DI-2000 Service Manual: Medifusion

MEDIFUSIONⓇ DI-2000 Service Manual Read this manual carefully before using the pump to use the pump correctly and safe

Views 213 Downloads 58 File size 5MB

Report DMCA / Copyright

DOWNLOAD FILE

Recommend stories

Citation preview

MEDIFUSIONⓇ

DI-2000 Service Manual

Read this manual carefully before using the pump to use the pump correctly and safely. Daiwha Corporation

Ⓒ Copyright 2002, DAIWHA Corporation, Ltd. All rights reserved.

◎ This manual provides technical descriptions of the infusion pump (Model: DI2000) manufactured by DAIWHA Corporation. ◎ Repair or circuit alteration of this equipment can only be carried out by person authorized by DAIWHA. The warranty shall not apply to defects or damage caused by repairs or disassembly by anyone other than person authorized by DAIWHA. ◎ Parts or circuits which are used for this equipment can be replaced to improve the safety and performance of this equipment. ◎ Contact DAIWHA if you have any questions regarding the equipment or the manual.

Operation

Address: Daiwha Corporation, Daiwha B/D, 733-18 Yeoksam-dong, Gangnam-gu, Seoul, Korea

Technology

Tel:

+82-2-558-1711∼6

Fax:

+82-2-558-1745

Address: Daiwha Corporation, Daiwha B/D, 733-18 Yeoksam-dong, Gangnam-gu, Seoul, Korea Tel:

+82-2-558-1711

Fax:

+82-2-558-1745

◎ Safety precautions are classified as below in accordance with expected danger, seriousness of damage, and urgency of danger occurrence. - Danger: A precaution which can result in critical personal injury or immediate loss of life if instructions are not followed. - Warning: A precaution which could result in critical personal injury or loss of life if instructions are not followed. - Caution: A precaution which could result in minor personal injury or damage to the product if instructions are not followed.

Table of Contents

1. Introduction 2. Appearance and Functions 3. Inspection Procedures 4. Operating Mechanism 5. Electrical Wiring Inspection 6. Power Supply Inspection 7. Troubleshooting 8. Disassembly and Adjustment 9. Assembly and Exploded Views 10. PCB Layout 11. Schematic ※ DI-2000, DI-2200 Firmware Upgrade Guide

1. Introduction 1.1 Overview This manual provides technical descriptions on MedifusionⓇ DI-2000 Volumetric Infusion Pump for hospitals' medical equipment technicians or person authorized by Daiwha Corporation for maintenance of the equipment. For more details on how to use the equipment, see the user manual.

1.2 Product Overview MedifusionⓇ DI-2000 Volumetric Infusion Pump is a medical instrument that injects a certain amount of drug into patients through the infusion set at the rate selected by the patient. The pump consists of the semi-transit linear peristaltic infusion part that infuses drug at a certain rate by the program and the circuit part that controls the infusion part.

The pump can be used with power supply of 50/60Hz between 100 and 240 VAC and includes self-charged built-in battery. The pump can be used in motion, and the infusion set designated for the pump is recommended. Make sure that the set is equipped with Luer lock adapter when drug is infused by central line catheter.

Available flow rates are 99.9 ml/h at 0.1 and 1,200ml/h at 100. Infused volume can be displayed between 1 and 9999 ml. When the total volume is set to infinite, the infused volume will be displayed as 0 at 9999 ml. Once the set volume is all infused, the pump will be automatically changed to Keep Vein Open (KVO) rate. The pump will stop infusion anytime when the stop key is pressed.

1.3 Technical Specifications Product Name Model

Volumetric Infusion Pump DI-2000

Pumping Mechanism

Peristaltic Transit Finger

Flow Rate

0.1 ~ 1200 ml/h

Micro(on): 0.1~99.9ml/h (0.1 ml/h step) 100~1200ml/h (1ml/h step) Micro(off): 1~1200ml/h (1ml/h step)

±5% ( Korea Vaccine CODE S203 Ⓣ type) Measure every hour at room temperature (25 ℃) using water or saline solution (above 1.0 ml/h of flow rate). * Accuracy may differ depending on the type of

Accuracy

medical fluid and IV set. Micro(on): 0.1~99.9ml (0.1 ml step) Total Vol. Infusion

0.1~9999 ml or

100~9999ml (1ml step)

infinite

Micro(off): 1~9999 ml (incremental by 1 ml)

Range Infused Vol.

0.0~9999 ml

0.1 ~ 99.9ml (0.1 ml step) 100 ~ 9999ml (1 ml step)

Bolus Rate (Purge)

700 ml/h (default); changeable to 1~1200 ml/h

Bolus Volume (Purge)

3 ml (default); changeable to 1~9999 ml At flow rate 1.0~1200 ml/h

K.V.O Rate

1 ml/h (default) 1~9ml/h 0.1 ml (fixed)

At flow rate 0.1~0.9 ml/h

Display Alarms

7 segments (4 digit * 3 lines) * Air bubble alarm * Occlusion alarm (detection capacity: 100~950 mmHg) - Seven adjustable steps for downstream occlusion * Door open alarm * Battery Low, Battery Empty alarm * Abnormal Drop alarm

* Infusion Complete alarm (automatic switch-over to KVO feature) * AC/DC Power Shut-off alarm * Start Reminder alarm (after the pre-set pause) * Standby alarm (at a 2-minute interval when not infusing) * Near Complete alarm * The pump stops operating when alarms go off except for battery low, start reminder, and infusion completion signals. * Malfunction

Safety Features

The status of the pump is displayed and an alarm goes off. * Door Open: Infusion is unavailable when the door is open. * Key Lock: Only the START, STOP and ON/OFF keys work. * Air Sensor: Detects air bubbles. * Occlusion Sensor: Detects clogging of the tube. * Drop Sensor: Detects medical fluid drops. * Nurse Call: Monitoring is possible if an alarm goes off.

Other Functions

* Time Mode: Calculates the flow rate by setting the infusion time. * Gtt Mode: Calculates the flow rate by setting the number of infusion drops. * Dosage Mode: Calculates the flow rate by setting the dosage. * Titration: Flow rate can be altered during infusion (available only when the remaining time for infusion is more than 5 minutes). * Battery Low, Empty: Alarms go off 30 minutes and 3 minutes before the battery goes flat. * Date and Time: The date and time can be checked

without a power supply, since a clock is installed. * Pause: Infusion resumes automatically after preset time elapses when the pause key is pressed and infusion is stopped (default 2 minutes, setting range from 1 minute to 24 hours in 1 minute units). * Event History: Saves up to 2,000 events. * The electronic data of the infusion pump are stored for over 2 years even if it has been turned off. Power Supply

* AC 100~240 VAC, 50~60Hz (Fuse: 250V, T3.15A) * DC 12V (500mA) * Ni-MH rechargeable battery - 2100mA/h, AA, 1.2V, 10EA - Operating time: about 4 hours (at the flow rate of 25ml/h) - Recharging time: about 10 hours

Power Requirements

35 VA

Protection Type

* Protection from electric shock: Class Ⅰ * Protection capacity for electric shock: Type CF * Protection from liquid infiltration: IPX 1 (drip-proof)

Dimensions (D×W×H)

120×130×206mm

Weight

About 1.7 kg

Operating Environment

Temperature: 5~40℃, Humidity: 20~90% Atmospheric pressure: 70~106kPa (525.04~795.06mmHg)

Storage Environment

Temperature: -10~60℃, Humidity: 10~95% Atmospheric pressure: 65~120kPa (487.54~900.07mmHg)

