this test. 84. Turn off the current source output. 85. Record PASS/FAIL 86. Send command: GFIR
87. Receive * 88. Set the current source to 27.5 mA (38.89 mA p-p), 60 Hz. 89. Send command: GFI=25MA
90. Receive * 91. Send command: EQCURR
92. Receive * 93. Turn on the current source output. 94. Send command: STAT3
95. Receive status word 3. The 4xxx bit will be set if the Analyzer passes this test (GFI tripped). 96. Turn off the current source output. 97. Record PASS/FAIL 98. Remove all testing connections. 99. Send command: GFIR 100. Receive *
3-64
Verification and Calibration Calibration Procedure
3
J. Equipment Current Calibration (Dependent on AC Meter) 1. Connect the current shunt to the DMM using the voltage inputs. Connect the EO hot to the DMM Input (+) side of the current shunt and the EO ground to one side of the variable current load. Connect the other side of the variable current load to the current shunt jack corresponding to the (-) DMM Input.
Fluke Biomedical ESA612
Fluke A40A Current Shunt
Fluke 8846A INPUT
SENSE 4W
V 2W/4W
HI
HI
1000 V CAT I 600V CAT II
300V
LO
LO 1V FUSED
100 mA
10A REAR
FRONT
Variable Load
Figure 3-36. Equipment Current Calibration Connection
2. Send command: EQCURR
gjc146.eps
3. Receive * 4. Establish an 8 A draw through the current load, with the shunt resistor in series with the load. 5. Note the actual current flowing in the circuit using the current shunt and DMM. 6. Send command: GAIN=XX.XXXX in amperes.
where XX.XXXX is the actual current calculated
7. Receive * 8. Establish an 3 A draw through the current load. 9. Send command: READ
10. Receive and record response. Compare it with the actual value of the current flowing through the current shunt. 11. Turn off the equipment outlet. Send command: PPL
12. Receive * 13. Remove all testing connections. 14. Clear GFI faults. Send command: GFIR
15. Receive * 3-65
ESA612 Service Manual
K. Mains Calibration Part I Note During the mains calibration, the differential current offset is also calculated and stored. Ensure nothing is connected to the Analyzer except mains power and the USB cable. 1. Send command: MAINS=L1-GND
2. Receive * 3. Send command: GAIN=XXX.X voltage (90 – 264 V ac).
where XXX.X is the actual value of the hot line
4. Receive * 5. Adjust the Elgar power supply to 100 V ac, 60 Hz. 6. Send command: READ
7. Receive and record response. L. Mains Calibration Part II 1. Send command: SAVE
2. Receive * 3. Send command: EXIT
4. Receive * 5. Send command: LOCAL
6. Receive * 7. Turn off the Analyzer. 8. Chang the power supply configuration such that: a. L2 → hot b. L1 → neutral c. GND → ground d. Adjust voltage back to 115 V, 60 Hz. 9. Turn the Analyzer on. The error code !55 will be received. 10. Send command: REMOTE=F2810 11. Receive * 12. Send command: CAL=F2810 13. Receive * 14. Send command: MAINS=L2-GND 15. Receive * 16. Send command: GAIN=XXX.X voltage (90 – 264 V ac).
where XXX.X is the actual value of the hot line
17. Receive * 18. Adjust the Elgar power supply to 100 V ac, 60 Hz. 19. Send command: READ 3-66
Verification and Calibration Calibration Procedure
3
20. Receive and record response. 21. Return the Elgar power supply to 115 V ac, 60 Hz. Do not turn off power to the Analyzer until section M is completed. M. Save Calibration 1. Write new calibration date in the format of MxxDyyYzzzz, where xx stands for the two digits of the current month, yy stands for the two digits of the current day, and zzzz stands for the four digits of the current year. 2. Send command: DATE=MxxDxxYxxxx 3. Receive * 4. Write calibration technician in the format of xxxxx, where xxxxx stands for the five digits of the technician’s employee number. 5. Send command: TECH=xxxxx
6. Receive * 7. Check Analyzer serial number, cal date, and cal technician are correct. 8. Send command: INFO=
9. Receive serial number, cal date, and cal technician in response. Record response. 10. Save the primary cal data. Send command: SAVE 11. Exit cal mode. Send command: EXIT
12. Receive * 13. Return to local mode. Send command: LOCAL
14. Receive * N. Conduct Verification
3-67
ESA612 Service Manual
3-68
Chapter 4
Remote Operations
Title Introduction.......................................................................................................... Serial Communication Connections .................................................................... Serial Communication ......................................................................................... Modes .................................................................................................................. Reading Formats .................................................................................................. Command Specifications ..................................................................................... General Commands ......................................................................................... Remote Commands ......................................................................................... ECG Commands.............................................................................................. Calibration Commands .................................................................................... Diagnostic Commands..................................................................................... Error Responses ................................................................................................... Status Word Definitions....................................................................................... Firmware Updates................................................................................................
Page 4-3 4-3 4-3 4-3 4-5 4-5 4-5 4-6 4-11 4-12 4-18 4-21 4-22 4-24
4-1
ESA612 Service Manual
4-2
Remote Operations Introduction
4
Introduction The Analyzer is designed for remote operation by sending it commands through its computer port. Measurement data and command responses are returned through the computer port for computer or user analysis. This remote capability can be employed to automate the verification, calibration, and troubleshooting procedures described in this manual. Extensive automation can be realized by writing a computer program that sends commands to the Analyzer and evaluates the responses to generate a user report. Or, individual commands can be typed and responses read using a program such as Hyperterminal. The degree of automation is up to the users needs. This chapter describes how to connect and setup the Analyzer for remote operation. Also included is a complete list of valid commands, each with a functional description, proper command syntax, Analyzer response, associated command parameters, and examples. A set of diagnostic commands are also included to help isolate problems in some functions and circuits while troubleshooting the Analyzer.
Serial Communication Connections When unprogrammed devices are connected to a PC, they look different to Windows. Without intervention, Windows will enumerate each device separately. To force Windows to ignore the serial numbers of FT232R devices, add a registry setting. •
Edit the registry to add an entry at: "HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\UsbFlags\".
•
Add a REG_BINARY entry "IgnoreHWSerNum04036001".
•
Set the binary value to "01".
•
Refer to FTDI Application Note AN232B-10 "Advanced Driver Options".
Note This only works if you add the registry entry before connecting any FTDI devices. You might have to uninstall FTDI device drivers with their "FTCLEAN" utility, then reload the drivers. You can force Windows (2000, XP, and above) to use the LocId of the USB port so all FTDI devices on a port are assigned to the same COM port number. Refer to FTDI Application Note AN232B-07 "Configuring FTDI's VCP Drivers to use Location IDs". Note If you properly set IgnoreHWSerNum04036001 as described above, you shouldn't have to use LocId.
Serial Communication These are common specifications for all serial ports: Baud rate: 115,200 baud Format: 8 data bits, no parity, 1 stop bit
Modes The following is a list of Analyzer modes with a brief description of each: 1. Power Up •
The Analyzer performs a number of self-diagnotic tests during this temporary 4-3
ESA612 Service Manual
start-up mode. •
The display shows a “splash” screen during this period.
2. Local •
This is the default mode entered after power-up.
•
The user controls the Analyzer by pressing buttons and viewing the display.
•
The computer interface is active at any of the user menus, but the UI can only execute a few commands: those that put the Analyzer into remote mode or request instrument status. All other commands are illegal.
3. Remote •
The display indicates that the Analyzer is under remote control.
•
The UI receives commands from the USB interface to control the Analyzer.
•
The Analyzer can be put back to local mode by a command or by pressing a key.
•
The UI will directly execute some commands that it can without contacting the Meter.
•
Other commands not executable by the UI will be forwarded to the Meter exactly as received. Then, the UI will get the response from the Meter and pass it back to the computer as received.
4. Cremote •
Same as Remote Mode, except that responses from the Analyzer are enclosed in ‘packets.’ Commands from the PC to the Analyzer must also follow this format.
•
This protocol also applies to sub-modes within remote control.
•
When returning to Local Mode, the communication protocol is restored to normal operation.
•
The packet format is defined below: STX
Byte Count
Data
ETX
Checksum
NULL
1 byte
4 bytes
N bytes
I byte
2 bytes
1 byte
STX = ‘0x02’ character Byte count =
Number of bytes in the Data field only (N bytes)
Data =
Command to or response from the Analyzer
ETX =
‘0x03’ character
Checksum =
Sum of bytes in the byte count and data fields only (4 + N bytes)
NULL =
‘0x00’ character
Note Special characters that do not need to be followed by carriage return/line feed, such as ESC and BACKSPACE characters, must be sent outside of packets. 5. Error
4-4
•
The Analyzer enters the error mode when it detects an internal fault.
•
The display shows an error screen where the user is prompted to take a given
Remote Operations Reading Formats
4
action to correct the error. 6. Test •
This is a special mode for Engineering testing purposes only.
•
The Analyzer is placed in an infinite loop, where it continuously takes measurements that are shown on the display.
7. Sub-modes •
The Analyzer can be put in certain sub-modes within remote control.
•
Sub-modes can allow only certain commands.
•
UI sub-modes for the Analyzer are Diagnostic and Calibration.
Reading Formats Readings returned by the meter microprocessor have a fixed number of decimal places, which may very from test to test. Table 4-1 lists the tests and their associated returned reading formats. Note Readings do not have leading zeroes, except when preceding the decimal point. Table 4-1. Reading Formats Character Preceding Reading Value
Maximum Number of Digits before Decimal Point
Number of Decimal Places
Example [1]
Voltage
V
3
1
“V221.2”
Resistance
O
1
3
“O1.001”
Insulation Resistance
M
2
1
“M5.3”
Current
A
2
1
“A10.4”
Leakage (mA)
L
Depends on range
Depends on range
Depends on range
Leakage (μA)
U
Depends on range
Depends on range
Depends on range
Test
[1]
All reading values are preceded by a letter that specifies the type of reading.
Command Specifications The following sections list the available commands which remotely control the Anaylzer. The commands are divided into the following functional analyzer categories: General, Remote, ECG, Calibration, Diagnostic, and Error. General and global setup commands ready the Analyzer for remote operation from a computer. The calibration commands are used to calibrate the Analyzer remotely. The diagnostic category of commands are for isolating problems in certain Analyzer functions and circuits. General Commands General commands set the Analyzer up for remote operation and are used to extract Analyzer information such as its calibration and current mode of operation.
4-5
ESA612 Service Manual
CREMOTE Description Puts the Analyzer in computer remote control. Parameters
Examples
P = Packet mode
CREMOTE=P
Note To escape from packet mode, return the Analyzer to local mode and reenter remote mode. IDENT Description Queries the Analyzer for identification. Query
IDENT
Returns the Analyzer model, UI firmware version, and meter processor firmware version.
Note When in calibration or diagnostic mode, the UI firmware and meter microprocessor firmware version are listed separately . LOCAL Description Exits remote control and returns to local control. REMOTE Description Sets the Analyzer to remote control mode. RESEND Description Resends the last response to the PC. RSTUI Description Resets the Analyzer. RSTM Description Resets the meter microprocessor. STAT Description Returns the UI status word. Remote Commands The remote commands are used to select or retrieve the tests and perform the necessary setup conditions for the various tests. Some control is also available for retrieving readings from the Analyzer through these commands. ALTEARTH Description Opens or closes the ground to the equipment outlet during Alternative Equipment Leakage function. Parameters
C closes equipment outlet ground and O opens it.
Examples
Closes equipment outlet ground.
ALTEARTH=C
AP Description Selects applied parts to connect to meter (+), applied parts to and connect to meter (-), and where to connect the remaining applied parts. Parameters
4-6
parts+ = list of parts to connect to meter (+) parts- = list of parts to connect to meter (–)
Remote Operations Command Specifications
4
parts remaining = where to connect remaining parts. Examples
AP=RL,LL/RA,V1/GND Connects RL and LL to meter (+), RA and V1 to
meter (-), and all others to ground (GND).
APINS Description Selects applied parts insulation test. AUX Description Select patient auxiliary leakage test. CAL Description Enters Analyzer calibration mode. Parameters
F2810 = Password for the Analyzer.
Examples
Puts Analyzer in calibration mode.
CAL=F2810
DIAG Description Enters Analyzer diagnostic mode. Parameters
F2810 = Password for the Analyzer.
Examples
Puts Analyzer in diagnostic mode.
DIAG=F2810
DIFF Description Selects differential current test. DIRL Description Selects direct leakage test. DMAP Description Selects direct applied part leakage test. Uses MAP voltage. EARTH Description Opens or closes the ground to the equipment outlet. Parameters
C closes equipment outlet ground and O opens it.
Examples
Closes equipment outlet ground.
EARTH=C
EARTHL Description Selects earth leakage test. ECG Description Enters ECG simulation mode and connects all applied parts to ECG. ENCL Description Selects Enclosure Leakage measurement. EOGNULL Description Opens or closes the connection between the NULL jack and the the equipment outlet ground. Parameters
C closes connection between NULL jack and equipment outlet ground and O opens it.
Examples
Closes connection between NULL jack and equipment outlet ground.
EARTH=C
4-7
ESA612 Service Manual
EQCURR Description Selects equipment current test. ERES Description Selects earth resistance 200 mA current. FN Description Returns the current function number. Examples
FN
0 = No function selected 1 = Mains voltage 2 = Equipment current 3 = Earth resistance 4 = Mains to earth insulation 5 = Applied parts to earth insulation 6 = Earth leakage 7 = Enclosure leakage 8 = Patient leakage 9 = Patient auxiliary leakage 10 = Direct equipment leakage 11 = Direct applied parts leakage 12 = MAP leakage 13 = Alternative applied parts leakage 14 = Alternative equipment leakage 15 = Differential leakage 16 = Not used 17 = Point to point leakage 18 = Not used 19 = Point to point voltage 20 = Point to point resistance 21 = Mains to neutral insulation 22 = Applied parts to neutral insulation 23 = Mains to applied parts insulation 24 = Lead isolation leakage
GFI Description Selects ground fault interrupt (GFI) trip level. Parameters
nn = 5, 10, or 25
Examples
Sets ground fault interrupt to 5 mA.
GFI=5MA
GFIR Description Resets ground fault interrupt ATTENTION. HIGH_RES Description Controls the number of digits in the displayed reading. Parameters Examples
HIGH_RES=ON
Increases the number of digits in the displayed reading.
IDLE Description Turns off all relays and ASWs, clears faults and status words. INS Description Controls the insulation resistance test voltage. 4-8
Remote Operations Command Specifications
Parameters
xxxx = Insulation voltage variable. LOW = Selects 250 V insulation voltage HIGH = Selects 500 V insulation voltage (default)
Examples
Sets insulation voltage to 250 V.
INS=LOW
4
INSB Description Selects insulation test B. INSD Description Selects insulation test D. INSE Description Selects insulation test E. LEAD_ISO Description Selects lead isolation function. LOAD Description Selects the load for the meter input. Parameters
xxxx = Meter load variable. 601 = IEC 60601 patient load AAMI = ANSI/AAMI patient load NONE = IEC 60601 patient load
Examples
Selects the IEC 60601 patient load.
LOAD=601
LOADDSP Description Puts the meter microprocessor in boot loader mode. LOCAL Description Puts the Analyzer in local control mode. MAINS Description Selects the mains voltage measurement test. Parameters
L1-L2 = Measures ac, hot to neutral. L1-GND = Measures ac, hot to ground. L2-GND = Measures ac, neutral to ground.
Examples
Sets measurement for ac, hot to neutral.
MAINS=L1-L2
MAP Description Selects the mains on applied parts test. Parameters
LOW = Switches to MAP level of 100 %. NORM = Switches to normal polarity MAP voltage. REV = Switches to reverse polarity MAP voltage. 1MA = Sets MAP limit to 1 mA. 3.5MA = Sets MAP limit to 3.5 mA. 7.5MA = Sets MAP limit to 7.5 mA.
Examples
Sets MAP voltage to 100 %.
MAP=LOW
MINS Description Selects mains insulation test. 4-9
ESA612 Service Manual
MODE Description Selects voltage measurement mode. Parameters
DC = Selects dc measurement mode. AC = Selects ac measurement mode. ACDC = Selects ac+dc measurement mode.
Examples
Selects ac measurement mode.
MODE=AC
MREAD Description Returns meter readings continuously (~400 ms on a single range) until an escape character is received. NEUT Description Opens or closes the neutral to the equipment outlet. Parameters
C = Closes equipment outlet neutral and O opens it.
Examples
Closes neutral to equipment outlet.
NEUT=C
NOMINAL Description Causes all leakage except differential to be multiplied by the ratio of Nominal Mains/Actual Mains. Parameters
ON = leakage multiplied by ratio Nominal Mains/Actual Mains. OFF = leakage without multiplier. xxx = Nominal value to be stored by user.
Examples
Leakage value multiplied by Nominal Mains/Actual Mains.