Accessories

User Manual 1 ea (Manual No.: Refer to the front cover of the manual.) Infusion Pole Clamp 1 ea (Model: PoleClamp-A) Drop Sensor (Model: DropSensor-A) (optional) AC Power Cable 1 ea (4 meters, 0.75 ㎟ x 3C, KS C

IEC 60227-5 300/500V 60227 KSIEC 53) Communication

USB, RS485 (optional)

2. Appearance and Functions 2.1. Front Panel

ⓐ Occlusion alarm LED ⓑ Key lock LED ⓒ Air (bubble) alarm LED ⓓ Door open alarm LED ⓔ Infusion completion alarm LED ⓕ Battery low alarm LED ⓖ Micro LED ⓗ Total volume (ml) display ⓘ Infused volume (ml) display, Remaining time display ⓙ Flow rate (ml/h) display ⓚ AC/DC power LED ⓛ Battery level display LED

① SEL key ② MODE key ③ Hundreds digit up key ④ Hundreds digit down key ⑤ Tens digit up key ⑥ Tens digit down key ⑦ Units digit up key ⑧ Units digit down key ⑨ BOLUS key ⑩ PAUSE key ⑪ STOP/CLEAR key ⑫ START key ⑬ ON/OFF key ⑭ Remaining time key

2.2. Display ⓐ ⓑ ⓒ ⓓ ⓔ

Displays OCCL LED LOCK LED AIR LED DOOR LED COMP LED

ⓕ BATT LED ⓖ Micro LED ⓗ T.Vol (ml) display ⓘ I.Vol (ml) display, R.TIME display ⓙ F.Rate(ml/h) display ⓚ AC/DC LED ⓛ BATT STATUS LED

Functions To be illuminated if the IV set is clogged. To be illuminated if keys are locked (Key Lock). To be illuminated if air is flowed into the pump body. To be illuminated if the door is open. To be illuminated when infusion is completed. To be illuminated when the remaining battery capacity is 30 minutes and turned off when the battery capacity is three minutes. To be illuminated with the pump is operated in the Micro mode. To display the desired infusion volume. To display the volume infused into the patient and the remaining volume. To display flow rate, alarm messages, and error codes. To be illuminated when electric power (AC/DC) is supplied. To display the battery status in three levels (high, middle, low).

2.3. Keys Keys ① SEL key (

Functions Selects total volume, flow rate, and infusion time in

)

infusion mode 1. ② MODE key (

Switches the mode among TIME, GTT, and DOSAGE

)

in infusion mode 2 by long pressing the MODE key. ③ Hundreds digit up key

Sets total volume, flow rate, and Set Mode. Whenever

(

the key is pressed, the hundreds digit increases by 1.

)

④ Hundreds digit down

Sets total volume, flow rate, and Set Mode. Whenever

key (

the key is pressed, the hundreds digit decreases by 1.

)

⑤ Tens digit up key (

)

Sets total volume, flow rate, and Set Mode. Whenever the key is pressed, the tens digit increases by 1.

⑥ Tens digit down key

Sets total volume, flow rate, and Set Mode. Whenever

(

the key is pressed, the tens digit decreases by 1.

)

⑦ Units digit up key (

Sets total volume, flow rate, and Set Mode. Whenever )

the key is pressed, the units digit increases by 1.

⑧ Units digit down key

Sets total volume, flow rate, and Set Mode. Whenever

(

the key is pressed, the units digit decreases by 1.

)

If the BOLUS key is pressed during infusion or pause, infusion will be carried out at the rate and volume set in the Set Mode (BOLUS volume and rate). While the key is being pressed, infusion continues at the rate set in the ⑨ BOLUS key (

Set Mode (BOLUS rate). If the key is short pressed in )

infusion mode 2 or setting modes 1 & 2, the pump will be switched to the standby mode. (While BOLUS is operating during infusion or pause, only the Occlusion Sensor will work and the Air Sensor will not work.)

⑩ PAUSE key (

If the PAUSE key is pressed during infusion, infusion )

will stop for the length of time set in the Set Mode and

resume after the lapse of the pre-set time. (In order to resume infusion before the pre-set time has elapsed, press the STOP/CLEAR key. The pre-set time automatically passes and the pump will be switched to a normal state.) This key is used to stop drug infusion, BOLUS, and K.V.O. If it is long pressed in standby for infusion, the ⑪ STOP/CLEAR key (

)

blinking display information (total volume, flow rate, and cumulative infusion amount) will be initialized. This key is used to start infusion after infusion setting is

⑫ START key (

completed. However, infusion starts only when the total

)

volume is bigger than the volume to be infused. If the ON key is pressed, the pump is immediately ⑬ ON/OFF key (

turned on. Long press the OFF key to turn off the pump. )

(The key will not work during infusion and only work in the infusion standby mode.) Once infusion starts, the infusion volume will be

⑭ R.TIME key (

)

displayed. If the R.TIME key is pressed, the remaining infusion time will be displayed.

2.4. Inside the Door ① Air Detector – Prevents the patient from getting injected with air by detecting the air in the IV set. ② Auxiliary Door Lock – Used when opening the auxiliary door. ③ Pumping Unit – Infuses medical fluid by pressing the tube of the IV set. ④ Door Lock Lever – Opens or closes the door. ⑤ Occlusion Detector – Detects whether the tube at the bottom of the equipment is clogged. ⑥ Auxiliary Door – Holds the IV set in position for safe infusion of medical fluid. ⑦ Door Lock – Prevents the door from opening.

2.5. Rear Panel ① RS485 Port – Used when downloading History Log and communicating via RS485. ② Drop Sensor Port – The Drop Sensor is connected to this port. ③ Nurse Call Port – Connected to the Nurse Call equipment with a Nurse Call cable and warning signals are sent to the Nurse Call equipment if necessary during the infusion. ④ USB Port – Used when downloading History Log and communicating via USB. ⑤ DC Power Port - Provides 12 ~ 15 VDC power supply. ⑦ Key Lock – All keys but START ( STOP/CLEAR (

), and ON/OFF (

), ) will be

locked. ⑧ Power Switch - Turns power on or off. ⑨ Power Inlet - The power cord is plugged into this inlet. ⑩ Battery – Built-in battery

3. Inspection Procedures

In order to prevent sudden problems of Infusion Pump DI-2000 and ensure safe use, monthly inspections need to be carried out in accordance with the following procedures.

3.1. Power Supply Inspection 1) Connect the infusion pump to the external power. Make sure that the infusion set is not mounted and the door is closed. 2) Check whether the AD/CD indicator is on. 3) Turn on the power switch on the rear side of the pump. 4) Check whether the three batter level indicators on the front panel are all illuminated. 5) Disconnect the external power supply and check whether the AC/DC indicator is off. Also, check whether at least one battery level indicator is illuminated. 6) Connect the external power supply again and check whether the AC/DC indicator is illuminated.

3.3. Dropping Test 1) Turn on the power. 2) Fill the infusion set with a solution and remove air inside the tube. Make sure that the infusion set container is filled to the half level so that dropping of solution drops can be checked. 3) Open the door and mount the infusion set to fit the sensors and the pumping unit. 4) Mount the infusion set into the device such that the distance between the top of the pump and the solution bottle mouth is about 1 meter. Keep the infusion set clamp closed. 5) Close the pump door and open the infusion set clamp. Observe for 60 seconds whether the solution is dropping from the end of the set. 6) Open the pump door and lower the pumping unit lever to open the pumping unit. Once free drops are generated, close the door. Leave the pump for one minute and check for one minute whether drops are generated from the end of the infusion set.

3.4. Accuracy Test 1) Turn on the power and mount the infusion set without air bubbles and clogging in the tube. 2) Set flow rate and total volume to 100 ml/h and 100 ml respectively. 3) Check whether the flow rate and the total volume are set to 100 ml/h and 100 ml respectively and make sure that the infused volume is 0 ml. Otherwise, set the infused volume to 0 in the infusion mode. 4) Start infusion by pressing the START key. Make sure that air bubbles or clogging would not occur during infusion. 5) When the infusion completion alarm sounds when infusion is completed, stop infusion by pressing the STOP key. Check infusion time (1 hour) and infused volume (100 ml).

3.5. Door Open Alarm Test 1) Turn on the power and mount the infusion set without air bubbles and clogging inside the tube. 2) Open the door and check whether the door open indicator is illuminated. 3) Start infusion by pressing the START key. Check whether the door indicator blinks with an alarm sound. 4) Press the STOP key and check whether the alarm is off. Check whether the door open indicator is still illuminated. 5) Close the door and check whether the door open indicator is turned off. 6) Press the START key to start infusion. Check whether the infusion indicator blinks. 7) Open the door again and check whether the infusion indicator stops blinking and the door open indicator blinks. 8) Close the door and press the STOP key to stop the alarm sound.

3.6. Air Alarm Test 1) Turn on the power and mount the infusion set without air bubbles and clogging inside the tube. 2) Start infusion by pressing the START key. Check whether the STOP indicator is turned off and the infusion indicator blinks.

3) Create air bubbles intentionally inside the tube and make them flow into the pump. 4) Check whether the air indicator blinks and alarm sounds when air bubbles reach the air sensor. The infusion indicator stops blinking, and the stop indicator is turned on. 5) Press the STOP key and check whether the alarm sound stops and the air indicator is still illuminated. 6) Open the door and remove air bubbles from the air sensor. Mount the infusion set again. 7) Close the door and check whether the air indicator is turned off.

3.7. Occlusion Alarm Test 1) Turn on the power and mount the infusion set without air bubbles and clogging inside the tube. 2) Start infusion by pressing the START key. Check whether the infusion indicator blinks. 3) Block the tube of the infusion set about 15 cm away from the lower part of the pump and check whether the occlusion indicator blinks with an alarm sound. Now, the infusion indicator stops blinking and the stop indicator is turned on. Press the STOP key and check whether the alarm sound stops and the occlusion indicator is still illuminated.

3.8. Battery Low Alarm Test 1) Turn on the power and mount the infusion set without air bubbles and clogging inside the tube. 2) Set flow rate and total volume to 25 ml/h and 100 ml respectively. 3) Start infusion by pressing the START key. Check whether the micro function indicator is turned on, the STOP indicator is turned off, and the infusion indicator blinks. 4) Disconnect external power supply by pulling the plug out and check whether the pump is operated without problems. The AC/DC indicator will be turned off about one to two seconds after the external power supply is removed. 5) If more than one battery level indicator is not illuminated and the battery indicator blinks with an alarm sound, the pump needs to be recharged. If the external power supply is connected when the battery low alarm sounds, the battery low alarm will stop,

all three battery level indicators will be turned on, and the AC/DC indicator will be turned on. 6) Plug in again and check whether the device is operated without problems. Repeat 3) and 4) above. 7) As the pump continues infusion, the battery level indicators will be turned off one after another and eventually the battery low alarm will sound. Connect to the external power supply and check whether the alarm stops and infusion continues.

3.9. Electrical Safety Test The measurement of leakage current and ground impedance of DI-2000 should follow the international standards: IEC 60601-1 and IEC60601-2-24. Leakage current should not exceed one mA. For the ground impedance, make sure that the clamp assembly should not exceed 0.1Ω (expected to check).

3.10. User Maintenance 3.11.1. Replacement of fuse of the power inlet Open the fuse box at the top of the power inlet and replace the fuse with a rated fuse (250V T3.15A)

3.11.2. Replacement of the built-in battery Remove the battery box cover on the rear panel and replace the battery with a Ni-MH (Nickel Hybrid Battery (12V, 2100 mAh)) pack with the same size. Make sure that the battery is place in the right position.

3.11.3. Recharging of battery A storage battery can be recharged regardless of power switch. When the external power supply is connected, the AC/DC indicator is turned on and recharging starts. The pump needs to be recharged at least once a month.

3.11.4. Cleaning In order to minimize hardening or accumulation of solution onto the pump, clean the pump immediately after use. Wash the external case and the front panel with a soft cloth

or cotton soaked with warm water. Do not use thinner, solvent, benzene, ammonium, or acetone. Also, do not disinfect the pump by autoclave or with EO (Ethylene Oxide) gas.

3.12. User Mode No

Parameter

Setting Value

Factory Setting Value

K001(Korea Vaccine T1

Maker IVSET

type), T001(Turkey IVSET),

K001

expected to add 2

DROP Set

15, 20, 60

20

3

Micro Set

ON, OFF

OFF

4

Calibration Set

5

Bolus Flow Rate(Purge)

6

Bolus Volume(Purge)

7

K.V.O Set

00:01~ 24:00 (1 min)

1ml/h

8

Pause Timer

ON, OFF

2min

9

Standby Alarm

00:01~24:00 (1min), OFF

OFF

10

Near Complete Alarm

11

Air Sensor Set

12

Occl Sensor Set

ON, OFF

7 (insensitive)

13

Nurse Call

0~11 (11 High Sound)

OFF

14

Sound Volume Set

15

Date Set

16

Time Set

17

Save Set

18

Set Factory Initialization

1~1200 ml/h ( Resolution 0.1 ml/h ) 1~100ml 1~9 ml/h (Resolution 1 ml/h )

Low (sensitive), Middle (ordinary), High (insensitive) 1~7 (The smaller the number is, the more sensitive.)

Year: 2010~2099, Month: 01~12, Day: 01~31 Hour: 01~24, Minute: 00~59, Year: 2010~2099, Month: 01~12, Day: 01~31

0.0 % 700 ml/h 5

OFF

Hi (insensitive)

11 (Max) Year: 2010 Month: 05, Day: 18 Hour: 01 Minute: 1,

ON, OFF

ON

Initializing by the four digit

Master password

password

maintained

4. Operating Mechanism 4.1 Block Diagram

4.2. Power Supply The SMPS (Switching Mode Power Supply) changes AC phase current (100-240 VAC, 50/60 Hz) to DC current (24VDC, 1A) and charges the built-in battery. Through the regulator inside the main board, it generates 3.3 V for each electronic element and 5 V for the air sensor and displays.

The built-in battery is used as an emergency power supply when AC power supply is disconnected. The phase power is provided from the inlet with an EMC filter. It is rectified by the SMPS and switched to DC power. The AC/DC power supply LED is illuminated regardless of the power switch, and the built-in battery is automatically charged. When the phase power is disconnected, the built-in battery is automatically on, supplying power to each circuit ceaselessly.

The inlet and the SMPS have a separate fuse (250V T3.15) to prevent sudden overcurrent. Also, the SMPS has a feature to prevent over-current and over-voltage. The main power supply is provided to each part by the SMPS as 24V/1A power and used to recharge the built-in battery. The power supply is divided by the regulator into 24V for the motor and 5V and 3.3V for circuits.

4.3. Main Board The main board is composed of memory that stores MCU and the program, EEPROM that stores data, the source program that controls each part, MCU oscillator that is the basis of all signals, and the reset circuit that initializes MCU.

All system signals are controlled by 8-bit MCU, and signals required for control are sent to parts or received by the address decoder. The clock pulse of MCU is supplied with 8MHz at the oscillator part and provides pulse for the A/D converter by the pulse dispenser. The program is embedded in the memory and drives the pump. Major data are stored in EEPROM and preserved semi-permanently even if power supply is disconnected.