NOMINAL=ON
NOMINAL? Description Returns the current stored value for nominal. NOSHOW Description Return reading only during MREAD command (default at power-on). OVR Description Resets over voltage ATTENTION. PAT Description Selects patient leakage test. POL Description Sets equipment outlet polarity. Parameters
OFF = Sets equipment outlet polarity off. N = Sets equipment outlet polarity to normal. R = Sets equipment outlet polarity to reverse.
Examples
Sets equipment outlet polarity to reverse.
POL=R
PPL Description Selects point to point leakage test. PPR Description Selects point to point resistance test and set the test current value. Parameters < LOW > 4-10
LOW = Selects low current (200 mA).
Remote Operations Command Specifications
4
Note A PPR command without a parameter just selects the test. Examples
PPR=LOW
Selects the point to point resistance test.
PPV Description Selects point to point voltage test. READ Description Returns a single meter reading. RPTIME Description Sets the equipment outlet (EO) polarity switch time between zero and five seconds. Parameters
n = Time in seconds (0 – 5).
Examples
Sets EO polarity switch time to 3 seconds.
RPTIME=3
SAF Description Selects substitute appliance fault leakage test. SHOWALL Description Returns range, ADC count, and reading during MREAD command. SN Description Returns ESA612 serial number. SPAT Description Selects substitute patient leakage test. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3. STD Description Selects the test standard for tests. Based on the standard, the Analyzer automatically sets the load, GFI trip, and MAP voltage and current values. Parameters
xxxx = Desired test standard. 601 = IEC 60601 standard AAMI = ANSI/AAMI standard ASNZ = AS/NZS 3551 standard 353 = IEC 62353 standard
Examples
Selects the ANSI/AAMI standard.
STD=AAMI
ZERO Description Zero the resistance meter. $ Description Puts the UI microprocessor in boot loader mode. ECG Commands The following commands control the ECG function of the Analyzer. 4-11
ESA612 Service Manual
CPL30 Description Runs ECG complex wave @ 30 Hz. CPL60 Description Runs ECG complex wave @ 60 Hz. CPL120 Description Runs ECG complex wave @ 120 Hz. CPL180 Description Runs ECG complex wave @ 180 Hz. CPL240 Description Runs ECG complex wave @ 240 Hz. EXIT Description Exits ECG simulation mode and disconnects all applied parts. PLS30 Description Runs ECG 63 ms pulse @ 30 Hz. PLS60 Description Runs ECG 63 ms pulse @ 60 Hz. RESEND Description Resends the last response to the PC. SN10 Description Runs ECG sine wave @ 10 Hz. SN40 Description Runs ECG sine wave @ 40 Hz. SN50 Description Runs ECG sine wave @ 50 Hz. SN60 Description Runs ECG sine wave @ 60 Hz. SN100 Description Runs ECG sine wave @ 100 Hz. SQ125 Description Runs ECG square wave @ 0.125 Hz. SQ2 Description Runs ECG square wave @ 2.0 Hz. TR2 Description Runs ECG triangle wave @ 2 Hz. VFIB Description Runs ECG ventricular fibrillation. Calibration Commands The following commands are used to remotely calibrate the Analyzer. APINS 4-12
Remote Operations Command Specifications
4
Description Selects applied parts insulation test. CAL_GFI Description The GFI level must be selected before running this command. Current must be supplied from AC Hot to GND, and CAL_GFI must be executed. This will calibrate the selected GFI (5 mA, 10 mA, or 25 mA) to trip at the current value supplied. CLEAREE Description Clears all data stored in EEPROM (both UI and meter data). COMP Description Generates MAP comp DAC offset value. DATE Description Programs calibration date into Analyzer EEPROM. Parameters
Mxx = Month number Dnn = Day number Ynnnn = Year number
Examples
Sets Calibration date to 11/23/2009.
M11D23Y2009
DEFAULTS Description Sets calibration constants to defaults. DIFF Description Selects differential current test. DIS_FAULT Description Enables or disables GFI and OV faults. Parameters
0 = Enables GFI and OV faults 1 = Disables GFI and OV faults
Examples
Enables GFI and OV faults.
DIS_FAULT=0
EARTH Description Opens or closes the ground to the equipment outlet. Parameters
C = Closes equipment outlet ground and O opens it.
Examples
Closes ground to equipment outlet.
EARTH=C
EQCURR Description Selects equipment current test. ERES Description Selects earth resistance test. Parameters
LOW = 200 mA test current
Examples
Sets earth resistance test current to 200 mA.
ERES=LOW
EXIT Description Exits the calibration mode. FN Description Returns the current function number. 4-13
ESA612 Service Manual
Examples
FN
0 = No function selected 1 = Mains voltage 2 = Equipment current 3 = Earth resistance 4 = Mains to earth insulation 5 = Applied parts to earth insulation 6 = Earth leakage 7 = Enclosure leakage 8 = Patient leakage 9 = Patient auxiliary leakage 10 = Direct equipment leakage 11 = Direct applied parts leakage 12 = MAP leakage 13 = Alternative applied parts leakage 14 = Alternative equipment leakage 15 = Differential leakage 16 = Not used 17 = Point to point leakage 18 = Not used 19 = Point to point voltage 20 = Point to point resistance 21 = Mains to neutral insulation 22 = Applied parts to neutral insulation 23 = Mains to applied parts insulation 24 = Lead isolation leakage 25 = Meter ac 26 = Meter dc
GAIN Description Allows the Analyzer to record current meter value and calculate counts per unit gain, which is then stored in EEPROM. Parameters
nnnn = Actual value of applied signal.
Examples
Command to calibrate current meter value to an applied signal (in this case, mains voltage).
GAIN=115.0
GFI Description Selects ground fault interrupt (GFI) trip level. Parameters
nn = 5, 10, or 25
Examples
Sets ground fault interrupt to 5 mA.
GFI=5MA
GFIR Description Resets ground fault interrupt ATTENTION.
HIGH_RES Description Increases the number of digits displayed in the reading. This command is only used by automated calibration.
4-14
Parameters
ON = increase the number of display digits. OFF = normal number of display digits..
Examples
Increase the number of display digits.
HIGH_RES=ON
Remote Operations Command Specifications
4
INFO Description Returns serial number, calibration date, and technician number. INS Description Controls the insulation resistance test voltage. Parameters
xxxx = Insulation voltage variable. LOW = Selects 250 V insulation voltage HIGH = Selects 500 V insulation voltage (default)
Examples
Sets insulation voltage to 250 V.
INS=LOW
INSWAP Description Swaps MAINS input neutral and hot. Parameters
0 = MAINS neutral and hot to normal polarity 1 = MAINS neutral and hot to reverse polarity.
Examples
Sets MAINS neutral and hot to normal polarity.
INSWAP=0
LEAD_ISO Description Selects the lead isolation function. LOAD Description Selects the load for the meter input. Parameters
xxxx = Meter load variable. 601 = IEC 60601 patient load AAMI = ANSI/AAMI patient load NONE = IEC 60601 patient load
Examples
Selects the IEC 60601 patient load.
LOAD=601
MAINS Description Selects the mains voltage measurement test. Parameters
L1-L2 = Measures ac, hot to neutral. L1-GND = Measures ac, hot to ground. L2-GND = Measures ac, neutral to ground.
Examples
Sets measurement for ac, hot to neutral.
MAINS=L1-L2
MAP Description Selects the mains on applied parts test. Parameters
LOW = Switches to MAP level of 100 %. NORM = Switches to normal polarity MAP voltage. REV = Switches to reverse polarity MAP voltage. 1MA = Sets MAP limit to 1 mA. 3.5MA = Sets MAP limit to 3.5 mA. 7.5MA = Sets MAP limit to 7.5 mA.
Examples
Sets MAP voltage to 100 %.
MAP=LOW
METER_AC Description Selects a test that measures ac voltage on red and black jacks without the potential divider, i.e. connects directly to the meter. 4-15
ESA612 Service Manual
Note Like the PPV test, this command forces IEC filtering. METER_DC Description Selects a test that measures dc voltage on red and black jacks without the potential divider, i.e. connects directly to the meter. Note Like the PPV test, this command forces IEC filtering. MODE Description Selects voltage measurement mode. Parameters
DC = Selects dc measurement mode. AC = Selects ac measurement mode. ACDC = Selects ac+dc measurement mode.
Examples
Selects ac measurement mode.
MODE=AC
MREAD Description Returns meter readings continuously (on a single range) until an escape character is received. NOSHOW Description Return reading only during MREAD command (default at power-on). OFFSET Description Sets offsets to zero all ranges for the current function. OVR Description Resets over voltage ATTENTION. POL Description Sets equipment outlet polarity. Parameters
OFF = Sets equipment outlet polarity off. N = Sets equipment outlet polarity to normal. R = Sets equipment outlet polarity to reverse.
Examples
Sets equipment outlet polarity to reverse.
POL=R
PPL Description Selects point to point leakage test. PPR Description Selects point to point resistance test and set the test current value. Parameters
LOW = Selects low current (200 mA).
Note A PPR command without a parameter just selects the test. Examples
PPR=LOW
Selects the point to point resistance test.
PPV Description Selects point to point voltage test. PSN Description Sets serial number into EEPROM. Parameters 4-16
nnnnnnn = serial number up to seven digits.
Remote Operations Command Specifications
Examples
4
PSN=1234567
RANGE Description Sets the meter range. Parameters
n = range and function numbers (0 – 4).
Query
RANGE?
Returns the number of the meter range.
Examples
RANGE=0,0,1
Sets the ac meter range to range 1.
READ Description Returns a single meter reading with no autoranging. RPTIME Description Sets the equipment outlet (EO) polarity switch time between zero and five seconds. Parameters
n = Time in seconds (0 – 5).
Examples
Sets EO polarity switch time to 3 seconds.
RPTIME=3
RSTM Description Resets the meter processor. RSTUI Description Resets the Analyzer. RSTEECAL Description Clears the meter calibration constants stored in EEPROM. SAVE Description Saves the current calibration constants into EEPROM. SHOWALL Description Returns range, ADC count, and reading during MREAD command. STAT Description Returns the UI ASCII hex status word. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3. STD Description Selects the test standard for tests. Based on the standard, the Analyzer automatically sets the load, GFI trip, and MAP voltage and current values. Parameters
xxxx = Desired test standard. 601 = IEC 60601 standard AAMI = ANSI/AAMI standard ASNZ = AS/NZS 3551 standard 353 = IEC 62353 standard
Examples
Selects the ANSI/AAMI standard.
STD=AAMI
4-17
ESA612 Service Manual
TECH Description Programs technician code into EEPROM. Twenty alphanumeric characters allowed. TIMEOUT Description Turns the UI to DSP communications timeout on and off. Parameters Examples
ON = Turns on timeout. OFF = Turns off timeout
TIMEOUT=ON
UPLOAD Description DSP sends formatted EEPROM data to USB port. ZERO Description Zero the resistance meter. Diagnostic Commands The following commands set specific conditions within the Analyzer and are used strictly for troubleshooting Analyzer problems. These commands allow isolation of, and access to, functional blocks of the Analyzer. These commands can also be used during Analyzer verification and calibration. BEEP Description
Sounds the beeper once.
CONT Description Sets the contrast to the specified value (in ASCII hex). Range is 00 to 11. Examples
CONT=0F
Sets contrast.
DAC Description Sets ECG DAC register to a specified hex value Parameters
xxx = Hex value
Examples
Sets ECG DAC to 0F0.
DAC=0F0
DSPECHO Description Runs continuous DSP communication test. (“ESC” to stop). Characters received from USB are shown on the display and sent to the DSP. The DSP echoes the character back to the UI, which then sends it to the PC. DUPDATE Description Turns the display update on or off. Parameters Examples
DUPDATE=ON
Display update on.
EEPROMV Description Returns the UI EEPROM values. ENCL Description Selects Enclosure Leakage measurement. EXIT Description Exits diagnostic mode. 4-18
Remote Operations Command Specifications
4
FLASH Description Returns the UI flash checksum [XXXX]. FRAMT Description Runs an FRAM test and returns “PASS” or “FAIL”. GFIR Description Resets ground fault interrupt ATTENTION. IDENT Description Queries the Analyzer for identification. Query
IDENT
Returns the Analyzer model, UI firmware version, and meter processor firmware version.
Note When in calibration or diagnostic mode, the UI firmware and meter microprocessor firmware version are listed separately . KEY Description Runs continuous keypad test (“ESC” to stop). Pressed key number is shown in the display. KRL Description Disables all relays by asserting the RELAY_DISABLE line. LED Description Turns LED on or off. Parameters Examples
LED=ON
Turns LED on.
LITE Description Turns backlight on or off. Parameters Examples
LITE=ON
Turns backlight on.
MCS Description Returns the meter microprocessor flash checksum [xxxx]. OTEMP Description Returns the over temperature status bit. OVR Description Resets over voltage ATTENTION. POFF Description Turns off all display pixels. PON Description Turns on all display pixels. PXL Description Write pixel column (8-bit) to the display, given the address and value (all in ASCII hex). Parameters 4-19
ESA612 Service Manual
Examples
Writes FF to row 5 and column 0A7.
PXL=05,0A7,FF
R Description Repeat last command. READ_AD Description Returns A/D value. Takes up to 100 ms. RESEND Description Resends the last response to the PC. RFRAM Description Reads the value (8-bit) from the specified FRAM location. Parameters Examples
RFRAM=0023
xxxx = Memory address (0000 – 7FFF). Returns the 8-bit word at address 0023.
RXFLASH Description Reads the value (8-bit) from the specified external FLASH location. Parameters
xxxx = Memory address (000000 – 20FFFF).
Examples
Returns the 8-bit word at address 000023.
RXFLASH=000023
SET_GFI Description Sets GFI DAC register to a specified hex value Parameters
xxx = Hex value
Examples
Sets GFI DAC register to 00A.
SET_GFI=00A
SET_MAP Description Sets MAP DAC register to a specified hex value Parameters
xxx = Hex value
Examples
Sets MAP DAC register to 002.
SET_MAP=002
SPI Description Writes single SPI byte. Parameters
x = SPI byte (0 – 9) y = value (0x00 – 0xFF) z = loop (Z=1 means loop until ESC).
Examples
Write 01 to SPI byte 5 repeatedly until escape character is received.
SPI=5,01,1
SPIR Description Returns SPI 0-9 values. STAT1 Description Returns ASCII hex status word 1. STAT2 Description Returns ASCII hex status word 2. STAT3 Description Returns ASCII hex status word 3.
4-20
Remote Operations Error Responses
4
UIECHO Description Runs continuous USB test (“ESC” to stop). Characters received from USB are shown on the display and echoed back to the PC. WFRAM Description Writes the value (8-bit) to the specified external FRAM location. Parameters ,
xxxx = Memory address (0000 – 7FFF) dd = data.
Examples
Writes A3 to address 0023.
WFRAM=0023, A3
WXFLASH Description Writes the value (8-bit) from the specified external FLASH location. Parameters Examples
xxxxxx = Memory address (000000 – 20FFFF) dd = data.
WXFLASH=000023, A3 Writes A3 to address 000023.
XFLASHT Description Runs a test on the external FLASH chip and returns “PASS” or “FAIL”. ZOMBIE Description Places the Analyzer in test mode. Parameters Examples
Turns test mode on.
ZOMBIE=ON
Error Responses Table 4-2 list the error codes and their descriptions, that may be returned from the Analyzer. Table 4-2. Error Responses Code
Description
00
No commands allowed now
01
Unknown command
02
Illegal command, not allowed in current mode
03
Illegal parameter
04
Receive buffer overrun
05
General failure
06
Option not installed
21
Analog to Digital Converter out of range
30
Test passed
31
Test failed
32
No current
33
Can not null
37
No reading available
4-21
ESA612 Service Manual
Table 4-2. Error Responses (cont.) Code
Description
38
Load discharge timeout
40
Over temperature
41
CREMOTE protocol error
42
Initialization error
50
GFI
51
Overvoltage
52
Analyzer out of calibration
53
Mains out of range
54
Open ground
55
Reverse voltage
56
Polarity time wait
58
External memory error
Status Word Definitions Table 4-3 lists the status word bit definitions. Table 4-3. Status Word Definitions Description
Status Word
Definition
UI Status Word Bit Definitions
4-22
POWER_UP
0x0001
Device is in power-up mode
LOCAL
0x0002
Device is in local mode
REMOTE
0x0004
Device is in remote mode
CREMOTE
0x0008
Device is in cremote mode
DIAG
0x0010
Device is in diagnostic mode
CAL
0x0020
Device is in cal mode
ERROR
0x0040
Device is in error mode
TEST
0x0080
Device is in test mode
OVER_TEMP
0x0100
Overtemp condition
SPARE
0x0200
Spare
SPARE
0x0400
Spare
SPARE
0x0800
Spare
SPARE
0x1000
Spare
SPARE
0x2000
Spare
SPARE
0x4000
Spare
Remote Operations Status Word Definitions
4
Table 4-3. Status Word Definitions (cont.) Description SPARE
Status Word
Definition
0x8000
Spare
REMOTE
0x0001
Device is in remote mode
DIAG
0x0002
Device is in diagnostic mode
CAL
0x0004
Device is in cal mode
ECG
0x0008
Device is in ECG mode
Spare
0x0010
Spare
SVOLTS
0x0020
Measure from 0 to 300 volts
SLEAK
0x0040
Measure from 0 to 10,000 μA
SOHMS
0x0080
Measure from 0 to 2 Ω @ 200 mA
Spare
0x0100
Spare
SMEG
0x0200
Measure from 0 to 100 MΩ
SEQUIP
0x0400
Measure from 0 to 20 A ac
SDIFF
0x0800
Measure from 0 to 10 mA ac
AC_ONLY
0x1000
Measure ac only
DC_ONLY
0x2000
Measure dc only
ACDC
0x4000
Measure ac+dc
SPARE
0x8000
Spare
LDAAMI
0x0001
AAMI load selected
Spare
0x0002
Spare
LD601
0x0004
601 load selected
EO
0x0008
Equipment outlet on
Spare
0x0010
Spare
MAPR
0x0020
MAP reverse selected
MAPON
0x0040
MAP voltage on
L2OPEN
0x0080
Neutral open
EOPEN
0x0100
Earth open
POLR
0x0200
EO polarity reversed
GFIL
0x0400
GFI low selected
GFIH
0x0800
GFI high selected
INS_ON
0x1000
Insulation voltage (500 V) on
RCURON
0x2000
Resistance current on
Status Word 1 Bit Definitions
Status Word 2 Bit Definitions
4-23
ESA612 Service Manual
Table 4-3. Status Word Definitions (cont.) Description
Status Word
Definition
MAINS0
0x4000
MAINS0-MAINS1 = Mains parameter selection.