4.4. Key Controls The key controls are structured as the key matrix of tact switches. Signals input by pressing each key are received by the MCU port and processed by scanning keys. For keys input properly, proper operations are carried out.

4.5. Displays The display part consists of the controller, 4-digit 7 segments, and LED. The displays are controlled by three display controllers and divided into flow rate display and other displays. Each digit of 7 segments is illuminated sequentially to display one after another. Warning lights and indicators, battery level indicators, and AC/DC indicator are controlled and illuminated by MCU.

4.6. Motor Control

Motor control is composed of motor, motor driver, and motor checker. For precise control, uni-polar stepping motor was adopted and 1/16 micro step control is carried out through the motor driver for more precise control.

A motor checker is attached to the pumping unit axis, which is sensed by the interrupt switch to check operation of the pumping unit (or motor) and to calculate infused amount. For detection of the pumping unit backlash and quick infused amount error compensation, a dual pattern checker is used. The number of rotations of the pumping unit is identified by comparison with the number of pulses input to the motor. Low or high rotation of the pumping unit is also detected.

4.7. Sensors 1) Air sensor The air sensor transmitter is composed of the oscillator that generates oscillation signals of ceramic and piezo ceramic that changes the electric signals into mechanical vibration. The receiver has piezo ceramic that changes mechanical vibration into electric signals. The signals are amplified, rectified by the diode, and provided to the A/D converter. If air bubbles pass in the set between the two piezo ceramics, transmission of ultrasonic waves will be interrupted and voltage will drop. This is sensed and alarm sounds. This feature also senses an empty infusion set.

2) Occlusion sensor The occlusion sensor detects clogging or expansion inside the infusion set by means of the hall sensor. The occlusion device moves up and down, and the sensor's output voltage varies depending on the device movement. When output voltage fluctuates beyond a certain value, warning sound is generated. Also, the occlusion device can be used to check whether or not the infusion set is mounted.

3) Battery low sensor The input power is monitored by the monitoring IC, and the current power level is measured and sent to MCU. If voltage drops below a certain level, warning sound is

generated. If voltage declines to a level that the device cannot operate, the display is turned off and the motor stops.

4) Door open sensor When the door is opened, the distance between the magnet attached to the door and the hall sensor is widened and voltage drop is sensed, generating a warning sound.

5) Drop sensor (optional) Liquid drops falling from the chamber of the infusion set are sensed by the photoelectric sensor. When free drops (solution drops that fall by the force of gravity, not controlled by the device) are generated, the pump generates a warning sound. For error compensation, the infusion set needs to be set clearly before the drop sensor is used given that the amount of one drop varies depending on infusion sets.

5. Electrical Wiring Inspection

5.1. SMPS 1) Check the wiring between Connector CN81 and the power inlet. 2) Check the wiring between Connector CN91 and Connector J5 of the main board.

5.2. Main board 1) Check the wiring between Connector J1 and Connector J1 of the checker. 2) Check the wiring between Connector J4 and EXT Power Switch. 3) Check the wiring between Connector J5 and Connector CN91 of SMPS. 4) Check the wiring between Connector J7 and FFC (Flexible Flat Cable) of Connector J4 of the display board. 5) Check the wiring between Connector J9 and the step motor. 6) Check the wiring between Connector J10 and FFC (Flexible Flat Cable) of Connector J9 of the display board. 7) Check the wiring between Connector J11 and Connector J1 of the battery gauge.

5.2. Display board 1) Check the wiring between Connector J2 and Connector J1 of the alarm board.

2) Check the wiring between Connector J4 and FFC (Flexible Flat Cable) of Connector J7 of the main board. 3) Check the wiring between Connector J6 and Connector J1 of the door. 4) Check the wiring between Connector J7 and Connector J1 of the air sensor. 5) Check the wiring between Connector J8 and speaker. 6) Check the wiring between Connector J9 and FFC (Flexible Flat Cable) of Connector J10 of the main board. 7) Check the wiring between Connector J11 and Connector J1 of the OCCL sensor.

5.3. Air 1) Check the wiring between Connector J1 and Connector J7 of the display board.

5.4. Alarms 1) Check the wiring between Connector J1 and Connector J2 of the display board.

5.5. Battery Gauge 1) Check the wiring between Connector J1 and Connector J11 of the main board. 2) Check the wiring between Connector J2 and battery pack.

5.6. Checker 1) Check the wiring between Connector J1 and Connector J1 of the main board.

5.7. Door 1) Check the wiring between Connector J1 and Connector J6 of the display board.

5.7. Occl 1) Check the wiring between Connector J1 and Connector J11 of the display board.

6. Power Supply Inspection 6.1. Main Board

< Bottom of the main board> 1) TP8: Check the voltage of 24V. 2) TP31: Check the voltage of 5V 3) TP30: Check the voltage of 3.3V (VCC1). 4) TP35: Check the voltage of 3.3V (VCC2) (with front switch (SW14) on).

6.2. Display Board

1) TP1: Check the voltage of 5V. 2) TP2: Check the voltage of 3.3V (VCC1). 3) TP3: Check the voltage of 3.3V (VCC2) (with front switch (SW14) on).

6.3. Air Sensor

1) TP1: Check the voltage of 5V. 2) TX1: Check 2.2 Mhz frequency. 3) TP2: Check the voltage of 2V.

6.4. Alarm

1) Check J1's 1Pin VCC.

6.5. Battery Gauge

1) TP2: Check the voltage of 3.3V. 2) TP3: Check the current battery voltage. 3) TP5: Check the voltage between 3.0V and 6.5V. 4) TP7: Check the voltage of HDQ 3.3V.

6.6. Checker

< Bottom of the checker board> 1) TP1: Check the voltage of 3.3V. 2) TP2: Check Checker1 output voltage of 0~3.3. 3) TP3: Check Checker2 output voltage of 0~3.3.

6.7. Door Sensor

1) TP1: Check the voltage of 3.3V . 2) TP2: Check output voltage (Close:0V, Open: 3.3V).

6.8. Occlusion Sensor

1) TP1: Check the voltage of 3.3V. 2) TP2: Check output voltage (0.8V~2.0V). 3) TP3: Check temperature voltage (1.0V~2.0V (changes depending on temperature; voltage increases as temperature rises)).

7. Troubleshooting 7.1. Troubleshooting Table 7.1.1. Power supply and battery problems Table 1. Power & Battery Block No

Symptoms

Causes

Corrective Actions 1. Use phase power of

AC/DC indicator is not 1

illuminated when phase power is connected.

1. Phase power and power

100VAC-240VAC.

line have problems.

2. Check power cord and

2. Fuse and SMPS have

fuse.

problems.

3. Check problems with the

3. AC/DC indicator has

AC/DC indicator and LED1

problems.

and a wrong component of

4. Power circuit has

AL9.

problems.

4. (Contact the Technical Support Team.)

AC/DC indicator is not 2

turned off when phase power is disconnected.

1. Battery power is supplied to the AC/DC indicator.

1. Check wrong components of R7 (M) and R8 (M) or replace them if necessary.

1. Power is not supplied to the circuit. The battery level indicator is not 3

illuminated when the power switch is turned on with phase power.

2. Power ADC value is not detected. 3. The battery level indicator is out of order. 4. The display circuit has problems. 5. MCU is not working.

1. Check J4 (M) connection. 2. Check wrong components of D6 (M) and D9 (M). 3. Replace the display board. 4. Replace the display board. 5. Replace the main board. 6. Replace the main board.

6. MCU circuit has problems.

4

Part of the battery level

1.1. Check battery

indicators are not on, or

connection and charging

the battery warning

1. Battery ADC value

state.

light is illuminated

detection is not clear.