MAINS1
0x8000
MAINS0-MAINS1: 00 = Not Used 01 = L2-GND 10 = L1-GND 11 = L1-L2
RPT0
0x0001
RPT0 – RPT2 reverse EO polarity switch time = 0 – 3.5 sec
RPT1
0x0002
RPT0 – RPT2 is a count of 1 sec increments
RPT2
0x0004
GFIM
0x0008
GFI medium level (10 mA) selected
SHOWALL
0x0010
SHOWALL selected
NOMINAL
0x0020
NOMINAL selected
INS_LOW
0x0040
1 – insulation volts = 250 V; 0 – insulation volts = 500 V
MAP3MA
0x0080
MAP test – current limit = 3.5 mA if on or 1 mA if off
Spare
0x0100
Spare
MAINS
0x0200
0 = 115 V ac, 1 = 230 V ac Mains
EEP_CS_ERR
0x0400
EEPROM checksum error
VOLT_BAD
0x0800
MAINS – voltage L1-L2 out of range
BAD_GND
0x1000
MAINS – AC GND bad
REV_PWR
0x2000
MAINS – L1 and L2 reversed
GFITRIP
0x4000
GFI interrupt has occurred
FAULT
0x8000
FAULT interrupt has occurred
Status Word 3 Bit Definitions
Firmware Updates To update the firmware in the Analyzer, two firmware updates must be done: one for the meter and one for the UI.
4-24
Remote Operations Firmware Updates
4
W Caution To prevent damage to the Analyzer, do not interrupt the firmware download process. If this should happen, then both boards will have to be returned to the factory for reinitialization. Do not attempt to return the Analyzer to an earlier firmware version. This could result in a corrupt firmware load, and both boards will have to be returned to the factory for reinitialization. 1. With the Analyzer powered off, connect your PC to the Analyzer and run AnsurMLC.exe. 2. Click Tools->Upload Firmware. 3. Select ESA612, select the appropriate COM port, and click NEXT. 4. Browse to the appropriate DSP and UI firmware HEX files, and click NEXT. 5. Power the Analyzer on. 6. Click UPLOAD. 7. The LCD will blank but AnsurMLC will begin downloading the new firmware. When both downloads are complete, cycle power to the Analyzer and verify the new firmware has been correctly uploaded.
4-25
ESA612 Service Manual
4-26
Chapter 5
Maintenance
Title Introduction.......................................................................................................... Testing and Replacing the Fuses.......................................................................... Cleaning the Analyzer ......................................................................................... Disassembly/Assembly ........................................................................................ Troubleshooting ................................................................................................... Power on Failure.............................................................................................. Output Failures ................................................................................................ Meter Failures.................................................................................................. MAINS Power............................................................................................. Differential Current and Equipment Current .............................................. Meter Troubleshooting ....................................................................................
Page 5-3 5-3 5-3 5-4 5-4 5-4 5-5 5-5 5-5 5-5 5-5
5-1
ESA612 Service Manual
5-2
Maintenance Introduction
5
Introduction This chapter covers the general maintenance as well as troubleshooting information for isolating failed Analyzer components. The Analyzer needs little maintenance or special care. However, treat it as a calibrated measuring instrument. Avoid dropping or other mechanical abuse that could cause a shift in the calibrated settings.
Testing and Replacing the Fuses XW Warning To prevent electric shock, remove all power cords and test leads from the Analyzer before opening the fuse door. For electrical protection of the equipment outlet, the Analyzer uses two fuses, one in the live (L1) line and one in the neutral (L2) line. To test the fuses, do the following while referring to Figure 5-1: 1. Turn the Analyzer so the case bottom is facing up. 2. Flip up the tilt stand. 3. Remove the fuse door from the Analyzer by removing the screw holding the fuse door with a #2 Phillips head screwdriver and lifting the fuse door from the Analyzer. 4. Remove the fuses from the Analyzer.
F1 - F2
Figure 5-1. Fuse Access
fis111.eps
5. Using a multimeter, measure the continuity of each fuse. If one or both fuses do not show continuity, replace the fuse(s) with fuses that have the same current and voltage rating. Appropriate fuse ratings are posted on the case bottom label of the Analyzer. The final assembly table in the Replaceable Parts chapter lists available fuses with Fluke Biomedical part numbers. 6. Reinstall the fuse door and secure it with the screw.
Cleaning the Analyzer XW Warning To avoid electric shock, do not clean the Analyzer plugged into mains or attached to a DUT. W Caution Do not pour fluid onto the Analyzer surface; fluid seepage into the electrical circuitry may cause the Analyzer to fail. 5-3
ESA612 Service Manual
W Caution Do not use spray cleaners on the Analyzer; such action may force cleaning fluid into the Analyzer and damage electronic components. Clean the Analyzer occasionally utilizing a damp cloth and mild detergent. Take care to prevent the entrance of liquids. Wipe down the adapter cables with the same care. Inspect them for damage to and deterioration of the insulation. Check the connections for integrity before each use.
Disassembly/Assembly To disassemble the Analyzer, do the following: 1. Turn the Analyzer so the case bottom is facing up. 2. Remove the six screws in the case bottom with a Philips screwdriver 3. Lift the case bottom off of the case top and disconnect the cables that connect the two circuit boards together. The A1 Power PCA will remain attached to the case bottom and the A2 Meter PCA will remain with the case top. To reassemble the Analyzer, follow the disassembly steps in reverse order.
Troubleshooting The verification procedure contained in this manual is an effective troubleshooting tool; based upon the section of the test that has failed, or on the failure mode, the component that has caused the malfunction can be isolated. Power on Failure If the Analyzer fails to power up, verify fuses F1 and F2 inside the rear fuse-panel door have not blown. If they have not blown, verify power is applied to the power inlet assembly and that the wiring on the power switch internal to the device is correct. Failures during power up are usually related to mismatched firmware or to power supply problems. To verify firmware, if the device will power up and gives an “Initialization Error,” it may be possible to reach the remote mode through the USB port. Establish hyperterminal communication with the device and send the following command: REMOTE=F2810
If the device displays “Remote Mode” on the front panel and responds with a “*,” determine the firmware status. CAL=F2810 Receive * IDENT
Receive the meter and UI firmware builds. Verify that the two are matched; if they are not, upload the latest firmware to the device and try again. If the firmware matches and is up to date, verify the power and data cables connecting the A1 Power PCA and A2 Meter PCA are connected at J4 and J7 on the power board and at J4 and J20, respectively, on the meter board; also inspect the soldering of the connectors to the pcas. If these connections are not at fault, verify the power supplies for the processors and for the device as a whole are functional. First verify U9, the power supply module at the bottom of the power board, is functional and that CR4, the red LED, is on. Then verify the outputs of U12 and U28 are correct. On the meter board, check that Q9 and U48 – U53 are all operating correctly. Finally, check the power inputs and resets to the microprocessors.
5-4
Maintenance Troubleshooting
5
Output Failures If sourcing a current or voltage through the electrical outlet, such as during an Earth Resistance, or powering a device with normal or reversed polarity, such as during a leakage test, and no signal is applied to the electrical outlet, failures could reside in the physical construction of the electrical outlet, in the reversing relays, or in the signal generation circuit. Verify the electrical outlet wiring is in accordance with the J1 annotation on the pca. Also, if troubleshooting an Australian Analyzer, measure the impedance of the ground connector of the electrical outlet; some outlets have been found to have more than 0.5 Ω. These should be replaced. For US electrical outlets, ensure the backing plate is fully installed; this forces the insulation displacement crimp connection to engage. If this backing plate has been separated, replace the entire electrical outlet. If sourcing a mains power source through the electrical outlet, measure the voltage at the electrical outlet in normal and reversed polarity. If the power configuration is incorrect, inspect the reversing relays on the power PCB: K1-K2, K4-K7. If GFCI trips are also a symptom, verify proper operation of the solid-state relay. K9 and K10 deliver the insulation resistance and MAP signals to the equipment outlet, respectively, and are a good place to half-split the source circuit. The protective earth resistance signal is selected via K12. For leakage measurements, the return path from the electrical outlet is selected by K14. The MAP signal is generated by transformer T1 and regulated by Q1, Q2, and the supporting resistors and is routed through K16 and K17. The insulation resistance voltage is created by transformer T2 and rectified by the nearby circuitry, including Q7; the resultant signal is sensed by U24. Meter Failures While most measurements made by the Analyzer are made by the A2 Meter PCA directly, several are detected or analyzed directly on the A1 Power PCA. MAINS Power Most Mains voltage measurements that are erroneous are the result of power supply failures in U9, isolated supply U18, or in calibration failures. If a calibration error is suspected, calibrate the Mains voltage to determine if this can correct the problem. Otherwise, evaluate F3, the resistor network R2 through R7, U3, and U4 for failures. Differential Current and Equipment Current Both differential and equipment currents are measured at the outlet of the reversing relays, just prior to the Equipment Outlet using L1 and L3, respectively, installed on the backside of the A1 Power PCA beneath the SSR. If all the solder connections on the inductors are good, verify U1, U8, and the supporting circuitry. Meter Troubleshooting If the source signal is functional for a particular measurement, but no measurement is read by the Analyzer, then the meter inputs are suspect. K13 and K15-K18 connect the meter jacks, Red, Black, and Null, to the positive and negative sides of the meter for measurement or resistance nulling. The patient load is the circuit by which a human body is simulated for patient leakage measurements, and is comprised of R78 and R79 (800 and 200 Ω, respectively). For greater leakage currents of 2 mA or more, the meter input is shifted from R78 to R79. To troubleshoot leakage measurements, for ac measurements, half split the circuit at the output of the rms converter, U23 pin 5 or the input of the low pass filter, U31 pin 3. The input to the analog-to-digital converter for ac measurements is at TP95. For dc measurements, use the output of the low pass filter, TP 91. 5-5
ESA612 Service Manual
Resistance measurements are driven by the power supplies on the A1 Power PCA, but is inverted by U15. Applied parts problems should be troubleshot separately than leakage measurements, if possible. First, determine if a single ECG post is affected, or the entire bank. If a single ECG has a failure, inspect the diodes CR24, 29, and 46, L9, or relays K9, K10, and K22 (or respective components with other ECG posts). If all applied parts are affected equally, verify K11 or inspect the signal source (MAP, 500 V dc, etc). For ECG waveform problems, the buffer U13 or resistor network of R5361 drive the entire circuit; the waveform itself is generated by the meter microprocessor. U/I and meter microprocessor failures will normally result in complete nonfunctionality of the Analyzer. Calibration data table and configuration information failures indicate a problem in U44; a loss or corruption of stored test data indicates a failure in U1 or U59. Display board failures are investigated through the driver, which consists of U3 and the power supply of U5, U6, R22 and Q3. Otherwise, the display is replaced as a unit.
5-6
Chapter 6
Replaceable Parts
Title
Page
Introduction.......................................................................................................... 6-3 How to Obtain Parts............................................................................................. 6-3 Parts List .............................................................................................................. 6-4
6-1
ESA612 Service Manual
6-2
Replaceable Parts Introduction
6
Introduction This section contains an illustrated list of replaceable parts for the Analyzer. Parts are listed by assembly, alphabetized by reference designator. Each assembly is accompanied by an illustration showing the location of each part and its reference designator. Refer to Table 6-1. The parts lists give the following information: • • • •
Reference designator (for example, “R52”) An indication if the part is subject to damage by static discharge (* near the part description) Description Fluke stock number WCaution A * symbol indicates a device that may be damaged by static discharge.
How to Obtain Parts Electronic components may be ordered directly from Fluke Biomedical and its authorized representatives by using the part number under the heading Fluke Stock No. Parts price information is available from Fluke Biomedical or its representatives. To contact Fluke Biomedical, call one of the following telephone numbers: • • •
North America 1-800-850-4608 (Email: [email protected]) Europe, Middle East, and Africa +31-402-675-300 (Email: [email protected]) All other countries 1-440-248-9300 (Email: [email protected])
In the event that the part ordered has been replaced by a new or improved part, the replacement will be accompanied by an explanatory note and installation instructions, if necessary. To ensure prompt delivery of the correct part, include the following information when you place an order: • • • • • •
Instrument model and serial number Part number and revision level of the pca (printed circuit assembly) containing the part Reference designator Fluke stock number Description (as given under the Description heading) Quantity
6-3
ESA612 Service Manual
Parts List Tables 6-1 though 6-3 lists the replaceable parts of the Analyzer and their location identified in Figures 6-1 though 6-3. Table 6-1. Final Assembly Item
Description
Fluke Stock No.
A1
ESA612-4002,PCA, POWER
3316505
A2
ESA612-4001,PCA, METER
3316497
C1
ESA612-4401,CABLE ASSY, AC INPUT
3316689
C2
ESA612-4402,CABLE ASSY, POWER -METER
3316692
C3
ESA612-4403,CABLE ASSY, POWER TO METER SUPPLY
3316704
US Japan
WFUSE,20A,250V,3AB,TIME LAG,.25X1.25,BULK
2183691
WFUSE,16A,250V,TIME LAG,6.3X32,CERAMIC BODY,BULK
3321245
WFUSE ,FUSE,.25X1.25,10A,250V,SLOW
109298
France Europe F1, F2
Israel United Kingdom Thailand Australia Switzerland
6-4
H1
SCREW,5-14,.750,PAN,PHILLIPS,STEEL,BLACK CHROMATE,THD FORMING
832246
H2
SCREW,4-14,.375,PAN,PHILLIPS,STEEL,ZINC-ROHS CLEAR,THREAD FORM
448456
H3
SCREW,M3X0.5,6MM,PAN,PHILLIPS,STEEL,ZINC-BLACK CHROMATE
2032792
H4
SCREW,M3X0.5,5MM,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK
2032811
H5
SCREW,6-32,.312,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK
152157
H6
SCREW,8-32,.375,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK
114124
MP1
ESA612-2008,CASE TOP ASSY
3274779
MP2
ESA612-2010,CASE BOTTOM ASSY
3276744
MP3
ESA612-2007-02,FEET SET, LEFT, RIGHT
3359523
MP4
ESA612-2004,DOOR, FUSE ACCESS
3236200
MP5
ESA612-2009,PLATE, TOP END
3236274
MP6
ESA612-2006,BRACKET, DISPLAY
3236239
MP7
ESA612-2019,INSULATOR, INPUT RECEPTACLE
3383167
MP8
ESA612-8002,KEYPAD, UPPER
3258971
MP9
ESA612-8002-01,KEYPAD, LOWER
3258980
MP10
CONTRACT MFG ITEM, RELAY SOLID STATE 75A VDECSA
2077927
Replaceable Parts Parts List
6
Table 6-1. Final Assembly (cont.) Item
Description
Fluke Stock No.
MP11
LCD MODULE,320X240,FSTN,3.8 IN MONO GRAPHIC,TRANSFLCTV,WHITE LED,0.22MM PITCH
3299802
MP12
POWER ENTRY MODULE,A/C INLET,ON/OFF SWITCH,2POLE,20A,250VAC,SNAP-IN,BULK
2791306
MP13
LABEL,SERIAL NUMBER
3097856
US
ESA612-8009,DECAL, BOTTOM CASE, 20 AMP
3343933
ESA612-8009-01,DECAL, BOTTOM CASE, 16 AMP
3343940
ESA612-8009-02,DECAL, BOTTOM CASE, 10 AMP
3343957
ESA612-4405,WIRE ASSY, EQUIPMENT OUTLET
3316728
France
ESA612-4411,WIRE ASSY, EQUIPMENT OUTLET
3343887
Europe
ESA612-4412,WIRE ASSY, EQUIPMENT OUTLET
3343893
Israel
ESA612-4413,WIRE ASSY, EQUIPMENT OUTLET
3343902
Australia
ESA612-4415,WIRE ASSY, EQUIPMENT OUTLET
3343916
United Kingdom
ESA612-4410,WIRE ASSY, EQUIPMENT OUTLET
3316762
Switzerland
ESA612-4416,WIRE ASSY, EQUIPMENT OUTLET
3343925
Thailand
ESA612-4418,WIRE ASSY, EQUIPMENT OUTLET
3474220
LINE CORD,N AMER,20A/125V,5-20/C19,12-3 SJT,2.5M,BLK
2238680
75026,PWRCORDSET EUROPE W/IEC320
2238615
LINE CORD,ISRAEL,16A/250V,SI 32 STRAIGHT OR RT ANG/C19,BLACK,2.5 METERS
2434122
75025,PWRCORD SET AUST W/IEC320
2238603
United Kingdom
75024,PWRCORD SET UK W/IEC320
2238596
Switzerland
FBC-ESA620-4420,SWISS 10A POWER CORD ASSY WITH C-19 CONNECTOR
3379149
Thailand
75033,POWERCORD US WITH C19 PLUG
2238644
Japan France Europe MP14
Israel United Kingdom Thailand Australia Switzerland US Japan
MP15
US Japan France Europe Not Israel Shown Australia
Not ESA612-2016,BJ2ECG ADAPTER ASSY Shown
3359538
6-5
ESA612 Service Manual
H1 (6x) MP3 MP15
MP14 MP13
MP4 H3
MP10, H5 (2x), H6 (2x) A2 C1 MP2
MP5
F2 F1
MP12
C2 MP7 (5x)
H2 (7x) C3 H3 (3x)
H2 (2x)
H2
A1
MP6
MP11 MP8 MP9
MP1
Figure 6-1. Final Assembly
6-6
gjc145.eps
Replaceable Parts Parts List
6
Table 6-2. A1 Power PCA
Item
Description
Fluke Stock No.
C2 C4 C6 C7 C11 C14 C17 C19-26 CAPACITOR,CERAMIC,0.1UF,+-10%,50V,X7R,0805,TAPE C32 C35 C36 C41 C43
690500
C3 C12 C15 C18 C28 C53 C55 C60
CAPACITOR,ELECTROLYTIC,TANTALUM,10UF,+-20%,16V,3216,TAPE
3093654
C5 C38-40
CAPACITOR,CERAMIC,1UF,+-20%,25V,X5R,0805,TAPE
2530275
C8 C9 C37
CAPACITOR,CERAMIC,0.047UF,+-10%,50V,X7R,0805,TAPE
1708909
C10
CAPACITOR SMR,CAP,CER,4700PF,+-10%,50V,X7R,0805
604231
C13 C54
CAPACITOR,CERAMIC,0.01UF,+-10%,50V,X7R,0805,TAPE
106146
C27 C74
CAPACITOR,CERAMIC,10PF,+-5%,100V,C0G,0805,TAPE
3331432
C29 C42
CAPACITOR,CERAMIC,0.047UF,\+-10%,500V,X7R,1812,TAPE
644077
C30 C33 C44 C57 C73
CAPACITOR SMR,CAP,CER,1000PF,+-10%,50V,X7R,0805
484378
C31
CAPACITOR SMR,CAP,CER,470PF,+-1%,50V,C0G,0805
929476
C34
560180002224,CAP-SM,0805,16V,10%,X7R,.22UF
2331299
C45-52 C56 C58 C59 C61-71
CAPACITOR SMR,CAP,CER,100PF,+-1%,50V,C0G,0805
867650
C72
CAPACITOR,ELECTROLYTIC,TANTALUM,47UF,+-10%,16V,200MOHM ESR,7343,TAPE
3352434
CR1
ZENER,UNCOMP,MMBZ5235B,6.8V,5%,20MA,225MW,SOT-23,TAPE
837195
CR2 CR3 CR11 CR19
ZENER,UNCOMP,MMBZ5242B,12V,5%,20MA,225MW,SOT-23,TAPE
2044054
CR4
LED,RED,AM2520,280MCD,20MA,2.5V,20 DEG VIEW ANGLE,GULL WING PREP,T3/4,TAPE
2281407
CR5 CR10
DIODE,SI,PN,S1M,1KV,1A,75NS,SMA,TAPE
2087962
CR6 CR1618 CR20
DIODE,SI,BAS16,75V,200MA,6NS,SOT-23,TAPE
830489
CR7 CR1215
DIODE,SI,PN,S1M,1KV,1A,1.8US,DO-214AC,TAPE
2646221
F3
FUSE,2.5A,250V,TIME LAG,8X8.5MM,RADIAL
3331539
H1 H2
SPACER,ROUND,4.2MM ID,2.0MM HEIGHT,8.74MM OD,STEEL,TIN,SMD,TAPE
3350760
H3 H4
SPACER,ROUND,3.6MM ID,2.0MM HEIGHT,7.14MM OD,STEEL,TIN,SMD,TAPE
3352172
J1
CONNECTOR,TERMINAL BLOCK,6.35MMCTR,VERT PWB,30A,300V,3 POS
3342141
J2
CONNECTOR,SOCKET,2 ROW,3.0MM CTR,VERT SMD,6 POS,TAPE
3343846
6-7
ESA612 Service Manual
Table 6-2. A1 Power PCA (cont.)
Item
Description
J4
CONNECTOR,HEADER,2 ROW,0.1 CTR,VERT SMT,SHROUDED,20 POS,TAPE
2791599
J5
CONNECTOR,HEADER,1 ROW,.25 CTR,RT ANG PWB,LOCKING,W/EARS,5 POS,BULK
3351138
K1-7 K11
RELAY,ARMATURE,1 FORM C,12V,1 C/O CONTACT.16A,29X12.7MM,BULK
2152365
K8 K9 K13 K17
RELAY,ARMATURE,2 FORM C,12VDC,SINGLE STABLE,15X7.4MM,TAPE
2082903
K10 K12 K14 RELAY,ARMATURE,2 FORM C,12V,15X9MM,SMT,TAPE K16
6-8
Fluke Stock No.
3301088
L1
TRANSFORMER,SIGNAL,200:1,DIFFERENTIAL,SHIELDED,VERTICAL TOROID,26X27X11MM,4-PIN,BULK
3405268
L3
TRANSFORMER,SIGNAL,CURRENT SENSOR,2-20A,100 OHM,50/60HZ,3PIN,BULK
2750107
Q1
MOSFET,SI,N,IRFBF20S,900V,1.7A,8 OHMS,3.1W,D2-PAK,TAPE
2742520
Q2
TRANSISTOR,SI,NPN,MMBT5089,30V,50MA,50MHZ,200MW,SOT-23,TAPE
820902
Q3
MOSFET,SI,N,IRLR3410,100V,17A,105 MOHMS,3W,D-PAK,TAPE
1579273
Q4
MOSFET,SI,N,FQB5N90,900V,5.4A,2.3 OHMS,3.13W,D2-PAK,TAPE
3343629
Q5 Q6
MOSFET,SI,N,2N7002E,60V,240MA,3 OHMS,350MW,SOT-23,TAPE
1756473
Q7-9
TRANSISTOR,SI,NPN,MUN2216,50V,100MA,DIGITAL,R1=4.7K,R2=INF,230M 2002486 W,SC-59,TAPE
R2-7 R23 R24 R27
RESISTOR,METAL FILM,330K,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE
3326205
R9 R16 R20 R43 R58
RESISTOR,CERMET,100,+-1%,0.125W,100PPM,0805,TAPE
928937
R12
RESISTOR,CERMET,10,+-1%,0.125W,100PPM,0805,TAPE
928924
R13
RESISTOR,CERMET,16.2K,+-1%,0.125W,100PPM,0805,TAPE
2005953
R14 R15 R25 R30 R31 R33 R34 R39 R44 R54 R62 R65 R69-74 R7983 R88-91 R96-99 R101-113
RESISTOR,CERMET,1K,+-1%,0.1W,100PPM,0805,TAPE
928713
R17 R18 R21 R22
RESISTOR,CERMET,6.2K,+-1%,0.125W,100PPM,0805,TAPE
3336652
R19 R52 R67 R68 R128
RESISTOR,CERMET,100K,+-1%,0.125W,100PPM,0805,TAPE
928866
R26
RESISTOR,CERMET,7.5K,+-1%,0.125W,100PPM,0805,TAPE
2278492
R10
RESISTOR SMR,RES,CERM,332,+-1%,0.1W,100PPM,0805
604926
R11 R100
RESISTOR,CERMET,33,+-1%,0.125W,100PPM,0805,TAPE
2803420
Replaceable Parts Parts List
6
Table 6-2. A1 Power PCA (cont.)
Item
Description
Fluke Stock No.
R29
RESISTOR SMR,RES,CERM,1.82K,+-1%,0.1W,100PPM,0805
930172
R32 R37 R49 R51 R55 R61
RESISTOR,CERMET,10K,+-1%,0.125W,100PPM,0805,TAPE
928791
R41 R56 R57
RESISTOR,METAL FILM,1M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE
3326231
R42 R53 R59 R76 R85 R86
RESISTOR,CERMET,1,+-1%,0.125W,400PPM,0805,TAPE
2005982
R45
RESISTOR,CERMET,56K,+-1%,0.125W,100PPM,0805,TAPE
3331360
R46
RESISTOR,CERMET,11.5K,+-1%,0.1W,100PPM,0805,TAPE
1997330
R47
RESISTOR,CERMET,1M,\+-1%,0.125W,100PPM,0805,TAPE
928945
R48
RESISTOR,CERMET,4.7K,+-1%,0.125W,100PPM,0805,TAPE
3326333
R50
RESISTOR SMR,RES,CERM,332K,+-1%,0.1W,100PPM,0805
602695
R60
RESISTOR,CERMET,470,+-1%,0.125W,100PPM,0805,TAPE
2803447
R63 R64
RESISTOR,CERMET,68K,+-5%,1W,200PPM,2512,TAPE
3331385
R66
RESISTOR,CERMET,23.2K,+-1%,0.125W,100PPM,0805,TAPE
3331397
R77
RESISTOR,CERMET,6.81K,+-1%,0.1W,100PPM,0805,TAPE
2087241
R78
RESISTOR,CERMET,3.6K,+-1%,0.125W,100PPM,0805,TAPE
3331415
R92-95
RESISTOR,CERMET,200,+-1%,0.125W,100PPM,0805,TAPE
3372234
R114-118
RESISTOR,CERMET,2M,+-5%,1W,200PPM,2512,TAPE
3310288
R119-127
RESISTOR,CERMET,JUMPER,0,+0.05 MAX,0.125W,1206,TAPE
810747
RV1 RV2
VARISTOR
706838
RV40 RV75
THERMISTOR,POSITIVE,1.1K,+-20%,COATED,RADIAL,TAPE
1277360
T1
TRANSFORMER,POWER,115/230V,50/60HZ,[email protected],0.51VA,EI30/1 2.5,PC MOUNT,BULK
3342134
T2
TRANSFORMER,FLY-BACK,6V,20KHZ,1:10:10,1.5W,HV,FLUKE165X,RM6,SMT8,TAPE
2032204
TP12 TP29 TP33-37
CONNECTOR,TERMINAL,TEST POINT,SMD,510 PH BRONZE,TAPE
602125
U1
IC,OP AMP,AD8510,+-4.5V TO +-18V,0.9MV OFFSET,8MHZ,PRECISION,JFET,S08,TAPE
2434633
U2 U5
IC,DIGITAL ISOLATOR,ADUM1401,QUAD,3/1,10MHZ,2500V,3V/5V,SOICW16,TAPE
3331444
U3
IC,ADC,MCP3202,2.7-5.5V,12BIT,2-CHANNEL,SPI,S08,TAPE
3331459
U4
IC,OP AMP,TLV2372,2.7V TO 16V,4.5MV OFFSET,3MHZ,DUAL,R/R,S/S,SO8,TAPE
3329934
U6 U10 U11 U19-21 U29
OPTICAL,OPTOCOUPLER,PHOTO TRANSISTOR,SMD DIP4,TUBE
3328473
U7
IC,COMPARATOR,LMV7271,1.8V TO 5V,4MV OFFSET,R/R INPUT,LO PW,800NS PD,SOT-23-5,TAPE
3332950
U8
IC,ANALOG SWITCH,ADG431,5-44V,24 OHMS,QUAD,SPST,NO,SO16,TAPE
3367437
R05R,VARISTOR,430V,+-10%,1MA
6-9
ESA612 Service Manual
Table 6-2. A1 Power PCA (cont.)
Item
6-10
Description
Fluke Stock No.
U9
POWER SUPPLY,SW,15W,85-264VAC,[email protected],32X64MM,PCB MOUNT,BULK
3331480
U12 U28
IC,VOLTAGE REGULATOR,LINEAR,LP2985,3.3V,150MA,LDO,LOW POWER,SOT-23-5,TAPE
2559443
U13 U15
TRANSISTOR,SI,NPN,ULN2003,50V,500MA,HEPTADARLINGTON,3.5W,SO16,TAPE
2113871
U14 U16 U25
IC,LOGIC,74HC595,2.0V-6.0V,8-BIT SHIFT REGISTER,TRISTATE,SOIC16,TAPE
904388
U17
IC,PULSE WIDTH MODULATOR,UCC3805,4V,35MA,1MHZ,10V SUPPLY,SO8,TAPE
1541378
U18
POWER SUPPLY,DC-DC,2W,12VDC,+-5V@200MA,3KV ISO,SIP5,TUBE
3338710
U22
603022-008,VOLTAGE REF 1.2V LM4041CM3
1790839
U23 U24
IC,OP AMP,TLV2371,2.7V TO 16V,4.5MV OFFSET,3MHZ,R/R,S/S,SOT-235,TAPE
3331500
U26 U27
IC,LOGIC,NC7SZ32,1.65V-5.5V,SINGLE OR GATE,NON-INV,SOT23-5,TAPE
1604581
VR2 VR3
TVS DIODE,SMA6J6.5A,6.5V,5%,UNIPOLAR,SMA,TAPE
3385450
W1
WIRE,PVC,UL1015,600V,105C,14AWG,STRANDED,BLU
2198323
W2
WIRE,PVC,UL1015,14AWG,STRANDED,INDIVIDUAL TINNED,BRN
2148907
XAF1 XAF2 XBF1 XBF2
FUSE CLIP,FUSE CLIP PCB
756460
Replaceable Parts Parts List
6
Assembly Top View (CKT 1)
Assembly Bottom View (CKT 4)
ESA612-4002
Figure 6-2. A1 Power PCA
gjc134.eps
6-11
ESA612 Service Manual
Table 6-3. A2 Meter PCA
Item
6-12
Description
Fluke Stock No.
C1 C73 C103 C116 C132 C157 C194
CAPACITOR SMR,CAP,CER,1000PF,+-10%,50V,X7R,0805
484378
C2
CAPACITOR,CERAMIC,0.33UF,+-10%,16V,X7R,0805,TAPE
1745562
C3-8 C10 C11 C14-21 C35-45 C47 C48 C63 C64 C69 C70 C72 C74 C75 C79-83 C93 C94 C98 C99 C105 C107-112 C114 C119 C123 C124 C126 C127 C129-131 C133 C135137 C141145 C155 C163-165 C167 C168 C170 C174 C177-179
CAPACITOR,CERAMIC,0.1UF,+-10%,50V,X7R,0805,TAPE
690500
C9
CAPACITOR,CERAMIC,4.7UF,+-10%,10V,X7R,0805,TAPE
3092853
C12
CAPACITOR,CERAMIC,2.2UF,+-10%,10V,X5R,0805,TAPE
3093102
C13
CAPACITOR SMR,CAP,CER,470PF,+-1%,50V,C0G,0805
929476
C22 C46 C50 C57 C58 C60 C61 C104 C117 C122 C146-150 C171
CAPACITOR,ELECTROLYTIC,TANTALUM,10UF,+-20%,16V,3216,TAPE
3093654
C24 C25
CAPACITOR SMR,CAP,CER,22PF,+-5%,200V,C0G,0805
643459
C26 C27 C29-34 C51 C62 C67 C106 C113 C115 C120 C121 C169 C189-191
CAPACITOR,CERAMIC,1UF,+-20%,25V,X5R,0805,TAPE
2530275
C28 C96 C97
CAPACITOR SMR,CAP,CER,4.7PF,+-0.25PF,50V,C0G,0805
806760
C49 C52 C118 C139
CAPACITOR,CERAMIC,0.01UF,+-10%,50V,X7R,0805,TAPE
106146
Replaceable Parts Parts List
6
Table 6-3. A2 Meter PCA (cont.)
Item
Description
Fluke Stock No.