1.2. Check wrong

when the power switch

components of R16 (M) and

is turned on with phase

R17 (M) and replace them if

Table 1. Power & Battery Block No

Symptoms

Causes

power.

necessary.

The battery level

5

indicators are on but

1. The On/Off key has

the pump is not turned

problems.

on by the on/off key

2. Key circuits have

when the power switch

problems.

is turned on. 1. Infusion is going on. 6

The pump is not turned

2. Warning or alarm sound is

off by the on/off key.

generated. 3. The key has problems.

The pump is not turned 7

off when the power switch is turned off.

Corrective Actions

1. Check problems with SW14 (D) and wrong components of AL1 and AL2. 2. Replace the display board. 1&2. Press the stop key during infusion or when warning or alarm sound is generated. 3. Replace the display board. 1.1. Check the power switch

1. The power switch has

and involved cables.

problems.

1.2. Check J4 (M) and short circuit of adjoining circuits.

The battery level

8

indicators are not

1. The battery was not

illuminated when the

charged.

1. Check battery connection.

power switch is turned

2. The battery circuit has

2. Replace the battery board.

on in case the device is

problems.

powered by battery. The battery full

9

recharging indicator is

1. Recharge was not

1. Must recharge for more

not on after the battery

sufficient.

than 12 hours.

has been fully

2. The battery is dead.

2. Replace the battery.

recharged. Battery life is short 10

even after the battery

1. The same reason as

recharged. 11

The battery will not be

1. The battery has problems.

1. Replace the battery. 1.2 (Contact the Technical Support Team.) 1. Replace the battery.

Table 1. Power & Battery Block No

Symptoms recharged.

Causes

Corrective Actions

2. The battery recharging

2.1. Check DQ Line (High).

circuit has problems.

2.2. Check MCU 10-pin, 29pin (High). 2.3. Replace the battery board.

1. The DC power jack has Power is not on by 12

external DC power jack.

problems. 2. The connector has problems. 3. The power supply circuit has problem.

1. Check DC power jack and replace if necessary. 2. Check DJ1 state. 3. Check wrong components of D3.

7.1.2. Display problems Table 2. Display Block (Except AC/DC LED) No. 1

Symptoms

Causes

Corrective Actions

The entire display is

1. Power is not supplied to the

1. Check power supply to

blank.

circuit.

the main board.

2

1. Check J7 (M), J10 (M), J4 Only FNK indicators

1. The connector has

(D), and J9 (D) and involved

are not illuminated or

connection problems.

cables.

working abnormally.

2. FND has problems.

2. Check FND and possibilities of short circuit.

3

Part of the figures or characters of the flow rate or volume displays

1. The displays have problems.

1. Replace the display board.

are not accurate. 4

1. The flow rate display has Only the flow rate

problems.

display is not on.

2. The display processor has problems.

5

All digits of the flow

1. The display processor has

1. Check the display (FND1). 2. Replace the display board. 1. Replace the display board.

Table 2. Display Block (Except AC/DC LED) No.

Symptoms rate display show the

Causes

Corrective Actions

problems.

same number. 6

Part of the digits of the flow rate display are not displayed.

7

All displays but flow

2. The display processor has problems.

1. Check the display (FND1). 2. Replace the display board.

1. Check LED and replace if

problems.

necessary.

Brightness of displays

1. FND or LED has problems.

1. Check FND1, 2, and 3 (D)

or indicators is not

2. The display circuit has

or LED (D).

regular.

problems.

2. Replace the display board.

The volume display

1. To show that infusion

shows "----".

volume is infinite.

work normally.

9

problems.

1. The display (LED) has

rate and volume do not

8

1. The flow rate display has

1. Set a desired infusion volume by the Up or Down key.

7.1.3. Key problems Table 3. Keypad Block No

Symptoms

Causes 1. The device is in the pause mode.

1

All keys do not work.

2. The key circuit has problems. 3. The MCU circuit has problems.

2

Some keys do not work.

Corrective Actions 1. Check whether Lock Key was set. 2. Replace the display board. 3. Check MCU and replace if necessary.

1. Infusion is going on.

1. Up and Down key will not

2. Warnings or alarms have

work during infusion.

been generated.

2. Take proper measures for

3. Some keys have problems.

the warnings and alarms.

4. The key circuit has

Only the STOP key can be

Table 3. Keypad Block No

Symptoms

Causes problems.

Corrective Actions used. 3. Check keys with problems and replace them if necessary. 4. Replace the display board.

Keys are working, but 3

speaker sound is not generated.

1.1. Check whether the sound 1. The speaker or the speaker

volume is too low.

circuit has problems.

1.2. Check speaker connection.

1. Warnings or alarms have been generated. 2. Flow rate or total volume

4

Infusion does not start by the Start key.

was set to "0." 3. Infusion has been completed. 4. The key has problems. 5. The key circuit has problems.

1. Take proper measures for the warnings and alarms, and start infusion. 2. Set the flow rate and the total volume properly, and start infusion. 3. Clear the infused volume and start infusion. 4. Check the Start key. 5. Replace the display board.

7.1.4. Occlusion sensor problems Table 4. Occlusion Sensor No.

1

2

Symptoms

Causes

Corrective Actions

1. A general infusion set is

1. Use a recommended

Occlusion warning sound

being used.

infusion set.

or indicator is not on and

2. The infusion set is not

2. Re-mount the infusion set.

infusion continues even

mounted properly.

3. Check the occlusion

when the tube is clogged.

3. The occlusion circuit has

circuit and replace if

problems.

necessary.

Occlusion warning sound

1. The infusion set is not

1. Re-mount the infusion set.

or indicator is on and

mounted properly.

2. Replace the infusion set.

Table 4. Occlusion Sensor No.

Symptoms

Causes

Corrective Actions

infusion stops even when

2. The tube around the

3. Check the occlusion

the tube is not clogged.

occlusion device was

circuit and replace if

severely deformed because

necessary.

of the use of the infusion set for long periods of time. 3. The occlusion circuit has problems. 1. Occlusion sensitivity is 1 or 7. 2. The recommended 3

Tube occlusion detection

infusion set was not used.

is too fast or slow.

3. Infusion set was not mounted properly. 4. The occlusion circuit has problems.

Occlusion warning light 4

and sound are not off by pressing the stop key. Occlusion warning light

5

is not off and the warning sound is off by pressing the stop key.

1. Check the occlusion sensitivity and re-set if necessary. 2. Use the recommended infusion set. 3. Re-mount the infusion set. 4. Check the occlusion circuit and replace if necessary.

1. The stop key has problems. 2. The key circuit has problems. 1. Alarm and warning indicators are on, and infusion will not resume.

1. Check the stop key and replace if necessary. 2. Replace the display board.

1. Open and close the door and press the start key.

7.1.5. Air sensor problems Table 5. Air Detector Block No

1

Symptoms

Causes

Corrective Actions

Air warning light is off

1. The air warning light has

1. Check whether Air Level

when the tube is not

problems.

is off.

mounted.

2. The air sensor circuit has

2. Check the air sensor board

Table 5. Air Detector Block No

Symptoms

Causes problems.

The air warning light is 2

on when the tube is mounted. Air warning is not generated and infusion

3

continues when air bubbles pass around the air sensor.

Corrective Actions and replace if necessary.

1. The infusion set is not mounted properly. 2. The air sensor circuit has problems. 1. The air sensor is insensitive. 2. The air sensor circuit has problems.

1. Re-mount the infusion set. 2. Check the air sensor board and replace if necessary.

1. Set Air Level to Lo. 2. Check the air sensor board and replace if necessary.

1. The infusion set is not mounted properly. Air warning light and 4

sound are generated and infusion stops even when air bubbles do not exist.