C53-56 C134 C140 C151154 C156 CAPACITOR SMR,CAP,CER,100PF,+-1%,50V,C0G,0805 C159-162 C166
867650
C65
560380008223,CAP-SM,0805,50V,10%,X7R,.022UF
2332024
C68
560380008333,CAP-SM,.033UF,50V,10%,X7R,0805
2332036
C71 C76
CAPACITOR SMR,CAP,CER,4700PF,+-10%,50V,X7R,0805
604231
C77 C78 C87 C88 C90 C91
CAPACITOR,CERAMIC,0.47UF,+-5%,16V,X7R,0805,TAPE
3341246
C86 C92
CAPACITOR,CERAMIC,0.047UF,+-5%,50V,X7R,0805,TAPE
3341222
C89
CAPACITOR,CERAMIC,0.1UF,+-5%,50V,X7R,0805,TAPE
3349793
C95 C125
CAPACITOR,CERAMIC,15PF,+-5%,50V,C0G,0805,TAPE
514174
C101 C102
CAPACITOR,CERAMIC,33PF,+-5%,50V,C0G,0805,TAPE
603172
C175 C193
CAPACITOR,CERAMIC,0.027UF,+-10%,1000V,X7R,1812,TAPE
1579931
C180 C181 C187 C188
CAPACITOR,CERAMIC,100PF,+-5%,50V,C0G,0402,TAPE
2813217
C182-186
CAPACITOR,CERAMIC,0.1UF,+-20%,16V,X7R,0402,TAPE
2812858
CR1 CR8 CR15 CR17
DIODE,BAT54A SMR,DIODE,SI,SCHOTT,DUAL,30V,200MA,SOT-23
942594
CR2 CR5 CR10 CR11
DIODE,SI,BAS16,75V,200MA,6NS,SOT-23,TAPE
830489
CR3 CR4
DIODE,MBR0530 SMR,DIODE,SI,SCHOTTKY,30V,0.5A,SOD-123
691717
CR6 CR2029 CR33-37
TVS DIODE,SMAJ440CA,440V,5%,BIPOLAR,SMA,TAPE
2636264
CR9 CR41
DIODE,SI,PN,MMBD1503A,150V,200MA,DUAL,SERIES,LOW LEAKAGE,SOT-23,TAPE
928143
CR18
TVS DIODE,SMAJ5.0CA,5V,5%,BIPOLAR,SMA,TAPE
2073208
CR19 CR3032 CR38
VARISTOR,0603 3.6 V VARISTOR (CLAMP @ 10 V), VC060303A100
2764916
CR42-46
DIODE,SI,PN,BAV199,70V,215MA,3US,DUAL,SERIES,SOT-23,TAPE
605805
DS1
LED,RED,AM2520,280MCD,20MA,2.5V,20 DEG VIEW ANGLE,GULL WING PREP,T3/4,TAPE
2281407
F1
THERMISTOR,PTC,50,+-20%,300V,3A MAX,8MM RADIAL,BULK
3339126
H1
CABLE ACCESSORY ,CABLE ACCESS,TIE,5.50L,.10W,1.25 DIA
530360
J1
CONNECTOR,FLAT-FLEX,RECEPTACLE,SINGLE SIDE,0.5MM PITCH,VERTICAL SMD,20 POS,TAPE
2739499
J2 J21
CONNECTOR,FLAT FLEX,1MMCTR,VERTICAL SMD,ZIF,LOCKING,10 POS,TAPE
3408166
J3
CONNECTOR,MICRO-RIBBON, USB,SERIES B, RECEPTACLE,RT ANG PWB,4 POS,BULK
1541856
6-13
ESA612 Service Manual
Table 6-3. A2 Meter PCA (cont.)
Item
6-14
Description
Fluke Stock No.
J4
CONNECTOR,HEADER,2 ROW,0.1 CTR,VERT SMT,SHROUDED,20 POS,TAPE
2791599
J20
CONNECTOR,SOCKET,2 ROW,3.0MM CTR,VERT SMD,6 POS,TAPE
3343846
J60-64
CONNECTOR,HOLDER,FUSE,5MM DIA,RT ANG SURFACE MOUNT,TAPE
1620687
L1 L2
INDUCTOR,10UH,20%,1.8ADC,140MOHMS,6X6X3MM,SMT,TAPE
3338177
L3-10
INDUCTOR,BEAD,95 OHMS@100MHZ,1ADC,1MOHM,3612,TAPE
867734
LS1
AUDIO TRANSDUCER,MAGNETIC,SOUNDER,2.4KHZ,5V,TOP PORT,13MM SQ,TAPE
690271
P12 P13 P24 ESA612-8003,STRAP, INPUT
3258998
Q1
MOSFET,SI,N,2N7002,60V,115MA,7.5 OHMS,225MW,SOT-23,TAPE
927538
Q2-4
NPN,MMBT3904 SMR,TRANSISTOR,SI,NPN,60V,350MW,SOT-23
742676
Q7
MOSFET,SI,P,FDN338P,20V,1.6A,115 MOHMS,500MW,SOT-23,TAPE
2538867
Q8
TRANSISTOR,SI,NPN,MMUN2211,50V,100MA,DIGITAL,R1=10K,R2=10K,246 3331929 MW,SOT-23,TAPE
R2
RESISTOR,CERMET,130,+-1%,0.125W,100PPM,0805,TAPE
690644
R3 R5 R1115 R33 R4043 R50 R7072 R74 R82 R90 R91 R102 R119 R133 R135140 R146 R147 R168171 R194 R195 R243245 R249257 R260 R404 R405 R408 R412 R413
RESISTOR,CERMET,10K,+-1%,0.125W,100PPM,0805,TAPE
928791
R4 R21 R23 R36-39 R4548 R77 R95 R96 R118 R129 R131 R132 R134 R141 R148 R149 R172 R190-193 R196-199 R203 R215 R218 R225 R230-232 R236-239 R246-248 R270 R271 R345 R346
RESISTOR,CERMET,1K,+-1%,0.1W,100PPM,0805,TAPE
928713
Replaceable Parts Parts List
6
Table 6-3. A2 Meter PCA (cont.)
Item
Description
Fluke Stock No.
R7 R29-32 R73 R144 R150 R157167 R206214 R217 R233-235 R240 R407
RESISTOR,CERMET,100,+-1%,0.125W,100PPM,0805,TAPE
928937
R16 R20 R76 R80 R81 R86 R92 R117 R121 R142 R143 R145 R179 R180 R223
RESISTOR,CERMET,100K,+-1%,0.125W,100PPM,0805,TAPE
928866
R19
RESISTOR,CERMET,2M,+-1%,0.125W,100PPM,0805,TAPE
3326041
R22
RESISTOR,CERMET,24K,\+-1%,0.125W,100PPM,0805,TAPE
2499314
R24 R25 R27 R28
RESISTOR,CERMET,121K,+-1%,0.125W,100PPM,0805,TAPE
801560
R26 R53
RESISTOR,CERMET,1.5M,+-1%,0.125W,100PPM,0805,TAPE
3326090
R34 R35
RESISTOR SMR,RES,CERM,150,+-1%,0.1W,100PPM,0805
930086
R44
RESISTOR,CERMET,16.2K,+-1%,0.125W,100PPM,0805,TAPE
2005953
R49 R226229
RESISTOR,CERMET,3.01K,+-1%,0.125W,100PPM,0805,TAPE
1591278
R51
40034,RES,9.53K,1/10W,1%,0805
2499039
R52
RESISTOR SMR INAC,RES,CERM,13.7K,+-1%,0.1W,100PPM,0805
928812
R54
576186043120,RES-SM,0805,1/10W,1%,TF,604K
2338933
R55
RESISTOR,CERMET,732K,+-1%,0.125W,100PPM,0805,TAPE
1727162
R56
RESISTOR,CERMET,487K,+-1%,0.1W,100PPM,0805,TAPE
1665340
R57-61
RESISTOR,CERMET,499,+-1%,0.125W,100PPM,0805,TAPE
928978
R62 R104
RESISTOR,CERMET,30.1K,+-1%,0.125W,100PPM,0805,TAPE
2005994
R63 R182
RESISTOR SMR,RES,CERM,11K,+-1%,0.1W,100PPM,0805
928796
R64-66 R98100
RESISTOR,METAL FILM,330K,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE
3326205
R67
RESISTOR,CERMET,6.81K,+-1%,0.1W,100PPM,0805,TAPE
2087241
R68
RESISTOR,WIREWOUND,10K,+-1%,3W,20PPM,AXIAL,TAPE
2076480
R75
RESISTOR,CERMET,4.99K,+-1%,0.125W,100PPM,0805,TAPE
928767
R78
RESISTOR,WIREWOUND,800,+-1%,5W,20PPM,AXIAL,TAPE
2741174
R79
RESISTOR,WIREWOUND,200,+-1%,3W,20PPM,AXIAL,TAPE
2741195
R83-85
RESISTOR,METAL FILM,1M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE
3326231
R87-89 R127 RESISTOR,CERMET,150K,+-1%,0.125W,100PPM,0805,TAPE
1612759
6-15
ESA612 Service Manual
Table 6-3. A2 Meter PCA (cont.)
Item
6-16
Description
Fluke Stock No.
R93 R94 R101 R103
RESISTOR,CERMET,3.92K,+-1%,0.125W,100PPM,0805,TAPE
1591284
R97
RESISTOR,CERMET,8.06K,+-1%,0.125W,100PPM,0805,TAPE
928788
R105
RESISTOR,CERMET,360K,+-1%,0.125W,100PPM,0805,TAPE
3326268
R106
RESISTOR,CERMET,43.2K,+-1%,0.125W,100PPM,0805,TAPE
2499017
R107
RESISTOR,CERMET,432K,+-1%,0.125W,100PPM,0805,TAPE
1641918
R108 R110 R112 R122 R124 R154
RESISTOR,CERMET,2.43K,+-1%,0.125W,100PPM,0805,TAPE
928754
R109 R152
RESISTOR,CERMET,931,+-1%,0.125W,100PPM,0805,TAPE
3339167
R111 R113 R123 R153
RESISTOR,CERMET,1.1K,+-1%,0.125W,100PPM,0805,TAPE
691014
R114 R115 R125 R126
RESISTOR,CERMET,51.1K,+-1%,0.125W,100PPM,0805,TAPE
2011360
R116
RESISTOR,CERMET,4.7K,+-1%,0.125W,100PPM,0805,TAPE
3326333
R120
RESISTOR,CERMET,23.2K,+-1%,0.125W,100PPM,0805,TAPE
3331397
R128 R347349
RESISTOR,CERMET,10M,+-1%,0.125W,400PPM,0805,TAPE
943659
R156
RESISTOR,CERMET,100,+-1%,0.1W,100PPM,0603,TAPE
642196
R173-175 R409-411
RESISTOR,METAL FILM,3.32M,+-1%,0.25W,50PPM,CYLINDRICAL 0204,MELF,TAPE
3408749
R176
RESISTOR,CERMET,39.2K,+-1%,0.1W,100PPM,0805,TAPE
943092
R177 R178 R185 R186 R200 R201 R204 R205 R219 R242 R352
RESISTOR,CERMET,33,+-1%,0.125W,100PPM,0805,TAPE
2803420
R18 R151 R187-189 R220 R221
RESISTOR,CERMET,10,+-1%,0.125W,100PPM,0805,TAPE
928924
R181 R258
RESISTOR,CERMET,1M,\+-1%,0.125W,100PPM,0805,TAPE
928945
R183 R184
RESISTOR,CERAMIC COMPOSITION,27,+-10%,2W,-1300 +300PPM,AXIAL,TAPE
3454161
R216
RESISTOR,CERMET,51.1,+-1%,0.125W,100PPM,0805,TAPE
2005930
R350 R351
RESISTOR SMR,RES,CERM,249,+-1%,0.1W,100PPM,0805
606092
R353 R354 R362 R363 R366-368 R373 R374 R379 R388 R389 R392 R395-399
122421,RES - 102R, 1%, 1/16W, 0402, PB-FREE
3093309
R355-358 R401
RESISTOR,CERMET,JUMPER,0,+0.05 MAX,0.125W,0805,TAPE
928705
Replaceable Parts Parts List
6
Table 6-3. A2 Meter PCA (cont.)
Item
Description
Fluke Stock No.
R359-361 R364 R365
RESISTOR,CERMET,10K,+-1%,0.063W,100PPM,0402,TAPE
1706323
R370-372 R406
RESISTOR SMR,RES,CERM,261K,+-1%,.125W,100PPM,1206
877014
R375-378 R380-387 R390 R391 R393 R394 R402
RESISTOR,CERMET,499,+-1%,0.063W,100PPM,0402,TAPE
3254825
RT1
THERMISTOR,NTC,10K,+-5%,NON-LINEAR,0805,TAPE
2030909
RV1
VARISTOR,VARISTOR,430V,+-10%,1.0MA
105871
TP86 TP87 TP89-92 TP94 TP95 CONNECTOR,TERMINAL,TEST POINT,SMD,510 PH BRONZE,TAPE TP97 TP124127
602125
U1
IC,MEMORY,FLASH,45DB161,16MB,2MX8,2.73.6V,20MHZ,SERIAL,SPI,SO8,TAPE
3328276
U2
IC,MICROPROCESSOR,AT32UC3A,1.65-1.95V,32-BIT,66MZH,512K FLASH,64K SRAM,LQFP144,TAPE
3337006
U3
IC,LCD CONTROLLER,3.3V,32K SRAM,8-BIT,TQFP64
3328311
U4
OSCILLATOR,HCMOS CLOCK,40MHZ,3.3V,100PPM,15PF LOAD,TRISTATE,3.2X2.5MM SMT,TAPE
3328327
U5
IC,VOLTAGE REGULATOR,SWITCHING,1615-1,ADJ,100MA,STEP UP,SOT23-5,TAPE
1791133
U6
IC, OP AMP,LP324,+-3V TO +-32V,+-2MV OFFSET,100KHZ,QUAD,BIPOLAR,SOIC14,TAPE
2841280
U7
IC,COMM,FT232RL,3.3V-5.25V,USB TO RS232 UART,SSOP28,TAPE
2670615
U8 U42
IC,DIGITAL ISOLATOR,ADUM1402,QUAD,2/2,2MHZ,2500V,3V/5V,SOICW16,TAPE
2528074
U9
IC,COMM,SN75240,USB TRANSIENT SUPPRESSOR 15KV,7VBR,TSSOP8,TAPE
2044695
U10
IC,VOLTAGE REGULATOR,LINEAR,LP2980,3V,50MA,LDO,LO PWR,W/SHUT 3328412 DOWN,SOT-23-5,TAPE
U11
IC,VOLTAGE REGULATOR,SWITCHING,LT1616,ADJ,600MA,1.4MHZ,STEP DOWN,SOT-23-6,TAPE
3328447
U12 U43
OPTICAL,OPTOCOUPLER,PHOTO TRANSISTOR,SMD DIP4,TUBE
3328473
U13 U15 U54
IC,OP AMP,AD8510,+-4.5V TO +-18V,0.9MV OFFSET,8MHZ,PRECISION,JFET,S08,TAPE
2434633
U14
IC,DIGITAL,SUPERVISORY RESET IC,ACTIVE HIGH,SOT23-3,TAPE
3328499
U16 U21
IC,DAC,TLV5623,2.7V-5.5V,8-BIT,SPI,VOUT,SOIC8,TAPE
2761956
U17
589085015001,IC-SM,LMV331,SGL COMPTR,SOT235
2343218
U18
IC,COMPARATOR,LM393,+-1V TO +-18V,5MV OFFSET,DUAL,LOW PWR,SO8,TAPE
837211
6-17
ESA612 Service Manual
Table 6-3. A2 Meter PCA (cont.)
Item
6-18
Description SMR,IC,CMOS,QUAD INPUT NOR GATE,SOIC
Fluke Stock No.
U19
CMOS 74HC02
U22
IC,OP AMP,TLV2372,2.7V TO 16V,4.5MV OFFSET,3MHZ,DUAL,R/R,S/S,SO8,TAPE
3329934
U23
IC,RMS CONVERTER,LTC1967,5V,LOW POWER,R/R INPUT,MS0P8,TAPE
3385912
U24
IC,ADC,AD7799,2.7V-5.25V,24-BIT,3-CH,500HZ,SIGMADELTA,SERIAL,TSSOP16,TAPE
2630817
U25 U27 U39
IC,ANALOG SWITCH,ADG431,5-44V,24 OHMS,QUAD,SPST,NO,SO16,TAPE
3367437
U26 U28
IC,OP AMP,AD8034,5V TO 24V,2MV OFFSET,80 MHZ,DUAL,FET INPUT,SO8,TAPE
2558152
U29 U31
IC,OP AMP,AD8539,2.7 TO 5.5V,15UV OFFSET,430KHZ,R/R,S/S,DUAL,S08,TAPE
3385920
U30
CONTRACT MFG ITEM, IC 2.5V REF LM4040CIM3-2.5
2073591
U32 U34 U37 U38 U55
IC,LOGIC,74HC595,2.0V-6.0V,8-BIT SHIFT REGISTER,TRISTATE,SOIC16,TAPE
904388
U33 U35 U36 U56
TRANSISTOR,SI,NPN,ULN2003,50V,500MA,HEPTADARLINGTON,3.5W,SO16,TAPE
2113871
U40
IC,MICROCONTROLLER,LPC2138,3.3V,16/32 BIT,512KB FLASH,ARM7,25MHZ,LQFP64,TAPE
3329792
U41
IC,VOLTAGE MONITOR,MAX803,VTH=2.93V,MICROPOCESSOR RESET CIRCUIT,OPEN-DRAIN OUTPUT,SOT23-3,TAPE
3350940
U44
IC,MEMORY,EEPROM,93LC66A,4KB,256X16,2.55.5V,2MHZ,SERIAL,SO8,TAPE
3344058
U45 U46
CMOS 74HC00
830703
U47
CMOS 74HC132 SMR,IC,CMOS,QUAD 2 IN NAND W/SCHMT,SOIC
837245
U48
POWER SUPPLY,DC-DC,2W,12VDC,12V@167MA,3KV ISO,SIP4,TUBE
3338731
U49
POWER SUPPLY,DC-DC,2W,12VDC,+-12V@83MA,3KV ISO,SIP5,TUBE
3338746
U50
IC,VOLTAGE REGULATOR,LINEAR,LP2985,3.3V,150MA,LDO,LOW POWER,SOT-23-5,TAPE
2559443
U51
POWER SUPPLY,DC-DC,2W,12VDC,5V@400MA,3KV ISO,SIP4,TUBE
3338754
U52
589480084001,IC-SM,MC79L05A,-5V REG,8-SOIC
2343646
U53
BIPOLAR 78L05 SMR,IC,VOLT REG,FIXED,5 VOLTS,0.1 A,SOIC8
604105
U59
IC,MEMORY,FRAM,FM25L256B,256KB,32KX8,2.7-3.6V,SERIAL,SO8,TAPE
3374127
U60
IC,VOLTAGE CONVERTER,LTC1144,SWITCH CAP,WIDE INPUT RANGE,SO8,TAPE
2609823
VR3 VR4
TVS DIODE,SMBJ15A,15V,5%,UNIPOLAR,SMB,TAPE
2813982
VR6 VR7
ZENER,UNCOMP,MMBZ5236B,7.5V,5%,20MA,225MW,SOT-23,TAPE
837138
VR12-14 VR16
ZENER,UNCOMP,MMBZ5242B,12V,5%,20MA,225MW,SOT-23,TAPE
2044054
W2
WIRE,PVC,UL1015,600V/105C,14AWG,STRAND,INDIVIDUAL TIN,GRN/YEL
2279979
Y1 Y2
CRYSTAL,12MHZ,50/100PPM,20PF,PLASTIC ENCAPSULATED,SMD,TAPE
2041235
SMR,IC,CMOS,QUAD INPUT NAND GATE,SOIC
830711
Replaceable Parts Parts List
6
ESA612-4001 Figure 6-3. A2 Meter PCA
gjc133.eps
6-19
ESA612 Service Manual
6-20
Chapter 7
Schematics
Title
Page
Introduction.......................................................................................................... 7-3
7-1
ESA612 Service Manual
7-2
Schematics Introduction
7
Introduction The schematic and pca layout diagrams for the ESA612 are shown in Figures 7-1 through 7-3.
7-3
ESA612 Service Manual
7-4
Electrical Safety Analyzer Schematic Diagrams
Assembly Top View (CKT 1)
6
Assembly Bottom View (CKT 4)
ESA612-4002
Figure 7-1. A1 Power PCA
7-5
ESA612 Service Manual
RV2
706838 1 2 1MA
XAF2
XBF2
12VDC
XAF1
8
2152365
E1
GND
2.5A 250V
CHGND
3C6>
CRL1+2
VSS
7 2 5 4
1
3331539
CRL3
K3
3 6 12VDC
1
IN
IN
3C6>
K2
430V 2
+12V
2
F3
8
2152365
12VDC
1
8
LOAD+
LOAD3326205
R2
1% 330K 3326205
9
3
+12V CRL8
IN
3326205
3326205
MELF0204
6.2K
6.2K
5
X7R
R18
6.2K
3336652
C9
6
0.047U
+ -
1708909
U4
X7R
V4
GND 3A6>
7
SO8
3
R17
C8
0.047U
1708909
2
+ -
SO8
V4
IN
GND
12VDC
1
DIP8
2152365
IN
8
12
+12V CRL12
TP19
10
9
8 3
4
1 TX2SA 2082903
IN
1
CLK 7 CS 1
ADC_DOUT ADC_DCK
OUT
ADC_/CS
IN
IN
V+
5
SOT23-5
+12V
K10
MAP_RTN 500V
CRL10
IN
2C1>
R12
3C2> 4D2>
GNDMS
10.0
+12VM
3
1U
R128
100K
2
+
GND
2
GND
C6
C19
0.1U
-
7 V+
U1
V4
AD8510AR
GNDMS
LMV7271
3332950 C10
0.0047U
R19
100K
928866
OUT
1
1C4>
+12V
+3.3VMD
VDD
VL
SO16
0.1U
3
GNDMS
1
/EQU_CUR_ASW
IN
VSS GND
4
5
-12VM
GNDMS
PWR GRP TP29
GNDMS
C7
0.1U
-12VM
690500 X7R
13
R9
GNDMS
100
10P
3B6>
2
C21
690500
2434633
C27
3B6>
690500
GNDMS
6
SO8
928866
3A6
IN
3D1>
OUT
3B2>
5
GFI_SENSE
SOLID STATE RELAY SPACERS UNNAMED_1_RLY1C1COIL_I12_W1
U7 OUT
8
13 12
U8 ADG431B 3367437
0.1U
UNNAMED_1_RLY1C1COIL_I12_NO2
MAINS_ZC
METER
10
0.1U
OUT
SO8
4
TP38
+12VM
C20
GNDMS 6
3 5
12
METER CIRCUITS
TP20
5
+12V
K14
CRL9
3C6>
CRL14
IN
1
12VDC
12VDC
+12V
3D8>
3B2> 4A3
IN
TP21
12VDC
8
4
5
12
+12V
IN
9
K10
OUT
TP17
400MA_RTN
10
9
8
3 5
9 METER
+3.3VPD
928791
K12 4
3 6
3D6>
MAINS ZERO CROSSINGS
3
CRL11
GNDMS
GND
928791
R32
IN
CHGND
+12V
C5
DOUT
R37
1
3D6>
928937
3
4
10.0K
3329934
100
7 2 5 4
X7R
VSS
10.0K
GND
3C6>
3331459
TLV2372
BAS16
TERM-BLOCK
3342141
K11
2
R16
830489
CR18 SOT23
P
5304
690500
3 CH1
1
BAS16
1
830489
8 U3 VDD/VREF
2 CH0
U4
CR17 SOT23
1
GND
MCP3202
3329934
GND
8 V+
DIP8
1
200:1 3
1
332
5 DIN
ADC_DIN
TLV2372
+3.3VPD
GND
4
8
0.1U
GND 8 V+
12VDC
1
D
O
R10
690500
0.1U
+3.3VPD
3 6
WP4 W2 L3 2750107
L1
GNDMS
C2
C17
330K 1%
CRL6+7
BROWN
+3.3VPD
+3.3VPD
R22
6.2K
AP_METER
IN
K9
TP18
12VDC
TX2SA 2082903
MAINS VOLTAGE A TO D
R5
K7
2152365
1
3326205
R21
3336652
5
12
330K 1%
330K 1%
10 8 3
4
R7
X7R
+3.3VPD
K8
3C2>
R6
IN
+12V
J1
3 EO_N 2 EO_PE 1 EO_L
14AWG
SH1 SH2
MAINS CIRCUITS
SSR_DRV
X7R
1% 330K
MELF0204
R3
MELF0204
3326205
MELF0204
1% 330K
+12V
NOLOAD
E2 R4
3D6>
D
EQUIPMENT OUTLET
WP2
4
MECCA
L2_IN
OUT
DIP8
2152365
3C6>
MELF0204
4A7< 2C7
2152365
12VDC
2152365
TP16
DIP8
+12V
RV1
3 6
1
CRL4
IN
3C6>
7 2 5 4
3 6
DIP8
1
10% 1MA
DIP8
XBF1
K4
7 2 5 4
3 6
1
F1
HEADER
3351138
K1
7 2 5 4
X7R
CHGND EOGS PE N L
430V 10%
F2
1 2 3 4 5
X7R
J5
X7R
MAINS INLET
3B3< 13
11 9 4
5
R14
H5 H6
1.00K
R112
604231 X7R
13
12
7
U8 ADG431B 3367437
6 8 4
5
PAD-375C-WSP PAD-375C-WSP PAD-300C-WSP PAD-300C-WSP
252-HOLE 252-HOLE 215-HOLE 215-HOLE
3350760 3350760 3352172 3352172
COMMENTS COMMENTS COMMENTS COMMENTS
POWER INLET CONN SPACERS
12
10
U8 ADG431B 3367437
H1 H2 H3 H4
PAD-325C-WSP PAD-325C-WSP
215-HOLE 215-HOLE
3376173 3376173
NOLOAD=TRUE NOLOAD=TRUE
1.00K
R113
ESA612-1002 (1 of 4)
1.00K
GNDMS
Figure 7-1. A1 Power PCA (cont.)
7-6
3D1>
Electrical Safety Analyzer Schematic Diagrams
6
MAP CURRENT LIMITS CR12
CR13
S1M 1KV 1A
S1M 1KV 1A
SMA
T1
4A7< 1C8>
IN
L2_IN
5
S1 O
O
102V
115V
P2
S2 O
O
R24
4
3326205
1
NOLOAD
7 9
C16
1000P 20% 2083392 300V Y5P
R27
10P
Q2
3 SOT23
1
2 3 1
TP22 TP23
3331432
820902
S
2742520 3
3326205
10
MMBT5089
K16
900V D2PAK
330K 1%
C74
3342134
Q1
IRFBF20S
C0G
4
P1
2646221
3326205
6 MELF0204
2
102V
115V
330K 1%
SOT23
1
T394
IN
L1_IN_F
THRU-HOLE
4A7< 1C8>
B0710099
MELF0204
MAP TRANSFORMER
R23
330K 1%
MELF0204
2646221
SMA
R25 1.00K
4
CR1
6.8V 5% 5235 837195
R26
7.5K
2278492
930172
3
U10
DIP4S
HCPL-817
3328473
+12V
1 2
R31 1.00K
4
MAP_3MA
IN
3C6>
3
9
3 5
4
12
8
1
IN
OUT
1C4>
5
CRL16 12
1
12VDC
K16
+12V
MAP_RTN
3
4 3C6>
10
+12V
GNDMS
+12V
U29
1
2 R104
DIP4S
1.00K
HCPL-817
3328473
CR14 SMA
CR15
S1M 1KV 1A
S1M 1KV 1A
2646221
10 8
TX2SA 2082903
1.82K
3.01K
1591278
K17
12VDC
R29
R28
928713
CRL17
IN
3C6>
9
Q8
2002486
4.7K 1
INF
SMA
2646221
MUN2216
3
2
MAP_7.5MA
IN
3C7>
SC59
GND
ESA612-1002 (2 of 4)
Figure 7-1. A1 Power PCA (cont.)
7-7
ESA612 Service Manual
CR19 SOT23
CR20 SOT23
MMBZ5242B 2044054 12V 5%
75V 200MA
CRL14
Q9
3
+12V
BAS16
+3.3VMD
830489
R100
4.7K 1 INF
33 16 VCC
SC59
MUN2216
2002486
R107
GND
928713
C57
R105
3B7
R110 C53
16V 10U 3093654
1.00K
3
IN
OUT
LP2985
3.3V
SOT23-5
ON
BYP GND 2
5 4
GNDMS
GNDMD
GNDMD
Figure 7-1. A1 Power PCA (cont.)