2. Small air bubbles are gathered around the air sensor. 3. The air sensor is too sensitive. 4. The air sensor circuit has

1&2. Remove air bubbles and re-mount the infusion set. 3. Set Air Level to Hi. 4. Check the air sensor board and replace if necessary.

problems. The air sensor seems to 5

be too sensitive or insensitive.

1. It is necessary to adjust

1. Check the air sensor board

the air sensor circuit.

and replace if necessary.

7.1.6. Door sensor problems Table 6. Door Block No

Symptoms

Causes 1. The door warning light

1

The door open warning

has problems.

light is off when the door

2. The door circuit has

is open.

problems. 3. The display circuit has

Corrective Actions 1. Check LED7. 2. Check the door board and replace if necessary. 3. Replace the display board.

Table 6. Door Block No

Symptoms

Causes

Corrective Actions

problems. The door open warning 2

light is on when the door is closed.

1. The door circuit has problems. 2. The display circuit has problems.

The door open warning is not generated and 3

infusion continues when the door is opened during

1. Check the door board and replace if necessary. 2. Replace the display board.

1. Check the presence of the 1. The door circuit has

door handle magnet, or

problems.

replace the door board if necessary.

infusion.

7.1.7. Drop sensor problems Table 7. Free-Drop Detector Block No

Symptoms

Causes

Corrective Actions

1. The drop sensor is exposed to direct sunlight.

1

Even if free drops do not

2. The infusion set chamber

occur during infusion,

is tilted.

warning message and

3. The drop sensor plug has

sound are generated.

problems. 4. The drop sensor has problems.

1. Do not expose the device to direct sunlight. 2. Re-install the drop sensor. 3. Check plug disconnection or short-circuit. 4. Replace the drop sensor if necessary.

When free drops occur 2

during infusion, drop

1. The same reason as

1. Check the drop sensor

warning is not generated



installation.

and infusion continues.

7.1.8. Speaker sound problems Table 8. Speaker Block No

Symptoms

Causes

Corrective Actions

Table 8. Speaker Block No

Symptoms Speaker sound is not generated (when a

1

warning sound is generated or keys are selected.)

Causes 1. The speaker has

1. Check the speaker and J8

problems.

connection on the display

2. The speaker circuit has

board.

problems.

2. Check TP36 and Q5.

1. Alarm sound volume 2

Speaker volume is too

setting is too high or low.

high or low.

2. The speaker circuit has problems.

3

Speaker sound is broken. Speaker sound is

4

generated during infusion without any problems.

5

1. The speaker has problems. 1. The remaining battery level is low.

Speaker sound is

1. No keys were selected

generated in the standby

for two minutes in the

mode.

standby mode. 1 The stop key has

6

Corrective Actions

Speaker sound is not off

problems.

by pressing the stop key.

2. The key circuit has problems.

1. Adjust volume to a proper level. 2. Check TP36, Q5, and D20.

1. Replace the speaker.

1. Recharge the battery.

1. Press the stop key.

1. Check SW11 and AL8 (3, 6) connection on the display board. 2. Replace the display board if necessary.

7.1.9. Others Table 9. Others No

1

Symptoms

Infusion volume is too high or low.

Causes

Corrective Actions

1. The infusion set was too

1. Re-mount the infusion set.

loosened or tightened.

2. Adjust the calibration

2. Calibration setting is not

setting in the setting mode 2.

correct.

3. Check the pumping unit

3. The pumping unit

assembly.

Table 9. Others No

Symptoms

Causes assembly is not proper. 1. In case of purge infusion, air occlusion warning will

2

Warning and alarm are

not be generated. Also,

not generated during

infusion will continue even

purge infusion.

after infusion is completed once the purge key is pressed.

Corrective Actions

8. Disassembly and Adjustment

Only some part of Infusion Pump DI-2000 can be disassembled and replaced. The warranty shall not apply to disassembly or repairs arbitrarily carried out. Reassembly can be conducted in the reverse order of disassembly. Disconnect the external power supply before disassembly, and leave the power switch off. Since the electronic elements of the infusion pump are sensitive to electrostatic discharge, disassembly should be conducted on the antistatic workbench.

8.1. Separation of Case-Front from Case-Rear (See 9.2.) 1) Remove the bolts from the rear case with power off. Ensure that the front case should not be damaged by other elements. 2) Lift the rear case to separate the front part from the rear part. 3) When disassembling, make sure that the battery pack will not be separated or damaged and that the O-ring will not be lost. 4) When reassembling, ensure that the connector of each cable should be placed correctly. 5) When reassembly is completed, carry out electric safety test and make sure that the device is working normally.

8.2. Separation of Case-Front from Base Frame (See 9.4.) 1) Open the door and remove the bolt on the front. 2) Separate door latch, pumping unit, and door. 3) Remove the base frame bolt and separate the flat cable from the main PCB before complete separation from the front. 4) Make sure that PCB components will not be damaged when the battery pack is separated. 5) Be careful of the location of the plate spring when reassembling.

8.3. Replacement of Main PCB Ass'y (See 9.5.) 1) Separate the case-front from the base frame completely.

2) Remove the bolt fixing the main PCB and separate PCB by removing the cable connected. 3) Be careful about organization of cables when reassembling.

8.4. Separation of Pumping Unit (See 9.6.) 1) Lift the hook of the finger cover to separate the finger case. 2) Make sure that the spring is mounted properly when reassembling. 3) Coat grease on the working surface when reassembling.

8.5. Separation of Door (See 9.7.) 1) Remove the door pin to separate the auxiliary door. 2) Separate the auxiliary door and the presser block while taking precautions against dislocation of the spring. 3) After removing the bolt, remove the door handle pin and then disassemble the door cover completely. 4) Make sure that the door upper seal is placed completely when reassembling. 5) Make sure that the spring will not be displaced when reassembling.

8.6. Replacement of Air Sensor Ass'y (See 9.8.) 1) Separate the case front and the base frame completely. 2) Separate the air sensor connector from the display PCB assembly and remove the bolt from the air sensor assembly. 3) Push the air sensor assembly to separate it from the case front. Be careful that the assembly is attached to the case front with silicon. 4) Put silicon on the adjoining part. 5) Once the air sensor assembly and the case front are assembled, leave them for one to two hours so that silicon is completely hardened.

8.7. Replacement of Occlusion Sensor (See 9.8.) 1) Separate the case front from the base frame completely. 2) Separate the connector from the display PCB assembly.

3) Remove the occlusion sensor assembly from the case front and separate the occlusion sensor PCB assembly.

8.8. Replacement of Door Sensor PCB Ass'y (See 9.8.) 1) Separate the case front from the base frame completely. 2) Remove the connector of the door sensor PCB assembly from the display PCB assembly. 3) Loosen the bolt and remove the door sensor PCB assembly. 4) Make sure that the bolt tap will not be damaged when reassembling.

8.9. Replacement of Motor (See 9.5.) 1) Separate the case front and base frame completely. 2) Separate the main PCB assembly. 3) Separate the pumping unit pulley and remove the timing belt. 4) Remove the fixing bolt to separate the motor and the dustproof JIG. 5) When reassembling, make sure that the timing belt will not be too loosened or tightened.

8.10. Replacement of Checker and Checker PCB Ass'y (See 9.5.) 1) Separate the case front and the base frame completely. 2) Separate the main PCB assembly. 3) Remove the checker PCB assembly connector from the main PCB assembly. 4) Remove the checker PCB assembly. 5) Loosen the checker bolt to separate the checker. 6) Make sure the checker and the checker PCB assembly will be placed correctly when reassembling.