7-8
RLY_SDI_M RLY_SCK_M RLY_RCK_M RLY_SDO_M
ADC_DCK_M ADC_CS_M ADC_DIN_M 400MA_FREQ RELAY_EN_M MAINS_ZC_M ADC_DOUT_M GFI_SENSE
PWR_GRP
GNDMA
2791599
+3.3VMD
C54
0.01U
106146
C55
10U 16V 3093654
E4 MECCA
C60
10U 16V 3093654
500V_RTN
+12VM -12VM GNDM GNDM
E12
C46
100P
400MA_RTN
+3.3VMD U28
2559443
928713
867650
HEADER
GNDMD
+12VM 1
928713
GNDMD
C4
ADC_DCK_M
928713
VI C 5
ADUM1401
GNDMD
GNDMD
3343846
VSS
400MA_RTN 500V_RTN
OUT OUT
928713
VI A 3
GND 10U 16V 3093654
867650
VDD1 1
U5
3331444
C68
C47
100P
200
3372234
16 VDD2
1.00K
TANT 3216
+12V
OUT
ADC_/CS
R95
3372234
C0G
R94
200
C0G
ADC_DCK
C67
100P
867650
928713
1.00K
MAINS_ZC
E13
R33
2
DIP4S
GND
C0G
OUT
C66
100P
867650
TANT 3216
1B5
3328473
GND
R39
AP_METER
OUT
J2
HEADER
1 400MA 2 METER 3 +12V 4 AP_METER 5 0V 6 GND
GNDMS
GND
C12
U11
HCPL-817
X7R
4
4 3
928713
C0G
RLY_SCK_G
RLY_RCLK
GNDMD
C0G
OUT
OUT
METER
TP35
METER
OUT
+3.3VMD
R30
1.00K
HCPL-817
3C6>
C0G
3D8
R15
C0G
+3.3VPD
C23
3D8< 3B7
C28
GNDMD
U26 1
5
4
GNDMD
GND
X7R
X7R
/RELAY_EN_M
TP34
0.1U
C43
690500
4C7
SC59
2
GNDMD
R55
10.0K
R50
928791
332K
-5VCM
3
4
2N7002E
R46
1 2
R69
HV_MEN_500V
1.00K
IN
3D1>
IN
3D1>
928713
11.5K
2
S
3328473
GNDMD
3331397
3
60V 240MA
1756473
DIP4S
23.2K
1997330
Q5
U20
HCPL-817
R66
Q6 3
2N7002E
1
60V 240MA
1756473
R68
R86
2
S
4
U21
DIP4S
1 2
R70
HV_MEN_250V
1.00K
928713
HCPL-817
R67
3328473
100K
928866
-5VCM
3
1
100K
2005982
2005982
-
3
1 400PPM
1 400PPM
+
-5VCM
R47
2005982
R59
V2
3
484378
R85
MUN2216
3
928791
X7R 50V 10%
1 400PPM
3B2>
+12VM
0.001U
-5VCM
2005982
1
CR6
BAS16
C33
SOT23-5
3331500
830489
928945
1 400PPM
-5VCM
R48
1.00M
R76
1
2
929476
OUT
10.0K
5 V+
TLV2371
3326333
C31
484378
1M 1%
3326231
4.7K
SO8
C30
R57
4
-5VCM
470
X7R
X7R
R121
484378
0.047U
1708909
S
C73
S
R44
Q4
4
5 V+
-5VCM
3331415
3
S1M 1KV 1A
D2PAK
6.81K
3
X7R
10.0K
928713
X7R
R77
1579273
2646221
1.00K
R61
IRLR3410
100V 17A .1 OHMS 3W
56K
MELF0204
1.1K 20%
400MA
+5VCM
R62
RV75
THERM-THRU
+t
7
-5VCM
SOT23
GNDMD
TP36
Q3 4
500V_RTN
R49
U24
SOT23
HCPL-817
3
CR16
BAS16
C35
0.1U 50V 690500
SOT23
DIP4S
400MA
R45
GNDCM
C0G
2
4
X7R
1.00K
928713
1
690500
DPAK
R65
C32
0.1U
+5VCM
3
830489
0.22U
3326231
1277360
Q7
5% 12V MMBZ5242B 2044054
10.0K
690500
1C4
3 5
SMA
0.0
-5VCM
U19
9
1
MELF0204
1812
C42
0.047U
10%
10 8
SOT23
47U 16V
.2OHM ESR
3352434
X7R
7343
GNDCM
0.0
6
Z
500V X7R 644077
R123
2
O
C41
0.1U
2087962
0.047U
RV40
THERM-THRU
K13
TP26
MELF0204
O
X
8
C72
Y
C29
1812
4
US1M 1KV 1A
X7R
O
68K 5%
3331385
GNDMS
SMA
0.0
1
2512 1W
68K 5%
CR5
R122
T2
2512 1W
3331385
SOT23
INSULATION INVERTER
TP37
R63
928866
GNDMD
-5VCM
+5VCM E16
+12V
MAINS POWER SUPPLY 85-264VAC,15W,12VDC
2C7< 1C8> 2C7< 1C8>
IN IN
L2_IN
2
L1_IN_F
1
U9
ACL +VO 32X64MM ACN -VO
ECL15US12
3331480
3 4
R120
0.0
R38
3.01K
0.0
+12V
20%
C40
RED
VSS
0.0
VOUT+
2 VIN-
VOUT-
SIP7-5P
GND
D215EI 3338710
5
+5VCM
C39
VR2
1U 25V 20% 2530275
7
COM 6
CR4
AM2520 2281407 LED-SMD
VIN+
25V 1U 2530275
1591278
VSS
U18 1
R119
R125
GNDCM
TP12
6.5V
E5
R126 0.0
R127
C38
MECCA
VR3
1U 25V 20%
2530275
6.5V
GNDCM -5VCM
0.0
400MA_RTN
IN
3B2> 1D1
METER
3
IN
METER INPUTS
METER JACKS RED 1
2
3
4
3
TP168
L4
BLACK 1
2
P13
3
CR6
2
1
440V
CR36
10
440V
L10
4
95_OHM
1
NULL 1
2
CR34
2
440V
P24
3
1
CR37
10
TP100
2
4
5
R85
1M
R65
R66
330K
330K
330K
3
TP107
5 8
6.81K
1
2
1
440V NULL WIRE FROM POWER PCA GREEN
C175
CR35
2
TP105
440V
TP169
R370
261K
3 10
R184
2
27
F1 C
3A
0.027U
C135 0.1U
CHGND WIRE FROM POWER PCA GREEN/YELLOW
GHGND WIRE TO PAGE 4 GREEN/YELLOW
4B7> 400MA
E27
261K
5 8
W2 D
14AWG
4
3 5
9
10 8
E28
K18
2 4 9
GNDMA
R88
IN
IN
1 2
TP101 TP102
10
4
5 8
TP103
3
TP104
10
5 8
C139
0.01U
6
R86
100K
K21 4
GND
R89
GNDMA GNDMA
V+
8
4
150K
9
-5VMA
VDD
C145 0.1U
GNDMA
VL
R182
3
-12VMA
PWR_GRP
4A6>
C138
VSS GND
4
IN
11.0K
1
0.015U
5
GNDMA
GNDMA
IN
GNDMA
PWR_GRP_SEL
3B6>
-12VMA
1
OUT
/FAULT_SENSE
3B8
3
3
8
3C6>
R71
C165
+5VMA
150K
TP106
+12VMA
ADG431B U27
+3.3VMA
VR7
7.5V
AAMI_LOAD
13 12
7.5V
+5VMA
+5VMA
R87
R406
IN
5
2
VR6
CR9
FAULT SENSE
27
R372
0.027U
16 4
10.0K
R183
3C6>
12
+12VMA
MMBD1503 1
IEC_LOAD
IN
5
14
U27 ADG431B
GNDMA
0.0047U
261K
C193
9 4
15
C71
R371
261K
11
13
9
GNDMA
9
8
12
100
4
4
5
2C8
5
R70
K16
NO_LOAD
OUT
GNDMA
R67
METER_AMP_IP
OUT
R156
K20
95_OHM
CR33
4
U27 ADG431B
200
R64
8
9
8
440V
L3
4
3
TP99
6
U27 ADG431B
10
R79
10
K15
C100
0.015U
12
7
13
1M
9
8
2
R84
10K
9
8
800
4
5
95_OHM
1
K13
10
R78
1M
13
R68
4
5
R83
P12
K25
6
3
TP108
U18 7
LM393DT
R100
R99
TP109 R98
330K
330K
330K
MAP NULL
K24 4
5 10
+5VMA
R219
9
8
0.1U
-5VMA
GNDMA
-5VMA
R97
R103
8.06K
3.92K
C74
0.1U
GNDMA
C75
0.1U
-5VMA
U22 7
TLV2372
8 V+ V4
-5VMA
+
R94
6
3.92K
5
R96
1.00K
GNDMA
7
R93 3.92K
+5VMA
+5VMA
9
33
C141
U22 1
TLV2372
4
+5VMA 8 V+ V4
-5VMA
3 +
2 3
R101 3.92K
R95
1.00K
8
R102
VDD OUT FS
U21
VREF DIN
TLV5623
CS-
SCLK
6 1 2
10.0K
R117
100K
IN IN
ADC16_SDO ADC16_SCK
5C6> 2B2< 3A8< 5C6> 2B2< 3A8
5C6> 3A8
METER_AMP_IP
3
IN
NC
+
NC
5
U54
V4
NC
U39
13
6
11
AD8510AR
4
3
R111
+12VMA 1.10K +3.3VMA
5
ADG431B
-12VMA
2
2.43K
9
8
2A2< 5C6> CONT_GATE
R108
8 V+
+
2
VDD
U25 ADG431B
-12VMA
R154
1
6
2.43K
5
R153
V4
+
1.10K
8 V+
U28
R110
7
14
1
U25 ADG431B
16
U25 ADG431B
5
1
GNDMA 13
9 5
+
V4
TP87
U26
8 V+
U23
+5VMA
+5VMA
1.10K
AD8034
-5VMA
C77
R129
0.47U
1.00K
7
AD8034
-5VMA
R127
LTC1967
VDD
2
IN1
RTN 6
IN2
8
3
150K
GND 1
6
U25 ADG431B
8 4
IN
E118
E225
MECCA
MECCA
7
VDD
U39 ADG431B
-12VMA
VL
6
0.47U
C143
VSS GND
13
0.1U
5
12
15
14 16
GNDMA
GNDMA
4
R114
2
51.1K
3
+
R126
5
ADG431B
2
1
2
-12VMA
3
NC
+
NC
5
GNDMA
V4
NC
-12VMA
R105 360K
C89
IN
0.47U
+5VMA
6 5
+
8 V+
U29
V4
AD8539
0.1U
5D5< 2C2< 2B6< 2A2> 2V5REF
6
8
5
ADG431B
7
R106
R107
43.2K
432K
C86
6 5
+
8 V+ V4
3
U46 74HC00
AD8510AR
2
74HC00 6
U46
U46 74HC00
AD8539
4.7K
11
4 +3.3VMD
1
2
U46 74HC00
12 13
U30 LM4040 2.5V
CR19
2
3.3V
+3.3VMD
14 DVDD
U24
R119
DOUT/RDY
AD7799
3 AIN3+/P1
R116
GNDMA
10
1
8 AIN2-
+5VMA
GNDMA
9
5
0.1U
2
7 AIN2+
7
0.047U
GNDMA
CR18
6 AIN1-
GNDMA
8
2B5< 2B6< 5D5
GNDMA
249
10 REFIN-
DC INPUT
TP91
+5VMA
0.1U
VCC=+3.3VMD;GND=GNDMD
R351
249
13 AVDD 9 REFIN+
2.5V REF
3
4
U15
7 V+
R104
0.1U
5V
TP90
C87
C88 0.47U
30.1K
C82
0.1U
R350
1
HV MEAS
1
+12VMA
0.1U
GNDMA
1
1
+3.3VMD
C81
0.1U
12
10.0K
0.1U
AD8539
AD8539
GNDMA
C177
GNDMA
U39
13
R168
C72
V4
GNDMA
10.0K
C70
U29
C92
U31
V4
0.047U
+5VMA
TP89
51.1K
R171
+12VMA
8 V+
+
8 V+
GNDMA
AC INPUT
51.1K
+5VMA
U39
-
GNDMA
4
TP95
GNDMA
3
1.00K
NC
GNDMA
0.1U
R115
+3.3VMA
8 4
R125
51.1K
13 12
100K
R118
GNDMS
GNDMD
C80 +3.3VMA
23.2K
2
METER ADC
+5VMA
+12VMA
0.47U
R121
IN
0.1U
METER DC ONLY
C78
C90
R120
GNDMA
GNDMA
IN
-EN
0.47U
10.0M
5
C99
5
IN
5D5< 2C2< 2B5< 2A2> 2V5REF
VOUT
C91
R128
7
12
7
GNDMA 3C6> RMS_GAIN_X1P5
5
R123
12
4
6
2.43K
V4
11
3C6> RMS_GAIN_X10
0.1U
GNDMA
+
GNDMA 13
C142
-
R124
+5VMA
GNDMA
R349
10.0M
2.43K
U26
0.1U
10.0M
R122
8 V+
C83
R348
4.7P
+5VMA
1.10K
-5VMA
12
4
3
R113
3
+3.3VMA
2
2.43K
V4
AD8034
GNDMA
-5VMA
-5VMA
10.0M
GNDMA
C97
+5VMA
R347
0.1U
R112
10
5
C98
0.1U
GNDMA
2.43K
+5VMA
15
VSS GND
4
U28
13
VL
4.7P
931
AD8034
GNDMA
13 12
IN
15P
+5VMA
12
10
C96
R152
931
7 V+
C125
+5VMA
+5VMA
0.1U
-5VMA
15P
-12VMA
C126
0.1U
GNDMA
C95
+5VMA
4 AIN3-/P2 GND
15
DIN 16
SCLK 1 CS 2 PSW 11
10.0K
IN IN
ADC16_SDI
5C6
1B1< 3A8
1B1< 3A8
12
GNDMA
IN OUT
HV_MEAS 4B6> 2V5REF 2B5
IN
CONT_GATE
5C6> 2C7
ADC16_SDO
IN
2B2< 1B1< 5C6> ADC16_SCK
IN
1 2
+5VMD
VDD
DIN
10.0K
1A1
GFI_SENSE
IN
IN
FS
3
V+
5
R77
U17
1.00K
4
GND
2
LMV331
GNDMD
Q9
R401
SI2312B
0
S
VOUT-
2
GNDMD
0.1U
GNDMD 2
CR8
3
C68
0.033U
BAT54A
R76
100K
10U
+12VM
E18
GNDMA
5
C120
COM 6
GND GND
E20
3
1U
+5VMA
5
NC
10.0
-12VMA
C148
C123 0.1U
6
10U
R220 10.0
GNDMA
NOLOAD
6
10U
E11
GNDMA
1
E12
C67 1U
1
3
NC NC GND
1 8
C124
2
10.0
5
10U
U47
100 10
8
R235 C132
0.001U
74HC132
GNDMD
12 13
R407
U47 74HC132
100
U47
4 5
6
GNDMD
1 2
U47
GNDMD
100K
VCC=+3.3VMD;GND=GNDMD
2
74HC02
U19
5
1
6
+3.3VMD
C155
74HC132
0.1U
VCC=+3.3V_U47;GND=GNDMD
11
+3.3VMD
3
VCC=+3.3V_U47;GND=GNDMD
3
GNDMD
74HC02
U19
4
8 9
IN
74HC02
U19
11
10
12
GNDMD
RELAY_DISABLE
5C2>
74HC02
U19
13
R170
OUT IN
/RELAY_EN 4A6> RELAY_ENABLE
3D8< 4D8< 5C1>
10.0K
GNDMD
OUT
/FAULT 5C6
/CLR_FAULT
100
3
CS-
R75
IN
+5VMD
R72
1
RST MCP101T GND 3
3
R172
100K
DAC_FS 5C6>
4
R73
GND 5
R80
5C6> /GFI_DAC_CS
7
OUT
TLV5623
SCLK
R195
BAS16
2 VCC
R74
10.0K
CR10
+5VMD
+3.3VMD
+12VMA
C146
MC79L05A
10.0K
+5VMD
GNDMD
U16
CR40
1.00K
4
2
U14
8
VREF
C121
IN
R91
0.1U
6
8
7
5C1
GFI DAC
33
10U
R151
+3.3VMD
+5VMD
+12VMA
10U
C122
0.01U
VSS
R90
10.0K
VIN+
2 VIN-
NOLOAD
R403
IN
1C1
ECG_SDO
+12VMRLY
33
1.00K
IN
4D8< 5C2> METER_RCK
R186
R192
IN
4D8< 5C2> 4A6> RELAY_SCK
U48
D213EI
+12V
6
/GFI_TRIP
5C6
4A6> RELAY_SCK
3D8< 5C2> METER_RCK
5C1> 4A6> RELAY_SDO
3D8< 4A6> 3B1> /RELAY_EN
33
1.00K 1.00K
11 SHCP
R196
IN
12 STCP
1.00K
14 DS
R198
IN
ULN2003AD
U32
1.00K
74HC595
C156
13 OE 10 MR
100P
C134
15 1 2 3 4 5 6 7
Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7
TP12
GND
8
GND 8
R237
16 15 14 13 12 11 10
2 3 4 5 6 7
R239
GNDMD
1.00K
1.00K
100P
12V
9
TP10 1
9
Q7S
+12VMRLY
VR12
U33
16 VCC
R199
IN
+12VMRLY
R178
R197
IN
GNDMD
C151 100P
+12VMRLY
C140
GNDMD
11 SHCP
ULN2003AD
U34
12 STCP 14 DS
C170 0.1U
GNDMD
74HC595
13 OE 10 MR
R238
VR13
U35
16 VCC
100P
15 1 2 3 4 5 6 7
Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7
TP11 1
16 15 14 13 12 11 10
2 3 4 5 6 7
TP13
GND
9
Q7S
12V
9
8
GND 8
GNDMD
1.00K 3D8< ECG_SDO
12
1
K11
3
K10
10
12
1
12
K8 1
TP172
3
12
K6 1
TP173
10
12
K4 1
12
K2 1
3 5 10
K22
12
1
12
1 K23
GNDMA
1
12
1 K9
12
1 K7
12
1 K5
12
1 K3
12
1 K1
8
E149 MECCA
ECG WAVEFORM AMP
E177
R176
MECCA
39.2K
OUT
9 3
TP174
K23
1B8
4D8
3D8< 4D8
MAINS_SDI
5C2
OUT
/ADC12_CS
5C1>
ADC12_SDO
5C6>
OUT
400MA_FREQ
OUT OUT OUT OUT
/RELAY_EN
OUT
ADC12_SDI
5C6
+12V
VR4
15V
E7
GHGND WIRE TO PAGE 1 GREEN/YELLOW
U60
GNDMA
1U
GNDPD
LTC1144
8
V+
TP166
1
BOOST
TP167
7
+12V
C190
VOUT 5
OSC -SD 6
5C1> 2A2
METER_RESET
R141
IN
U43
1
1.00K
2
OUT IN
OUT OUT
5A4> UI_RXD 3A1> /GFI_TRIP
3A1> /FAULT
OUT
IN
OUT
33P
61 XTAL2
R374
102
R375 R376 R377 R378
499 499 499 499
102
2B2< 1B1< ADC16_SCK 3A8< 2B2< 1B1< ADC16_SDO 3A8
METER_TXD
IN
6C6> /METER_ISP
IN
GNDPD
10.0K
R206
R140
10.0K
100
100
11 VI D
R208 100
C159 100P
U42
ADUM1402
R209 100
16 VDD2 14 VO A 13 VO B 12 VI C
R207
C160
TP119 TP120 TP121
R382 R383
499 499
R384 R385
499 499
R386 R387
499 499
OUT
TP118
OUT
TP124 TP122
OUT OUT
OUT
TP123
IN
U41
RST MAX803 GND 1
RELAY_SCK 4A6> 3D8
ADC12_SDI
IN
12MHZ
C102
0
C112
VDD
R203
Y2
GNDMD
GNDMD
0
TP22
VDD
33P
0
C111
VSS VSS VSS VSS VSS
1.00K
GNDMA
0
C110
R358 0.1U
GNDMD
VREF
R148
3
C101
8
+3.3VPD
10U
VDD
1 2 3 4
+3.3VPD
TP19
100K
C104
2
R357
VSSA
HEADER
1
R356
59
R145
J22
100
CR17
BAT54A
2
49
NOLOAD
R355
1U
63
DEBUG RS232
R144
IN
2C2< 2B6< 2B5< 2A2> 2V5REF
CR30 3.3V
C115
VBAT
+3.3VMD
1
6 18 25 42 50
METER CIRCUITS
MAINS CIRCUITS
+3.3VMD
METER PROCESSOR