8.11. Replacement of SMPS (See 9.5.) 1) Separate the case front and base frame completely. 2) Remove the cable from SMPS. 3) Loosen the bolt to separate SMPS.

4) Reassemble in the reverse order of disassembly. After replacing SMPS, check the ground wire connection.

8.12. Replacement of Power Inlet (See 9.5.) 1) Separate the case front and case rear completely. 2) Remove the cable from SMPS. 3) Loosen the bolt to separate the power inlet. 4) Remove the cable from the inlet. 5) Be careful about organization of cables when reassembling.

8.13. Replacement of Cam Ass'y (See 9.5.) 1) Separate the case front and base frame completely. 2) Remove the checker B/D and separate the pumping unit pulley. 3) Remove the bearing holder and separate the cam shaft. 4) Coat grease on the working surface when reassembling.

8.14. Replacement of Battery (See 9.2.) 1) Remove the battery cover from the rear panel. 2) Take out the battery pack from the battery box and separate the cable connector. Do not pull the cable extremely. 3) Replace the battery pack and close the battery cover. Be careful about the battery type and battery pole location.

9. Assembly and Exploded Views 9.1. Clamp Module

No

Title

Quantity (ea/set)

No

Title

Quantity (ea/set)

81

Rubber foot

4

99

Clamp Handle

1

92

Clamp Lever

1

100

E-Ring (∅4) SUS

5

1

101

Clamp

1

93

Clamp Lever Stopper Spring

Bolt (flat head M4X16L,

94

Clamp Lever Stopper

1

102

95

Clamp Latch

1

103

Clamp Body

1

96

Clamp Base Hinge

1

104

Clamp Lever Shaft

1

97

Clamp Base

1

105

Clamp Latch Shaft

1

98

Bolt (flat head M4X12L, SUS)

4

SUS)

4

9.2. Product Exploded View

·

No Bolt 15 83

Quantity

Title (M3X8L,

P.W:∅6, SUS) Button Overlay

Quantity

No

Title

6

86

Battery Cover

1

1

87

Rubber O-ring

1

(ea/set)

(ea/set)

SEMS,

84

Case-Front second Ass'y

1

88

ID name plate

1

85

Case-Rear Ass'y

1

91

Sponge

1

9.3. Case-Rear Exploded View

No 76 77

Title Handle Cover Tapping Screw (M3X14L, type 2, SUS)

Quantity (ea/set)

No

Title

Quantity (ea/set)

1

80

Rear panel name plate

4

3

81

Rubber foot

4

82

Case-Rear

1

78

Hole Cover-A

1

79

Hole Cover-B

1

9.4. Case-Front First Exploded View

No

Title

Quantity (ea/set)

No

Title

Quantity (ea/set)

Bolt (M3X8L, SEMS, 9

55 64

P.W:∅6, Ni-plated) Bolt (flat head M4X16L, SUS) Door Latch

3

69

Battery Pack

1

2

70

Door Ass'y

1

1

71

Bolt(M3X16L, SEMS,

2

P.W:∅6, SUS) 65

Sub Door Locker Pin

1

72

Pumping unit Ass'y

1

66

Locker Plate Spring

1

73

Door Hinge Spring

2

67

Case-Front First Ass'y

1

74

68

Base Frame Ass'y

1

75

Bolt(M3X20L, SEMS,

9.5. Base Frame Exploded View

P.W:∅6, SUS) Sub Door Locker

4 1

No

5

Title Tapping Screw (M3X6L, Type 2, Ni-plated)

Quantity (ea/set)

No

Title

Quantity (ea/set)

2

50

Pulley-A

1

4

51

Power Switch

1

6

52

Power Switch Cap

1

1

53

SMPS Insulated Sheet

1

2

54

SMPS

1

Tapping Screw 7

(M3X10L, Type 2, Niplated) Bolt (M3X8L, SEMS,

9 38 39

P.W:∅6, Ni-plated) Motor Tapping Screw (M3X8L, Type 2, Ni-plated)

Bolt (Flat Head

40

Main PCB Ass'y

1

55

41

Dustproof JIG

1

56

Power Inlet

1

42

Tooth Washer (Bore 3.2)

3

57

Pumping unit CAM

11

2

58

Bearing Holder

2

M3X8L, SUS)

2

Bolt(M3X10L, SEMS, 43

P.W:∅6, Ni-plated)

Pumping unit CAM

44

Set Screw (M3X6L)

4

59

45

Pulley-B

1

60

46

Checker PCB Ass'y

1

61

Teflon Washer

2

1

62

Bearing

2

Shaft Pumping unit CAMSubt

1

3

Bolt(M3X5L, SEMS, 47

P.W:∅6, Ni-plated)

48

Checker

1

63

Base Frame

1

49

Timing Belt

1

90

Battery PCB Ass'y

1

9.6. Pumping unit Exploded View

No

34

35

Quantity

Title Pumping

unit

(ea/set) Finger

Cover Pumping unit Finger

No

1

36

14

37

Quantity

Title Pumping

(ea/set)

unit

Finger

unit

Finger

Case Pumping Spring

2

14

9.7. Door Exploded View

No

Title

Quantity (ea/set)

No

Title

Quantity (ea/set)

Bolt (M3X8L, SEMS, 15

P.W:∅6, SUS)

4

26

Door Hinge

1

18

Magnet (∅4X5L)

1

27

Presser Block

1

19

Door Cover

1

28

Sub Door Stopper

2

No

Title

Quantity (ea/set)

No

Title

Quantity (ea/set)

Spring 20

Door Handle Pin

1

29

Presser Block Spring

3

21

Door Base

1

30

Sub Door

1

22

Door Upper Seal

1

31

Door Pin

1

23

Sub Door Stopper

1

32

Occl. Block

1

1

33

Door Handle

1

24

Plate Washer (Bore 3.2 Outside Diameter 7) Tapping Screw

25

(M2.6X4L, Type 2, Niplated)

1

9.8. Case-Front Second Exploded View

No

Title

Quantity (ea/set)

No

Title

Quantity (ea/set)

1

Case-Front

1

11

Occl. Case

1

2

Alarm Lamp Lens

1

12

Occl. PCB Ass'y

1

3

Alarm Lamp PCB Ass'y

1

13

Occl. Finger

1

2

14

Occl. Cover

1

4

Plate Washer (Bore 3.2 Outside Diameter 8)

No

5 6

Title Tapping Screw (M3X6L, Type 2, Ni-plated) Air Sensor Ass'y

Quantity (ea/set)

No

Title

Quantity (ea/set)

Bolt(M3X8L, SEMS, 5

15

1

16

Magnet (∅4X5L)

2

2

17

Occl. Finger Spring

1

1

89

Speaker

1

6

91

Sponge

P.W:∅6, SUS)

2

Tapping Screw 7

(M3X10L, Type 2, Niplated)

8

Display PCB Ass'y Bolt(M3X8L, SEMS,

9 10

P.W:∅6, Ni-plated) Door Sensor PCB Ass'y

1

10. PCB Layout 10.1 Main PCB



10.2 Display PCB



10.3 Air PCB









10.4 Alarm PCB

10.5 Battery PCB

10.6 Checker PCB









10.7 Door PCB

10.8 Occl PCB

11. Schematic

Contact Daiwha Corporation Tel: +82-2-558-1711 Fax: +82-2-558-1745 e-mail: [email protected] Daiwha B/D, 733-18 Yeoksam-dong, Gangnam-gu, Seoul, Korea

DI-2000, DI-2200 Firmware Upgrade Guide

Daiwha,. co. ltd

TABLE OF CONTENTS 1. Setup Procedure 1.1 Setup Procedure 1.2 PC Setup 1.3 USBto Serial Adapter Drivers Installation 1.3.1 Installing CDM Drivers 2. Firmware Upgrade 2.1 Firmware Upgrade

1. Setup Procedure 1.1 Setup Procedure For the firmware upgrade, you can refer to the below Figure 1.1 and follow the update.