6
10 VE2 9 15 GND2 GND2
R137
VDD1 1
C164 0.1U
R147
10.0K
10.0K
VI A 3 VI B 4 VO C 5
R210
100
R211
100
R212
100
GNDMD
VE1 7 GND1 2 GND1 8
R138
10.0K
100P
GNDMD
C161 100P
C162 100 100P
1
UI_TXD 5C6>
OUT
UI_RXD 5C6
OUT
/ISP 5C6
METER_ATTN
GNDPD
1
7D6>
KEYCOL_0
7D6>
KEYCOL_1
7D6>
KEYCOL_2
7D6>
KEYCOL_3
7D6>
GNDPA
TP131 TP132
5A7< /METER_ISP
OUT
IN
5C8< METER_RESET
OUT
OUT
R388
102
TP133 TP134 100P TP135 TP136 TP137 GNDPD TP138
C188
R5 10.0K
OUT
TP129 TP130
499 499 102 499 499
IN IN IN
IN IN
KEYCOL4
+3.3VPD
DS1
RED
AM2520
R2
Q1
0.1U
3C4< MPSO 7C3
7D3< /DISP_CLK_OFF
1.00K SI
IN
FM25L256
+3.3VPD
ISP
SO
R346
U59
USB_RXD
8C7> /USB_PRESENT
OUT
8C7< USB_RTS
C174
8D7>
OUT
VDDOUT
GNDPA
GNDANA
0.1U
1.8V 1
VBUS N/C
85 86 124 125 132 133
TP152 TP153
TP73
1 2 4 10 12 24 26 31 33 35 38 40 42 44 46 59 61 63 R399 65 67 87 89 91 95 97 99 101 103 105 107 110 112 114 118 120 135 137 140 142 144
TP156 TP157 TP158
23
12MHZ
7 6 5 4 3 2 1 0
TP127 102
TP160 TP161 TP162 TP163 TP164
GNDPD
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
/RD 7C5
/DISP_CLK_OFF
GNDPD
150
R260
IN
100K
TP77
29
BL_ENABLE
6B6>
150
6B6>
DISPLAY CONTRAST
E24
R181
S
1.00M
Q7
1
E25
U5
10K
5
VIN
SW
1
10K
A
CR3
MBR0530
R19
4 SHDN 10U
24K
1.00K
1U
R23 1.00K
121K
20V
LP324
1
R29
1
100
R20
+
4 V+
U6
V11
LP324
7
R30
4.7P
C26 1U
1U
R26
9 10
+
4 V+
U6
V11
LP324
8
C42
C45
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
0.1U
C36
C32
C35
R31
3
100
+3.0V
C38 C37
C27 1U
+3.3VPD 2
100
10.0
C28
1.5M
100K
6 5
E17 C29
R21
R22
3
GND 2
C46
GNDPD
FB
R25
+20V
R18
C
2M
2
V11
1U
1U
LT1615
3
+
U6
C30
C31
20%
Q8
-
4 V+
L1 10UH
MMUN2211
2 3
FDN338P
DISP_V
0
MECCA
R24
121K
+12V
100K
IN
R81
TP76
31
1
R34
NOLOAD
IN
-ST
DISP_V
6C3> DISP_OFF
34
40MHZ GND 2
GNDPD
GNDPD
30 26 23
OSC
U4
B1 B2
R180
6B6> CONTRAST
D0 D1 D2 D3 D4 D5 D6 D7
+3.3VPD
R253
10.0K
10.0K
TEST
+12V
6C3> ENABLE_20V
R252
CLKI CNF0 CNF1 CNF2 CNF3 CNF4
6B6>
S4
LEFT
C1 C2
6C7
USB_TXD
OUT OUT IN IN
100
10U
C53
100P
R47
16 VDD2
GNDPD
1.00K
14 VO A 13 VO B 12 VI C
R45 1.00K
R46
C55
10 VE2 9 15 GND2 GND2
C56
100P
100P
1.00K
VI A 3
R38
1.00K
R37
1.00K
VO C 5
R39
1.00K
VE1 7 GND1 2 GND1 8
C58 10U
R42
TP79 TP80
10.0K
10.0K
OUT
23 22 13 14 12
U12
1
3
2
USBDM
0.1U
21
10U
17
7
0.01U
16
USBDP
15
RESET
19
J3
C47
0.1U
OSCI 27
TP84
OSCO 28
TP85
USB
USB
2
+
GND
1
GND
3
U9 75240
TEST 26
GND
C49
1 2 3 4
FT232RL
CBUS0 CBUS1 CBUS2 CBUS3 CBUS4
GNDUSB
4
3V3OUT
U7
C60
1.00K 0.1U
VCC
TXD RXD RTS CTS DTR DSR DCD RI
C136
+3.3VPD
R50
6C5< /USB_PRESENT
TP113
10.0K
10.0K
1 5 3 11 2 9 10 6
51.1
R215 C48
20
VCCIO
R41
R40
GNDPD GNDPD
4
VO D 6
ADUM1402
R48
R36
VI B 4
11 VI D
1.00K 1.00K
VDD1 1
U8
R216
+5V_USB
MT1
C54
100P
MT2
C57
-
AGND
4
25
18
R49
NOLOAD
L11
95_OHM
OUT
NOLOAD
1
CR39
2
440V
POS_ESD 4B1
24 oz
B
>24 oz
C
>8 oz
Functional A.11
PASS
FAIL
A.13
PASS
FAIL
A,32
PASS
FAIL
A.36
PASS
FAIL
A.41
PASS
FAIL
A.46
PASS
FAIL
Electrical Outlet Polarity B.10
PASS
FAIL
B.13
PASS
FAIL
B.18
PASS
FAIL
B.27
PASS
FAIL
B.30
PASS
FAIL
B.33
PASS
FAIL
EO Ground to NULL Jack C.6
PASS
FAIL 0 to 2.5 Ω
0
+2.5 Ω
+2.5 Ω
D.8
112.5 – 117.5 V
115.0 V
±2.5 V
±2.0 % +0.2 V
D.12
112.5 – 117.5 V
115.0 V
±2.5 V
±2.0 % +0.2 V
C.8 Mains Voltage
GFI Verification
A-2
E.11
PASS
FAIL
E.23
PASS
FAIL
E.38
PASS
FAIL
E.50
PASS
FAIL
E.64
PASS
FAIL
E.76
PASS
FAIL
Datasheets Introduction
A
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
Point to Point Voltage F.9
0.78 to 1.22 V
1.00 V (120 Hz)
±0.22 V
±2.0 % +0.2 V
F.12
244.8 to 255.2 V
250 V (60 Hz)
±5.20 V
±2.0 % +0.2 V
F.15
60.10 to 81.32 V
70.71 V (633 Hz)
±10.61 V
±15.0 %
F.21
3.72 to 4.28 V
4.00 V
±0.28 V
±2.0 % +0.2 V
F.24
7.64 to 8.36 V
8.00 V
±0.36 V
±2.0 % +0.2 V
F.27
9.60 to 10.40 V
10.00 V
±0.40 V
±2.0 % +0.2 V
F.30
24.30 to 25.70 V
25.00 V
±0.70 V
±2.0 % +0.2 V
F.33
39.00 to 41.00 V
40.00 V
±1.00 V
±2.0 % +0.2 V
F.36
78.20 to 81.80 V
80.00 V
±1.80 V
±2.0 % +0.2 V
F.39
127.2 to 132.8 V
130.00 V
±2.80 V
±2.0 % +0.2 V
F.42
235.00 to 245.00 V
240.00 V
±5.00 V
±2.0 % +0.2 V
-0.015 to +0.015 Ω
0.0000 Ω
±0.015 Ω
±2.0 % +0.015 Ω
Resistance G.11
1.800 Ω
G.14 [1]
G.21
-0.015 to +0.015 Ω
0.0000 Ω
±2.0 % +0.015 Ω ±0.015 Ω
1.800 Ω
G.24 [1]
±2.0 % +0.015 Ω ±2.0 % +0.015 Ω
Insulation Resistance H.11
250.0 to 300.0 V
250.0 V
+50.0 V
+20.0 %
H.16
500.0 to 600.0 V
500 V
+100.0 V
+20.0 %
H.20
1.750 to 2.250 mA
2.000 mA
±0.250 mA
±12.5 %
A-3
ESA612 Service Manual
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
Mains to PE 9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
H.79.RA
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
LL
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
LA
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
RL
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
V1
9.6 to 10.4 MΩ
10.000 MΩ
±0.4 MΩ
±2.0 % +0.2 MΩ
250
0.486 to 0.914 MΩ
0.700 MΩ
±0.214 MΩ
±2.0 % +0.2 MΩ
500
0.486 to 0.914 MΩ
0.700 MΩ
±0.214 MΩ
±2.0 % +0.2 MΩ
250
0.780 to 1.220 MΩ
1.000 MΩ
±0.220 MΩ
±2.0 % +0.2 MΩ
500
0.780 to 1.220 MΩ
1.000 MΩ
±0.220 MΩ
±2.0 % +0.2 MΩ
250
6.170 to 6.830 MΩ
6.500 MΩ
±0.330 MΩ
±2.0 % +0.2 MΩ
500
6.170 to 6.830 MΩ
6.500 MΩ
±0.330 MΩ
±2.0 % +0.2 MΩ
250
17.440 to 18.560 MΩ
18.000 MΩ
±0.560 MΩ
±2.0 % +0.2 MΩ
500
17.440 to 18.560 MΩ
18.000 MΩ
±0.560 MΩ
±2.0 % +0.2 MΩ
AP to PE H.44 AP to NE H.57 Mains to NE H.68 Mains to AP
A-4
Datasheets Introduction
A
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
250
20.150 to 23.850 MΩ
22.000 MΩ
± 1.850 MΩ
± 7.5 % +0.2 MΩ
500
20.150 to 23.850 MΩ
22.000 MΩ
± 1.850 MΩ
± 7.5 % +0.2 MΩ
250
55.300 to 64.700 MΩ
60.00 MΩ
± 4.700 MΩ
± 7.5 % +0.2 MΩ
500
55.300 to 64.700 MΩ
60.00 MΩ
± 4.700 MΩ
± 7.5 % +0.2 MΩ
250
92.30 to 107.70 MΩ
100.00 MΩ
± 7.70 MΩ
± 7.5 % +0.2 MΩ
500
92.30 to 107.70 MΩ
100.00 MΩ
± 7.70 MΩ
± 7.5 % +0.2 MΩ
DC Leakage I.10
0.980 to 1.020 KΩ
1.000 KΩ
±0.02 KΩ
±2.0 %
I.23
8.900 to 11.100 μA
10.00 μA
±1.1 μA
±1.00 % +1.0 μA
I.27
48.500 to 51.500 μA
50.00 μA
±1.5 μA
±1.00 % +1.0 μA
I.31
98.000 to 102.000 μA
100.00 μA
±2.0 μA
±1.00 % +1.0 μA
I.35
157.40 to 162.600 μA
160.00 μA
±2.6 μA
±1.00 % +1.0 μA
I.39
335.60 to 344.40 μA
340.00 μA
±4.4 μA
±1.00 % +1.0 μA
I.43
494.00 to 506.00 μA
500.00 μA
±6.0 μA
±1.00 % +1.0 μA
I.47
0.989 to 1.011 mA
1.000 mA
±0.011 mA
±1.00 % +0.001 mA
I.51
1.583 to 1.617 mA
1.600 mA
±0.017 mA
±1.00 % +0.001 mA
I.55
3.356 to 3.444 mA
3.40 mA
±0.044 mA
±1.00 % +0.01 mA
I.59
4.940 to 5.060 mA
5.00 mA
±0.060 mA
±1.00 % +0.01 mA
I.63
6.020 to 7.080 mA
7.00 mA
±0.080 mA
±1.00 % +0.01 mA
A-5
ESA612 Service Manual
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
AC Filter Frequency Response J.8
J.9
J.15
Analyzer Current Measured Current 0.98801 – 1.00799
±0.00999
Analyzer Current Measured Current 0.67218 – 0.70982
J.19
0.99800
0.69100
±0.01882
Analyzer Current Measured Current 0.088669 – 0.102492
0.09558
±0.0069116
Differential Leakage
A-6
K.12 [1]
76.00 μA
±10.0 % +20.0 μA
K.13 [1]
160.00 μA
±10.0 % +20.0 μA
K.13 [1]
240.00 μA
±10.0 % +20.0 μA
K.13 [1]
500.00 μA
±10.0 % +20.0 μA
K.13 [1]
760.00 μA
±10.0 % +20.0 μA
K.14 [1]
1.600 mA
±10.0 % +0.02 mA
K.14 [1]
2.40 mA
±10.0 % +0.02 mA
K.14 [1]
5.00 mA
K.14 [1]
7.60 mA
±10.0 % +0.02 mA
K.15 [1]
16.00 mA
±10.0 % +0.02 mA
±10.0 % +0.02 mA
Datasheets Introduction
A
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
Leakage Functionality ECG Leakage Functionality RA L.15
PASS
FAIL
L.18
PASS
FAIL
L.21
PASS
FAIL
L.15
PASS
FAIL
L.18
PASS
FAIL
L.21
PASS
FAIL
L.15
PASS
FAIL
L.18
PASS
FAIL
L.21
PASS
FAIL
L.15
PASS
FAIL
L.18
PASS
FAIL
L.21
PASS
FAIL
L.15
PASS
FAIL
L.18
PASS
FAIL
L.21
PASS
FAIL
LL
LA
RL
V1
Direct Applied Part, Alternative Applied Part: Ground to RA M.20 [1]
M.38 [1]
1.000 mA
±1.00 % +1.0 μA
1.000 mA
±1.00 % +1.0 μA
1.000 mA
±1.00 % +1.0 μA
1.000 mA
±1.00 % +1.0 μA
Direct Applied Part, Alternative Applied Part: Red to RA N.20 [1]
N.38 [1]
A-7
ESA612 Service Manual
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
Alternative Equipment: Red to Hot O.20 [1]
1.000 mA
±1.00 % +1.0 μA
Alternate Equipment Leakage, Alternative Applied Part: RA to Hot P.20 [1]
P.38 [1]
1.000 mA
±1.00 % +1.0 μA
1.000 mA
±1.00 % +1.0 μA
1.000 mA
±1.00 % +1.0 μA
100.00 μA
±1.00 % +1.0 μA
100.00 μA
±1.00 % +1.0 μA
100.00 μA
±1.00 % +1.0 μA
100.00 μA
±1.00 % +1.0 μA
100.00 μA
±1.00 % +1.0 μA
Alternative Applied Part Patient: RA to Neutral Q.18 [1] Patient Auxiliary: RA to RL R.22 [1]
Direct Equipment, Patient: Ground to RA S.20 [1]
S.44 [1] Direct Equipment, Enclosure: Ground to Red T.20 [1]
T.42 [1] Filter Tests
A-8
U.20 [1]
0.000 μA
±1.00 % +1.0 μA
U.26 [1]
0.000 μA
±1.00 % +1.0 μA
Datasheets Introduction
A
Table A-1. ESA612 Verification Datasheet (cont.) Step No.
Measured Value
Limits
Nominal (Actual) [1]
Range
Percentage
MAP Voltage and Current Limits V.9
115.0 V
±5.0 %
V.21
230.0 V
±5.0 %
V.31
3.500 mA
±25.0 %
V.40
7.500 mA
±25.0 %
V.52
1.000 mA
±25.0 %
ECG Waveforms W.9
PASS
FAIL
W.11
0.674 mV
±5.0 %
LL
1.673 mV
±5.0 %
LA
1.384 mV
±5.0 %
V1
2.074 mV
±5.0 %
X.9 [1]
8.000 A
±5.0 % +0.2 A
X.12 [1]
3.000 A
±5.0 % +0.2 A
X.15 [1]
1.000 A
±5.0 % +0.2 A
Equipment Current
[1]
These steps require measurement of the standard prior to testing the UUT. The value of the standard is entered into the “Actual” block, and the range and limits are filled in based on this value.
Serial Number on Analyzer: UI Firmware Revision: Meter Firmware Revision: Nominal Voltage: Final Data Serial Number: Cal Date: Technician Number: Technician Signature:
A-9
ESA612 Service Manual
Table A-2. Calibration Datasheet Step No. A.15
Measured Value PASS
Limits
(Actual) [1]
Range
Percentage
FAIL
A.50
1470.00 – 1530.00 mV
1.500 V
±30 mV
±2 %
A.53
490.00 – 510.00 mV
0.5000 V
±10 mV
±2 %
A.58
147.00 – 153.00 mV
150.00 mV
±3 mV
±2 %
A.63
14.70 – 15.30 mV
15.00 mV
±0.3 mV
±2 %
B.11
1470.00 – 1530.00 mV
1.500 V
±30 mV
±2 %
0.5000 mA
C.13 [1]
D.9
244.8 – 255.2 V
F.13 [1]
G.11 [1]
H.13 [1] I.18
PASS
FAIL
I.30
PASS
FAIL
I.52
PASS
FAIL
I.64
PASS
FAIL
I.85
PASS
FAIL
PASS
FAIL
I.97 J.12
[1]
250.00 V
±1 % +0.1 μA ±5.2 V
97.8 – 102.2 v
L.20
97.8 – 102.2 v
±2 % +0.2 V
10.000 MΩ
±2 % +0.2 MΩ
3.000 Ω
±2 % +0.015 Ω
11.000 mA
±10 % +0.02 mA
3.000 A
K.7
[1]
±1 % +0.1 μA
3.0000 mA
C.22 [1]
A-10
Nominal
±5 % +0.2 A
100 V
±2.2 V
±2.0 % +0.2 V
100 V
±2.2 V
±2.0 % +0.2 V
These steps require measurement of the standard prior to testing the UUT. The value of the standard is entered into the “Actual” block, and the range and limits are filled in based on this value.
Datasheets Introduction
A
Serial Number on Analyzer: UI Firmware Revision: Meter Firmware Revision: Nominal Voltage: Final Data Serial Number: Cal Date: Technician Number: Technician Signature:
A-11
ESA612 Service Manual
A-12