Figure 1.1 The following are the steps to set-up up the system fully. - Connect th power supply, DI-2000 2000 or DI2200 - PC( Windowxp) - USB port (A-A type) - USB to Serial Adapter Driver - Firmware Downloader( Megaload

1.2 PC Setup The following steps need to be performed to ssetup etup your PC for the Firmware Upgrade task (assumes an XP environment) 1. Create a directory/folder on your PC called “C: “C:\Daiwha Downloads”; 2. Create a subdirectory in this directory called “DI2000 or DI2200 hex files”; 3. Create a second subdirectory in this directory called “MegaLoad”; 4. (While you are at it, create another subdirectory called “Daiwha DI2000 or DI2200 Docs” for manuals, etc); 5. Go to: http://www.microsyl.com/megaload/megaload.zip 6. Click on OK to the request “Would you like to open or save this file?” To the request “Do you want to open or save this file?” click on “Save”. Then navigate to “C:\Daiwha Daiwha Downloads/MegaLoad” and save the file there. 7. Click on “Open”. When the “megaload.zip” window appears, double double-click on “Setup.msi” (If a “Publisher could not be verified ….” warning appears, click on “Run”;) 8. At this point, the installer seeks to use the “.NET Framework” run run-time time utilities, which are often on systems from previous use.

If they ARE found, the installer proceeds automatically to the “MegaLoad Setup Wizard” described in the next main paragraph (Para 9 below).

a. If

NOT, you will be prompted with a small window with text starting: “This setup requires the .NET

framework …” Click on “Yes” to go to Microsoft’s “netframework” site. (Ignore the “Silverlight” box by X’ing it). b. Click on “Download x86 version” in the “.NET Framework 3.5 Redistributable Package” area (lower left of screen). c. On the x86 screen which appears, click on the “Download” button mid-screen, then click on “Save” in the “File Download” window. d. In the “Save As” window which appears, move to the “Megaload” window and save the “dotnetfx.exe” file there. (If you just downloaded the MegaLoad file, you are probably already there). Click on “Save”. e. Go get a coffee … this is a 22 Mbyte file! f. Click on “Run” in the “Download Complete” window, and then “Run” again in the “Security Warning” window, (if one appears). g. In the “Microsoft .NET Framework 3.5 Setup” window, click “Next”, tick the “I accept the terms …” and then click on “Install”. When it shows “Setup Complete”, click on “Finish”. h. At this point, close the Internet browser, and you will be back in the “megaload.zip” window. i. Go back to step 7, above. 9. In the “Welcome to the MegaLoad setup wizard” window, click on “Next”; 10. Use the default location, usually “C:\Program Files\MicroSyl\MegaLoad\”. Click on Next, twice. 11. Close the window when the installation is complete.

1.3 USBto Serial Adapter Drivers Installation 1.3.1 Installing CDM Drivers

To install CDM drivers for an FTDI device under Windows XP, follow the instructions below: - If a device of the same type has been installed on your machine before and the drivers that are about to be installed are different from those installed already, the original drivers need to be uninstalled. Please refer to the Uninstalling CDM Drivers section of this document for further details of this procedure. - Download the latest available CDM drivers from the FTDI web site and unzip them to a location on your PC. Goto: http://www.ftdichip.com/Drivers/VCP.htm - If you are running Windows XP or Windows XP SP 1, temporarily disconnect your PC from the

Internet. This can be done by either removing the network cable from your PC or by disabling your network card by going to the "Control Panel\Network and Dial-Up Connections", right-clicking on the appropriate connection and selecting "Disable" from the menu. The connection can be re-enabled after the installation is complete. This is not necessary under Windows XP SP 2 if configured to ask before connecting to Windows Update. Windows XP SP 2 can have the settings for Windows Update changed through "Control Panel\System" then select the "Hardware" tab and click "Windows Update". - Connect the device to a spare USB port on your PC. If the device is based on the FT2232, the Microsoft composite device driver is automatically loaded in the background. Once the composite driver has been installed Windows Found New Hardware Wizard will launch. If there is no available Internet connection or Windows XP SP 2 is configured to ask before connecting to Windows Update, the screen shown in Figure 1.2s displayed. Select "No, not this time" from the options available and then click "Next" to proceed with the installation. If there is an available Internet connection, Windows XP will silently connect to the Windows Update website and install any suitable driver it finds for the device in preference to the driver manually selected.

Figure 1.2 -Select "Install from a list or specific location (Advanced)" as shown in Figure 1.3 below and then click "Next".

Figure 1.3 -Select "Search for the best driver in these locations" and enter the file path in the combo-box ("C:\CDM 2.02.04" in Figure 1.4 below) or browse to it by clicking the browse button. Once the file path has been entered in the box, click next to proceed

Figure 1.4 -If Windows XP is configured to warn when unsigned (non-WHQL certified) drivers are about to be installed, the message dialogue shown in Figure 1.5 will be displayed unless installing a Microsoft WHQL certified driver. Click on "Continue Anyway" to continue with the installation. If Windows XP is configured to ignore file signature warnings, no message will appear.

Figure 1.5 -The screen shown in Figure 1.6 will be displayed as Windows XP copies the required driver files.

Figure 1.6 -Windows should then display a message indicating that the installation was successful (Figure 1.7). Click "Finish" to complete the installation for the first port of the device.

Figure 1.7

2. Firmware Upgrade 2.1 Firmware Upgrade 1. Email

Daiwha at [email protected] and ask for the latest “DI2000S_xxxx_Revxx.hex” file. (xxx is the

current version number). Save it in the “DI2000 or DI2200 hex files” data directory just created on your PC: i.e. “C:\Daiwha downloads\DI2000 hex files” 2. Click on the Megaload.NET icon on your desktop to run the downloader:

Figure 2.1 3. In the first window, “Files to be programmed in the Flash”, click on “Open”, move to the directory: “C:\Daiwha downloads\DI2000 or DI2200 hex files”, and double-click on the latest DI2000S_xxxx_Revxx.hex file there. (Note: There is no corresponding EEprom file to be opened); The “Messages” window should show “Flash File ok”; 4. Select the COM port number on your PC that you have connected to the RS232 cable, in the “CommPort” window (or the simulated COM port number that a USB -> serial converter has been installed as. This number can be determined by going to “Start”, right-clicking on “My Computer”, and left clicking on “Properties”.

Click on the “Hardware” tab, then click on the “Device Manager” button. Click on the “+” sign next to “Ports(COM & LPT). Under this heading you should see the name of the USB-> RS232 device, e.g. “Belkin Serial on USB (COMxx)”, where xx is the port number you use in the window in MegaLoad);

Figure 2.2

Go to Start → Control Panel → System → Hardware → Device Manager → Ports

Click on the “+” sign to drop down the ports connected to your computer.

Figure 2.3 5. Set the baud rate to 38.4kbps in the “Speed” window. (Don’t click any “Command” or “Lock bits” buttons); 6. The status window should show “Ready, Waiting for Target”; 7. Plug in the DI2000 or DI2200 cable into the USB port. Press the “POWER ON/OFF” switch; 8. In the “Messages” window, the “Sending Page #nnn” scrolls as each 256 byte page is erased, sent and verified. The final message should be: “Flash Prog Done! Sending LockBits”.

Figure 2.4 9. Unplug, setup the switches and install. If doing a number, just plug the next one in and toggle its yellow switch.(If it fails to program or stops prematurely, push the small RESET switch to retry. It should show Target Mega128 at the end).